Design and Analysis of a Novel Six-Component F/T Sensor based


on CPM for Passive Compliant Assembly
Qiaokang Liang1, Dan Zhang2, Yaonan Wang1, Yunjian Ge3
1
College of Electric and Information Technology, Hunan University, Changsha, Hunan 410082, China,
qiaokang@mail.ustc.edu.cn
2
Faculty of Engineering and Applied Science, University of Ontario Institute of Technology, Oshawa, Ontario,
Canada L1H 7K4, dan.zhang@uoit.ca
3
State Key Laboratories of Transducer Technology, Institute of Intelligent Machines, Chinese Academy of Science, Hefei,
Anhui 230031, China, yjge@iim.ac.cn

This paper presents the design and analysis of a six-component Force/Torque (F/T) sensor whose design is based on the
mechanism of the Compliant Parallel Mechanism (CPM). The force sensor is used to measure forces along the x-, y-, and z-axis
(Fx, Fy and Fz) and moments about the x-, y-, and z-axis (Mx, My and Mz) simultaneously and to provide passive compliance during
parts handling and assembly. Particularly, the structural design, the details of the measuring principle and the kinematics are
presented. Afterwards, based on the Design of Experiments (DOE) approach provided by the software ANSYS®, a Finite Element
Analysis (FEA) is performed. This analysis is performed with the objective of achieving both high sensitivity and isotropy of the
sensor. The results of FEA show that the proposed sensor possesses high performance and robustness.

Keywords: Compliant Parallel Mechanism, Passive Compliant Assembly, Six-component F/T sensor.

1. INTRODUCTION 1. Relative maturity of the theory system: all the related


theory, such as position analysis, statics, stiffness analysis,
S INCE the Oak Ridge and Argonne National Laboratories
applied force feedback on the load the slave manipulator
encountered to their master-slave manipulator during the
dynamics, have been investigated for a long time and gotten
well known;
late 1940s, F/T sensing with appropriate control techniques 2. Possibility to provide decoupling F/T information: unlike
has played an important role in the dexterous and reliable traditional force-sensing elements that sense all F/T
robotic manipulation. After F/T control has been recognized components with a single monolithic structure, force-
as a key scheme for using robots in advanced applications sensing elements based on parallel mechanisms employ their
[1], various types of F/T sensors have been designed and limbs to sense the components and show a limited cross talk;
developed [2-6]. Several six-axis force sensors are 3. Isotropy: based on the statics analysis of parallel
commercially available at present, such as force sensors of mechanisms, a global stiffness matrix denoting the
ATI [7] and JR3. relationship between the undergoing load and the
A major problem in developing force sensors is the design infinitesimal movement of the mobile platform can be
of the force-sensing element, which has been done
obtained, which could be used to pursue measurement
heuristically, dependent on the experience of designers [1].
To detect three orthogonal forces and three orthogonal isotropy among components;
torques simultaneously, the force-sensing elements of multi- 4. High stiffness and high sensitivity: as the stiffness of the
component force sensors are always featured by complicated robotic manipulator depends on the component with
geometrical structures, which make accurate calculation of minimum stiffness, which is always the F/T sensor, the
strains and deflection difficult or even impossible with stiffness of a F/T sensor is a critical performance factor.
current techniques. Another critical drawback of the However, it is impossible for a traditional force-sensing
traditional multi-component F/T sensor is the existence of element of high stiffness to detect F/T with high sensitivity.
significant measurement couplings among components, On the contrary, for the force-sensing element based on
especially between component Mx and component Fy, parallel mechanisms, the limb stiffness can be much lower
component My and component Fx. Additionally, as there are compared to the overall required stiffness of the F/T sensor
six components to be measured and processed due to their parallel arrangement. Therefore, the trade-off
simultaneously, approximately equal measurement between stiffness and sensitivity can be solved.
sensitivity for each component is expected [8], [9].
Recently, some researchers developed force/torque sensors
Furthermore, the stiffness and the sensitivity of the F/T
based on parallel mechanisms to meet the requirement and
sensor is always a trade-off, which means users have to
make a sacrifice of stiffness to get a high-sensitivity escape the mentioned shortcoming [6]. Gailet and Reboulet
performance and vice versa. developed a force sensor of SP (Stewart Platform) based on
Taking into account the above-mentioned drawbacks, octahedral structure [10]. Dwarakanath and Bhaumick
researchers found that parallel mechanisms are preferred to designed and implemented a force/torque sensor based on
be considered as a candidate for a force-sensing element of a the SP structure in 1999, and the analysis deals with
multi-component F/T sensor due to their following kinematic design, leg design and optimization of the form
advantages [6]. of the leg and the aspects of integration were presented [11].

