Вы находитесь на странице: 1из 57

HORIZONTAL DEFLECTION

Deflection Factor 5mV to 20V/DIV on 12 ranges in 1-2-5 step with fine control. DC
Frequency Responde lo lMHz(—3dB)
Input Inpedance Max 1MÍ2 shouted by 20pF ±3pF
Input Voltage X-Y 300V DC + AC peak or 600Vp-p
Operation X-Y mode is selected by SWEEP TIME/DIV switch
CH-A: Y axis
CH-B: Xaxis
Intensity Modulalion Z Axis: TTL Level (3Vp-p ~ 50V) + bright, — dark.

OTHER SPECIFICATIONS
CRTHV APPROX-2KV
Calibration Voltage 0.5Vp-p ±5%, 1 kllz Square Wave
Power Requrements AC: 100V/120V/220/V240V/, 50/60Hz, 19W
Weight Dimensions 7kg approx.
162(H) X 294(W) X 352(D)mm

SECTION 3 OPERATION

3-1 1NITIAL OPERATION


Inspect the cartón for serious damage which might have caused failure of the instrument during transportaron. If damage is noted,
notify the agent you bought from before turning on.

INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
a) Set the POWER switch to OFF.
b) Tum all the three POSITION Controls to mid-position.
c) Tum INTENSITY control to mid-position.
d) Push the PULL 5X MAC control for normal.
e) Pulí TR1GGERING LEVEL control for AUTO.
f) The rest of the Controls remain at any position for normal operation.
g) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
a) Set the POWER switch to OFF.
b) Tum all the three POSITION Controls to mid-position.
c) Tum INTENSITY control to mid-position.
d) Push the PULL 5X MAC control for normal.
e) Pulí TR1GGERING LEVEL control for AUTO.
f) The rest of the Controls remain at any position for normal operation.
g) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
a) Set the POWER switch to OFF.
b) Tum all the three POSITION Controls to mid-position.
c) Tum INTENSITY control to mid-position.
d) Push the PULL 5X MAC control for normal.
e) Pulí TR1GGERING LEVEL control for AUTO.
f) The rest of the Controls remain at any position for normal operation.
g) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
a) Set the POWER switch to OFF.
b) Tum all the three POSITION Controls to mid-position.
c) Tum INTENSITY control to mid-position.
d) Push the PULL 5X MAC control for normal.
e) Pulí TR1GGERING LEVEL control for AUTO.
f) The rest of the Controls remain at any position for normal operation.
g) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
a) Set the POWER switch to OFF.
b) Tum all the three POSITION Controls to mid-position.
c) Tum INTENSITY control to mid-position.
d) Push the PULL 5X MAC control for normal.
e) Pulí TR1GGERING LEVEL control for AUTO.
f) The rest of the Controls remain at any position for normal operation.
h) Check the line voltage. POWER switch to OFF.
i) Tum all the three POSITION Controls to mid-position.
j) Tum INTENSITY control to mid-position.
k) Push the PULL 5X MAC control for normal.
l) Pulí TR1GGERING LEVEL control for AUTO.
m) The rest of the Controls remain at any position for normal operation.
n) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
h) Set the POWER switch to OFF.
i) Tum all the three POSITION Controls to mid-position.
j) Tum INTENSITY control to mid-position.
k) Push the PULL 5X MAC control for normal.
l) Pulí TR1GGERING LEVEL control for AUTO.
m) The rest of the Controls remain at any position for normal operation.
n) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
g) Set the POWER switch to OFF.
h) Tum all the three POSITION Controls to mid-position.
i) Tum INTENSITY control to mid-position.
j) Push the PULL 5X MAC control for normal.
k) Pulí TR1GGERING LEVEL control for AUTO.
l) The rest of the Controls remain at any position for normal operation.
m) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
o) Set the POWER switch to OFF.
o) POWER switch to OFF.
p) Tum all the three POSITION Controls to mid-position.
q) Tum INTENSITY control to mid-position.
r) Push the PULL 5X MAC control for normal.
s) Pulí TR1GGERING LEVEL control for AUTO.
t) The rest of the Controls remain at any position for normal operation.
u) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
p) Set the POWER switch to OFF.
q) Tum all the three POSITION Controls to mid-position.
r) Tum INTENSITY control to mid-position.
s) Push the PULL 5X MAC control for normal.
t) Pulí TR1GGERING LEVEL control for AUTO.
u) The rest of the Controls remain at any position for normal operation.
v) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
n) Set the POWER switch to OFF.
o) Tum all the three POSITION Controls to mid-position.
p) Tum INTENSITY control to mid-position.
q) Push the PULL 5X MAC control for normal.
r) Pulí TR1GGERING LEVEL control for AUTO.
s) The rest of the Controls remain at any position for normal operation.
t) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
w) Set the POWER switch to OFF.
v) POWER switch to OFF.
w) Tum all the three POSITION Controls to mid-position.
x) Tum INTENSITY control to mid-position.
y) Push the PULL 5X MAC control for normal.
z) Pulí TR1GGERING LEVEL control for AUTO.
aa) The rest of the Controls remain at any position for normal operation.
bb) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
x) Set the POWER switch to OFF.
y) Tum all the three POSITION Controls to mid-position.
z) Tum INTENSITY control to mid-position.
aa) Push the PULL 5X MAC control for normal.
bb) Pulí TR1GGERING LEVEL control for AUTO.
cc) The rest of the Controls remain at any position for normal operation.
dd) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
u) Set the POWER switch to OFF.
v) Tum all the three POSITION Controls to mid-position.
w) Tum INTENSITY control to mid-position.
x) Push the PULL 5X MAC control for normal.
y) Pulí TR1GGERING LEVEL control for AUTO.
z) The rest of the Controls remain at any position for normal operation.
aa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ee) Set the POWER switch to OFF.
cc) POWER switch to OFF.
dd) Tum all the three POSITION Controls to mid-position.
ee) Tum INTENSITY control to mid-position.
ff) Push the PULL 5X MAC control for normal.
gg) Pulí TR1GGERING LEVEL control for AUTO.
hh) The rest of the Controls remain at any position for normal operation.
ii) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ff) Set the POWER switch to OFF.
gg) Tum all the three POSITION Controls to mid-position.
hh) Tum INTENSITY control to mid-position.
ii) Push the PULL 5X MAC control for normal.
jj) Pulí TR1GGERING LEVEL control for AUTO.
kk) The rest of the Controls remain at any position for normal operation.
ll) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bb) Set the POWER switch to OFF.
cc) Tum all the three POSITION Controls to mid-position.
dd) Tum INTENSITY control to mid-position.
ee) Push the PULL 5X MAC control for normal.
ff) Pulí TR1GGERING LEVEL control for AUTO.
gg) The rest of the Controls remain at any position for normal operation.
hh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
mm) Set the POWER switch to OFF.
jj) POWER switch to OFF.
kk) Tum all the three POSITION Controls to mid-position.
ll) Tum INTENSITY control to mid-position.
mm) Push the PULL 5X MAC control for normal.
nn) Pulí TR1GGERING LEVEL control for AUTO.
oo) The rest of the Controls remain at any position for normal operation.
pp) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
nn) Set the POWER switch to OFF.
oo) Tum all the three POSITION Controls to mid-position.
pp) Tum INTENSITY control to mid-position.
qq) Push the PULL 5X MAC control for normal.
rr) Pulí TR1GGERING LEVEL control for AUTO.
ss) The rest of the Controls remain at any position for normal operation.
tt) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ii) Set the POWER switch to OFF.
jj) Tum all the three POSITION Controls to mid-position.
kk) Tum INTENSITY control to mid-position.
ll) Push the PULL 5X MAC control for normal.
mm) Pulí TR1GGERING LEVEL control for AUTO.
nn) The rest of the Controls remain at any position for normal operation.
oo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
uu) Set the POWER switch to OFF.
qq) POWER switch to OFF.
rr) Tum all the three POSITION Controls to mid-position.
ss) Tum INTENSITY control to mid-position.
tt) Push the PULL 5X MAC control for normal.
uu) Pulí TR1GGERING LEVEL control for AUTO.
vv) The rest of the Controls remain at any position for normal operation.
ww) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
vv) Set the POWER switch to OFF.
ww) Tum all the three POSITION Controls to mid-position.
xx) Tum INTENSITY control to mid-position.
yy) Push the PULL 5X MAC control for normal.
zz) Pulí TR1GGERING LEVEL control for AUTO.
aaa) The rest of the Controls remain at any position for normal operation.
bbb) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
pp) Set the POWER switch to OFF.
qq) Tum all the three POSITION Controls to mid-position.
rr) Tum INTENSITY control to mid-position.
ss) Push the PULL 5X MAC control for normal.
tt) Pulí TR1GGERING LEVEL control for AUTO.
uu) The rest of the Controls remain at any position for normal operation.
vv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ccc) Set the POWER switch to OFF.
xx) POWER switch to OFF.
yy) Tum all the three POSITION Controls to mid-position.
zz) Tum INTENSITY control to mid-position.
aaa) Push the PULL 5X MAC control for normal.
bbb) Pulí TR1GGERING LEVEL control for AUTO.
ccc) The rest of the Controls remain at any position for normal operation.
ddd) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ddd) Set the POWER switch to OFF.
eee) Tum all the three POSITION Controls to mid-position.
fff) Tum INTENSITY control to mid-position.
ggg) Push the PULL 5X MAC control for normal.
hhh) Pulí TR1GGERING LEVEL control for AUTO.
iii) The rest of the Controls remain at any position for normal operation.
jjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ww) Set the POWER switch to OFF.
xx) Tum all the three POSITION Controls to mid-position.
yy) Tum INTENSITY control to mid-position.
zz) Push the PULL 5X MAC control for normal.
aaa) Pulí TR1GGERING LEVEL control for AUTO.
bbb) The rest of the Controls remain at any position for normal operation.
ccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkk) Set the POWER switch to OFF.
eee) POWER switch to OFF.
fff) Tum all the three POSITION Controls to mid-position.
ggg) Tum INTENSITY control to mid-position.
hhh) Push the PULL 5X MAC control for normal.
iii) Pulí TR1GGERING LEVEL control for AUTO.
jjj) The rest of the Controls remain at any position for normal operation.
kkk) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
lll) Set the POWER switch to OFF.
mmm) Tum all the three POSITION Controls to mid-position.
nnn) Tum INTENSITY control to mid-position.
ooo) Push the PULL 5X MAC control for normal.
ppp) Pulí TR1GGERING LEVEL control for AUTO.
qqq) The rest of the Controls remain at any position for normal operation.
rrr) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ddd) Set the POWER switch to OFF.
eee) Tum all the three POSITION Controls to mid-position.
fff) Tum INTENSITY control to mid-position.
ggg) Push the PULL 5X MAC control for normal.
hhh) Pulí TR1GGERING LEVEL control for AUTO.
iii) The rest of the Controls remain at any position for normal operation.
jjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
sss) Set the POWER switch to OFF.
lll) POWER switch to OFF.
mmm) Tum all the three POSITION Controls to mid-position.
nnn) Tum INTENSITY control to mid-position.
ooo) Push the PULL 5X MAC control for normal.
ppp) Pulí TR1GGERING LEVEL control for AUTO.
qqq) The rest of the Controls remain at any position for normal operation.
rrr) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ttt) Set the POWER switch to OFF.
uuu) Tum all the three POSITION Controls to mid-position.
vvv) Tum INTENSITY control to mid-position.
www) Push the PULL 5X MAC control for normal.
xxx) Pulí TR1GGERING LEVEL control for AUTO.
yyy) The rest of the Controls remain at any position for normal operation.
zzz) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkk) Set the POWER switch to OFF.
lll) Tum all the three POSITION Controls to mid-position.
mmm) Tum INTENSITY control to mid-position.
nnn) Push the PULL 5X MAC control for normal.
ooo) Pulí TR1GGERING LEVEL control for AUTO.
ppp) The rest of the Controls remain at any position for normal operation.
qqq) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
aaaa) Set the POWER switch to OFF.
sss) POWER switch to OFF.
ttt) Tum all the three POSITION Controls to mid-position.
uuu) Tum INTENSITY control to mid-position.
vvv) Push the PULL 5X MAC control for normal.
www) Pulí TR1GGERING LEVEL control for AUTO.
xxx) The rest of the Controls remain at any position for normal operation.
yyy) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bbbb) Set the POWER switch to OFF.
cccc) Tum all the three POSITION Controls to mid-position.
dddd) Tum INTENSITY control to mid-position.
eeee) Push the PULL 5X MAC control for normal.
ffff) Pulí TR1GGERING LEVEL control for AUTO.
gggg) The rest of the Controls remain at any position for normal operation.
hhhh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrr) Set the POWER switch to OFF.
sss) Tum all the three POSITION Controls to mid-position.
ttt) Tum INTENSITY control to mid-position.
uuu) Push the PULL 5X MAC control for normal.
vvv) Pulí TR1GGERING LEVEL control for AUTO.
www) The rest of the Controls remain at any position for normal operation.
xxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiii) Set the POWER switch to OFF.
zzz) POWER switch to OFF.
aaaa) Tum all the three POSITION Controls to mid-position.
bbbb) Tum INTENSITY control to mid-position.
cccc) Push the PULL 5X MAC control for normal.
dddd) Pulí TR1GGERING LEVEL control for AUTO.
eeee) The rest of the Controls remain at any position for normal operation.
ffff) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
jjjj) Set the POWER switch to OFF.
kkkk) Tum all the three POSITION Controls to mid-position.
llll) Tum INTENSITY control to mid-position.
