Вы находитесь на странице: 1из 6

British Heart Journal, I972, 34, 930-935.

Exercise studies in congenital heart block


Mervyn R. H. Taylor and Simon Godfrey
From The Institute of Diseases of the Chest, London

Bicycle exercise studies were performed on 4 patients with congenital heart block. Cardiac output
was measured using the indirect (CO2) Fick technique. Two patients were found to have a normal
capacity for work but they both had low cardiac outputs at high work loads. The 2 patients who
were unable to perform a normal amount of work had normal cardiac outputs on exercise at
relatively low work loads. The degree to which a patient increased his ventricular rate or stroke
volume was not related to the maximum work which he could perform. It is suggested that in these
patients with very slow heart rates at work normal amounts of work can only be done if the patient
can tolerate a low cardiac output, a large arteriovenous oxygen difference, and a low venous
oxygen tension.

There have been many studies of cardiac func- spirometer and analysed continuously for 02 and
tion in patients with congenital heart block C02. When pulse, minute ventilation, and expired
(Holmgren, Karlberg, and Pernow, I959; gas concentration were all steady (after 24-31
Wright, Adams, and Anderson, I959; Paul, minutes), a collection of expired gas was made in
the spirometer over at least one minute and
Rudolph, and Nadas, 1958; Nakamura and analysed immediately. While the expired gas was
Nadas, I964; Moss, I96I; Wallgren and Win- being collected a sample of arterialized blood was
blad, I937; Campbell and Thorne, 1956). taken from an ear lobe for estimation of pH,
However, there have been few reports of Po2, and Pco2 (Godfrey et al., I97ib). The mixed
exercise tolerance, or of cardiac output mea- venous Pco2 was estimated by rebreathing a C02
surements during exercise in patients with in oxygen mixture to obtain equilibration of CO2
this condition who do not have artificial tension between the gas mixture and the mixed
cardiac pacemakers (Ikkos and Hanson, I960; venous blood (Jones et al., 1967; Denison et al.,
Moss, I96I; Holmgren et al., I959). Cardiac I969). A downstream correction was not applied
output at rest has been reported both as (Jones et al., 1967; Godfrey et al., I97ia).
Carbon dioxide production was calculated
low (Wright et al., I959) and normal (Paul et from the analysis of expired air, and the carbon
al., I958; Moss, I96I). On exercise, too, the dioxide content of arterial and mixed venous
reports of cardiac output have been conflicting blood was calculated from the gas tensions using
(Ikkos and Hanson, I960; Holmgren et al., a C02 dissociation curve and a digital computer
'959). programme (Godfrey, 1970). Cardiac output was
This communication reports the exercise calculated using the indirect (C02) Fick method:
tolerance and cardiac performance of 4 pa- carbon dioxide production
tients with congenital heart block who do not mixed venous - arterial carbon
have artificial cardiac pacemakers. dioxide content difference
Mixed venous oxygen content was calculated
Materials and methods from the cardiac output, the oxygen consumption,
Exercise tests were performed in the upright and the arterial oxygen content.
position on an electrically braked cycle ergometer The arterial Pco2 used for the estimation of
as described elsewhere (Godfrey et al., I97ia). cardiac output was obtained either from arterial-
The exercise test was composed of two parts. A ized ear lobe blood samples or derived from the
progressive exercise test was performed first, in Bohr equation assuming a normal dead space
which measurements were made of ventilation (Godfrey and Davies, I970). Arterial blood lactate
and heart rate while the work load was increased rise was calculated by the method of Clode and
every minute up to a maximum that the subject Campbell (I969).
could tolerate. After a 30-minute rest, steady state The normal values given are those which have
measurements were made in each subject at rest, been obtained in this laboratory using the same
one-third, and two-thirds of the maximum work methods as in the present study (Godfrey et al.,
load reached in the progressive test. I97ia). 'Normal range' in all cases refers to two
Expired gas was flushed through a large Tissot standard errors either side of the estimate for
Received 17 January 1972. normal subjects.
Exercise in congenital heart block 931

