Вы находитесь на странице: 1из 17

FieldCare

Device Report for Selected Devices

Name:

Department:

Company:

Address:

Telephone:

Email:
FieldCare

\Network Topology\COM6\--------

Created by: Administrator

Created on: 2/20/2018 12:21:41 PM

Plant tag: -

Device address: 0

Device tag: --------

Device type: Promass / 80 / V3.01.0x

Manufacturer: Endress+Hauser

Type class: Flow

Serial number: 5448793

Version: 1.4.178.486

Page: 2 of 17
FieldCare

Offline Parameterize

Offline Parameters

GROUP SELECT.

MEASURING VALUES

Parameter Value
MASS FLOW 0.000000 kg/h
VOLUME FLOW 0.000000 m³/h
COR. VOLUME FLOW 0.0000 Nm³/h
DENSITY 0.000000 kg/l
REF. DENSITY 0.0000 kg/Nl
TEMPERATURE 0.000000 °C

SYSTEM UNITS

Parameter Value
UNIT MASS FLOW kg/h
UNIT MASS kg
UNIT VOLUME FLOW m³/h
UNIT VOLUME m³
UNIT COR.VOL.FL Nm³/h
UNIT COR.VOLUME Nm³
UNIT DENSITY kg/l
UNIT REF.DENSITY kg/Nl
UNIT TEMPERATURE °C (CELSIUS)
UNIT LENGTH MILLIMETER
UNIT PRESSURE BAR G
FORMAT DATE/TIME DD.MM.YY 24H

OPERATION

Parameter Value
LANGUAGE ENGLISH
ACCESS CODE 0
DEF.PRIVATE CODE 80
Page: 3 of 17
STATUS ACCESS ACCESS CUSTOMER
ACCESS CODE CNTR 0

USER INTERFACE

Parameter Value
ASSIGN LINE 1 MASS FLOW
ASSIGN LINE 2 TOTALIZER 1
FORMAT x.xxxx
DISPLAY DAMPING 1s
CONTRAST LCD 50 %
BACKLIGHT 50 %
TEST DISPLAY OFF

TOTALIZER 1

Parameter Value
ASSIGN TOTALIZER MASS FLOW
SUM 1 0.000000 kg
OVERFLOW 0.000000 E7 kg
UNIT TOTALIZER kg
TOTALIZER MODE BALANCE
RESET TOTAL. NO

TOTALIZER 2

Parameter Value
ASSIGN TOTALIZER MASS FLOW
SUM 2 0.000000 kg
OVERFLOW 0.000000 E7 kg
UNIT TOTALIZER kg
TOTALIZER MODE FORWARD
RESET TOTAL. NO

HANDLING TOTAL.

Parameter Value
RESET ALL TOTAL. NO
FAILSAFE ALL TOT STOP

CURRENT OUTPUT 1

Parameter Value
ASSIGN CURRENT MASS FLOW
CURRENT SPAN 4-20mA HART NAM.
VALUE 20mA 36000.000000 kg/h
TIME CONSTANT 1.000000 s
FAILSAFE MODE MIN. CURRENT
ACTUAL CURRENT 0.000000 mA
Page: 4 of 17
SIMULATION CURR. OFF

COMMUNICATION

Parameter Value
TAG NAME --------
TAG DESCRIPTION ----------------
FIELDBUS ADDRESS 0
HART PROTOCOL ON
MANUFACT ID ENDRESS+HAUSER
DEVICE ID 50 (hex)
DEVICE REVISION 9

PROCESSPARAMETER

Parameter Value
ASSIGN LF-CUTOFF MASS FLOW
ON-VAL.LF-CUTOFF 0.000000 kg/h
OFFVAL.LF-CUTOFF 50.000000 %
EMPTY PIPE DET. OFF
FIX. REF.DENSITY 1.0000 kg/Nl
ZEROPOINT ADJUST CANCEL
DENSITY SET VAL 0.000000 kg/l
MEASURE FLUID CANCEL
DENSITY ADJUST CANCEL
RESTORE ORIGINAL 0
PRESSURE MODE OFF