Unauthenticated
253 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

Table 1. Comparisons on the multi-dimensional F/T sensor. based on a hyperstatic Stewart Platform. Parameter
optimization of the sensor structure was performed with
Size genetic algorithms, and the sensor shows good isotropy and
Sensitivity & Sensing
Year & Fabrication (mm) sensitivity performances [15]. Recently, Jia, Lin, and Liu
Measurement principl
developer technology & No.
of axes
range e introduced another type of Stewart Platform dedicated to
heavy six-axis load sharing measurement [16], [17]. With
1 mN and 2 mN
Bulk silicon
mm for force
single-component force piezoelectric sensor fixed in each
micro- 4.5 × limb, the sensor could measure six-component F/T with
1998 & Jin and moment Piezo-
machining and 4.5 ×
and Mote
wafer-bonding 1.2 & 6
components resistive linear and repeatability error of 1%. The comparisons of
(resolutions) & main features and characterization methods using a selection
technologies
n.a. of multi-dimensional F/T sensors up to the date of the
13mV/N in the publication are shown in Table 1.
CMOS Z-axis and Unfortunately, in force-sensing applications, the presence
process, about 2.3mv/N Piezo-
1999 & Mei
silicon bulk
4×4×2
in X- and Z-axis resistive of clearance, friction and backlash at the rigid mechanical
et al. &3 joints of parallel mechanisms will modify the axial forces
micro- & 0—50 N in
machining Z-axis, ±10N in that the limbs are subjected to, therefore, disturb the
X- and Y-axis performance of the sensor in unpredictable ways. To avoid
16 PF/N in Fz these problems, force-sensing candidate based on parallel
Bulk silicon 9×9×1
2012 & and 2.7PF/N mechanism is preferred to be designed with flexural joints,
Micro-
Brookhuis, machining and (PCB mm in Mx and Capaci- which eliminate the friction, backlash and wear and possess
chip) My & 50N in tive sub-micron accuracy with high resolution, continuous and
Lammerink silicon fusion
&3 Fz, ±25 N mm
bonding
in Mx and My
smooth displacement.
In this paper, modeling and performance evaluation of a
Four chip
novel six-component force/torque sensor based on
inductors, 3.2× 0.06 N
2009 & silicone gel, Inducti- compliant parallel mechanism is presented, which consists
2.5 × (resolution) & -
Takenawa, S. and a 2.2 & 3 40 N to 40 N
ve of three identical sensing limbs of type SPS (Spherical-
neodymium Prismatic-Spherical) and one central leg with three degrees
magnet of freedom. The compliant joints of the four limbs are
Three spoke 95%
structure, 40.25× (uncertainty) &
employed as active force-sensing portions. By the electric
2011 & M. Strain measurement technique of strain gauge, the developed
Gobbi et al.
standard 16 × 16 10 kN (Fz), 5
gauge sensor can provide high-performance force/torque
mechanical &6 kN (Fx, Fy) and
machining 0.5 kNm (M) information, and the calibration experiment shows that the
Parallel Fx=Fy=20kgf, presented sensor possesses positive characteristics such as
170×10
2004 & Liu, support
5×26.5
Fz=l00kgf, and Strain high stiffness and sensitivity, weak couplings, and good
Inoue et al. mechanism & Mx=My=Mz=l0 gage isotropy.
&6
machining 0 Nm
Three 2. ARCHITECTURE DESCRIPTION OF THE SENSOR
elastomers in a 0.3% & 100 kgf
flexed spoke and 300kgfcm The proposed force-sensing element is based on spatial
1990 & parallel compliant mechanism with three identical sensing
shape & Φ76×4 for parallel and
Hirose and Optical
Yoneda
milling and 0&6 rotating limbs of type SPS and one central limb with three DOFs
electric directions, connecting the mobile platform and the base. Each sensing
discharge respectively
machining limb consists of two three-axis flexural joints as spherical
joints at its both ends and a one-axis flexural joint as a
linear error and
the repeatability
prismatic joint in the middle. After neglecting the redundant
Stewart error of 1%, degrees of freedom about the lines passing through the two
small
platform & interference spherical joints, the total degree of freedom of each limb is
2010 & Jia, than Piezo-
standard error of 4.5% equal to six. Therefore, these limbs impose no constraint on
Lin, and Liu Φ160 × electric
mechanical & 40kN and the moving platform. The central limb consists of a large-
210
machining 5kNm for force
and moment displacement compliant universal joint and a large-
components displacement compliant revolute joint, and it has
connectivity of three and imposes three constraints on the
moving platform. Fig.1. shows a 3D model of the proposed
Ranganath, Mruthyunjaya and Ghosal analyzed and sensor.
designed a SP-based force/torque sensor in a near-singular
configuration with high sensitivity [12]. Nguyen, Antrazi
and Zhou presented the kinematic analysis of a 6 DOF
force/torque sensor based on the mechanism of the Stewart
Platform and composed of two platforms coupled together
by 6 spring-loaded pistons [13]. Dasgupta, Reddy and
Mruthyunjaya presented a design methodology for the
Stewart Platform sensor structure based on the optimal
conditioning of the force transformation matrix [14]. Hou, Fig.1. Six-component force/torque sensor based on compliant
Zeng, Yao et al. presented a six-component F/T sensor parallel mechanism.

Unauthenticated
254 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

Fig.2. shows the link-pair relationship diagram for the 3- Fz Et 3/2


DOF mechanism. The blue boxes represent passive joints ≈ (4)
while the red boxes represent sensing joints, which are used δz 2 R1/2
as the elastic element of the sensor. The number of degrees
of freedom, Fd, is given by the Chebyshev-Grubler- Where Fn and δn are the force and translation displacement
Kutzbach criterion: in n-axis, respectively; t and R are the inter-hole space and
common radius of notches, respectively; and Mn and θn are
j the moment and rotational displacement around n-axis,
Fd = λ ( n − j − 1) + ∑ f i − f r (1) respectively; the E is modulus of longitudinal elasticity of
i =1
= 6 × (9 − 11 − 1) + 24 − 3 = 3 the material.
The majority of the existing compliant translational joints
are based on a parallel four-bar building block [29]. Though
Where λ denotes the dimension of the space, n and j
they can deliver a pure translational motion with acceptable
denote the number of links and the number of joints,
off-axis stiffness, the range of motion is very limited and the
respectively, fi and fr represent the degree of freedom
construction is complicated. A novel compliant translational
associated with joint i and the number of redundant degrees
joint based on a simple cantilever is proposed and shown in
of freedom in the mechanism.
Fig.3(b).
Due to its geometry character (as shown in Fig.4.), the
applied force Fy is transmitted to the beam through a drive
bar with an additional bending moment Mz. The bending
equation for this particular leaf spring is given by

d 2δ y L L (5)
EI = Fy − Fy ( L − x ) = Fy ( x − )
dx 2 2 2

Integrating twice with boundary conditions and assuming


small displacements, then,
Fig.2. Schematic diagram of the compliant parallel mechanism.
1 L (6)
EIθ = Fy ( x 2 − x )
The notch hinges and the leaf type flexure are probably the 2 2
most popular element in applications of compliant
mechanisms. All compliant joints used in this design are 1 L
EI δ y = Fy ( x 3 − x 2 ) (7)
shown in Fig.3. 6 4