The physical characteristics and diagnostic TABLE I Patients


details of the patients are given in Table I.
Congenital heart block was diagnosed in Case i Case Age Sex Height Weight Associated cardiac abnormalities
at the age of 6 years. She was instructed never to No. (yr) (cm) (kg)
exert herself and so she never played games at
school and always led a sedentary life. Cardiac I 23 F I6o 51.3 Nil
* catheterization revealed a normal cardiovascular 2 21 M I72-5 73-4 Dextrocardia and corrected trans-
system apart from the conduction defect. Con- position of great vessels
genital, heart block was diagnosed at birth in 3 I5i M 169 580o Nil
Case 2. He has had exercise intolerance from 4 9+ M 124 2-55 Dextrocardia
childhood. When playing football at school he
had to stop about every I0 minutes because of
dyspnoea. He complains of being tired in his
atrial and ventricular rate, but the atrial rate
present work as a waiter and suffers from mild
rose faster than the ventricular rate with in-
4 bronchitis. Cardiac catheterization revealed only
creasing work (Fig. i). The rate of rise was
minimally increased right ventricular and pul-
greatest in the youngest patient. The atrial
monary arterial pressures in addition to the
rate in all cases followed the same overall
anatomical lesions given in Table i. Congenital
trend as in normal subjects and only in Case 3
heart block was diagnosed in Case 3 at the age of
did it fall outside the normal range at some
I0 years when he attended hospital because of a
chest infection. Case 4 was delivered by caesarian
work loads.
section because of bradycardia. His heart rate
The increase in ventricular rate on work
^ remained between go and ioo beats a minutevaried from patient to patient. From rest to
after delivery and the diagnosis of congenital heart
maximum work Cases i and 3 increased their
block was confirmed by an electrocardiogram.
ventricular rate by 6o beats a minute, Case 2
Cases 3 and 4 both take part in athletics and foot-
ball without any difficulty whatsoever. by 3I beats a minute, and Case 4 by 20 beats
a minute. The maximum work load that a
None of the patients had ever suffered from
Adams-Stokes attacks, nor were any of thempatient was able to complete was not related
to his rise in ventricular rate. Case 4, with the
receiving any treatment at the time of the study.
> All. of them had complete atrioventricular dis-
sociation on their electrocardiograms.
smallest rise in ventricular rate (20 beats/
minute), reached a normal maximum work
for his size. Case 3 also reached a normal
Results maximum work load, while Cases i and 2 did
The progressive exercise test On exer- not reach their normal maximum loads (i40
cise all patients showed an increase in both and 210 W, respectively).
FIG. I The atrial rate (0) and the ventricular rate (*) for each patient in the progressive
exercise test plotted against the work load. The shaded areas represent 2 standard errors either
side of the regression line for normal subjects of the same height.
220 Patient 1 Patient 2 Patient 3
200 0 010 101
180 (watts)
160~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
140~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
120~~~~~~~~~~~~~~~~~~~~~~~~~~0
140'
80~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~
60~~~~~~~~~~~~~~

