Академический Документы
Профессиональный Документы
Культура Документы
Overview 11-3
Objective 11-3
Purpose 11-3
Instructions 11-3
System Verification Checklist 11-4
Mechanical and Visual 11-4
Electrical 11-4
Functional 11-4
Diagnostics 11-4
External System Verification 11-5
System Mechanical & Visual 11-5
System Cleaning 11-6
Electrical & Safety 11-7
Transducers 11-8
Hardware Verification 11-10
System Power Supplies and Thermal Conditions Analysis 11-10
System History Analysis 11-11
Diagnostic Verification 11-12
Operational Verification 11-13
System Functionality 11-13
Peripherals 11-15
Functionality Verification 11-16
Image Quality Verification 11-17
User Programs 11-17
Transducer Selection 11-17
2-D Performance 11-17
Wrap Up 11-22
Verification Questions 11-23
Module 11 - Verifying Aspen™ Ultrasound System Acuson Confidential
Revision History
OVERVIEW
OBJECTIVE Be able to verify to perform a system verification check on a Aspen
system.
FUNCTIONAL
Transducers • Verify all transducers used or at least linear and vector.
• Verify all ports.
• Test TEE thermal sense operation.
2-D Modality • Use Phantom and RES to check for channel failures.
• Use Phantom to verify lateral and axial resolution of pins.
• Adjust Dynamic Range and notice noise change in cysts.
Doppler • Image carotid artery to verify PW, Color, and Audio.
AEGIS • Start exam; verify Static and Dynamic clip capture and review.
• Use MO and exam utilities to verify storage to MO and Hard
disk.
Miscellaneous • Verify functionality of M-mode, CW Doppler and ECG/Physio.
• Backup user programs to hard-copy or VCR tape.
• Test battery voltage on SVC.
• Save copy of BBM to laptop (if available).
• Verify all Peripherals. Use Diagnostic Test Patterns as desired.
SYSTEM CLEANING
WARNING! - If damage to the socket is apparent, bring this to the attention of the
customer for correction.
- If the socket is damaged such that it may result in loss of ground
connection or could cause damage to the system, inform the
customer before completing any further work, and request that the
fault be corrected.
NOTE: Line voltage outside these limits can adversely affect system
performance and reliability.
HARDWARE VERIFICATION
SYSTEM POWER 1 Verify the power supplies and system operating temperature.
SUPPLIES AND For full information, refer to the Diagnostic Module.
THERMAL • At the Manufacturing Menu of Diagnostics,
CONDITIONS
• Select Utilities
ANALYSIS
• Select Power Supply Measurement
• Verify that all voltages and temperatures are within the
following limits.
NOTE: Ensure that you have a valid copy of user parameters and setup
information before replacing the battery.
NOTE: Do not clean the keyboard when diagnostics are running, to avoid
accidentally interrupting the tests or hanging the system.
1 Run the System Verification Tests Suite (or the NI suite if you have
more time.) This takes approximately 30 minutes.
2 At completion of the test suite, type STAT [RETURN] to report any
failures.
If a diagnostic test fails when run from a test suite, run the failing
test from Manual Diagnostics. If the test passes when run from
Manual diagnostics, ignore the original failure from the test suite.
Always assume a diagnostic failure is real unless:
• Manual Diagnostics proves otherwise.
• The failing test is shown in a known false failure list for that
Aspen system release level.
• You’ve verified with the Help Desk or that it is a known false
failure.
• If there are any failures, note and correlate them with any
reports made by the customer or entries in the system logs.
• Troubleshoot any failures and correct before proceeding.
• Ensure that you complete FSR and FRR details and return
diagnostic failure information with any replaced FRU.
OPERATIONAL VERIFICATION
SYSTEM 1 Verify the transmitters.
FUNCTIONALITY • Follow the instructions from the Diagnostic Module for running
the Transmitter Test.
2 Verify that the switch indicators illuminate correctly.
• Press all buttons on the User Interface which have an LED, and
ensure that the LED illuminates as expected.
• Replace the complete UI assembly or UIP as necessary.
3 Verify the system panel lights.
• Visually inspect the keyboard downlights located in the disk
box underneath the system monitor (qty 5) and ensure that they
are illuminated. Replace keyboard downlights as necessary.
• Visually inspect the transducer port downlights immediately
above the connectors (qty 5). Ensure that they are illuminated.
Replace transducer port downlights as necessary.
4 Verify the system switches and controls using the keyboard test.
• Select Keyboard Functionality Check from the Diagnostics
Utilities menu.
• The screen will display a representation of the keyboard.
• Push, rotate and slide all UI/keyboard keys and ensure that the
display changes appropriately.
• Press CODE+C to return to the main menu.
• Perform appropriate repairs if any controls or keys are
inoperative.
