Вы находитесь на странице: 1из 6

Outline

 Field shaping
Treatment planning III: ◦ Custom blocks
field shaping, skin dose, and ◦ Use of independent jaws
field separation ◦ Multileaf collimators
 Skin dose
Chapter 13 ◦ Maintaining skin sparing of MV beams
F. M. Khan “The Physics of Radiation Therapy”  Filed separation techniques
◦ Geometric
◦ Dosimetric

Field shaping Field blocks


 The shaping of a treatment field is 1
n
1
dictated by two factors:    n  0.05
2 2
◦ Complete coverage of the tumor with 1
2n   20
prescription dose, including local and distal 0.05
disease log 20
n  4.32
◦ Dose to normal tissue should be minimized; log 2
vital organ tolerance observed
 Typical figure of merit: primary beam
transmission of 5% or less through a  Thickness of a block should be large enough to transmit
only 5% of the primary photon beam
blocked region
 n is the number of HVL’s, thickness depends on the beam
quality

Custom blocks Custom blocks


 The blocks should be shaped  Custom blocks are made of
so that their sides follow the melted Cerrobend based
geometric divergence of the on shapes cut from
beam, minimizing the block Styrofoam
transmission penumbra
◦ “Positive” or “negative”
(partial transmission of the
beam at the edges of the  Mounted on a block tray in
block) the accelerator head
 Straight blocks are used for  Transmission is ~3.5%
beams with large geometric
penumbra (60Co source)
Styrofoam block cutter

1
Independent jaws Multileaf collimators
 Used for rectangular filed  MLC consists of a large
blocking (asymmetric number of collimating blocks
(leaves) that can be driven
fields) automatically, independent of
 Transmission ~1% each other
 The effect on the isodose  Can generate field of almost
distribution is very close any shape
to that of a custom block  Typical MLC consists of 40
◦ Close agreement as well as pairs (80 leaves)
the tilt of the isodose curves ◦ Made of tungsten
toward the blocked edge ◦ Each leaf 1 cm wide at the
isocenter
◦ Thickness 6 to 7.5 cm
Comparison of isodose distribution with half ◦ Leaf transmission ~2%
the beam blocked by an independent jaw ◦ Interleaf transmission ~3%
versus a block on a tray

Multileaf collimators Multileaf collimators


 Adjustable collimator
systems are designed to
follow the beam divergence
(“focused” design)
 Difficult to implement with
independent leaves
 Leaf edges of different
design; most common –
rounded edge
 Single and double-focused
The penumbra at any position is
• Micro-MLC’s for stereotactic treatments: leaf within 1–3 mm of that obtained
(in both x and y directions)
width is 2 mm at the isocenter with a focused system or for alloy systems
• Interleaf leakage is minimized with tongue- blocks with divergent sides
and-groove design “Basic applications of multileaf collimators”, Report of TG-50, AAPM
2001

Multileaf collimators  Custom blocks


◦ Cerrobend
 MLC produces larger penumbra than lung blocks
used when
either jaws or custom blocks (leaf edges MLC’s may not
of different design) be able to
cover all
 MLC is ideally suited for multi-field
critical
treatments, and complex multi-field structures
treatments (IMRT) adequately

 Limitation of filed shaping: cannot produce


blocked islands

2
Measurement of dose distribution in the
Skin dose build-up region
 Skin sparing is highly desirable feature of  Due to steep dose gradient
MV beams 100 the accurate measurement
is difficult
 Secondary electron contamination of 80
 Extrapolation chamber is

PDD (%)
photon beams may reduce this effect 60
the best tool, but is not
40 Monte Carlo
 Sources of secondary electrons: beam 20
Extrapolation chamber readily available, and is
difficult to use
Parallel plate chamber

shaping equipment and air 0


0.125 cc cylindrical chamber

0.0 0.5 1.0 1.5 2.0


 Next best is plane-parallel
 Contamination depends on several Depth (cm)
plate chamber
factors: photon energy, field size, SSD, 6 MV beam, FS=10x10
 TLD can be used for
angle of incidence measuring skin dose due to
their small thickness

Skin dose Skin dose


 The block tray
(shadow tray)
increases the
secondary electron
scatter and thus the
skin dose
 The amount of skin
dose depends on the
distance of the tray
from the skin
 Skin dose decreases with increasing photon energy Figure 13.6.Effect of Lucites hadow tray on dose  It is best to have the
buildup for 10-MV x-rays. Percent depth dose tray as far from the
 For high-energy beams significant sparing is achieved for distribution is plotted for various tray to surface

subcutaneous layers
distances (d). 10-MV x-rays, tray thickness =1.5 patient as possible
g/cm2, field size =15 x15 cm, SSD =100 cm, and
source to diaphragm distance =50 cm.