40-
20

0 204060801000' i240&)O801I0012'010 6 2040 6b80 I00I 40C


WVork (watts)
932 Taylor and Godfrey

TABLE 2 Results for steady state exercise test


l6 * Patient
in 4 patients (mean values for normal subjects
A Pati!ent2
are given in brackets)
c 14 A Patient 3
X Patient 4
.E Case Work Minute Oxygen Arterial Cardiac Ven-
a) 12 No. (W) ventilation consumption PCo2 output tricular
(I./mmin B TPS) (I./mmin STPD) (mmHg) (1./min) rate/min
m
10
0-
I 0 io-8 (io) 2I4 (330) 2I.5 5-4 (4-5) 63
8 x~~~~~~~~~~~~~~~~~~~
33 32-9 (25) 827 (720) 3I.2 9-8 (8-5) 87 (I43)
66 52-0 (35) I2I3 (II20) 28-2 II-4 (II-0) 114 (I90)
6-
o 2 0 I2 (I0) 334 (320) 24.8 3-6 (5-4) 52
L-
a
4. 47 36-7 (32) II05 (900) 34.2 8-7 (I-2) 55 ('34)
95 56-8 (42) I498 (I460) 33.9 I2-2 (I2-8) 72 (I55) 4
2-
3 0 I29 (Io) 312 (320) 32.3 3-6 (5-I) 56
0' 66 37-7 (36) i268 (1120) 3-0 7-0 (II-4) 63 (146)
0 0&25 0*50 0.75 1I00 1-25 1-50 1-75 2-OD 132 54-I (50) i8io (I880) 32.5 IO-5 (14-8) 87 (7I)
Oxygen consumption (litres/min) 4 0 5-3 (8) I58 (i50) 33-8 2-0 (4-0) 59
FIG. 2 The cardiac output for each patient 20 I8-9 (i8) 528 (550) 33-0 5-0 (6-4) 70 (I39)
40 26-0 (23) 7I2 (8oo) 30°3 5-I (7-8) 82 (I60)
plotted against oxygen consumption. The heavy
line represents the regression line for normal
subjects and the shaded area two standard
errors either side. Steady state tests The results of the steady
state tests are given in Table 2.
Most normal subjects have a heart rate in
the range I75-195 beats/minute at their maxi- Cardiac output The results for cardiac
mum work load (Godfrey et al., I97ia). Three output in relation to oxygen consumption are
of the patients in this study had atrial rates in given in Fig. 2. Cardiac output was normal at
excess of i 8o beats a minute at their maximum rest in all cases except Case 4, in whom it was
work load. In general the atrial response to low. On exercise cardiac output increased in
exercise was normal but the ventricular in- all cases. The results for Cases i and 2 were
crease in rate was below normal. normal at both levels of exercise. Case 4 had
Minute ventilation was normal throughout a cardiac output within normal limits at the
the progressive exercise test except in Case I first work load but a low output at the second
where it was increased by i 8 1./minute above work load. Case 3 had a low cardiac output
the normal range during the last work load. at both work loads.
FIG. 3 The stroke volume for each patient plotted against the oxygen consumption. The shaded
area represents two standard errors either side of the regression line for normal subjects of the
same height.
180 Patient 1 Patient 2 Patient 3 Patient 4
160.
140
E 120
-2E 100.
> 80.
0
40 ...... -..
20-
0- 0k0-5 ..
0 0 25 0.50 0 75 1I00 1I25 0
0-25
0-25 0450 0.75 1I0
I
.*25
0 025
0 0-25 0z50 0.75
110jWi5
1*00 I*25 1*50 02
1.50 5
07lllli
7
Oxygen consumption (litres/min)
Exercise in congenital heart block 933

Stroke volume In Fig. 3 stroke volume less apprehensive, the figures were II9 beats
for each patient is plotted against oxygen con- a minute and 63 beats a minute, respectively.
sumption and the normal ranges given. It appears that emotion affects the atrial but
All patients had a large rise in stroke volume not the ventricular rate.
between rest and exercise with only a small Moss and Adams (I968) have stated that the
further change between the first and second ventricular and atrial rates generally parallel
work loads. Only in Case i was the stroke each other. It can be seen in Fig. i that this
volume abnormally large at rest. At work all only occurred in i of the 4 patients in this
patients had a stroke volume greater than nor- study (Case i). They also state that heart rate
mal except Case 4 whose stroke volume was remains quite constant (40-80) from before
just within normal limits during the second birth to the adult age group. Case 4 had a ven-
i work load. tricular rate of go to ioo beats/minute at birth
and a rate of 80 to go beats/minute after 40
4 Minute ventilation During the second hours of age. His ventricular rate at rest is
steady state work load, Cases i and 2 both now 62 beats a minute.
had a raised minute ventilation. At rest and Case 2 had no extrasystoles at any time.
at the first work load they both had a normal Case 3 developed ventricular extrasystoles at
minute ventilation. In Cases 3 and 4 minute the ioo watt load and they remained present
V ventilation was normal. All the patients had at all higher work loads. The extrasystoles
moderate alveolar hyperventilation, as shown disappeared almost immediately he stopped
4 by the low arterial Pco2 values. work. Case i followed the same pattern but
the extrasystoles first appeared at the 80 watt
Other investigations Venous admixture load. Case 4 had frequent ventricular extra-
was normal in all patients. Calculated lactate systoles both at rest and on exercise. Ikkos
rise never increased above normal values in and Hanson (I960) noted ventricular extra-
any subject. systoles on exercise in 6 of their ii patients
while Holmgren et al. (I959) found them in
Discussion 2 out of 4 patients and Moss and Adams (I968)
Progressive exercise test Ikkos and Han- in 2 out of I0 patients during exercise.
son (I960) assessed the maximum working
capacity of I0 patients with congenital heart Steady state test The 2 patients (Cases i
block. They found that 6 of the patients had and 2) who failed to reach a normal maximum
, a normal working capacity and in 4 it was work load in the progressive test had normal
slightly reduced. Holmgren et al. (i959) cardiac outputs in the steady state test at
; studied 4 patients with this condition and relatively low work loads. Both of the patients
found that the maximum working capacity who reached a normal maximum work load
was low in 3 of them and normal in only one. in the progressive test had a low cardiac out-
Moss (I96I) found a normal working capacity put at the higher work load in the steady state
in all of the 4 children he studied with con- test. The increase in stroke volume which
genital heart block. In the present study 2 occurred on work, though greater than normal,
patients had a normal maximum working was insufficient to compensate for the slow
capacity as measured in the progressive test ventricular rate. In order to reach a normal
and in the other 2 it was low. maximum work load these patients had to
tolerate a low cardiac output and, as a con-
Heart rate and rhythm Ikkos and Hanson sequence, an extremely low oxygen content
(I960), using a two work load test, noted that of their venous blood (Table 3). The calcu-
the rise in ventricular rate with work was not lated arteriovenous oxygen differences in
smooth. They found a large increase between these two patients were greater than the
rest and the first work load and a smaller in- highest values reported by Ekblom and
crease between the first and second work Hermansen (I968) in a study of Swedish
loads. This occurred in the present study athletes. The small changes in the calculated
z during the steady state test, but in the pro- lactate rise suggest that anaerobic metabolism
gressive test where small work increments did not play a large part in energy production.
were used, the ventricular rate rose smoothly That the venous oxygen saturation may be
as the work load increased (Fig. i). a limiting factor for work in patients with
Case i was somewhat apprehensive at the heart block is suggested by the work of Edhag
4 start of her progressive test and had an atrial and Zetterquist (I968) in adults with artificial
rate of I44 beats a minute and a ventricular cardiac pacemakers. They found that the mean
X rate of 59 beats a minute at rest. At rest, femoral venous oxygen saturation at which
during the steady state study, when she was their patients stopped work was similar whe-
934 Taylor and Godfrey

