Model 3080-Series
Electrostatic Classifiers
ince 1976, TSI Electrostatic Classifiers have been
S helping aerosol scientists generate and size submi-
crometer particles. These instruments have been used Choose the right DMA for your
in a broad variety of aerosol research and have earned
a well-deserved reputation for being highly reliable and
application, or switch DMAs as
extremely versatile. your measurement needs change!
Our latest Electrostatic Classifiers – the Model
3080-series – continue that tradition, but with an ex-
panded definition of versatility. Designed as a replace-
ment for Models 3071 and 3071A, the current design In an aerosol generation system, the Electrostatic
gives you a choice of differential mobility analyzers Classifier selects airborne particles of uniform size from
(DMAs). The same controller platform works with both a polydisperse source, resulting in a highly monodis-
DMAs, greatly expanding your research possibilities! perse aerosol. With TSI’s Long DMA, you’re able to
Collectively, our Electrostatic classify particles in the range from 10 to 1000 nanome-
Classifiers produce highly monodis- ters in diameter. The Nano DMA offers classification
perse, submicrometer aerosols in the from 2 to 150 nanometers.
range from 2 to 1000 nanometers in In a particle sizing system, the Electrostatic Classi-
diameter. The Electrostatic Classifier is fier separates particles by size for high-resolution mea-
most often used as an integral part of a surements of particle size distribution. When used in
monodisperse-aerosol generation system TSI Scanning Mobility Particle Sizer (SMPS) systems,
or a submicrometer-particle sizing system. for example, monodisperse aerosol exiting the Elec-
trostatic Classifier passes to a Condensation Particle
Counter (CPC), which measures particle number con-
centration. By scanning quickly through the size range
from 3 to 1000 nanometers (size range varies by SMPS
configuration), the SMPS measures the size distribu-
tion of the aerosol precisely.
The United States Department of Commerce,
National Institute of Standards and Technology
(NIST) uses a TSI Long DMA for sizing 0.1-microme-
ter standard reference materials (Kinney et al. 1991).
In fact, no other technique provides concentration and
size-resolution measurements with such a high degree
of accuracy.
Model 3080L
REVOLUTIONARY DESIGN Order” section later in this brochure. Our new Elec-
The 3080-series Electrostatic Classifiers offer a trostatic Classifiers offer this open-platform design to
modern design consisting of two basic components: a allow you to easily interchange DMAs. The result is
controller platform and a differential mobility analyzer much greater flexibility when selecting an instrument
(DMA). The platform (Model 3080) contains a power for individual research needs.
supply, blowers, Aerosol Neutralizer, LCD display, and In addition to interchangeable differential mobili-
all instrument controls. The DMA, however, is the ty analyzers, the 3080-series instruments offer these ad-
heart of the Electrostatic Classifier. It selects particles vances over our previous design:
of a specific size out of the aerosol sample. You have a • Recirculating flow for precise match of sheath and
choice of two DMAs with the 3080-series: excess flows
• Long DMA (Model 3081) for the broadest particle • Microprocessor-controlled volumetric flow with lam-
size range, from 10 to 1000 nanometers inar flow element
• Nano DMA (Model 3085) for the best performance • Convenient front-panel design with control knob
at particle sizes below 150 nanometers and built-in display
• Electronic control of flow, voltage, particle-size, gas
The DMAs may be purchased individually, or ordered type, and instrument functions
already installed in a platform, as described in the “To
3080
3081
3085
SPECIFICATIONS
3080 Controller Platform 100
0.01
0.02
Particle 0
Dimensions (LWH): 45.7 cm × 41.4 cm × 40.6 cm
0.04
0.06
0.08
+2
0.1
Charge
0.2
(18 in. × 16.3 in. × 16 in.)
