Вы находитесь на странице: 1из 7

Scandinavian Journal of Clinical & Laboratory Investigation

Vol. 69, No. 5, September 2009, 585–591

ORIGINAL ARTICLE

Potential preanalytical errors in whole-blood analysis: Effect of


syringe sample volume on blood gas, electrolyte and lactate values

PIRJO HEDBERG1,∗, AILA MAJAVA1,∗, KAI KIVILUOMA2 & PASI OHTONEN3


1Laboratory, Oulu University Hospital, Department of Clinical Chemistry, University of Oulu, 2Department of
Anesthesiology, Oulu University Hospital, and 3Department of Anesthesiology and Surgery, Oulu University Hospital,
Scand J Clin Lab Invest Downloaded from informahealthcare.com by CDL-UC Santa Cruz on 10/26/14

Oulu, Finland

Abstract
Background: Arterial blood samples are sensitive to bias because of the physiological properties of blood. Several errors
can occur in the preanalytical phase leading to incorrect diagnosis and improper treatment of patients. Collection of a blood
specimen, as well as its handling and transport, belong to the key factors to affect the accuracy and good quality of clinical
laboratory analysis. Methods: The aim of this study was to validate the effect of different sample volumes on the blood
gas, electrolyte and lactate values using 3 mL Rapidlyte plastic syringes with filter cap and Rapidlab 865 blood gas analyser.
Also, the stability of blood gas analyser parameters with different sample volume was studied. Results: No substantial
change in blood gas, electrolyte and lactate parameters was found when the results of 3 mL, 1.8 mL sample volumes in the
For personal use only.

3 mL syringes were compared. The sample volume of 1.0 mL or 1.5 mL in the 3 mL syringe is not suitable for the measure-
ment of oxygen tension, especially when accurate results of pO2 and arterial blood is needed for patient’s diagnosis.
Conclusions: The minimum sample volume when blood gases, electrolytes and lactate are all measured with the
Rapidlab system should be 1.8 mL using 3 mL Rapidlyte plastic syringe with filtercap. According to this study  1.8 mL
sample volumes can provide inaccurate results and can impose biases on measurements.

Key Words: Blood gas analyser, preanalytical, sample volume, syringe, whole-blood analysis

Introduction
conditions for treatment, and preventing adverse
Blood-gas analysis is widely used in the diagnosis of events to human or system errors [1].
diseases and clinical follow-up. Alterations in values The preanalytical phase plays a major role in
give a clue to the nature of the disturbance, if the the source of errors in whole-blood analysis.
primary problem is in gas changes or metabolism. Arterial blood sample is sensitive to bias because
Successful treatment can be seen as returning the blood physiologically changes after it leaves
values towards normal. the body [2,3].
Blood-gas analysis has pronounced value in The arterial specimen collection and analysis
diagnostics and follow-up in intensive care treatment. guidelines published by the Clinical and Laboratory
Successful ventilator treatment is based on reliable Standards Institute (CLSI; Previous NCCLS; National
blood-gas analysis. Trends in blood-gas analysis give Committee for Clinical Laboratory Standards) has
the basis for the follow-up of life-threatening situations provided specific recommendations regarding speci-
like sepsis. men collection devices, sample handling, specimen
In diagnostics it is usually important to know transport and storage conditions [4,5]. Arterial
the level, where the values are. In follow-up it is specimens should be collected in a plastic syringe, left
obligatory to see reliable changes in absolute values. at room temperature and analysed within 30 minutes.
It is the responsibility of health organizations to guar- Blood collected for special analyses, like shunt analy-
antee a high level of healthcare by using adequate ses, should be analysed within 5 minutes. Actually,
methodologies and instruments, creating secure over the past several years the practice has led to a

∗P.Hedberg and A. Majava contributed equally to this work.


