Вы находитесь на странице: 1из 4

Pharmacopeial Forum

Vol. 33(3) [May–June 2007] 1

Class 1 System Suitability Solution—Transfer 1.0 mL of Class 1


Standard Stock Solution to an appropriate headspace vial, add 5.0 mL
of Test Stock Solution, apply the stopper, cap, and mix.
Chromatographic System (see Chromatography h621i)—The gas
chromatograph is equipped with a flame-ionization detector, a
h467i RESIDUAL SOLVENTS 0.32-mm 6 30-m fused-silica column coated with a 1.8-mm layer
of phase G43 or a 0.53-mm 6 30-m wide-bore column coated with
a 3.0-mm layer of phase G43. The carrier gas is nitrogen or helium
(Chapter under this new title—to become official July 1, 2007) with a linear velocity of about 35 cm per second, and a split ratio
(Current chapter title is h467i Organic Volatile Impurities) of 1 : 5. .[NOTE—Split ratio can be modified in order to optimize sen-
sitivity.].3 The column temperature is maintained at 408 for 20 min-
utes, then raised at a rate of 108 per minute to 2408, and maintained at
Change to read: 2408 for 20 minutes. The injection port and detector temperatures are
maintained at 1408 and 2508, respectively. Chromatograph the Class
IDENTIFICATION, CONTROL, AND 1 Standard Solution, Class 1 System Suitability Solution, and Class
2 Mixture A Standard Solution, and record the peak responses as di-
QUANTIFICATION OF RESIDUAL SOLVENTS rected for Procedure: the signal-to-noise ratio of 1,1,1-trichloroeth-
. ane in the Class 1 Standard Solution is not less than 5; the signal-
Whenever possible, the substance under test needs to be dissolved to-noise ratio of each peak in the Class 1 System Suitability Solution
to release the residual solvent. Because the USP deals with drug prod- is not less than 3; and the resolution, R, between acetonitrile and
ucts, as well as active ingredients and excipients, it may be acceptable methylene chloride in the Class 2 Mixture A Standard Solution is
that in cases some of the components of the formulation will not dis- not less than 1.0.
solve completely. In those cases, the drug product may first need to be
pulverized into a fine powder so that any residual solvent that may be Procedure—Separately inject (following one of the headspace op-
present can be released. This operation should be as fast as possible to erating parameter sets described in the table below) equal volumes of
prevent the loss of volatile solvents during the procedure..3 headspace (about 1.0 mL) of the Class 1 Standard Solution, Class
Page 1 of 20 — Time:09:27 — Date:3/23/07 Instance: g:/Pf/Production/Final/m99226h.xml Template:s:/Pf/Template/pf_server_v8_pagesizes-v1.3f

