PHYSICS REVIEW
tractography, including optimizing the b-values, the number uncooperative patients, in whom all other sequences are too
and orientations of diffusion-weighted acquisitions, as well as motion-degraded to be useful.
the fiber tracking parameters. The undesirable effects of com- Shortcomings of SS-EPI DWI. However, limitations of SS-
mon problems such as low signal intensity–to-noise ratio EPI include low spatial resolution, blurring effects of T2* de-
(SNR) and pulsation artifact are reviewed. An overview is pro- cay and T2 decay occurring during image readout, and sensi-
vided of current methods for analyzing quantitative DTI data tivity to artifacts due to Nyquist ghosting, chemical shift,
for clinical research, including the reproducibility of DTI mea- magnetic field inhomogeneity, and local susceptibility effects.
surements. Throughout this review, the emphasis is on opti- With current MR imaging hardware and software limitations,
mizing the many technical factors needed for state-of-the-art the single-shot technique limits matrix size for a typical 2D
DWI and DTI based on the best available evidence in the DWI to 128 ⫻ 128, which is much less than that for standard
literature. T1- and T2-weighted scans, which can have matrix sizes of
256 ⫻ 192 or greater. Blurring of T2-weighted and T2*-
Technical Considerations for State-of-the-Art DWI weighted contrast is another shortcoming of the single-shot
approach; this also occurs in single-shot fast spin-echo (SS-
Echo-Planar DWI FSE) and half-Fourier acquired single-shot turbo spin-echo
Advantages of Single-Shot Echo-Planar DWI. Because (HASTE) sequences due to their extended echo-train lengths.
even minimal bulk patient motion during acquisition of DWIs The Nyquist ghost is an artifact that is specific to EPI, because
can obscure the effects of the much smaller microscopic water it results from the fact that EPI traverses k-space in opposite
motion due to diffusion, ultrafast imaging sequences are nec- directions on alternate echoes. Errors in the phase of the MR
essary for successful clinical DWI. Most commonly, diffusion imaging signal intensity from sources such as fat-water chem-
imaging is performed by using spin-echo single-shot echo- ical shift can cause a second “ghost” image to be overlaid on
planar imaging (SS-EPI) techniques. The term “single shot” the original image, where the ghost image is shifted by one half
means that an entire 2D image is formed from a single radio- of the FOV in the phase-encoding direction. Because of the
frequency excitation pulse. Images can be acquired in a frac- Nyquist ghost phenomenon, chemical shift can be particularly
tion of a second; therefore, artifact from physiologic cardiac troublesome in DWI. Fortunately for head imaging, fat-con-
and respiratory pulsatility and from patient motion is greatly taining regions are usually limited to the scalp, the orbits, and
the bone marrow of the calvarium, including the skull base.
From the Department of Radiology, University of California, San Francisco, San Francisco,
Calif.
These artifacts are minimized by frequency-selective fat-satu-
ration pulses incorporated into the SS-EPI sequence. Perhaps
Please address correspondence to Pratik Mukherjee, MD, PhD, Department of Radiology,
Box 0628, University of California, San Francisco, 505 Parnassus Ave, L-358, San Francisco, the worst artifacts inherent to SS-EPI, especially at high fields
CA 94143-0628; e-mail: pratik@radiology.ucsf.edu of 3T or greater, are those from magnetic field inhomogene-
DOI 10.3174/ajnr.A1052 ities, primarily caused by local susceptibility differences be-
tween adjacent structures. These often result in marked dis- skeletal muscle contractions. Taller patients would be more
tortion and signal intensity dropout near air-filled cavities likely to be affected due to the longer length of their peripheral
such as the paranasal and mastoid sinuses, particularly at the nerves. All clinical MR imaging scanners meet FDA safety
skull base and the posterior fossa, limiting the sensitivity of guidelines for dB/dt limits. The current generation of MR im-
DWI with SS-EPI in these areas. agers with 40 – 80 mT/m maximal gradient amplitude and
Optimizing SS-EPI DWI: b-Values. As reviewed previ- 150 –200 mT/m per millisecond maximal slew rate enables
