Вы находитесь на странице: 1из 8

ISO/IEC 17025:2017 Transition Webinar Questions & Answers

Contents
1. Complaints .......................................................................................................................... 1
2. Decision Rules .................................................................................................................... 2
3. General ............................................................................................................................... 3
4. MS Options ......................................................................................................................... 4
5. Risk..................................................................................................................................... 5
6. Sampling ............................................................................................................................. 6
7. Traceability ......................................................................................................................... 7
8. Transition ............................................................................................................................ 8

1. Complaints
Question Answer
Is the Complaints process expected to be A description of the process, yes. This may
available to any interested party on be a brief account rather than a copy of the
request? process.
7.9.2 If a description of the complaint We assess the lab’s compliance with 17025,
handling process has to be available to any not explicitly their compliance with Law. See
interested party. In some UK countries (e.g. other related question on this topic.
Wales) this would have to be bilingual for
public body laboratories. Would we rely on
the lab’s translation or would we have to
use an external translation resource?
Is it in your view acceptable for small labs to The requirement is that it is reviewed before
have the 2nd person for complaint handling giving to client, so probably generally not.
to come in annually to look over de Would need to be accessible on demand.
complaints as part of the internal auditing
process?
7.9.2 says, A description of the handling Yes, a description of the process is easier
process for complaints shall be available to than a copy of the procedure. It could just be
any interested party on request. Not a sentence to complainant?
mandatory to send to complainant.

Page 1 of 8
ISO/IEC 17025:2017 Transition Webinar Questions & Answers

2. Decision Rules
Question Answer
Fail is based on 95% Confidence level, The decision rule is to be stated. Fail will be
should this be stated in Compliance based on agreed criteria.
statement?
Will it be allowed on a test certificate to Absolutely yes, unless the lab has contracted
provide the results, together with the with you for a decision, or it is part of the test
specification, but leave it to the customer to spec. Depends on contract review.
decide pass/fail, rather than indicating it on
the test certificate?
Can the decision rule be declared as not It should take it into account (i.e. ensure it is
taking the uncertainty into account? not ridiculously large for example) Sometimes
the test specs suggest it can be ignored. So,
sometimes it can be accounted for but
dismissed and not included in the calculation.
This would be described in the statement on
test report.

If the customer says for example, they do Not a matter of documentation, the contract
not require Uncertainty to be taken into review will conclude what is to be done. The
account of a pass/fail criteria what level of lab must maintain its integrity. It should take it
documentation does this require? into account (ie ensure it is not ridiculously
large for example) Sometimes the test specs
suggest it can be ignored. So, sometimes it
can be accounted for but dismissed and not
included in the calculation. This would be
described in the statement on test report. If
the customer has supplied a decision rule, no
further assessment of risk is required, but it
must be sound.

It is worth pointing out that M3003 Appendix Yes, Thank you. The point here is that now
M contains specific detail on decision rules we appreciate that almost any decision rule is
and risk levels (probability) associated with acceptable rather than relying on the ILAC
false acceptance or reject. model.
Do the decision rules apply to observational If there are pass/fail criteria, Yes.
results e.g. positive identification of amosite
asbestos in a sample of lagging?

Page 2 of 8
ISO/IEC 17025:2017 Transition Webinar Questions & Answers

3. General
Question Answer
Should external Proficiency Testing be No such prescription. It depends on technical
done annually? and availability considerations.

Can opportunities be raised as mandatory Failure to cover opportunities and risks in the
findings? MS would be a finding.

Can an organisation holding ISO 17025 No. It should rely on the equivalence
accreditation legitimately claim ISO9001 statements in the Standard and the tripartite
certification? agreement.
The copy I have of the Final Draft – ISO/EC All references should have been to the ISO/IEC
FDIS 17025 does not have a Section under 17025:2017 and not to the DIS or FDIS. I am
7.8.1.3 as discussed in the first part of this aware of a couple of errors. Thanks.
talk.
5.3 Is it OK for the laboratory to simply say If that is true, I don’t see why the same
its scope is defined on the UKAS schedule? document should not be used, but bear in mind
that they must not depend on us for the
conformity with 17025. That is against 17011.
So, they would need to reproduce it and keep it
up to date, not just link to our website.

Will we expect the scope to cover all I take this to be about technical scoping for
activities, or just those relevant to ISO sampling, calibration and testing.
17025?
I read from clause 8.7.1 that the extent The requirement is, I agree, clearer now. I
analysis is more clearly required than the believe the intent is the same.
2005 version of the standard requires. Do
you agree?
7.8.8.2 - Does this mean that any change Reporting may be in various forms as agreed
[to the report] must be explained in the with client, but normally, I would say yes so
report? In annex for example but not in a that any reader may see this in the context of
mail associated with the report? the report.

