X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
• HV transformer
• HV multiplier
• HV cable
• X-ray tube
To figure out the bad component the following procedure should be carried out
• Connect the „test adapter“ to X530 on the PCB –B20 (X530=connector on top of PCB)
• Switch kV limit to position ON
• Switch mA clearance to position locked
Step 1
Step 2
• Setpoints for kV : 20 , mA : 5
- Manual Control Command: GV20 , GC5
• Switch the generator to HV ON
- Manual Control Command: SA.. , ..
Step 3
6-2
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Step 4
6-3
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
• Unscrew the 2 red, 1 black wire on the front side of the multiplier
red
black
red
• Check the varistors mounted on the tube side of the heat current cable.
There should be no short cut across the varistors.
The function of the varistors is to carry the discharge current to chassis in case of tube arcing.
WARNING
Take care on the HV cable when replacing the generator. The cable can nip.
6-4
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
⇓
⇒ Switch off generator, connect test adapter to X530
⇒ Test adapter : kV - Limit to " ON " (ea. kV-regulation
is
blocked at minimum)
switch mA release to "locked"
(preheating current)
⇒ Switch S4 to "operation"
⇒ Read heating current (HCU) in the S4-Tools
HCU ≠ 5,8....6,8 HCU = 5,8....6,8
mA = 0 mA > 0
⇒ Tube faulty
⇓
⇒ Switch off generator
⇒ Remove heating cable from plug X75 (rear panel)
⇒ Measure the tube resistance at the removed end of the
cable
6-5
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Diagram 1 (continued)
Low-resistance High-resistance
⇒ Measure resistance at
⇓ pins 1+2 at plug X24 on
inverter –B115
low-resistance high-
resistance
⇒ Secure heating cable to X75 → Connection → replace
again X 75 to X 24 -B115
⇒ Switch test adapter to "act. heat. open circuited
curr.", kV limit to "OFF" and mA
release to "OFF"
No voltage U ≈ 2,5 V
(Generator switches off again)
→
⇓ - replace -B 20
U = Umains = 230 V ∼ U = 0V
6-6
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
kV = 40 mA = 0 kV = 40 mA ≠ 0
⇓
⇒ Switch off generator
⇒ Insert HV-cable again
⇒ Connect test adapter to X530. kV-limit : ON, mA release :
locked
⇒ Potentiometer I - reg : 0 , V - reg : 0
⇒ Switch S4 to operating position
⇒ Select Generator in the S4-tools
kV = 13 - 19 mA = 0 kV ≠ 13 - 19 mA ≠ 0
Generator switches off with
Error Code1 or kV – indication
is pending
6-7
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Diagram 2 (continued)
kV = 20 mA = 0 kV ≠ 20 mA ≠ 0
Generator reacts as above
6-8
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
⇒
⇓ ⇒
Test max. heating current:
Switch off generator
⇒ Connect test adapter
⇒ Set test adapter :
mA - release : clear
Potentiometer I – reg. : I
⇒ Switch generator to heating position
⇒ Observe heating current
⇒ Rotate pot. I-reg. in clockwise
direction
Generator switches Generator does not
off at 7,5 .... 8,5 A switches off or at
incorrect heat.
current
⇒ Check tube ⇒ -B 20 replace
characteristic:
HCU ( 20kV/
50mA)< 11A
6-9
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
R = 100 R ± 1 % R ≠ 100 R ± 1%
6-10
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
→ Temperature monitor of
the inverter triggered
⇓
→ Check fan function and
replace fan if
necessary
→ Branch from diagram 10
⇒ Check fuse F 1 on -B 20 (Integralboard)
⇒ Check plug X 520 on -B 20 (Integralboard)
→ Replace fuse
⇓ → Replace B20 after
repeated faults
6-11
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Diagram 5 (continued)
⎯ X 523 (HV-multiplier)
⎯ X 522 (voltage regulation (B 4) )
⎯ X 524 (S4-internal)
⎯ X 521 (inverter)
A voltage disappears
All voltages present following the connection
of a plug
→ Replace
corresponding
module
⇓
⇒ Check wiring of PTC resistor on inverter:
Disconnect plug X205 from –B100 and measure the
resistance at the end of the cables, correct value: R =
120 .... 480 Ohm
6-12
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
WARNING
The diagram must be run through from the beginning again each time a component is replaced!