mmmm) Push the PULL 5X MAC control for normal.
nnnn) Pulí TR1GGERING LEVEL control for AUTO.
oooo) The rest of the Controls remain at any position for normal operation.
pppp) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyy) Set the POWER switch to OFF.
zzz) Tum all the three POSITION Controls to mid-position.
aaaa) Tum INTENSITY control to mid-position.
bbbb) Push the PULL 5X MAC control for normal.
cccc) Pulí TR1GGERING LEVEL control for AUTO.
dddd) The rest of the Controls remain at any position for normal operation.
eeee) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
qqqq) Set the POWER switch to OFF.
gggg) POWER switch to OFF.
hhhh) Tum all the three POSITION Controls to mid-position.
iiii) Tum INTENSITY control to mid-position.
jjjj) Push the PULL 5X MAC control for normal.
kkkk) Pulí TR1GGERING LEVEL control for AUTO.
llll) The rest of the Controls remain at any position for normal operation.
mmmm) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrrr) Set the POWER switch to OFF.
ssss)Tum all the three POSITION Controls to mid-position.
tttt) Tum INTENSITY control to mid-position.
uuuu) Push the PULL 5X MAC control for normal.
vvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwww) The rest of the Controls remain at any position for normal operation.
xxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ffff) Set the POWER switch to OFF.
gggg) Tum all the three POSITION Controls to mid-position.
hhhh) Tum INTENSITY control to mid-position.
iiii) Push the PULL 5X MAC control for normal.
jjjj) Pulí TR1GGERING LEVEL control for AUTO.
kkkk) The rest of the Controls remain at any position for normal operation.
llll) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyyy) Set the POWER switch to OFF.
nnnn) POWER switch to OFF.
oooo) Tum all the three POSITION Controls to mid-position.
pppp) Tum INTENSITY control to mid-position.
qqqq) Push the PULL 5X MAC control for normal.
rrrr) Pulí TR1GGERING LEVEL control for AUTO.
ssss)The rest of the Controls remain at any position for normal operation.
tttt) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
zzzz) Set the POWER switch to OFF.
aaaaa) Tum all the three POSITION Controls to mid-position.
bbbbb) Tum INTENSITY control to mid-position.
ccccc) Push the PULL 5X MAC control for normal.
ddddd) Pulí TR1GGERING LEVEL control for AUTO.
eeeee) The rest of the Controls remain at any position for normal operation.
fffff) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
mmmm) Set the POWER switch to OFF.
nnnn) Tum all the three POSITION Controls to mid-position.
oooo) Tum INTENSITY control to mid-position.
pppp) Push the PULL 5X MAC control for normal.
qqqq) Pulí TR1GGERING LEVEL control for AUTO.
rrrr) The rest of the Controls remain at any position for normal operation.
ssss)Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ggggg) Set the POWER switch to OFF.
tttt)
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
hhhhh) Set the POWER switch to OFF.
iiiii) Tum all the three POSITION Controls to mid-position.
jjjjj) Tum INTENSITY control to mid-position.
kkkkk) Push the PULL 5X MAC control for normal.
lllll) Pulí TR1GGERING LEVEL control for AUTO.
mmmmm) The rest of the Controls remain at any position for normal operation.
nnnnn) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
uuuu) Set the POWER switch to OFF.
vvvv) Tum all the three POSITION Controls to mid-position.
wwww) Tum INTENSITY control to mid-position.
xxxx) Push the PULL 5X MAC control for normal.
yyyy) Pulí TR1GGERING LEVEL control for AUTO.
zzzz) The rest of the Controls remain at any position for normal operation.
aaaaa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ooooo) Set the POWER switch to OFF.
ppppp) Tum all the three POSITION Controls to mid-position.
qqqqq) Tum INTENSITY control to mid-position.
rrrrr) Push the PULL 5X MAC control for normal.
sssss) Pulí TR1GGERING LEVEL control for AUTO.
ttttt) The rest of the Controls remain at any position for normal operation.
uuuuu) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bbbbb) Set the POWER switch to OFF.
ccccc) Tum all the three POSITION Controls to mid-position.
ddddd) Tum INTENSITY control to mid-position.
eeeee) Push the PULL 5X MAC control for normal.
fffff) Pulí TR1GGERING LEVEL control for AUTO.
ggggg) The rest of the Controls remain at any position for normal operation.
hhhhh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
vvvvv) Set the POWER switch to OFF.
wwwww) Tum all the three POSITION Controls to mid-position.
xxxxx) Tum INTENSITY control to mid-position.
yyyyy) Push the PULL 5X MAC control for normal.
zzzzz) Pulí TR1GGERING LEVEL control for AUTO.
aaaaaa) The rest of the Controls remain at any position for normal operation.
bbbbbb) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiiii) Set the POWER switch to OFF.
jjjjj) Tum all the three POSITION Controls to mid-position.
kkkkk) Tum INTENSITY control to mid-position.
lllll) Push the PULL 5X MAC control for normal.
mmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnn) The rest of the Controls remain at any position for normal operation.
ooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
cccccc) Set the POWER switch to OFF.
dddddd) Tum all the three POSITION Controls to mid-position.
eeeeee) Tum INTENSITY control to mid-position.
ffffff) Push the PULL 5X MAC control for normal.
gggggg) Pulí TR1GGERING LEVEL control for AUTO.
hhhhhh) The rest of the Controls remain at any position for normal operation.
uuuu) Check the line voltage. Set the POWER switch to OFF.
vvvv) Tum all the three POSITION Controls to mid-position.
wwww) Tum INTENSITY control to mid-position.
xxxx) Push the PULL 5X MAC control for normal.
yyyy) Pulí TR1GGERING LEVEL control for AUTO.
zzzz) The rest of the Controls remain at any position for normal operation.
aaaaa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiiiii) Set the POWER switch to OFF.
jjjjjj) Tum all the three POSITION Controls to mid-position.
kkkkkk) Tum INTENSITY control to mid-position.
llllll) Push the PULL 5X MAC control for normal.
mmmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnnn) The rest of the Controls remain at any position for normal operation.
oooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ppppp) Set the POWER switch to OFF.
qqqqq) Tum all the three POSITION Controls to mid-position.
rrrrr) Tum INTENSITY control to mid-position.
sssss) Push the PULL 5X MAC control for normal.
ttttt) Pulí TR1GGERING LEVEL control for AUTO.
uuuuu) The rest of the Controls remain at any position for normal operation.
vvvvv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
pppppp) Set the POWER switch to OFF.
qqqqqq) Tum all the three POSITION Controls to mid-position.
rrrrrr) Tum INTENSITY control to mid-position.
ssssss) Push the PULL 5X MAC control for normal.
tttttt) Pulí TR1GGERING LEVEL control for AUTO.
uuuuuu) The rest of the Controls remain at any position for normal operation.
vvvvvv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
wwwww) Set the POWER switch to OFF.
xxxxx) Tum all the three POSITION Controls to mid-position.
yyyyy) Tum INTENSITY control to mid-position.
zzzzz) Push the PULL 5X MAC control for normal.
aaaaaa) Pulí TR1GGERING LEVEL control for AUTO.
bbbbbb) The rest of the Controls remain at any position for normal operation.
cccccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
wwwwww) Set the POWER switch to OFF.
xxxxxx) Tum all the three POSITION Controls to mid-position.
yyyyyy) Tum INTENSITY control to mid-position.
zzzzzz) Push the PULL 5X MAC control for normal.
aaaaaaa) Pulí TR1GGERING LEVEL control for AUTO.
bbbbbbb) The rest of the Controls remain at any position for normal operation.
ccccccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
dddddd) Set the POWER switch to OFF.
eeeeee) Tum all the three POSITION Controls to mid-position.
ffffff) Tum INTENSITY control to mid-position.
gggggg) Push the PULL 5X MAC control for normal.
hhhhhh) Pulí TR1GGERING LEVEL control for AUTO.
iiiiii) The rest of the Controls remain at any position for normal operation.
jjjjjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ddddddd) Set the POWER switch to OFF.
eeeeeee) Tum all the three POSITION Controls to mid-position.
fffffff) Tum INTENSITY control to mid-position.
ggggggg) Push the PULL 5X MAC control for normal.
hhhhhhh) Pulí TR1GGERING LEVEL control for AUTO.
iiiiiii) The rest of the Controls remain at any position for normal operation.
jjjjjjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkkkkk) Set the POWER switch to OFF.
llllll) Tum all the three POSITION Controls to mid-position.
bbbbb) Set the POWER switch to OFF.
ccccc) Tum all the three POSITION Controls to mid-position.
ddddd) Tum INTENSITY control to mid-position.
eeeee) Push the PULL 5X MAC control for normal.
fffff) Pulí TR1GGERING LEVEL control for AUTO.
ggggg) The rest of the Controls remain at any position for normal operation.
hhhhh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkkkkkk) Set the POWER switch to OFF.
lllllll) Tum all the three POSITION Controls to mid-position.
mmmmmmm) Tum INTENSITY control to mid-position.
nnnnnnn) Push the PULL 5X MAC control for normal.
ooooooo) Pulí TR1GGERING LEVEL control for AUTO.
ppppppp) The rest of the Controls remain at any position for normal operation.
qqqqqqq) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
mmmmmm) Set the POWER switch to OFF.
nnnnnn) Tum all the three POSITION Controls to mid-position.
oooooo) Tum INTENSITY control to mid-position.
pppppp) Push the PULL 5X MAC control for normal.
qqqqqq) Pulí TR1GGERING LEVEL control for AUTO.
rrrrrr) The rest of the Controls remain at any position for normal operation.
ssssss) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrrrrrr) Set the POWER switch to OFF.
sssssss) Tum all the three POSITION Controls to mid-position.
ttttttt) Tum INTENSITY control to mid-position.
uuuuuuu) Push the PULL 5X MAC control for normal.
vvvvvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwwwwww) The rest of the Controls remain at any position for normal operation.
xxxxxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
tttttt) Set the POWER switch to OFF.
uuuuuu) Tum all the three POSITION Controls to mid-position.
vvvvvv) Tum INTENSITY control to mid-position.
wwwwww) Push the PULL 5X MAC control for normal.
xxxxxx) Pulí TR1GGERING LEVEL control for AUTO.
yyyyyy) The rest of the Controls remain at any position for normal operation.
zzzzzz) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyyyyyy) Set the POWER switch to OFF.
zzzzzzz) Tum all the three POSITION Controls to mid-position.
aaaaaaaa) Tum INTENSITY control to mid-position.
bbbbbbbb) Push the PULL 5X MAC control for normal.
cccccccc) Pulí TR1GGERING LEVEL control for AUTO.
dddddddd) The rest of the Controls remain at any position for normal operation.
eeeeeeee) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
aaaaaaa) Set the POWER switch to OFF.
bbbbbbb) Tum all the three POSITION Controls to mid-position.
ccccccc) Tum INTENSITY control to mid-position.
ddddddd) Push the PULL 5X MAC control for normal.
eeeeeee) Pulí TR1GGERING LEVEL control for AUTO.
fffffff) The rest of the Controls remain at any position for normal operation.
ggggggg) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ffffffff) Set the POWER switch to OFF.
gggggggg) Tum all the three POSITION Controls to mid-position.
hhhhhhhh) Tum INTENSITY control to mid-position.
iiiiiiii) Push the PULL 5X MAC control for normal.
jjjjjjjj) Pulí TR1GGERING LEVEL control for AUTO.
kkkkkkkk) The rest of the Controls remain at any position for normal operation.
llllllll) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
hhhhhhh) Set the POWER switch to OFF.
iiiiiii) Tum all the three POSITION Controls to mid-position.
iiiii) Set the POWER switch to OFF.
jjjjj) Tum all the three POSITION Controls to mid-position.
kkkkk) Tum INTENSITY control to mid-position.
lllll) Push the PULL 5X MAC control for normal.
mmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnn) The rest of the Controls remain at any position for normal operation.
ooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
mmmmmmmm) Set the POWER switch to OFF.
nnnnnnnn) Tum all the three POSITION Controls to mid-position.
oooooooo) Tum INTENSITY control to mid-position.
pppppppp) Push the PULL 5X MAC control for normal.
qqqqqqqq) Pulí TR1GGERING LEVEL control for AUTO.
rrrrrrrr) The rest of the Controls remain at any position for normal operation.
ssssssss) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
jjjjjjj) Set the POWER switch to OFF.
kkkkkkk) Tum all the three POSITION Controls to mid-position.
lllllll) Tum INTENSITY control to mid-position.
mmmmmmm) Push the PULL 5X MAC control for normal.
nnnnnnn) Pulí TR1GGERING LEVEL control for AUTO.
ooooooo) The rest of the Controls remain at any position for normal operation.
ppppppp) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
tttttttt) Set the POWER switch to OFF.
uuuuuuuu) Tum all the three POSITION Controls to mid-position.
vvvvvvvv) Tum INTENSITY control to mid-position.
wwwwwwww) Push the PULL 5X MAC control for normal.
xxxxxxxx) Pulí TR1GGERING LEVEL control for AUTO.
yyyyyyyy) The rest of the Controls remain at any position for normal operation.
zzzzzzzz) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
qqqqqqq) Set the POWER switch to OFF.
rrrrrrr) Tum all the three POSITION Controls to mid-position.
sssssss) Tum INTENSITY control to mid-position.
ttttttt) Push the PULL 5X MAC control for normal.
uuuuuuu) Pulí TR1GGERING LEVEL control for AUTO.
vvvvvvv) The rest of the Controls remain at any position for normal operation.
wwwwwww) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
aaaaaaaaa) Set the POWER switch to OFF.
