For the following steps, ensure the power is off and the power cord is disconnected.
1. Locking screw
Remove the Locking Screw either from the back as the instrument has already been worked with or from the
side when it becomes installed the first time.
Push the top cover from back to the front and lift
it up
Take care of the hooks that fix the top cover with the main cover
4. Demount the Main Cover
Remove the 4 screws at the side (2 screws each
side)
Put the instrument to the upright position and
remove the screws as shown below.
When the control cover is loosen, disconnect the plug from the heat plate
6. Release the Heat lid block
To release the heat lid block the screws on the
left and on the right side need to be removed
NOTE! The 2 screws marked in red have a shorter length than the others.
Be careful when reassemble the cover.
If the wrong screws will be used, the drawer can not be closed properly anymore.
Dismounting the cover might be bit tricky. Lifts
up the rear side with one hand while the other
hand pushes the cover from the front to the back.
The small hooks on the left and on the right side
have to be loosen from its mounting without
braking them
NOTE! Upon replacement of the CPU board X origin calibration has to be performed. Refer to section
1.2 X-axis calibration in the Service manual.
Turn the board down and disconnect the flat cable from the optical control board (PMT module)
12. Demount the PMT module
Remove the 4 screws