Skin dose Electron filters


 The larger is the
field size, the more
secondary
electrons are
emitted from the
collimator and air
 Skin sparing is
 Materials with medium Z produce less forward scattered
significantly reduced electrons, Z=50 (tin) is the best
Figure 13.7.Percent surface dose as a function of
field size. 60Co, Theratron 80, source to surface for the larger field  Electron filters can be used when skin dose becomes
distance(SSD) =80 cm, source to diaphragm
distance (SDD) =59 cm. 4 MV, Clinac 4, SSD =80 sizes excessive:
cm. 10 MV, LMR 13, SSD =100 cm, SDD =50 cm. ◦ Large field size is large
60Co and 4-MV.
◦ The block tray distance is 15-20 cm from the skin

3
Electron filters Oblique incidence
 The angle of
 Tin filter should face incidence of a beam
the patient surface has an effect on skin
 The thickness of filter dose and the depth
should be equal to of dmax
the range of  High energy photon
secondary electrons beams generate
(0.9 mm of tin for secondary electrons
Co-60) in the air around
Figure 13.10.The use of electron range them
surface (ERS) to determine surface dose
buildup at point P. A:Perpendicular beam
incidence. B:Oblique beam incidence.
C:Tangential beam incidence.

Oblique incidence Separation of adjacent fields


 Obliquity factor: a  In some cases there is a need for
ration of doses at a treatments involving adjacent fields
point on CAX in ◦ Hodgkin’s disease (lymphoma)
phantom for oblique to
◦ Craniospinal fields in treatment of
normal incidence
medulloblastoma
 For tangential beams
◦ Some head and neck treatment fields
maximum skin dose can
be estimated as  Problem: when photon fields are placed
(entrance dose – in next to one another the divergence
normal incidence): causes hot spots at depths and cold areas
Figure 13.11.Obliquity factor at the surface
%Skin dose 
1
100%  entrance dose
near the surface
plotted as a function of beam angle for various
energy beams. Jackson formula for tangential 2
beam incidence is based on Equation 13.1.

Separation of adjacent fields Methods of field separation


 Various techniques used for  Two basic approaches: geometric and
field matching: dosimetric
◦ A: Angling the beams away
from each other so that the  In geometric approach fields are joined at 50%
two beams abut and are isodose line, producing 100% at the junction
aligned vertically
point
◦ B: Fields separated at the skin
surface. The junction point is ◦ The lateral dose distribution at the junction depth
at a depth where dose is can be more or less uniform, depending on the
uniform across the junction interfield scatter contribution
◦ C: Isocentric split-beam
technique for head and neck
 In dosimetric approach the goal is to produce a
tumors composite isodose distribution which is uniform
◦ D: Craniospinal irradiation at the desired depth
using penumbra generators

4
Separation of adjacent fields Separation of adjacent fields
 An example gap calculation
Figure 13.13.
Geometry of two
adjacent beams,
separated by a
distance S1+S2 on the = 5cm
surface and
junctioning at depth d.
= 10cm

 Finding distances S1 and S2 from similar


SSD1 for para-aortic = 95cm, SSD2 for pelvis = 98cm
triangles, obtain the total filed separation: Matching fields at 8cm depth gives:
d d  5cm   10cm 
S  S1  S 2  1 L1  1 L2 gap    8cm     8cm   1.24cm
2 SSD 2 SSD
1 2  95cm   98cm 