2 Mixture A Standard Solution, Class 2 Mixture B Standard Solution,


NOTE—The organic-free water specified in the following proce- and the Test Solution into the chromatograph, record the chromato-
dures produces no significantly interfering peaks when chromato- grams, and measure the responses for the major peaks. .If a peak re-
graphed. sponse of any peak, other than a peak for 1,1,1-trichloroethane, in the
Test Solution is greater than or equal to a corresponding peak in either
the Class 1 Standard Solution or either of the two Class 2 Mixture
Standard Solutions, or a peak response of 1,1,1-trichloroethane is
Class 1 and Class 2 Residual Solvents greater than or equal to 150 times the peak response corresponding
.The following procedures are useful to identify and quantify re- to 1,1,1-trichloroethane in the Class 1 Standard Solution, proceed to
sidual solvents when the information regarding which solvents are Procedure B to verify the identity of the peak; otherwise the article
likely to be present in the material is not available. When the infor- meets the requirements of this test..3
mation about the presence of specific residual solvents is available,
only Procedure C is needed to quantify the amount of residual sol- Table 5. Headspace Operating Parameters
vents present..3
Headspace Operating
Parameter Sets
WATER-SOLUBLE ARTICLES 1 2 3
Equilibration temperature (8) 80 105 80
Procedure A— Equilibration time (min.) 60 45 45
Class 1 Standard Stock Solution—Transfer 1.0 mL of USP Class 1 Transfer-line temperature (8) 85 110 105
Residual Solvents Mixture RS to a 100-mL volumetric flask, add 9 Carrier gas: nitrogen or helium at an appropriate pressure
mL of dimethyl sulfoxide, dilute with water to volume, and mix. Pressurization time (s) 30 30 30
Transfer 1.0 mL of this solution to a 100-mL volumetric flask, dilute Injection volume (mL) 1 1 1
with water to volume, and mix. Transfer 1.0 mL of this solution to a
10-mL volumetric flask, dilute with water to volume, and mix.
Class 1 Standard Solution—Transfer 1.0 mL of Class 1 Standard Procedure B—
Stock Solution to an appropriate headspace vial, add 5.0 mL of water, Class 1 Standard Stock Solution, Class 1 Standard Solution, Class
apply the stopper, cap, and mix. 2 Standard Stock Solutions, Class 2 Mixture A Standard Solution,
Class 2 Standard Stock Solutions—Transfer 1.0 mL of USP Resid- Class 2 Mixture B Standard Solution, Test Stock Solution, Test Solu-
ual Solvents Class 2—Mixture A RS to a 100-mL volumetric flask, tion, and Class 1 System Suitability Solution—Prepare as directed for
dilute with water to volume, and mix. This is Class 2 Standard Stock Procedure A...3
Solution A. Transfer 1.0 mL of USP Residual Solvents Class 2—Mix- Chromatographic System (see Chromatography h621i)—The gas
ture B RS to a 100-mL volumetric flask, dilute with water to volume, chromatograph is equipped with a flame-ionization detector, a
and mix. This is Class 2 Standard Stock Solution B. 0.32-mm 6 30-m fused-silica column coated with a 0.25-mm layer
Class 2 Mixture A Standard Solution—Transfer 1.0 mL of Class 2 of phase G16, or a 0.53-mm 6 30-m wide-bore column coated with a
Standard Stock Solution A to an appropriate headspace vial, add 5.0 0.25-mm layer of phase G16. The carrier gas is nitrogen or helium
mL of water, apply the stopper, cap, and mix. with a linear velocity of about 35 cm per second and a split ratio of
1 : 5. .[NOTE—Split ratio can be modified in order to optimize sensi-
Class 2 Mixture B Standard Solution—Transfer 5.0 mL of Class 2 tivity.].3 The column temperature is maintained at 508 for 20 minutes,
Standard Stock Solution B to an appropriate headspace vial, add 1.0 then raised at a rate of 68 per minute to 1658, and maintained at 1658
mL of water, apply the stopper, cap, and mix. for 20 minutes. The injection port and detector temperatures are main-
Test Stock Solution—Transfer about 250 mg of the article under tained at 1408 and 2508, respectively. Chromatograph the Class 1
test, accurately weighed, to a 25-mL volumetric flask, dissolve in Standard Solution and the Class 1 System Suitability Solution, ..3
and dilute with water to volume, and mix. and record the peak responses as directed for Procedure: the sig-
Test Solution—Transfer 5.0 mL of Test Stock Solution to an appro- nal-to-noise ratio of benzene in the Class 1 Standard Solution is
priate headspace vial, add 1.0 mL of water, apply the stopper, cap, not less than 5; the signal-to-noise ratio of each peak in the Class 1
and mix. System Suitability Solution is not less than 3; and the resolution, R,
between acetonitrile and .cis-dichloroethene in the Class 2 Mixture A
Standard Solution.3 is not less than 1.0.

# 2007 The United States Pharmacopeial Convention, Inc. All Rights Reserved.
Pharmacopeial Forum
2 Vol. 33(3) [May–June 2007]

Procedure—Separately inject (following one of the headspace op- WATER-INSOLUBLE ARTICLES


