Вы находитесь на странице: 1из 7

154 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, MARCH 1976

[151 P. Bach-y-Rita, "Sensory plasticity," Acta. Neurol. Scandinav, [19] -, "Localization of tonal stimuli in the vertical plane," JASA,
vol. 43, 1967, pp. 417-426. vol. 43, pp. 1260-1266, 1968.
[16] P. Bach-y-Rita, C. C. Collins, L. Scadden, G. W. Holmlund, and [20] G. H. Harris, "Binaural interactions of impulsive stimuli and pure
B. K. Hart, "Display techniques in a tactile vision-substitution tones," JASA, vol. 32, pp. 685-692, 1960.
system," Medical and Biological Illustration, vol. 20, 1970, pp. [21] A. W. Mills, "On the minimum audible angle," JASA, vol. 30,
6-12. pp. 237-246, 1958.
[171 S. K. Roffler, "Sound localization in the vertical plane," unpub- [22] S. S. Stevens, "Neural events and the psychophysical law,"
lished Ph.D. dissertation, University of Chicago, 1967. Science, vol. 170, pp. 1043-1050, 1970.
[18] S. K. Roffler and R. A. Butler, "Factors that influence the local- [23] D. Rowell, "Auditory display of spatial information," unpub-
ization of sound in the vertical plane," JASA, vol. 43, 1968, lished Ph.D. dissertation, University of Canterbury, Christchurch,
pp. 1255-1259. New Zealand, 1970.

Communications.

A Digital QRS Detector Based on the Principle of


Contour Limiting A D
/--
IA
/"."
I

Threshold Th I Th I

HENK G. GOOVAERTS, HENK H. ROS,


THOMAS J. VAN DEN AKKER, AND I,g I/

HANS SCHNEIDER Detector L Det. ,, - v/ Det.


.t- , Output
t b.
I'I t t
Abstract-The article describes a detector for physiological phe- 1 _-
nomena, e.g., the QRS-complex, having a trigger accuracy of 0.5 ms.
The configuration to be recognized first is preprocessed then fed into a (a) (b) (c)
pair of amplifiers with adjustable gain and offset, representing the Fig. 1. Trigger errors due to alterations in the triggering signal. (a)
upper and lower contour-limit derived from the actual signaL These Alterations in steepness. (b) Alterations in amplitude. (c) Alterations
signals are sampled and A/D converted, then stored in two memories due to baseline shift.
during the "read"-operation.
The actual signal, which is continuously sampled and A/D converted, However, detection of slightly different configurations is not
will be fed into a third memory. The contents of this memory will be possible with these methods.
compared with those of the aforementioned memories. Continuous Introducing additional criteria increases the discrimination
refreshing of the third memory shifts the actual signal through the of the method of detection, but at the same time a decreasing
memory, thus obtaining a situation in which the actual signal fits into
number of configurations is detected owing to the presence of
both contour-limits. If so, the comparators, operating between the noise and artifacts in the signal. Besides, the accuracy of the
samples, will generate a trigger signal. derived detection signal is not optimal because concessions
Averaging methods requiring a high accuracy are performed for non- have been made in the set-up of conditions.
invasive investigations on the His-bundle activity. Here the accuracy is Alterations in the steepness of the signal will introduce a
quantitatively determined by a small counter reading, the number of detection error while the same holds good for variations in
samples for which the actual signal fits into both contour-limits. amplitude because the latter may be regarded as changes in
steepness.
INTRODUCTION In Fig. l(a) and (b) the resulting detection error of both
components is shown. Detection errors due to baseline shift,
The on-line detection of a specific configuration of electro- which are shown in Fig. 1(c), can easily be eliminated through
physiological signals is an important aid for the processing of the use of a high-pass filter.
these signals. One can roughly distinguish two methods of In short, several conditions have to be satisfied in order to
which one method deals with the threshold detection of the derive a detection signal from a given signal configuration.
filtered input signal [1], [2] while the other method demands Even so, there are artifacts that will satisfy these conditions
some rigidly defined detection criteria which have to be ful- for a given signal configuration. This brings us to the conclu-
filled by part of the configuration of the input signal [3]. sion that the desired signal configuration itself has to specify
the conditions on which a detection signal will be generated in
Manuscript received January 23, 1975; revised April 28, 1975. order to obtain optimal accuracy.
The authors are with the Department of Medical Physics, Free Uni- Summarizing, all the aforementioned approaches deal with a
versity, Amsterdam, The Netherlands. principle based upon average conditions to which the trigger-