bbbbbbbbb) Tum all the three POSITION Controls to mid-position.
ccccccccc) Tum INTENSITY control to mid-position.
ddddddddd) Push the PULL 5X MAC control for normal.
eeeeeeeee) Pulí TR1GGERING LEVEL control for AUTO.
fffffffff) The rest of the Controls remain at any position for normal operation.
ggggggggg) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
xxxxxxx) Set the POWER switch to OFF.
yyyyyyy) Tum all the three POSITION Controls to mid-position.
zzzzzzz) Tum INTENSITY control to mid-position.
aaaaaaaa) Push the PULL 5X MAC control for normal.
bbbbbbbb) Pulí TR1GGERING LEVEL control for AUTO.
cccccccc) The rest of the Controls remain at any position for normal operation.
dddddddd) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
hhhhhhhhh) Set the POWER switch to OFF.
iiiiiiiii) Tum all the three POSITION Controls to mid-position.
jjjjjjjjj) Tum INTENSITY control to mid-position.
kkkkkkkkk) Push the PULL 5X MAC control for normal.
lllllllll) Pulí TR1GGERING LEVEL control for AUTO.
mmmmmmmmm) The rest of the Controls remain at any position for normal operation.
nnnnnnnnn) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
eeeeeeee) Set the POWER switch to OFF.
ffffffff) Tum all the three POSITION Controls to mid-position.
ppppp) Set the POWER switch to OFF.
qqqqq) Tum all the three POSITION Controls to mid-position.
rrrrr) Tum INTENSITY control to mid-position.
sssss) Push the PULL 5X MAC control for normal.
ttttt) Pulí TR1GGERING LEVEL control for AUTO.
uuuuu) The rest of the Controls remain at any position for normal operation.
vvvvv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ooooooooo) Set the POWER switch to OFF.
ppppppppp) Tum all the three POSITION Controls to mid-position.
qqqqqqqqq) Tum INTENSITY control to mid-position.
rrrrrrrrr) Push the PULL 5X MAC control for normal.
sssssssss) Pulí TR1GGERING LEVEL control for AUTO.
ttttttttt) The rest of the Controls remain at any position for normal operation.
uuuuuuuuu) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
gggggggg) Set the POWER switch to OFF.
hhhhhhhh) Tum all the three POSITION Controls to mid-position.
iiiiiiii) Tum INTENSITY control to mid-position.
jjjjjjjj) Push the PULL 5X MAC control for normal.
kkkkkkkk) Pulí TR1GGERING LEVEL control for AUTO.
llllllll) The rest of the Controls remain at any position for normal operation.
mmmmmmmm) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
vvvvvvvvv) Set the POWER switch to OFF.
wwwwwwwww) Tum all the three POSITION Controls to mid-position.
xxxxxxxxx) Tum INTENSITY control to mid-position.
yyyyyyyyy) Push the PULL 5X MAC control for normal.
zzzzzzzzz) Pulí TR1GGERING LEVEL control for AUTO.
aaaaaaaaaa) The rest of the Controls remain at any position for normal operation.
bbbbbbbbbb) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
nnnnnnnn) Set the POWER switch to OFF.
oooooooo) Tum all the three POSITION Controls to mid-position.
pppppppp) Tum INTENSITY control to mid-position.
qqqqqqqq) Push the PULL 5X MAC control for normal.
rrrrrrrr) Pulí TR1GGERING LEVEL control for AUTO.
ssssssss) The rest of the Controls remain at any position for normal operation.
tttttttt) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
cccccccccc) Set the POWER switch to OFF.
dddddddddd) Tum all the three POSITION Controls to mid-position.
eeeeeeeeee) Tum INTENSITY control to mid-position.
ffffffffff) Push the PULL 5X MAC control for normal.
gggggggggg) Pulí TR1GGERING LEVEL control for AUTO.
hhhhhhhhhh) The rest of the Controls remain at any position for normal operation.
iiiiiiiiii) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
uuuuuuuu) Set the POWER switch to OFF.
vvvvvvvv) Tum all the three POSITION Controls to mid-position.
wwwwwwww) Tum INTENSITY control to mid-position.
xxxxxxxx) Push the PULL 5X MAC control for normal.
yyyyyyyy) Pulí TR1GGERING LEVEL control for AUTO.
zzzzzzzz) The rest of the Controls remain at any position for normal operation.
aaaaaaaaa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
jjjjjjjjjj) Set the POWER switch to OFF.
kkkkkkkkkk) Tum all the three POSITION Controls to mid-position.
llllllllll) Tum INTENSITY control to mid-position.
mmmmmmmmmm) Push the PULL 5X MAC control for normal.
nnnnnnnnnn) Pulí TR1GGERING LEVEL control for AUTO.
oooooooooo) The rest of the Controls remain at any position for normal operation.
pppppppppp) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bbbbbbbbb) Set the POWER switch to OFF.
ccccccccc) Tum all the three POSITION Controls to mid-position.
wwwww) Set the POWER switch to OFF.
xxxxx) Tum all the three POSITION Controls to mid-position.
yyyyy) Tum INTENSITY control to mid-position.
zzzzz) Push the PULL 5X MAC control for normal.
aaaaaa) Pulí TR1GGERING LEVEL control for AUTO.
bbbbbb) The rest of the Controls remain at any position for normal operation.
cccccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
qqqqqqqqqq) Set the POWER switch to OFF.
rrrrrrrrrr) Tum all the three POSITION Controls to mid-position.
ssssssssss) Tum INTENSITY control to mid-position.
tttttttttt) Push the PULL 5X MAC control for normal.
uuuuuuuuuu) Pulí TR1GGERING LEVEL control for AUTO.
vvvvvvvvvv) The rest of the Controls remain at any position for normal operation.
wwwwwwwwww) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ddddddddd) Set the POWER switch to OFF.
eeeeeeeee) Tum all the three POSITION Controls to mid-position.
fffffffff) Tum INTENSITY control to mid-position.
ggggggggg) Push the PULL 5X MAC control for normal.
hhhhhhhhh) Pulí TR1GGERING LEVEL control for AUTO.
iiiiiiiii) The rest of the Controls remain at any position for normal operation.
jjjjjjjjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
xxxxxxxxxx) Set the POWER switch to OFF.
yyyyyyyyyy) Tum all the three POSITION Controls to mid-position.
zzzzzzzzzz) Tum INTENSITY control to mid-position.
aaaaaaaaaaa) Push the PULL 5X MAC control for normal.
bbbbbbbbbbb) Pulí TR1GGERING LEVEL control for AUTO.
ccccccccccc) The rest of the Controls remain at any position for normal operation.
ddddddddddd) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkkkkkkkk) Set the POWER switch to OFF.
lllllllll) Tum all the three POSITION Controls to mid-position.
mmmmmmmmm) Tum INTENSITY control to mid-position.
nnnnnnnnn) Push the PULL 5X MAC control for normal.
ooooooooo) Pulí TR1GGERING LEVEL control for AUTO.
ppppppppp) The rest of the Controls remain at any position for normal operation.
qqqqqqqqq) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
eeeeeeeeeee) Set the POWER switch to OFF.
fffffffffff) Tum all the three POSITION Controls to mid-position.
ggggggggggg) Tum INTENSITY control to mid-position.
hhhhhhhhhhh) Push the PULL 5X MAC control for normal.
iiiiiiiiiii) Pulí TR1GGERING LEVEL control for AUTO.
jjjjjjjjjjj) The rest of the Controls remain at any position for normal operation.
kkkkkkkkkkk) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrrrrrrrr) Set the POWER switch to OFF.
sssssssss) Tum all the three POSITION Controls to mid-position.
ttttttttt) Tum INTENSITY control to mid-position.
uuuuuuuuu) Push the PULL 5X MAC control for normal.
vvvvvvvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwwwwwwww) The rest of the Controls remain at any position for normal operation.
xxxxxxxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
lllllllllll) Set the POWER switch to OFF.
mmmmmmmmmmm) Tum all the three POSITION Controls to mid-position.
nnnnnnnnnnn) Tum INTENSITY control to mid-position.
ooooooooooo) Push the PULL 5X MAC control for normal.
ppppppppppp) Pulí TR1GGERING LEVEL control for AUTO.
qqqqqqqqqqq) The rest of the Controls remain at any position for normal operation.
rrrrrrrrrrr) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyyyyyyyy) Set the POWER switch to OFF.
zzzzzzzzz) Tum all the three POSITION Controls to mid-position.
dddddd) Set the POWER switch to OFF.
eeeeee) Tum all the three POSITION Controls to mid-position.
ffffff) Tum INTENSITY control to mid-position.
gggggg) Push the PULL 5X MAC control for normal.
hhhhhh) Pulí TR1GGERING LEVEL control for AUTO.
iiiiii) The rest of the Controls remain at any position for normal operation.
jjjjjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
sssssssssss) Set the POWER switch to OFF.
ttttttttttt) Tum all the three POSITION Controls to mid-position.
uuuuuuuuuuu) Tum INTENSITY control to mid-position.
vvvvvvvvvvv) Push the PULL 5X MAC control for normal.
wwwwwwwwwww) Pulí TR1GGERING LEVEL control for AUTO.
xxxxxxxxxxx) The rest of the Controls remain at any position for normal operation.
yyyyyyyyyyy) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
aaaaaaaaaa) Set the POWER switch to OFF.
bbbbbbbbbb) Tum all the three POSITION Controls to mid-position.
cccccccccc) Tum INTENSITY control to mid-position.
dddddddddd) Push the PULL 5X MAC control for normal.
eeeeeeeeee) Pulí TR1GGERING LEVEL control for AUTO.
ffffffffff) The rest of the Controls remain at any position for normal operation.
gggggggggg) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
zzzzzzzzzzz) Set the POWER switch to OFF.
aaaaaaaaaaaa) Tum all the three POSITION Controls to mid-position.
bbbbbbbbbbbb) Tum INTENSITY control to mid-position.
cccccccccccc) Push the PULL 5X MAC control for normal.
dddddddddddd) Pulí TR1GGERING LEVEL control for AUTO.
eeeeeeeeeeee) The rest of the Controls remain at any position for normal operation.
ffffffffffff) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
hhhhhhhhhh) Set the POWER switch to OFF.
iiiiiiiiii) Tum all the three POSITION Controls to mid-position.
jjjjjjjjjj) Tum INTENSITY control to mid-position.
kkkkkkkkkk) Push the PULL 5X MAC control for normal.
llllllllll) Pulí TR1GGERING LEVEL control for AUTO.
mmmmmmmmmm) The rest of the Controls remain at any position for normal operation.
nnnnnnnnnn) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
gggggggggggg) Set the POWER switch to OFF.
hhhhhhhhhhhh) Tum all the three POSITION Controls to mid-position.
iiiiiiiiiiii) Tum INTENSITY control to mid-position.
jjjjjjjjjjjj) Push the PULL 5X MAC control for normal.
kkkkkkkkkkkk) Pulí TR1GGERING LEVEL control for AUTO.
llllllllllll) The rest of the Controls remain at any position for normal operation.
mmmmmmmmmmmm) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
oooooooooo) Set the POWER switch to OFF.
pppppppppp) Tum all the three POSITION Controls to mid-position.
qqqqqqqqqq) Tum INTENSITY control to mid-position.
rrrrrrrrrr) Push the PULL 5X MAC control for normal.
ssssssssss) Pulí TR1GGERING LEVEL control for AUTO.
tttttttttt) The rest of the Controls remain at any position for normal operation.
uuuuuuuuuu) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
nnnnnnnnnnnn) Set the POWER switch to OFF.
oooooooooooo) Tum all the three POSITION Controls to mid-position.
pppppppppppp) Tum INTENSITY control to mid-position.
qqqqqqqqqqqq) Push the PULL 5X MAC control for normal.
rrrrrrrrrrrr) Pulí TR1GGERING LEVEL control for AUTO.
ssssssssssss) The rest of the Controls remain at any position for normal operation.
tttttttttttt) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
vvvvvvvvvv) Set the POWER switch to OFF.
wwwwwwwwww) Tum all the three POSITION Controls to mid-position.
kkkkkk) Set the POWER switch to OFF.
llllll) Tum all the three POSITION Controls to mid-position.
mmmmmm) Tum INTENSITY control to mid-position.
nnnnnn) Push the PULL 5X MAC control for normal.
oooooo) Pulí TR1GGERING LEVEL control for AUTO.
pppppp) The rest of the Controls remain at any position for normal operation.
qqqqqq) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
uuuuuuuuuuuu) Set the POWER switch to OFF.
vvvvvvvvvvvv) Tum all the three POSITION Controls to mid-position.
wwwwwwwwwwww) Tum INTENSITY control to mid-position.
xxxxxxxxxxxx) Push the PULL 5X MAC control for normal.
yyyyyyyyyyyy) Pulí TR1GGERING LEVEL control for AUTO.
zzzzzzzzzzzz) The rest of the Controls remain at any position for normal operation.
aaaaaaaaaaaaa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
xxxxxxxxxx) Set the POWER switch to OFF.
yyyyyyyyyy) Tum all the three POSITION Controls to mid-position.
zzzzzzzzzz) Tum INTENSITY control to mid-position.
aaaaaaaaaaa) Push the PULL 5X MAC control for normal.
bbbbbbbbbbb) Pulí TR1GGERING LEVEL control for AUTO.
ccccccccccc) The rest of the Controls remain at any position for normal operation.
ddddddddddd) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bbbbbbbbbbbbb) Set the POWER switch to OFF.
ccccccccccccc) Tum all the three POSITION Controls to mid-position.