The slope and deflection at the free end of the beam can be
obtained

Fy 1 2 L
θ x=L
= ( x − x) = 0 (8)
EI 2 2

Fy 1 3 L 2 L3
(a) (b) (c) δy x= L
= ( x − x ) = − Fy (9)
EI 6 4 12 EI
Fig.3. Compliant joints used in the design: (a) compliant spherical
joint; (b) compliant prismatic joint; (c) compliant spherical joint for The compliant spherical joint for central limb used in this
central limb. study, which was designed by Trease [25], [26], is shown in
Fig.3(c).

Toroidal notch hinge, as shown in Fig.3(a)., is commonly


utilized in spatial compliant mechanisms to provide 3-axis
motions [25], [26]. The equation for the stiffness in the
freedom axes can be approximated from [27], [28].

My Mx Et 7 /2 (2)
= ≈
θy θx 20 R1/ 2

Fx Fy Et 7/ 2
= ≈ (3)
θy θx 20 R 3/ 2 Fig.4. The configuration of the cantilever.

Unauthenticated
255 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

3. KINEMATICS The angle αi, βi represent the angle between x-axis and the
The pseudo-rigid-body model provides a simple method of line OAi and the angle between u-axis and the line PPi,
analyzing compliant systems that undergo large, nonlinear respectively. So, the position vector Ai with respect to frame
deflections and is used to model the deflection of flexible O and Pi with respect to frame G can be written as
members using rigid-body components that have equivalent
force-deflection characteristics. Therefore, rigid-link Ai = [e ⋅ cαi e ⋅ sα i 0]T , for i=1, 2, 3 (10)
mechanism theory may be used to analyze the compliant
mechanism [31], [32]. Pi = [q ⋅ cβi q ⋅ sβi 0]T , for i=1, 2, 3 (11)

A. Inverse and forward kinematics


As the platform of the mechanism has three degrees of
Fig.5. shows a kinematically equivalent diagram of the freedom, only three of the six Cartesian coordinates of the
proposed micro-manipulator, in which squares and cylinders platform are independent, which have been chosen for
represent prismatic and revolute joints, respectively. The convenience as (ψ, θ, Φ).
lower ends of the actuated limbs, points A1, A2, A3, are The position of the moving platform is defined by the
connected to fixed base through universal joints, and the vector, G,
upper end of the actuated limbs, points P1, P2, P3, are
connected to the mobile platform through spherical joints. T
G = OG = Gx Gy Gz  (12)
Mobile
P3 platform w P2 And the orientation of the moving platform is defined by
S-joint P3 G the rotation matrix,
u v
P1 Sensing
O
RP = Rz (φ ) Ry (θ ) Rx (ψ ) (13)
limb
P-joint
R-joint Where ψ, θ, Φ (roll, pitch and yaw angles) represent three
Central successive rotations of the moving frame about the fixed x, y
limb and z-axis [33]. As the passive limb is a 3-DOF serial chain,
U-joint z its posture can be described by the three joint variables, θ1,
y A2 θ2, θ3. Referring to Fig.7., the coordinate frames are
A3 A3 O established and the corresponding Denavit-Hartenberg
Fixed x parameters are given in Table 2.
base
A1

Fig.5. Kinematically equivalent diagram of the sensor.

As shown in Fig.6., the fixed base reference frame O{x, y, z}


and the mobile coordinate frame G{u, v, w} are attached to the
fixed base and the mobile platform, respectively. The
original of the O{x, y, z} is located at the centroid O of the
triangle A1A2A3, which is the original placement without
displacement, and the x- and y-axis lying on the triangle
plane with y-axis parallel to A3A2 and x-axis pointing to A1.
The original of the G{u, v, w} is located at the centroid G of the
triangle P1P2P3 with the u and v-axis lying on the triangle
plane and u-axis pointing to P1 and v-axis parallel to P3P2.
Triangles P1P2P3 and A1A2A3 are set equilateral with the
lengths q and e, respectively.

P3 P2 A2
q
G β2 v y
u O α2
e x
P1 A3 Fig.7. Coordinate frames of the central limb.
A1
(a) (b)
Hence, the D-H transformation matrices from the moving
Fig.6. Top views of the mobile platform (a) and the base (b). platform to the fixed base can be obtained [33].

Unauthenticated
256 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

Table 2. The D-H parameters of the passive constraining limb. B. Jacobian matrix
Since the mechanism possesses only three rotational
i ai di αi θi
degrees of freedom, the input vector can be written as the
1 0 l1 0 θ1 T
2 0 0 -π/2 θ2 extension rate of the sensing limbs d& = d&1 , d&2 , d&3  , and the
 