TABLE 3 Calculated values for arterial son (I960) both had low maximum working
oxygen content in ml/ioo ml blood (CaO2), capacities and normal cardiac outputs (assum-
mixed venous oxygen content in ml/Ioo ml blood ing a normal oxygen consumption) at low
(CQo2), mixed venous partial pressure of levels of work. Their findings agree with the
oxygen in mmHg (Pijo2), and mixed venous findings in the 2 patients in this study who
oxygen saturation per cent (SWo,) at rest and had normal cardiac outputs at work.
on exercise in 4 patients It appears that subjects with congenital
heart block who can produce only a small in-
Case Work CaO2 C;o2 PPo2 S;o2 crease in ventricular rate can only achieve a
No. (W) normal work output if they can tolerate a low
cardiac output relative to the work done.
I 0 215 I7-5 86 82 The difference between the patients reach-
33 2I13 12-8 3I 60
66 211I io-8 33 50 ing a normal maximum work load and those
failing to do so is presumably due to the re-
2 0 20 9 ii*6 33 53 sponse of the working muscle to increasing
47 2I23 8-6 25 40 difficulty in extracting oxygen from the circu-
95 2II 8-8 26 4I lation. Had the patients with normal cardiac
3 0 20 5 II-9 26 57 outputs reached a normal maximum work
66 20 6 2-5 12 12 load without any further increase in cardiac
I33 20 6 3-4 I3 i6 output, they would have had to tolerate a low
4 0 I7-9 IO1O 30 55 venous oxygen content. The ability to tolerate
20 i8-i 7T5 22 41 low venous oxygen contents may be the factor
40 i8-i 42I I5 22 separating the two types of response to exer-
cise and it is possible that this can be acquired
by training.
ther the ventricular rate was 47 or 72 beats a
minute, though the duration and severity of The authors would like to thank the consultants
work performed was higher at the higher ven- of Brompton, Guy's, and Hillingdon Hospitals
tricular rate. It may be that the ability to who allowed their patients to be studied, their
tolerate a low venous oxygen tension has to technical and medical colleagues, and the pa-
be acquired by training and this could explain tients themselves for their co-operation. This
the differences in exercise tolerance in the work was carried out under grants from the British
Heart Foundation and the Chest and Heart
present study. Both the subjects who had a Foundation. Dr. M. Taylor was in receipt of the
normal exercise tolerance were very active and Eden Fellowship in Paediatrics during the course
took part in field sports but this could be a of the study.
reflection of their normal exercise tolerance
rather than the cause of it. References
The fact that 2 patients were able to con- Campbell, M., and Thorne, M. G. (1956). Congenital
tinue work with low cardiac outputs suggests heart block. British Heart3Journal, I8, 90.
that cardiac output alone is not a limiting Clode, M., and Campbell, E. J. M.(I969). The rela-
factor for work in congenital heart block, and tionship between gas exchange and changes in
so heart rate and stroke volume will not be blood lactate concentrations during exercise.
Clinical Science, 37, 263.
limiting factors either. In these 4 patients the Denison, D., Edwards, R. H. T., Jones, G., and Pope,
ability to reach a normal maximum work load H. (I969). Direct and rebreathing estimates of the
was not related to the increase in ventricular 02 and C02 pressures in mixed venous blood.
rate, or in stroke volume on exercise. Respiration Physiology, 7, 326.
Edhag, 0., and Zetterquist, S. (I968). Peripheral circu-
Holmgren et al. (I959) have reported one latory adaptation to exercise in restricted cardiac
patient with congenital heart block who had output. Scandinavian Journal of Clinical and
a normal exercise tolerance and a normal car- Laboratory Investigation 21, 123.
diac output at work but this patient was able Ekblom, B., and Hermansen, L. (I968). Cardiac output
in athletes. Journal of Applied Physiology, 25, 6I9.
to increase his ventricular rate by 75 beats a Godfrey, S. (I970). The manipulation of the indirect
minute between rest and exercise which is Fick principle by a digital computer programme for
greater than the increase of any of the patients the calculation of exercise physiology results.
in the present study. The second subject they Respiration, 27, 513.
studied had a low maximum working capacity Godfrey, S., and Davies, C. T. M. (I970). Estimates
of arterial PCO2 and their effect on the calculated
and a slightly low cardiac output at work. This values of cardiac output and dead space on exercise.
subject was only able to increase his ventricu- Clinical Science, 39, 529.
lar rate by 34 beats a minute between rest and Godfrey, S., Davies, C. T. M., Wozniak, E., and
exercise. Barnes, C. A. (197Ia). Cardio-respiratory response
to exercise in normal children. Clinical Science, 40,
The 2 patients studied by Ikkos and Han- 4I9.
Exercise in congenital heart block 935