SYSTEM PARAMETER

Parameter Value
INSTL.DIR.SENSOR NORMAL FORWARD
MEASURING MODE STANDARD
POS. ZERO RETURN OFF
DENSITY DAMPING 0.000000 s
FLOW DAMPING 0.000000 s

SENSOR DATA

Parameter Value
K-FACTOR 1.000000
ZEROPOINT 0.000000
NOMINAL DIAMETER DN25
KM 0.000000 e-6
KM2 0.000000 e-9
KT 0.000000 e-6
KD1 0.000000 e-6
KD2 0.000000 e-9
C0 0.000000
C1 0.000000 e9
C2 0.000000 e6
C3 0.000000 e6
C4 0.000000 e15
Page: 5 of 17
C5 0.000000 e-3
MIN.TEMP MEAS. 100000002004087730000.000000 °C
MAX.TEMP MEAS. -100000002004087730000.000000 °C
MIN.TEMP CARR. 100000002004087730000.000000 °C
MAX.TEMP CARR. -100000002004087730000.000000 °C

SUPERVISION

Parameter Value
ACTUAL SYS.COND
PREV.SYS.COND
ASSIGN SYS.ERROR CANCEL
ASSIGN PROC.ERR. CANCEL
ALARM DELAY 0s
SYSTEM RESET NO
OPERATION HOURS 0.00
PERMANENT STORAG ON

SIMULAT. SYSTEM

Parameter Value
SIM.FAILSAFEMODE OFF
SIM. MEASURAND OFF

SENSOR VERSION

Parameter Value
SERIAL NUMBER
SENSOR TYPE
HW-REV. SENSOR
SW-REV. S-DAT

AMP. VERSION

Parameter Value
DEVICE SOFTWARE
HW-REV. AMP.
SW-REV. AMP.
LANGUAGE GROUP UNKNOWN
I/O TYPE UNKNOWN
HW-REV. I/O
SW-REV. I/O no IO

Device Data

Parameter Value
Distributor Endress+Hauser
TAG NAME --------
TAG DESCRIPTION ----------------
Page: 6 of 17
Message --------------------------------
Date 1/1/2000
Dev id 0
WRITE PROTECT OFF
Final asmbly num 0

Page: 7 of 17
FieldCare

Configuration

Communication

Parameter Value
TAG NAME --------
Poll addr 0
Dynamic Parameter Refresh after 10 sec
Static Parameter Refresh No Refresh
Refresh on Connect Refresh Data Set

Page: 8 of 17
FieldCare

Diagnosis

Simulation

Parameter Value
Device status A modification has been made to the configuration of the field device

Transmitter Status

Parameter Value
Device status A modification has been made to the configuration of the field device
STATUS 0
STATUS 1
STATUS 2
STATUS 3
STATUS 4
STATUS 5
STATUS 6
STATUS 7
STATUS 8
STATUS 9
STATUS 10
STATUS 11
STATUS 12
STATUS 13
STATUS 14
STATUS 15
STATUS 16
STATUS 17
STATUS 18
STATUS 19
STATUS 20
STATUS 21
STATUS 22
STATUS 23
STATUS 24

Page: 9 of 17
FieldCare

Observe

Process Variables

Parameter Value
MASS FLOW -0.114069 t/h
VOLUME FLOW -0.111903 m³/h
DENSITY 996.968750 kg/m³
TEMPERATURE 24.665924 °C
ACTUAL CURRENT 7.946511 mA
SUM 1 787525.500000 kg
SUM 2 787529.437500 kg

Page: 10 of 17
FieldCare

Online Parameterize

Online Parameterize

GROUP SELECT.

