Вы находитесь на странице: 1из 8

The Anesthesia Ventilator

Why is the piston replacing the bellows?

For many decades, and millions of


anesthetics, the bellows anesthesia

I N S I G H T
ventilator has been a safe and
effective clinical device. Indeed,
Draeger anesthesia ventilators
based upon the bellows design
continue to be used in all parts of
the world. More recently, Draeger

M E D I C A L
Medical has been producing
anesthesia ventilators using the
piston design. Why would a
company with decades of
investment in bellows ventilation
technology decide to base future
anesthesia ventilator products on a
piston design? The answer lies in
the advantages inherent to the
piston design for producing a
versatile, reliable anesthesia
ventilator now and in the future.

The Clinical Requirements of an


Anesthesia Ventilator

The clinical needs for mechanical squeezing the reservoir bag,


ventilation in the operating room sometimes for several hours. The
have changed significantly over earliest anesthesia ventilators were
time. The earliest anesthesia bellows designs that essentially
delivery systems were open inhalers automated the process of squeezing
intended to deliver volatile the reservoir bag, freeing the
anesthetics while patients breathed anesthesia provider from this
spontaneously throughout the repetitive manual activity.
surgical procedure. Breathing
circuits with reservoir bags were Given the variety of patients that
developed to contain the anesthetic require anesthesia for surgery today,
gases and allow clinicians to the performance demands on the
ventilate patients manually. With the anesthesia ventilator have increased
advent of muscle relaxants and dramatically. The demand for
narcotics, positive pressure performance equivalent to that of an
ventilation became essential and intensive care ventilator, while
was accomplished by manually maintaining the ability to deliver

Emergency Care · Perioperative Care · Critical Care · Perinatal Care · Home Care Because you care
The introduction of the Laryngeal
Mask Airway led to a reemergence of
spontaneous ventilation during
anesthesia. Ventilation modes
commonly used in the ICU to
augment or support spontaneous
ventilation include Synchronized
Intermittent Mandatory Ventilation
(SIMV) and Pressure Support
Ventilation (PSV). Implementation of
these modes requires that the
ventilator controller sense either a
pressure or flow change in the
breathing circuit associated with
inspiration to trigger ventilator
Fig. 1: The volume delivered by the piston is determined by the distance the
piston moves. When a volume is set to be delivered (eg. 750 mls), the piston is support. Once inspiration is
moved the distance required to deliver the set volume to the patient. detected, the preset amount of
ventilator support begins. In the
case of SIMV, a volume or pressure
controlled breath is delivered
anesthetic gases, has been the measuring the pressure in the synchronized with the start of
motivation for redesigning the breathing circuit with feedback inspiration. The breath that is
anesthesia ventilator. The clinical control of the ventilator during each delivered is very similar to the
needs for ventilation in the breath. This feedback control breaths given by the ventilator
operating room fall into two broad reduces the inspiratory flow as the during controlled mechanical
categories: controlled mechanical lungs fill resulting in a decelerating ventilation. In the case of PSV, the
ventilation and supported flow pattern. The rigid coupling trigger is used to adjust the constant
spontaneous ventilation. Both between the piston and its drive pressure in the breathing circuit
bellows and piston ventilators have mechanism allows for fine control during inspiration and expiration.
features designed to serve these over the movement of the piston and The volume that is delivered during
needs although the performance of continuous adjustment of PSV will depend upon the magnitude
these ventilators is not identical. inspiratory flow to maintain the of the patient’s effort and the degree
desired inspiratory pressure. of pressure support. When using an
The most common mode of
controlled mechanical ventilation
used in the operating room is
volume controlled ventilation
whereby a preset tidal volume is
delivered by the ventilator to the
patient. For a patient of average size
with healthy lungs, it is not difficult
to deliver the appropriate tidal
volume safely. The challenge is to
deliver tidal volume accurately when
lung compliance is very poor (eg.
patients with ARDS) and/or when
the patient is very small. One of the
major advantages of the piston
ventilator is the ability to deliver
tidal volume accurately to all
patients under a large variety of
clinical conditions.