ously (equation 5, Part 11), for clinical b-values in the range of DWI with better anatomic fidelity than older MR imaging
0 –1000 s/mm2, only 2 different b-values need to be measured systems. Some newer MR imaging scanners are equipped with
to estimate ADC: 1 at a very low b-value (or zero) and the even stronger and faster gradients that have a reduced FOV, to
other at a high b-value. Acquisition of DWIs at additional stay within dB/dt guidelines, and are suitable for head imaging
intermediate values of b is redundant. A high b-value of 1000 but not for spine or body imaging. Besides maximal amplitude
s/mm2 has become the standard for clinical DWI. The brains and slew rate, another important factor is the gradient duty
of neonates and infants have much longer T2 relaxation times cycle, which can be limited by heating constraints. Larger duty
and much higher ADC values than those of adults2-5; there- cycles permit more 2D diffusion-weighted sections to be ac-
fore, it is common practice to use less diffusion-weighting, for quired in a given TR. Modern MR imaging systems have wa-
example, b ⫽ 600 s/mm2 for premature neonates and b ⫽ 700 ter-cooled gradients with larger duty cycles than older sys-
s/mm2 for term neonates and infants younger than 1 year of tems, which allow faster DWI.
age. A simple rule of thumb is that the optimal b-value multi- As gradients become more powerful, they may exacerbate
plied by the ADC of the tissue under investigation should be problems such as eddy currents and mechanical vibration. All
close to 1. MR imaging gradient coils are self-shielded to prevent eddy
Optimizing SS-EPI DWI: Gradients. One of the most im- currents, which are residual magnetic fields induced by gradi-
portant hardware factors influencing the quality of DWI is the ent switching that persist after the gradients are turned off.
gradient performance of the MR imaging scanner, for both the However, self-shielding is inadequate for the large amplitude
diffusion gradients and the EPI readout gradients. Stronger and rapid onset and offset of diffusion-sensitizing gradients at
and faster gradients enable stronger diffusion-weighting in a high b-values. Eddy currents cause 3 different types of image
shorter period of time as well as reducing the time required to artifacts in DWI: scaling, shift, and shear.6 “Scaling” refers to
form an EPI image. This permits DWIs to be acquired at a expansion or contraction of the DWI. “Shift” describes dis-
shorter TE, which improves SNR and reduces geometric warp- placement of the image along the phase-encoding direction
ing artifacts from susceptibility effects. Gradient strength is (Fig 1). “Shear” denotes shifting of the image in opposite di-
often measured in milliteslas per meter. The switching speed rections on the left and right sides. Therefore, additional eddy
of gradients is referred to as “slew rate” and is measured in current compensation strategies are needed for DWI. Most
milliteslas per meter per millisecond. Higher gradient ampli- DWI sequences now use bipolar diffusion gradients, which
tudes and slew rates are desirable for DWI; however, there are have positive and negative lobes to cancel eddy currents.7 An-
US Food and Drug Administration (FDA) limits on the max- other option available on the current generation of MR imag-
imal rate at which the magnetic field can be changed, referred ing scanners is the twice-refocused spin-echo (TRSE) or dual
to technically as dB/dt. Gradient performance that exceeds the spin-echo (DSE) diffusion-weighted sequences, which use 2
federal dB/dt guidelines risks peripheral nerve stimulation due consecutive radio-frequency refocusing pulses, each with a
to induced electric currents, which can lead to involuntary pair of bipolar diffusion gradients to further break up the time
construction errors might occur with GRAPPA, which may be plex, and their joint effect on the quality of DWI and DTI has
less predictable. been explored in detail by Jaermann et al,15 who recom-
Another method for reducing the echo-train length of SS- mended an optimal acceleration factor of approximately 2 in
EPI is through partial-Fourier encoding of k-space. Due to combination with 60% partial-Fourier encoding for 3T acqui-
conjugate symmetry, only half of k-space actually needs to be sitions at b ⫽ 1000 s/mm2.
