Service Manual

HyLED Series
Surgical Lights
Copyright
Intellectual Property Statement

NANJING MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Nanjing


Mindray) owns the intellectual property rights to this product and this manual. This manual
may refer to information protected by copyrights or patents and does not convey any license
under the patent rights of Nanjing Mindray, nor the rights of others. Nanjing Mindray does not
assume any liability arising out of any infringements of patents or other rights of third parties.

, , are the registered trademarks or trademarks owned by


Shenzhen Mindray in China and other countries. Nanjing Mindray is authorized by Shenzhen
Mindray to use the trademarks.

© 2014-2015 Nanjing Mindray Bio-Medical Electronics Co., Ltd. All rights Reserved.

I
This page is intentionally blank.
Table of Contents
1 System Overview ································································ 1-1
1.1 Main Components ........................................................................ 1-1
1.2 Safety .......................................................................................... 1-3
1.2.1 Safety Notes ....................................................................... 1-3
1.2.2 Safety Instructions ............................................................... 1-4

2 General Maintenance··························································· 2-1


2.1 Daily Checks by User ................................................................... 2-1
2.2 Monthly Checks ........................................................................... 2-3
2.3 Annual Checks ............................................................................. 2-4
2.4 Servicing Checklist ....................................................................... 2-6

3 FRU List and Diagram ························································· 3-1


3.1 Flange Tube ................................................................................. 3-1
3.1.1 FRU (Field Repairable Unit) List .......................................... 3-1
3.1.2 Typical Diagram .................................................................. 3-2
3.2 Swivel Arm Assembly ................................................................... 3-6
3.2.1 FRU List .............................................................................. 3-6
3.2.2 Typical Diagram .................................................................. 3-8
3.3 Spring Arm Assembly ................................................................. 3-12
3.3.1 FRU List ............................................................................ 3-12
3.3.2 Typical Diagram ................................................................ 3-13
3.4 Gimbal Joint Assembly ............................................................... 3-18
3.4.1 FRU List ............................................................................ 3-18
3.4.2 Typical Drawing ................................................................. 3-19
3.5 Control Panel ............................................................................. 3-25
3.5.1 FRU List ............................................................................ 3-25
3.5.2 Typical Diagram ................................................................ 3-27
3.6 Lighthead ................................................................................... 3-31
3.6.1 FRU List ............................................................................ 3-31
3.6.2 Typical Diagram ................................................................ 3-32
3.7 Sterilizable Handle ..................................................................... 3-42
3.7.1 FRU List ............................................................................ 3-42
3.7.2 Typical Diagram ................................................................ 3-42
3.8 Backup Battery Module .............................................................. 3-47
3.8.1 FRU List ............................................................................ 3-47
3.8.2 Typical Diagram ................................................................ 3-47
3.9 Integrated Camera ..................................................................... 3-49
3.9.1 FRU List ............................................................................ 3-49
3.9.2 Typical Diagram ................................................................ 3-50

1
Table of Contents

3.10 Carrier-arm Camera ................................................................... 3-52


3.10.1 FRU List .......................................................................... 3-52
3.10.2 Typical Diagram ............................................................... 3-52
3.11 Swivel Arm for Display................................................................ 3-53
3.11.1 FRU List .......................................................................... 3-53
3.11.2 Typical Diagram ............................................................... 3-53
3.12 Ceiling Cover ............................................................................. 3-57
3.12.1 FRU List .......................................................................... 3-57
3.12.2 Typical Diagram ............................................................... 3-57
3.13 Wall Control ............................................................................... 3-60
3.13.1 FRU List .......................................................................... 3-60
3.13.2 Typical Diagram ............................................................... 3-60
3.14 LED Mobile Surgical Light .......................................................... 3-62
3.14.1 FRU List .......................................................................... 3-62
3.14.2 Typical Diagram ............................................................... 3-62
3.15 Other FRUs................................................................................ 3-65

4 Replacement ······································································ 4-1


4.1 Tools ............................................................................................ 4-1
4.2 HyLED 9700/9500/9500M Lighthead Replacement....................... 4-2
4.2.1 Replacing Sterilizable Handle .............................................. 4-2
4.2.2 Replacing Encircling Handles .............................................. 4-3
4.2.3 Replacing Glass .................................................................. 4-3
4.2.4 Replacing AICS Sensor ....................................................... 4-6
4.2.5 Replacing LED Unit ............................................................. 4-8
4.2.6 Replacing LED Drive Board ................................................. 4-9
4.3 HyLED 8600/8600M Lighthead Replacement ............................. 4-10
4.3.1 Replacing Sterilizable Handle ............................................ 4-10
4.3.2 Replacing Encircling Handles ............................................ 4-11
4.3.3 Replacing Glass ................................................................ 4-12
4.3.4 Replacing LED Unit ........................................................... 4-14
4.3.5 Replacing LED Drive Board ............................................... 4-16
4.4 HyLED 760/760M/730/730M Lighthead Replacement ................. 4-19
4.4.1 Replacing Sterilizable Handle ............................................ 4-19
4.4.2 Replacing Outer Handle .................................................... 4-20
4.4.3 Replacing Glass ................................................................ 4-21
4.4.4 Replacing LED Unit ........................................................... 4-23
4.4.5 Replacing LED Drive Board ............................................... 4-25
4.5 Replacing Control Panel ............................................................. 4-26
4.5.1 Replacing Front Cover Assembly ....................................... 4-26
4.5.2 Replacing Control Board.................................................... 4-27
4.5.3 Internal Wiring of Control Panel ......................................... 4-28
4.6 Replacing Touch Control Panel................................................... 4-30
4.6.1 Replacing Front Cover Assembly ....................................... 4-30

2
Table of Contents

4.6.2 Replacing Control Board.................................................... 4-31


4.6.3 Replacing Touch Screen .................................................... 4-33
4.6.4 Internal Wiring of Touch Control Panel ............................... 4-34
4.7 Replacing Gimbal Joint .............................................................. 4-35
4.7.1 Replacing Vertical Gimbal Joint and Horizontal Gimbal Joint4-35
4.7.2 Replacing Vertical Gimbal Joint and Horizontal Gimbal Joint
(for Lighthead with Integrated Camera) ...................................... 4-38
4.7.3 Replacing Gimbal Joint (LCH) ........................................... 4-39
4.7.4 Replacing Gimbal Joint for Lighthead with Integrated Camera
(LCH)............... .......................................................................... 4-43
4.8 Replacing Spring Arm ................................................................. 4-45
4.8.1 Replacing Standard/LCH Spring Arm ................................. 4-45
4.8.2 Replacing Standard/LCH Wired Spring Arm ....................... 4-46
4.8.3 Replacing Bearing Seat of Spring Arm ............................... 4-47
4.9 Replacing Swivel Arm Assembly................................................. 4-49
4.10 Replacing Terminal Block/Power Module/Backup Power Module
(optional)/Video Power Module (optional)/Video Box (optional)/Remote
Control Module (optional) ..................................................................... 4-51
4.11 Replacing Backup Battery Module .............................................. 4-53
4.11.1 Replacing Charging Board ............................................... 4-53
4.11.2 Replacing Battery ............................................................ 4-54
4.12 Replacing Infrared Remote Control Board .................................. 4-55
4.13 Replacing Control Board of Carrier-arm Camera ........................ 4-56
4.14 Replacing Wall Control Board..................................................... 4-58
4.15 Replacing Batteries of Mobile Surgical Light ............................... 4-61
4.15.1 Replacing Battery ............................................................ 4-61
4.15.2 Replacing Power Board ................................................... 4-63

5 Calibration········································································· 5-1
5.1 Control Panel Calibration.............................................................. 5-1
5.1.1 Lighthead Type and Software Illuminance Level Calibration
(HyLED 8600/8600M/9500/9700/9500M Lightheads).................... 5-1
5.1.2 Lighthead Type and Software Illuminance Level Calibration
(HyLED 730/730M/760/760M Lightheads) .................................... 5-5
5.1.3 Lighthead Type and Software Illuminance Level Calibration
(HyLED 8600 New Lightheads) .................................................... 5-8
5.1.4 Camera Type Calibration (HyLED 8600 New) .................... 5-11
5.2 Touch Screen Calibration ........................................................... 5-13
5.3 Touch Control Panel Calibration (New Hardware) ....................... 5-15
5.3.1 Camera Type Calibration (HyLED 730/760/9500/9700) ...... 5-15
5.3.2 Lighthead Number and Light System Number Calibration
(HyLED 730/760//9500/9700) ..................................................... 5-17
5.3.3 Software Illuminance Level Calibration (HyLED
730/730M/760/760M /9500/9700/9500M) ................................... 5-19

3
Table of Contents

5.4 Touch Control Panel Calibration (New User Interface) ................ 5-23
5.4.1 Camera Type Calibration ................................................... 5-23
5.4.2 Lighthead Number and Light System Number Calibration .. 5-24
5.4.3 Software Illuminance Level Calibration .............................. 5-27
5.5 Carrier-arm Camera Calibration.................................................. 5-30

6 Lighthead Internal Wiring ····················································· 6-1


6.1 LED Unit Layout and Wiring ......................................................... 6-1
6.1.1 HyLED 9500/9500M LED Unit Layout and Wiring ................ 6-1
6.1.2 HyLED 9700 LED Unit Layout and Wiring ............................ 6-2
6.1.3 HyLED 9500/9500M LED Unit Layout and Wiring (Color
Temperature Adjustment) (Old Hardware)..................................... 6-3
6.1.4 HyLED 9700 LED Unit Layout and Wiring (Color Temperature
Adjustment) (Old Hardware) ......................................................... 6-4
6.1.5 HyLED 9500/9500M LED Unit Layout and Wiring (Color
Temperature Adjustment) (New Hardware) ................................... 6-5
6.1.6 HyLED 9700 LED Unit Layout and Wiring (Color Temperature
Adjustment) (New Hardware) ....................................................... 6-6
6.1.7 HyLED 8600 LED Unit Layout and Wiring ............................ 6-7
6.1.8 HyLED 8600 (New) LED Unit Layout and Wiring.................. 6-8
6.1.9 HyLED 730/730M LED Unit Layout and Wiring .................... 6-9
6.1.10 HyLED 760/760M LED Unit Layout and Wiring ................ 6-10
6.2 LED Drive Board Wiring ............................................................. 6-11
6.2.1 HyLED 9500/9700/9500M Drive Board Wiring (Old
Hardware)................................................................... .................6-11
6.2.2 HyLED 9500/9700/9500M Drive Board Wiring (New
Hardware)................ .................................................................. 6-13
6.2.3 HyLED 8600/8600M Drive Board Wiring ............................ 6-13
6.2.4 HyLED 8600 (New) Drive Board Wiring ............................. 6-15
6.2.5 HyLED 730/730M/760/760M Drive Board Wiring ............... 6-16

7 Troubleshooting ································································· 7-1


7.1 Electrical Schematic Diagram ....................................................... 7-1
7.2 Common Errors............................................................................ 7-2
7.2.1 Lighthead Off and Abnormal Lighting ................................... 7-2
7.2.2 Broken LEDs ....................................................................... 7-7
7.2.3 LIN Communication Error .................................................... 7-7
7.2.4 Alarm of Control Panel/Touch Control Panel ...................... 7-10
7.2.5 Video System Error ........................................................... 7-13

8 Communication Protocol ····················································· 8-1


8.1 General ........................................................................................ 8-1
8.2 Selecting Communication Protocol ............................................... 8-2
8.2.1 Old User Interface of Touch Control Panel ........................... 8-2

4
Table of Contents

8.2.2 New User Interface of Touch Control Panel.......................... 8-5

9 Appendices ······································································· 9-1


A Differences between Old and New Hardware ............................... 9-1
B Electrical Connection Diagram ..................................................... 9-2
B.1 Terminal Block Connection .................................................. 9-2
B.2 HyLED 730/760/8600(New)/9500/9700 Series, Mains
Connection.................. ................................................................. 9-3
B.3 HyLED 730/760/8600(New)/9500/9700 Series with Mindray’s
Backup Power Module ................................................................. 9-4
B.4 HyLED 730/760/8600(New)/9500/9700 Series with Integrated
Camera........................................................................................ 9-5
B.5 HyLED 730/760/8600(New)/9500/9700 Series with Carrier-arm
Camera........................................................................................ 9-6
B.6 HyLED 8600(New)/9500/9700 Series with SYNC Function .. 9-7
B.7 HyLED 8600/9500/9700 (Old Hardware) Series, Mains
Connection............. ...................................................................... 9-8
B.8 HyLED 8600/9500/9700 (Old Hardware) Series with Mindray’s
Backup Power Module ................................................................. 9-9
B.9 HyLED 8600/9500/9700 (Old Hardware) Series with Carrier-arm
Camera...................................................................................... 9-10

5
This page is intentionally blank.
1 System Overview

1.1 Main Components

1
2
3

5
8 7
6

9
10

Figure 1-1 Surgical light main components

1 Flange tube 2 Ceiling cover


3 Swivel arm assembly 4 Standard spring arm
5 Vertical gimbal joint 6 Horizontal gimbal joint
7 Lighthead 8 Lighthead
9 Carrier-arm camera (optional) 10 Display (optional)

1-1
System Overview

1
2 5
6
3 4

Figure 1-2 Surgical light for low ceiling height

1 Flange tube 2 Ceiling cover


3 Swivel arm assembly 4 LCH (Low Ceiling Height) spring arm (optional)
5 Gimbal joint 6 Lighthead
7 Lighthead

Figure 1-3 LED mobile surgical light

1 Lighthead 2 Spring arm


3 Pole 4 Handle
5 Power module 6 Castor
7 Base 8 Gimbal joint

1-2
System Overview

1.2 Safety

1.2.1 Safety Notes

Table 1-1 Definition of safety notes

Note Meaning

Read the statement below the symbol. The statement alerts you
to an operating hazard that can cause personal injury.

Read the statement below the symbol. The statement alerts you
to possible damage to the device or other property.

Read the statement below the symbol. The statement alerts you
to information that requires your attention.

1-3
System Overview

1.2.2 Safety Instructions

 The surgical light is to be serviced by personnel authorized by Mindray only.


Non-authorized remove will compromise product safety and effectiveness
and invalidate the warranty.
 Before maintaining, servicing or removing the surgical light, make sure the
mains supply has been disconnected from the light.
 Use spare parts and accessories specified by the manufacture only.
Otherwise, safety and effectiveness of the light may be compromised.

 Exercise caution when transporting or moving the device. Do not damage


the device.
 The device is to be maintained or serviced regularly by personnel
authorized by Nanjing Mindray.
 During servicing, make sure that the surgical light is well protected from
mechanical collision.

 During servicing, do not place anything on the lighthead or hang anything


on the light arms or lighthead.
 Connector pins with an ESD warning symbol should not be touched and no
connections should be made between these connectors without
implementing ESD protective measures. Such precautionary procedures
may include antistatic clothing and shoes, the touch of a ground stud before
and during connecting the pins or the use of electrically isolating and
antistatic gloves. All staff involved in the above shall receive instruction in
these ESD precautionary procedures.

 Refer to the operator’s manual for more information on operating the light.

1-4
2 General Maintenance

2.1 Daily Checks by User

Check the lightheads and the arms for


chipped paint, impact marks and any other
damage.

Check the stability/drift of the swivel arms, the


spring arms and the lightheads.

Check the load capacity of the spring arm.


Three checkpoints: bottom, middle, top.

Check if the sterilizable handle clicks and


locks in place correctly; replace it if not.

Check if the status indicator is green when


the lighthead is turned on.
For the lighthead with a touch control panel,
/ check if the power switch turns green when
the lighthead is turned on.

Check if the LEDs operate correctly.

2-1
General Maintenance

Check if the lighthead switches correctly


among normal lighting and full illuminance
mode/MIS lighting mode/AICS mode (for
HyLED 9700).

Check if the backup power module (optional)


can switch over.
/

Check if the wall control functions (optional)


work.

2-2
General Maintenance

2.2 Monthly Checks

Check the lightheads and the arms for


chipped paint, impact marks and any other
damage.

Check if the backup power supply (optional)


turns on and the light operates correctly if a
power cut were to occur.

Check the battery (optional) service time.

2-3
General Maintenance

2.3 Annual Checks

Check the verticality of the flange tube.

Check if the screws fixing the flange tube and swivel


arm assembly are tight and the rubber junction seal
is in place.

Check the limit stops.

Check the screw fixing the sleeve of the spring arm.

Remove the lightheads and lubricate the sleeve.


Recommended Greases: DeoxIT L260Np

Check whether the spring arm is correctly balanced.

Check for any loose covers and caps.

2-4
General Maintenance

Check the locking mechanism for the sterilizable


handle.