ddddddddddddd) Tum INTENSITY control to mid-position.
eeeeeeeeeeeee) Push the PULL 5X MAC control for normal.
fffffffffffff) Pulí TR1GGERING LEVEL control for AUTO.
ggggggggggggg) The rest of the Controls remain at any position for normal operation.
hhhhhhhhhhhhh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
eeeeeeeeeee) Set the POWER switch to OFF.
fffffffffff) Tum all the three POSITION Controls to mid-position.
ggggggggggg) Tum INTENSITY control to mid-position.
hhhhhhhhhhh) Push the PULL 5X MAC control for normal.
iiiiiiiiiii) Pulí TR1GGERING LEVEL control for AUTO.
jjjjjjjjjjj) The rest of the Controls remain at any position for normal operation.
kkkkkkkkkkk) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiiiiiiiiiiii) Set the POWER switch to OFF.
jjjjjjjjjjjjj) Tum all the three POSITION Controls to mid-position.
kkkkkkkkkkkkk) Tum INTENSITY control to mid-position.
lllllllllllll) Push the PULL 5X MAC control for normal.
mmmmmmmmmmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnnnnnnnnnn) The rest of the Controls remain at any position for normal operation.
ooooooooooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
lllllllllll) Set the POWER switch to OFF.
mmmmmmmmmmm) Tum all the three POSITION Controls to mid-position.
nnnnnnnnnnn) Tum INTENSITY control to mid-position.
ooooooooooo) Push the PULL 5X MAC control for normal.
ppppppppppp) Pulí TR1GGERING LEVEL control for AUTO.
qqqqqqqqqqq) The rest of the Controls remain at any position for normal operation.
rrrrrrrrrrr) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ppppppppppppp) Set the POWER switch to OFF.
qqqqqqqqqqqqq) Tum all the three POSITION Controls to mid-position.
rrrrrrrrrrrrr) Tum INTENSITY control to mid-position.
sssssssssssss) Push the PULL 5X MAC control for normal.
ttttttttttttt) Pulí TR1GGERING LEVEL control for AUTO.
uuuuuuuuuuuuu) The rest of the Controls remain at any position for normal operation.
vvvvvvvvvvvvv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
sssssssssss) Set the POWER switch to OFF.
ttttttttttt) Tum all the three POSITION Controls to mid-position.
rrrrrr) Set the POWER switch to OFF.
ssssss) Tum all the three POSITION Controls to mid-position.
tttttt) Tum INTENSITY control to mid-position.
uuuuuu) Push the PULL 5X MAC control for normal.
vvvvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwwwww) The rest of the Controls remain at any position for normal operation.
xxxxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
wwwwwwwwwwwww) Set the POWER switch to OFF.
xxxxxxxxxxxxx) Tum all the three POSITION Controls to mid-position.
yyyyyyyyyyyyy) Tum INTENSITY control to mid-position.
zzzzzzzzzzzzz) Push the PULL 5X MAC control for normal.
aaaaaaaaaaaaaa) Pulí TR1GGERING LEVEL control for AUTO.
bbbbbbbbbbbbbb) The rest of the Controls remain at any position for normal operation.
cccccccccccccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
uuuuuuuuuuu) Set the POWER switch to OFF.
vvvvvvvvvvv) Tum all the three POSITION Controls to mid-position.
wwwwwwwwwww) Tum INTENSITY control to mid-position.
xxxxxxxxxxx) Push the PULL 5X MAC control for normal.
yyyyyyyyyyy) Pulí TR1GGERING LEVEL control for AUTO.
zzzzzzzzzzz) The rest of the Controls remain at any position for normal operation.
aaaaaaaaaaaa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
dddddddddddddd) Set the POWER switch to OFF.
eeeeeeeeeeeeee) Tum all the three POSITION Controls to mid-position.
ffffffffffffff) Tum INTENSITY control to mid-position.
gggggggggggggg) Push the PULL 5X MAC control for normal.
hhhhhhhhhhhhhh) Pulí TR1GGERING LEVEL control for AUTO.
iiiiiiiiiiiiii) The rest of the Controls remain at any position for normal operation.
jjjjjjjjjjjjjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bbbbbbbbbbbb) Set the POWER switch to OFF.
cccccccccccc) Tum all the three POSITION Controls to mid-position.
dddddddddddd) Tum INTENSITY control to mid-position.
eeeeeeeeeeee) Push the PULL 5X MAC control for normal.
ffffffffffff) Pulí TR1GGERING LEVEL control for AUTO.
gggggggggggg) The rest of the Controls remain at any position for normal operation.
hhhhhhhhhhhh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkkkkkkkkkkkkk) Set the POWER switch to OFF.
llllllllllllll) Tum all the three POSITION Controls to mid-position.
mmmmmmmmmmmmmm) Tum INTENSITY control to mid-position.
nnnnnnnnnnnnnn) Push the PULL 5X MAC control for normal.
oooooooooooooo) Pulí TR1GGERING LEVEL control for AUTO.
pppppppppppppp) The rest of the Controls remain at any position for normal operation.
qqqqqqqqqqqqqq) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiiiiiiiiiii) Set the POWER switch to OFF.
jjjjjjjjjjjj) Tum all the three POSITION Controls to mid-position.
kkkkkkkkkkkk) Tum INTENSITY control to mid-position.
llllllllllll) Push the PULL 5X MAC control for normal.
mmmmmmmmmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnnnnnnnnn) The rest of the Controls remain at any position for normal operation.
oooooooooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrrrrrrrrrrrrr) Set the POWER switch to OFF.
ssssssssssssss) Tum all the three POSITION Controls to mid-position.
tttttttttttttt) Tum INTENSITY control to mid-position.
uuuuuuuuuuuuuu) Push the PULL 5X MAC control for normal.
vvvvvvvvvvvvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwwwwwwwwwwwww) The rest of the Controls remain at any position for normal operation.
xxxxxxxxxxxxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
pppppppppppp) Set the POWER switch to OFF.
qqqqqqqqqqqq) Tum all the three POSITION Controls to mid-position.
yyyyyyyyyyyyyy)
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrrrrrrrrrrr) Set the POWER switch to OFF.
ssssssssssss) Tum all the three POSITION Controls to mid-position.
tttttttttttt) Tum INTENSITY control to mid-position.
uuuuuuuuuuuu) Push the PULL 5X MAC control for normal.
vvvvvvvvvvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwwwwwwwwwww) The rest of the Controls remain at any position for normal operation.
xxxxxxxxxxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
zzzzzzzzzzzzzz) Set the POWER switch to OFF.
aaaaaaaaaaaaaaa) Tum all the three POSITION Controls to mid-position.
bbbbbbbbbbbbbbb) Tum INTENSITY control to mid-position.
ccccccccccccccc) Push the PULL 5X MAC control for normal.
ddddddddddddddd) Pulí TR1GGERING LEVEL control for AUTO.
eeeeeeeeeeeeeee) The rest of the Controls remain at any position for normal operation.
fffffffffffffff) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyyyyyyyyyyy) Set the POWER switch to OFF.
zzzzzzzzzzzz) Tum all the three POSITION Controls to mid-position.
aaaaaaaaaaaaa) Tum INTENSITY control to mid-position.
bbbbbbbbbbbbb) Push the PULL 5X MAC control for normal.
ccccccccccccc) Pulí TR1GGERING LEVEL control for AUTO.
ddddddddddddd) The rest of the Controls remain at any position for normal operation.
eeeeeeeeeeeee) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ggggggggggggggg) Set the POWER switch to OFF.
hhhhhhhhhhhhhhh) Tum all the three POSITION Controls to mid-position.
iiiiiiiiiiiiiii) Tum INTENSITY control to mid-position.
jjjjjjjjjjjjjjj) Push the PULL 5X MAC control for normal.
kkkkkkkkkkkkkkk) Pulí TR1GGERING LEVEL control for AUTO.
lllllllllllllll) The rest of the Controls remain at any position for normal operation.
mmmmmmmmmmmmmmm) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
fffffffffffff) Set the POWER switch to OFF.
ggggggggggggg) Tum all the three POSITION Controls to mid-position.
hhhhhhhhhhhhh) Tum INTENSITY control to mid-position.
iiiiiiiiiiiii) Push the PULL 5X MAC control for normal.
jjjjjjjjjjjjj) Pulí TR1GGERING LEVEL control for AUTO.
kkkkkkkkkkkkk) The rest of the Controls remain at any position for normal operation.
lllllllllllll) Check the line voltage.
mmmmmmmmmmmmm)
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
nnnnnnnnnnnnn) Set the POWER switch to OFF.
ooooooooooooo) Tum all the three POSITION Controls to mid-position.
ppppppppppppp) Tum INTENSITY control to mid-position.
qqqqqqqqqqqqq) Push the PULL 5X MAC control for normal.
rrrrrrrrrrrrr) Pulí TR1GGERING LEVEL control for AUTO.
sssssssssssss) The rest of the Controls remain at any position for normal operation.
ttttttttttttt) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
nnnnnnnnnnnnnnn) Set the POWER switch to OFF.
yyyyyy) POWER switch to OFF.
zzzzzz) Tum all the three POSITION Controls to mid-position.
aaaaaaa) Tum INTENSITY control to mid-position.
bbbbbbb) Push the PULL 5X MAC control for normal.
ccccccc) Pulí TR1GGERING LEVEL control for AUTO.
ddddddd) The rest of the Controls remain at any position for normal operation.
eeeeeee) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ooooooooooooooo) Set the POWER switch to OFF.
ppppppppppppppp) Tum all the three POSITION Controls to mid-position.
qqqqqqqqqqqqqqq) Tum INTENSITY control to mid-position.
rrrrrrrrrrrrrrr) Push the PULL 5X MAC control for normal.
sssssssssssssss) Pulí TR1GGERING LEVEL control for AUTO.
ttttttttttttttt) The rest of the Controls remain at any position for normal operation.
uuuuuuuuuuuuuuu) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
uuuuuuuuuuuuu) Set the POWER switch to OFF.
vvvvvvvvvvvvv) Tum all the three POSITION Controls to mid-position.
wwwwwwwwwwwww) Tum INTENSITY control to mid-position.
xxxxxxxxxxxxx) Push the PULL 5X MAC control for normal.
yyyyyyyyyyyyy) Pulí TR1GGERING LEVEL control for AUTO.
zzzzzzzzzzzzz) The rest of the Controls remain at any position for normal operation.
aaaaaaaaaaaaaa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
vvvvvvvvvvvvvvv) Set the POWER switch to OFF.
fffffff) POWER switch to OFF.
ggggggg) Tum all the three POSITION Controls to mid-position.
hhhhhhh) Tum INTENSITY control to mid-position.
iiiiiii) Push the PULL 5X MAC control for normal.
jjjjjjj) Pulí TR1GGERING LEVEL control for AUTO.
kkkkkkk) The rest of the Controls remain at any position for normal operation.
lllllll) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
wwwwwwwwwwwwwww) Set the POWER switch to OFF.
xxxxxxxxxxxxxxx) Tum all the three POSITION Controls to mid-position.
yyyyyyyyyyyyyyy) Tum INTENSITY control to mid-position.
zzzzzzzzzzzzzzz) Push the PULL 5X MAC control for normal.
aaaaaaaaaaaaaaaa) Pulí TR1GGERING LEVEL control for AUTO.
bbbbbbbbbbbbbbbb) The rest of the Controls remain at any position for normal operation.
cccccccccccccccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bbbbbbbbbbbbbb) Set the POWER switch to OFF.
cccccccccccccc) Tum all the three POSITION Controls to mid-position.
dddddddddddddd) Tum INTENSITY control to mid-position.
eeeeeeeeeeeeee) Push the PULL 5X MAC control for normal.
ffffffffffffff) Pulí TR1GGERING LEVEL control for AUTO.
gggggggggggggg) The rest of the Controls remain at any position for normal operation.
hhhhhhhhhhhhhh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
dddddddddddddddd) Set the POWER switch to OFF.
mmmmmmm) POWER switch to OFF.
nnnnnnn) Tum all the three POSITION Controls to mid-position.
ooooooo) Tum INTENSITY control to mid-position.
ppppppp) Push the PULL 5X MAC control for normal.
qqqqqqq) Pulí TR1GGERING LEVEL control for AUTO.
rrrrrrr) The rest of the Controls remain at any position for normal operation.
sssssss) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
eeeeeeeeeeeeeeee) Set the POWER switch to OFF.
ffffffffffffffff) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiiiiiiiiiiiii) Set the POWER switch to OFF.
jjjjjjjjjjjjjj) Tum all the three POSITION Controls to mid-position.
kkkkkkkkkkkkkk) Tum INTENSITY control to mid-position.
llllllllllllll) Push the PULL 5X MAC control for normal.
mmmmmmmmmmmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnnnnnnnnnnn) The rest of the Controls remain at any position for normal operation.
oooooooooooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
gggggggggggggggg) Set the POWER switch to OFF.
ttttttt) POWER switch to OFF.
uuuuuuu) Tum all the three POSITION Controls to mid-position.
vvvvvvv) Tum INTENSITY control to mid-position.
wwwwwww) Push the PULL 5X MAC control for normal.
xxxxxxx) Pulí TR1GGERING LEVEL control for AUTO.
yyyyyyy) The rest of the Controls remain at any position for normal operation.
zzzzzzz) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
hhhhhhhhhhhhhhhh) Set the POWER switch to OFF.
iiiiiiiiiiiiiiii) Tum all the three POSITION Controls to mid-position.
jjjjjjjjjjjjjjjj) Tum INTENSITY control to mid-position.