measured, and the other half can be inferred. This cuts the Alternatives to SS-EPI DWI. Other pulse sequences have
SS-EPI echo train in half, entailing the same benefits as parallel also been applied to diffusion imaging, including variants of
imaging with an acceleration factor of 2. The interaction be- fast spin-echo (FSE) or turbo spin-echo imaging, multishot
tween partial-Fourier encoding and parallel imaging is com- EPI, spiral imaging, and line-scanning methods. Single-shot
methods other than SS-EPI, such as SS-FSE or HASTE, can be efficiency compared with SS-EPI; for this purpose, GRAPPA
used to perform DWI. Because these are also ultrafast se- was found to produce fewer off-resonance and motion arti-
quences, they share the relative immunity to bulk patient mo- facts than SENSE.30
tion like SS-EPI, but they do not have nearly as much suscep-
tibility or chemical shift artifacts. For this reason, they may Technical Considerations for State-of-the-Art DTI
represent a good alternative to SS-EPI in regions in which All of the factors reviewed previously for optimizing DWI also
susceptibility or chemical shift effects are particularly pro- apply for optimizing DTI acquisitions. However, there are a
found, such as the spine or neck.22,23 However, SS-FSE and number of additional considerations that are specific to DTI.
HASTE have not proved popular for brain DWI because of Because diffusion imaging is an SNR-limited technique and
their low SNR per unit of time compared with SS-EPI, result- DTI measures such as fractional anisotropy (FA) and relative
ing in longer scanning times. anisotropy (RA) are more affected by measurement noise than
Multishot methods also have much reduced susceptibility DWI measures such as ADC, the ability to acquire DTI with
artifacts compared with SS-EPI. However, they are not as fast adequate SNR for clinical applications is limited by time con-
as single-shot methods, and this also makes them intrinsically straints. Thus, optimizing image acquisition parameters is an
more sensitive to artifacts arising from bulk motion during essential step for producing high-quality DTI. There is no
image acquisition. Motion artifacts can be ameliorated by the fixed set of parameters optimal for every application; optimi-
use of navigator echoes, especially in combination with car- zation depends on the MR imaging hardware configuration,
diac and respiratory gating. One increasingly popular tech- available scanning time, anatomic coverage needed, and spe-
nique for DWI is a multishot FSE sequence called periodically cific anatomic structures to be investigated. For instance,
rotated overlapping parallel lines with enhanced reconstruc- when studying less compliant subjects such as children, min-
tion (PROPELLER), which continually oversamples the cen- imizing the scanning time is crucial, whereas high spatial res-
ter of k-space (ie, self-navigation) to mitigate motion artifacts olution (leading to long scanning times) is essential for delin-
without the need for gating.24 This has been shown to improve eating relatively small white matter tracts. Factors to consider
detection of small acute infarcts, especially at the skull base for application-specific optimization of DTI are emphasized
and in the posterior fossa, where SS-EPI has the greatest sus- below.
ceptibility-induced distortions.25,26 However, PROPELLER
has not overtaken SS-EPI for routine brain DWI, likely be- Optimizing Clinical DTI
cause of its much longer scanning times. Optimizing the Number of Diffusion-Encoding Direc-
Further improvements in speed, such as in the more re- tions. To estimate the diffusion tensor, one needs DWIs with
cently developed Turboprop sequence,27 may continue to nar- high b-values along at least 6 noncollinear directions in addi-
row this gap. Moreover, a similar self-navigated k-space tra- tion to a low-b DWI or a T2-weighted (b ⫽ 0 s/mm2) image.
jectory can be applied to multishot EPI to yield the desired However, for most applications, many more images are usu-
combination of high SNR per unit of time and reduced sus- ally required to boost SNR to acceptable levels. It has been a
ceptibility artifacts. This new technique is called PROPELLER common practice simply to repeat acquisition of the same
EPI28 and can be further enhanced by parallel imaging.29 Al- DWIs (ie, increasing NEX) to achieve this. However, acquiring
ternatively, parallel imaging can be directly incorporated into more distinct diffusion-encoding directions without any re-
multishot EPI for improved-quality DWI without loss of SNR peated acquisitions is becoming more widespread. There has