Shenzhen Kindle Medical Devices Co.

, Ltd

Operating Manual

KD760 Semi-auto Chemistry Analyzer


KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

CONTENT

Chapter 1 Introduction........................................................................................................................................2

1.1 Principle of measurement............................................................................................................................. 2

1.2 Technical Parameters................................................................................................................................... 2

Chapter 2 Instrument Installation.................................................................................................................... 3

1. Working condition.............................................................................................................................................3

2. Instrument Installation..................................................................................................................................... 3

Chapter 3 System Login..................................................................................................................................... 4

Chapter 4 Measurement..................................................................................................................................... 8

1.Measure Flow Chart..........................................................................................................................................9

2. Select test for measurement........................................................................................................................ 10

3. Edit patient information................................................................................................................................. 10

4. aspirate and measure....................................................................................................................................10

5. Result display..................................................................................................................................................11

6. Program test and edit result report in the measure interface..................................................................11

Chapter 5 Report................................................................................................................................................... 12

1. Edit the patient information...........................................................................................................................13

2. Edit sample info..............................................................................................................................................14

3. Edit result.........................................................................................................................................................15

4. Print report.......................................................................................................................................................16

Chapter 6 Program.............................................................................................................................................17

Chapter 7 Group Setup.....................................................................................................................................20

Chapter 8 Q.C......................................................................................................................................................22

Chapter 9 UTILITY.............................................................................................................................................. 24

Chapter 10 Maintenance...................................................................................................................................31

1. Maintenance................................................................................................................................................... 31

2. Troubles handle..............................................................................................................................................31

- I -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 1 Introduction

1.1 Principle of measurement

This Bio-chemical analyzer is mainly used for the quantitative analysis of the clinical bio-chemical items.
its working principle follows the Lamber-Beer law:
A=Kbc=lg(I0/I)
Among them:
A: Absorbance
K: Absorbent coefficient
b: Thickness of the Liquid
c: Solution’s Consistency
I0: Incident light’s Strength
I: Transmitting light’s Strength
According to the above formula, if the liquid thickness and the incident light’s strength is fixed, as long as
we get the intensity of transmitting light which goes through the liquid, we can calculate the consistency
of the solution.

1.2 Technical Parameters

Lamp: Quartz-halogen lamp, 6V 10W.


Flow cell: quartz, volume: 32uL.
Liquid thickness: 10mm.
Pump: peristaltic pump.
Filters: 340nm, 380nm, 405nm, 505nm, 546nm, 578nm, 620nm.
Photometric range: -0.2 – 2.3 Abs.
Blanking:Automatic zero setting.
Measure method: END POINT, KINETICS, MULTI STD, TWO-POINT, BICHROMATIC, FACTOR,
SUBTRATION, DIVISION, POLAR, ABSORBENCY.
Absorbing samples volume: ≥500ul, adjustable.
Printer: print with A4 dimension.
Display: big-screen display
CPU: Pentium
Hard Disk:80G
Keyboard: standard 101 keys keyboard
Mouse: photoelectric mouse
Memory: 256M DDR memory
Dimension:435mm in Length * 450mm in Width* 160.5mm in Height

- 1 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 2 Instrument Installation

1. Working condition

KD760 Biochemistry analyzer is a precise electronic instrument, in order to make the instrument working
normally and get the accurate results, the environment must follow with certain requirement, such as
below:
Environment temperature: 10°C~30°C
Relative humidity: 20%~80%
Atmospheric pressure: 86KPa~106Kpa
Ground-line: It should be a special ground-line for the medical instrument; while connecting to the earth,
the resistance must be less than 0.5Ω.
Additional requirements: the instrument must be put on a platform which has enough room, also be
stable and level without any vibration. It must avoid fans and air-conditions, and so on, in order not to
affect the accuracy of temperature control. It is forbidden to put in the environment with high temperature,
big sunshine or much dust. And do not put it away with centrifuges, refrigerators or thermostats or
share a same power socket with them.