0.4
0.6
+4
0.8
1
Weight: 17.6 kg (38.9 lb) Particle Diameter (µm)
Ports
Polydisperse aerosol: 1/4-in. OD Equilibrium-charge distribution graph
Sheath air inlet/outlet: 3/8-in. OD
Bypass air inlet/outlet: 3/8-in. OD
3080N Electrostatic Classifier
Power requirements: 85 to 260 VAC, 50/60 Hz, 200 W maximum
DMA included: Nano (Model 3085)
Particle type: Solids and nonvolatile liquids
3080L Electrostatic Classifier Particle size range (generation mode): Adjustable from
DMA included: Long (Model 3081) 2 to 150 nm
Particle type: Solids and nonvolatile liquids Maximum input concentration: 108 particles/cm3 at 10 nm
Particle size range (generation mode): Adjustable from Voltage: 10 to 10,000 VDC
10 to 1000 nm Flow rate
Maximum input concentration: 108 particles/cm3 at 10 nm Sheath air: 3 to 20 L/min
Voltage: 10 to 10,000 VDC Aerosol: 0.3 to 3 L/min
Flow rate Bypass: 0 to 15 L/min
Sheath Air: 2 to 15 L/min Physical measurements
Aerosol: 0.2 to 2 L/min Nano DMA
Bypass: None Height: 20.3 cm (8 in.)
Physical measurements Outside diameter: 7.9 cm (3.1 in.) excluding ports
Long DMA Weight: 2.2 kg (4.9 lb)
Height: 61 cm (24 in.) Polydisperse aerosol inlet: 3/8-in. OD
Outside diameter: 7.6 cm (3 in.) excluding ports Monodisperse aerosol outlet: 1/4-in. OD
Weight: 5.4 kg (11.8 lb) Sheath, excess, and bypass ports: 3/8-in. OD
Monodisperse and polydisperse aerosol ports: 1/4-in. OD Complete instrument
Sheath and excess air ports: 3/8-in. OD Dimensions (LWH): 45.7 cm × 41.4 cm × 40.6 cm
Complete instrument (18 in. × 16.3 in. × 16 in.)
Dimensions (LWH): 45.7 cm × 41.4 cm × 64.3 cm Weight: 20.1 kg (44.3 lb)
(18 in. × 16.3 in. × 25.3 in.)
*TSI is authorized by the United States Nuclear Regulatory Commission to dis-
Weight: 23.2 kg (51.2 lbs) tribute these Aerosol Neutralizers. If your location is within the United States,
no other federal license is required. Check local regulations for your own protec-
tion. Neutralizers are shipped separately from other system components. End-
user name and address are required.
The design of the Model 3085 Nano DMA is covered under U.S. Patent
Number 6,230,572. Specifications are subject to change without notice. TSI
and the TSI logo are registered trademarks of TSI Incorporated.
APPLICATIONS When used in tandem: Measuring small and large
The 3080-series classifiers are primary-standard changes in particle size due to coagulation, evaporation,
aerosol instruments. That means their performance is condensation, and chemical reactions
based on basic physical principals, and their results are When used with a TSI Model 3068A Aerosol
highly repeatable. Collectively, these instruments are Electrometer: Primary size and concentration cali-
suitable for a variety of applications. bration of Condensation Particle Counters and other
When used individually: aerosol instruments
• Aerosol research, including the study of particle When used with a TSI Condensation Particle Counter:
transport, diffusion, coagulation, nucleation, and High-resolution sizing of submicrometer particles
condensation
• Particle-charge and electrical-mobility studies The availability of two DMAs allows even further
• Filter-media tests for filter efficiency specialization. The Model 3081 Long DMA is intend-
• Enhancing the monodispersity of polystyrene latex ed for wide usage as a classifier of large particles up to
(PSL) aerosols used to calibrate optical particle 1000 nanometers in diameter. The Model 3085 Nano
counters (The removal of residue particles and mul- DMA is designed for high-resolution sizing of ultrafine
tiplets from the nebulized PSL enhances monodis- particles down to 2 nanometers in diameter. Its short
persity.) length and bypass inlet flow minimizes diffusional
broadening and particle loss, which is a critical re-
Controller Platform
Kr
Control 85
Knob Aerosol
Neutralizer
Display
Polydisperse
Aerosol In
Inlet
Impactor
P
Bypass
Out
Bypass Pump
P Optional
Filter Bypass
Orifice
Filter
Filter Excess DMA
Sheath
Pump
Temperature Absolute
Heat Exchanger
Sensor Pressure
Sheath
Flowmeter Monodisperse
Aerosol Out
High Voltage
Power Supply
quirement for classifying particles smaller than 20 MODEL 3081 LONG DMA
nanometers. Remember, the DMAs are available
individually, or preinstalled in a controller platform. Sheath
air in
See “To Order” for details. Variable
high-voltage
supply
OPERATION 20-10,000
1
Volts
Polydisperse, submicrometer aerosol passes (-)
Filter
BIBLIOGRAPHY
Chen D-R, DYH Pui, D Hummes, H Fissan, FR Quant, and
High-voltage GJ Sem, Design and Evaluation of a Nanometer Aerosol Dif-
rod
ferential Mobility Analyzer (Nano-DMA), J. Aerosol Sci.