Authorized licensed use limited to: PSG College of Technology. Downloaded on July 26,2010 at 06:33:28 UTC from IEEE Xplore. Restrictions apply.
COMMUNICATIONS 155

ing signal must match. In our design, it is the configuration


itself that is taken as a norm, thus specifying its own detect-
ing condition.

DESCRIPTION OF THE PRINCIPLE


In starting the design the idea was to recognize the actual
signal by means of pattern-recognition. The basic principle
of this system involves the generating of an upper limit and a (a) (b)
lower limit whose forms specify a certain area as shown in
Fig. 2(a). If a given input falls within this area, an indication Fig. 2. The principle of contour limiting. (a) The contour-limits (Cu)
and (Cl) derived from a previous triggering signal are adjusted while a
is given that the required signal has been received. new similar configuration (X) is moving from the right to the left into.
In this figure the actual signal meets the specifications to be the contour-limits. No trigger signal is generated. (b) The triggering
recognized if the following equation holds: signal fits into both contour-limits thus generating a trigger signal.
CU > X > Cl. (1)
In other words, if and only if the configuration of the actual
signal remains between the upper and lower contour-limits, a
recognition will be possible, and this recognition will occur
during the whole time that the actual signal is present in the
area bounded by the two contour-limits. The accuracy of the
time detection will then obviously be dependent on the width
of the bounded area.
METHOD OF OPERATION
Contour-limits (C") and (Cl) are present in a memory con-
taining these limits in n samples. A third memory contains
the last n samples of the actual signal (X). Before the next Fig. 3. Block diagram of the system. The upper and lower contour-
sample of (X) is taken, the contents of these three memories limits are stored in the upper and lower shift registers, respectively.
are compared according to the following equation: The actual signal shift through the middle register. The contents of all
three registers are compared for similarity and D/A converted for dis-
cu i > Xi > cli (2) play purposes.
in which Xi, Cu1 and Cl1 represent the sample number i of the
actual signal and the contour limits (i = 1,.* *, n). a. Input-Preprocessor
If equation (2) is not fulfilled, no detection signal will occur.
Then a new sample from the actual signal (X) is added to the The input preprocessing consists of a wide band amplifier
(X) memory while eliminating the earliest sample in this and a two stage multiple-feedback filter having a pass-band
memory. between 0.5 Hz and 500 Hz. At the output of the filter the
Again comparison takes place between the contents of the signal is distributed on three lines. Two of these lines are fed
three memories. After a certain number of samples the con- into amplifiers which can be connected to the upper and lower
figuration of (X) will fit into both contour-limits, thus gen- contour memories.
erating a detection signal. The accuracy of this detection Through the use of high-resolution potentiometers and a
signal only depends on the sampling frequency of the system. 10-bit A/D-converter, the resolution of the contour-limits
Adjustment of the contour-limits in such a way that retained in memory is ± 10 mV. In Fig. 4 the schematic
(Cu - Cl) increases will allow signals with a smaller signal-to- diagram of the preprocessing unit is given.
noise ratio to be detected. The contour-limits may be gen-
erated artificially or by using samples of the input signal. b. Memory and Comparator
The preprocessed signal now is converted into two contour-
Design Considerations limits. These limits are fed into a pair of 10-bit A/D con-
In order to build such a piece of equipment, certain charac- verters if push button "READ" is operated (Fig. 3). If the
teristics must be kept in mind: "READ" switch is disengaged only the preprocessed signal is
1. The (time-) accuracy of the detection depends upon the available on the input of one A/D converter.
sampling frequency of the system. The A/D converted samples of 10-bits are then stored in a
2. The width of the area (Cu - Cl) determines the range 1024 X 10 bit dynamic shift register. Sampling is done at a
of variability of the input signals which will be detected. rate of 1 or 2 kHz. Sample gates are open for approximately
3. It is not necessary that equation (2) be true for all 20 ,is. The last sample is fed into the (X)-buffer, whereupon
samples (example: normal QRS with slight artifact). the buffers are circulated in parallel through the comparators.
4. In the case of biological signals it is advantageous to be This comparison is performed at a clock frequency of 1.2 MHz
able to use the signal itself in setting the limits. or 2.4 MHz, and during each clock pulse equation (2) is per-
5. A visual display is necessary for the adjustment of limits. formed on the three binary words presented in the three
buffers.
DESCRIPTION OF THE SYSTEM Each time that equation (2) is satisfied a logical 1 will be
The total setup as given in Fig. 3 can be divided into four presented to the counter. The counter therefore contains the
number of times which equation (2) is fulfilled.
sub-systems: In Fig. 5 a simplified diagram of the A/D converter, memory
a. Input-preprocessor. and comparator section is given.
b. A/D-converter, memories and comparator. After completing the cycle, the counter is reset. In the
c. Output-display. instrument a small counter is incorporated which counts the
d. Clock. number of successive samples for which the equation holds.

