124 5.4 Inspecting the Prober before Installation 0057.

1
You should inspect the prober before installation to check for and assess any damage that may have occurred
TEL P-8XL/XLn Installation/Startup Manual Rev.8.0

during transport.

The following list describes the checks that should be performed before installation. The prober dimensions
are provided for reference.

• Check that there is approximately 1 m of space surrounding the prober on all sides. This space is
necessary to perform maintenance on the prober.
• Check that the floor is strong enough to support the weight of the prober.
• If the prober you are installing is equipped with an optional auto HF unit, check the specifications
of the auto HF when it is in the fully retracted (180°) position.

Installation Space

Prober Dimensions

Prober Unit Prober Dimensions

Main Unit Dimensions 1090 mm (W) x 1105 mm (D) x 930 mm (H)


The height when the signal tower is attached is 1831.5 mm.

Main Unit Weight Approx. 785 kg (Standard)

Dimensions when auto HF (optional) Approx. 1321 mm (W) x approx. 1492 mm (D) (TEL standard model)
is fully retracted (180°). These dimensions vary according to the specifications. Check the
auto HF specifications manual.

Auto HF Weight Approx. 350 kg (hinge and frame)


These dimensions vary according to the specifications. Check the
Auto HF specifications manual.