kkkkkkkkkkkkkkkk) Push the PULL 5X MAC control for normal.
llllllllllllllll) Pulí TR1GGERING LEVEL control for AUTO.
mmmmmmmmmmmmmmmm) The rest of the Controls remain at any position for normal operation.
nnnnnnnnnnnnnnnn) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
pppppppppppppp) Set the POWER switch to OFF.
qqqqqqqqqqqqqq) Tum all the three POSITION Controls to mid-position.
rrrrrrrrrrrrrr) Tum INTENSITY control to mid-position.
ssssssssssssss) Push the PULL 5X MAC control for normal.
tttttttttttttt) Pulí TR1GGERING LEVEL control for AUTO.
uuuuuuuuuuuuuu) The rest of the Controls remain at any position for normal operation.
vvvvvvvvvvvvvv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
oooooooooooooooo) Set the POWER switch to OFF.
aaaaaaaa) POWER switch to OFF.
bbbbbbbb) Tum all the three POSITION Controls to mid-position.
cccccccc) Tum INTENSITY control to mid-position.
dddddddd) Push the PULL 5X MAC control for normal.
eeeeeeee) Pulí TR1GGERING LEVEL control for AUTO.
ffffffff) The rest of the Controls remain at any position for normal operation.
gggggggg) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
pppppppppppppppp) Set the POWER switch to OFF.
qqqqqqqqqqqqqqqq) Tum all the three POSITION Controls to mid-position.
rrrrrrrrrrrrrrrr) Tum INTENSITY control to mid-position.
ssssssssssssssss) Push the PULL 5X MAC control for normal.
tttttttttttttttt) Pulí TR1GGERING LEVEL control for AUTO.
uuuuuuuuuuuuuuuu) The rest of the Controls remain at any position for normal operation.
vvvvvvvvvvvvvvvv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
wwwwwwwwwwwwww) Set the POWER switch to OFF.
xxxxxxxxxxxxxx) Tum all the three POSITION Controls to mid-position.
yyyyyyyyyyyyyy) Tum INTENSITY control to mid-position.
zzzzzzzzzzzzzz) Push the PULL 5X MAC control for normal.
aaaaaaaaaaaaaaa) Pulí TR1GGERING LEVEL control for AUTO.
bbbbbbbbbbbbbbb) The rest of the Controls remain at any position for normal operation.
ccccccccccccccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
wwwwwwwwwwwwwwww) Set the POWER switch to OFF.
hhhhhhhh) POWER switch to OFF.
iiiiiiii) Tum all the three POSITION Controls to mid-position.
jjjjjjjj) Tum INTENSITY control to mid-position.
kkkkkkkk) Push the PULL 5X MAC control for normal.
llllllll) Pulí TR1GGERING LEVEL control for AUTO.
mmmmmmmm) The rest of the Controls remain at any position for normal operation.
nnnnnnnn) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
xxxxxxxxxxxxxxxx) Set the POWER switch to OFF.
yyyyyyyyyyyyyyyy) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ddddddddddddddd) Set the POWER switch to OFF.
eeeeeeeeeeeeeee) Tum all the three POSITION Controls to mid-position.
fffffffffffffff) Tum INTENSITY control to mid-position.
ggggggggggggggg) Push the PULL 5X MAC control for normal.
hhhhhhhhhhhhhhh) Pulí TR1GGERING LEVEL control for AUTO.
iiiiiiiiiiiiiii) The rest of the Controls remain at any position for normal operation.
jjjjjjjjjjjjjjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
zzzzzzzzzzzzzzzz) Set the POWER switch to OFF.
oooooooo) POWER switch to OFF.
pppppppp) Tum all the three POSITION Controls to mid-position.
qqqqqqqq) Tum INTENSITY control to mid-position.
rrrrrrrr) Push the PULL 5X MAC control for normal.
ssssssss) Pulí TR1GGERING LEVEL control for AUTO.
tttttttt) The rest of the Controls remain at any position for normal operation.
uuuuuuuu) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
aaaaaaaaaaaaaaaaa) Set the POWER switch to OFF.
bbbbbbbbbbbbbbbbb) Tum all the three POSITION Controls to mid-position.
ccccccccccccccccc) Tum INTENSITY control to mid-position.
ddddddddddddddddd) Push the PULL 5X MAC control for normal.
eeeeeeeeeeeeeeeee) Pulí TR1GGERING LEVEL control for AUTO.
fffffffffffffffff) The rest of the Controls remain at any position for normal operation.
ggggggggggggggggg) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkkkkkkkkkkkkkk) Set the POWER switch to OFF.
lllllllllllllll) Tum all the three POSITION Controls to mid-position.
mmmmmmmmmmmmmmm) Tum INTENSITY control to mid-position.
nnnnnnnnnnnnnnn) Push the PULL 5X MAC control for normal.
ooooooooooooooo) Pulí TR1GGERING LEVEL control for AUTO.
ppppppppppppppp) The rest of the Controls remain at any position for normal operation.
qqqqqqqqqqqqqqq) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
hhhhhhhhhhhhhhhhh) Set the POWER switch to OFF.
vvvvvvvv) POWER switch to OFF.
wwwwwwww) Tum all the three POSITION Controls to mid-position.
xxxxxxxx) Tum INTENSITY control to mid-position.
yyyyyyyy) Push the PULL 5X MAC control for normal.
zzzzzzzz) Pulí TR1GGERING LEVEL control for AUTO.
aaaaaaaaa) The rest of the Controls remain at any position for normal operation.
bbbbbbbbb) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiiiiiiiiiiiiiiii) Set the POWER switch to OFF.
jjjjjjjjjjjjjjjjj) Tum all the three POSITION Controls to mid-position.
kkkkkkkkkkkkkkkkk) Tum INTENSITY control to mid-position.
lllllllllllllllll) Push the PULL 5X MAC control for normal.
mmmmmmmmmmmmmmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnnnnnnnnnnnnnn) The rest of the Controls remain at any position for normal operation.
ooooooooooooooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrrrrrrrrrrrrrr) Set the POWER switch to OFF.
sssssssssssssss) Tum all the three POSITION Controls to mid-position.
ttttttttttttttt) Tum INTENSITY control to mid-position.
uuuuuuuuuuuuuuu) Push the PULL 5X MAC control for normal.
vvvvvvvvvvvvvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwwwwwwwwwwwwww) The rest of the Controls remain at any position for normal operation.
xxxxxxxxxxxxxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ppppppppppppppppp) Set the POWER switch to OFF.
ccccccccc) POWER switch to OFF.
ddddddddd) Tum all the three POSITION Controls to mid-position.
eeeeeeeee) Tum INTENSITY control to mid-position.
fffffffff) Push the PULL 5X MAC control for normal.
ggggggggg) Pulí TR1GGERING LEVEL control for AUTO.
hhhhhhhhh) The rest of the Controls remain at any position for normal operation.
iiiiiiiii) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
qqqqqqqqqqqqqqqqq) Set the POWER switch to OFF.
rrrrrrrrrrrrrrrrr) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyyyyyyyyyyyyyy) Set the POWER switch to OFF.
zzzzzzzzzzzzzzz) Tum all the three POSITION Controls to mid-position.
aaaaaaaaaaaaaaaa) Tum INTENSITY control to mid-position.
bbbbbbbbbbbbbbbb) Push the PULL 5X MAC control for normal.
cccccccccccccccc) Pulí TR1GGERING LEVEL control for AUTO.
dddddddddddddddd) The rest of the Controls remain at any position for normal operation.
eeeeeeeeeeeeeeee) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
sssssssssssssssss) Set the POWER switch to OFF.
jjjjjjjjj) POWER switch to OFF.
kkkkkkkkk) Tum all the three POSITION Controls to mid-position.
lllllllll) Tum INTENSITY control to mid-position.
mmmmmmmmm) Push the PULL 5X MAC control for normal.
nnnnnnnnn) Pulí TR1GGERING LEVEL control for AUTO.
ooooooooo) The rest of the Controls remain at any position for normal operation.
ppppppppp) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ttttttttttttttttt) Set the POWER switch to OFF.
uuuuuuuuuuuuuuuuu) Tum all the three POSITION Controls to mid-position.
vvvvvvvvvvvvvvvvv) Tum INTENSITY control to mid-position.
wwwwwwwwwwwwwwwww) Push the PULL 5X MAC control for normal.
xxxxxxxxxxxxxxxxx) Pulí TR1GGERING LEVEL control for AUTO.
yyyyyyyyyyyyyyyyy) The rest of the Controls remain at any position for normal operation.
zzzzzzzzzzzzzzzzz) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ffffffffffffffff) Set the POWER switch to OFF.
gggggggggggggggg) Tum all the three POSITION Controls to mid-position.
hhhhhhhhhhhhhhhh) Tum INTENSITY control to mid-position.
iiiiiiiiiiiiiiii) Push the PULL 5X MAC control for normal.
jjjjjjjjjjjjjjjj) Pulí TR1GGERING LEVEL control for AUTO.
kkkkkkkkkkkkkkkk) The rest of the Controls remain at any position for normal operation.
llllllllllllllll) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
aaaaaaaaaaaaaaaaaa) Set the POWER switch to OFF.
qqqqqqqqq) POWER switch to OFF.
rrrrrrrrr) Tum all the three POSITION Controls to mid-position.
sssssssss) Tum INTENSITY control to mid-position.
ttttttttt) Push the PULL 5X MAC control for normal.
uuuuuuuuu) Pulí TR1GGERING LEVEL control for AUTO.
vvvvvvvvv) The rest of the Controls remain at any position for normal operation.
wwwwwwwww) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bbbbbbbbbbbbbbbbbb) Set the POWER switch to OFF.
cccccccccccccccccc) Tum all the three POSITION Controls to mid-position.
dddddddddddddddddd) Tum INTENSITY control to mid-position.
eeeeeeeeeeeeeeeeee) Push the PULL 5X MAC control for normal.
ffffffffffffffffff) Pulí TR1GGERING LEVEL control for AUTO.
gggggggggggggggggg) The rest of the Controls remain at any position for normal operation.
hhhhhhhhhhhhhhhhhh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
mmmmmmmmmmmmmmmm) Set the POWER switch to OFF.
nnnnnnnnnnnnnnnn) Tum all the three POSITION Controls to mid-position.
oooooooooooooooo) Tum INTENSITY control to mid-position.
pppppppppppppppp) Push the PULL 5X MAC control for normal.
qqqqqqqqqqqqqqqq) Pulí TR1GGERING LEVEL control for AUTO.
rrrrrrrrrrrrrrrr) The rest of the Controls remain at any position for normal operation.
ssssssssssssssss) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiiiiiiiiiiiiiiiii) Set the POWER switch to OFF.
xxxxxxxxx) POWER switch to OFF.
yyyyyyyyy) Tum all the three POSITION Controls to mid-position.
zzzzzzzzz) Tum INTENSITY control to mid-position.
aaaaaaaaaa) Push the PULL 5X MAC control for normal.
bbbbbbbbbb) Pulí TR1GGERING LEVEL control for AUTO.
cccccccccc) The rest of the Controls remain at any position for normal operation.
dddddddddd) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
jjjjjjjjjjjjjjjjjj) Set the POWER switch to OFF.
kkkkkkkkkkkkkkkkkk) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
tttttttttttttttt) Set the POWER switch to OFF.
uuuuuuuuuuuuuuuu) Tum all the three POSITION Controls to mid-position.
vvvvvvvvvvvvvvvv) Tum INTENSITY control to mid-position.
wwwwwwwwwwwwwwww) Push the PULL 5X MAC control for normal.
xxxxxxxxxxxxxxxx) Pulí TR1GGERING LEVEL control for AUTO.
yyyyyyyyyyyyyyyy) The rest of the Controls remain at any position for normal operation.
zzzzzzzzzzzzzzzz) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
llllllllllllllllll) Set the POWER switch to OFF.
eeeeeeeeee) POWER switch to OFF.
ffffffffff) Tum all the three POSITION Controls to mid-position.
gggggggggg) Tum INTENSITY control to mid-position.
hhhhhhhhhh) Push the PULL 5X MAC control for normal.
iiiiiiiiii) Pulí TR1GGERING LEVEL control for AUTO.
jjjjjjjjjj) The rest of the Controls remain at any position for normal operation.
kkkkkkkkkk) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
mmmmmmmmmmmmmmmmmm) Set the POWER switch to OFF.
nnnnnnnnnnnnnnnnnn) Tum all the three POSITION Controls to mid-position.
oooooooooooooooooo) Tum INTENSITY control to mid-position.
pppppppppppppppppp) Push the PULL 5X MAC control for normal.
qqqqqqqqqqqqqqqqqq) Pulí TR1GGERING LEVEL control for AUTO.
rrrrrrrrrrrrrrrrrr) The rest of the Controls remain at any position for normal operation.
ssssssssssssssssss) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
aaaaaaaaaaaaaaaaa) Set the POWER switch to OFF.
bbbbbbbbbbbbbbbbb) Tum all the three POSITION Controls to mid-position.
ccccccccccccccccc) Tum INTENSITY control to mid-position.
ddddddddddddddddd) Push the PULL 5X MAC control for normal.
eeeeeeeeeeeeeeeee) Pulí TR1GGERING LEVEL control for AUTO.
fffffffffffffffff) The rest of the Controls remain at any position for normal operation.
ggggggggggggggggg) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
tttttttttttttttttt) Set the POWER switch to OFF.
llllllllll) POWER switch to OFF.
mmmmmmmmmm) Tum all the three POSITION Controls to mid-position.