Use a photometer to check the maximum


illuminance:
HyLED 9700: about 160,000 lux at 1 metre
1m HyLED 9500/9500M: about 120,000 lux at 1 metre
HyLED 8600/8600M: about 160,000 lux at 1 metre
HyLED 760/760M: about 160,000 lux at 1 metre
HyLED 730/730M: about 130,000 lux at 1 metre

2-5
General Maintenance

2.4 Servicing Checklist

Result
No. Item Action
NA Pass Fail
Check the verticality of the flange
tube.
Check the rigidity between the flange
tube and swivel arm assembly.
All the visible screws are tightened.
The securing segment of the spring
Installation
1 arm is in place.
check
The gimbal joint, lighthead, camera
and the display are correctly
positioned.
All the end covers are tightened.
The sterilizable handle clicks and
locks into position.
The lighthead can stay at any desired
position.
The braking screws are adjusted.
Handling The spring arms are balanced.
2
check The height stop of the spring arm can
be adjusted.
Check the rotation stop of the swivel
arm, spring arm and gimbal joint.
Check the power supply connections,
including all the earth connections.
Check the cable clamp.
Check the
illuminance with the lux
batteries on at 1 meter
Electrical
3 (optional).
check
Check the lux
illuminance with AC. at 1 meter
Check earth bonding.
ohm
Check the tightening of the 3-pole
connector.
4 Optical check Check the maximum illuminance.
Operational Check the control panel function.
5
check The status indicator is green when

2-6
General Maintenance

AC is present. For touch control


panel, the power switch turns green.
The MIS mode can change over.
The AICS mode (for HyLED 9700
only) can change over.
The camera (optional, integrated
camera or carrier-arm camera) can
function normally.
The backup power supply (optional)
can change over.
The wall control (optional) functions
normally.
Check the cleanliness of the external
surfaces of lighthead and arms.
Check the cleanliness of the end
covers, and the covers are installed in
place.
Visual
6 No signs of wear on the sterilizable
inspection
handle
No signs of wear on the rubber
junction seal
No scratches or cracks on the glass
All labels are in place.

2-7
This page is intentionally blank.
3 FRU List and Diagram

3.1 Flange Tube

3.1.1 FRU (Field Repairable Unit) List

No. FRU number Description Remark

1 115-017954-00 Flange tube-225mm

2 115-017955-00 Flange tube-350mm

3 115-017956-00 Flange tube-500mm


4 115-017957-00 Flange tube-800mm
5 115-022058-00 Flange tube-1000mm
Flange tube-225mm (triple
6 115-017958-00
lighthead configuration)
Flange tube-350mm (triple
7 115-017959-00
lighthead configuration)
Flange tube-500mm (triple
8 115-017960-00 Flange tube (Integrated type)
lighthead configuration)
Flange tube-800mm (triple
9 115-017961-00
lighthead configuration)
Flange tube-1000mm (triple
10 115-022059-00
lighthead configuration)
11 115-022061-00 Flange tube-225mm (Φ270mm)

12 115-022062-00 Flange tube-350mm (Φ270mm)

13 115-022063-00 Flange tube-500mm (Φ270mm)


14 115-022064-00 Flange tube-800mm (Φ270mm)
15 115-022065-00 Flange tube-1000mm (Φ270mm)

16 115-021119-00 Separate flange tube-350mm

17 115-022060-00 Separate flange tube-500mm Separate flange tube

18 115-015837-00 Separate flange tube-800mm

19 041-008601-00 Insulation sleeve

20 041-008602-00 Insulation washer

3-1
FRU List and Diagram

3.1.2 Typical Diagram

3.1.2.1Flange Tube (Integrated Type)

No. FRU number Description Qty Remark

1 M16mm threaded bar 6

2 M16mm hexagon nut 24

3 Φ16mm plain washer 24


115-017954-00
4 Φ16mm spring washer 12

5 Flange tube-225mm 1
Protective ring, SB-1822A, nylon,
6 1
black

Figure 3-1 Flange tube

3-2
FRU List and Diagram

3.1.2.2Flange Tube (with Insulation Washer, Integrated Type)

No. FRU number Description Qty Remark

1 M16mm threaded bar 6

2 M16mm hexagon nut 24

3 Φ16mm plain washer 24


115-017954-00
4 Φ16mm spring washer 12

5 Flange tube-225mm 1
Protective ring, SB-1822A,
6 1
nylon, black
7 041-008601-00 Insulation sleeve 6

8 041-008602-00 Insulation washer 6

Figure 3-2 Flange tube

3-3
FRU List and Diagram

3.1.2.3Separate Flange Tube

No. FRU number Description Qty Remark

1 M16mm threaded bar 6


Hexagon socket head cap screw,
2 4
M8×35mm
3 Φ8mm plain washer 4

4 Φ8mm spring washer 4

5 Limit cover 1

6 115-021119-00 Protective ring, SB-1822A, nylon, black 1

7 M16mm hexagon nut 12

8 Φ16mm plain washer 12

9 Flange 1

10 Φ16mm spring washer 6

11 Threaded tube-350mm 1

Figure 3-3 Separate flange tube

3-4
FRU List and Diagram

3.1.2.4Separate Flange Tube (with Insulation Washer)

No. FRU number Description Qty Remark

1 M16mm threaded bar 6


Hexagon socket head cap screw,
2 4
M8×35mm
3 Φ8mm spring washer 4

4 Φ8mm plain washer 4

5 Limit cover 1

6 115-021119-00 Threaded tube-350mm 1

7 Protective ring, SB-1822A, nylon, black 1

8 M16mm hexagon nut 12

9 Φ16mm plain washer 12

10 Flange 1

11 Φ16mm spring washer 6

12 041-008601-00 Insulation sleeve 6

13 041-008602-00 Insulation washer 6

Figure 3-4 Separate flange tube

3-5
FRU List and Diagram

3.2 Swivel Arm Assembly

3.2.1 FRU List

No. FRU number Description Remark

1 115-021682-00 Column bottom cover

2 115-021683-00 Column cover

3 115-021688-00 Infrared remote control board assembly

4 0091-30-100509 One swivel arm (one lighthead, 3-pole)


One swivel arm (one lighthead, HyLED
5 115-022102-00
series integrated camera)
Two swivel arms (two lightheads,
6 0091-30-100504
3-pole)
Two swivel arms (one lighthead,
7 115-022104-00
carrier-arm camera)
Two swivel arms (one lighthead,
8 115-022072-00
carrier-arm camera)
Two swivel arms (one lighthead, 3-pole,
9 115-022086-00
single display)
Two swivel arms (one lighthead, 3-pole,
10 115-022087-00
double displays)
Two swivel arms (two lightheads,
11 115-018199-01
HyLED series integrated camera)
Two swivel arms (one lighthead, HyLED
12 115-022105-00 series integrated camera, HD single
display)
Two swivel arms (one lighthead, HyLED
13 115-022106-00 series integrated camera, HD double
displays)
Two swivel arms (one lighthead, HyLED
14 115-022107-00 series integrated camera, SD single
display)
Two swivel arms (one lighthead, HyLED
15 115-022108-00 series integrated camera, SD double
displays)
Three swivel arms (three lightheads,
16 0091-30-100501
3-pole)
17 115-017762-00 Three swivel arms (two lightheads,

3-6
FRU List and Diagram

carrier-arm camera)

Three swivel arms (two lightheads,


18 115-004712-01
carrier-arm camera)
Three swivel arms (two lightheads,
19 115-022088-00
3-pole, single display)
Three swivel arms (two lightheads,
20 115-022089-00
3-pole, double displays)
Three swivel arms (three lightheads,
21 115-022111-00
HyLED series integrated camera)
Three swivel arms (two lightheads,
22 115-011330-00 HyLED series integrated camera, HD
single display)
Three swivel arms (two lightheads,
23 115-022112-00 HyLED series integrated camera, HD
double displays)
Three swivel arms (two lightheads,
24 115-019631-01 HyLED series integrated camera, SD
single display)
Three swivel arms (two lightheads,
25 115-022113-00 HyLED series integrated camera, SD
double displays)
Three swivel arms (one lighthead,
26 115-022114-00
carrier-arm camera, single display)
Three swivel arms (one lighthead,
27 115-022115-00
carrier-arm camera, double displays)
Three swivel arms (one lighthead,
28 115-022077-00
carrier-arm camera, single display)
Three swivel arms (one lighthead,
29 115-022076-00
carrier-arm camera, double displays)
Four swivel arms (three lightheads,
30 115-022116-00
carrier-arm camera)
Four swivel arms (three lightheads,
31 115-022078-00
carrier-arm camera)
Four swivel arms (three lightheads,
32 115-022090-00
3-pole, single display)
Four swivel arms (three lightheads,
33 115-022091-00
3-pole, double displays)
Four swivel arms (three lightheads,
34 115-022117-00 HyLED series integrated camera, HD
single display)
Four swivel arms (three lightheads,
35 115-022118-00 HyLED series integrated camera, HD
double displays)

3-7
FRU List and Diagram

Four swivel arms (three lightheads,


36 115-022119-00 HyLED series integrated camera, SD
single display)
Four swivel arms (three lightheads,
37 115-022120-00 HyLED series integrated camera, SD
double displays)
Four swivel arms (two lightheads,
38 115-017761-01
carrier-arm camera, single display)
Four swivel arms (two lightheads,
39 115-022121-00
carrier-arm camera, double displays)
Four swivel arms (two lightheads,
40 115-002788-01
carrier-arm camera, single display)
Four swivel arms (two lightheads,
41 115-022080-00
carrier-arm camera, double displays)

3.2.2 Typical Diagram

3.2.2.1Swivel Arm Assembly

No. FRU number Description Qty Remark

1 / Central axis (four swivel arms) 1

2 / Deep groove ball bearing 8


Protective ring, SB-1822A, nylon,
3 / 2
black
4 / Braking screw on the main axis 8

5 / Plain washer 3
Hexagon socket head cap screw,
6 / 6
M3×12mm, stainless steel, passivated
7 / Plain washer, stainless steel, polished 6
Cross recessed countersunk head
8 / screw, M3×6mm, stainless steel, 8
passivated
9 / Side cover on the swivel arm 4

10 / Φ60mm circlip 1

11 / M60mm locking nut 1

3-8
FRU List and Diagram

Cross recessed countersunk head


12 screw, M4×8mm, stainless steel, 3
passivated
115-021682-00
13 Column bottom cover 1

14 Small plastic cover 1

15 / Φ8mm plain washer 6

16 / Φ8mm spring washer 2


Hexagon socket head cap screw,
17 / 2
M8×16mm, galvanized
18 / M4mm hexagon nut, passivated 2

19 / Swivel arm-1150mm 1

20 / 3-pole connector assembly 2

21 / Stator of 3-pole connector 2

22 / Wired swivel arm-1000mm 1


Hexagon socket countersunk head
23 / screw, M4×16mm, stainless steel, 6
polished
24 / Swivel arm (3-pole)-850mm 1

25 / Swivel arm-700mm 1

26 / End cover for swivel arm 4


Hexagon socket countersunk head
27 / screw, M5×12mm, stainless steel, 6
polished
28 / Braking screw for spring arm 8
Cross recessed small pan head
29 / combination screw, M3×8mm, 4
galvanized
30 / Cover for fixing the 3-pole plug 2

31 / Pin, Φ3×10mm, nickel plated 4

32 / 3-pole plug 2

3-9
FRU List and Diagram

Figure 3-5 Swivel arm assembly

3-10
FRU List and Diagram

3.2.2.2Infrared Remote Control Board Assembly

No. FRU number Description Qty Remark


Cross recessed pan head screw,
1 4
115-021688-00 M3×5mm, stainless steel, passivated
2 Infrared remote control board 1

3 115-021682-00 Column bottom cover 1

Figure 3-6 Infrared remote control board assembly

3-11
FRU List and Diagram

3.3 Spring Arm Assembly

3.3.1 FRU List

No. FRU number Description Remark


Standard spring arm assembly
1 115-004994-00 Rotation-unlimited spring arm
(AC2000, 3-pole, 12-18kg)
Wired spring arm assembly
2 115-004708-00 (AC2000, 7-12kg, for carrier-arm
camera)
Wired spring arm assembly
3 115-004709-00 (AC2000, 7-12kg, for HD
display)
Wired spring arm assembly
4 115-004710-00 (AC2000, 7-12kg, for SD
display)
Wired spring arm assembly
5 115-015798-00 (AC2000, 13.5-21kg, for HD Rotation-limited spring arm
display)
Wired spring arm assembly
6 115-015801-00 (AC2000, 13.5-21kg, for SD
display)
Wired spring arm assembly
7 115-017126-01 (AC2000, 13.5-21kg, for HyLED
series integrated camera)
Wired spring arm assembly
8 115-017758-00 (AC2000, 7-12kg, for carrier-arm
camera)
Wires spring arm assembly
Rotation-limited, heavy-load
9 115-011331-00 (AC2000, 13.5-21kg, heavy
spring arm
load)
LCH spring arm assembly Spring arm for low ceiling
10 115-005044-00
(AC2000, 176Nm, 3-pole) height use, rotation unlimited
LCH spring arm assembly Spring arm for low ceiling
11 / (AC2000, 176Nm, 3-pole, heavy height use, rotation unlimited,
load) heavy load
LCH spring arm assembly
Spring arm for low ceiling
12 / (AC2000, 176Nm, for HyLED
height use, rotation limited
series integrated camera)
13 / Wired spring arm assembly Ondal Space spring arm,

3-12
FRU List and Diagram

(Ondal Space, 20-40kg) rotation limited

Wired spring arm assembly


14 /
(Ondal Space, 20-40kg, HD)
Wired spring arm assembly
15 /
(Ondal Space, 20-40kg, SD)

3.3.2 Typical Diagram

3.3.2.1Standard Spring Arm

No. FRU number Description Qty Remark

1 Circlip 1

2 Washer 1 1

3 115-004994-00 Bearing seat 1

4 Washer 2 1

5 Standard spring arm (3-pole) 1

Figure 3-7 Standard spring arm

3-13
FRU List and Diagram

3.3.2.2LCH Spring Arm

No. FRU number Description Qty Remark

1 Circlip 1

2 Washer 1 1

3 115-005044-00 Bearing seat 1

4 Washer 2 1

5 LCH spring arm (3-pole) 1

Figure 3-8 LCH spring arm

3-14
FRU List and Diagram

3.3.2.3Wired Standard Spring Arm

No. FRU number Description Qty Remark

1 Circlip 1

2 Washer 1 1
115-004708-00
3 Bearing seat 1

4 Wired standard spring arm 1

Figure 3-9 Wired standard spring arm

3-15
FRU List and Diagram

3.3.2.4Wired LCH Spring Arm

No. FRU number Description Qty Remark

1 Circlip 1

2 Plain washer 1 1
/
3 Bearing seat 1

4 Wired LCH spring arm 1

Figure 3-10 Wired LCH spring arm

3-16
FRU List and Diagram

3.3.2.5Wired Standard Spring Arm (Ondal Space)

No. FRU number Description Qty Remark

1 Spring arm 1

2 Bearing seat 1
/
3 Plain washer 1 1

4 Circlip 1

Figure 3-11 Wired standard spring arm (Ondal Space)

3-17
FRU List and Diagram

3.4 Gimbal Joint Assembly

3.4.1 FRU List


No. FRU number Description Remark

1 / Horizontal gimbal joint (for HyLED 9500)

2 / Horizontal gimbal joint (for HyLED 9700)

3 / Horizontal gimbal joint (for HyLED 760/730)

4 / Horizontal gimbal joint (for HyLED 8600)

5 / Vertical gimbal joint (for HyLED 9500)

6 / Vertical gimbal joint (for HyLED 9700/8600)

7 / Vertical gimbal joint (for HyLED760/730)

8 / Wired vertical gimbal joint (for HyLED 9700)

9 / Wired vertical gimbal joint (for HyLED 760)

10 / Gimbal joint (for HyLED 760/730, LCH)

11 / Gimbal joint (for HyLED 8600, LCH)

12 / Gimbal joint (for HyLED 9500, LCH)

13 / Gimbal joint (for HyLED 9700, LCH)

14 / Wired gimbal joint (for HyLED 760, LCH)

15 / Wired gimbal joint (for HyLED 9500, LCH)

3-18
FRU List and Diagram

3.4.2 Typical Drawing

3.4.2.1Gimbal Joint Assembly

No. FRU number Description Qty Remark

1 / Horizontal gimbal joint (for HyLED 9500) 1

2 / End bearing, 28×48×1.5 1

3 / Vertical gimbal joint (for HyLED 9500) 1

4 / Plain washer for gimbal joint 1

5 / Φ25mm circlip 2

6 / Braking screw for gimbal joint 2

7 / 3-pole connector 1
Cross recessed small pan head
8 / combination screw, M3×8mm, 2
galvanized
9 / End cover for gimbal joint 4

10 / Limited pin, 4m6×12 2

Figure 3-12 Horizontal and vertical gimbal joints

3-19
FRU List and Diagram

3.4.2.2Gimbal Joint Assembly for Lighthead with Integrated Camera

No. FRU number Description Qty Remark

1 / Horizontal gimbal joint (for HyLED 9700) 1

2 / Limited pin, 4m6×12 2

3 / End bearing, 28×48×1.5 1

4 / Wired vertical gimbal joint (for HyLED 9700) 1

5 / Plain washer for gimbal joint 1

6 / End cover for gimbal joint 4

7 / Φ25mm circlip 2

8 / Braking screw for gimbal joint 2

9 / Φ4mm spring washer 1

10 / Hexagon socket head cap screw, M4×6mm 1

Figure 3-13 Gimbal joint assembly for the lighthead with an integrated camera

3-20
FRU List and Diagram

3.4.2.3Gimbal Joint Assembly (LCH)

No. P/N Description Qty Remark

1 / Gimbal joint (for HyLED 760/730, LCH) 1

2 / 3-pole connector 1
Cross recessed small pan head combination
3 / 2
screw, M3×8mm, galvanized
4 / Sealing ring 1