3 0 0 -π/2 θ3 output vector can be presented as the angular velocity of the
G 0 l4 0 0
moving platform, X & = ω , ω , ω T By using the velocity
 x y z
vector-loop method, we can obtain the Jacobian matrix of
the mechanism A without the passive leg,
 O R OG 
O
TP = T1T2T3TP =  P  & = d&
 0 1  AX (20)
 cθ1cθ 2 cθ 3 cθ sθ cθ1cθ 2 sθ 3 l4 (-cθ 3 sθ1 
 + sθ1 sθ 3 1 2
- sθ1cθ 3 + cθ1cθ 2 sθ 3 )  (14)
Where
 
sθ cθ cθ sθ cθ sθ l4 ( cθ1cθ 3 
=  -cθ sθ sθ1 sθ 2 1 2 3
1 2 3
sθ1cθ 3 + cθ 2 sθ1 sθ 3 ) 
-  s1T ( P1 × s1 ) 
T
 1 3
 sθ 2 cθ3 -cθ 2 sθ 2 sθ 3 l1 + l4 sθ 2 sθ 3   T
  A = s 2T ( P2 × s2 )  (21)
   s3T ( P3 × s3 ) 
0 0 0 1 T

Comparing (13) to (14), one can yield And si is the unit vector pointing along the ith limb.
The Jacobian matrix of the passive constraining leg of the
θ2 = cos −1 (cθ sψ ) (15) mechanism B can be expressed as

sθ cψ −sθ  e e2 e3 
θ3 = Atan2( , ) (16) B= 1 
sθ 2 cθ 2 e1 × r1 e2 × r2 e3 × r3 
l4 (−cθ1cθ3 l cθ sθ sθ l4 (cθ1cθ 2 cθ3 
 −cθ 2 sθ1sθ3 ) 4 1 2 3 + sθ1sθ3 ) 
sφ sθ sψ + cφ cψ sφ cθ   (22)
θ1 = Atan2( , ) (17)
l4 (− sθ1cθ3 −l4 sθ 2 sθ1sθ3 l4 (cθ 2 cθ3 sθ1 
sθ 2 sθ 2
 −cθ1cθ 2 sθ3 ) −cθ1sθ3 ) 
= 0 l4 cθ 2 sθ3 l4 cθ3 sθ 2 
 
The inverse kinematics problem can be simply stated as:  0 sθ1 cθ1sθ 2 
 0 −cθ1 cθ1sθ 2 
given the independent parameters, ψ, θ, Φ, to find the  
corresponding lengths of the actuated limbs:  1 0 −cθ 2 

d i = [G − Ai + O RP P Pi ]T [G − Ai + O RP P Pi ] (18) Hence,
cθ 2 cθ (23)
θ&1 = cθ1 ω1 + sθ1 2 ω2 + ω3
sθ 2 sθ2
For the forward kinematics, the limb lengths l1, l2, and l3
are given and the problem is to find the orientation of the θ&2 = sθ1ω1 − cθ1ω2 (24)
mobile platform. (18) is written in the following form [33]:
cθ1 sθ
θ&3 = ω1 + 1 ω2 (25)
sθ 2 sθ 2
µi1cθ1cθ2cθ3 + µi 2 sθ1cθ2cθ3 + µi3sθ1sθ3
+ µi 4cθ1sθ3 + µi 5 sθ1sθ2 + µi 6cθ1sθ2 for i=1, 2, 3 (19)
J, a 3x3 Jacobian matrix, relating the independent velocity
+ µi 7 cθ3 sθ3 + µi8cθ2 + µi 9 = 0 & , to the vector of sensing
variables of mobile platform, X
&.
limb rates, d
Where µij are constant coefficients that are determined by
d& = A  s  X
J & & = JX
&
= AJ t X (26)
the manipulator geometry and input leg lengths [33].  I 3×3 

Unauthenticated
257 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

Where, Jt, a 6×3 Jacobian matrix, relating the velocity state 4) Voltage output to F/T output: besides common signal
T conditioning and signal processing, nonlinear decoupling
of the moving platform, vG = G& x G& y G& z  , to the
  and calibration are necessary due to the nonlinear coupling
vector of sensing limb extension rates. And Js, a 3×3 error, which seriously decreases the sensor measurement
Jacobian matrix, relating the three independent velocities &
X precision..
to the three dependent velocities vG , which could be
obtained through (22) to (25).

4. MEASURING PRINCIPLE
Several measurement techniques have been used to
transduce the interaction between a contacting load and a
F/T sensor into multi-dimensional force and torque. The
transduction method which has received most attention in
F/T sensor design is concerned with the conductive and
resistive approach. Strain gauges commonly act as sensing
elements on devices for measuring force in conductive and
Fig.8. Structure of measurement chain.
resistive approach. The measurement chain consists of
several elements from true load to measured F/T, as shown
in Fig.8. The succession of conversions for a single strain When resistive moments are exerted on the mobile
gauge is the following. platform, the compliant prismatic joints in each sensing limb
1) Load to elastic strain: mechanical stress occurs when are compressed or extended. Therefore, the moving platform
load F acts on the elastic element of the sensor, and specific possesses three degrees of freedom in micro-scale spherical
deformation and elastic strain will occur accordingly, motion. Hence the force sensor can provide passive
following the Hook’s law of elasticity. The quality of the compliance to the mobile platform during parts assembly.
By the virtual work principle, we can obtain
occurred stain ε only relates to the applied force/torque F
after the dimensions of the force-sensing element are
decided. wT ∆X = f T ∆d (31)

ε = χ (F ) (27) where w=[Mx, My, Mz]T is the vector of the moment applied
to the platform, and f=[f1,f2,f3]T represents the force that the
2) Elastic strain to variation of resistance: occurred strain sensing limb is subjected to. Where ∆X , ∆d represent the
will result in variation of relative resistance of strain gauges vector of virtual displacements associated with the mobile
bonded on the force-sensing element. The gauge factor, platform and sensing limbs, respectively. (31) can be
denoted by Gf, is defined mathematically as follows [34]. rewritten as

∆R R (28) wT X& = f T d& (32)


Gf =
ε
Substituting (26) into (32) yields
where R is the original resistance of the strain gauge.
Therefore, the resistance change ∆R of the resistance of the w = JT f (33)
strain gauge is:
Hence the moments applied on the moving platform can
∆R = G f ⋅ R ⋅ χ ( F ) (29) be calculated from the actuated limb forces, and vice versa.