2. Instrument Installation

1) Open the instrument’s packing box. Check carefully to sure the instrument,fittings in the box match
with the packing list . If there is any absence or damage, contact with the suppliers in time;
2) Take out the instrument carefully after checking, and then put it on a stable, smooth working platform;
3) The apparatus back-panel sketch map is shown below; connect the display to the VGA port, keyboard
to the KB port, and mouse to the mouse port.
4) Take out the power lines, connect one side to the power jack which is in the middle of the back panel,
then connect the other side to the three phase power socket which has the earth-line;
5) Find a waste bottle; put the waste tube which is in the back panel into the waste bottle;
6) Use a beaker or other container to get a cup of DI water;
7) Turn on the display and the power switch in the back panel, press ‘Power’ key in the right side of the
mainframe to turn on the apparatus.
8) Double click the “KD760 Biochemistry analyzer” icon in the desktop, or click “Start”, then “All
Programs”, then click “KD760 Biochemistry analyzer”.
9) Incubate at least 15 minutes after turn on the instrument;
10) While entering the measure interface, please follow the English instruction.

- 2 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 3 System Login

Before you enter the “KD760 Biochemistry analyzer” program, system will initialize at first.

After initialization, login interface will appear as below:

The user name lists all the doctors in the clinical labs (the information of the doctors can be edited in the
“doctor info” page of “utility” form )
When the “user Type” is administrator ,you can operate all the function of this apparatus.
When the “user Type” is operator, you can measure, edit and print the result report. But you havn’t the
right to edit items, groups, doctor information and basic information in “utility”, and some operation in the
measure interface, such as right click to edit item, check the result report.
If the doctor information(refer to chapter 9, “doctor info”) has not been set up, the system default user
would be “system administrator” and “system operator”.
After inputting the correct password(the default password is 123), you can enter the main interface.

- 3 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

In the main interface, you can see 8 buttons: ‘measure’, ‘report’, ‘Q.C.’, ‘Program’, ‘utility’, ‘help’,
‘exit’.

Click button, enter the measure interface, and measure.

Click button,enter into the result report interface, then there you can edit patient

information and print report.

Click button, enter into the ‘QC’ interface,there you can take the QC operation.

Click button,enter into the setup interface, where you can setup parameters of each
Test.

Click button,enter into the group edit interface,where you can combine the
function-related Tests, so that you can directly select them while measuring and adding results.

- 4 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Click button, then enter into the system maintenance interface, there you can set and

revision temperature, ASP Volume, observe ADC, measure absorbence, set alarm, set basic info, set
doctor info, set report format, etc.

- 5 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 4 Measurement

In the main interface, click button to enter into the measure interface, like below:

Patient information Kinetic result graph

Click by mouse
right to test Test information
selected

results

Group and
tests in the
group

Click by mouse
right to edit pick

Test list and group list on the left of the form; measure results on the right of the form , on the upside of
middle, there is patient information pane, including brief information of the patient, then below is brief
information of the selected test;on the downside of middle, there are all the results measured today.

- 6 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

1.Measure Flow Chart

Selected test

Input patient info

Is the first time measured


No
YES

Aspirate Clean solution

Aspirate DI Water

No
Setup blank in the item

Yes
No
Is the first time measured

Yes

Abosrb blank
No

Setup standard in the

No Yes

Is the first time measured

Yes

Absorb STANDARD

Absorb Sample

Measure

- 7 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

2. Select test for measurement


Select test in the tests list or select test in the tests list of group
3. Edit patient information
In the patient information column, please input or select a patient ID in the Patient ID ComboBox

If the patient information has been edited which displayed includes the

number, name, and type, sample No. and sample type. If there is a new ID, will display the default value.

4. aspirate and measure


1) Wash pipe

Select a new test, there will be a cue in the cue line in the bottom.

Click or press the ‘aspirate’ button under the aspirate needle, it will begin to aspirate the

clean solution to wash the pipe, if you want , you can click button to wash the pipe again.

2) aspirate DI water
While beginning to measure the first sample of the test, the cue line will suggest you that ‘Ready for

aspirate DI water’. Absorbence of DI water is need to adjust zero, click button, zero again.

3) aspirate Blank
If test requires measuring reagent blank or sample blank, the cue line suggest that ‘ready for reagent
blank’ or ‘ready for sample blank’, when test is operated the first time, or test parameter be changed,
Click ‘aspirate’ button to absorb blank. if you want to re-measure the blank, you can

click or .
4) aspirate Standard solution
The test require measure the standard solution when be measured the first time, or test parameter be
changed, the cue line suggest that ‘Ready for aspirate standard’; if the method of the test is
multi-standards, the cue line will suggest “ready for aspirate STD1”, “ready for aspirate absorb STD2” …

in turn. If you want to re-measure the standard,you can press button.

