Corneal Topography
and Contact Lenses
Michael Twa, Cleusa Coral-Ghanem,
and Breno Barth
Keratometry
Keratometry measures central corneal curvature at two points within
the central 2 to 4 mm (3 mm on average) of the cornea and does not
measure the mid-peripheral or peripheral cornea. Keratometry is a
rapid and accurate method of determining curvature of the central cor-
neal cap that is useful for estimating the overall optical power of the
cornea. Keratometry measurements are less repeatable with abnormal
corneal shapes distorted by injury or diseases that distort corneal shape
such as keratoconus. While keratometry normally provides no infor-
mation about the peripheral corneal shape, the normal clinical proce-
dure can be modified to measure peripheral corneal curvature.3
The principle of keratometry relies on the reflective optical properties
of the cornea. The reflective surface of the cornea produces a virtual
image the size of which is a function of the corneal radius of curvature.
The keratometer employs the doubling principle. The reflected image
is split into two images using a semisilvered mirror and a series of
prisms. The relative displacement of the two images is then adjusted
until the top of one image is coincident with the bottom of the second,
as described below.
The figure demonstrates the mires as viewed from the practitioner’s
side of a conventional keratometer. The task is to align the mires at the
9 o’clock position of the lower right hand circle to measure the hori-
zontal meridian and at the 12 o’clock position of the lower right hand
circle to measure the vertical meridian.
The specific procedure is as follows:
1. Ask the patient to place his or her chin in the chin rest and forehead
against the headrest.
2. Have the patient fixate on the reflected image of his or her eye or
the target in the instrument.
3. With the keratometer in the straight-ahead position, sight down the
instrument vertically until the leveling sight is aligned with the
patient’s temporal canthus.
4. Occlude the eye not being tested. Release the knob for locking the
instrument and rotate the instrument until it points directly at the