erating parameter sets described in Table 5) equal volumes of head-
space (about 1.0 mL) of the Class 1 Standard Solution, the Class .Procedure A—[NOTE—Dimethy sulfoxide may be substituted as
2 Mixture A Standard Solution, the Class 2 Mixture B Standard Solu- an alternative solvent to dimethylformamide.]
tion, and the Test Solution into the chromatograph, record the chro- Class 1 Standard Stock Solution—Transfer 1.0 mL of USP Class 1
matograms, and measure the responses for the major peaks. If the Residual Solvents Mixture RS to a 100-mL volumetric flask previ-
peak response(s) in the Test Solution of the peak(s) identified in Pro- ously filled with about 80 mL of dimethylformamide, dilute with di-
cedure A is/are greater than or equal to a corresponding peak(s) in methylformamide to volume, and mix. Transfer 1.0 mL of this
either the Class 1 Standard Solution or either of the two Class 2 Mix- solution to a 100-mL volumetric flask, previously filled with about
ture Standard Solutions, proceed to Procedure C to quantify the 80 mL of dimethylformamide, dilute with dimethylformamide to vol-
peak(s); otherwise the article meets the requirements of this test. ume, and mix. [NOTE—Reserve a portion of this solution for the Class
Procedure C— 1 System Suitability Solution.] Transfer 1.0 mL of this solution to a
Class 1 Standard Stock Solution, Class 1 Standard Solution, Class 10-mL volumetric flask, dilute with dimethylformamide to volume,
2 Standard Stock Solution A, Class 2 Mixture A Standard Solution, and mix.
Test Stock Solution, Test Solution, and Class 1 System Suitability So- Class 1 Standard Solution—Transfer 1.0 mL of Class 1 Standard
lution—Prepare as directed for Procedure A. Stock Solution to an appropriate headspace vial, containing 5.0 mL of
Standard Solution—[NOTE—.Prepare a separate Standard Solution water, apply the stopper, cap, and mix.
for each peak identified and verified by Procedures A and B. For the Class 2 Standard Stock Solutions—Transfer 1.0 mL of USP Resid-
Class 1 solvents other than 1,1,1-trichloroethane, prepare the first di- ual Solvents Class 2—Mixture A RS to a 100-mL volumetric flask,
lution as directed for the first dilution under Class 1 Standard Stock previously filled with about 80 mL of dimethylformamide, dilute with
Solution in Procedure A..3] Transfer an accurately measured volume dimethylformamide to volume, and mix. This is Class 2 Standard
of each individual USP Reference Standard corresponding to each Stock Solution A. Transfer 0.5 mL of USP Residual Solvents Class
residual solvent peak identified and verified by Procedures A and B 2—Mixture B RS to a 10-mL volumetric flask, dilute with dimethyl-
to a suitable container, and dilute quantitatively, and stepwise if nec- formamide to volume, and mix. This is Class 2 Standard Stock Solu-
essary, with water to obtain a solution having a final concentration of tion B.
Page 2 of 20 — Time:09:27 — Date:3/23/07 Instance: g:/Pf/Production/Final/m99226h.xml Template:s:/Pf/Template/pf_server_v8_pagesizes-v1.3f