5) aspirate Sample
After DI water, blank and standard (if required) solution had been measured, the cue line will display that
‘Ready for aspirate sample’, when sample is ok,click ‘aspirate’ button, the sample solution will be
measured. After measured, the cue line will display ‘ready for aspirate sample’ again; if select ‘ID
Increasing’, the patient ID will automatically add one. If selected ‘Fixed Patient Type’, the newly increased
patients will keep this type.

5. Result display
All the result will be displayed in the right side of the bottom of the interface.
Blank line: the result of reagent blank and sample blank using non-kinetic method is original value, while

- 8 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

using kinetic method is the difference.

Standard line: after the STD solution be measured, the ABS volume will displayed at the ABS column,
and factor will be calculated according to the standard absorbance, then display at factor line.

Sample line: after sample solution had be measured, absorbance and result will be displayed in the ABS
column and result column.
Table and curve
If the method of the test is Kinetic or Two-point. Table of curve will describe the abs volume of each sec

6. Program test and edit result report in the measure interface


Select test, click by mouse right , then you will get the pop-up menu, like below:

Select ‘Test setup’, you can program test, if you want to know how to program test refer to Chapter
6-‘Program’ .
Select ‘Results of the test’,a form will show all results of the test had been measured today.
select one result of today’s measured results , click right mouse , then you will get the pop-up menu,
like below:

- 9 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Click ‘Patient information’ and you can edit the report of the patient,you can refer to Chapter 5-‘Result
setup’ to know how to edit patient report.

- 10 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 5 Report
you can edit patient info, sample information and result in the report interface In the main interface click

to enter ‘report’. The report interface has two page one is test page and the other is patient
page
Test page: the table in the left is list test ID and test name, select and click one, information of the test will
be showed at the top of the right side; the table in the bottom will list result of the test had been
measured

Patient page: in the left of the screen, the table list the patient ID and name, select and click one , the
information of the patient, and the tests has been measured and the measure results are displayed in the
right side.

Edit report step by step


1. Edit the patient information
Select and click one at the patient ID list, then information of th patient including age,blood type,Bed
No.,Department,address,tel will be showed at the top of right. And the diagnose given by doctor will be
show underside.
If the patient ID is newly, the information of this patient is all blank except the patient number, so you can
edit the name, age, blood type, sickbed number, section office, patient address, clinic diagnosis of the
patient in the patient information column.

- 11 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Add report When the patient list does not have the required patient ID, press

(the shortcut is “Ctrl+A”), or right click in the patient list table, select ‘New Patient’ in the pop-up

menu , then you can edit information of the Patient ID,and you can

press to saved change.

Delete report Select the patient ID. which you want to delete in the patient list, press

(the shortcut is “Ctrl+D”), or click mouse right in the patient list table,and you will get the pop-up
menu,and click ‘Delete patient ’,you will delete the patient.

Save report press (the shortcut is “Ctrl+S”),or you can click ’Save patient’ at Pop-up menu

to save change.

2. Edit sample info


Several samples maybe refer to the same Patient ID. The samples refer to the same patient ID list at
the sample ID list ,you can select and Click one of sample ID in the sample list, then information of the
sample will be show at the panel , include sample ID, sample type, doctor, sample date,Referred date,
etc.

If you click ‘Add sample’ to add sample,the default ID. of the Sample is to add one of the large number of
the exisit sample ID.
If the sample ID is a newly sample type and the sample date,referred date should be all default volume.
There is a pane in front of the sample, while ticking this pane means this sample can be printed, contrarily
means not to print it.

Add sample press , or click mouse right to select ‘Add sample’ from the pop-up

menu
.

Delete Sample press , or right click the sample list, select delete sample from the

pop-up menu.

3. Edit result

- 12 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

choose and click one sample ID, the result table will list all the measure refer to the sample ID and the
patient ID.
Add result the results measured by the other apparatuss can be added to the report by ‘adding result’.

When press , or click mouse right in the result table, then select ‘add result’ in the

pop-up menu, so that can enter into the add result interface.

Delete result choose one or more result items to be deleted in the result table, press

, or click mouse right in the result table, then click ‘delete result’ in the pop-up menu.

Adjust result position If you want to adjust the item order in the result report, you can select the item,

press or , make the result table in a order you want.

After finish edit all the info, press button or use shortcut ‘Ctrl +S’ to save the modification.