3) Resistance change to output voltage variation: Full- A. Attachment of position of stain gauges
bridge circuits, as a ratiometric device, increase further the The central limb restricts the sensor’s freedom of
sensitivity of the circuit. And its measurement sensitivity is: displacements along x-, y-, and z-axis. Therefore, the force
Fx, Fy, and Fz applied on the mobile platform will be
∆U transmitted through the central limb. Finite Element
= −G f ⋅ ε = −G f ⋅ χ ( F ) (30)
U Analysis (FEA) via software ANSYS® was performed and
the distributions of the normal elastic strains occurring on
Where ∆U and U are output of the circuit and voltage the CPM under sign-dimensional force/torque are shown in
excitation source, respectively. the Fig.9.

Unauthenticated
258 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

Spots that have the maximum strain are selected to bond


strain gauges to detect the corresponding loads, as shown in
Fig.10.

(a)

(a)

(b)

(b)

Fig.10. Strain gauges arrangement: (a) arrangement on the sensing


limbs; (b) arrangement on the central limb.

(c) B. Connection mode of strain gauges


According to the arrangement scheme of strain gauges, the
Wheatstone bridges connection mode of the sensor is
determined, as shown in Fig.11. All of the gauges used in
the present study are Y series linear strain gauges (1-LY11-
3/120) made by HBM Inc.

(d)
Fig.11. Wheatstone bridges connection mode.

Fig.9. The strain produced in the CPM by the force Fx (a), When the sensor is applied to single force/torque Fx, Fy, Fz
the force Fz (b), the moment Mx (c) and Mz (d). respectively, corresponding output of each bridge is:

Unauthenticated
259 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

U  ∆R41 ∆R42 ∆R43 ∆R44  5. COMPLIANCE/STIFFNESS ANALYSIS


∆U Fz =  − + − 
4  R41 R42 R43 R44  The compliance of the CPM can be established via a
(34) proper investigation by taking into account the flexibilities
U   ∆R41   ∆R42  
=  2   − 2   of every compliant joint [27], [28], [29], [30]. In what
4   R41 ε  R42 ε  follows, the compliance of the developed CPM based on the
UK stiffness matrix method is presented.
= (ε 41 + ε 42 ) With the definition of local z-axis in Fig.3., the
2
infinitesimal translational and rotational displacements X of
mobile platform are formulated when the load F is exerted
U  ∆R51 ∆R52 ∆R53 ∆R54 
∆U Fx =  − + −  on a certain point, given the linear relation between the
4  R51 R52 R53 R54  applied load and deformation.
  ∆R51 
U  ∆R52   (35)
=  2   − 2   X = [δ x, δ y, δ z,θ x,θ y,θ z ]T
4  R51 ε  R52 ε  (40)
= CF = C[ Fx, Fy, Fz, Mx, My, Mz ]T
UK
= (ε 51 + ε 52 )
2
where C is the compliance matrix of the CPM.
Each limb of a parallel mechanism can be considered as a
U  ∆R61 ∆R62 ∆R63 ∆R64  serial chain that comprises m joints, which are assumed to
∆U Fy =  − + − 
4  R61 R62 R63 R64  be flexure members with the local compliance matrix cij -
(36) established in the local frame. Given the deformations only
U   ∆R61   ∆R62  
=  2   − 2   relate to the compliant joints, we can obtain the lumped
4   R61 ε  R62 ε  deformation at the tip of the ith limb
UK
= (ε 61 + ε 62 ) m m
2 X i = ∑ xijR = ∑ J ij xij (41)
j =1 i =1

Where K is the sensitivity coefficient of the strain gauges,


εi is the elastic strain at the spot where the Ri is bonded on where xj and xjR are the elastic deformations of the flexure
the diaphragms, U is the excitation voltage, and ∆Ri/Ri member of the chain with respect to the local and the
means change rate of the resistance of the strain gauge Ri reference frame, respectively. And Jij can be derived as
due to strain variation. follows:
The corresponding output of each sensing limb can be
calculated as  R -R S( r )
J ij =  ij ij ij  (42)
0 Rij 
U  ∆R11 ∆R12 ∆R13 ∆R14 
∆U limb1 =  − + − 
4  R11 R12 R13 R14  which requires knowledge of the position rij of the origin of
(37) reference frame with respect to local frame as well as the
U   ∆R11   ∆R12   orientation Rij of reference frame with respect to local frame.
=  2   − 2  
4   R11 ε  R12 ε  And S(*) is the skew-symmetric operator.
UK Similarly, the wrench Fi applied on the tip of the ith limb
= (ε11 + ε12 ) described in the reference frame can be distributed to
2
wrenches fij on the jth flexure member of the ith limb:
U  ∆R21 ∆R22 ∆R23 ∆R24 
∆Ulimb2 =  − + −  fij = J Fij Fi (43)
4  R21 R22 R23 R24 
U   ∆R21   ∆R22   (38) Where JFij = JijT. From (26) and (29)
=  2  − 2  
4   R21 ε  R22 ε  m m m
(44)
UK
X i = C i Fi = ∑ J ij x ij = ∑ J ij c ij f ij = ∑ J ij c ij J ijT F i
= (ε 21 + ε 22 ) j =1 j =1 j =1

2
Therefore, the compliance of the ith limb is given by
U  ∆R ∆R ∆R ∆R 
∆Ulimb3 =  31 − 32 + 33 − 34  m
4  R31 R32 R33 R34  Ci = ∑ J ij cij J ijT (45)
j =1
U   ∆R31   ∆R32   (39)
=  2   − 2  
4   R31 ε  R32 ε  From the kinematic analysis, the mobile platform and tips
of limbs have the same angular displacements but different
UK
= (ε 31 + ε 32 ) linear displacements. Particularly, the displacement vector
2 of the CPM in X and the displacement vector of the tip of
the ith limb Xi can be transformed reciprocally,

Unauthenticated
260 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

X = Ji X i (46) Table 3. Normal strains in the locations of strain gauges.