Volume controlled ventilation is


appropriate for most patients, but
pressure controlled ventilation offers
advantages to some patients.
Pressure controlled ventilation
requires that the preset inspiratory
pressure is maintained throughout
the inspiratory cycle. Proper
implementation of pressure
controlled ventilation requires
ICU ventilator, the volume that can
be delivered is unlimited whereas
the volume of both bellows and
piston ventilators is limited by the
maximum size of the bellows and
the piston chambers respectively.
Modern bellows and piston
ventilators are designed with
sufficient volume capacity to meet
the needs of virtually all patients.

The clinical need for an anesthesia


ventilator that can provide the
capabilities of an intensive care unit
ventilator, while maintaining the
ability to deliver anesthetic gases
efficiently, is a major challenge to
ventilator designers. Inherent
limitations of the bellows design to
meet the clinical needs for advanced
ventilation in the operating room Fig. 2: To deliver a certain volume to a patient, the desired volume (eg
led to a decision to base future 750 mls) is delivered into the bellows compartment displacing the bellows
anesthesia ventilator designs on and pushing circuit gas to the patient. Since the pressure in the bellows
compartment will vary from patient to patient, the volume displaced by the
piston rather than bellows bellows will also vary. Furthermore, since the final position of the bellows is
technology. The piston design offers not known, the volume that was delivered by the ventilator is also not
advantages of more accurate volume known.
delivery and the ability to serve as a
platform for future development.
This monograph describes the major circuit gas into the breathing Superior Control of the Ventilator
advantages of the piston design in circuit. One common bellows
detail and addresses frequently ventilator design begins inspiration Irrespective of the type of ventilator
asked questions about piston with the bellows at its maximum being used, the volume delivered by
ventilators. volume and is calibrated to deliver a the ventilator into the breathing
volume of drive gas into the bellows circuit and the volume the patient
More Accurate Volume Delivery compartment equal to the volume receives are not identical. One
set to be delivered to the patient. As major determinant of the difference
The most common mode of the volume of drive gas enters the between the two volumes is the
ventilation used during anesthesia is bellows compartment, the bellows compliance of the breathing system.
volume controlled ventilation where moves to displace gas into the As the ventilator delivers gas to the
the clinician sets a specific tidal breathing circuit. For a given set breathing circuit, the pressure
volume to be delivered to the tidal volume, the pressure that increases. The increased pressure
patient. The piston ventilator design results in the breathing circuit is will compress the gas in the system
is uniquely suited to deliver tidal determined by the resistance and
volume accurately. Since the area of compliance of the breathing circuit
the piston is fixed, the volume and the patient’s lungs. Since the
delivered by the piston is directly pressure in the bellows compartment
related to the linear movement of will vary between patients (or even
the piston. When the user sets a between breaths), the gas driving
volume to be delivered to the the bellows will be subject to varying