4. Print report
Select samples and results to be printed.
There is a small pane in front of each sample item or result. item

- 13 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Click the pane, then a ‘√’ will appear or disappear.

When the pane has a ‘√’ in it, this sample item or result item will appear in the print and print preview,
contrarily they won’t appear.

Select the report to be printed


Select the patient ID in the Patient ID list, or press ‘SHIFT’, then click the report to be printed, and then

press ,or select Print’ in the pop-up menu while right

clicking the report. And if you want to preview the report, press .

- 14 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 6 Program
This chapter mainly introduces how to programe parameter of test while first time to use it. every time
change the reagent.,we suggests that you should follow the reagent instruction to programe parameters
again.

Press in the main menu to enter this interface.

1. Parameter setup
In the left side, there is a test list, choose and click one, the concrete definition is displayed in the right
side. If this test parameter has already been programed, the concrete parameters will be displayed, these
parameters are default values.

Test
information

Test list Referrence


range volume

STD
volume

The Test ID, test name and the test method are all should be input, while select different test methods,
parameter require to be changed.Test can be measured in the apparatus when it’s method is KINETICS,
END POINT, NON-LINEAR END-POINT, MULTI STD, TWO-POINT, SAMPLE BLANK, DUAL
WAVELENGTH, CALCULATE method should depend on other item’s results to calculate, while if test
method is qualitative,it should measure by other apparatuss, then add result in report
The concrete steps of inputting parameters are shown below:
select Supplier, batch No. and validity of reagent into the line.
Select methods to measure this test.

- 15 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

In the drop-down menu, select the test method.


Input the test parameters. Like the qualitative analysis, only need to input the normal value range. Like
the CALCULATE, only need to input the equation and the referrence value range; like KINETICS, END
POINT, DUAL WAVELENGTH, MULTI STD, FACTOR, TWO-POINT, they all need to input the right
parameters, these parameters can be obtained from the reagent instructions.
3) Decimal digits: the decimal fraction digits the result can have.

Fuse KINETICS, please find the test coefficientand fiil in , ranging from

‘0--59999’.

5) Select or input the right unit in .

6) Choose the right temperature volume in , you can choose ‘37oC’, ‘30oC’,

‘25oC’ ‘room temperature’ three temperature control states.

7) choose the right Primary filter in there are 340nm, 380nm, 405nm,

505nm, 546 nm, 578nm, and 620 nm 7 filters .

8) Input the right ASP Volume in according to the instructions.

9) If test require reagent blank and sample blank, click the pane in front of the ‘reagent blank’ or ‘sample

blank’, then a ‘√’ appears in the pane. 10) postponed time,Interval


time, Data number and linearity range.
11) Input referenced range of test;
12) Standard: if method of the test is END POINT, one standard is required; KINETICS, don’t require
standard volume; TWO-POINT, one standard is required.and MULTI STANDARD ,the number of
standard is required by instruction of reagent.
About other parameters: the print level is a random number among 1-99999, test arrange of report print
or preview according to print level from low to high, the tests which have the same level must follow the
item ID.
2. Add, delete, and save Test

If want to add a new test, click , or click mouse right click at test list, then click ‘add

test’ at the pop-up menu , or use shortcut ‘Ctrl+A’. The default test ID. is to add

- 16 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

one to the current highest test ID, the default test method is END POINT, All the parameters but test ID
can be modified, the Test ID once been saved cannot be modified.

When to delete test, press , or click mouse right click at test list, select ‘delete test’ in the

pop-up menu, or use shortcut ‘Ctrl+D’, then can delete test selected.

save test, press , or click mouse right at test list, select ‘Delete test’ in the pop-up menu, or use

shortcut ‘Ctrl+S’, then can save the test.

- 17 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 7 Group Setup

click at the main interface, can enter into the group setup interface.

Test list

Test in the
group

Group list

There are three list, group list at left, list of all test at middle and list of test in the group selected at right
side.

1. Define the group


Select one group in the group list, if want to modify the group name, input a new name

in .

The list at right side shows all tests of the group, if you want to add new test, please select one test at

middle list, then double click mouse at the test, or click ;

delete test of the group, select test,then click mouse double or click . Then you can

click or use shortcut ‘Ctrl+S’ to save change

2. Add, save and delete group

- 18 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

add new group, click , or use shortcut ‘Ctrl+A’. The new group ID is the highest group ID

plus 1, group name is blank and no test in the new group em. The group ID can be modified, but cannot
overlap with the ID already exist.

save new group or modified group, click , or use shortcut ‘Ctrl+S’ in the group list to save

the definition.

delete a group, select a group and click , or use shortcut ‘Ctrl+D’.