nnnnnnnnnn) Tum INTENSITY control to mid-position.
oooooooooo) Push the PULL 5X MAC control for normal.
pppppppppp) Pulí TR1GGERING LEVEL control for AUTO.
qqqqqqqqqq) The rest of the Controls remain at any position for normal operation.
rrrrrrrrrr) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
uuuuuuuuuuuuuuuuuu) Set the POWER switch to OFF.
vvvvvvvvvvvvvvvvvv) Tum all the three POSITION Controls to mid-position.
wwwwwwwwwwwwwwwwww) Tum INTENSITY control to mid-position.
xxxxxxxxxxxxxxxxxx) Push the PULL 5X MAC control for normal.
yyyyyyyyyyyyyyyyyy) Pulí TR1GGERING LEVEL control for AUTO.
zzzzzzzzzzzzzzzzzz) The rest of the Controls remain at any position for normal operation.
aaaaaaaaaaaaaaaaaaa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
hhhhhhhhhhhhhhhhh) Set the POWER switch to OFF.
iiiiiiiiiiiiiiiii) Tum all the three POSITION Controls to mid-position.
jjjjjjjjjjjjjjjjj) Tum INTENSITY control to mid-position.
kkkkkkkkkkkkkkkkk) Push the PULL 5X MAC control for normal.
lllllllllllllllll) Pulí TR1GGERING LEVEL control for AUTO.
mmmmmmmmmmmmmmmmm) The rest of the Controls remain at any position for normal operation.
nnnnnnnnnnnnnnnnn) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bbbbbbbbbbbbbbbbbbb) Set the POWER switch to OFF.
ssssssssss) POWER switch to OFF.
tttttttttt) Tum all the three POSITION Controls to mid-position.
uuuuuuuuuu) Tum INTENSITY control to mid-position.
vvvvvvvvvv) Push the PULL 5X MAC control for normal.
wwwwwwwwww) Pulí TR1GGERING LEVEL control for AUTO.
xxxxxxxxxx) The rest of the Controls remain at any position for normal operation.
yyyyyyyyyy) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ccccccccccccccccccc) Set the POWER switch to OFF.
ddddddddddddddddddd) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ooooooooooooooooo) Set the POWER switch to OFF.
ppppppppppppppppp) Tum all the three POSITION Controls to mid-position.
qqqqqqqqqqqqqqqqq) Tum INTENSITY control to mid-position.
rrrrrrrrrrrrrrrrr) Push the PULL 5X MAC control for normal.
sssssssssssssssss) Pulí TR1GGERING LEVEL control for AUTO.
ttttttttttttttttt) The rest of the Controls remain at any position for normal operation.
uuuuuuuuuuuuuuuuu) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
eeeeeeeeeeeeeeeeeee) Set the POWER switch to OFF.
zzzzzzzzzz) POWER switch to OFF.
aaaaaaaaaaa) Tum all the three POSITION Controls to mid-position.
bbbbbbbbbbb) Tum INTENSITY control to mid-position.
ccccccccccc) Push the PULL 5X MAC control for normal.
ddddddddddd) Pulí TR1GGERING LEVEL control for AUTO.
eeeeeeeeeee) The rest of the Controls remain at any position for normal operation.
fffffffffff) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
fffffffffffffffffff) Set the POWER switch to OFF.
ggggggggggggggggggg) Tum all the three POSITION Controls to mid-position.
hhhhhhhhhhhhhhhhhhh) Tum INTENSITY control to mid-position.
iiiiiiiiiiiiiiiiiii) Push the PULL 5X MAC control for normal.
jjjjjjjjjjjjjjjjjjj) Pulí TR1GGERING LEVEL control for AUTO.
kkkkkkkkkkkkkkkkkkk) The rest of the Controls remain at any position for normal operation.
lllllllllllllllllll) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
vvvvvvvvvvvvvvvvv) Set the POWER switch to OFF.
wwwwwwwwwwwwwwwww) Tum all the three POSITION Controls to mid-position.
xxxxxxxxxxxxxxxxx) Tum INTENSITY control to mid-position.
yyyyyyyyyyyyyyyyy) Push the PULL 5X MAC control for normal.
zzzzzzzzzzzzzzzzz) Pulí TR1GGERING LEVEL control for AUTO.
aaaaaaaaaaaaaaaaaa) The rest of the Controls remain at any position for normal operation.
bbbbbbbbbbbbbbbbbb) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
mmmmmmmmmmmmmmmmmmm) Set the POWER switch to OFF.
ggggggggggg) POWER switch to OFF.
hhhhhhhhhhh) Tum all the three POSITION Controls to mid-position.
iiiiiiiiiii) Tum INTENSITY control to mid-position.
jjjjjjjjjjj) Push the PULL 5X MAC control for normal.
kkkkkkkkkkk) Pulí TR1GGERING LEVEL control for AUTO.
lllllllllll) The rest of the Controls remain at any position for normal operation.
mmmmmmmmmmm) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
nnnnnnnnnnnnnnnnnnn) Set the POWER switch to OFF.
ooooooooooooooooooo) Tum all the three POSITION Controls to mid-position.
ppppppppppppppppppp) Tum INTENSITY control to mid-position.
qqqqqqqqqqqqqqqqqqq) Push the PULL 5X MAC control for normal.
rrrrrrrrrrrrrrrrrrr) Pulí TR1GGERING LEVEL control for AUTO.
sssssssssssssssssss) The rest of the Controls remain at any position for normal operation.
ttttttttttttttttttt) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
cccccccccccccccccc) Set the POWER switch to OFF.
dddddddddddddddddd) Tum all the three POSITION Controls to mid-position.
eeeeeeeeeeeeeeeeee) Tum INTENSITY control to mid-position.
ffffffffffffffffff) Push the PULL 5X MAC control for normal.
gggggggggggggggggg) Pulí TR1GGERING LEVEL control for AUTO.
hhhhhhhhhhhhhhhhhh) The rest of the Controls remain at any position for normal operation.
iiiiiiiiiiiiiiiiii) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
uuuuuuuuuuuuuuuuuuu) Set the POWER switch to OFF.
nnnnnnnnnnn) POWER switch to OFF.
ooooooooooo) Tum all the three POSITION Controls to mid-position.
ppppppppppp) Tum INTENSITY control to mid-position.
qqqqqqqqqqq) Push the PULL 5X MAC control for normal.
rrrrrrrrrrr) Pulí TR1GGERING LEVEL control for AUTO.
sssssssssss) The rest of the Controls remain at any position for normal operation.
ttttttttttt) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
vvvvvvvvvvvvvvvvvvv) Set the POWER switch to OFF.
wwwwwwwwwwwwwwwwwww) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
jjjjjjjjjjjjjjjjjj) Set the POWER switch to OFF.
kkkkkkkkkkkkkkkkkk) Tum all the three POSITION Controls to mid-position.
llllllllllllllllll) Tum INTENSITY control to mid-position.
mmmmmmmmmmmmmmmmmm) Push the PULL 5X MAC control for normal.
nnnnnnnnnnnnnnnnnn) Pulí TR1GGERING LEVEL control for AUTO.
oooooooooooooooooo) The rest of the Controls remain at any position for normal operation.
pppppppppppppppppp) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
xxxxxxxxxxxxxxxxxxx) Set the POWER switch to OFF.
uuuuuuuuuuu) POWER switch to OFF.
vvvvvvvvvvv) Tum all the three POSITION Controls to mid-position.
wwwwwwwwwww) Tum INTENSITY control to mid-position.
xxxxxxxxxxx) Push the PULL 5X MAC control for normal.
yyyyyyyyyyy) Pulí TR1GGERING LEVEL control for AUTO.
zzzzzzzzzzz) The rest of the Controls remain at any position for normal operation.
aaaaaaaaaaaa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyyyyyyyyyyyyyyyyyy) Set the POWER switch to OFF.
zzzzzzzzzzzzzzzzzzz) Tum all the three POSITION Controls to mid-position.
aaaaaaaaaaaaaaaaaaaa) Tum INTENSITY control to mid-position.
bbbbbbbbbbbbbbbbbbbb) Push the PULL 5X MAC control for normal.
cccccccccccccccccccc) Pulí TR1GGERING LEVEL control for AUTO.
dddddddddddddddddddd) The rest of the Controls remain at any position for normal operation.
eeeeeeeeeeeeeeeeeeee) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
qqqqqqqqqqqqqqqqqq) Set the POWER switch to OFF.
rrrrrrrrrrrrrrrrrr) Tum all the three POSITION Controls to mid-position.
ssssssssssssssssss) Tum INTENSITY control to mid-position.
tttttttttttttttttt) Push the PULL 5X MAC control for normal.
uuuuuuuuuuuuuuuuuu) Pulí TR1GGERING LEVEL control for AUTO.
vvvvvvvvvvvvvvvvvv) The rest of the Controls remain at any position for normal operation.
wwwwwwwwwwwwwwwwww) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ffffffffffffffffffff) Set the POWER switch to OFF.
bbbbbbbbbbbb) POWER switch to OFF.
cccccccccccc) Tum all the three POSITION Controls to mid-position.
dddddddddddd) Tum INTENSITY control to mid-position.
eeeeeeeeeeee) Push the PULL 5X MAC control for normal.
ffffffffffff) Pulí TR1GGERING LEVEL control for AUTO.
gggggggggggg) The rest of the Controls remain at any position for normal operation.
hhhhhhhhhhhh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
gggggggggggggggggggg) Set the POWER switch to OFF.
hhhhhhhhhhhhhhhhhhhh) Tum all the three POSITION Controls to mid-position.
iiiiiiiiiiiiiiiiiiii) Tum INTENSITY control to mid-position.
jjjjjjjjjjjjjjjjjjjj) Push the PULL 5X MAC control for normal.
kkkkkkkkkkkkkkkkkkkk) Pulí TR1GGERING LEVEL control for AUTO.
llllllllllllllllllll) The rest of the Controls remain at any position for normal operation.
mmmmmmmmmmmmmmmmmmmm) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
xxxxxxxxxxxxxxxxxx) Set the POWER switch to OFF.
yyyyyyyyyyyyyyyyyy) Tum all the three POSITION Controls to mid-position.
zzzzzzzzzzzzzzzzzz) Tum INTENSITY control to mid-position.
aaaaaaaaaaaaaaaaaaa) Push the PULL 5X MAC control for normal.
bbbbbbbbbbbbbbbbbbb) Pulí TR1GGERING LEVEL control for AUTO.
ccccccccccccccccccc) The rest of the Controls remain at any position for normal operation.
ddddddddddddddddddd) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
nnnnnnnnnnnnnnnnnnnn) Set the POWER switch to OFF.
iiiiiiiiiiii) POWER switch to OFF.
jjjjjjjjjjjj) Tum all the three POSITION Controls to mid-position.
kkkkkkkkkkkk) Tum INTENSITY control to mid-position.
llllllllllll) Push the PULL 5X MAC control for normal.
mmmmmmmmmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnnnnnnnnn) The rest of the Controls remain at any position for normal operation.
oooooooooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
oooooooooooooooooooo) Set the POWER switch to OFF.
pppppppppppppppppppp) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
eeeeeeeeeeeeeeeeeee) Set the POWER switch to OFF.
fffffffffffffffffff) Tum all the three POSITION Controls to mid-position.
ggggggggggggggggggg) Tum INTENSITY control to mid-position.
hhhhhhhhhhhhhhhhhhh) Push the PULL 5X MAC control for normal.
iiiiiiiiiiiiiiiiiii) Pulí TR1GGERING LEVEL control for AUTO.
jjjjjjjjjjjjjjjjjjj) The rest of the Controls remain at any position for normal operation.
kkkkkkkkkkkkkkkkkkk) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
qqqqqqqqqqqqqqqqqqqq) Set the POWER switch to OFF.
pppppppppppp) POWER switch to OFF.
qqqqqqqqqqqq) Tum all the three POSITION Controls to mid-position.
rrrrrrrrrrrr) Tum INTENSITY control to mid-position.
ssssssssssss) Push the PULL 5X MAC control for normal.
tttttttttttt) Pulí TR1GGERING LEVEL control for AUTO.
uuuuuuuuuuuu) The rest of the Controls remain at any position for normal operation.
vvvvvvvvvvvv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrrrrrrrrrrrrrrrrrrr) Set the POWER switch to OFF.
ssssssssssssssssssss) Tum all the three POSITION Controls to mid-position.
tttttttttttttttttttt) Tum INTENSITY control to mid-position.
uuuuuuuuuuuuuuuuuuuu) Push the PULL 5X MAC control for normal.
vvvvvvvvvvvvvvvvvvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwwwwwwwwwwwwwwwwwww) The rest of the Controls remain at any position for normal operation.
xxxxxxxxxxxxxxxxxxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
lllllllllllllllllll) Set the POWER switch to OFF.
mmmmmmmmmmmmmmmmmmm) Tum all the three POSITION Controls to mid-position.
nnnnnnnnnnnnnnnnnnn) Tum INTENSITY control to mid-position.
ooooooooooooooooooo) Push the PULL 5X MAC control for normal.
ppppppppppppppppppp) Pulí TR1GGERING LEVEL control for AUTO.
qqqqqqqqqqqqqqqqqqq) The rest of the Controls remain at any position for normal operation.
rrrrrrrrrrrrrrrrrrr) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyyyyyyyyyyyyyyyyyyy) Set the POWER switch to OFF.
wwwwwwwwwwww) POWER switch to OFF.
xxxxxxxxxxxx) Tum all the three POSITION Controls to mid-position.
yyyyyyyyyyyy) Tum INTENSITY control to mid-position.
