Академический Документы
Профессиональный Документы
Культура Документы
Collimator
Control type Manual
Lamp brightness (Lux) > 160 lux
External filter rail Yes
Patient Table
Type 4-Ways movement
Table top dimensions (L x W) 2,200 mm x 750 mm
Detector longitudinal travel 300 mm
Detector longitudinal travel method Manual
Longitudinal movement range
± 500 mm
(Table top, Left&Right)
Transverse movement range
± 125 mm
(Table top, Left&Right)
Table height 660 mm
Max. patient weight (kg) 250 kg
Scatter radiation grid 215 lp/inch, SID 100/180, Ratio 10:1, Removable
Auto tracking Longitudinal tracking of detector with longitudinal travel and rotation of
Wall Stand
Up/Down range (mm) 1,350 mm (420 ~ 1,770 mm, floor to Bucky center)
Up/Down method Manual
Scatter radiation grid 215 lp/inch, SID 180, Ratio 10:1, Removable
Automatic Exposure Control (AEC) AEC function prevents excessive radiation exposure
Network Specifications
DICOM functions
- DICOM Verification
- DICOM Modality worklist Interface with HIS / RIS with Auto refresh option
- DICOM MPPS Send the status of exams to HIS / RIS
- DICOM Storage Send Image (DX or CR IOD) and GSPS to PACS
- DICOM Storage commitment
- DICOM Grayscale print Support DICOM printers
- DICOM Query / retrieve Query and retrieve DX and CR images from PACS
- DICOM GSPS Send overlay information along with the Image
- DICOM Media exchange (DICOM DIR) Patient images export to DVD/CD
Installation Data
Power connection 3-Phase, 380/400 VAC, 50/60 Hz
Power rating 90 kVA
Wall stand 1,927 mm x 510 mm x738 mm
(W x D x H)
Patient table 4-Way movement
2,200 x 750 x 660 mm
(W x D x H)
Examination room Floor-Mounted
4,500 x 3,000 mm
Minimum Ceiling Height : 2,200 mm
ky center)
able
on exposure
option
to PACS
om PACS
mage