Page 3 of 8
ISO/IEC 17025:2017 Transition Webinar Questions & Answers

4. MS Options
Question Answer
There has been no reference made to the If taking Option B then the lab will be
context of the organisation which is a key expected to ensure it adequately covers
ISO 9001 requirement; is this applicable to 17025 requirements. That’s all, so I suppose
ISO 17025? not. However, although the word context is
not used, the intention is much the same.

If a laboratory has opted for an option B Yes sections 4 to 7 in all cases to be covered
quality system, should the 17025 assessor by the MS, but issues are raised against 8.1.3
assess that the laboratory meet the for MS issues, not 8.2.
requirements of section 4 to 7?

If the lab take Option B can you confirm that We assess that the lab’s MS adequately
we still have to assess the management covers 17025 cl 4 to 7. This may be more or
system. I am already getting labs asking me less than if taking option A. It depends on the
if the assessment effort can go down as we complexity of the MS. Initially I would expect
don’t need to assess this area anymore. no change but later effort may go up or down.

For option B Do we allow non-accredited At present there is no policy on this. It is not


certification? for us to allow or not, Accredited certification
is likely to need less assessment by us to see
that it is adequate.

Having assessed labs which were part of Yes, we assess the suitability of the MS to
an ISO 9001 certified manufacturer I have a cover 17025 requirements. Deficiencies in
certain amount of scepticism about the the MS would be 8.1.1 or 8.1.3. Technical
quality of some ISO 9001 audits. With deficiencies would be in Cl 4 to 7.
Option B are [we] going to be able to raise
NCs against the management system as It is not a matter of “accepted”. The MS may
accepted under ISO9001 e.g. in the quality not be certificated anyway, or may have used
and effectiveness of their audits, document a non-accredited registrar. There is presently
control process etc? no policy on this.

Could a lab easily apply for ISO 9001 and Probably not. Some things in 9001 like
not have to change anything in their development, research etc don’t appear in
management system if they hold ISO 17025. Anyway, if it is only a lab there is no
17025:2017? point. See the tripartite agreement and joint
communique.

Page 4 of 8
ISO/IEC 17025:2017 Transition Webinar Questions & Answers

5. Risk
Question Answer
A labs own perception of risk may be very By persuasion and being a great assessor. The
different to ours. How do you deal with this? decision is the lab’s but whether the handling of
risk and opportunity is adequate is ours.
7.8.6 - How will the laboratory be in a By contract review, especially when
position to quantify the risk to the client? considering decision rules.
The laboratory may not be aware of the
clients’ requirements, and their risk
strategy.
Should the process of evaluation of a It depends on the complexity of the lab. In
specific risk be documented? order to achieve this, they will need to know
what the process is and it may therefore need
documentation.
In order to assess the suitability of the labs It certainly would help but ultimately the
determination of risk, does this mean that decision is that of the lab anyway.
the AM or Lead Assessor must have
knowledge of the industry sector the
laboratory is involved in?
So, the risk based approach will require Yes, but it is clear that the lab decides which
assessors to make judgement calls on the risks to attack. We comment on degrees of
laboratory's own judgement calls. Won't that compliance with 17025 and these judgement
lead to greater 'disagreement' about calls comprise the difference between
whether requirements of the Standard are assessing and auditing.
met or not?
Do you see any changes from a monitoring No, but I expect the necessary measures to
perspective as to what is being accepted as vary greatly, but it is very clear that this should
an acceptable risk management process? be ongoing.

Are UKAS going to look at the risks Interesting point. I would say that the
associated with us assessing risk and assessment is now more skilled and about
opportunity in the new standard? It’s not competence rather than conformity so findings
about impartiality but about consistency of will vary between labs for good reason. This is
assessment. essentially the difference between skilled
assessing and simple auditing.

Page 5 of 8
ISO/IEC 17025:2017 Transition Webinar Questions & Answers

6. Sampling
Question Answer
In sampling e.g. for deposition of pollutants, Remember that they are to report components
a soil sampling grid can have a very large relevant to later establishing the uncertainty of
uncertainty due to patchy deposition. Does measurement in the subsequent testing. This
this large uncertainty need to be (a) defined is not an uncertainty of sampling which is
and (b) combined into the overall covered by the very nature of the sampling,
uncertainty of the result of the pollutant in sampling plan design, agreement and
'concentration units per square metre'? understanding of the client etc. The laboratory
is required to report items affecting the
uncertainty of measurement and anything else
that could affect the validity of the test
results. Comments perhaps on the adequacy
of the sampling plan if supplied by the client,
comments on the need or not to homogenise or
sub sample, features of the sampling site in
practice, observation of adjacent features,
weather conditions, etc. The important thing is
that the contract review establishes the
needs. This will vary; sometimes the sampling
lab is the main contractor, sometimes the test
lab or the client. Discussions and agreement
are vital. Lessened, of course, if this is covered
in a comprehensive specification containing it
all.