zzzzzzzzzzzz) Push the PULL 5X MAC control for normal.
aaaaaaaaaaaaa) Pulí TR1GGERING LEVEL control for AUTO.
bbbbbbbbbbbbb) The rest of the Controls remain at any position for normal operation.
ccccccccccccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
zzzzzzzzzzzzzzzzzzzz) Set the POWER switch to OFF.
aaaaaaaaaaaaaaaaaaaaa) Tum all the three POSITION Controls to mid-position.
bbbbbbbbbbbbbbbbbbbbb) Tum INTENSITY control to mid-position.
ccccccccccccccccccccc) Push the PULL 5X MAC control for normal.
ddddddddddddddddddddd) Pulí TR1GGERING LEVEL control for AUTO.
eeeeeeeeeeeeeeeeeeeee) The rest of the Controls remain at any position for normal operation.
fffffffffffffffffffff) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
sssssssssssssssssss) Set the POWER switch to OFF.
ttttttttttttttttttt) Tum all the three POSITION Controls to mid-position.
uuuuuuuuuuuuuuuuuuu) Tum INTENSITY control to mid-position.
vvvvvvvvvvvvvvvvvvv) Push the PULL 5X MAC control for normal.
wwwwwwwwwwwwwwwwwww) Pulí TR1GGERING LEVEL control for AUTO.
xxxxxxxxxxxxxxxxxxx) The rest of the Controls remain at any position for normal operation.
yyyyyyyyyyyyyyyyyyy) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ggggggggggggggggggggg) Set the POWER switch to OFF.
ddddddddddddd) POWER switch to OFF.
eeeeeeeeeeeee) Tum all the three POSITION Controls to mid-position.
fffffffffffff) Tum INTENSITY control to mid-position.
ggggggggggggg) Push the PULL 5X MAC control for normal.
hhhhhhhhhhhhh) Pulí TR1GGERING LEVEL control for AUTO.
iiiiiiiiiiiii) The rest of the Controls remain at any position for normal operation.
jjjjjjjjjjjjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
hhhhhhhhhhhhhhhhhhhhh) Set the POWER switch to OFF.
iiiiiiiiiiiiiiiiiiiii) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
zzzzzzzzzzzzzzzzzzz) Set the POWER switch to OFF.
aaaaaaaaaaaaaaaaaaaa) Tum all the three POSITION Controls to mid-position.
bbbbbbbbbbbbbbbbbbbb) Tum INTENSITY control to mid-position.
cccccccccccccccccccc) Push the PULL 5X MAC control for normal.
dddddddddddddddddddd) Pulí TR1GGERING LEVEL control for AUTO.
eeeeeeeeeeeeeeeeeeee) The rest of the Controls remain at any position for normal operation.
ffffffffffffffffffff) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
jjjjjjjjjjjjjjjjjjjjj) Set the POWER switch to OFF.
kkkkkkkkkkkkk) POWER switch to OFF.
lllllllllllll) Tum all the three POSITION Controls to mid-position.
mmmmmmmmmmmmm) Tum INTENSITY control to mid-position.
nnnnnnnnnnnnn) Push the PULL 5X MAC control for normal.
ooooooooooooo) Pulí TR1GGERING LEVEL control for AUTO.
ppppppppppppp) The rest of the Controls remain at any position for normal operation.
qqqqqqqqqqqqq) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkkkkkkkkkkkkkkkkkkkk) Set the POWER switch to OFF.
lllllllllllllllllllll) Tum all the three POSITION Controls to mid-position.
mmmmmmmmmmmmmmmmmmmmm) Tum INTENSITY control to mid-position.
nnnnnnnnnnnnnnnnnnnnn) Push the PULL 5X MAC control for normal.
ooooooooooooooooooooo) Pulí TR1GGERING LEVEL control for AUTO.
ppppppppppppppppppppp) The rest of the Controls remain at any position for normal operation.
qqqqqqqqqqqqqqqqqqqqq) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
gggggggggggggggggggg) Set the POWER switch to OFF.
hhhhhhhhhhhhhhhhhhhh) Tum all the three POSITION Controls to mid-position.
iiiiiiiiiiiiiiiiiiii) Tum INTENSITY control to mid-position.
jjjjjjjjjjjjjjjjjjjj) Push the PULL 5X MAC control for normal.
kkkkkkkkkkkkkkkkkkkk) Pulí TR1GGERING LEVEL control for AUTO.
llllllllllllllllllll) The rest of the Controls remain at any position for normal operation.
mmmmmmmmmmmmmmmmmmmm) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrrrrrrrrrrrrrrrrrrrr) Set the POWER switch to OFF.
rrrrrrrrrrrrr) POWER switch to OFF.
sssssssssssss) Tum all the three POSITION Controls to mid-position.
ttttttttttttt) Tum INTENSITY control to mid-position.
uuuuuuuuuuuuu) Push the PULL 5X MAC control for normal.
vvvvvvvvvvvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwwwwwwwwwwww) The rest of the Controls remain at any position for normal operation.
xxxxxxxxxxxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
sssssssssssssssssssss) Set the POWER switch to OFF.
ttttttttttttttttttttt) Tum all the three POSITION Controls to mid-position.
uuuuuuuuuuuuuuuuuuuuu) Tum INTENSITY control to mid-position.
vvvvvvvvvvvvvvvvvvvvv) Push the PULL 5X MAC control for normal.
wwwwwwwwwwwwwwwwwwwww) Pulí TR1GGERING LEVEL control for AUTO.
xxxxxxxxxxxxxxxxxxxxx) The rest of the Controls remain at any position for normal operation.
yyyyyyyyyyyyyyyyyyyyy) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
nnnnnnnnnnnnnnnnnnnn) Set the POWER switch to OFF.
oooooooooooooooooooo) Tum all the three POSITION Controls to mid-position.
pppppppppppppppppppp) Tum INTENSITY control to mid-position.
qqqqqqqqqqqqqqqqqqqq) Push the PULL 5X MAC control for normal.
rrrrrrrrrrrrrrrrrrrr) Pulí TR1GGERING LEVEL control for AUTO.
ssssssssssssssssssss) The rest of the Controls remain at any position for normal operation.
tttttttttttttttttttt) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
zzzzzzzzzzzzzzzzzzzzz) Set the POWER switch to OFF.
yyyyyyyyyyyyy) POWER switch to OFF.
zzzzzzzzzzzzz) Tum all the three POSITION Controls to mid-position.
aaaaaaaaaaaaaa) Tum INTENSITY control to mid-position.
bbbbbbbbbbbbbb) Push the PULL 5X MAC control for normal.
cccccccccccccc) Pulí TR1GGERING LEVEL control for AUTO.
dddddddddddddd) The rest of the Controls remain at any position for normal operation.
eeeeeeeeeeeeee) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
aaaaaaaaaaaaaaaaaaaaaa) Set the POWER switch to OFF.
bbbbbbbbbbbbbbbbbbbbbb) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
uuuuuuuuuuuuuuuuuuuu) Set the POWER switch to OFF.
vvvvvvvvvvvvvvvvvvvv) Tum all the three POSITION Controls to mid-position.
wwwwwwwwwwwwwwwwwwww) Tum INTENSITY control to mid-position.
xxxxxxxxxxxxxxxxxxxx) Push the PULL 5X MAC control for normal.
yyyyyyyyyyyyyyyyyyyy) Pulí TR1GGERING LEVEL control for AUTO.
zzzzzzzzzzzzzzzzzzzz) The rest of the Controls remain at any position for normal operation.
aaaaaaaaaaaaaaaaaaaaa) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
cccccccccccccccccccccc) Set the POWER switch to OFF.
ffffffffffffff) POWER switch to OFF.
gggggggggggggg) Tum all the three POSITION Controls to mid-position.
hhhhhhhhhhhhhh) Tum INTENSITY control to mid-position.
iiiiiiiiiiiiii) Push the PULL 5X MAC control for normal.
jjjjjjjjjjjjjj) Pulí TR1GGERING LEVEL control for AUTO.
kkkkkkkkkkkkkk) The rest of the Controls remain at any position for normal operation.
llllllllllllll) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
dddddddddddddddddddddd) Set the POWER switch to OFF.
eeeeeeeeeeeeeeeeeeeeee) Tum all the three POSITION Controls to mid-position.
ffffffffffffffffffffff) Tum INTENSITY control to mid-position.
gggggggggggggggggggggg) Push the PULL 5X MAC control for normal.
hhhhhhhhhhhhhhhhhhhhhh) Pulí TR1GGERING LEVEL control for AUTO.
iiiiiiiiiiiiiiiiiiiiii) The rest of the Controls remain at any position for normal operation.
jjjjjjjjjjjjjjjjjjjjjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
bbbbbbbbbbbbbbbbbbbbb) Set the POWER switch to OFF.
ccccccccccccccccccccc) Tum all the three POSITION Controls to mid-position.
ddddddddddddddddddddd) Tum INTENSITY control to mid-position.
eeeeeeeeeeeeeeeeeeeee) Push the PULL 5X MAC control for normal.
fffffffffffffffffffff) Pulí TR1GGERING LEVEL control for AUTO.
ggggggggggggggggggggg) The rest of the Controls remain at any position for normal operation.
hhhhhhhhhhhhhhhhhhhhh) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkkkkkkkkkkkkkkkkkkkkk) Set the POWER switch to OFF.
mmmmmmmmmmmmmm) POWER switch to OFF.
nnnnnnnnnnnnnn) Tum all the three POSITION Controls to mid-position.
oooooooooooooo) Tum INTENSITY control to mid-position.
pppppppppppppp) Push the PULL 5X MAC control for normal.
qqqqqqqqqqqqqq) Pulí TR1GGERING LEVEL control for AUTO.
rrrrrrrrrrrrrr) The rest of the Controls remain at any position for normal operation.
ssssssssssssss) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
llllllllllllllllllllll) Set the POWER switch to OFF.
mmmmmmmmmmmmmmmmmmmmmm) Tum all the three POSITION Controls to mid-position.
nnnnnnnnnnnnnnnnnnnnnn) Tum INTENSITY control to mid-position.
oooooooooooooooooooooo) Push the PULL 5X MAC control for normal.
pppppppppppppppppppppp) Pulí TR1GGERING LEVEL control for AUTO.
qqqqqqqqqqqqqqqqqqqqqq) The rest of the Controls remain at any position for normal operation.
rrrrrrrrrrrrrrrrrrrrrr) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiiiiiiiiiiiiiiiiiiii) Set the POWER switch to OFF.
jjjjjjjjjjjjjjjjjjjjj) Tum all the three POSITION Controls to mid-position.
kkkkkkkkkkkkkkkkkkkkk) Tum INTENSITY control to mid-position.
lllllllllllllllllllll) Push the PULL 5X MAC control for normal.
mmmmmmmmmmmmmmmmmmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnnnnnnnnnnnnnnnnnn) The rest of the Controls remain at any position for normal operation.
ooooooooooooooooooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ssssssssssssssssssssss) Set the POWER switch to OFF.
tttttttttttttt) POWER switch to OFF.
uuuuuuuuuuuuuu) Tum all the three POSITION Controls to mid-position.
vvvvvvvvvvvvvv) Tum INTENSITY control to mid-position.
wwwwwwwwwwwwww) Push the PULL 5X MAC control for normal.
xxxxxxxxxxxxxx) Pulí TR1GGERING LEVEL control for AUTO.
yyyyyyyyyyyyyy) The rest of the Controls remain at any position for normal operation.
zzzzzzzzzzzzzz) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
tttttttttttttttttttttt) Set the POWER switch to OFF.
uuuuuuuuuuuuuuuuuuuuuu) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ppppppppppppppppppppp) Set the POWER switch to OFF.
qqqqqqqqqqqqqqqqqqqqq) Tum all the three POSITION Controls to mid-position.
rrrrrrrrrrrrrrrrrrrrr) Tum INTENSITY control to mid-position.
sssssssssssssssssssss) Push the PULL 5X MAC control for normal.
ttttttttttttttttttttt) Pulí TR1GGERING LEVEL control for AUTO.
uuuuuuuuuuuuuuuuuuuuu) The rest of the Controls remain at any position for normal operation.
vvvvvvvvvvvvvvvvvvvvv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
vvvvvvvvvvvvvvvvvvvvvv) Set the POWER switch to OFF.
aaaaaaaaaaaaaaa) POWER switch to OFF.
bbbbbbbbbbbbbbb) Tum all the three POSITION Controls to mid-position.
ccccccccccccccc) Tum INTENSITY control to mid-position.
ddddddddddddddd) Push the PULL 5X MAC control for normal.
eeeeeeeeeeeeeee) Pulí TR1GGERING LEVEL control for AUTO.
fffffffffffffff) The rest of the Controls remain at any position for normal operation.
ggggggggggggggg) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
wwwwwwwwwwwwwwwwwwwwww) Set the POWER switch to OFF.
xxxxxxxxxxxxxxxxxxxxxx) Tum all the three POSITION Controls to mid-position.
yyyyyyyyyyyyyyyyyyyyyy) Tum INTENSITY control to mid-position.
zzzzzzzzzzzzzzzzzzzzzz) Push the PULL 5X MAC control for normal.
aaaaaaaaaaaaaaaaaaaaaaa) Pulí TR1GGERING LEVEL control for AUTO.
bbbbbbbbbbbbbbbbbbbbbbb) The rest of the Controls remain at any position for normal operation.
ccccccccccccccccccccccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
wwwwwwwwwwwwwwwwwwwww) Set the POWER switch to OFF.
xxxxxxxxxxxxxxxxxxxxx) Tum all the three POSITION Controls to mid-position.