Diagram 6
6-13
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Diagram 6 (continued)
U - AK ≈ 125 (125...135 V ) U - AK = 0
U - AK = 0
U - AK ≈ 125 (125-135 V) or
U - AK ≈ 150 V
→ inverter faulty
⇓
U - AK ≈ 120 V U - AK = 0
6-14
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
R = 56 k ± 0,1% R ≠ 56 k ± 0,1%
→ Replace –B650
6-15
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
⇓
⇒ Switch test adapter to "UAK"
⇒ Test adapter: Switch "mA release" to "locked"
Switch "kV - Limit" to "ON"
⇒ Switch generator to operating position and test with following
kV - setpoint (mA-setpoint = 5mA) ;
read voltage at DVM : (1 V = 100 V UAK)
- kV : 20 UAK = 122 V
6-16
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
The diagram shows Imains dependent upon x-ray tube current and tension
(Umains = 230V and f = 50Hz).
U/ I/
kV mA 5 10 20 25 33 50
40 3,8A 11,4A
50 4,2A 10,4A
6-17
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
R = 6 ....10 Ω otherwise
6-18
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Diagram 10
Answer = 1 Answer ≠ 1
→ replace -B 20 (Integralboard)
⇓
⇒ Switch generator to heating position
⇒ Send GR2 (= Status) , GR4 (= Relays) and GR5 (= Diag)
and check answers:
6-19
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Diagram 10 (continued)
Status = 15 Status = 16
→ Status 16 is the character
for switching off
→ The last number output
⇓ before status=16 is the
cause of switching off .
Diag = 31 Diag = 15
→ Contactor K 52 is faulty
or
⇓ → -B 20 is faulty
6-20
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Component Layouts
6-21
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
6-22
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
6-23
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
6-24
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Overview K751
B20 :
B115 :
X 24 . 1 + 2 X 23 . 1 + 4 X1 X2
B 100 :
X 40 X 23 X 15
B4:
X3 X4
6-25
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
The numbers in the picture refer to the spare parts list above.
1
5
16 6
9
2
4
10
3
11 14
8
12 13
17 15
6-26
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Error-Code
1 heating current too small
2 heating current too large
3 kV overshooting
4 mA overshooting
6 kV > 58
7 DC voltage regulation error (UAK > 480V)
10 Shut-off circuit open
15 Key switch HV-On was not turned until READY-LAMP was aktiv.
20 Generator switched off via key-switch
21 Power cut
22 Circuit breaker F1 switched off
23 Mains current too large
28 No sample in spectrometer chamber
29 Short circuit to ground in shut-off circuit
30 Safety circuit open
31 Short circuit to ground in safety circuit
33 Relay K11 open
37 Relay K12 open
38 Relay K13 open
40 Tube surges
41 High voltage in stand-by mode
42 DC voltage in stand-by mode
50 Stand-by function not parametrized
51 Generator type not identified
52 Eeprom type not identified
STATUS
0 -B20 board defective
1 Wait for key-switch "ON"
2 Check relay contacts K3 + K5
3 Check relay contacts K12
4 Check power supply
5 Check relay contacts K11 + K13 Check key-switch "ON"
6 Check safety circuit
7 Wait for "high voltage ON"
8...14 Internal use
15 Generator ready
RELAY
1 Relais K3 closed
2 Relais K5 closed
4 Relais K12 closed
8 Relais K13 closed
16 Relais K10 closed
64 Relais K14 closed
WARN
16 Current regulation out of range or generator in stand-by
6-27
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
RS 232 Commands
GS-commands
GS1 Operating mode control word
(decimal number resulting from the following bit significances ; meaning if set to 1)
Significance Meaning
1: Generator switched to position ON
2: XRAY ON (K52 + K53 closed)
4: kV actual value = kV setpoint
8: mA actual value = mA setpoint
16 : Remote control
32 : High voltage flashover
64 : Generator warning (see GR6)
128 : Generator in Standby mode (switch off circuit open, GV0 command)
6-29
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
GS5 0 (reserved)
GS6 Voltage at capacitor battery (UAK)
GS7 High voltage setpoint value (kV)
GS8 Tube current setpoint value (mA)
GS9 Main input current (A)
GS10 Operating hours
GS11 High voltage flashover counter (tube surges)
GR-commands
GR1 Program version (Firm:V..)