patient, the piston moves the degrees of compression that cannot
distance necessary to deliver the be predicted. Variable compression
required volume into the breathing of the drive gas is a fundamental
circuit. Furthermore, since the obstacle to accurate volume delivery
connection between the piston and by a bellows ventilator. This is
the drive motor is rigid, the position particularly true for small tidal
of the piston is always known and volumes and high inspiratory
the volume delivered by the piston is pressures. As opposed to the piston
also known. (FIGURE 1) design, the position of the bellows in
the bellows compartment at the end
When using a bellows ventilator, the of inspiration is not known. As a
movement of the bellows is result, volume delivered by the
controlled by drive gas which enters ventilator for a given breath is not
the bellows chamber and pushes known. (FIGURE 2)
system since the plateau pressure is
essentially constant throughout.
Control of the bellows ventilator
based upon pressure is difficult due
to variable compression of the drive
gas from patient to patient. Bellows
ventilators with compliance
compensation utilize a flow sensor
in the breathing circuit to measure
the volume delivered and to tell the
ventilator to increase the volume
delivered to offset the effects of gas
compression. Since flow sensors
ultimately measure volume, they
work best to ensure delivery of set
tidal volume when located at the
patient’s airway. In this location,
flow sensors are prone to inaccuracy
due to accumulation of moisture or
secretions. If the flow sensor is
Fig. 3: Effect of compliance on delivered tidal volume without
compliance compensation. Ventilator set to deliver 750 mls but located at the beginning of the
only 550 mls reaches the patient due to a compliance factor of inspiratory limb to reduce the
5 mls/cmH2O and peak pressure of 40 cmH2O. (Schematic of impact of moisture and secretions,
Ohmeda Excel) the set tidal volume is not delivered
* Modified from the Virtual Anesthesia Machine by permission from the Department
to the airway. Furthermore, if the
of Anesthesiology, University of Florida College of Medicine. For more information, flow sensor should fail or become
visit www.simanest.org unreliable, the ventilator must
revert to volume controlled
ventilation without compliance
and also expand the circuit tubing, the set tidal volume to the patient’s compensation.
therefore reducing the volume that airway irrespective of changes in
reaches the patient. Every breathing lung compliance. (FIGURE 5) Facilitate Advanced Ventilation
circuit has a certain compliance The ability to deliver volume Modes
factor which defines the amount of accurately simply based upon a
volume stored in the circuit for a pressure measurement is a unique The trend in anesthesia ventilator
given change in pressure. During advantage of the piston ventilator. technology is to eliminate the
volume controlled ventilation, the Pressure sensors are simple devices disadvantages of traditional
pressure that results when a set that are easily calibrated and can be anesthesia ventilator technology and
volume is delivered by the ventilator located anywhere in the breathing to increase the availability of
will vary between patients. Without
some means of compensating for the
effect of circuit compliance, as
pressure in the circuit increases, the
volume the patient receives will
decrease. (FIGURE 3)