yyyyyyyyyyyyyyyyyyyyy) Tum INTENSITY control to mid-position.
zzzzzzzzzzzzzzzzzzzzz) Push the PULL 5X MAC control for normal.
aaaaaaaaaaaaaaaaaaaaaa) Pulí TR1GGERING LEVEL control for AUTO.
bbbbbbbbbbbbbbbbbbbbbb) The rest of the Controls remain at any position for normal operation.
cccccccccccccccccccccc) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ddddddddddddddddddddddd) Set the POWER switch to OFF.
hhhhhhhhhhhhhhh) POWER switch to OFF.
iiiiiiiiiiiiiii) Tum all the three POSITION Controls to mid-position.
jjjjjjjjjjjjjjj) Tum INTENSITY control to mid-position.
kkkkkkkkkkkkkkk) Push the PULL 5X MAC control for normal.
lllllllllllllll) Pulí TR1GGERING LEVEL control for AUTO.
mmmmmmmmmmmmmmm) The rest of the Controls remain at any position for normal operation.
nnnnnnnnnnnnnnn) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
eeeeeeeeeeeeeeeeeeeeeee) Set the POWER switch to OFF.
fffffffffffffffffffffff) Tum all the three POSITION Controls to mid-position.
ggggggggggggggggggggggg) Tum INTENSITY control to mid-position.
hhhhhhhhhhhhhhhhhhhhhhh) Push the PULL 5X MAC control for normal.
iiiiiiiiiiiiiiiiiiiiiii) Pulí TR1GGERING LEVEL control for AUTO.
jjjjjjjjjjjjjjjjjjjjjjj) The rest of the Controls remain at any position for normal operation.
kkkkkkkkkkkkkkkkkkkkkkk) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
dddddddddddddddddddddd) Set the POWER switch to OFF.
eeeeeeeeeeeeeeeeeeeeee) Tum all the three POSITION Controls to mid-position.
ffffffffffffffffffffff) Tum INTENSITY control to mid-position.
gggggggggggggggggggggg) Push the PULL 5X MAC control for normal.
hhhhhhhhhhhhhhhhhhhhhh) Pulí TR1GGERING LEVEL control for AUTO.
iiiiiiiiiiiiiiiiiiiiii) The rest of the Controls remain at any position for normal operation.
jjjjjjjjjjjjjjjjjjjjjj) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
lllllllllllllllllllllll) Set the POWER switch to OFF.
ooooooooooooooo) POWER switch to OFF.
ppppppppppppppp) Tum all the three POSITION Controls to mid-position.
qqqqqqqqqqqqqqq) Tum INTENSITY control to mid-position.
rrrrrrrrrrrrrrr) Push the PULL 5X MAC control for normal.
sssssssssssssss) Pulí TR1GGERING LEVEL control for AUTO.
ttttttttttttttt) The rest of the Controls remain at any position for normal operation.
uuuuuuuuuuuuuuu) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
mmmmmmmmmmmmmmmmmmmmmmm) Set the POWER switch to OFF.
nnnnnnnnnnnnnnnnnnnnnnn) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
kkkkkkkkkkkkkkkkkkkkkk) Set the POWER switch to OFF.
llllllllllllllllllllll) Tum all the three POSITION Controls to mid-position.
mmmmmmmmmmmmmmmmmmmmmm) Tum INTENSITY control to mid-position.
nnnnnnnnnnnnnnnnnnnnnn) Push the PULL 5X MAC control for normal.
oooooooooooooooooooooo) Pulí TR1GGERING LEVEL control for AUTO.
pppppppppppppppppppppp) The rest of the Controls remain at any position for normal operation.
qqqqqqqqqqqqqqqqqqqqqq) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ooooooooooooooooooooooo) Set the POWER switch to OFF.
vvvvvvvvvvvvvvv) POWER switch to OFF.
wwwwwwwwwwwwwww) Tum all the three POSITION Controls to mid-position.
xxxxxxxxxxxxxxx) Tum INTENSITY control to mid-position.
yyyyyyyyyyyyyyy) Push the PULL 5X MAC control for normal.
zzzzzzzzzzzzzzz) Pulí TR1GGERING LEVEL control for AUTO.
aaaaaaaaaaaaaaaa) The rest of the Controls remain at any position for normal operation.
bbbbbbbbbbbbbbbb) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ppppppppppppppppppppppp) Set the POWER switch to OFF.
qqqqqqqqqqqqqqqqqqqqqqq) Tum all the three POSITION Controls to mid-position.
rrrrrrrrrrrrrrrrrrrrrrr) Tum INTENSITY control to mid-position.
sssssssssssssssssssssss) Push the PULL 5X MAC control for normal.
ttttttttttttttttttttttt) Pulí TR1GGERING LEVEL control for AUTO.
uuuuuuuuuuuuuuuuuuuuuuu) The rest of the Controls remain at any position for normal operation.
vvvvvvvvvvvvvvvvvvvvvvv) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
rrrrrrrrrrrrrrrrrrrrrr) Set the POWER switch to OFF.
ssssssssssssssssssssss) Tum all the three POSITION Controls to mid-position.
tttttttttttttttttttttt) Tum INTENSITY control to mid-position.
uuuuuuuuuuuuuuuuuuuuuu) Push the PULL 5X MAC control for normal.
vvvvvvvvvvvvvvvvvvvvvv) Pulí TR1GGERING LEVEL control for AUTO.
wwwwwwwwwwwwwwwwwwwwww) The rest of the Controls remain at any position for normal operation.
xxxxxxxxxxxxxxxxxxxxxx) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
wwwwwwwwwwwwwwwwwwwwwww) Set the POWER switch to OFF.
cccccccccccccccc) POWER switch to OFF.
dddddddddddddddd) Tum all the three POSITION Controls to mid-position.
eeeeeeeeeeeeeeee) Tum INTENSITY control to mid-position.
ffffffffffffffff) Push the PULL 5X MAC control for normal.
gggggggggggggggg) Pulí TR1GGERING LEVEL control for AUTO.
hhhhhhhhhhhhhhhh) The rest of the Controls remain at any position for normal operation.
iiiiiiiiiiiiiiii) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
xxxxxxxxxxxxxxxxxxxxxxx) Set the POWER switch to OFF.
yyyyyyyyyyyyyyyyyyyyyyy) Tum all the three POSITION Controls to mid-position.
zzzzzzzzzzzzzzzzzzzzzzz) Tum INTENSITY control to mid-position.
aaaaaaaaaaaaaaaaaaaaaaaa) Push the PULL 5X MAC control for normal.
bbbbbbbbbbbbbbbbbbbbbbbb) Pulí TR1GGERING LEVEL control for AUTO.
cccccccccccccccccccccccc) The rest of the Controls remain at any position for normal operation.
dddddddddddddddddddddddd) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyyyyyyyyyyyyyyyyyyyyy) Set the POWER switch to OFF.
zzzzzzzzzzzzzzzzzzzzzz) Tum all the three POSITION Controls to mid-position.
aaaaaaaaaaaaaaaaaaaaaaa) Tum INTENSITY control to mid-position.
bbbbbbbbbbbbbbbbbbbbbbb) Push the PULL 5X MAC control for normal.
ccccccccccccccccccccccc) Pulí TR1GGERING LEVEL control for AUTO.
ddddddddddddddddddddddd) The rest of the Controls remain at any position for normal operation.
eeeeeeeeeeeeeeeeeeeeeee) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
eeeeeeeeeeeeeeeeeeeeeeee) Set the POWER switch to OFF.
jjjjjjjjjjjjjjjj) POWER switch to OFF.
kkkkkkkkkkkkkkkk) Tum all the three POSITION Controls to mid-position.
llllllllllllllll) Tum INTENSITY control to mid-position.
mmmmmmmmmmmmmmmm) Push the PULL 5X MAC control for normal.
nnnnnnnnnnnnnnnn) Pulí TR1GGERING LEVEL control for AUTO.
oooooooooooooooo) The rest of the Controls remain at any position for normal operation.
pppppppppppppppp) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ffffffffffffffffffffffff) Set the POWER switch to OFF.
gggggggggggggggggggggggg) Tum all the three POSITION Controls to mid-position.
Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
fffffffffffffffffffffff) Set the POWER switch to OFF.
ggggggggggggggggggggggg) Tum all the three POSITION Controls to mid-position.
hhhhhhhhhhhhhhhhhhhhhhh) Tum INTENSITY control to mid-position.
iiiiiiiiiiiiiiiiiiiiiii) Push the PULL 5X MAC control for normal.
jjjjjjjjjjjjjjjjjjjjjjj) Pulí TR1GGERING LEVEL control for AUTO.
kkkkkkkkkkkkkkkkkkkkkkk) The rest of the Controls remain at any position for normal operation.
lllllllllllllllllllllll) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
hhhhhhhhhhhhhhhhhhhhhhhh) Set the POWER switch to OFF.
qqqqqqqqqqqqqqqq) POWER switch to OFF.
rrrrrrrrrrrrrrrr) Tum all the three POSITION Controls to mid-position.
ssssssssssssssss) Tum INTENSITY control to mid-position.
tttttttttttttttt) Push the PULL 5X MAC control for normal.
uuuuuuuuuuuuuuuu) Pulí TR1GGERING LEVEL control for AUTO.
vvvvvvvvvvvvvvvv) The rest of the Controls remain at any position for normal operation.
wwwwwwwwwwwwwwww) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
iiiiiiiiiiiiiiiiiiiiiiii) Set the POWER switch to OFF.
jjjjjjjjjjjjjjjjjjjjjjjj) Tum all the three POSITION Controls to mid-position.
kkkkkkkkkkkkkkkkkkkkkkkk) Tum INTENSITY control to mid-position.
llllllllllllllllllllllll) Push the PULL 5X MAC control for normal.
mmmmmmmmmmmmmmmmmmmmmmmm) Pulí TR1GGERING LEVEL control for AUTO.
nnnnnnnnnnnnnnnnnnnnnnnn) The rest of the Controls remain at any position for normal operation.
oooooooooooooooooooooooo) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
mmmmmmmmmmmmmmmmmmmmmmm) Set the POWER switch to OFF.
nnnnnnnnnnnnnnnnnnnnnnn) Tum all the three POSITION Controls to mid-position.
ooooooooooooooooooooooo) Tum INTENSITY control to mid-position.
ppppppppppppppppppppppp) Push the PULL 5X MAC control for normal.
qqqqqqqqqqqqqqqqqqqqqqq) Pulí TR1GGERING LEVEL control for AUTO.
rrrrrrrrrrrrrrrrrrrrrrr) The rest of the Controls remain at any position for normal operation.
sssssssssssssssssssssss) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
pppppppppppppppppppppppp) Set the POWER switch to OFF.
xxxxxxxxxxxxxxxx) POWER switch to OFF.
yyyyyyyyyyyyyyyy) Tum all the three POSITION Controls to mid-position.
zzzzzzzzzzzzzzzz) Tum INTENSITY control to mid-position.
aaaaaaaaaaaaaaaaa) Push the PULL 5X MAC control for normal.
bbbbbbbbbbbbbbbbb) Pulí TR1GGERING LEVEL control for AUTO.
ccccccccccccccccc) The rest of the Controls remain at any position for normal operation.
ddddddddddddddddd) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
qqqqqqqqqqqqqqqqqqqqqqqq) Set the POWER switch to OFF.
rrrrrrrrrrrrrrrrrrrrrrrr) Tum all the three POSITION Controls to mid-position.
ssssssssssssssssssssssss) Tum INTENSITY control to mid-position.
tttttttttttttttttttttttt) Push the PULL 5X MAC control for normal.
uuuuuuuuuuuuuuuuuuuuuuuu) Pulí TR1GGERING LEVEL control for AUTO.
vvvvvvvvvvvvvvvvvvvvvvvv) The rest of the Controls remain at any position for normal operation.
wwwwwwwwwwwwwwwwwwwwwwww) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
ttttttttttttttttttttttt) Set the POWER switch to OFF.
uuuuuuuuuuuuuuuuuuuuuuu) Tum all the three POSITION Controls to mid-position.
vvvvvvvvvvvvvvvvvvvvvvv) Tum INTENSITY control to mid-position.
wwwwwwwwwwwwwwwwwwwwwww) Push the PULL 5X MAC control for normal.
xxxxxxxxxxxxxxxxxxxxxxx) Pulí TR1GGERING LEVEL control for AUTO.
yyyyyyyyyyyyyyyyyyyyyyy) The rest of the Controls remain at any position for normal operation.
zzzzzzzzzzzzzzzzzzzzzzz) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
xxxxxxxxxxxxxxxxxxxxxxxx) Set the POWER switch to OFF.
eeeeeeeeeeeeeeeee) POWER switch to OFF.
fffffffffffffffff) Tum all the three POSITION Controls to mid-position.
ggggggggggggggggg) Tum INTENSITY control to mid-position.
hhhhhhhhhhhhhhhhh) Push the PULL 5X MAC control for normal.
iiiiiiiiiiiiiiiii) Pulí TR1GGERING LEVEL control for AUTO.
jjjjjjjjjjjjjjjjj) The rest of the Controls remain at any position for normal operation.
kkkkkkkkkkkkkkkkk) Check the line voltage.
INITIAL AC OPERATION
1. Prior to any kind of operation of the instrument, proceed as follows to get familiarized with the instrument.
yyyyyyyyyyyyyyyyyyyyyyyy) Set the POWER switch to OFF.
zzzzzzzzzzzzzzzzzzzzzzzz) Tum all the three POSITION Controls to mid-position.

Вам также может понравиться