5 / End cover for gimbal joint 1

6 / Limited pin, 4m6×12 1

7 / End bearing, 28×48×1.5 1

8 / O-ring 1

9 / Needle rolling bearing 2

10 / Braking screw for gimbal joint 2

11 / Plain washer for gimbal joint 1

12 / Φ25mm circlip 1

13 / Washer of LED board 2


Cross recessed countersunk head screw,
14 / 2
M3×12mm, galvanized
15 / End cover for control panel seat 1

16 / Control panel seat 1


Cross recessed countersunk head screw,
17 / 1
M5×10mm, stainless steel, passivated
18 / Steel ball 2

19 / Spring 2

20 / Spring screw cover 2

3-21
FRU List and Diagram

Figure 3-14 Gimbal joint assembly (LCH)

3-22
FRU List and Diagram

3.4.2.4Gimbal Joint Assembly for Lighthead with Integrated Camera


(LCH)

No. FRU number Description Qty Remark

1 / End cover for gimbal joint 1

2 / Limited pin, 4m6×12 1

3 / End bearing, 28×48×1.5 1

4 / O-ring 1

5 / Braking screw for gimbal joint 2

6 / Needle rolling bearing 2

7 / Plain washer for gimbal joint 1

8 / Φ25mm circlip 1

9 / Control panel seat 1

10 / Washer of LED board 2

11 / End cover for control panel seat 1


Cross recessed countersunk head
12 / 2
screw, M3×12mm, galvanized
Cross recessed countersunk head
13 / screw, M5×10mm, stainless steel, 1
passivated
14 / Steel ball 2

15 / Spring 2

16 / Spring screw cover 2


Wired gimbal joint (for HyLED 760,
17 / 1
LCH)

3-23
FRU List and Diagram

Figure 3-15 Gimbal joint assembly for lighthead with an integrated camera (LCH)

3-24
FRU List and Diagram

3.5 Control Panel

3.5.1 FRU List

No. FRU number Description Remark

1 115-021698-00 Control board (HyLED 9700, touch control panel)


Front cover assembly (HyLED 9700, touch control
2 115-021699-00
panel)
Control board (HyLED 9700, touch control panel,
3 115-021643-00
with color temperature adjustment)
Front cover assembly (HyLED 9700, touch control
4 115-021644-00
panel, with color temperature adjustment)
Control board (HyLED 9700 with an integrated
5 115-021645-00
camera, touch control panel)
Front cover assembly (HyLED 9700 with an
6 115-021646-00
integrated camera, touch control panel)
Control board (HyLED 9700 with an integrated
7 115-021647-00 camera, touch control panel, with color
temperature adjustment)
Front cover assembly (HyLED 9700 with an
8 115-021648-00 integrated camera, touch control panel, with color
temperature adjustment)
9 115-021649-00 Control board (HyLED 9500, touch control panel)
Front cover assembly (HyLED 9500, touch control
10 115-021650-00
panel)
Control board (HyLED 9500, touch control panel,
11 115-021651-00
with color temperature adjustment)
Front cover assembly (HyLED 9500, touch control
12 115-021700-00
panel, with color temperature adjustment)
13 115-021701-00 Control board (HyLED 9700, control panel)

14 115-021702-00 Control board (HyLED 9500, control panel)

15 115-021704-00 Control board (HyLED 9700, touch control panel)


Front cover assembly (HyLED 9700, touch control
16 115-021705-00
panel) New
Control board (HyLED 9700 with an integrated hardware
17 115-021706-00
camera, touch control panel)
Front cover assembly (HyLED 9700 with an
18 115-021707-00
integrated camera, touch control panel)

3-25
FRU List and Diagram

Control board (HyLED 9700, touch control panel,


19 115-021708-00
carrier-arm camera)
Front cover assembly (HyLED 9700, touch control
20 115-021718-00
panel, carrier-arm camera)
21 115-021719-00 Control board (HyLED 9500, touch control panel)
Front cover assembly (HyLED 9500, touch control
22 115-021720-00
panel)
Control board (HyLED 9500 with an HD integrated
23 115-021721-00
camera, touch control panel)
Touch control panel (HyLED 9500 with an HD
24 115-021722-00
integrated camera)
Control board (HyLED 9500, touch control panel,
25 115-021723-00
carrier-arm camera)
Front cover assembly (HyLED 9500, touch control
26 115-021724-00
panel, carrier-arm camera)
27 115-021661-00 Control board (HyLED 760, control panel)

28 115-021662-00 Control board (HyLED 730, control panel)

29 115-021663-00 Control board (HyLED 760, touch control panel)

30 115-021664-00 Touch control panel (HyLED 760)


Control board (HyLED 760 with an HD integrated
31 115-021665-00
camera, touch control panel)
Front cover assembly (HyLED 760 with an HD
32 115-021666-00
integrated camera, touch control panel)
Control board (HyLED 760 with an SD integrated
33 115-021667-00
camera, touch control panel)
Front cover assembly (HyLED 760 with an SD
34 115-021668-00
integrated camera, touch control panel)
Control board (HyLED 760, touch control panel,
35 115-021669-00
carrier-arm camera)
Touch control panel (HyLED 760, carrier-arm
36 115-021670-00
camera)
37 115-021671-00 Control board (HyLED 730, touch control panel)
Front cover assembly (HyLED 730, touch control
38 115-021672-00
panel)
Control board (HyLED 730, touch control panel,
39 115-021673-00
carrier-arm camera)
Front cover assembly (HyLED 730, touch control
40 115-021674-00
panel, carrier-arm camera)
41 115-021679-00 Control board (HyLED 8600, control panel)

42 115-021680-00 Control board (HyLED 8600S, control panel)

3-26
FRU List and Diagram

3.5.2 Typical Diagram

3.5.2.1Control Panel

No. FRU number Description Qty Remark

1 / Front cover of control panel 1

2 Control board 1
Cross recessed pan head screw, M3×8mm,
3 4
passivated
115-021701-00
4 Micro-switch cap 5
Cross recessed small pan head combination
5 7
screw, M3×8mm, galvanized
6 / Back cover of control panel 1
Sticker of control panel (HyLED
7 / 1
9500/9500M)

Figure 3-16 Control panel

3-27
FRU List and Diagram

3.5.2.2Touch Control Panel

No. FRU number Description Qty Remark

1 AM1808 CPU module 1


Cross recessed small pan head
2 2
screw, M2.5×6mm, galvanized
Control board of touch control
3 115-021651-00 1
panel
115-021649-00 Stud screw, M2.5×7+6-6, nickel
4 6
plated 1. The assemblies
Cross recessed pan head screw, (115-021651-00 and
5 M3×6mm, stainless steel, 4 115-021700-00)
passivated have the color
Cross recessed small pan head temperature
6 combination screw, M3×8mm, 8 adjustment function.
galvanized 2. The assemblies
Back cover of touch control (115-021649-00 and
7 1
panel 115-021650-00) are
8 Support 1 for HyLED 9500
lighthead use,
Cross recessed countersunk
without an integrated
9 115-021700-00 head screw, M3×6mm, stainless 2
camera and color
115-021650-00 steel, passivated
temperature
10 Touch screen 1 adjustment.
11 Sealing strip 1

12 Washer 1
Front cover of touch control
13 1
panel
14 Mindray label (small) 1

3-28
FRU List and Diagram

Figure 3-17 Touch control panel

3-29
FRU List and Diagram

3.5.2.3Control Panel (New Hardware)

No. FRU number Description Qty Remark


Cross recessed small pan
1 / head screw with washer, 4
M3×12mm, galvanized
2 / Washer of LED board 4

3 / Back cover of control panel 1


Cross recessed pan head 1. The assembly
4 screw, M3×6mm, 2 (115-021661-00) is for
galvanized HyLED 760 use.
5 115-021661-00 Control board 1 2. The assembly
115-021662-00 Cross recessed small pan (115-021662-00) is for
6 head combination screw, 6 HyLED 730 use.
M3×8mm, galvanized
7 Micro-switch cap 7

8 / Front cover of control panel 1

9 / Mindray label (small) 1

10 / Sticker of control panel 1

Figure 3-18 Control panel (new hardware)

3-30
FRU List and Diagram

3.6 Lighthead

3.6.1 FRU List

No. FRU number Description Remark

1 049-000181-00 Screw cap

2 049-000184-00 Screw cap for top cover


Screw cap on the lamp housing
3 043-003048-00 hood (for HyLED
730/730M/760/760M)
4 115-014794-00 LED unit Old LED unit (Seoul):
115-014795-00 and
5 115-014795-00 Nature white LED unit
115-014796-00: for color
6 115-014796-00 Warm white LED unit temperature adjustment use

7 115-021693-00 LED unit New LED unit (Nichia):


115-021694-00 and
8 115-021694-00 Cold white LED unit
115-021695-00: for color
9 115-021695-00 Warm white LED unit temperature adjustment use
Encircling handle (for HyLED
10 115-021691-00
9700)
Encircling handle (for HyLED
11 115-021692-00
9500/9500M)
Encircling handle (HyLED
12 115-021678-00
8600/8600M)
13 115-014800-00 Glass (for HyLED 9700)

14 115-014801-00 Glass (for HyLED 9500/9500M)


Glass (for HyLED
15 115-021757-00
730/730M/760/760M)
16 115-015726-00 Glass (for HyLED 8600/8600M)
Drive board (for HyLED
17 115-021675-00
730/730M/760/760M)
Drive board (for HyLED
18 115-021703-00 Old hardware
9700/9500/8600)
Drive board (for HyLED
19 115-021725-00 New hardware
9700/9500)
20 115-006128-00 Sterilizable handle assembly

3-31
FRU List and Diagram

3.6.2 Typical Diagram

3.6.2.1HyLED 730/730M Lighthead

No. P/N Description Qty Remark


Lamp housing hood (for HyLED
1 / 1
760/760M/730/730M)
2 LED unit 1

3 Fixing piece for LED unit 20

4 115-021693-00 Washer of LED board 44


Cross recessed small pan head
5 combination screw, M3×10mm, 44
galvanized
6 / Magnetic core 1

7 / Metal piece for magnetic core 1


Cross recessed small pan head
8 / combination screw, M4×8mm, 1
galvanized
Screw cap on the lamp housing hood (for
9 043-003048-00 1
HyLED 730/730M/760/760M)
Cross recessed pan head screw,
10 / 1
M3×16mm, galvanized
11 / Support pin 4

12 / Black pad 2
Drive board (for HyLED
13 115-021675-00 1
730/730M/760/760M)
Cross recessed pan head screw,
14 / 4
M3×6mm, galvanized
15 / Plastic cover (for HyLED 730/730M) 1

16 Inner sealing strip 2

17 115-021757-00 Outer sealing strip 1

18 Glass (for HyLED 730/730M/760/760M) 1


Cross recessed small pan head
19 / combination screw, M3×8mm, 10
galvanized

3-32
FRU List and Diagram

20 / Outer handle 1

21 115-006128-00 Sterilizable handle assembly 1

Figure 3-19 HyLED 730/730M lighthead

3-33
FRU List and Diagram

3.6.2.2HyLED 760/760M Lighthead

No. FRU number Description Qty Remark


Lamp housing hood (for HyLED
1 / 1
760/760M/730/730M)
2 LED unit 1

3 Fixing piece for LED unit 20

4 115-021693-00 Washer of LED board 47


Cross recessed small pan head
5 combination screw, M3×10mm, 46
galvanized
6 / Metal piece for magnetic core 1
Cross recessed small pan head
7 / combination screw, M4×8mm, 1
galvanized
8 / Magnetic core 1
Screw cap (for HyLED
9 043-003048-00 1
730/730M/760/760M)
Cross recessed pan head screw,
10 / 1
M3×16mm, galvanized
11 / Support pin 4

12 / Black pad 2
Drive board (for HyLED
13 115-021675-00 1
730/730M/760/760M)
Cross recessed pan head screw,
14 / 4
M3×6mm, galvanized
15 / Plastic cover (for HyLED 760/760M) 1

16 Inner sealing strip 2

17 115-021757-00 Outer sealing strip 1


Glass (for HyLED
18 1
730/730M/760/760M)
Cross recessed small pan head
19 / combination screw, M3×8mm, 10
galvanized
20 / Outer handle 1

21 115-006128-00 Sterilizable handle assembly 1

3-34
FRU List and Diagram

Figure 3-20 HyLED 760/760M lighthead

3-35
FRU List and Diagram

3.6.2.3HyLED 8600/8600M Lighthead

No. FRU number Description Qty Remark

1 049-000181-00 Screw cap 12


Lamp housing hood (for HyLED
2 / 1
8600/8600M)
Plastic sleeve (for HyLED
3 / 1
8600/8600M)
4 LED unit 51

5 Washer of LED board 102


115-014794-00
Cross recessed small pan head
6 combination screw, M3×10mm, 102
galvanized
Plastic cover (for HyLED
7 / 3
8600/8600M)
8 / Sealing hose 1
Cross recessed countersunk head
9 / 4
screw, M3×8mm, galvanized
10 049-000184-00 Screw cap for top cover 3
Cross recessed countersunk head
11 / screw, M4×12mm, stainless steel, 3
polished
12 / Top cover 1

13 / Metal plate 1

14 / Black pad 4
Cross recessed pan head screw,
15 / 3
M3×8mm, galvanized
Drive board (for HyLED
16 115-021703-00 1
9700/9500/8600)
Stud screw, M3×10+8-8, nickel
17 / 3
plated
18 Glass (for HyLED 8600/8600M) 1

115-015726-00 Cross recessed small pan head


19 screw with washer, M3×12mm, 3
galvanized

3-36
FRU List and Diagram

Cross recessed pan head screw,


20 12
M3×16mm, galvanized
Encircling handle (right opening, for
21 1
HyLED 8600/8600M)
115-021678-00
Encircling handle (for HyLED
22 1
8600/8600M)
Encircling handle (left opening, for
23 1
HyLED 8600/8600M)
24 115-006128-00 Sterilizable handle assembly 1

Figure 3-21 HyLED 8600/8600M lighthead

3-37
FRU List and Diagram

3.6.2.4HyLED 9500/9500M Lighthead

No. FRU number Description Qty Remark

1 049-000181-00 Screw cap 16


Lamp housing hood (for
2 / 1
HyLED 9500/9500M)
3 / Plastic sleeve 1

4 LED unit 1

5 Washer of LED board 1 For HyLED 9500/9700


115-014795-00
lighthead with color
115-014796-00 Cross recessed small pan
temperature adjustment
6 head combination screw, 1
M3×10mm, galvanized
Plastic cover (for HyLED
7 / 4
9500/9500M)
8 / Sealing hose 1
Cross recessed countersunk
9 / head screw, M3×8mm, 4
galvanized
10 049-000184-00 Screw cap for top cover 3
Cross recessed countersunk
11 / head screw, M4×12mm, 3
stainless steel, polished
12 / Top cover 1

13 / Metal plate 1

14 / Black pad 4
Cross recessed pan head
15 / 3
screw, M3×8mm, galvanized
Drive board (for HyLED
16 115-021703-00 1
9700/9500/8600)
Stud screw, M3×10+8-8,
17 / 3
nickel plated
Glass (for HyLED
18 1
9500/9500M)
115-014801-00 Cross recessed small pan
19 head screw with washer, 4
M3×12mm, galvanized

3-38
FRU List and Diagram

Cross recessed pan head


20 screw, M3×16mm, 16
galvanized
Encircling handle (right
21 opening, for HyLED 1
115-021692-00 9500/9500M)
Encircling handle (for HyLED
22 2
9500/9500M)
Encircling handle (left
23 opening, for HyLED 1
9500/9500M)
24 115-006128-00 Sterilizable handle assembly 1

Figure 3-22 HyLED 9500/9500M lighthead

3-39
FRU List and Diagram

3.6.2.5HyLED 9700 Lighthead

No. FRU number Description Qty Remark

1 049-000181-00 Screw cap 16


Lamp housing hood (for
2 / 1
HyLED 9700)
3 / Plastic sleeve 1

4 LED unit 1

5 Washer of LED board 1 For HyLED 9500/9700


115-014794-00
lighthead with color
115-014795-00 Cross recessed small pan
temperature adjustment
6 head combination screw, 1
M3×10mm, galvanized
Plastic cover (for HyLED
7 / 4
9700)
8 / Sealing hose 1
Cross recessed
9 / countersunk head screw, 4
M3×8mm, galvanized
10 049-000184-00 Screw cap for top cover 3
Cross recessed
11 / countersunk head screw, 3
M4×12mm, polished
12 / Top cover 1

13 / Metal plate 1

14 / Black pad 4
Cross recessed pan head
15 / screw, M3×8mm, 3
galvanized
Drive board (for HyLED
16 115-021703-00 1
9700/9500/8600)
Stud screw, M3×10+8-8,
17 / 3
nickel plated
18 Glass (for HyLED 9700) 1

115-014800-00 Cross recessed small pan


19 head screw with washer, 4
M3×12mm, galvanized

3-40
FRU List and Diagram

Cross recessed pan head


20 screw, M3×16mm, 16
galvanized
Encircling handle (right
21 1
115-021691-00 opening, for HyLED 9700)
Encircling handle (for
22 2
HyLED 9700)
Encircling handle (left
23 1
opening, for HyLED 9700)
Sterilizable handle
24 115-006128-00 1
assembly

Figure 3-23 HyLED 9700 lighthead

3-41
FRU List and Diagram

3.7 Sterilizable Handle

3.7.1 FRU List

No. FRU number Description Remark


Central handle assembly (without integrated
1 115-014793-00
camera)
2 115-006128-00 Sterilizable handle assembly