Advanced piston ventilator designs


are able to compensate for the
compliance of the breathing system
by delivering enough additional
volume with each breath to ensure
that the patient receives the volume
set to be delivered. (FIGURE 4)
Draeger piston ventilators measure
the compliance of the breathing
system during the pre-use self-test
procedure. Once the compliance
factor is determined, only a pressure
sensor is needed to determine how
much additional volume should be Fig. 4: Effect of compliance compensation on delivered tidal
delivered with each breath to volume. Ventilator delivers 1000 mls to insure that 750 mls
reaches the patient due to a compliance factor of 5 mls/cmH2O.
compensate for the breathing system Note that peak pressure has increased to 50 cmH2O due to the
compliance. The result is delivery of additional delivered volume. (Schematic of Fabius GS)
Fig. 5: Plot of tidal volume
measured at the airway versus
airway pressure. Data obtained
using both adult and pediatric
lung simulators and both a
Draeger Medical Fabius GS
equipped with a piston
ventilator as well as a
traditional Narkomed 2B.
Airway pressure increases due
to reduced lung compliance.
Note constant volume delivery
to the airway as airway pressure
increases when using the
Fabius GS. When using the NM
2B, the delivered tidal volume
decreases as plateau pressure
increases.

intensive care modes of ventilation throughout the inspiratory cycle. flow is required to achieve the
in the operating room. The ability of The volume delivered to the patient desired pressure. Piston ventilators
the piston ventilator to deliver will depend upon the lung offer adjustable inspiratory flow
volume accurately enables the compliance. (FIGURE 6) settings. The default or initial flow
clinician to use volume controlled setting is adequate for most patients.
ventilation for all types of patients. Both piston and bellows ventilators For patients with relatively large
From neonates requiring very small can be designed to meet the needs of lung compliance, inspiratory flow
tidal volumes to adults with ARDS PCV. In both cases, the pressure in can be increased to ensure that
where accurate tidal volume is the circuit is measured and used to inspiratory pressure is rapidly
critical to ensuring oxygenation, control the movement of the attained. Limiting the maximum
Draeger piston ventilators are ventilator. As pressure builds in the inspiratory flow is useful to avoid
capable of meeting the clinical breathing circuit, the flow delivered overshooting the target pressure
needs. by the ventilator is progressively especially when lung compliance is
reduced generating the low.
The demand for modes of ventilation characteristic decelerating flow
in the operating room other than waveform. Since the goal of PCV is to SIMV has found application in the
traditional volume controlled develop the desired inspiratory operating room to facilitate
ventilation is also increasing. pressure as rapidly as possible, emergence from anesthesia as the
Pressure controlled ventilation bellows ventilators require a greater patient transitions from controlled
(PCV) has found application in initial flow than a piston design to to spontaneous ventilation. Both
children and adults who require overcome compression of the drive piston and bellows ventilators can
increased pressure to achieve gas. The flow required to achieve offer this mode of ventilation. As the
adequate ventilation during the desired inspiratory pressure in procedure is concluding, SIMV can
anesthesia. PCV requires that the the breathing circuit will vary with be used to ensure a minimal
ventilator deliver sufficient gas to lung compliance. When the lung amount of ventilation until the
achieve the desired pressure compliance is low, relatively little patient begins spontaneous
25.0 25.0
Fig. 6: Plots of pressure, flow
and volume obtained using a
20.0 20.0 Draeger Medical Fabius GS
to ventilate an adult test lung
15.0 15.0 using Pressure mode at
different lung compliance
10.0 10.0
settings. Note constant
pressure and less tidal
5.0 5.0
volume as lung compliance is
reduced. Also note
2.0 2.0
decelerating flow pattern.
0.00 0.00

60.0 60.0

30.0 30.0

15.0 15.0

0.00 0.00

-15.0 -15.0

-30.0 -30.0

-60.0 -60.0

1.250 1.250

0.750 0.750

0.250 0.250

0.00 0.00
2.00 3.00 4.00 5.00 6.00 7.00 8.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00

breathing efforts. As the patient ventilation on an anesthesia


begins to breathe, the ventilator will ventilator requires a means to
be triggered to begin inspiration in monitor for the onset of inspiration
concert with the spontaneous and exhalation and to maintain the
breaths. The clinician is freed from desired pressures throughout each
the task of periodically ventilating respiratory cycle. Both bellows and
the patient by hand. The ventilator piston ventilators are limited by the
will begin each breath from its volume of the ventilator chamber
maximum volume capability so that but maximum volume capabilities
sufficient volume is available to the are adequate for virtually all
patient. patients.
The use of laryngeal mask airways
has led to a dramatic increase in Due to the accuracy with which the
spontaneous ventilation in the piston can be controlled, advanced
operating room. Pressure Support ventilation modes are implemented
ventilation is used in the ICU to through software enhancements to
reduce the work of breathing the piston ventilator. The basic
associated with the breathing circuit piston design has proven itself to be
and endotracheal tube and also to a versatile platform for anesthesia
impose varying degrees of ventilator design.
respiratory muscle exercise.
Implementing this mode of
FREQUENTLY ASKED QUESTIONS 4. If advanced ventilation is to the airway. Furthermore, most
important, why not just use an exhaled volume monitors have an
1. What is fresh gas decoupling? intensive care ventilator? inherent accuracy of only +/-15%.

