(290306-H) Service Report
I3ST No.: 300554835968
xakisg a 8-0-8, Endah Promenade, No. 5, Jalan 3/149E
Taman Sri Endah, Bandar Baru Sri Petaling Date 03.01.18
57000 Kuala Lumpur, Malaysia
Fax: +603 9057 3999
Tel: +603 9057 8999
Time 8:05AM
SG www.tekmarkgroup.com
Contact Person: Mr. Billy Tay H.S Technician: Ammar lmran Bin Shamsudd
55200 KUL MY
Service Details
Calibration with data report.
"/,,/,,I"'
7
0 LABO
MS ISO/IEC 17025
CALIBRATION
SAMM NO. 051
CALIBRATION CERTIFICATE
Customer : LPC Electrical & Plumber,
No 2C, Jalan Perniagaan Ceria 3,
Pusat Perniagaan Ceria,
83000 Batu Pahat, Johor.
Instrument Condition:
Received: Physically in good condition and all measured values were within calibration limits.
Returned: Calibrated, measured values were within calibration limits for the parameters tested.
Comments: Nil
This Certificate is issued in accordance with the conditions of accreditation granted by the SAMM which has assessed the measurement
capability of the laboratory and its traceability to recognised national standards and units of measurement realised at the corresponding
national standards laboratory. Copyright of the this certificate is owned by the issuing laboratory and may not be reproduced other
than in full except with the prior written approval of the Head of the issuing laboratory.
Page 1 of 2
MS !SOIIEC 17025
CALIBRATION
SAMM NO. 051
CALIBRATION DATA
Instrument : Insulation Tester Certificate No TMK48527AL1217
Model No. : 3132A Cal. Date : 26 December 2017
Serial No. W8276493 Data Type : Before Adjustment
AC Voltage Check
Continuity Range
Page 2 of 2