- 19 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 8 Q.C.

Click to enter into Q.C. interface. the aim of quality control is to check and control the

precision of the instrument while working, monitor stability; heighten the consistency of the test results,
so that to ensure the reliability of the measure result of each sufferer sample.

Q.C. parameter

Test list

Q.C. Curve

Print mode

Result of the
Result of Q.C.
test

Test list in the left of the Q.C interface, all tests which can do quality control, and in the right side, it shows
the QC parameters and QC results.

1. QC parameter
Choose and click test in the test list, the parameters of QC test will be showed in the top of right side.
The QC parameters are divided to QC!,QC2 and QC3, parameter including Lot, target,Deviation and

Supplier. After that, click to save the QC parameter.

- 20 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

2. Refer to QC result
When select a Lot , the results refer to the Lot will be shown by a table and a curve (when the number of
results over four)

While you click ‘table’, the results will be shown from the first to the last one by a table format.
Click ‘curve’, the curve mode regards the average of the QC value plus SD as upper limit, average
subtracting SD as lower limit. Connecting all the QC value as a curve, you can observe the QC value
falling in which area.

3. The procedure of QC measure


In the QC test table, select one test and select Lot,QC Type (QC1, QC2 and QC3) cue column in the
bottom of the interface will suggest:

click button or shortcut ‘F2’ to aspirate clean solution, after that,cue line will displayed
‘Ready for DI water’, after that it will displayed ‘Measuring’, then will suggest ‘Ready for aspirate QC
solution’, and then begins to measure. At last it will show the absorbance and result in the result line in
the bottom of the left side.

If there are more than 4 QC results in one batch, it will show the times, mean,SD and CV.

4. Operation to the QC result


If you want to delete a QC result, in the QC result table, then select the result, click or

click mouse right then click in the pop-up menu or use shortcut ‘Ctrl+D’.
If you want to delete the whole results refer to the same Lot and same test, choose the lot, click

or use shortcut ‘Ctrl+F’.

5. Report setup and preview

choose a right print mode in

click to print the QC report,or click to preview the report.

- 21 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 9 UTILITY
There are ‘temperature’, ‘ASP Volume’, ‘ADC’, ‘Absorbence’, ‘Alarm’, ‘basic info’, ‘doctor info’, ‘report list’
8 pages of utility .

1. Temperature
The set temperature is the temperature that the flow cell should keep when it isn’t at measure.cell
temperature is temperature measured by sensor
thermometer temperature is the temperature measured by thermometer.

When you want change set temperature, input new volume, then click .

Revise temperature: If you want to revise temperature measured by sensor and thermometer

click , and then it will suggest you inputting the protecting password(123). This

is to protect the temperature revise factor from being modified at one’s will. After you input the
password,temperature page will appear like blow:

- 22 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Please fill in the thermometer temperature blank line with the temperature value measured by the
thermometer, and then you press ‘Revise Temperature’ button to calculate the temperature revising factor.
The system will adjust the temperature and make it approach to the actual temperature.

2. ASP Volume

Revise the ASP Volume: input the quantity you want to aspirate theoretically in the ‘theory’, then

click , after that you can survey real volume has been aspirated, fill it in the ‘real’, then

click , that can revise the aspirate factor.

Update Clean asp volume: input the volume at the edit column then click .

3. ADC

It displays the instrument’s internal AD convert parameters, for the repairman to use.

- 23 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

4. Absorbence

The aim of The absorbence measurement is to examine the variety of the absorbence at different period,
and then the system will revise it in order to make the result more credible.

Click to begin measure absorbence, the screen will display ‘ready for aspirate DI

water’,click or use shortcut ‘F2’ to aspirate DI water, you can get Raw data and

absorbence after having measured at first filter. if raw data in normal range, cue will suggest ‘OK!’, else if
raw data is abnormal cue will display ‘failed’.then the next filter will be examined, the process finished at
last filter be examined.

5. Alarm
The aim of set Alarm is to give user there may appear some abnormal instance in his operation.

The Deviation between Blank and Std ABS—the Deviation between blank and standard absorbence:
When measuring the standard in the measurement interface, if the deviation between blank and standard
absorbence is less than the set alarm volume, it will suggest ‘Std Absorbency over range,try again? ’.