Correspondence: Pirjo Hedberg, PhD, Laboratory, Oulu University Hospital, Oulu, PO Box 500, FI-90029 OYS, Finland. Tel: 358 8 3155453.
Fax: 358 8 5134409. E-mail: pirjo.hedberg@ppshp.fi

(Received 22 August 2008; accepted 4 March 2009)

ISSN 0036-5513 print/ISSN 1502-7686 online © 2009 Informa UK Ltd. (Informa Healthcare, Taylor & Francis AS)
DOI: 10.1080/00365510902878716
586 P. Hedberg et al.

change of collection device from glass syringes to sample aspiration or processing faults. All results
plastic ones. Using the right sampling equipment were processed anonymously.
and improving the instructions to medical staff on All the patients granted written permission for
handling blood samples including the possible sources blood samples to be collected, and the local hospital
of errors are essential for the sampling quality. ethical committee approved the study.
To our knowledge, no exact studies of the effect
of syringe sample volume to the blood gas, electro-
Analytical methods
lyte and lactate values using Siemens Rapidlab
865 analyser and Siemens Rapidlyte plastic syringes All analyses were done using the Rapidlab 865
have been published.Very often the blood gas syringes analyser (Siemens Medical Solutions Diagnostics,
sent to the core laboratory from the hospital wards are Tarrytown, NY 10591-5097, USA).
insufficiently filled with patient blood. The aim of The analytical performance of the analyser was
this study was to examine this bias and to give more followed with the protocol used in the blood gas
information and instructions to medical staff on analysers in Oulu University Hospital laboratory.
handling blood samples which should be applied in The performance of the analyser was checked with
Scand J Clin Lab Invest Downloaded from informahealthcare.com by CDL-UC Santa Cruz on 10/26/14

daily practice in the hospital laboratories, critical care the three quality controls with different levels of
units and emergency rooms. parameters (RapidQC Complete 1, 2 and 3, Bayer
Health Care). It was performed by analysing one
level in turn in the morning and afternoon and a
Materials and methods patient pool in the morning.
Samples
The study protocol
Arterial venting 3 mL syringe with filter cap (04180028,
Bayer Corporation, East Walpole, MA 02032-1597, Before sample collection the full grant for permissions
USA) with total balanced heparin concentration of were asked from the patients. After permission four
about 70 IU/3 mL for blood gas, electrolyte and lactate specimens were collected from each patient; 3.0 mL,
For personal use only.

analysis were collected from patients in the critical care 1.8 mL, 1.5 mL and 1.0 mL in 3.0 mL syringe samples.
unit in Oulu University Hospital. The arterial samples The sample volumes were chosen as following; the
were obtained from an arterial cannula and the mixed 3.0 mL is a fully filled syringe and is a reference sam-
venous samples from a pulmonary artery catheter ple volume. The 1.8 mL is a volume, which has been
(PAC). The total amount of sample serials was 127, of orally quoted by the manufacturer and is used in the
which 71 were arterial and 56 pulmonary. The samples Oulu University Hospital, the last two volumes were
were analysed as soon as possible after collection. The chosen to find out the exact limit of volume where
samples were kept at 4C to slow down the metabo- the possible changes in the results would appear.
lism keeping the maximum storage time under 15 min- The sample volume 1.0 mL in the 3 mL syringe
utes. The samples were mixed very thoroughly would have been insufficient for the analyser. Samples
according to the manufacturer of the syringes before were mixed thoroughly and analysed as soon as
they were analysed to ensure that a homogeneous part possible. The results from 1.0 mL, 1.5 mL and 1.8 mL
of the samples are introduced into the analyser. Bub- samples were compared to the results from full-filled
bles in the samples were avoided as much as possible, (3 mL) samples.The data was processed in two groups,
but if bubbles had formed in the syringe, they were venous and arterial blood samples.
removed immediately after sampling. To avoid air bub-
bles, the Rapidlyte syringes have a filter cap to remove
Statistical methods
them after sample collection. The filter cap device is
also used as a stopper tip cap during transport to the All data was expressed as mean with standard devia-
analyser. In the stability test the samples in the first tion (SD), unless otherwise indicated. Statistical
group were stored at room temperature (232C) analyses were performed with SPSS for Windows
and mixed thoroughly before analysis and in the software package (SPSS 15.0.1, SPSS Inc., Chicago,
second group they were stored at cooled conditions Illinois, USA). Since the 3.0 mL, 1.8 mL, 1.5 mL and
(4C) before mixing and analysing. 1.0 mL blood samples were collected from the same
person the results were analysed using analysis of
variance for repeated measurements (R-ANOVA), if
Patients
the sphericity assumption did not hold the p-value
The effect of sample volume on the results was studied according to Greenhouse-Geisser is presented. When
with a total of 127 patients. Four samples were the R-ANOVA showed a p-value  0.05 a pair wise
collected from each patient in plastic heparin bal- comparison between 3.0 mL (reference volume) and
anced syringes. No sample selection criteria were other volumes was done, mean differences with 95%
used. Data exclusions were only made when there confidence intervals and p-values with Sidak’s
was clear evidence of incomplete or suboptimal multiple comparison corrections were presented.
Sample volume needed for blood gas parameters 587