5 Verify the system monitor.
NOTE: Image processing for B-Mode and Color is optimized when the monitor
is set to the calibrated contrast and brightness positions (knobs out).
Additionally, general video levels to peripherals are set so that the
closest match between peripherals and the system monitor display
occurs in this mode. You should inform and encourage the customer to
use the preset positions.
NOTE: After viewing a test pattern, press CODE+C to return to the menu.
• CW Doppler:
• The spectral signal is clean with no frequency or noise
artifacts.
• The audio is clean and free from distortion and extraneous
signals.
• Calipers can be selected and appropriate measurements
taken.
• Aux. CW Doppler:
• The spectral signal is clean with no frequency or noise
artifacts.
• The audio is clean and free from distortion and extraneous
signals.
• Calipers can be selected and appropriate measurements
taken.
• CDI:
• Color is clean and free from dropout.
• Vessel fill-in is good, without excessive holes.
• No random or coherent color spikes are seen (color/noise
spikes).
• Calipers can be selected and appropriate measurements taken.
8 Verify the ECG module (Cardiac or Shared Service systems only).
Using an ECG simulator, or by hookup to yourself, verify that a
clean ECG waveform can be acquired and displayed on the system.
NOTE: The connectors are marked RA (Right Arm), RL (Right Leg) and LA
(Left Arm). You may connect them as specified, or onto your chest/
abdomen in the same general orientation.
9 Verify AEGIS system static and dynamic capture and review to the
hard disk and MO disk.
• Insert an MO disk into the disk drive.
• Press the Begin/End key to create a study.
• Capture a static image and dynamic clip in 2-D, CDI Color
Doppler Imaging, and Doppler modalities.
• Press the Begin/End key to end the study.
• Press the Study Util key, and recall/review the study from the
hard disk, and then from the MO disk.
TRANSDUCER
SELECTION
◆ To Check the ability of the system to recognize all available
transducers
1 Insert the transducer connector into one of the three active
transducer ports on the front Panel of the system and rotate the
locking knob clockwise to lock the transducer in place.
2 Press the softkey corresponding to the transducer you want to
check.
3 Press the MULTIHERTZ® toggle up or down to change the imaging
frequency available to you on the transducer by your system
4 Select modalities from User interface to check available modes for
each transducer
5 Repeat the procedure for all available transducers.
2-D PERFORMANCE Since the invention of modern Ultrasound equipment,
measurement techniques have changed very little. Typical tests
involve measurements using external reference such as a tissue
equivalent test phantom.
We also know that the characteristics of the Phantoms vary with
time and is dependent on several factors. Hence the use of
phantoms is best only as a compromise between need for
consistency and the near representation of human anatomy and
pathology.
The following tests provide practical Image Quality checks on 2-D
image.
The test phantom used here is the RMI 403 GS.
10
Axial Resolution Group
12
14
16
18
Figure 11-1 The RMI 403 Test Phantom
Channel failures The following test is recommended for initial diagnosis of any
channel failures or drop outs. This test is also valid for testing a
transducer for delamination.
◆ To test channel related failures
• Scan a tissue equivalent test phantom and slide the transducer
along the surface of the phantom and watch the image for any
dark vertical bands below the skin line which move relative to
the detail in the phantom i.e. they are static with respect to the
screen.
• The above mentioned test can be confirmed by complementing
it with Diagnostic tests e.g.
• Transmit Chain Test/ Transmit Delay Test
• Receive-Apodization Test
• Mixer Clock Test
For details of these diagnostic tests refer to the Diagnostic Tools
Module.
Detail Resolution The Detail or Spatial resolution is the ability of the ultrasound
system, to discriminate between two closely spaced objects and is
specified as the distance in mm. at which two discrete objects can be
differentiated. The resolution figures are commonly measured at
the focal point and represent best attainable values. Resolution in
the axial and lateral dimensions are usually specified
independently.
Axial
Lateral
Dynamic Range Cysts are fluid filled structures that are anechoic and usually
weakly attenuating. Its acoustic image has a distinctive pattern that
should represent the size, shape, and the consistency of the cyst.
This is also a good way of observing signal-to-noise ratio.
◆ To check Dynamic Range
1 Scan the bottom set of cysts in a test phantom and increase
DYNAMIC RANGE key.
2 Note the value of dynamic range when the noise floor is seen.
3 Look out for excessive noise within the cysts as dynamic range is
increased.
WRAP UP
1 Verify that, for the system and peripherals:
• All panels and covers have been replaced correctly.
• Cables and installation wiring are clean and tidy and connected
correctly.
• Peripherals, transducers and other equipment are situated as
the customer wishes.
• All debris and tools are removed from the installation area.
2 Verify that the system boots correctly.
• Power on the system and ensure that it boots to imaging and
that the fans are running at low RPM.
3 Instruct the customer of maintenance performed.
VERIFICATION QUESTIONS
1 Which logs should be checked at every system verification?