GR2 Generator status
Repetitive GR2-command will return a decimal number from 0 to 16 depending on the generator
status.
1. On board check when powering up : GR2 should return 1 after some seconds,
if not remove –B20 board
2. On board check when switching to ON: GR2 should return 7 after some seconds
3. Switching to XRAY ON : GR2 should return 15 after some seconds
GR3 Error-Code number
GR4 Status of relays on –B20 board
GR5 Status of supply voltage on –B20 board
(decimal number resulting from the following bit significances ; meaning if set to 1)
Significance Meaning
1: +15V present
2: -15V present
4: -5V present
8: +14V present
16 : +24V present
Significance Meaning
16 : Current controller out of range (tube current actual value < setpoint value), for example in standby mode
(switch off circuit open).
6-30
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
GP- commands
GP1 10 High voltage minimal value (kV)
GP2 50 High voltage maximal value (kV)
GP3 5 Tube current minimal value (mA)
GP4 50 Tube current maximal value (mA)
GP5 Tube power maximal value (see GF8 command)
GP Complete string GP1,GP2,....,GP5 will be transmitted
GF-commands
GF1,1 Start tube conditioning program.
GF-command will return:
GF3 as long as the conditioning program runs
GF130 will be returned. if an error occurs
GF66 will be returned, if conditioning program has successful been finished
GF1,0 Stop tube conditioning program
GF2,1 Display calibration: kV=27 and mA=40 for 1-2 min., when the calibration was successful :
kV=20 and mA=10 (always after replacing a B424 board)
GF2,0 End of calibration program
GF3,1 Disable error detection of Error-Code:1, Error-Code:22, Error-Code:27
GF3,0 Enable of error detection
GF4,1 Lock kV + mA regulator (High voltage locked to about 12kV and tube current = 0mA)
GF4,0 Normal operation
GF4 commands requires of the generator to be in standby mode to be executed or ?23 will be returned.
GF8,nnnn Input for tube power (in Watt) maximal value with nnnn = 0400 ... 4000
GF9,7 reset operating hours
GF9,8 reset tube surge counter
6-31
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
• PCB C79249-A3054-B424
• PCB C79249-A3053-B404
• Inverter
• HV tank
• HV cable
• X-ray tube
To figure out the bad component the following procedure should be carried out
STEP2
Setpoints for kV : 20 , mA : 5
- Manual Control Command: GV20 , GC5
Switch the generator to HV ON
- Manual Control Command: SA.. , ..
6-32
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
STEP3
Switch OFF the generator
- Manual Control Command: SA0
- Turn key switch to OFF
Remove the HV cable on the generator side
Switch the generator to HV ON
- Manual Control Command: SA.. , ..