Authorized licensed use limited to: PSG College of Technology. Downloaded on July 26,2010 at 06:33:28 UTC from IEEE Xplore. Restrictions apply.
156 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, MARCH 1976

Input

Fig. 4. Simplified schematic diagram of the preprocessing unit. AU switch functions are performed by means of digitally
controlled FET switches.

Trigger

Clock 2

Fig. 5. Simplified diagram of the A/D converter, memory and comparator section. Clock 1 represents the high frequent
clock of 1.2-2.4 MHz which recirculates the shift registers. Clock 2 is the low frequent clock of 1-2 kHz used for sam-
pling of the actual signal.
Authorized licensed use limited to: PSG College of Technology. Downloaded on July 26,2010 at 06:33:28 UTC from IEEE Xplore. Restrictions apply.
| rJ~ rCbock2
COMMUNICATIONS 157

From Ll __
n
Cu men or

X nmeowry

C~~~1
memor

00 100

Fig. 6. Simplified diagram of the display section. The Z-axis part of this section is designed to be used with a Tektronix
7904 oscilloscope. Using different oscilloscopes, the Z-output may cause insufficient blanking and unblanking. Supply-
ing both 2N3904's from +15 V and -15 V may correct this.

c. Output-Display Unit At the same time a counter stores the number of words
In order to read the desired signal and for the case of adjust- shifted. When the word number 1024 is reached, the clock
is inhibited. In the meantime a new sample is taken during the
ing the contour-limits, it is necessary to visualize the contents cycling procedure and is presented at the outputs of the A/D
of all three buffers. This can easily be done by just adding converter. Then the read pulse is derived which brings us to
three D/A converters to the system, because for comparing the beginning of the main cycle. In Fig. 7(a) the timing dia-
all three buffers the necessity for circulating exists. After D/A gram is shown.
conversion one has a continuous signal that is repeated with If a read pulse is presented to the buffers at the same time,
each cycle. Because of the relative high sampling frequency the inputs to the buffers must be enabled. Use is made of dy-
it is possible to alternate the analog converted contents of all
three buffers. namic shift registers requiring a two-phase clock-pulse ob-
An appropriate time base is derived from the clock pulses tained by means of a flip-flop.
used for cycling of the buffers. This clock is fed into a 10-bit In order to reduce the power dissipation of the shift-registers
counter which is followed by a D/A converter. The output of in the logical-0 state of the clock-pulse, the duty-cycle is re-
this D/A converter presents a ramp voltage during cycling of duced by means of a R.C. network. This configuration is
the buffers facilitating in this way the display of the buffer shown in Fig. 7(b).
contents. In Fig. 6 the simplified diagram of the display sec- Horizontal shifting of the contour-limits is achieved by using
tion is given. Furthermore this section provides blanking and the read clock pulse without enabling the input gates. This
unblanking of the displayed signal. shifting, which is only possible in the time-scaling mode, ob-
It is possible to shift both contour-limits to the right side of structs a proper positioning of these limits due to the fast mov-
the screen in such a way that the end of the configuration to
be recognized ends the display. In this way a minimum delay ing display. Improvement of this shifting speed is easily ob-
is achieved. It should be noted that the display may be cali- tained by omitting a certain number of read pulses by means
brated in real-time figures. So it will be possible to adjust of gating.
precisely the delay in milliseconds while looking at the time-
scaling display presented at a clock rate of 1 or 2 kHz.
APPLICATIONS, RESULTS
d. Clock
Timing is achieved through a very simple cyclic approach 1. Fig. 8 shows the results which are obtained with an input
consisting of one pass for sampling and the next for cycling signal having a large noise artifact. Line B shows the trigger
the buffers. From the end of the sampling pass a read pulse signal resulting from the contour-limit detection while line C
feeds this sample into the buffer. Then memory cycling starts shows the output of a conventional QRS detector. It can
at a frequency 1200 times higher than the sampling frequency. clearly be seen that no trigger signal is gathered in response to
Authorized licensed use limited to: PSG College of Technology. Downloaded on July 26,2010 at 06:33:28 UTC from IEEE Xplore. Restrictions apply.
158 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, MARCH 1976