3 115-011156-00 Removable handle assembly

3.7.2 Typical Diagram

3.7.2.1Sterilizable Handle without Light Field Adjustment (for HyLED


8600/8600M series)

No. FRU number Description Qty Remark

1 / Mechanism without light field adjustment 1


Cross recessed countersunk head
2 / 4
screw, M4×20mm
3 / Central cover 1
Cross recessed countersunk head
4 / screw, M4×12mm, stainless steel, 3
polished
5 Central handle 1

6 115-014793-00 Φ4mm spring washer 2

7 M3mm expansion screw 2

8 115-006128-00 Sterilizable handle assembly 1

3-42
FRU List and Diagram

Figure 3-24 Sterilizable handle without light field adjustment

3.7.2.2Sterilizable Handle with Light Field Adjustment (for HyLED


9500/9700/9500M series)

No. FRU number Description Qty Remark

1 / Focusing mechanism 1
Cross recessed countersunk head
2 / 4
screw, M4×20mm
3 / Central cover 1
Cross recessed countersunk head
4 / screw, M4×12mm, stainless steel, 3
polished
5 Central handle 1

6 115-014793-00 Φ4mm spring washer 2

7 M3mm expansion screw 2

8 115-006128-00 Sterilizable handle assembly 1

3-43
FRU List and Diagram

Figure 3-25 Sterilizable handle with light field adjustment

3-44
FRU List and Diagram

3.7.2.3Sterilizable Handle without Light Field Adjustment (for HyLED


730/760/730M/760M series)

No. P/N Description Qty Remark


Cross recessed small pan head
1 / combination screw, M5×12mm, 2
galvanized
2 / Installation plate for handle 1

3 / Sealing hose 2
Cross recessed countersunk head
4 / 4
screw, M4×20mm
5 / Installation shaft for central handle 1

6 / Central cover 1
Cross recessed countersunk head
7 / screw, M4×12mm, stainless steel, 3
polished
8 Central handle 1

9 115-014793-00 Φ4mm spring washer 2

10 M3mm expansion screw 2

11 115-006128-00 Sterilizable handle assembly 1

Figure 3-26 Sterilizable handle for HyLED 730/760/730M/760M series

3-45
FRU List and Diagram

3.7.2.4Sterilizable Handle with Integrated Camera Interface

No. FRU number Description Qty Remark

1 / Female connector assembly 1

2 115-011156-00 Removable handle assembly 1

3 Central handle 1

4 115-014793-00 Φ4mm spring washer 2

5 M3mm expansion screw 2

6 115-006128-00 Sterilizable handle assembly 1

Figure 3-27 Sterilizable handle with integrated camera interface

3-46
FRU List and Diagram

3.8 Backup Battery Module

3.8.1 FRU List

No. FRU number Description Remark

1 115-021717-00 Battery module

2 115-005026-00 Battery assembly Old hardware

3 051-000298-00 Charging board

4 115-021639-00 Battery module

5 115-018675-00 Battery assembly New hardware

6 051-001477-00 Charging board

3.8.2 Typical Diagram

No. P/N Description Qty Remark


Cross recessed countersunk head screw,
1 4
M4×8mm, stainless steel, passivated
2 115-021717-00 Charging board 1 051-000298-00
Cross recessed small pan head
3 4
combination screw, M3×8mm, galvanized
4 Upper cover assembly 1

5 Metal bar 1
Cross recessed small pan head
6 6
combination screw, M5×12mm, galvanized
7 Lead acid batteries, 12V,15Ah 2
115-005026-00
8 Shock absorber 1

9 Lower cover 1

10 Washer for the battery module 4


Cross recessed small pan head
11 4
combination screw, M4×12mm, galvanized

3-47
FRU List and Diagram

Figure 3-28 Backup battery module

3-48
FRU List and Diagram

3.9 Integrated Camera

3.9.1 FRU List

No. FRU number Description Remark


Rotatable HD integrated camera with sterilizable
1 115-020559-00
handle
2 115-017660-00 Rotatable HD integrated camera assembly
Sterilizable handle for rotatable integrated
3 115-017661-00
camera
Rotatable SD integrated camera with sterilizable
4 115-020560-00
handle
5 115-017659-00 Rotatable SD integrated camera assembly

6 115-009813-00 Sterilizable handle for integrated camera

7 115-022362-00 Removable integrated camera assembly

8 115-021635-00 Control assembly for rotatable integrated camera

9 115-021697-00 Control assembly for HD integrated camera

3-49
FRU List and Diagram

3.9.2 Typical Diagram

3.9.2.1Rotatable Integrated Camera

No. FRU number Description Qty Remark

1 / Female connector assembly 1

2 Rotatable HD integrated camera 1 115-017660-00


115-020559-00 Sterilizable handle for rotatable integrated
3 1 115-017661-00
camera

Figure 3-29 Rotatable integrated camera

3-50
FRU List and Diagram

3.9.2.2Integrated Camera

No. FRU number Description Qty Remark

1 / Fixing ring 1

2 115-022362-00 Removable integrated camera assembly 1

3 115-009813-00 Sterilizable handle for integrated camera 1

Figure 3-30 Integrated camera

3-51
FRU List and Diagram

3.10 Carrier-arm Camera

3.10.1 FRU List

No. FRU number Description Remark

1 115-022363-00 Camera assembly


Old hardware
2 115-021711-00 Control board of carrier-arm camera

3 115-017756-00 Camera assembly


New hardware
4 115-021726-00 Control board of carrier-arm camera

3.10.2 Typical Diagram

No. FRU number Description Qty Remark


Cross recessed pan head screw, M3×5mm,
1 3
115-021726-00 stainless steel, passivated
2 Control board of carrier-arm camera 1

3 / Cover 1
Cross recessed countersunk head screw,
4 / 1
M4×8mm, stainless steel, passivated
5 115-006128-00 Sterilizable handle assembly 1

Figure 3-31 Carrier-arm camera

3-52
FRU List and Diagram

3.11 Swivel Arm for Display

3.11.1 FRU List

No. FRU number Description Remark

1 / Swivel arm for single display

2 / Swivel arm for double displays

3.11.2 Typical Diagram

3.11.2.1Swivel Arm for Single Display

No. FRU number Description Qty Remark

1 / Display 1

2 / Braking screw 1
Hexagon socket head cap screw,
3 / 12
M4×12mm
4 / Vertical arm assembly 1

5 / Front cover of storage box 1

6 / Cover of vertical arm 1


Cross recessed countersunk head
7 / 14
screw, M3×6, passivated
8 / Φ4mm plain washer 12

9 / Φ4mm spring washer 12


Hexagon socket head cap screw,
10 / 4
M4×6mm, galvanized
11 / Power cable distribution box 1
Cross recessed small pan head
12 / combination screw, M4×8mm, 7
galvanized
13 / Balancing weight 1

14 / Back cover of storage box 1

3-53
FRU List and Diagram

15 / Fixing support for the adaptor 1

16 / White plug, Φ12.7×10.3mm, nylon 1

17 / Horizontal arm 1

18 / Black plug, round, Φ30×11.5mm 1

19 / Central handle 1

20 / M3mm expansion screw 2

21 / Sterilizable handle 1

22 / Mounting plate for display 1


Cross recessed countersunk head
23 / 8
screw, M5×10mm, passivated

3-54
FRU List and Diagram

Figure 3-32 Swivel arm for single display

3.11.2.2Swivel Arm for Double Displays

No. FRU number Description Qty Remark


Right horizontal
arm, also used in
1 / Horizontal arm 1
swivel arm for
single display

3-55
FRU List and Diagram

Hexagon socket head cap screw,


2 / 2
M4×6mm, galvanized
3 / Plain washer for gimbal joint 2

4 / Φ25mm circlip 4

5 / End cover for gimbal joint 2

6 / Extension arm 1
Wired spring arm assembly (Ondal
7 / 1
Space, 20-40kg, HD)
8 / End cover for swivel arm 1

9 / Vertical arm 2

10 / Display 2

11 / Left horizontal arm 1

Figure 3-33 Swivel arm for double displays

3-56
FRU List and Diagram

3.12 Ceiling Cover

3.12.1 FRU List

No. FRU number Description Remark

1 0091-30-102274 Big ceiling cover assembly

2 0091-30-102273 Small ceiling cover assembly

3 115-022070-00 Square ceiling cover assembly

3.12.2 Typical Diagram

3.12.2.1Big ceiling cover assembly

No. FRU number Description Qty Remark

1 Rubber junction seal 1


Cross recessed countersunk head screw,
2 8
M4×12mm, stainless steel, polished
3 Big ceiling cover-half 2

4 0091-30-102274 White plug, Φ12.7×10.3mm, nylon 4


Hexagon socket head cap screw, M4×16mm,
5 2
stainless steel, passivated
6 Φ4mm spring washer 2

7 Fixing ring of ceiling cover 2

3-57
FRU List and Diagram

Figure 3-34 Big ceiling cover assembly

3.12.2.2Small ceiling cover assembly

No. FRU number Description Qty Remark

1 Small ceiling cover 1 1

2 Small ceiling cover 2 1


Cross recessed countersunk head screw,
3 6
M4×12mm, stainless steel, polished
0091-30-102273
4 Fixing ring of ceiling cover 2

5 Φ4mm spring washer 2


Hexagon socket head cap screw, M4×16mm,
6 2
stainless steel, passivated

Figure 3-35 Small ceiling cover assembly

3-58
FRU List and Diagram

3.12.2.3Square ceiling cover assembly

No. FRU number Description Qty Remark

1 Square ceiling cover 2


Hexagon socket head cap screw, M4×16mm,
3 2
stainless steel, passivated
4 115-022070-00 Φ4mm spring washer 2

5 Fixing ring of ceiling cover 2


Cross recessed countersunk head screw,
6 12
M4×12mm, stainless steel, polished

Figure 3-36 Square ceiling cover assembly

3-59
FRU List and Diagram

3.13 Wall Control

3.13.1 FRU List

No. FRU number Description Remark

1 115-021689-00 Wall control (one lighthead)

2 115-021709-00 Wall control (one lighthead, camera)

3.13.2 Typical Diagram

No. P/N Description Qty Remark

1 / Wall control panel sticker (3 lightheads + camera) 1

2 Screw cap 2
Cross recessed countersunk head screw,
3 2
M3×10mm, galvanized
4 Lighthead keypad 3

5 Φ3mm plain washer 6

6 115-021709-00 Φ3mm spring washer 6

7 Stud screw, M3×10+8-8, nickel plated 6

8 Camera keypad 1

9 Main control board (wall control) 3


Cross recessed small pan head screw with
10 8
washer, M3×8mm, galvanized
11 / Front cover 1

12 / Back cover 1

3-60
FRU List and Diagram

Figure 3-37 Wall control

3-61
FRU List and Diagram

3.14 LED Mobile Surgical Light

3.14.1 FRU List

No. FRU number Description Remark

1 115-007786-00 Spring arm assembly

2 115-021681-00 Power module

3 115-014813-00 Battery assembly

3.14.2 Typical Diagram

3.14.2.1LED Mobile Surgical light

No. FRU number Description Qty Remark

1 / HyLED 8600M mobile LED surgical light 1

2 115-007786-00 Spring arm assembly 1

3 / Braking screw for the gimbal joint 2


Hexagon socket countersunk head screw,
4 / 4
M5×12mm, stainless steel, polished
5 / Limit screw 1

6 / Pole 1

7 / Φ6mm plain washer 2

8 / Φ6mm spring washer 2


Hexagon socket head cap screw, M6×16mm,
9 / 2
stainless steel, passivated
10 / Base 1

11 / Φ12mm plain washer 3

12 / Φ12mm spring washer 3

13 / Hexagon socket head cap screw, M12×25mm 3


115-021681-00 +
14 Power module 1
115-014813-00

3-62
FRU List and Diagram

15 / M10mm hexagon nut 2

16 / Balancing weight 1

17 / Φ10mm plain washer 2

18 / Φ10mm spring washer 2


Hexagon socket head cap screw, M10×40mm,
19 / 2
galvanized

Figure 3-38 LED mobile surgical light

3-63
FRU List and Diagram

3.14.2.2Power Module of LED Mobile Surgical Light

No. FRU number Description Qty Remark


Cross recessed countersunk head screw,
1 / 3
M5×10mm, stainless steel, passivated
2 / Back cover 1
Cross recessed countersunk head screw,
3 6
M4×8mm, stainless steel, passivated
Cross recessed small pan head combination
4 7
115-021681-00 screw, M3×8mm, galvanized
5 Power board 1

6 Power indicator board 1

7 / Battery assembly 1

8 / Front cover 1

9 / Power module label 1

10 / Main switch 1

11 / Power connector 1
Cross recessed countersunk head screw,
12 / 2
M3×8mm, galvanized

Figure 3-39 Power module of LED mobile surgical light

3-64
FRU List and Diagram

3.15 Other FRUs

No. FRU number Description Remark

1 115-021642-00 Video power module

2 115-017757-00 Video box assembly New hardware


Video box assembly (for integrated
3 115-021696-00 Old hardware
camera)
Video box assembly (for carrier-arm
4 115-004329-00 Old hardware
camera)
5 115-021713-00 Power module Old hardware

6 115-021714-00 Backup power module Old hardware

7 115-021640-00 Power module assembly New hardware:


The assembly
(115-021640-00) contains
8 115-020469-00 Power module a power module
(115-020469-00) and
M4×12mm fixing screw.
9 115-021641-00 Backup power module assembly New hardware:
The assembly
(115-021641-00) contains
10 115-020470-00 Backup power module a backup power module
(115-020470-00) and
M4×12mm fixing screw.
11 115-020471-00 Backup power supply system
Remote control module (one
12 115-021636-00
lighthead)
Remote control module (two
13 115-021637-00
lightheads)
Remote control module (three
14 115-021638-00
lightheads)
The spare parts contain
15 115-014806-00 Standard spare parts (HyLED series) common screws and screw
caps.
16 A01-000018--- Glue

17 023-000114-00 Infrared remote control

18 M07-00104F--- Fuse

3-65
This page is intentionally blank.
4 Replacement

 Make sure the mains supply has been disconnected before replacing any
component.

4.1 Tools

Table 4-1 Tools needed

Tool Qty Model/Type

Lift 1 Not specified

Electric saw 1 Not specified

Measuring tape 1 5m

Hammer 1 Not specified

Torque wrench 1 Up to 250Nm

Drill 1 Not specified

Spirit level 1 Magnetic with horizontal and vertical bubbles


2mm/2.5mm/3mm/4mm/5mm/8mm (one for each
Allen key 1 set
type)
Open-end wrench 2 24mm

Adjustable wrench 1 0-25mm

Socket wrench 1 5mm/8mm

Flat-bladed screwdriver 1 5mm (longer than 200mm)

Philips screwdriver 1 5mm (longer than 200mm)

Lifting equipment / /

Knife 1 Not specified

Tweezers 1 Not specified

Multimeter 1 Not specified

4-1
Replacement

4.2 HyLED 9700/9500/9500M Lighthead Replacement

4.2.1 Replacing Sterilizable Handle


1. Grab the lighthead with one hand.

2. Press the button at the bottom of the sterilizable handle and pull it out.

Central handle

Sterilizable handle

Button

Figure 4-1 Removing the sterilizable handle

3. Slide a new sterilizable handle onto the central handle of the lighthead. Push the
sterilizable handle upward until it clicks. Pull the sterilizable handle downward to make
sure it is not loose.

Figure 4-2 Reinstalling a new sterilizable handle

 The sterilizable handle is to be used by sterile personnel only.

 Be sure to sterilize the sterilizable handle each time before using it.

4-2
Replacement

4.2.2 Replacing Encircling Handles


1 Place the lighthead as the figure below shows. Remove the 16 screw caps and then use
a Philips screwdriver to remove the 16 M3×16mm cross recessed pan head screws.

Screw cap

M3×16mm
cross recessed
pan head screw
Figure 4-3 Removing the screws

2 Remove the 4 encircling handles.

Encircling
handle

Figure 4-4 Removing encircling handles

3 Reinstall the new encircling handle(s) by following the above instructions in reverse
order.

4.2.3 Replacing Glass


1. Remove the sterilizable handle and the encircling handles as instructed in 4.2.1 and
4.2.2 . Place the lighthead as the figure below shows.

4-3
Replacement

Figure 4-5 Removing encircling handles

2. Use a 4mm Allen key to remove the 2 M3mm expansion screws and remove the central
handle.

M3mm expansion
screw

Central handle

Figure 4-6 Removing the central handle

3. Use a Philips screwdriver to remove the 3 M4×12mm cross recessed countersunk head
screws and remove the central cover.

M4×12mm cross recessed


Central cover
countersunk head screw

Figure 4-7 Removing the central cover

4. Use a Philips screwdriver to remove the 4 M4×20mm cross recessed countersunk head
screws and remove the handle seat.

4-4
Replacement

M4×20mm cross recessed


Handle seat
countersunk head screw

Figure 4-8 Removing the handle seat

 Be sure to keep all the screws and removed components well for later
reinstallation.

5. Use a Philips screwdriver to remove the 4 M3×12mm cross recessed small pan head
screws with washers and remove the damaged glass.
M3×12mm cross
recessed small pan head
Glass screw with washer

Figure 4-9 Removing the glass

6. Reinstall a new glass and other removed components by following the above
instructions in reverse order.

 During the reinstalling process, be sure to protect the new glass from
damage. Do not directly touch the glass with your hands. Otherwise, it may
become dirty.