MEASURING VALUES

Parameter Value
MASS FLOW -0.114069 t/h
VOLUME FLOW -0.111903 m³/h
COR. VOLUME FLOW -0.1117 Nm³/h
DENSITY 996.968750 kg/m³
REF. DENSITY 1000.0000 kg/Nm³
TEMPERATURE 24.665924 °C

SYSTEM UNITS

Parameter Value
UNIT MASS FLOW t/h
UNIT MASS kg
UNIT VOLUME FLOW m³/h
UNIT VOLUME m³
UNIT COR.VOL.FL Nm³/h
UNIT COR.VOLUME Nm³
UNIT DENSITY kg/m³
UNIT REF.DENSITY kg/Nm³
UNIT TEMPERATURE °C (CELSIUS)
UNIT LENGTH MILLIMETER
UNIT PRESSURE BAR G
FORMAT DATE/TIME DD.MM.YY 24H

OPERATION

Parameter Value
LANGUAGE ENGLISH
ACCESS CODE 4685
DEF.PRIVATE CODE 80
Page: 11 of 17
STATUS ACCESS ACCESS SERVICE
ACCESS CODE CNTR 73

USER INTERFACE

Parameter Value
ASSIGN LINE 1 MASS FLOW
ASSIGN LINE 2 TOTALIZER 1
FORMAT x.xxxx
DISPLAY DAMPING 1s
CONTRAST LCD 50 %
BACKLIGHT 50 %
TEST DISPLAY OFF

TOTALIZER 1

Parameter Value
ASSIGN TOTALIZER MASS FLOW
SUM 1 787525.500000 kg
OVERFLOW 0.000000 E7 kg
UNIT TOTALIZER kg
TOTALIZER MODE FORWARD
RESET TOTAL. NO
SUM INVENTORY 2804258.250000 kg
OVERFL. INVENT. 0 E7 kg

TOTALIZER 2

Parameter Value
ASSIGN TOTALIZER MASS FLOW
SUM 2 787529.437500 kg
OVERFLOW 0.000000 E7 kg
UNIT TOTALIZER kg
TOTALIZER MODE FORWARD
RESET TOTAL. NO
SUM INVENTORY 0.000000 kg
OVERFL. INVENT. 787526 E7 kg

HANDLING TOTAL.

Parameter Value
RESET ALL TOTAL. NO
FAILSAFE ALL TOT STOP

CURRENT OUTPUT 1

Parameter Value
ASSIGN CURRENT TEMPERATURE
CURRENT SPAN 4-20mA HART NAM.
Page: 12 of 17
VALUE 4mA 0.000000 °C
VALUE 20mA 100.000000 °C
TIME CONSTANT 1.000000 s
FAILSAFE MODE MIN. CURRENT
ACTUAL CURRENT 7.948544 mA
SIMULATION CURR. OFF
CURRENT ADJUST OFF

PULS/FREQ.OUT.

Parameter Value
OPERATION MODE FREQUENCY
ASSIGN FREQUENCY MASS FLOW
END VALUE FREQ 1000.000000 Hz
VALUE f HIGH 1.000000 t/h
OUTPUT SIGNAL PASSIVE/POSITIVE
TIME CONSTANT 0.00 s
FAILSAFE MODE FALLBACK VALUE
ACTUAL FREQ. 0.000000 Hz
SIMULATION FREQ. OFF

COMMUNICATION

Parameter Value
TAG NAME --------
TAG DESCRIPTION ----------------
FIELDBUS ADDRESS 0
HART PROTOCOL ON
MANUFACT ID ENDRESS+HAUSER
DEVICE ID 50 (hex)
DEVICE REVISION 9

PROCESSPARAMETER

Parameter Value
ASSIGN LF-CUTOFF MASS FLOW
ON-VAL.LF-CUTOFF 0.020000 t/h
OFFVAL.LF-CUTOFF 50.000000 %
EMPTY PIPE DET. OFF
FIX. REF.DENSITY 1000.0000 kg/Nm³
ZEROPOINT ADJUST CANCEL
DENSITY SET VAL 1000.000000 kg/m³
MEASURE FLUID CANCEL
DENSITY ADJUST CANCEL
RESTORE ORIGINAL 0
PRESSURE MODE FIX
PRESSURE 0.3000 bar g