Fresh gas decoupling eliminates any Anesthesia ventilators are different 6. What is the difference in
interaction between fresh gas flow from intensive care unit ventilators compressed gas requirements
and the volume delivered to the in that they must be able to deliver between a piston and bellows
patient. One can adjust fresh gas inhalation anesthesia in addition to ventilator?
flow freely or even press the oxygen provide mechanical ventilation.
flush button during ventilation Whereas intensive care ventilators The piston ventilator does not
without concern for altering the can function in an open circuit require compressed gas as a source
volume delivered to the patient. Fresh configuration, the need to deliver of power whereas the bellows
gas decoupling is accomplished by inhaled anesthetics efficiently ventilator is completely dependent
the breathing circuit design and is requires that anesthesia ventilators upon compressed gas to function.
not a feature of the piston ventilator contain patient gases within the When using a cylinder source of
per se. In the case of the Fabius GS, breathing circuit. The purpose of compressed gas, the duration of
fresh gas decoupling is accomplished the bellows is to separate the gases time the anesthesia machine can be
by placing a decoupling valve driving the ventilator from the gases used will be significantly greater
between the fresh gas inlet and the being delivered to the patient. In a when using a piston ventilator since
breathing circuit. When the circuit similar fashion, the piston chamber the only gas consumption by the
is pressurized during inspiration, of a piston ventilator isolates the piston ventilator is from the fresh
the decoupling valve closes and gases that the patient will receive. gas flow. A full E cylinder contains
fresh gas is directed towards the In both cases, the total volume that 625 liters of gas. If fresh gas flow is
reservoir bag.(Figure 4) can be delivered per breath is set at 1 liter per minute, there will
limited by the maximum volume of be a supply for more than 10 hours.
2. Are all Draeger piston the bellows and piston chambers. The bellows ventilator will typically
ventilators identical? Standard designs offer sufficient not function for more than one hour
volume capability to meet the needs on an E cylinder due primarily to
Although all of the newer Draeger of virtually all patients. the compressed gas used to power
anesthesia workstation designs the ventilator.
utilize piston ventilators, these 5. Why does the exhaled volume
ventilators are not identical. Each measurement differ from the 7. Can I still ventilate the patient
ventilator is fully integrated with a set tidal volume? when using a piston ventilator
specific workstation and designed to if the power fails?
complement the functions available The set tidal volume is the volume
in that workstation. the clinician desires the patient to All Draeger anesthesia workstations
receive. In the case of a piston are equipped with battery supplies
3. How do I know the ventilator is ventilator with compliance to provide at least 30 minutes of
working if I cannot see it? compensation, the volume delivered power in the case of AC power
to the patient’s airway will equal the failure. If a total electrical power
Studies on safety in anesthesia have volume set to be delivered. For a failure occurs, the piston ventilator
documented that human vigilance bellows ventilator with a flow sensor will cease to function but manual or
alone is inadequate to insure patient at the inspiratory valve, the set spontaneous ventilation and delivery
safety and have underscored the volume will equal the volume of anesthetic gases will still be
important of monitoring devices. passing through that sensor. possible. Modern bellows ventilators
These studies have been reinforced Exhaled volume measurement in a are microprocessor driven and also
by standards for equipment design, circle system is typically performed require a source of electrical power
guidelines for patient monitoring at the expiratory limb adjacent to to function.
and reduced malpractice premiums the expiratory valve. This sensor
for the use of capnography and measures exhaled gases plus the gas 8. How can I detect a leak in the
pulse oximetry during anesthesia. that is compressed in the breathing circuit when using a piston
Draeger anesthesia workstations circuit during inspiration. When ventilator?
integrate ventilator technology with inspiratory pressure is high, the
patient monitors and alarms to help difference between measured In the case of a bellows ventilator, a
prevent patient injury in the exhaled volume and actual exhaled leak is recognized when the bellows
unlikely event of a ventilator failure. volume can be significant due to the fail to return to their starting
Furthermore, since the reservoir compliance of the breathing circuit. position and (instead) progressively
bag is part of the circuit during Draeger anesthesia workstations can fall in the bellows compartment.
mechanical ventilation, the visible use the compliance factor of the The leak may be observed but low
movement of the reservoir bag is breathing circuit to subtract the pressure and volume alarms are
confirmation that the ventilator is impact of compliance and obtain a required on all anesthesia machines
functioning. better estimate of volume delivered to eliminate the need for vigilance
Selected References
and the potential for failing to
1. Stayer et. al. Comparison of NAD 6000 and recognize a leak or disconnect. With
Servo 900C Ventilators in an Infant Lung
Model. Anesth Analg 2000;90:315-321. a piston ventilator, similar alarms
alert the user to a potential leak.
In vitro study comparing delivery of small
tidal volumes (30 and 100 mls) to a
Furthermore, the reservoir bag will
pediatric lung model by the NAD 6000 be observed to collapse and cause a
Divan ventilator and the Siemens Servo low fresh gas alarm.
900C ICU ventilator using a variety of
settings for respiratory rate, lung
compliance and PEEP. The NAD 6000 9. Do piston ventilators require
which uses the Draeger Divan piston
ventilator was found to be equivalent to the more maintenance than
Siemens 900C ICU ventilator for delivering bellows ventilators?
small tidal volumes.