1/20 of the value stated in Table 1 or 2 (under Concentration Limit). Class 2 Mixture A Standard Solution—Transfer 1.0 mL of Class 2
Transfer 1.0 mL of this solution to an appropriate headspace vial, add Standard Stock Solution A to an appropriate headspace vial, contain-
5.0 mL of water, apply the stopper, cap, and mix. ing 5.0 mL of water, apply the stopper, cap, and mix.
Spiked Test Solution—[NOTE—Prepare a separate Spiked Test Solu- Class 2 Mixture B Standard Solution—Transfer 1.0 mL of Class 2
tion for each peak identified and verified by Procedures A and B.] Standard Stock Solution B to an appropriate headspace vial, contain-
Transfer 5.0 mL of Test Stock Solution to an appropriate headspace ing 5.0 mL of water, apply the stopper, cap, and mix.
vial, add 1.0 mL of the Standard Solution, apply the stopper, cap, and Test Stock Solution—Transfer about 500 mg of the article under
mix. test, accurately weighed, to a 10-mL volumetric flask, dissolve in
Chromatographic System (see Chromatography h621i)—[NOTE— and dilute with dimethylformamide to volume, and mix.
If the results of the chromatography from Procedure A are found to be Test Solution—Transfer 1.0 mL of the Test Stock Solution to an ap-
inferior to those found with Procedure B, the Chromatographic Sys- propriate headspace vial, containing 5.0 mL of water, apply the stop-
tem from Procedure B may be substituted.] The gas chromatograph is per, cap, and mix.
equipped with a flame-ionization detector, a 0.32-mm 6 30-m fused-
silica column coated with a 1.8-mm layer of phase G43 or a 0.53-mm Class 1 System Suitability Solution—Mix 5 mL of the Test Stock
6 30-m wide-bore column coated with a 3.0-mm layer of phase G43. Solution with 0.5 mL of the intermediate dilution reserved from Class
The carrier gas is nitrogen or helium with a linear velocity of about 35 1 Standard Stock Solution. Transfer 1.0 mL of this solution to an ap-
cm per second, and a split ratio of 1 : 5. .[NOTE—Split ratio can be propriate headspace vial, containing 5.0 mL of water, apply the stop-
modified in order to optimize sensitivity.].3 The column temperature per, cap, and mix.
is maintained at 408 for 20 minutes, then raised at a rate of 108 per Chromatographic System (see Chromatography h621i—The gas
minute to 2408, and maintained at 2408 for 20 minutes. The injection chromatograph is equipped with a flame-ionization detector, a
port and detector temperatures are maintained at 1408 and 2508, 0.53-mm 6 30-m wide-bore column coated with a 3.0-mm layer of
respectively. Chromatograph the Class 1 Standard Solution, the Class phase G43. The carrier gas is helium with a linear velocity of about
1 System Suitability Solution, and the Class 2 Mixture A Standard So- 35 cm per second, and a split ratio of 1 : 3. [NOTE—Split ratio can be
lution, and record the peak responses as directed for Procedure: the modified in order to optimize sensitivity.] The column temperature is
signal-to-noise ratio of 1,1,1-trichloroethane in the Class 1 Standard maintained at 408 for 20 minutes, then raised at a rate of 108 per mi-
Solution is not less than 5; the signal-to-noise ratio of each peak in the nute to 2408, and maintained at 2408 for 20 minutes. The injection
Class 1 System Suitability Solution is not less than 3; and the resolu- port and detector temperatures are maintained at 1408 and 2508,
tion, R, between acetonitrile and methylene chloride in the Class respectively. Chromatograph the Class 1 Standard Solution, Class
2 Mixture A Standard Solution is not less than 1.0. 1 System Suitability Solution, and Class 2 Mixture A Standard Solu-
Procedure—Separately inject (following one of the headspace op- tion, and record the peak responses as directed for Procedure: the sig-
erating parameters described in Table 5) equal volumes of headspace nal-to-noise ratio of 1,1,1-trichloroethane in the Class 1 Standard
(about 1.0 mL) of the Standard Solution, the Test Solution, and the Solution is not less than 5; the signal-to-noise ratio of each peak in
Spiked Test Solution into the chromatograph, record the chromato- the Class 1 System Suitability Solution is not less than 3; and the re-
grams, and measure the responses for the major peaks. Calculate solution, R, between acetonitrile and methylene chloride in the Class
the amount, in ppm, of each residual solvent found in the article under 2 Mixture A Standard Solution is not less than 1.0.
test by the formula: Procedure—Separately inject (use headspace operating parameters
3 in Table 5 with a vial pressure of 10 psi) equal volumes of head-
5(C/W)[rU /(rST – rU)] space (about 1.0 mL) of the Class 1 Standard Solution, Class 2 Mix-
in which C is the concentration, in .mg per mL,.3 of the appropriate ture A Standard Solution, Class 2 Mixture B Standard Solution, and
USP Reference Standard in the Standard Solution; W is the weight, in the Test Solution into the chromatograph, record the chromatograms,
g, of the article under test taken to prepare the Test Stock Solution; and and measure the responses for the major peaks. If a peak response of
rU and rST are the peak responses of each residual solvent obtained any peak, other than a peak for 1,1,1-trichloroethane, in the Test So-
from the Test Solution and the Spiked Test Solution, respectively. lution is greater than or equal to a corresponding peak in either the
Class 1 Standard Solution or either of the two Class 2 Mixture Stan-
dard Solutions, or a peak response of 1,1,1-trichloroethane is greater
than or equal to 150 times the peak response corresponding to 1,1,1-
trichloroethane in the Class 1 Standard Solution, proceed to Proce-
dure B to verify the identity of the peak; otherwise the article meets
the requirements of this test.

# 2007 The United States Pharmacopeial Convention, Inc. All Rights Reserved.
Pharmacopeial Forum
Vol. 33(3) [May–June 2007] 3