Fx=1N Fy=1N Fz=1N Mx=10N· My=10N· Mz=10


where Ji is the transformation matrix from the reference of mm mm N·mm
ith limb, whose reference point is set at the tips of limbs, to ε11 -2.57 2.259 -4.80 2.856 -2.821 -1.967
the reference of the CPM. Similarly, the wrench F applied e-5 e-5 e-5 e-4 e-4 e-4
on the platform described in the reference frame of CPM ε12 2.497 -2.22 4.801 -2. 781 2.827 1.9736
can be obtained via distributed wrench Fi described in the e-5 e-5 e-5 e-4 e-4 e-4
reference frame of the limbs ε13 -2.61 2.301 -4.80 2.861 -2.812 -2.105
e-5 e-5 e-5 e-4 e-4 e-4
n ε14 2.526 -2.29 4.798 -2.859 2.819 2.0982
F = KX = ∑ J Fi Fi (47) e-5 e-5 e-5 e-4 e-4 e-4
i =1 ε21 -1.09 -1.10 -4.80 -2.732 2.601 -2.132
e-5 e-5 e-5 e-4 e-7 e-4
where JFi is the transformation matrix of the applied wrench ε22 1.102 1.101 4.879 2.728 -2.594 1.9861
from the local frame to the reference frame: e-5 e-5 e-5 e-4 e-7 e-4
ε23 -1.12 -1.11 -4.80 -2.709 2.610 -2.110
e-5 e-5 e-5 e-4 e-7 e-4
J Fi = J i T = 
Ri 0 (48) ε24 1.241 1.239 4.799 2.173 -2.598 2.1092
 S( ri )R Ri  e-5 e-5 e-5 e-4 e-7 e-4
ε31 2.261 -2.56 -4.80 2.571 2.863 -1.985
Let K and ki be the stiffness matrix of the CPM and the e-5 e-5 e-5 e-7 e-4 e-4
stiffness of the ith limb, respectively. Form (42) and (43), ε32 -2.21 2.486 4.802 -2.497 -2. 871 1.9842
one can obtain e-5 e-5 e-5 e-7 e-4 e-4
ε33 2.312 -2.61 -4.80 2.586 2.874 -2.098
n n n
e-5 e-5 e-5 e-7 e-4 e-4
F = KX = ∑ J Fi Fi = ∑ J Fi K i X i = ∑ J iT K i J i−1 X (49) ε34 -2.29 2.524 4.797 -2.579 -2.869 2.1025
i =1 i =1 i =1 e-5 e-5 e-5 e-7 e-4 e-4
ε41 -5.43 4.481 3.157 -5.892 2.502 3.586
Then, the stiffness of the CPM becomes e-5 e-5 e-4 e-8 e-7 e-6
ε42 -4.74 4.931 -2.95 5.886 -2.509 -3.58
n
e-5 e-5 e-4 e-8 e-7 e-6
K = ∑ J iT K i J i−1 (50) ε43 -5.63 4.627 3.207 7.028 -2.839 -3.22
i =1 e-5 e-5 e-4 e-8 e-7 e-6
ε44 -5.70 4.692 -3.14 -7.019 2.849 3.211
6. FINITE ELEMENT ANALYSIS e-5 e-5 e-4 e-8 e-7 e-6
ε51 -3.25 1.590 1.134 2.497 1.598 -1.56
Finite Element Analysis (FEA) via software ANSYS® was e-4 e-5 e-4 e-7 e-6 e-5
performed to certify that the sensor possesses appropriate ε52 2.999 1.621 1.136 -2.499 -1.611 1.556
sensitivity and compliance. The normal strains in the e-4 e-5 e-4 e-7 e-6 e-5
locations of strain gauges are listed in Table 3. when the ε53 -2.97 -2.19 1.141 -2.847 1.798 -1.07
sensor undergoes single component load. e-4 e-5 e-4 e-7 e-6 e-5
ε54 2.982 -2.70 1.142 2.851 -1.792 1.110
The measurement sensitivity defined as the ratios of the
e-4 e-5 e-4 e-7 e-6 e-5
output voltages to the input voltages can be obtained by ε61 1.611 -3.23 1.129 1.644 -5.901 -1.07
using the strain results listed in Table 3. [4]. Sufficient e-5 e-4 e-4 e-6 e-8 e-5
measuring isotropy among components ensures that the ε62 1.601 3.109 1.139 -1.651 5.890 1.112
electrical circuit of a multi-component sensor possesses e-5 e-4 e-4 e-6 e-8 e-5
amplification symmetry, high degree of integration and ε63 -2.17 -3.09 1.140 1.838 7.031 -1.57
simple decoupling methods. The Anisotropy index of the e-5 e-4 e-4 e-6 e-8 e-5
ε64 -2.06 3.072 1.139 -1.829 -7.021 1.561
sensor is calculated by using the obtained measurement
e-5 e-4 e-4 e-6 e-8 e-5
sensitivities. These are listed in Table 4.
When resistive or gravitational forces/torques are exerted
Table 4. Measurement sensitivity and anisotropy index of the
on the mobile platform, the compliant spherical joint in the sensor (Gauge Factor K=2.1).
central leg and the compliant prismatic joints in each Measurement
Fx Fy Fz f1 f2 f3
sensing limb are compressed or extended. As a sensitivities (Si)
consequence, the CPM is also able to provide passive
compliance to the mobile platform during parts handling and ∆U/U (mV/V) 1.281 1.318 1.267 1.192 1.184 1.207
assembly. The displacements of the sensor are indicated in
Fig.12. when it is subjected to the single rated component
Isotropy index
force/torque. By calculating, the maximum Max{Si } − Min{Si }
displacements/rotations it can provide are 0.071 mm along 0.107846
x-axis, 0.008 mm along z-axis, and 0.085° about x-axis, and ∑S
6
i
2
6

0.018° about z-axis.