- 24 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

The Deviation between sample and set temperature: if the deviation between the sample temperature
and set temperature over set alarm volume ,the temperature will be show at measure or Q.C. interface
in red color .

6. Base info
The base information has saved the parameters of instrument set

In this interface you can edit all the info, including adding or deleting unit, department, blood, sample
type, duty, the title of technical post, test name and so on. Or you can add new items, sub-items or delete
items and sub-items.

click to save after any modification, otherwise the modification would be invalid.If the

database happens to serious trouble, it needs to adjust the state of the database to the leaving factory

settings, just click . Attention, all data of the patient information and item information

would all be lost.

7. Doctor info

- 25 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

To edit doctor information, input or select the name, sex, duty, title of the doctor . doctors of clinical lab
it needs to input password because they need input the password to log in this program.

8. Report list setup

you can set Header, hospital address, remark,fax and tel of report.input these, if it’s blank the print and
preview of the report won’t be display. Front setup: the report title and the text will be printed and
displayed following the setting front.

- 26 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

Chapter 10 Maintenance

1. Maintenance
It is important suit reagent to instrument,so that instrument can work normally and measure result reliable
In this chapter. The method of how to maintaining of the KD760 Biochemistry analyzer will be introduced,
please follow , to be sure the instrument work normally
1) Maintenance for every day
Clean the cell and piping with DI water repeatedly after measuring sample every day ,so to get rid of
residual reagent and protein of surm especially remains of flow cell. Finally,flow cell and tube should be
full of DI water.

2) Maintenance for every week


Clean the piping twice with DI water after measuring sample, then aspirate the CLEAN solution three
times (Main Component: NaClO, diluted to 5%), let the CLEAN solution stay at the tube for 10 minutes,
and then clean the piping with DI water for 3 times. Finally, it should be full of DI water.

3) Cleaning instrument’s appearance


Wipe the instrument appearance with wet cloth (Turn power off, and pull out the power socket).
Attention: Clean the instrument with clean water,neutral or non-causticity clean solution, but do not use
organic solution or hot water, otherwise instrument will be distorted or damaged.

2. Troubles handle
phenomena cause method
a) The aspirate pump doesn’t work
a) Tighten the screw of the pump.
No b)Pump pipe is loosen
b) set new ASP volume
aspiration c) Aspiratevolume is zero
d) replace new Pump pipe
d) Pump pipe is damaged
a) Turn on the power.
a) The power is off
No Display b) Check the power plug of display
b) Signal line of display is loosened
c) Check Signal cable of display.

a) examine again with DI water


a) do not aspirate DI water
Failure of b) aspirate solution to vent the bubble from
b) bubble in the flow cell.
Absorbance flow cell
c) flow cell is dirty
examination c) Clean the flow cell with clean solution
d) The lamp is aging
d) replace new lamp

a) No absorbance examination a) examine absorbance again


b) replace with valid reagent
b) Invalidate reagent
results is c) Adjust the ASP volume
c) aspirated solution Inadequate
unstable d) measure BLANK and STANDARD again
d) No STD.
e)set another filter to test in program
e)filter does not match reagent instruct
f) carefully ,standard operation
f) air in flow cell

- 27 -
KD760 Semi auto Chemistry Analyzer Operating Manual Rev: A/0

a) Exit process
b) Reset
c) Power off and restart
d)click mouse Right at “My computer” in the
desktop, select ” property”, then “hardware”,
1.Multi COM messages send out at one
then “equipment management”, click “+” in
time
front of the “port”, then click mouse right
2.Dead-lock occurs while processing

COM error COM messages


“COM1”, select “uninstall”, then press
3. COM driver malfunction
to scan the hardware driver, reship the COM
4. COM communication parts
driver.
malfunction
e) The above steps should go step after
step, one cannot solve the problem then go
the next.
f) You’d better not do other operations when
measuring and Q.C.interface.
If the product trouble has not been fixed yet or has other troubles, please contact with our after service
department.

深圳市坤德生物电子有限公司
Shen Zhen Kindle Medical Devices Co., Ltd.

Address:4F,Buliding 2, The Hong Zhen Gang Industrial Area Gushu, Xixiang, Bao'an
District Shenzhen,China
Tel: 86-0755-27512545 Fax: 86-0755-27512545

- 28 -