Clock 2

Read
enable

Convert
command

1/2 SN 7473

(b)

Fig. 7. (a) Tining diagram of the connection between clock 1 and clock 2 and the derived sampling pulses. Shifting one
contour-limit with respect to the other is obtained by omitting the read enable pulse (lower trace of the extracted area).
(b) Schematic diagram of the duty-cycle reducing circuit applied to clock 1.

Fig. 8. Strip chart recording presenting the actual signal (trace A) and the derived trigger signal (trace B). For comparison,
the output of a conventional detector is recorded (trace C). It should be noted that due to the cycle time of the detector
a certain time delay between the occurrence of the triggering signal and the derived trigger pulse is generated.

the artifacts. This can be of advantage in cases where false similarity of the triggering signal with the adjusted contour-
positive triggers must be avoided. limits, it is possible that an artifact will inhibit the appropriate
2. Fig. 9 shows detection of extrasystoles having a specific trigger though the configuration without artifact would meet
configuration. Since it is well known that the configuration of the specifications presented in these contour-limits. Therefore
the extrasystole as seen on the ECG is dependent upon the the design provides the possibility of generating a trigger signal
point of origin of the extrasystole, such an instrument could at different similarities of the compared samples. In our case,
be used to monitor patients suspected of having ectopic foci. 100%=n, 75% 3 n, 50%= 2 n and 25%= 1 n.
It can clearly be seen that an artifact (R), similar to the The trigger accuracy will depend on the time elapsed be-
extrasystole but not having the correct configuration, will be tween entrance of the configuration in the bounded area and
rejected. departure from this area. Shifting one contour-limit with
In Fig. 10 a long term recording is shown with the detector respect to the other will narrow this area. The trigger accuracy
adjusted to extrasystoles. will then be improved because the configuration will remain in
the bounded area during a shorter period of time. This will
DISCUSSION allow the configuration to vary in amplitude within both
A point of discussion is the necessity for all samples to contour-limits without considerable reduction in accuracy.
satisfy equation (2).
Authorized For,
licensed use if theto:trigger
limited is generated
PSG College of Technology. 100% on July
atDownloaded the same
In 26,2010 at 06:33:28 it isfrom
way UTC possible to adjust
IEEE Xplore. theapply.
Restrictions baseline of both
COMMUNICATIONS 159

Fig. 9. Strip chart recording showing trigger signals (trace C) derived from the instrument with contour-limits adjusted to
an extrasystole. It should be noted that an artifact similar to an extrasystole (R in trace A) is rejected. Trace B shows
the filtered signal, while trace D shows the triggers obtained from a conventional QRS detector.

