• The sensor can not be calibrated at zero, because this is set at fabrication. The Zero
calibration tools in the sensor will only do a check of the reading of the sensor, and
not a calibration as such. Ref. page 35 in vendor manual.
• The span can be calibrated with instrument air containing 20,9Vol% oxygen.
1. Turn the selector valve to SPAN position. Supply span gas (instrument air).
Check measurement.
⇒ If the reading is not as expected; go to paragraph 3
⇒ If the reading is OK; go to paragraph 2
2. Turn the selector valve to Zero position. Supply Zero gas or any other test reference
gas. Check the low point measurement.
⇒ If the reading is not as expected; replace the sensor cell
⇒ If the reading is OK; the calibration check is complete. Turn the selector valve
back to SAMPLE position. No further actions required.
← Back to measure
Push the OK button.
Enter password ↓
Standard calibration procedure for
Draeger Polytron 7000
Rev. 01 Date: 05. May 2008
Password Push:
- 3 x OK
___2 - 3 x ▲ button to enter the value 2 as the last suffix
- 1 x OK button.
← Back to measure
+ Information Push:
+ Calibration - 2 x ▼ to select Calibration
+ Settings - 1 x OK button.
- Calibration
+ Zero calibration Push:
+ Span calibration - 2 x ▼ to select Span calibration
+ Auto Calibration - 1 x OK button.
- Span calibration
+ EC-O2 Push:
- 1 x ▼ to select EC-O2
- 1 x OK button.
Push:
- 1 x ▲ button and then OK button
- 2 x ▲ button and then OK button
- 2 x ▲ button and then OK button