Results The change was especially seen in higher pO2


values. To clarify whether this difference is due to the
Changes observed in the blood gas, electrolyte
stability of the 1 mL sample, we investigated the
and lactate results of blood syringe samples
effect of storage of samples at  4C and 23  2C
with different sample volumes are summarized in
with different time points (Table III). This study was
Tables I and II.
performed with 25 1 mL specimens taken from
When comparing the 1.8 mL sample volume
5 different patients. The first sample of each patient
results to the baseline values established using a
was analysed at a time point of one to five minutes
3.0 mL sample volume, the bias % between the groups
(baseline value) and the other samples were analysed
ranged between  7.8% and  8.7%. The standard
at the time points of 15, 30, 60 and 90 minutes. The
deviations of biases for the base excess in blood
mean changes were less than 10% for pO2 parameter
(BEvt) parameter in all groups were high, which may
(in the tested range of 4.1–20.6 kPa). The mean
be due to the low absolute values (negative vs posi-
changes for hemoglobin derivatives (COHb and
tive). Also, the standard deviations for the fraction of
MetHb) and BEvt were over  10% at both storage
carboxyhemoglobin in total hemoglobin (COHb) and
temperatures over the time inconsistently. The stan-
Scand J Clin Lab Invest Downloaded from informahealthcare.com by CDL-UC Santa Cruz on 10/26/14

the fraction of methemoglobin in total hemoglobin


dard deviations of bias%s were high and the tested
(MetHb) parameters were high in all groups and the
values of these parameters in the samples were very
amounts of COHb and MetHb were very low.
low. Lactate values increased also over the time
All the analytes in the group 1.5 mL arterial samples producing mean percentage changes over 10% within
were found to change less than 10%, except COHb in 30 minutes at room temperature and within 60 min
arterial samples, which showed a decrease of 14.8% at  4C. The mean change for pCO2 was 10.8% at
from the 3 mL sample value. In other parameters the 23  2C during 90 minutes storage.
mean percentage changes varied between  8.7% and
7.3%. The bias % for pO2 was  8.6%.
Among the arterial blood samples over 10%
changes between the results of 3 mL and 1.0 mL Discussion
For personal use only.