Procedure B— Spiked Test Solution—[NOTE—Prepare a separate Spiked Test Solu-


Class 1 Standard Stock Solution, Class 1 Standard Solution, Class tion for each peak identified and verified by Procedures A and B.]
1 System Suitability Solution, Class 2 Standard Stock Solutions, Class Transfer 1.0 mL of the Test Stock Solution to an appropriate head-
2 Mixture A Standard Solution, and Class 2 Mixture B Standard So- space vial, add 1 mL of the Standard Stock Solution and 4.0 mL of
lution, Test Stock Solution, and Test Solution—Proceed as directed for water, apply the stopper, cap, and mix.
Procedure A. Chromatographic System—Proceed as directed for Procedure C
Chromatographic System—Proceed as directed for Procedure B under Water-Soluble Articles.
under Water-Soluble Articles with a split ratio of 1 : 3. [NOTE—Split Procedure—Separately inject (use headspace operating parameters
ratio can be modified in order to optimize sensitivity.] 3 in Table 5 with a vial pressure of 10 psi) equal volumes of head-
Procedure—Separately inject (use headspsce operating parameters space (about 1.0 mL) of the Standard Solution, the Test Solution, and
3 in Table 5 with a vial pressure of 10 psi) equal volumes of head- the Spiked Test Solution into the chromatograph, record the chromat-
space (about 1.0 mL) of the Class 1 Standard Solution, Class 2 Mix- ograms, and measure the responses for the major peaks. Calculate the
ture A Standard Solution, Class 2 Mixture B Standard Solution, and amount, in ppm, of each residual solvent found in the article under
the Test Solution, into the chromatograph, record the chromatograms, test by the formula:
and measure the responses for the major peaks. If the peak respon- 10 (C / W)[rU / (rST – rU)]
se(s) in the Test Solution of the peak(s) identified in Procedure A
is/are greater than or equal to a corresponding peak(s) in either the in which C is the concentration, in mg per mL, of the appropriate USP
Class 1 Standard Solution or any of the two Class 2 Mixture Standard Reference Standard in the Standard Solution; W is the weight, in g, of
Solutions, proceed to Procedure C to quantify the peak(s); otherwise the article under test taken to prepare the Test Stock Solution; and rU
the article meets the requirements of this test. and rST are the peak responses of each residual solvent obtained from
Procedure C— the Test Solution and the Spiked Test Solution, respectively..3
Class 1 Standard Stock Solution, Class 1 Standard Solution, Class
1 System Suitability Solution, Class 2 Standard Stock Solution A, and
Class 2 Mixture A Standard Solution—Proceed as directed for Proce- Class 3 Residual Solvents
Page 3 of 20 — Time:09:27 — Date:3/23/07 Instance: g:/Pf/Production/Final/m99226h.xml Template:s:/Pf/Template/pf_server_v8_pagesizes-v1.3f

dure A. .
Standard Stock Solution—[NOTE—Prepare a separate Standard So- If Class 3 solvents are present, the level of residual solvents may
lution for each peak identified and verified by Procedures A and B.] be determined as directed under Loss on Drying h731i when the
Transfer an accurately measured volume of each individual USP Ref- monograph for the article under test contains a loss on drying proce-
erence Standard corresponding to each residual solvent peak identi- dure, or a specific determination of the solvent may be made. If there
fied and verified by Procedures A and B to a suitable container, and is no loss on drying procedure in the monograph for the article under
dilute quantitatively, and stepwise if necessary, with water to obtain a test or if.3 a Class 3 solvent limit in an individual monograph is great-
solution having a final concentration of 1/20 of the value stated in er than 50 mg per day (corresponding to 5000 ppm or 0.5% under
Table 1 or Table 2 (under Concentration Limit). Option 1), .the individual Class 3 residual solvent or solvents present
in the article under test.3 should be identified and quantified, and the
Standard Solution—Transfer 1.0 mL of the Standard Stock Solu- procedures as described above, with appropriate modifications to the
tion to an appropriate headspace vial, containing 5.0 mL of water, standard solutions, are to be applied wherever possible. Otherwise an
apply the stopper, cap, and mix. appropriate validated procedure is to be employed. ..3 USP Refer-
Test Stock Solution—Proceed as directed for Procedure A. ence Standards, where available, should be used in these procedures.
Test Solution—Transfer 1.0 mL of the Test Stock Solution to an ap- A flow diagram for the application of residual solvent limit tests is
propriate headspace vial, containing 5.0 mL of water, apply the stop- shown in Figure 1.
per, cap, and mix.

# 2007 The United States Pharmacopeial Convention, Inc. All Rights Reserved.
Pharmacopeial Forum
4 Vol. 33(3) [May–June 2007]
Page 4 of 20 — Time:09:27 — Date:3/23/07 Instance: g:/Pf/Production/Final/m99226h.xml Template:s:/Pf/Template/pf_server_v8_pagesizes-v1.3f

Figure 1. Diagram relating to the identification of residual solvents and the application of limit tests.

# 2007 The United States Pharmacopeial Convention, Inc. All Rights Reserved.

Вам также может понравиться