Must the subsequent calibration/testing be The standard is not about accreditation, so


accredited? fundamentally no, but EA policy and our
ACC1000 requires the subsequent testing to
be accredited, as far as the sampling lab can
ensure. So, it depends on the contractual
setup.

Does sampling include the taking of sub Effectively yes, but that may be included in the
samples in the lab? test regime sometimes. Either way, any
sampling undertaken shall comply.

Reporting sampling – of [sic] [if] a lab The lab is expected to pass on all information
declares that it has sampled in accordance on things that could affect uncertainty of
to a known standard/standard requirement, measurement, not to evaluate measurement
would that negate the need to evaluate uncertainty because it is not making
measurement uncertainty? measurements. The use of a known standard
handling such matters certainly helps.

Can an organisation be accredited Yes, provided it is intended for subsequent


according to ISO 17025 for sampling testing or calibration, not necessarily by the
alone? same entity.

Page 6 of 8
ISO/IEC 17025:2017 Transition Webinar Questions & Answers

7. Traceability
Question Answer
Is there guidance for accredited No, Labs shall comply with all the clauses if
laboratories on what sections of the new they are claiming conformity with 17025. A
standard are to be covered when supplier of calibrations shall be competent;
determining the technical competence for ILAC MRA accredited, CIPM NMI listed or
traceability requirements? “the third way”. If the third way then they
would need to be assessed as competent.
Conformity with 17025 is an indicator of such
competence.

Traceability of test results are still the key Traceability is provided by features that do
added value of a third party lab to me. If not require independence. See Annex A and
customer is taken to court, the lab shall be JCGM, VIM etc. It is for the customer to be
able to provide traceable and reliable happy with lack of independence. Some 1st
results. party manufacturer’s labs are the very best as
they are the experts in their kit and have a
direct interest in getting the right answer.

6.5.2 - Does a lab supplying traceability It does not need to be accredited, nor CIPM
need to be accredited? If not, how to MRA listed but if not, establishing
ensure it complies with ISO 17025? competence would be difficult. See Appendix
to 17025, and the help in ILAC G8 Appendix
continues to apply.

Page 7 of 8
ISO/IEC 17025:2017 Transition Webinar Questions & Answers

8. Transition
Question Answer
How will the transition plan be managed for The Case Manager acts as AM, and will call for
Enterprise labs where they do not have an help as required.
AM?

Can new customers be accredited against See transition plan. Possibly, if our resources
the new standard i.e. for an initial are available and signed off.
assessment in May 2018?

After March 2018, will all applications have No, but this is advised as assessments after 1
to be for the new standard? Jan 2019 are only to new standard.

Will the IAR form have two dropdowns, one The Accreditation standard column auto
for each version, so that IAs and populates with the standard only, not the year,
Observations can be documented? so the next column, Additional accreditation
standards, will also need to be used where
assessment is against both 2005 and 2017
versions (free text fields).

Will the charges be relevant to the changes I would say, complexity, rather than changes or
made by the lab rather than the size? size. It is determined initially by simple UKAS
man day size rule but you may represent to
your manager if you disagree in a case.
Complex cases should be charged more.

I have been asked - will there be specific There is help from ISO/CASCO and we are
guidance provided for lab customers? putting some information on our website. We
are also offering awareness seminars.

When assessing to the new version of the We expect the site time to be as normal for that
standard, should this be an Initial visit in the cycle. WE would adjust the other
assessment approach, i.e. longer time on topics covered to suit. The extra work is in the
site, more depth covered. This could be transition template analysis etc. as desktop.
during the labs SU cycle.

If the assessment against the new standard I would say that we handle it as we would any
is a surveillance visit and recommendation findings and it should not be a problem. If we
is not positive and only includes one have messed up then the corrective activity
assessor could this not be an issue and would be at our expense so maybe best to
how will we handle this? have an experienced assessor or AM attend
that visit. I think it unlikely that we would need
all the technical assessors present to conclude
if the transition had been made, but the use of
decision rules in different areas needs to be
watched.

Page 8 of 8