4-5
Replacement

4.2.4 Replacing AICS Sensor


1. Place the lighthead as the figure below shows.

Figure 4-10 Placing the lighthead

2. Remove the 3 screw caps. Use a Philips screwdriver to remove the 3 M4×12mm cross
recessed countersunk head screws and remove the top cover.
Screw cap
M4×12mm cross recessed
Top cover countersunk head screw

Figure 4-11 Removing the top cover

3. Use a Philips screwdriver to remove the 4 M3×8mm cross recessed countersunk head
screws and remove the metal plate.
M3×8mm cross recessed
Metal plate countersunk head screw

Figure 4-12 Removing the metal plate

4-6
Replacement

 Be sure to keep all the screws and screw caps well for later reinstallation.

4. Remove the black pads.

Black pad

Figure 4-13 Removing the black pads

5. Remove the sterilizable handle and the encircling handles as instructed in 4.2.1 and
4.2.2 .

6. Remove the glass together with the plastic cover as instructed in 4.2.3 .

Glass and
plastic cover

Figure 4-14 Removing the glass and plastic cover

 Be sure to protect the removed glass and plastic cover from damage and
dirty.

4-7
Replacement

7. Disconnect the AICS sensor connection wire connected to the drive board. Use a Philips
screwdriver to remove the 2 M3×10mm cross recessed small pan head screws and
remove the AICS sensor.
M3×10mm cross recessed small
pan head screw

AICS sensor

Figure 4-15 Removing the AICS sensor

8. Reinstall the AICS sensor(s) and other components by following the above instructions
in reverse order.

4.2.5 Replacing LED Unit


1. Remove the glass and the plastic cover as instructed in 4.2.3 .
2. Disconnect the damaged LED unit connection. Use a Philips screwdriver to remove the
2 M3×10mm cross recessed small pan head combination screws. Remove the washer
of the LED board and the damaged LED unit.

M3×10mm cross recessed


small pan head
combination screw
Washer of the LED
LED unit board

Figure 4-16 Removing the LED unit

4-8
Replacement

3. Reinstall the new LED unit(s) and other components by following the above instructions
in reverse order.

 Please refer to 6.1.1 , 6.1.2 , 6.1.3 , 6.1.4 , 6.1.5 or 6.1.6 for LED unit
reconnection.

4.2.6 Replacing LED Drive Board


1. Remove the top cover, metal plate and the black pads as instructed in steps 1, 2, 3 and 4
of 4.2.4 .
2. Disconnect the cables connected to the LED drive board. Use a Philips screwdriver to
remove 3 M3×8mm cross recessed pan head screws and then remove the LED drive
board.

M3×8mm cross recessed


pan head screw
LED drive board

Figure 4-17 Removing the LED drive board

3. Reinstall a new drive board by following the above instructions in reverse order.

 Please refer to 6.2.1 or 6.2.2 for the drive board reconnection.

4-9
Replacement

4.3 HyLED 8600/8600M Lighthead Replacement

4.3.1 Replacing Sterilizable Handle


1. Grab the lighthead with one hand.

2. Press the button at the bottom of the sterilizable handle and pull it out.

Central handle

Sterilizable handle

Button

Figure 4-18 Removing the sterilizable handle

3. Slide a new sterilizable handle onto the central handle of the lighthead. Push the
sterilizable handle upward until it clicks. Pull the sterilizable handle downward to make
sure it is not loose.

Figure 4-19 Reinstalling a new sterilizable handle

 The sterilizable handle is to be used by sterile personnel only.


 Be sure to sterilize the sterilizable handle each time before using it.

4-10
Replacement

4.3.2 Replacing Encircling Handles


1. Place the lighthead as the figure below shows. Remove the 12 screw caps and then use
a Philips screwdriver to remove the 12 M3×16mm cross recessed pan head screws.

Screw cap

M3×16mm
cross recessed
pan head screw
Figure 4-20 Removing screws

2. Remove the 3 encircling handles.

Encircling
handle

Figure 4-21 Removing encircling handles

3. Reinstall the new encircling handle(s) by following the above instructions in reverse
order.

4-11
Replacement

4.3.3 Replacing Glass


1. Remove the sterilizable handle and encircling handles as instructed in 4.3.1 and 4.3.2 .
Place the lighthead as the figure below shows.

Figure 4-22 Removing encircling handles

2. Use a 4mm Allen key to remove the 2 M3mm expansion screws and remove the central
handle.

M3mm expansion
screw

Central handle

Figure 4-23 Removing the central handle

3. Use a Philips screwdriver to remove the 3 M4×12mm cross recessed countersunk head
screws and remove the central cover.

4-12
Replacement

M4×12mm cross recessed


Central cover
countersunk head screw

Figure 4-24 Removing the central cover

4. Use a Philips screwdriver to remove the 4 M4×20mm cross recessed countersunk head
screws and remove the handle seat.

M4×20mm cross recessed


countersunk head screw Handle seat

Figure 4-25 Removing the handle seat

 Be sure to keep all the screws and removed components well for later
reinstallation.

5. Use a Philips screwdriver to remove the 3 M3×12mm cross recessed small pan head
screws with washers and remove the damaged glass.

4-13
Replacement

M3×12mm cross
recessed small pan head
screw with washer

Glass

Figure 4-26 Removing the glass

6. Reinstall a new glass and other removed components by following the above
instructions in reverse order.

 During the reinstalling process, be sure to protect the new glass from
damage. Do not directly touch the glass with your hands. Otherwise, it may
become dirty.

4.3.4 Replacing LED Unit


1. Remove the glass and the plastic cover as instructed in 4.3.3 .

Glass and plastic cover

Figure 4-27 Removing the glass and plastic cover

4-14
Replacement

2. Disconnect the damaged LED unit connection. Use a Philips screwdriver to remove the
2 M3×10mm cross recessed small pan head combination screws. Remove the washers
of the LED board and the damaged LED unit.

M3×10mm cross recessed


small pan head combination
screw

Washer of the LED


board LED unit

Figure 4-28 Removing the LED unit

3. Reinstall the new LED unit(s) and other components by following the above instructions
in reverse order.

 Please refer to 6.1.7 for LED unit reconnection.

4-15
Replacement

4.3.5 Replacing LED Drive Board


1. Place the lighthead as the figure below shows.

Figure 4-29 Placing the lighthead

2. Remove the 3 screw caps. Use a Philips screwdriver to remove the 3 M4×12mm cross
recessed countersunk head screws and remove the top cover.

Screw cap

M4×12mm cross recessed


Top cover countersunk head screw

Figure 4-30 Removing the top cover

3. Use a Philips screwdriver to remove the 4 M3×8mm cross recessed countersunk head
screws and remove the metal plate.

4-16
Replacement

M3×8mm cross recessed


Metal plate
countersunk head screw

Figure 4-31 Removing the metal plate

 Be sure to keep all the screws and screw caps well for later reinstallation.

4. Remove the black pads.

Black pad

Figure 4-32 Removing the black pads

5. Disconnect the cables connected to the LED drive board. Use a Philips screwdriver to
remove the 3 M3×8mm cross recessed pan head screws and then remove the LED drive
board.

4-17
Replacement

M3×8mm cross recessed


pan head screw
LED drive board

Figure 4-33 Removing the LED drive board

6. Reinstall a new drive board by following the above instructions in reverse order.

 Please refer to 6.2.3 for the drive board reconnection.

4-18
Replacement

4.4 HyLED 760/760M/730/730M Lighthead


Replacement

4.4.1 Replacing Sterilizable Handle


1. Grab the lighthead with one hand.

2. Press the button at the bottom of the sterilizable handle and pull it out.

Central handle

Sterilizable handle

Button

Figure 4-34 Removing the sterilizable handle

3. Slide a new sterilizable handle onto the central handle of the lighthead. Push the
sterilizable handle upward until it clicks. Pull the sterilizable handle downward to make
sure it is not loose.

Figure 4-35 Reinstalling a new sterilizable handle

4-19
Replacement

 The sterilizable handle is to be used by sterile personnel only.

 Be sure to sterilize the sterilizable handle each time before using it.

4.4.2 Replacing Outer Handle


1. Place the lighthead as the figure below shows. Remove the screw cap and use a Philips
screwdriver to remove the M3×16mm cross recessed pan head screw.

Screw cap

M3×16mm
cross recessed
pan head screw

Figure 4-36 Removing the screw

2. Carefully turn the lighthead upside-down. Turn the outer handle clockwise and remove it.

Outer handle

Figure 4-37 Removing the outer handle

 Excise caution when turning the lighthead upside-down.

3. Reinstall the outer handle by following the above instructions in reverse order.

4-20
Replacement

4.4.3 Replacing Glass


1. Remove the sterilizable handle and the outer handle as instructed in 4.4.1 and 4.4.2 .
Place the lighthead as the figure below shows.

Figure 4-38 Removing the sterilizable handle and outer handle

2. Use a 4mm Allen key to remove the 2 M3mm expansion screws and remove the central
handle.

M3mm expansion
screw

Central handle

Figure 4-39 Removing the central handle

3. Use a Philips screwdriver to remove the 3 M4×12mm cross recessed countersunk head
screws and remove the central cover.

4-21
Replacement

M4×12mm cross
recessed countersunk Central cover
head screw

Figure 4-40 Removing the central cover

4. Use a Philips screwdriver to remove the 4 M4×20mm cross recessed countersunk head
screws and remove the handle seat.

M4×20mm cross
recessed countersunk Handle seat
head screw

Figure 4-41 Removing the handle seat

 Be sure to keep all the screws and removed components well for later
reinstallation.

5. Use a Philips screwdriver to remove the 10 M3×8mm cross recessed small pan head
combination screws and remove the damaged glass.

4-22
Replacement

M3×8mm cross recessed


small pan head combination
screw

Glass

Figure 4-42 Removing the glass

6. Reinstall a new glass and other removed components by following the above
instructions in reverse order.

 During the reinstalling process, be sure to protect the new glass from
damage. Do not directly touch the glass with your hands. Otherwise, it may
become dirty.

4.4.4 Replacing LED Unit


1. Remove the glass as instructed in 4.4.3 . Remove the outer sealing strip.

Outer sealing strip

Figure 4-43 Removing the outer sealing strip

4-23
Replacement

2. Remove the plastic cover.

Plastic cover

Figure 4-44 Removing the plastic cover

3. Disconnect the damaged LED unit connection. Use a Philips screwdriver to remove the
2 M3×10mm cross recessed small pan head combination screws. Remove the washers
of the LED board and the damaged LED unit.

M3×10mm cross recessed small


pan head combination screw

LED unit
Washer of the LED board

Figure 4-45 Removing the LED unit

4. Reinstall the new LED unit(s) and removed components by following the above
instructions in reverse order.

4-24
Replacement

 Please refer to 6.1.9 and 6.1.10 for LED unit reconnection.

4.4.5 Replacing LED Drive Board


1. Remove the sterilizable handle, outer handle, glass and plastic cover as instructed in
4.4.1 , 4.4.2 and 4.4.3 .
2. Disconnect the cables connected to the LED drive board. Use a Philips screwdriver to
remove 4 M3×6mm cross recessed pan head screws. Remove the LED drive board.

M3×6mm cross recessed


pan head screw

LED drive board

Figure 4-46 Removing the LED drive board

3. Reinstall a new drive board and other removed components by following the above
instructions in reverse order.

 Please refer to 6.2.5 for the drive board reconnection.

4-25
Replacement

4.5 Replacing Control Panel

4.5.1 Replacing Front Cover Assembly


1. Use a Philips screwdriver to remove the 2 M3×8mm cross recessed small pan head
combination screws fixing the back cover of the control panel.

M3×8mm cross recessed


small pan head
combination screws
Figure 4-47 Removing the fixing screws

2. Remove the back cover.

Back cover

Figure 4-48 Removing the back cover

3. Disconnect all the cables connected to the control board.

4-26
Replacement

Figure 4-49 Disconnecting the cables

4. Remove the 5 M3×8mm cross recessed small pan head combination screws fixing the
front cover assembly. Remove the front cover assembly.

M3×8mm cross recessed


small pan head
combination screw

Figure 4-50 Removing the front cover assembly

5. Reinstall the new front cover assembly and the removed back cover by following the
above instructions in reverse order.

 Please refer to 4.5.3 for the control board reconnection.

4.5.2 Replacing Control Board


1. Remove the back cover and disconnect the cables as instructed in steps 1, 2 and 3 of
4.5.1 .
2. Remove the 4 M3×8mm cross recessed pan head screws fixing the control board.
Remove the control board.

4-27
Replacement

M3×8mm cross recessed


pan head screw

Figure 4-51 Removing the control board

3. Reinstall the new control board and other removed components by following the above
instructions in reverse order.

 Please refer to 4.5.3 for the control board reconnection.

4.5.3 Internal Wiring of Control Panel

Please refer to the figures below to connect the cables.

Figure 4-52 Control panel for HyLED 8600/8600M/9500/9700/9500M (old hardware)

4-28
Replacement

Figure 4-53 Control panel for HyLED 9500/9700/9500M (new hardware)

Figure 4-54 Control panel for HyLED 730/760/730M/760M

Figure 4-55 Control panel for HyLED 8600 (New)

4-29
Replacement

4.6 Replacing Touch Control Panel

4.6.1 Replacing Front Cover Assembly


1. Use a Philips screwdriver to remove the 2 M3×8mm cross recessed small pan head
combination screws fixing the back cover of the touch control panel.

M3×8mm cross recessed


small pan head combination
screws
Figure 4-56 Removing the fixing screws

2. Lift the back cover slightly and push it forward and remove it.

Figure 4-57 Removing the back cover

3. Disconnect all the cables connected to the control board.

4-30
Replacement

Figure 4-58 Disconnecting the cables

4. Remove the 6 M3×8mm cross recessed small pan head combination screws fixing the
front cover assembly. Remove the front cover assembly.

M3×8mm cross recessed


small pan head combination
screws

Figure 4-59 Removing the front cover assembly

5. Reinstall the new front cover assembly and the removed back cover by following the
above instructions in reverse order.

 Please refer to 4.6.4 for the control board reconnection.

4.6.2 Replacing Control Board


1. Remove the back cover and disconnect the cables as instructed in steps 1, 2 and 3 of
4.6.1 .
2. Remove the 2 M2.5×6mm cross recessed small pan head screws fixing the CPU module.
Remove the CPU module.

4-31
Replacement

M2.5×6mm cross recessed


small pan head screw

CPU module

Figure 4-60 Removing the CPU module

3. Loosen the pins fixing the flat cable. Remove the flat cable.

Pin

Figure 4-61 Removing the flat cable

4. Remove the M2.5×7+6-6 stud screws with a 5mm socket wrench and then remove the 4
M3×6mm cross recessed pan head screws fixing the control board. Remove the
damaged control board.

4-32
Replacement

M3×6mm cross recessed


pan head screw

M2.5×7+6-6 stud screws

Figure 4-62 Removing the control board

5. Reinstall the new control board and other components by following the above
instructions in reverse order.

 Please refer to 4.6.4 for the control board reconnection.

4.6.3 Replacing Touch Screen


1. Remove the back cover, CPU module and the control board as instructed in steps 1, 2, 3
and 4 of 4.6.2 .
2. Remove the 2 M3×6mm cross recessed countersunk head screws fixing the support.
Remove the support.

Touch screen

Support

M3×6mm cross recessed


countersunk head screw

Figure 4-63 Removing the support

3. Remove the damaged touch screen.


4. Reinstall the new touch screen and other components by following the above

4-33
Replacement

instructions in reverse order.

 Please refer to 4.6.4 for the control board reconnection.

4.6.4 Internal Wiring of Touch Control Panel

Please refer to the figures below to connect the cables.

For integrated
camera use

Figure 4-64 Touch control panel for HyLED 9500/9700/9500M (old hardware)

For integrated
camera use

Figure 4-65 Touch control panel for HyLED


9500/9500M/9700/8600(new)/730/730M/760/760M (new hardware)

4-34
Replacement

4.7 Replacing Gimbal Joint

 When removing the lighthead from the spring arm, be sure to secure the
spring arm to the horizontal position. Ask an extra helper to hold the spring
arm and slightly release the spring arm to the horizontal position after the
lighthead is removed. Otherwise, the spring arm may jump up suddenly and
cause personal injury.

4.7.1 Replacing Vertical Gimbal Joint and Horizontal Gimbal

Joint
1. Secure the spring arm to the horizontal position. Remove the M3×4mm cross recessed
pan head screw fixing the sleeve.

M3×4mm cross recessed


pan head screw

Sleeve

Figure 4-66 Removing the fixing screw

2. Slide the sleeve up and remove the securing segment. Then remove the lighthead.

4-35
Replacement

Securing segment

Figure 4-67 Removing the securing segment

Figure 4-68 Removing the lighthead

 Excise caution when removing the lighthead. Do not drop it down.

 Be sure to keep the removed lighthead well.

3. Use a flat-bladed screwdriver to remove the end cover.

4-36
Replacement

End cover

Sealing ring
Figure 4-69 Removing the end cover

 For HyLED 760/730 series, do not forget to remove a sealing ring after
removing the end cover.

4. Use a 5mm Allen key to remove the braking screw connecting the lighthead and the
horizontal gimbal joint. Use a pair of circlip pliers to remove the two circlips.