Unauthenticated
261 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

7. CALIBRATION METHOD AND DECOUPLING MATRIX


Calibration is used to adjust the output voltages of the
electrical circuit to agree with the value of the applied
standard within a specified accuracy [35]. Besides, the
majority of the existing multi-component force/torque
sensors are confronted by high coupled interference errors
among their six components, especially between component
Fx and component My, component Fy and component Mx
respectively, which need a complicated calibration and
decoupling test to get the decoupling matrix. With the
(a) analysis mentioned in Section 4 and the method proposed in
[35], [36], [37], the decoupling matrix C of the proposed
sensor can be obtained.

λ I C1 
F = CS =  3×3 S (51)
 C2 λ J 

Where the S is the output of the bridge circuits, λ is the


linear coefficient between output of the bridge circuits to
standard load, and C1 and C2 represent the influence between
the three moment components and the three force
components.
(b)

8. CONCLUSIONS AND FUTURE WORK


This study has endeavored to design and analyze a novel
six-component force/torque sensor based on the mechanism
of CPM. The sensor is used to measure six-component
force/torque simultaneously and to provide passive
compliance during in parts handling and assembly. The
sensing element of the sensor is newly modeled using a
monolithic CPM. The kinematic transformations for the
force sensor, the measuring principle and compliance
analysis are introduced. The FEA is performed and its
results show that the proposed sensor possesses high
sensitivity, good isotropy, simple decoupling method and
(c) moderate compliance.
Further activities will be devoted to the dynamic analysis
and automatic manipulation based on real-time computation
of forces and torques. Moreover, fabrication and extensive
experimentation of the six-component F/T sensor based on
the proposed CPM is planned. Finally, an experimental
platform integrated with driver, the developed sensor, and
control algorithm will be developed.

ACKNOWLEDGMENT
This work was supported in part by the National Nature
Science Foundation of China (NSFC 61203207), specialized
Research Fund for the Doctoral Program of Higher
Education (No. 20120161120015), Hunan province Science
(d)
and Technology Planning Project of China (2012RS4046)
and China Postdoctoral Science Foundation (2012M510189,
Fig.12. The flexible deformation of the CPM under applied force 2013T60768). The second author appreciates the financial
Fx=1N (a), the force Fz=1N (b), the moment Mx =0.05N·m(c) and support from Natural Sciences and Engineering Research
Mz =0.05 N·m (d). Council of Canada (NSERC).

Unauthenticated
262 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

REFERENCES [14] Dasgupta, B., Reddy, S., Mruthyunjaya, T.S. (1994).