Fig. 10. Long term recording showing triggers generated with the instrument adjusted to extrasystoles (trace D). Trace A
shows the actual signal, trace B the filtered signal and trace C shows triggers generated with a conventional detector.

contour-limits in such a manner that the area in which the more than one configuration with individual limit adjustment.
baseline of the configuration has to remain will be increased. Thus a trigger output can be produced with the occurrence of
In this case a certain amount of baseline shift will be allowed. any one of several stored configurations.
It will be clear that any possible combination of amplification
factor, baseline adjustment and contour shifting can be CONCLUSIONS
achieved. It should be clearly pointed out that depending on the pur-
Reducing the number of samples raises the possibility of a pose for which the instrument will be used, adjustment of
higher sampling frequency, facilitating the detection of con- contour-limits has to be made for maximum detectability or
figurations of shorter duration. Detection of such phenomena maximum accuracy of the generated trigger signals.
would require a higher trigger accuracy, which is automatically maximum detectability As mentioned before, maximum accuracy contraindicates
due to the presence of noise in the sig-
available due to the increased sampling rate. After each cycle nal. Configurations with slight variations in similarity can be
the comparator/counter contains the number of samples that detected, but due to the allowance for these variations, with a
satisfied equation (2). If the system is not operated at 100% decreased accuracy. An interesting application will be the
agreement it is possible to apply some statistical analysis to the field of signal averaging.
contents of the counter in order to detect possible repetitive a. Averaging methods applied to the ECG in order to derive
dropouts. These dropouts could be the result of periodic noise a representative configuration from the ECG taken under
mixed with the input signal. stress will be improved by a higher trigger accuracy.
A slight modification of the instrument enables one to store b. Due to the accuracy it will be possible to use averaging
Authorized licensed use limited to: PSG College of Technology. Downloaded on July 26,2010 at 06:33:28 UTC from IEEE Xplore. Restrictions apply.
160 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, MARCH 1976