Braking screw

Circlips

Figure 4-70 Removing the circlips and braking screw

5. Remove the glass and plastic cover as instructed in 4.2 , 4.3 , 4.4 . Disconnect the
cables connected to the drive board.
6. Slightly remove the horizontal and vertical gimbal joints with the connection cables from
the lighthead.

4-37
Replacement

Figure 4-71 Removing the gimbal joints

7. As instructed in 4.5 and 4.6 , open the back cover of the control panel (touch control
panel) and disconnect the cables. Remove the front cover assembly.
8. Reinstall the control panel (touch control panel) to the new gimbal joints by following the
above instructions in reverse order. Reinstall the lighthead.

4.7.2 Replacing Vertical Gimbal Joint and Horizontal Gimbal

Joint (for Lighthead with Integrated Camera)


1. Secure the spring arm to the horizontal position. Remove the M3×4mm cross recessed
pan head screw fixing the sleeve and the securing segment as instructed in steps 1 and
2 of 4.7.1 .

Figure 4-72 Removing the fixing screw and the securing segment

2. Disconnect the cables between the spring arm and the gimbal joint. Remove the

4-38
Replacement

lighthead.

Figure 4-73 Disconnecting the cables

Figure 4-74 Removing the lighthead

 Excise caution when removing the lighthead. Do not drop it down.

 Be sure to keep the removed lighthead well.

3. Replace the gimbal joint as instructed in 4.7.1 and then reinstall the lighthead.

4.7.3 Replacing Gimbal Joint (LCH)


1. Secure the spring arm to the horizontal position. Use a flat-bladed screwdriver to remove
the 2 fixing screws and braking screw. Remove the lighthead.

4-39
Replacement

Fixing screws

Braking screw

Figure 4-75 Removing the fixing screws and braking screw

Figure 4-76 Removing the lighthead

 Excise caution when removing the lighthead. Do not drop it down.

 Be sure to keep the removed lighthead well.

2. Remove the 2 M3×10mm cross recessed countersunk head screws fixing the end cover.
Remove the end cover.

End cover
M3×10mm cross
recessed countersunk
head screws

Figure 4-77 Removing the end cover

3. Use a 5mm Allen key to remove the braking screws and spring screw cover. Use a
Philips screwdriver to remove the M5×10mm cross recessed countersunk head screw.

4-40
Replacement

Braking screws

M5×10mm cross recessed


Spring screw covers
countersunk head screw
Figure 4-78 Removing the braking screw and spring screw cover

4. Remove the back cover of the control panel (touch control panel) and disconnect the
cables connected to the control board.

Back cover

Figure 4-79 Removing the back cover

5. Pull the disconnected cable out of the front cover assembly and then pull them out of the
rotatable control panel (touch control panel) seat. Remove the front cover assembly with
the control panel seat.

Front cover assembly with the


control panel seat
Figure 4-80 Removing the front cover assembly with the control panel seat

 For the control panel/touch control panel as Figure 4-81 and Figure 4-82
show, be sure to loosen the screws on the connector and pull out the wires
before pulling them out of the rotatable control panel seat.

4-41
Replacement

Connector

Figure 4-81 HyLED 8600/8600M/9500/9700/9500M control panel (old hardware)

Connector

Figure 4-82 HyLED 9500/9700/9500M touch control panel (old hardware)

6. Use a pair of circlip pliers to remove the two circlips. Remove the gimbal joint.

Circlips

Figure 4-83 Removing circlips

Figure 4-84 Removing the gimbal joint

4-42
Replacement

7. Reinstall the control panel (touch control panel) to the new gimbal joint by following the
above instructions in reverse order. Reinstall the lighthead.

4.7.4 Replacing Gimbal Joint for Lighthead with Integrated

Camera (LCH)
1. Secure the spring arm to the horizontal position. Remove the two fixing screws and the
braking screw as instructed in step 1 of 4.7.3 .

Fixing screws

Braking screw

Figure 4-85 Removing the fixing screws and braking screw

2. Remove the lighthead in the direction as the figure below shows. Remove the end cover
and disconnect the cables. Remove the lighthead.

End cover

Disconnecting the cables

Figure 4-86 Removing the end cover and disconnecting the cables

Figure 4-87 Removing the lighthead

4-43
Replacement

 Excise caution when removing the lighthead. Do not drop it down.

 Be sure to keep the removed lighthead well.

3. Replace the gimbal joint as instructed in 4.7.3 and then reinstall it.

4-44
Replacement

4.8 Replacing Spring Arm

 When removing the lighthead from the spring arm, be sure to secure the
spring arm to the horizontal position. Ask an extra helper to hold the spring
arm and slightly release the spring arm to the horizontal position after the
lighthead is removed. Otherwise, the spring arm may jump up suddenly and
cause personal injury.

4.8.1 Replacing Standard/LCH Spring Arm


1. Remove the lighthead as instructed in steps 1 and 2 of 4.7.1 or step 1 of 4.7.3 .
2. Remove the 4 M5×12mm hexagon socket countersunk head screws and the 2 braking
screws fixing the spring arm and the swivel arm. Slightly pull the spring arm down.

Braking screws

M5×12mm hexagon socket


countersunk head screw

Figure 4-88 Removing the fixing screws

3. Remove the 2 M3×25mm cross recessed small pan head screws fixing the cover, and
pull the 3-pole plug out of the spring arm.

4-45
Replacement

M3×25mm cross recessed


small pan head screws

3-pole plug Cover

Figure 4-89 Removing the screws fixing the cover

4. Reinstall the new spring arm and removed lighthead by following the above instructions
in reverse order.

4.8.2 Replacing Standard/LCH Wired Spring Arm


1. Remove the lighthead as instructed in steps 1 and 2 of 4.7.2 or steps 1 and 2 of 4.7.4 .
2. Remove the 4 M5×12mm hexagon socket countersunk head screws and the 2 braking
screws fixing the spring arm and the swivel arm. Slightly pull the spring arm down.

Braking screws

M5×12mm hexagon socket


countersunk head screw

Figure 4-90 Removing the fixing screws

3. Disconnect the cables and remove the spring arm.

4-46
Replacement

Figure 4-91 Removing the spring arm

4. Reinstall the new spring arm and removed lighthead by following the above instructions
in reverse order.

4.8.3 Replacing Bearing Seat of Spring Arm


1. Remove the lighthead and the spring arm as instructed in 4.8.1 or 4.8.2 .Use a pair of
circlip pliers to remove the circlip.

Circlip

Figure 4-92 Removing the circlip

2. Remove the washer.

Washer

Figure 4-93 Removing the washer

3. Remove the damaged bearing seat and install the new bearing seat.

Bearing seat

Figure 4-94 Replacing the bearing seat

4-47
Replacement

4. Reinstall the spring arm and removed lighthead by following the above instructions in
reverse order.

4-48
Replacement

4.9 Replacing Swivel Arm Assembly


1. Remove the lighthead and the spring arm as instructed in 4.8 .

2. Disconnect all the cables above the flange tube. Remove the 8 M8×20mm hexagon
socket countersunk head screws fixing the swivel arm assembly.

M8×20mm hexagon
socket countersunk
head screw

Figure 4-95 Removing fixing screws

3. Carefully remove the swivel arm assembly.

Figure 4-96 Removing the swivel arm assembly

4-49
Replacement

 The swivel arm assembly is heavy and high installed. Do not remove it
alone. Always use extra helpers and a lift for removing and exercise caution
during the whole process.

4. Reinstall the new swivel arm assembly, spring arm and lighthead by following the above
instructions in reverse order.

4-50
Replacement

4.10 Replacing Terminal Block/Power Module/Backup


Power Module (optional)/Video Power Module
(optional)/Video Box (optional)/Remote Control
Module (optional)
1. Disconnect the cables connected to the terminal block/power module/backup power
module (optional)/video power module (optional)/video box (optional)/remote control
module (optional).
2. Remove the M4×12mm cross recessed pan head combination screw fixing the
protective earth wire.

3. Remove the M4×12mm cross recessed pan head combination screws fixing them.

4-51
Replacement

Power module/backup
power module (optional)
M4×12mm cross
recessed pan head
combination screw Protective earth wire

Remote control module


Video power module

Protective earth wire


Protective
earth wire

Video box

Terminal block

Protective earth wire

Power module/backup
power module (optional)

Figure 4-97 Removing the terminal block/power module/backup power module


(optional)/video power module (optional)/video box (optional)/remote control module
(optional)

4. Reinstall them by following the above instructions in reverse order.

 Please refer to B of chapter 9 for the electrical reconnection.

4-52
Replacement

4.11 Replacing Backup Battery Module

4.11.1 Replacing Charging Board


1. Remove the 4 M4×8mm cross recessed countersunk head screws on the battery
module.
M4×8mm cross recessed countersunk head screws

M4×8mm cross recessed


countersunk head screws
Figure 4-98 Removing the fixing screws

2. Remove the upper cover assembly and disconnect the cables connected to the
batteries.

Upper cover assembly

Figure 4-99 Removing the upper cover assembly

3. Remove the 4 M3×8mm cross recessed small pan head combination screws fixing the
charging board. Replace the charging board.

M3×8mm cross recessed small


pan head combination screws

Figure 4-100 Replacing the charging board

4-53
Replacement

4. Reinstall the battery module by following the above steps in reverse order.

4.11.2 Replacing Battery


1. Remove the upper cover assembly as instructed in steps 1 and 2 of 4.11.1 .
2. Remove the 6 M5×12mm cross recessed pan head combination screws fixing the metal
bar.

M5×12mm cross recessed pan head


combination screw

Metal bar

Figure 4-101 Removing the screws fixing the metal bar

3. Remove and replace the battery.

Figure 4-102 Removing the battery

4. Reinstall the battery module by following the above steps in reverse order.

4-54
Replacement

4.12 Replacing Infrared Remote Control Board


1. Remove the 3 M4×8mm cross recessed countersunk head screws fixing the column
bottom cover and then remove the cover.

M4×8mm cross recessed


countersunk head screws
Figure 4-103 Removing the cover

2. Disconnect the cable connected to the infrared remote control board.


3. Remove the 4 M3×5mm cross recessed pan head screws fixing the infrared remote
control board. Remove the control board.
M3×5mm cross recessed pan
head screws

Figure 4-104 Removing the fixing screws

4. Reinstall a new infrared remote control board by following the above instructions in
reverse order.

4-55
Replacement

4.13 Replacing Control Board of Carrier-arm Camera


1. Remove the M4×8mm cross recessed countersunk head screw at the bottom, and
slightly remove the cover with the control board.

Cover with control


board
M4×8mm cross recessed
countersunk head screw

Figure 4-105 Removing the cover with the control board

2. Disconnect all the cables connected to the control board. To disconnect the flat cable,
loosen the fixing pins and then slightly remove the cable.

Flat cable

Figure 4-106 Disconnecting the cables

4-56
Replacement

3. Remove the 3 M3×5mm cross recessed pan head screws fixing the control board and
then remove the control board.

M3×5mm cross recessed


pan head screw

Control board

Figure 4-107 Removing the control board

4. Reinstall the new control board by following the above instructions in reverse order.

Micro-switch cap

Figure 4-108 Micro-switch cap

 Make sure the micro-switch caps are properly positioned before installing
the new control board.
 Be sure to reconnect the flat cable correctly and secure it in position.

Figure 4-109 Flat cable reconnection

4-57
Replacement

4.14 Replacing Wall Control Board


1. Remove the two screw caps and then use a Philips screwdriver to remove the 2
M3×10mm cross recessed countersunk head screws.

M3×10mm cross
recessed countersunk
head screws

Screw caps

Figure 4-110 Removing the fixing screws

2. Slide the front cover downwards and then remove it.

Front cover

Figure 4-111 Removing the front cover

3. Disconnect all the wall control cables. Use a Philips screwdriver to remove the 5
M3×8mm cross recessed small pan head screws with washers fixing the camera keypad,
and then remove the keypad.

4-58
Replacement

M3×8mm cross recessed


small pan head screw with
washer

Camera keypad

Figure 4-112 Removing the camera keypad

4. Use a Philips screwdriver to remove the 2 M3×8mm cross recessed small pan head
screws with washers fixing the main control board, and then remove the board. Remove
the other two control boards in the same way.

M3×8mm cross recessed


small pan head screw with
washer

Main control board

Figure 4-113 Removing main control board

5. Use a socket wrench to remove the 2 M3×10+8-8 stud screws and then use a Philips
screwdriver to remove the M3×8mm cross recessed small pan head screw with washer
fixing the lighthead keypad, and then remove the keypad. Remove the other two
keypads in the same way.

4-59
Replacement

M3×8mm cross recessed


small pan head screw with
washer

M3×10+8-8 stud screws

Lighthead keypad

Figure 4-114 Removing the lighthead keypad

6. Reinstall the wall control boards by following the above instructions in reverse order.

4-60
Replacement

4.15 Replacing Batteries of Mobile Surgical Light

 When removing the lighthead from the spring arm, be sure to secure the
spring arm to the horizontal position. Ask an extra helper to hold the spring
arm and slightly release the spring arm to the horizontal position after the
lighthead is removed. Otherwise, the spring arm may jump up suddenly and
cause personal injury.

4.15.1 Replacing Battery


1. Secure the spring arm to the horizontal position and remove the lighthead.

2. Disconnect the power cable connectors.

Connectors

Figure 4-115 Disconnecting power cable connectors

3. Remove the 6 M4×8mm cross recessed countersunk head screws fixing the front cover.

4-61
Replacement

Front cover

M4×8mm cross recessed


countersunk head screw

Figure 4-116 Removing the screws fixing the front cover

4. Pull out the front cover slightly. Remove the M4×6mm cross recessed combination
screw fixing the earth wire. Disconnect all the red and black battery wires. Remove the
front cover.

M4×6mm cross recessed pan


head combination screw

Figure 4-117 Removing the front cover

5. Remove the 3 M5×10mm cross recessed countersunk head screws fixing the battery
assembly to the back cover. Remove the battery assembly.

4-62
Replacement

Back cover

M5×10mm cross recessed


countersunk head screws
Figure 4-118 Removing the battery assembly

6. Remove the M5×12mm cross recessed pan head combination screw fixing the metal bar.
Remove the metal bar and replace the battery or batteries.

M5×12mm cross recessed pan


head combination screw

Metal bar

Figure 4-119 Removing the metal bar

Figure 4-120 Removing the battery

7. Reinstall the power module by following the above instructions in reverse order.

4.15.2 Replacing Power Board


1. Follow the steps as instructed in 4.15.1 to remove the front cover.

4-63
Replacement

Figure 4-121 Removing the front cover

2. Remove the 4 M3×8mm cross recessed small pan head combination screws fixing the
power board and replace it with a new one.

M3×8mm cross recessed small pan


head combination screws

Power board

Figure 4-122 Replacing the power board

3. Reinstall the power module by following the above instructions in reverse order.

4-64
5 Calibration

5.1 Control Panel Calibration

5.1.1 Lighthead Type and Software Illuminance Level

Calibration (HyLED 8600/8600M/9500/9700/9500M Lightheads)

 If the maximum illumination of the lighthead is lower or higher than the


desired illumination, the software illuminance level can be adjusted.

1. The control panels of the HyLED 8600/8600M, 9500/9500M and 9700 lightheads are
shown in the figures below.

MIS lighting

Power switch Illuminance up

Illuminance down Illuminance indicator

Figure 5-1 HyLED 8600/8600M/9500/9500M control panel

5-1
Calibration

AICS/MIS
lighting

Power switch Illuminance up

Illuminance down Illuminance indicator

Figure 5-2 HyLED 9700 control panel

2. Press the “Power switch” button to turn on the lighthead. Press the “Illuminance down”
and “MIS Lighting” (“AICS/MIS lighting” button for HyLED 9700 lighthead) buttons
together and hold for more than 5 seconds to enter the calibration mode. The
illuminance indicator is defined and numbered as the figure below shows.

Lighthead LED illuminance


type level

Figure 5-3 Calibration screen

3. Press the “Illuminance down” button to define the lighthead type.


 The lighthead is recognized as the HyLED 9500/9500M when the first indicator is
on.
 The lighthead is recognized as the HyLED 9700 when the second indicator is on.
 The lighthead is recognized as the HyLED 8600/8600M when the third indicator is
on.

Figure 5-4 HyLED 9500/9500M lighthead

5-2
Calibration

Figure 5-5 HyLED 9700 lighthead

Figure 5-6 HyLED 8600/8600M lighthead

4. Press the “Illuminance up” button to select the software illuminance level. The higher the
level is, the higher the lighthead illuminance is.

The first level

The second level

The third level

The fourth level

The fifth level

5-3
Calibration

The sixth level

The seventh level

Figure 5-7 Software illuminance level

5. Press the “MIS lighting” or “AICS/MIS lighting” (for HyLED 9700 lighthead) button to quit
the calibration mode.

6. Shut off the mains supply.

7. Switch on the mains supply and turn on the lighthead.

5-4
Calibration

5.1.2 Lighthead Type and Software Illuminance Level

Calibration (HyLED 730/730M/760/760M Lightheads)

 If the maximum illumination of the lighthead is lower or higher than the


desired illumination, the software illuminance level can be adjusted.

1. The control panel of the HyLED 730/730M/760/760M lightheads is shown in the figure
below.

MIS lighting

Illuminance indicator

Power switch

Illuminance down Illuminance up

Figure 5-8 HyLED 730/730M/760/760M control panel

2. Press the “Power switch” button to turn on the lighthead. Press the “Illuminance down”
and “MIS lighting” buttons together and hold for more than 5 seconds to enter the
calibration mode. The illuminance indicator is defined and numbered as the figure below
shows.