Separation of adjacent fields


Figure 13.15.Geometric
separation of fields with all
 A high-dose region of the four beams intersecting at
three-field overlap is midpoint. Adjacent field sizes:
30 x30 cm and 15 x15 cm;
created when bigger fields source to surface distance
diverge into opposing (SSD) =100 cm;
anteroposterior thickness =20
smaller fields cm; 4-MV x-ray beams; each
beam weighted 100 at its
 The maximum overlap depth of Dmax.
region is at the surface A: Field separation at surface
=2.3 cm. A three-field overlap
L1 SSD1 exists in this case because
DS  S1  S d  0 if  the fields have different sizes
L2 SSD2 but the same SSD.
 Overlap can be avoided by B: The adjacent field
adjusting SSD’s of adjacent separation increased to 3 cm
to eliminate three-field
Figure 13.14.Two pairs of parallel opposed fields. fields overlap on the surface.
C: Field separation adjusted
Adjacent fields are separated on the surface so that
they all join at a point on the midline.  Another approach: increase 2.7 cm to eliminate three-field
A:Ideal geometry in which there is no three-fields
overlap. B:Arrangement in which there are two of DS (produces cold spots) overlap at the cord at a 15-
cm depth from anterior
regions(shaded) of three-fields overlap.

Orthogonal field junctions Orthogonal field junctions


 An arrangement in which the central  Figure 13.16.
axes of the adjacent fields are ◦ A: A general diagram
orthogonal showing the separation of
orthogonal fields.
 A geometrical method of field ◦ B: An example of
separation with orthogonal fields used for
craniospinal irradiation.
d ◦ C: A lateral view of B,
S1 L illustrating the geometry of
2 SSD orthogonal field separation.

 d is the depth of field junction

5
Orthogonal field junctions Orthogonal field junctions
Figure 13.17. Craniospinal
irradiation technique.
◦ A:Patient setup showing Styrofoam
blocks and Alpha Cradle mold to
provide stable position for abdomen,
chest, and head.
◦ B: Lateral view of fields showing
cranial field rotated to align with the
diverging border of the spinal field.
◦ C: Couch rotated to provide match
between the spinal field and the
diverging border of the cranial field.
◦ D: Elimination of cranial field  An example of orthogonal field
divergence by using an independent
jaw as a beam splitter. This provides junctions. The AP field (red) is
an alternative to couch rotation in C. the s.clav or yoke.
 The is abutted to a LT LAT and
RT LAT (blue)

Guidelines for field matching Guidelines for field matching


1. The site of field matching should be chosen over 3. For deep-seated tumors, the fields may be separated on
an area that does not contain tumor or a critical the skin surface so that the junction point lies at the
midline. Care must be taken in regard to a critical
structure structure near the junction region.
2. If the tumor is superficial at the junction site, the 4. It is not necessary anatomically to reproduce the line of
fields should not be separated because a cold field matching every day because variation in its location
spot on the tumor will risk recurrence. will only smear the junction point, which is desirable. For
◦ They will overlap at depth, which may be clinically the same reason some advocate moving the junction site
acceptable, provided the excessive dosage delivered to two or three times during a treatment course.
the underlying tissues does not exceed their tolerance. 5. A field-matching technique must be verified by actual
◦ In the case of a superficial tumor with a critical organ isodose distributions before it is adopted for general
located at depth, one may abut the fields at the surface clinical use. In addition, beam alignment with the light field
but eliminate beam divergence using a beam splitter or and the accuracy of isodose curves in the penumbra
by tilting the beams region are essential prerequisites.

Key points Key points


 Thickness of lead required to give 5% primary beam  Dose at the surface or in the buildup region is best
transmission is 4.3 half-value layer measured with an extrapolation or a plane-parallel
chamber
 Half-beam blocking gives rise to tilting of the isodose
curves toward the blocked edge. This effect is due to  Surface dose depends on beam energy, field size, SSD, and
tray to surface distance
missing electron and photon scatter from the blocked
part of the field into the open part of the field  Electron filters are medium-atomic-number absorbers
(Z~50) that reduce the surface dose by scattering
 Physical penumbra with MLC is wider than that with the contaminant electrons more than generating them
collimator jaws or Cerrobend blocks  Surface dose increases with increasing angle of obliquity
 Surface dose in MV beams is predominantly due to the  Separation of adjacent fields, when needed, may be
electron contamination of the incident photon beam accomplished geometrically. Hot and cold spots in the
resultant dose distribution must be assessed by viewing
composite isodose curves

Вам также может понравиться