29(5):497-509 (1998).
Bypass
aerosol out Kinney PK, DYH Pui, GW Mulholland, and NP Bryner, Use
of the Electrostatic Classification Method to Size 0.1-µm
SRM Particles—a Feasibility Study, J. Res. Natl. Inst. Stand.
Technol. 96(2):147 (1991). TSI paper A74
Excess
air out
Pourprix M and J Daval, Electrostatic Precipitation of Aero-
Monodisperse
aerosol out sols on Wafers, a New Mobility Spectrometer, Proceedings of
Sheath the 3rd International Aerosol Conference 2:797-800 (1990).
air in
TO ORDER The Submicrometer Monodisperse Aerosol Generation System
includes an Electrostatic Classifier with a Long DMA. See Model
Electrostatic Classifiers and Components 3940 Submicrometer Monodisperse Aerosol Generation System product
Specify Description information sheet for additional information.
3080L Electrostatic Classifier, includes 3080 Controller Specify Description
Platform (with 3077 Aerosol Neutralizer), 3081 Long
3940 Submicrometer Monodisperse Aerosol Generation
DMA, and 1035900 Inlet Impactor
System, includes 3080L, 3012, 3062, 3076, and 3074B
3080N Electrostatic Classifier, includes 3080 Controller
Platform (with 3077 Aerosol Neutralizer), 3085 The Nanometer Monodisperse Aerosol Generation System in-
Nano DMA, and 1035900 Inlet Impactor cludes an Electrostatic Classifier with a Nano DMA. Contact your
3080 Controller Platform with 3077 Aerosol Neutralizer TSI representative for additional information.
(DMA and 1035900 Inlet Impactor not included) Specify Description
308003 Controller Platform without 3077 Aerosol Neutralizer 3942 Nanometer Monodisperse Aerosol Generation System,
(DMA and 1035900 Inlet Impactor not included) includes 3080N, 3012, 3062, 3480, and 3074B
3081 Long DMA (10 to 1000 nm)
3085 Nano DMA (2 to 150 nm) Optional Accessories
3077 Aerosol Neutralizer Specify Description
1035900 Inlet Impactor 3010 or 3010-S
Compact Condensation Particle Counter
Systems 3022A or 3022A-S
General-purpose Condensation Particle Counter
The Scanning Mobility Particle Sizer (SMPS) is available in the
following configurations. All include an Electrostatic Classifier, 3025A or 3025A-S
Aerosol Neutralizer, Condensation Particle Counter (CPC), and Ultrafine Condensation Particle Counter
SMPS software. See Model 3936-Series Scanning Mobility Particle 3012 Aerosol Neutralizer
Sizers product information sheet for additional information. 3032 Vacuum Pump
Specify Description 3062 Diffusion Dryer
3936NL25 SMPS system, includes 3080, 3081, 3085, 3025A-S, 3068A Aerosol Electrometer
software, and interconnecting hardware
3074B Filtered Air Supply
3936N25 SMPS system, includes 3080N, 3025A-S, software, and
interconnecting hardware 3076 Constant Output Atomizer
3936L25 SMPS system, includes 3080L, 3025A-S, software, and 3480 Electrospray Aerosol Generator
interconnecting hardware These optional accessories are available in several system configurations described above.
Other configurations are possible. Ask your TSI representative about recommended
3936L22 SMPS system, includes 3080L, 3022A-S, software, and Electrostatic Classifier / Condensation Particle Counter combinations.
interconnecting hardware
When ordering an Electrostatic Classifier containing an Aerosol Neutralizer, or systems
3936L10 SMPS system, includes 3080L, 3010-S, 3032, software, containing such an Electrostatic Classifier, the name and address of the end user must be
and interconnecting hardware provided to comply with nuclear regulatory requirements.
P/N 1930097 Rev. A Copyright © 2001 TSI Incorporated Printed in U.S.A. (082001)