[1] Nakamura, Y., Yoshikawa, T., Futamata, I. (1998). Synthesis of a force–torque sensor based on the
Design and signal processing of six-axis force sensors. Stewart platform mechanism. In Proceedings of the
National Convention of Industrial Problems in
In Robotics Research : The Fourth International
Machines and Mechanisms, Bangalore, India, 14-23.
Symposium. MIT Press, 75-81.
[15] Hou, Y., Zeng, D., Yao, J., Kang, K., Lu, L., Zhao, Y.
[2] Kim, J.H., Kang, D.I., Shin, H.H., Park, Y.K. (2003).
(2009). Optimal design of a hyperstatic Stewart
Design and analysis of a column type multi- platform-based force/torque sensor with genetic
component force/moment sensor. Measurement, 33, algorithms. Mechatronics, 19 (2), 199-204.
213-219. [16] Jia, Z.Y., Lin, S., Liu, W. (2010). Measurement
[3] Kim, G.S. (2001). The design of a six-component method of six-axis load sharing based on the Stewart
force/moment sensor and evaluation of its uncertainty. platform. Measurement, 43 (3), 329-335.
Measurement Science and Technology, 12 (9), 1445- [17] Liu, W., Li, Y.J., Jia, Z.Y., Zhang, J., Qian, M. (2011).
1455. Research on parallel load sharing principle of
[4] Hashimoto, K. et al. (2013). Overload protection piezoelectric six-dimensional heavy force/torque
mechanism for 6-axis force/torque sensor. In Romansy sensor. Mechanical Systems and Signal Processing, 25
19 – Robot Design, Dynamics and Control. Springer, (1), 331-343.
Vol. 544, 383-390. [18] Jin, W.L., Mote, C.D., Jr. (1998). A six-component
[5] Liu, S.A., Tzo, H.L. (2002). A novel six-component silicon micro force sensor. Sensors and Actuators A:
force sensor of good measurement isotropy and Physical, 65 (2-3), 109-115.
sensitivities. Sensors and Actuators A: Physical, 100 [19] Mei, T., Ge, Y., Chen, Y., Ni, L., Liao, W.H., Xu, Y.,
(2-3), 223-230. Li, W.J. (1999). Design and fabrication of an
[6] Liang, Q., Zhang, D., Wang, Y., Coppola, G., Ge, Y. integrated three-dimensional tactile sensor for space
(2013). PM based multi-component F/T sensors—state robotic applications. In Micro Electro Mechanical
Systems (MEMS ’99) : 12th IEEE International
of the art and trends. Robotics and Computer-
Conference, 17-21 January 1999. IEEE, 112-117.
Integrated Manufacturing, 29 (4), 1-7.
[20] Brookhuis, R.A., Lammerink, T.S.J., Wiegerink, R.J.,
[7] ATI Industrial Automation. Multi-axis force / torque
de Boer, M.J., Elwenspoek, M.C. (2012). 3D force
sensors. sensor for biomechanical applications. Sensors and
http://www.ati-ia.com/products/ft/sensors.aspx. Actuators A: Physical, 182, 28-33.
[8] Baki, P., Szekely, G., Kosa, G. (2012). Miniature tri- [21] Takenawa, S. (2009). A soft three-axis tactile sensor
axial force sensor for feedback in minimally invasive based on electromagnetic induction. In Mechatronics
surgery. In Biomedical Robotics and Biomechatronics 2009. ICM 2009 : IEEE International Conference, 14-
(BioRob) : 4th IEEE RAS & EMBS International 17 April 2009. IEEE, 1-6.
Conference, 24-27 June 2012. IEEE, 805-810. [22] Liu, T., Inoue, Y., Shibata, K., Yamasaki, Y.,
[9] Mastinu, G., Gobbi, M., Previati, G. (2011). A new Nakahama, M. (2004). A six-dimension parallel force
six-axis load cell. Part I: Design. Experimental sensor for human dynamics analysis. In Robotics,
Mechanics, 51 (3), 373-388. Automation and Mechatronics, 1-3 December 2004.
[10] Gailler, A., Reboulet, C. (1983). An isostatic six IEEE, 208-212.
component force and torque sensor. In 13th [23] Hirose, S., Yoneda, K. (1990). Development of optical
International Symposium on Industrial Robots and six-axial force sensor and its signal calibration
Robots 7, 17-21 April 1983. Robotics International of considering nonlinear interference. In Robotics and
SME. Automation, 13-18 May 1990. IEEE, 46-53.
[11] Dwarakanath, T.A., Bhaumick, T.K., Venkatesh, D. [24] Gobbi, M., Previati, G., Guarneri, P., Mastinu, G.
(1999). Implementation of Stewart platform based (2011). A new six-axis load cell. Part II: Error
analysis, construction and experimental assessment of
force-torque sensor. In Multisensor Fusion and
performances. Experimental Mechanics, 51 (3), 389-
Integration for Intelligent Systems (MFI ’99) :
399.
IEEE/SICE/RSJ International Conference, 15-18
[25] Trease, B.P., Moon, Y.M., Kota, S. (2005). Design of
August, 1999, 32-37. large-displacement compliant joints. ASME Journal of
[12] Ranganath, R., Nair, P.S., Mruthyunjaya, T.S., Ghosal, Mechanical Design, 127, 788-798.
A. (2004). A force-torque sensor based on a Stewart [26] Zhu, Z.H., Meguid, S.A. (2008). Vibration analysis of
platform in a near-singular configuration. Mechanism a new curved beam element. Journal of Sound and
and Machine Theory, 39 (9), 971-998. Vibration, 309 (1), 86-95.
[13] Nguyen, C., Antrazi, S., Zhou, Z. (1991). Analysis and [27] Wu, T., Chen, J., Chang, S. (2008) A six-DOF
implementation of a 6 DOF Stewart platform-based prismatic-spherical-spherical parallel compliant
force sensor for passive compliant robotic assembly. In nanopositioner. IEEE Transactions on Ultrasonics
IEEE Proceedings of Southeastcon ‘91, 7-10 April Ferroelectrics and Frequency Control, 55 (12), 2544-
1991. IEEE, 880-884. 2551.

Unauthenticated
263 Download Date | 11/23/17 11:12 PM
MEASUREMENT SCIENCE REVIEW, Volume 13, No. 5, 2013

[28] Man Bok Hong, Yung-Ho Jo. (2012). Design and [34] Ouyang, P.R. (2005). Hybrid intelligent machine
evaluation of 2-DOF compliant forceps with force- systems: Design, modeling and control. Ph.D.
sensing capability for minimally invasive robot dissertation, University of Saskatchewan, Canada.
surgery. IEEE Transactions on Robotics, 28 (4), 932- [35] Liang, Q., Zhang, D., Song, Q., Ge, Y. (2010).
941. Micromanipulator with integrated force sensor based
[29] Dong, W., Sun, L., Du, Z. (2008). Stiffness research on compliant parallel mechanism. In Robotics and
on a high-precision, large-workspace parallel Biomimetics (ROBIO 2010), 14-18 December 2010.
mechanism with compliant joints. Precision IEEE, 709-714.
Engineering, 32 (3), 222-231. [36] Puangmali, P. et al. (2012). Miniature 3-axis distal
[30] Paros, J.M., Weisbord, L. (1965). How to design force sensor for minimally invasive surgical
flexure hinges. Machine Design, 37, 151-156. palpation. IEEE/ASME Transactions on Mechatronics,
[31] Smith, S. (2000). Flexures: Elements of Elastic 17 (4), 646-656.
Mechanisms. New York: Gordon and Breach Science [37] Bicchi, A. (1992). A criterion for optimal design of
Publishers. multi-axis force sensors. Robotics and Autonomous
[32] Boyes, W. (2009). Instrumentation Reference Book, Systems, 10 (4), 269-286.
3rd Edition. Burlington, MA: Elsevier.
[33] Sameer A. Joshi. (2002). A comparative study of two
classes of 3-DOF parallel manipulators. Ph.D.
dissertation, Department of Mechanical Engineering, Received January 14, 2013.
Accepted October 17, 2013.
University of Maryland, College Park, MD.

Unauthenticated