2. Stayer et.al., Pressure Control Ventilation:


The piston ventilator technology in
Three Anesthesia Ventilators Compared the Fabius GS and Narkomed 6000
Using an Infant Lung Model. Anesth Analg series uses an innovative rolling seal
2000;91:1145-50.
that dramatically reduces the friction
In vitro study comparing the ability to deliver between the piston and the cylinder.
pressure controlled ventilation to an infant
lung model by the Siemens Servo 900C The rolling seal is inexpensive, and is
ICU ventilator, the NAD 6000 Divan replaced as part of the preventative
Ventilator and the Ohmeda Aestiva 3000
ventilator. Ventilator settings were selected maintenance schedule every 2 years.
to mimic near maximum conditions of The motor drive for the piston is a
ventilation for a neonate (30 cmH2O) and
an adult (60 cmH2O). The tidal volumes
brushless system designed to operate
I N S I G H T

delivered by the NAD 6000 and Aestiva for 10 years without maintenance.
ventilators were 5.8 mls and 18.9 mls less Unlike the servo valves in the bellows
than the volume delivered by the 900 C
respectively. The 900 C tended to designs, the piston is not affected by
overshoot the set inspiratory pressure dust or dirt in the compressed gas
during low lung compliance conditions
supplies and is more fault tolerant.

90 49 447 / 08.05-2 / gm-ls-dw / Printed in Germany / Chlorine-free - environmentally compatible / Subject to modifications / © 2005 Dräger Medical AG & Co. KGaA
underscoring the value of an inspiratory flow
setting. The Aestiva did not achieve the set
inspiratory pressure with shorter inspiratory
10. Are all bellows ventilators
M E D I C A L

times whereas the 6000 was not greatly


affected by changes in inspiratory time. subject to the same limitations?
3. Stayer et.al. Volume Ventilation of Infants
with Congenital Heart Disease: There are some differences in
A Comparison of Drager, NAD 6000 and bellows ventilator designs that
Siemens Servo 900C Ventilators. Anesth
Analg 2001;92:76-79. influence the accuracy of volume
delivery. Draeger bellows ventilators
In vivo study of 20 patients undergoing
complex repair of congenital heart defects
fill only to the desired preset tidal
were ventilated in a random fashion with the volume and the user can set
Drager 6000 Divan ventilator and Siemens sufficient inspiratory flow to ensure
900C ICU ventilator. Both machines
provided adequate volume controlled that the bellows empties completely
ventilation. with each breath. Compression of
4. Feldman JM. Smith J. Compliance
drive gas does not influence the
Compensation of the Narkomed 6000 desired tidal volume in that case.
Explained. Anesth. 2001;543-544.

Technical explanation of the manner in


which the Narkomed 6000 Divan ventilator
accomplishes compliance compensation to
deliver the set tidal volume to the patient’s
airway.

USA:

Draeger Medical, Inc. Manufacturer:


3135 Quarry Road Draeger Medical, Inc.
Telford, PA 18969 Telford, PA 18969, USA
USA
Tel: (215) 721-5400 The quality management system at
Toll-free: (800) 437-2437 Draeger Medical, Inc. is certified according
Fax: (215) 723-5935 to ISO 13485, ISO 9001 and Annex II of
E-mail: wwwinfo@draegermed.com Directive 93/42/EEC (Medical devices).

www.draegermedical.com

Вам также может понравиться