methods in order to obtain high frequent components present and about 12 percent of the population report pain on jaw
in the QRS complex. For instance averaging of "slurs and opening. This syndrome appears to be a problem of some
notches" [41, [5]. complexity, and therefore a quantitative diagnosis would be
c. Also averaging of the complete ECG will be possible. valuable in assessing symptom severity, treatment progress,
This enables non-invasive investigations on the conduction and competing treatment plans.
system of the heart to be carried out [61. Our laboratory has taken a multi-faceted approach to the
problem of quantifying the diagnosis of TMJ problems [ 12] -
ACKNOWLEDGMENT [15]. The goal is to improve the curative and preventive
The authors are very grateful to Mr. J. G. Meijer of the De- techniques of the dentist by providing quantitative measure-
partment of Physiology for the photographic assistance in ments for diagnosis.
the preparation of the illustrations. They also wish to thank One quantitative measurement technique reported to have
Mrs. W. M. U. van Barschot for the accurate typing of the diagnostic significance is the duration of the electromyographic
manuscript. (EMG) silent period following the jaw-jerk or myotatic reflex.
Bessette, Bishop, and Mohl [ 16] have reported that the silent
REFERENCES period duration is greater than 32 ms in diagnosed TMJ
[1] W. P. Holsinger, K. M. Kempner and M. H. Miller. A QRS prepro- patients and less than 28 ms in asymptomatic patients.
cessor based on digital differentiation. IEEE Trans. Biomed. Eng., Further, they report that successful treatment results in a
vol. BME-18, nr. 3, pp. 212-217, May 1971. decrease of the silent period duration to within the normal
[2] A. R. Hambley, R. L. Moruzzi and C. L. Feldman. The use of range [16]. Studies from our laboratory have confirmed and
intrinsic components in an ECG filter. IEEE Trans. Bionm Eng., extended these results [14], [15].
vol. BME-21, nr. 6, pp. 469-473, November 1974. Previous techniques for measuring the duration of the silent
[3] A. Sandman, D. W. Hill and A. H. Wilcock. Analogue preproces- period whether using a mechanical device [ 16], optical scanner
sor for the measurement by a digital computer of R-R intervals [17], or photographic projection [18] all involve slow, post-
and R wave widths. Med. Biol. Eng., vol. 11, nr. 2, pp. 191- experiment hand measurements with the attendant possibilities
200, March 1973. for bias and error. The importance of the phenomena, the
[4] N. C. Flowers and L. G. Horan. Diagnostic impact of QRS notch-
ing in high-frequency electrocardiograms of living subjects with number of measurements required, and the possibility of
heart disease. Circulation, vol. XLIV, pp. 605-611, October 1971. human error all suggest that a rapid, automated technique for
[5] P. H. Langner, D. B. Geselowitz and S. A. Briller. Wide band measuring silent period duration would significantly enhance
recording of the electrocardiogram and coronary heart disease. research progress. The purpose here is to describe a device that
An. Heart Journ., vol. 86, nr. 3, pp. 308-317, September 1973. implements such a technique. A preliminary report has been
[6] E. J. Berbari, R. Lazarra, P. Samet and B. J. Scherlag. Noninvasive presented [19].
technique for detection of electrical activity during the P-R seg-
ment. Circulation, vol. XLVIII, pp. 1005-1013, November 1973. DESIGN CONSIDERATIONS
Input Signal Constraints: While physiologically the silent
period is comprised of a complex interaction of disfacilitation
and inhibition at the motoneuron, operationally a simple
definition must be used. Experience suggested that a defini-
tion including both an amplitude criterion and duration
criterion would be satisfactory. An amplitude window is
An Interactive Computer Peripheral to Measure the necessary because noise in the recording system precludes an
Electromyographic Silent Period absolutely "silent" silent period. A time window, constraining
the duration of acceptable silent periods to within reference
S. S. ALTUG, C. T. CHILDRESS, JR., S. M. DEMUND, minimum and maximum durations, is necessary to exclude
W. D. McCALL, JR., MEMBER, IEEE, AND M. M. ASH, JR. zero crossings and other short periods of low activity on the
minimum side and rest and other periods of low activity on the
maximum side.
Abstract-A device to provide automated measurement of the electro- Even with the two windows, some parts of EMG traces were
myographic silent period with real time capability uflizing a voltage- expected to fit the operational definition of the silent period
time window is described. This device is being used in the clinical and cause false alarms, so a monitoring technique would be
dental diagnosis and treatment of functional disorders of the masti- necessary. This would retain the clinical judgement of the
catory system. The design, performance, utilization, and other applica- clinician as to what constituted a silent period while relieving
tions of this device are discussed. him of the measurement chore. The intent is not to replace
the dentist but to relieve him of non-clinically oriented tasks.
Another feature of the EMG signal is variability of signal and
INTRODUCTION noise amplitude from patient to patient. Hence the ability to
Controversy abounds in clinical dentistry over the no- control the windows was considered necessary. Variations on
menclature, etiology [1]-[4], and treatment [5]-[9] of pain the order of milliseconds were expected from the nature of the
associated with the muscles of mastication and the temporo- waveform as the EMG signal entered and left the silent period.
mandibular joint (TMJ). DeBoever [10] has reviewed these Therefore device accuracy on the order of ± 1 percent was
controversies while Agerberg and Carlsson [ 11 ] have reported considered adequate. This would amount to ± 80 Mus for the
epidemiological evidence that up to 57 percent of the adult shortest silent period expected. The EMG signal has a
population report one or more symptoms of TMJ dysfunction spectrum that extends to a few hundred Hz but most of the
power is concentrated between 30 and 100 Hz [201, [211.
Manuscript received October 22, 1974; revised April 25, 1975. This Hence commercially available devices would give adequate
work was supported by the USPHS under Grant DE 02731. frequency response.
The authors are with the Bioengineering and Occlusion aboratory, Laboratory Environment: This design must interface to
Department of Occlusion, Dental Research Institute, School of Den- other items of laboratory equipment. The flow of information
tistry, University of Michigan, Ann Arbor, MI 48104. is shown in Fig. 1. The EMG amplifiers and polygraph were

Authorized licensed use limited to: PSG College of Technology. Downloaded on July 26,2010 at 06:33:28 UTC from IEEE Xplore. Restrictions apply.

Вам также может понравиться