Lighthead LED illuminance


type level

Figure 5-9 Calibration screen

3. Press the “Illuminance down” button to define the lighthead type.

5-5
Calibration

 The lighthead is recognized as the HyLED 730/730M when the first indicator is on.
 The lighthead is recognized as the HyLED 760/760M when the second indicator is
on.

Figure 5-10 HyLED 730/730M lighthead

Figure 5-11 HyLED 760/760M lighthead

4. Press the “Illuminance up” button to select the software illuminance level. The higher the
level is, the higher the lighthead illuminance is.

The first level

The second level

The third level

The fourth level

The fifth level

5-6
Calibration

The sixth level

The seventh level

Figure 5-12 Software illuminance level

5. Press the “MIS lighting” button to quit the calibration mode.

6. Shut off the mains supply.

7. Switch on the mains supply and turn on the lighthead.

5-7
Calibration

5.1.3 Lighthead Type and Software Illuminance Level

Calibration (HyLED 8600 New Lightheads)

 If the maximum illumination of the lighthead is lower or higher than the


desired illumination, the software illuminance level can be adjusted.

1. The control panel of the HyLED 8600 (new) lighthead is shown in the figure below.

Color
Integrated camera
temperature
adjustment

MIS lighting

Illuminance indicator

Power switch Full illuminance

Illuminance down Illuminance up

Figure 5-13 HyLED 8600 new control panel

2. Press the “Power switch” button to turn on the lighthead.

3. Press the “Illuminance down” and “MIS lighting” buttons together and hold for more than
5 seconds to enter the calibration mode. The illuminance indicator is defined and
numbered as the figure below shows.

5-8
Calibration

Color temperature
1 10
Lighthead type LED illuminance level

Figure 5-14 Calibration screen

4. Press the “Illuminance down” button to define the lighthead type. The lighthead is
recognized as the HyLED 8600 (New) when the first indicator is on.

Figure 5-15 HyLED 8600 (New) lighthead

5. Press the “Illuminance up” button to select the software illuminance level. The higher the
level is, the higher the lighthead illuminance is.

The first level

The second level

The third level

The fourth level

5-9
Calibration

The fifth level

The sixth level

The seventh level

Figure 5-16 Software illuminance level

6. Press the “Full Illuminance” button to choose the color temperature adjustment function.
If the light is on, it means the color temperature function is activated. If off, it means the
function is not activated.
7. Press the “MIS lighting” button to quit the calibration mode.

8. Shut off the mains supply.


9. Switch on the mains supply and turn on the lighthead.

5-10
Calibration

5.1.4 Camera Type Calibration (HyLED 8600 New)

 If the camera system does not match the display, calibrate the camera as
instructed below.

1. Press the “Power switch” button to turn on the lighthead.

2. Press the “Illuminance up” and “MIS lighting” buttons together and hold for more than 5
seconds to enter the calibration mode. The illuminance indicator is defined and
numbered as the figure below shows.

Integrated/Carrier-arm camera
1 10
Camera type Camera property

Figure 5-17 Calibration screen

3. Press the “Illuminance down” button to define the camera type.

Camera type Light Status

HD CCD

SD CCD

5-11
Calibration

4. Press the “Illuminance up” button to define the HD CCD Property.

HD CCD Property Light Status

CCD_HD_ATTR_1080_60HZ

CCD_HD_ATTR_720_60HZ

CCD_HD_ATTR_1080_50HZ

CCD_HD_ATTR_720_50HZ

5. Press the “MIS lighting” button to quit the calibration mode.

6. Shut off the mains supply.

7. Switch on the mains supply and turn on the lighthead.

5-12
Calibration

5.2 Touch Screen Calibration

 If the button fails to work when pressing, calibrate the touch screen as
instructed below.

1. Power off the surgical light. Recover the mains supply and immediately press anywhere
of the touch screen and hold and then move your finger to the center of the crosshair
until the crosshair appears on the top left corner of the screen as the figure below shows.

Figure 5-18 Crosshair appearing on the top left corner

2. Touch the center of the crosshair that appears on the top right corner of the screen.

Figure 5-19 Crosshair appearing on the top right corner

3. Touch the center of the crosshair that appears on the bottom right corner of the screen.

5-13
Calibration

Figure 5-20 Crosshair appearing on the bottom right corner

4. Touch the center of the crosshair that appears on the bottom left corner of the screen.

Figure 5-21 Crosshair appearing on the bottom left corner

5. Touch the center of the crosshair that appears in the central area of the screen to
complete the calibration.

Figure 5-22 Crosshair appearing in the central area

5-14
Calibration

5.3 Touch Control Panel Calibration (New Hardware)

5.3.1 Camera Type Calibration (HyLED 730/760/9500/9700)

 If the camera system does not match the display, calibrate the camera as
instructed below.

1. Power on the surgical light and press the “Power Switch” to turn on the lighthead. Press
the “Advanced Functions” button.

Power Switch Advance Functions

Figure 5-23 Touch control panel

2. Press the “Service Settings” button.

Service Settings

Figure 5-24 Service settings

5-15
Calibration

3. Enter the password “666666”.

Figure 5-25 Entering the password

 The “ ” delete key can be used if the entered password is wrong.

4. Select the camera type and press the “ ” button.

Figure 5-26 Selecting the camera type

5-16
Calibration

5. Press “ ” to save the configuration.

Figure 5-27 Saving the configuration

5.3.2 Lighthead Number and Light System Number

Calibration (HyLED 730/760//9500/9700)

 For HyLED 9500/HyLED 9700 series surgical lights configured with the
SYNC function, follow the steps below to calibrate the lighthead number and
light system number.
 If the lightheads need to be remotely controlled, please refer to 错误!未找到
引用源。to select the light system number and lighthead number.

1. Power on the surgical light and press the “Power Switch” to turn on the lighthead. Press
the “Advanced Functions” button.

Advance Functions

Figure 5-28 Touch control panel

5-17
Calibration

2. Press the “Service Settings” button.

Service Settings

Figure 5-29 Advanced functions

3. Enter the password “666666”.

Figure 5-30 Entering the password

4. Press “Next”.

Figure 5-31 Pressing “Next”

5-18
Calibration

5. Select the lighthead number according to the mark on the swivel arm. The default light

system number is “ ”. Press the “ ” button.

Selecting lighthead number

Selecting light system number

Figure 5-32 Selecting lighthead number and light system number

6. Press “ ” to save the configuration.

Figure 5-33 Saving the configuration

5.3.3 Software Illuminance Level Calibration (HyLED

730/730M/760/760M /9500/9700/9500M)

 If the maximum illumination of the lighthead is lower or higher than the


desired illumination, the software illuminance level can be adjusted.

1. Power on the surgical light and press the “Power Switch” to turn on the lighthead. Press
the “Advanced Functions” button.

5-19
Calibration

Advanced
Power switch
Functions

Figure 5-34 Entering the “Advanced functions”

2. Press the “Service Settings” button to access a password-protected screen. Enter the

password “666666” and then press .

Service Settings

Figure 5-35 Pressing the “Service Settings”

666666

Figure 5-36 Entering the password

3. Press “Next”.

5-20
Calibration

Figure 5-37 Pressing “Next”

4. Press “Next”.

Figure 5-38 Pressing “Next”

5. Press “ ” or “ ” to decrease or increase the software illuminance level and a dialog


box will pop out prompting you that improper illuminance setting may cause patient burn.
Click “ ” to continue to select the software illuminance level.

Figure 5-39 Selecting the software illuminance level

5-21
Calibration

Figure 5-40 Dialog box

6. Click “ ” to save the configuration.

Figure 5-41 Saving the configuration

5-22
Calibration

5.4 Touch Control Panel Calibration (New User


Interface)

5.4.1 Camera Type Calibration

 If the camera system does not match the display, calibrate the camera as
instructed below.

1. Power on the surgical light and press the “Power Switch” to turn on the lighthead. Press
the "Setting control” button.

Service Settings

Setting control

Figure 5-42 Setting control

2. Press the “Service Settings” button and enter the password “666666” to access the
service setting interface.

3. Select the camera type.

5-23
Calibration

Figure 5-43 Selecting the camera type

4. Press " " for three times and press “ ” to save the configuration.

Figure 5-44 Saving the configuration

5.4.2 Lighthead Number and Light System Number

Calibration

 For HyLED 9500/HyLED 9700/HyLED 8600 (New) series surgical lights


configured with the SYNC function, follow the steps below to calibrate the
lighthead number and light system number.

 If the lightheads need to be remotely controlled, please refer to 错误!未找到


引用源。to select the light system number and lighthead number.

1. Power on the surgical light and press the “Power Switch” to turn on the lighthead. Press
the "Setting control” button.

5-24
Calibration

Service Settings

Setting control

Figure 5-45 Setting control

2. Press the “Service Settings” button and enter the password “666666” to access the
service setting interface.

3. Press the " " button.

Figure 5-46 Pressing “ ”

4. Select the lighthead number and light system number according to the mark on the
swivel arm.

5-25
Calibration

Selecting light
system number
Selecting lighthead
number

Figure 5-47 Selecting lighthead number and light system number

5. Press the " " button.

Figure 5-48 Pressing “ ”

6. Press “ ” to save the configuration.

Figure 5-49 Saving the configuration

5-26
Calibration

5.4.3 Software Illuminance Level Calibration

 If the maximum illumination of the lighthead is lower or higher than the


desired illumination, the software illuminance level can be adjusted.

1. Power on the surgical light and press the “Power Switch” to turn on the lighthead. Press
the "Setting control” button.

Service Settings

Setting control

Figure 5-50 Setting control

2. Press the “Service Settings” button and enter the password “666666” to access the
service setting interface.

3. Press the " " button.

Figure 5-51 Pressing “ ”

5-27
Calibration

4. Press the " " button.

Figure 5-52 Pressing “ ”

5. Press “ ” or “ ” to decrease or increase the software illuminance level. And then

press " " button. The higher the level is, the higher the lighthead illuminance is.

Figure 5-53 Selecting the software illuminance level

6. Press “ ” to save the configuration.

5-28
Calibration

Figure 5-54 Saving the configuration

5-29
Calibration

5.5 Carrier-arm Camera Calibration

2 3 4

5 6 7

8 9 10

11 13 12

Figure 5-55 Control panel

1 Press and hold buttons “3” and “11” for more than 5s to calibrate the camera to
“1080i/50”.
2 Press and hold buttons “3” and “12” for more than 5s to calibrate the camera to
“1080i/60”.

5-30
6 Lighthead Internal Wiring

6.1 LED Unit Layout and Wiring

 All the LEDs marked with the same number shall be connected together.
 Be sure to tidy all the connection wires. Make sure the wires do not squeeze
the surface of the lens.

6.1.1 HyLED 9500/9500M LED Unit Layout and Wiring

Figure 6-1 HyLED 9500/9500M LED layout and wiring

6-1
Lighthead Internal Wiring

Short wire (001165)

Short connection wire (001162)

Medium connection wire (001163))

6.1.2 HyLED 9700 LED Unit Layout and Wiring

Figure 6-2 HyLED 9700 LED layout and wiring

Short wire (001165)

Short connection wire (001162)

Medium connection wire (001163)

Long connection wire (001164)

6-2
Lighthead Internal Wiring

6.1.3 HyLED 9500/9500M LED Unit Layout and Wiring (Color

Temperature Adjustment) (Old Hardware)

Figure 6-3 HyLED 9500/9500M LED layout and wiring (with color temperature
adjustment)

Short wire (001165)

Short connection wire (001162)

Medium connection wire (001163)

6-3
Lighthead Internal Wiring

6.1.4 HyLED 9700 LED Unit Layout and Wiring (Color

Temperature Adjustment) (Old Hardware)

Figure 6-4 HyLED 9700 LED layout and wiring (with color temperature adjustment)

Short wire (001165)

Short connection wire (001162)

Medium connection wire (001163)

6-4
Lighthead Internal Wiring

6.1.5 HyLED 9500/9500M LED Unit Layout and Wiring (Color

Temperature Adjustment) (New Hardware)

Figure 6-5 HyLED 9500 LED layout and wiring

Short wire (001165)

Short connection wire (003562)

Medium connection wire (003563)

Long connection wire (003564)

6-5
Lighthead Internal Wiring

6.1.6 HyLED 9700 LED Unit Layout and Wiring (Color

Temperature Adjustment) (New Hardware)

Figure 6-6 HyLED 9700 LED layout and wiring

Short wire (001165)

Short connection wire (003562)

Medium connection wire (003563)

Long connection wire (003564)

6-6
Lighthead Internal Wiring

6.1.7 HyLED 8600 LED Unit Layout and Wiring

Figure 6-7 HyLED 8600 LED layout and wiring

Short wire (001165)

Short connection wire (001162)

Medium connection wire (001163)

6-7
Lighthead Internal Wiring

6.1.8 HyLED 8600 (New) LED Unit Layout and Wiring

Figure 6-8 HyLED 8600 (New) LED layout and wiring

Short wire (001165)

Short connection wire (003562)

Medium connection wire (003563)

Long connection wire (003564)

6-8
Lighthead Internal Wiring

6.1.9 HyLED 730/730M LED Unit Layout and Wiring

Figure 6-9 HyLED 730/730M LED layout and wiring

Short wires (001165)

Short connection wires (001162)

Long connection wires (001164)

Connection wire between LED


unit and drive board

6-9
Lighthead Internal Wiring

6.1.10 HyLED 760/760M LED Unit Layout and Wiring

Figure 6-10 HyLED 760/760M LED layout and wiring

Short wires (001165)

Short connection wires (001162)

Long connection wires (001164)

Connection wire between LED


unit and drive board

6-10
Lighthead Internal Wiring

6.2 LED Drive Board Wiring

6.2.1 HyLED 9500/9700/9500M Drive Board Wiring (Old

Hardware)
1. Connect the 4 8-pin cable connectors respectively to the nearby ports marked LED_A,
LED_B, LED_C, LED_D.

Figure 6-11 Connecting the 8-pin cable

2. Connect the 3-pin cable connectors (for HyLED 9700) from the LED_A side to the port
marked SNR_A, the connector from LED_B side to the port marked SNR_B, the
connector from LED_C side to the port marked SNR_C and the connector from LED_D
side to the port marked SNR_D.

Figure 6-12 Connecting the 3-pin cable

6-11
Lighthead Internal Wiring

3. Connect other cable connectors as the figures below show.

Figure 6-13 Connecting other cables

6-12
Lighthead Internal Wiring

6.2.2 HyLED 9500/9700/9500M Drive Board Wiring (New

Hardware)
1. Connect the 6 8-pin cable connectors respectively to the nearby ports marked LED_A,
LED_B, LED_C, LED_D, LED_AD, LED_BC (For a lighthead with light field adjustment
and without color temperature adjustment, LED_AD and LED_BC ports are not
connected with the 8-pin cable connectors.).
2. Connect the 3-pin cable connectors (for HyLED 9700) from the LED_A side to the port
marked SNR_A, the connector from LED_B side to the port marked SNR_B, the
connector from LED_C side to the port marked SNR_C and the connector from LED_D
side to the port marked SNR_D.

3. Connect other cable connectors as the figures below show.

Figure 6-14 Connecting drive board cables

6.2.3 HyLED 8600/8600M Drive Board Wiring


1. Connect the 3 8-pin cable connectors respectively to the nearby ports marked LED_B,
LED_C, LED_D.

 LED_A is for HyLED 9500/9700/9500M series.

6-13
Lighthead Internal Wiring

Figure 6-15 Connecting the 8-pin cable connector

2. Connect other cable connectors as the figures below show.

Figure 6-16 Connecting other cables

6-14
Lighthead Internal Wiring

6.2.4 HyLED 8600 (New) Drive Board Wiring


1. Connect the 3 8-pin cable connectors respectively to the nearby ports marked LED_B,
LED_C, LED_D.

 LED_A is for HyLED 9500/9700/9500M series.

Figure 6-17 Connecting the 8-pin cable connector

2. Connect other cable connectors as the figures below show.

Figure 6-18 Connecting other cables

6-15
Lighthead Internal Wiring

6.2.5 HyLED 730/730M/760/760M Drive Board Wiring


1. Connect the 2 8-pin cable connectors respectively to the nearby ports.

2. Connect other cable connectors.

Figure 6-19 Connecting drive board cables

6-16
7 Troubleshooting

 Device servicing is to be performed by personnel authorized by Nanjing


Mindray only. Unauthorized device servicing may result in personal injury
and/or damage to the device or other property.
 Device servicing should be strictly based on technical data supplied by
Nanjing Mindray. If you are in need of more technical data, contact the
customer service department authorized by Nanjing Mindray.

7.1 Electrical Schematic Diagram

Power
LIN communication
SPI communication

Terminal
block

Power
module

Control
board

Drive board

LED unit

Figure 7-1 Electrical schematic diagram

7-1
Troubleshooting

7.2 Common Errors

7.2.1 Lighthead Off and Abnormal Lighting


1. Root cause: the power return circuit breaks down.

2. Main cause: terminal block, power module, connector between power module and
control board, control board, drive board

3. Solution:

1) Check if the control panel works normally.

2) Check if the connection cables in the gimbal joint and the spring arm are damaged.

