Standard calibration procedure for

Draeger Polytron 7000


Rev. 01 Date: 05. May 2008

CCR Pswd: 4700 (N2 Plant).......


CALIBRATING OXYGEN ANALYSER POLYTRON 7000

• The sensor can not be calibrated at zero, because this is set at fabrication. The Zero
calibration tools in the sensor will only do a check of the reading of the sensor, and
not a calibration as such. Ref. page 35 in vendor manual.

• The span can be calibrated with instrument air containing 20,9Vol% oxygen.

To calibrate the analyzer, complete the following steps:

1. Turn the selector valve to SPAN position. Supply span gas (instrument air).
Check measurement.
⇒ If the reading is not as expected; go to paragraph 3
⇒ If the reading is OK; go to paragraph 2

2. Turn the selector valve to Zero position. Supply Zero gas or any other test reference
gas. Check the low point measurement.
⇒ If the reading is not as expected; replace the sensor cell
⇒ If the reading is OK; the calibration check is complete. Turn the selector valve
back to SAMPLE position. No further actions required.

3. View display (value as example)

To access the analyzer menu:


Push the ▼m button until the next view shows up.
19.5
O2 Vol %

← Back to measure
Push the OK button.

Enter password ↓
Standard calibration procedure for
Draeger Polytron 7000
Rev. 01 Date: 05. May 2008

Password Push:
- 3 x OK
___2 - 3 x ▲ button to enter the value 2 as the last suffix
- 1 x OK button.

← Back to measure
+ Information Push:
+ Calibration - 2 x ▼ to select Calibration
+ Settings - 1 x OK button.

- Calibration
+ Zero calibration Push:
+ Span calibration - 2 x ▼ to select Span calibration
+ Auto Calibration - 1 x OK button.

- Span calibration
+ EC-O2 Push:
- 1 x ▼ to select EC-O2
- 1 x OK button.

- Span Cal. EC-O2 1/7


+ Back to menu This view shows expected values.
Cal. Gas: O2 Push:
Cal. Uni : Vol% - 1 x OK button.
Consentr.: 20.9
Next ↓
Standard calibration procedure for
Draeger Polytron 7000
Rev. 01 Date: 05. May 2008

- Span Cal. EC-O2 2/7


+ Previous Turn the sample selector valve to SPAN position to supply
Apply span gas: span gas.
O2 Push:
with flow rate - 1 x OK button.
500 ml/min!
Next ↓

- Span Cal. EC-O2 3/7


↑ back Wait until the value is stabilized.
Gas flow: 04:12 Push:
- 1 x OK button.
Value: 20.0 Vol%
Value stabilized?
Calibrate ↓

- Span Cal. EC-O2 4/7


↑ redo You can now view expected and actual measured value.
Vital: Remove span gas by turning the sample selector valve.
Gas: 20.9 Vol%
Value: 20.9 Vol% Push:
Remove gas! - 1 x OK button.
Accept value ↓

- Span Cal. EC-O2 5/7


Next date for calibration is displayed.
Calibration interval: Push:
270 days - 1 x OK button.
Next cal. date:
17.01.2009
Next ↓
Standard calibration procedure for
Draeger Polytron 7000
Rev. 01 Date: 05. May 2008

- Span Cal. EC-O2 6/7


Sensor data updates.
Write sensor data Next display comes automatically.
Please wait…….

- Span Cal. EC-O2 7/7


↑ Back to menu Push:
Value out of - 1 x OK button.
Alarm rage?

Value: 20.9 Vol%

4. To return to main view:

Push:
- 1 x ▲ button and then OK button
- 2 x ▲ button and then OK button
- 2 x ▲ button and then OK button

The Calibration is now complete.