Godfrey, S., Wozniak, E. R., Courtenay Evans, R. J., Moss, A. J., and Adams, F. H. (I968). Heart Disease in
and Samuels, C. S. (I97Ib). Ear lobe blood samples Infants, Children and Adolescents. Williams and
for blood gas analysis at rest and during exercise. Wilkins, Baltimore.
British Journal of Diseases of the Chest, 65, 58. Nakamura, F. F., and Nadas, A. S. (I964). Complete
Holmgren, A., Karlberg, P., and Pernow, B. (i959). heart block in infants and children. New England
Circulatory adaption at rest and during muscular Journal of Medicine, 270, 126I.
work in patients with complete heart block. Acta Paul, M. H., Rudolph, A. M., and Nadas, A. S. (1958).
Medica Scandinavica, I64, 119. Congenital complete atrioventricular block: prob-
Ikkos, D., and Hanson, J. S. (I960). Response to exer- lems of clinical assessment. Circulation, I8, I83.
cise in complete atrioventricular block. Circulation, Wallgren, A., and Winblad, S. (I937). Congenital heart
22, 583. block. Acta Paediatrica, 20, 175.
Jones, N. L., Campbell, E. J. M., McHardy, G. J. R., Wright, F. S., Adams, P., and Anderson, R. C. (1959).
Higgs, B. E., and Clode, M. (I967). The estimation Congenital atrioventricular dissociation due to
of carbon dioxide pressure of mixed venous blood complete or advanced atrioventricular heart block.
during exercise. Clinical Science, 32, 3II. American Journal of Diseases of Children, 98, 72.
Moss, A. J. (I96I). Congenital complete atrioventricu-
lar block. Clinical feature, haemodynamic findings Requests for reprints to Dr. Mervyn Taylor,
and physical working capacity. Journal-Lancet, 8i, Institute of Diseases of the Chest, Brompton
542. Hospital, London S.W.3.

Вам также может понравиться