STEP4
6-33
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
⇓
⇒ Switch off generator
⇒ Send GF4,1 to generator
⇒ Switch S4 to "operation"
⇒ Read heating current (HCU) in the S4-Tools or send GS3
⇒ Tube faulty
⇓
Switch off generator
⇒ Remove heating cable from socket X75 (rear panel)
⇒ Measure the tube resistance at the removed end of the
cable
Low resistance High resistance
6-34
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Diagram 1 continued
Low-resistance
No voltage U ≈ 2,5 V
(Generator switches off again)
→
⇓ - replace -B 424
U = Umains = 230 V ∼ U = 0V
6-35
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
kV = 40 mA = 0 kV = 40 mA ≠ 0
6-36
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Diagram 2 continued
kV = 20 mA = 0 kV ≠ 20 mA ≠ 0
Generator reacts as above
kV = 40 mA ≠ 0
kV = 40 mA = 0 Generator switches off with
Error Code 1
6-37
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
→ replace -B 424
6-38
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Diagram 4
R = 100 R ± 1 % R ≠ 100 R ± 1%
6-39
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
⇓ → Replace fuse
→ replace - B 424
⇒ One after another connect all plugs to –B424 again, and measure
the voltages again after each plug
6-40
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Diagram 5 continued
⎯ X 523 (HV-multiplier)
⎯ X 524 (S4-internal)
⎯ X 521 (inverter)
A voltage disappears
following the connection of a
All voltages present plug
→ Replace
corresponding module
6-41
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Circuit breaker
F1 tripped or
mains current U - AK < 10 V U - AK ≈ 70 V U-AK ≈ 70V
becomes too or
kV ≈ 8 (6......9) kV = 92
large
U - AK ≈ 180V or pending
⇒ Check run-up
6-42
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Continued Diagram 6
U - AK ≈ 125 (125...135 V ) U - AK = 0
U - AK = 0
U - AK ≈ 120 (115-135 V) or
U - AK ≈ 150 V
⇓ → inverter faulty
⇒ Reconnect X 15
U - AK ≈ 120 V U - AK = 0
6-43
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
R = 20 k ± 0,1% R ≠ 20 k ± 0,1%
→ Replace HV tank
R = 20 k ± 1% R > 20 k ± 1%
6-44
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Diagram 7
6-45
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
The diagram shows Imains dependent upon x-ray tube current and tension
(Umains = 230V and f = 50Hz).
U/ I/
kV mA 5 10 20 50 80 100
20
30
40
50
6-46
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
⇒ Check wiring of K52 and K53 ⇒ Check relays K54 and K55
6-47
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Diagram 9
R = 6 ....10 Ω otherwise
6-48
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Diagram 10
Answer = 1 Answer ≠ 1
Diag = 31 Diag = 15
→ Contactor K 52 is faulty
or
⇓ → -B 424 is faulty
6-49
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Component Layout
6-50
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
6-51
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
6-52
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
6-53
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Appendix :
Overview K774
9 8
7 6
10
14
13
12
11
6-54
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Generator Parts:
1. Capacitor bank C79249-A3028-B74
2. Thyristor module W79024-A6122-T483
3. PCB: integral board C79249-A3053-B424
4. PCB: DC voltage regulation C79249-A3053-B404
5. Contactors K52 + K53 C79298-A3230-C39
6. Relays K7, K54 and K55 C79298-A3200-C34
7. Contactor K51 W75053-B1001-N402
8. Fuses F2 (1.6AT), F6 (4AM); 250Vac
9. Transformer T51 W75040-B14-A55
10. Fan (230V, 50/60Hz) W79087-A2001-B50
11. Mains choke W79041-A1505-C1
12. Inverter C79249-A3053-B120
13. Step switch W79024-A6122-T483
14. HV tank C79249-A3053-B125
15. Line filter Z1 K180-C15
16. Circuit breaker F1 K180-C12
17. 50Hz / 60Hz switch X33 at the rear panel
16
17
15
6-55
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
Circuit Diagrams
Page 1:
6-56
S4 EXPLORER & PIONEER Service Manual X-ray Generators K751 and K774
Page 8
6-57
X-ray Generators K751 and K774 S4 EXPLORER & PIONEER Service Manual
6-58