SYSTEM PARAMETER

Parameter Value
INSTL.DIR.SENSOR NORMAL FORWARD
MEASURING MODE STANDARD
Page: 13 of 17
POS. ZERO RETURN OFF
DENSITY DAMPING 2.000000 s
FLOW DAMPING 1.000000 s

SENSOR DATA

Parameter Value
K-FACTOR 3.346000
ZEROPOINT 67.000000
NOMINAL DIAMETER DN80
KM 420.000000 e-6
KM2 0.000000 e-9
KT 0.000000 e-6
KD1 0.000000 e-6
KD2 0.000000 e-9
C0 -2286.243652
C1 1.392890 e9
C2 -0.545468 e6
C3 0.000000 e6
C4 0.000000 e15
C5 0.000000 e-3
MIN.TEMP MEAS. 11.850006 °C
MAX.TEMP MEAS. 56.850006 °C
MIN.TEMP CARR. 726.849976 °C
MAX.TEMP CARR. -273.149994 °C
C0 Orig. -2286.243652
C1 Orig. 1.392890 e9
C2 Orig. -0.545468 e6
C3 Orig. 0.000000 e6
C4 Orig. 0.000000 e15
C5 Orig. 0.000000 e-3

CALIBR. HISTORY

Parameter Value
COUNTER HISTORY 1
SELECTION HISTORY 1
DATE HISTORY 01.04.16 11:10
CAL.FACTOR HIST. 3.346

SUPERVISION

Parameter Value
ACTUAL SYS.COND SYSTEM OK
PREV.SYS.COND RANGE FRQ.OUT1
ASSIGN SYS.ERROR CANCEL
ASSIGN PROC.ERR. CANCEL
ALARM DELAY 0s
SYSTEM RESET NO
TROUBLESHOOTING CANCEL
OPERATION HOURS 13666.82
HRS. SINCE RESET 0.13
PERMANENT STORAG ON

Page: 14 of 17
SIMULAT. SYSTEM

Parameter Value
SIM.FAILSAFEMODE OFF
SIM. MEASURAND OFF

SENSOR VERSION

Parameter Value
SERIAL NUMBER L3045902000
SENSOR TYPE PROMASS E
HW-REV. SENSOR V1.00.00
HW-IDENT. SENSOR 500?????
PROD-NO. SENSOR 5011218
SW-REV. S-DAT V3.00.00

AMP. VERSION

Parameter Value
DEVICE SOFTWARE V3.01.02
HW-REV. AMP. V2.00.00
HW-IDENT. AMP. 71106147
SW-REV. AMP. V3.01.02
SW-IDENT. AMP. 71259124
PROD-NO. AMP. 0110197904
LANGUAGE GROUP ASIA
I/O TYPE I/O X0
HW-REV. I/O V1.00.00
HW-IDENT. I/O 71142657
SW-REV. I/O V1.04.XX
SW-IDENT. I/O 71134637
PROD-NO. I/O 0110538240

SERVICE&ANALYSIS

Parameter Value
OSCILL. FREQ. 648.189209 Hz
EXCITER CURR. 2.896367 mA
CARR. TUBE TEMP. 0.000000 °C
MONITORING MODE ON
OSC.AMPLITUDE 100.000000 %
GAIN AMPLIFIER 100.000000 %
MONITOR.INTERVAL 10.000000 s
GAS MODE MODE 1

Device Data

Parameter Value
Page: 15 of 17
Distributor Endress+Hauser
TAG NAME --------
TAG DESCRIPTION ----------------
Message --------------------------------
Date 1/1/2000
Dev id 5448793
WRITE PROTECT OFF
Final asmbly num 0

Page: 16 of 17
FieldCare

Simulation

Online Parameterize

Parameter Value
SIM.FAILSAFEMODE OFF
SIM. MEASURAND OFF

CURRENT OUTPUT 1

Parameter Value
SIMULATION CURR. OFF

Page: 17 of 17