3) Check if the terminal block is broken.


4) Check if the power module is broken.
5) Check if the 3-pole connector between the swivel arm and the spring arm is broken.

6) Check if the connection between the drive board and the control board is broken.

7) Check if the drive board breaks down.

8) Check if the connections between LED units are disconnected or break down.

4. Repair methods:

1) If the lighthead is off,


a) Test the voltage on the control board to judge which circuit (Upper circuit: from
the power module to the control board. Lower circuit: from the control board to
the lighthead) breaks down.
 If the input voltage is abnormal, it means the upper circuit is broken.
 If the input voltage is normal and the output voltage is abnormal, it means
the control board breaks down.
 If the input and output voltage is normal, it means the lighthead breaks
down.

The input voltage is


equal to the output
voltage.

Input voltage:
HyLED 9700: 22-24V
HyLED 9500: 16-26V Output voltage: 5V
HyLED 760/730: 16-26V
Figure 7-2 Control board

7-2
Troubleshooting

The input voltage is


equal to the output
voltage.

Input voltage:
HyLED 9700: 22-24V
HyLED 9500: 16-26V
HyLED 760/730: 16-26V

Figure 7-3 Control board

b) If the upper circuit breaks down, check if the connections in the swivel arm and
the spring arm are broken.
 Check the connections in the spring arm: Remove the end cover of the
vertical gimbal joint. Visually check and slightly drag the 3-pole connector
to see if the connections are broken.
 Check the connections in the swivel arm: Remove the end cover of the
swivel arm and check if the connections are broken.

Check the connections in the spring arm.

Check the connections in the swivel arm.

Figure 7-4 Checking connections

c) If the connections in the swivel arm and the spring arm are well, test the output
voltage of the terminal block (90-240V).
 If the voltage is 0, the fuse module may be broken. Replace the broken
fuse module as instructed in 4.10 .
 If the voltage is normal, check the power module.

7-3
Troubleshooting

Figure 7-5 Testing output voltage of the terminal block

 The output voltage of the terminal block is high. Excise caution when
testing.

d) Check the output voltage of the power module.


 Shut off the mains supply and remove the connector as the figure below
shows. Recover the mains supply and test the voltage of the pins on the
board for red and black wire. Normally, the voltage is 23-26V.
 If the output voltage of the power module is abnormal, it means the power
module is broken. Replace the broken power module as instructed in
4.10 .

Connector

Figure 7-6 Removing the connector

e) If the output voltage of the power module is normal, use a multimeter to test if
the 3-pole connector in the swivel arm is broken. Shut off the mains supply and
remove the connector as the figure below shows. Pull the 3-pole connector out
of the swivel arm. Use a multimeter to test if the red wire, black wire and yellow
wire is respectively shorted or if there is a short circuit among them.

7-4
Troubleshooting

Connector

Figure 7-7 Removing the connector

3-pole connector

Figure 7-8 Pulling the 3-pole connector out

Outside

Middle side

Inner side

Figure 7-9 3-pole connector

 Broken circuit test: use the omh band of a multimeter to test if the
following two ends can be conducted (Resistance<1Ω). If not, the wires
are broken.
 Black wire and the outside of the 3-pole connector
 Yellow wire and the middle side
 Red wire and the inner side
 Short circuit test: use the omh band of a multimeter to test if the following
two ends are conducted. Normally, the resistance is more than 10Ω if the
two ends are not conduced. If conducted, the wires are shorted.
 Black wire and the middle side, black wire and the inner side
 Yellow wire and the outside, yellow wire and the inner side
 Red wire and the outside side, red wire and the middle side

 Buzzer band can be used for easy short or broken circuits test. If the buzzer
sounds, it means conductivity. If not, there may be short or broken circuits.

7-5
Troubleshooting

 Replace the swivel arm assembly if the cables in the central axis are broken.

f) If the lower circuit breaks down, first check if the cables are broken or the
connections are well.
 Test the input voltage (24V and 5V) of the drive board. If the voltage is
normal, it means the drive board is broken. Replace the drive board as
instructed in 4.2.6 ,4.3.5 ,4.4.5 .
 If the input voltage is abnormal, the connections between the control
board and the drive board may be broken. Replace the gimbal joint as
instructed in 4.7 .

Input voltage:
16-26V

Figure 7-10 Drive board (HyLED 9500/9700)

Input voltage:
16-26V

Figure 7-11 Drive board (HyLED 730/760)

Input voltage 1 Input voltage 2


(16-26V) (5V)

Figure 7-12 Drive board (HyLED 8600/9500/9700 old hardware)

7-6
Troubleshooting

7.2.2 Broken LEDs

Input voltage of LED units (of the same number):1-22V

Figure 7-13 Input voltage of LED unit

1. Single LED cannot be lit: The LED breaks down. Replace the broken LED unit as
instructed in 4.2.5 , 4.3.4 , 4.4.4 .
2. A string of LEDs cannot be lit: The drive board breaks down or the broken LEDs cause
the breakdown of the return circuit. Use a multimeter to test the input voltage between
the black and white wires of the same connector. Normally, the voltage is 1-22V. If
abnormal, the drive board may be broken. Replace the broken drive board. If normal,
use the multimeter to test the voltage between the black and white wires until the broken
LED is found.

7.2.3 LIN Communication Error


Follow the steps below to check the error.

1. Test the signal input voltage (between the yellow wire and the black wire). Normally, the
voltage is between 0-12V. If normal, the control board or the power module may break
down. Replace the control board or the power module as instructed in 4.5 ,4.6 or 4.10 .

Voltage between
the yellow and
black wire: 0-12V

Figure 7-14 Testing the signal Input voltage

7-7
Troubleshooting

2. If the signal input voltage is abnormal, the upper communication circuit may break down.
Check if the connections in the swivel arm and the spring arm are broken.
 Check the connections in the spring arm: Remove the end cover of the vertical
gimbal joint. Visually check and slightly drag the 3-pole connector to see if the
connections are broken.
 Check the connections in the swivel arm: Remove the end cover of the swivel arm
and check if the connections are broken.

Check the connections in the spring arm.

Check the connections in the swivel arm.

Figure 7-15 Connections in the spring arm and swivel arm

3. If the connections in the swivel arm and the spring arm are well, the power module may
break down.
 Shut off the mains supply and remove the connector as the figure below shows.
Recover the mains supply and test the voltage of the pins on the board for yellow
and black wire. Normally, the voltage is 0-12V.
 If the output voltage of the power module is abnormal, the power module is broken.
Replace the broken power module. If normal, follow the next step to check.

Connector

Figure 7-16 Power module

4. If the output voltage of the power module is normal, use a multimeter to test if the 3-pole
connector between the swivel arm and the spring arm is broken. Shut off the mains
supply and remove the connector as the figure below shows. Pull the 3-pole connector

7-8
Troubleshooting

out of the swivel arm. Use a multimeter to test if the red wire, black wire and yellow wire
is respectively shorted or if there is a short circuit among them.

Connector

Figure 7-17 Removing the connector

Figure 7-18 Pulling the 3-pole connector out

Outside

Middle side

Inner side

Figure 7-19 3-pole connector

a) Broken circuit test: use the omh band of a multimeter to test if the following two
ends can be conducted (Resistance<1Ω). If not, the wires are broken.
 Black wire and the outside of the 3-pole connector
 Yellow wire and the middle side
 Red wire and the inner side
b) Short circuit test: use the omh band of a multimeter to test if the following two
ends are conducted. Normally, the resistance is more than 10Ω if the two ends
are not conduced. If conducted, the wires are shorted.
 Black wire and the middle side, black wire and the inner side
 Yellow wire and the outside, yellow wire and the inner side
 Red wire and the outside side, red wire and the middle side

 Buzzer band can be used for easy short and broken circuits test. If the
buzzer sounds, it means conductivity. If not, there may be short or broken

7-9
Troubleshooting

circuits.

 Replace the swivel arm assembly if the cables in the central axis are broken.

7.2.4 Alarm of Control Panel/Touch Control Panel

The indicators on the control board and touch control board are shown in the figure below.

Full illuminance indicator

MIS lighting
indicator

Status Indicator Illuminance indicator

Figure 7-20 HyLED 8600/9500 control panel

Full illuminance indicator

MIS lighting
indicator

AICS Indicator

Illuminance indicator
Status Indicator

Figure 7-21 HyLED 9700 control panel

7-10
Troubleshooting

AC indicator
MIS lighting
Battery indicator indicator

Full illuminance
indicator

Illuminance indicator

Figure 7-22 HyLED 730/760 control panel

Error indicator Battery indicator

Figure 7-23 HyLED 730/760/9500/9700 touch control panel

Control Control
Touch
panel panel-Illumi Lighthead
control Error type Solution
-Status nance status
panel
indicator indicator
LIN
communication
error: The
Orange illuminance Solve the error as
Normal
/ and Normal adjustment is instructed in
working
flashing normal, but the 7.2.3 .
wall control and
remote control do
not function.

7-11
Troubleshooting

Check if the
connections
between the
control board and
SPI the drive board
communication are well. If well,
error: The replace the
Red and Unstable lighthead cannot control board
/ Flashing
flashing working be turned on or and/or the drive
the illuminance board. If the error
cannot be cannot be solved,
adjusted. the connections
inside the gimbal
joint may be
broken. Replace
the gimbal joint.
Solve the error as
Red and Normal
/ Normal The voltage is low. instructed in
flashing working
7.2.1 .
LIN
communication
error: The
illuminance Solve the error as
Normal
/ / adjustment is instructed in
working
normal, but the 7.2.3 .
wall control and
remote control do
not function.
Check if the
connections
between the
control board and
SPI the drive board
communication are well. If well,
error: The replace the
Unstable lighthead cannot control board
/ /
working be turned on or and/or the drive
the illuminance board. If the error
cannot be cannot be solved,
adjusted. the connections
inside the gimbal
joint may be
broken. Replace
the lighthead.

7-12
Troubleshooting

Green
and
flashing
The lighthead is
(status Normal
Normal working on the Normal.
indicator); working
backup batteries.
Green
(battery
indicator)

7.2.5 Video System Error

7.2.5.1No Image
1. Root cause: the video power or the video cables break down.

2. Common error location: cables, power module (video power module), video box

3. Solution:

1) Check if the camera functions normally:


a) Check if the power cable of the camera and the connections on the video box
are normal.
b) Press the “Zoom in” or “Zoom out” on the control panel to check if the camera
lens moves.

2) If the camera moves, the video cables may be broken. Test if the video in and video
out cables connected to the video box are shorted or broken.
a) Remove the camera or the lighthead with an integrated camera and check if
the connectors are broken.
b) Remove the spring arm and check if the cable connections are broken.
c) Check if the cables connected to the video box are broken.

3) If the camera does not move, check if the output voltage of the power module or the
video power module is normal. For all the video boxes, the voltage between the red
and black wire of the connector as the figure below shows is 24V. To test the
voltage,
a) Shut off the mains supply and remove the connector.
b) Recover the mains supply and test the voltage between the red and black wire.

7-13
Troubleshooting

Connector

Figure 7-24 Connector

4) If the cables are well, the video box may breaks down. Replace the video box as
instructed in 4.10 .

7.2.5.2Noise/Interference
1. Root cause: The return circuit resistance of the video signal is abnormal.
2. Common error locations: poorly welded wires, cable connections or earth connection,
interference between the light source frequency and the camera frequency

3. Solution:

1) Check if the wires are poorly welded or the connectors are poorly connected.
a) Remove the camera or the lighthead with an integrated camera and check
visually if the connectors are broken or contact poorly.
b) Remove the spring arm and check if the connections are broken or contact
poorly.
c) Check if the cables connected to the video box are broken.

2) Check if the earth connections are well: Make sure the earth connections for the
power module, video power module, video box, and the display and terminal blocks
are well connected and conducted.
3) Check if the light source frequency interferes with that of the camera: If the light
source frequency is not multiple of the default frequency of the camera (50Hz), the
camera image may flash or the ripple may occur. Do not direct other light sources
except Mindray’s lightheads towards the camera.

7-14
8 Communication Protocol

8.1 General

The surgical light system provides three communication ports: RS232, RS485 and Olympus
communication ports.
The remote control module and the touch control panel shall be configured if the remote
control function is chosen.

8-1
Communication Protocol

8.2 Selecting Communication Protocol

8.2.1 Old User Interface of Touch Control Panel


1. Power on the surgical light and press the “Power Switch” to turn on the lighthead. Press
the “Advanced Functions” button.

Power switch Advanced


Functions

Figure 8-1 Entering the “Advanced functions”

2. Press the “Service Settings” button to access a password-protected screen. Enter the

password “666666” and then press “ ”.

Service Settings

Figure 8-2 Service Settings

8-2
Communication Protocol

666666

Figure 8-3 Entering the password

 Use the “ ” key to clear the wrong password.

3. Press “Next”.

Figure 8-4 Pressing “Next”

4. Press “Next”.

Figure 8-5 Pressing “Next”

8-3
Communication Protocol

5. Select the communication protocol and click “ ” to save the configuration.

Figure 8-6 Selecting the communication port

Figure 8-7 Saving the configuration

8-4
Communication Protocol

8.2.2 New User Interface of Touch Control Panel

Follow the steps as instructed in 5.4 to enter the screen as the figure below shows. Select the

type of the communication protocol and click “ ” to save the configuration.

Figure 8-8 Selecting the communication protocol

8-5
This page is intentionally blank.
9 Appendices

A Differences between Old and New Hardware

Figure 9-1 Differences between old and new hardware

1. New hardware: The video power is isolated from the lighthead power. The color
temperature and light field adjustments can be available on a lighthead.
2. Old hardware: The video power and lighthead power are from a common power supply.
The color temperature adjustment and light field adjustment are not available at the
same time.

9-1
Appendices

B Electrical Connection Diagram

B.1 Terminal Block Connection

To mains supply

N N N N N P P L L L L L
E A

Figure 9-2 Terminal block connection

9-2
Appendices

B.2 HyLED 730/760/8600(New)/9500/9700 Series, Mains

Connection

Figure 9-3 HyLED 730/760/HyLED 9500/9700 series, mains connection

9-3
Appendices

B.3 HyLED 730/760/8600(New)/9500/9700 Series with

Mindray’s Backup Power Module

BATTERY

~ ~

~ ~

Figure 9-4 HyLED 730/760/HyLED 9500/9700 series with Mindray’s backup power
module

9-4
Appendices

B.4 HyLED 730/760/8600(New)/9500/9700 Series with

Integrated Camera

L L L N N N

N N
L

C0027

Figure 9-5 HyLED 730/760/HyLED 9500/9700 series with an integrated camera

9-5
Appendices

B.5 HyLED 730/760/8600(New)/9500/9700 Series with

Carrier-arm Camera

L L L N N N

C0027

Figure 9-6 HyLED 730/760/HyLED 9500/9700 series with a carrier-arm camera

9-6
Appendices

B.6 HyLED 8600(New)/9500/9700 Series with SYNC

Function

Figure 9-7 HyLED 9500/9700 series with SYNC function

9-7
MAINS SUPPLY INPUT
N L
HyLED 8600/9500/9700 (Old Hardware) Series, Mains

Figure 9-8 HyLED 8600/9500/9700 (old hardware) series, mains connection


SWIVEL SPRING GIMBAL
ARM ARM JOINT
YE BK RD
LIN
LIGHTHEAD
BU BR RD BK YE 3 2 1 1 2 3
N L LIN
Appendices

LIN
CENTRAL AXIS LIN Node 1
AC IN C0004
C0003

9-8
C0001
1
H8
1
J2 J8 CONTROLLER WALL
H8 MAIN 6 6
CONTROL
POWER
BOARD H8 MAIN POWER
BK = BLACK
BN = BROWN
BU = BLUE
Connection
GN = GREEN
GNYE = GREEN/YELLOW
OG = ORANGE
RD = RED
B.7 WH = WHITE
YE = YELLOW
MAINS SUPPLY INPUT

Figure 9-9 HyLED 8600/HyLED 9500/9700 (old hardware) series with Mindray’s backup
N L
HyLED 8600/9500/9700 (Old Hardware) Series with

SWIVEL SPRING GIMBAL


ARM ARM JOINT
YE BK RD
LIN
LIGHTHEAD
BU BR RD BK YE 3 2 1 1 2 3
N L LIN LIN
AC IN CENTRAL AXIS LIN Node 1

power module
C0004
Appendices

C0001 C0003
H8 MAIN POWER
Mindray’s Backup Power Module

J1 WALL

9-9
1~6 BOARD H8
1~6 J1 1 1 CONTROLLER CONTROL
J2 J8
H8 BACKUP 6 6
POWER BOARD 1 1
J3
J11
6 6
C0016
BAT IN
H8 MINDRAY’S BACK UP
POWER MODULE
1 2 3 4
CHARGING
1 2 3 4 INPUT BK = BLACK
3 BN = BROWN
1
BAT OUT 2 BU = BLUE
2 C0017 1
GN = GREEN
3
4 H8 CHARGING BOARD
GNYE = GREEN/YELLOW
H8 BATTERY MODULE OG = ORANGE
RD = RED
B.8 WH = WHITE
YE = YELLOW
Appendices

B.9 HyLED 8600/9500/9700 (Old Hardware) Series with

Carrier-arm Camera

Figure 9-10 HyLED 8600/9500/9700 (old hardware) series with a carrier-arm camera

9-10
This page is intentionally blank.
P/N: 046-006478-00(2.0)