samples were observed for pO2 (13.5%), BEvt The typical preanalytical errors are caused by the
(11.3%) and COHb ( 19.8%). BEvt increased air bubbles in the sample, inhomogeneous samples,
14.8% and COHb decreased 13.5% with 1.0 mL hemolysis in samples, improper storage of samples,
sample volume compared to 3.0 mL sample volume. clots in the specimen, dilution of sample due to saline
In other parameters the bias % varied between  8.9 in arterial lines and liquid heparin, heparin inter-
and  8.2% in both of the groups. ference, arterial samples mixed with venous blood
The R-ANOVA showed a p-value  0.05 only for and patient in an unsteady state of ventilation. The
hemoglobin, hematocrit, and Ca2 (7.4) among the impact of each of these causes of error must be
arterial samples, but the pair wise comparison was clearly acknowledged by those who are collecting
also made for those. Although there were changes in the specimens, analysing them and interpreting
the bias % with a few parameters, the statistically the results to avoid making false decisions in
significant changes were observed more often. The patient managements.
significant changes (p  0.05) were found for pO2, An air bubble left in the sample for a few minutes
O2Sat and Na in both of groups and with all the can affect the pO2 values significantly [6,7]. An air
sample volumes compared to 3 mL sample volume. bubble with a relative volume of 0.5–1% of the sample
Among the arterial samples p-values were  0.05 for may seriously affect the pO2 value of the sample. In
Ca2(7.4) between the results of 3 mL and 1.5 mL our study the bubbles in the samples were avoided,
samples. In the comparison of 3 mL and 1 mL sam- but if bubbles had formed in the syringe, they were
ples the p-values were  0.05 for COHb, MetHb, K removed immediately after sampling. To avoid air
and Ca2 (7.4) and lactate. bubbles, the Rapidlyte syringes have a filter cap to
The R-ANOVA showed a p-value 0.05 only for remove them after sample collection.
hemoglobin, hematocrit and pH among the venous Significant differences have been found between
samples, but the pair wise comparison was also made the type of collection device, storage conditions and
for those. In our investigation, significant changes time to analysis in several studies. Mahoney et al. [8]
were found with several parameters in the venous examined the changes in oxygen measurements in
sample group when the results of 3.0 mL and 1.0 mL blood using plastic and glass syringe samples stored ice.
samples were compared. The changes were significant The authors concluded that samples collected in a
for carbon dioxide tension (pCO2), HCO3 act, BEvt, plastic syringe should not be stored in ice. The glass
COHb, Kand Ca2 (7.4).The last parameter changed syringes may still be more appropriate in these
significantly also with the 1.8 mL sample volume. specific conditions. Wu et al. [9] according to
Additionally, when comparisons were made between their studies recommended storage at room temper-
3 mL and all the other sample volumes, the significant ature when specimens are collected in plastic syringes
changes were found for lactate. and immediate analysis of the specimen. If the
Scand J Clin Lab Invest Downloaded from informahealthcare.com by CDL-UC Santa Cruz on 10/26/14
For personal use only.

588

Table I. Changes of arterial blood gas, electrolyte and lactate values induced by different syringe sample volume. Data are presented as mean with standard deviation (SD), differences between the mean
values (in absolute values and bias %) and statistical significances (a pair wise comparison; a p value  0.05 was considered statistically significant and is marked with asterisks∗). The 95% confidence
interval (95%CI) was presented, when a p value was  0.05.
P. Hedberg et al.

Difference between Difference between Difference between


3.0 mL and 1.8 mL 3.0 mL and 1.5 mL 1.0 mL 3.0 mL and 1.0 mL
3.0 mL (n  71) 1.8 mL (n 71) 1.5 mL (n  71) (n 71)
Arterial sample Mean (SD) Mean (SD) Bias % p-value Mean (SD) Bias % p-value Mean (SD) Bias % p-value

pH 7.37 (0.05) 7.37 (0.05) 0.1 0.03∗, 0.000 7.37 (0.05) 0.02 0.90 7.37 (0.05)  0.01  0.9
to 0.007
pCO2 (kPa) 5.4 (0.6) 5.4 (0.6)  0.4 0.93 5.4 (0.5) 0.5 0.79 5.3 (0.5) 1.3 0.07
pO2 (kPa) 14.4 (4.1) 15.0 (4.2)  3.9  0.001∗,  0.883 15.7 (4.3)  8.6  0.001∗,  1.690 16.4 (4.5)  13.5  0.001∗,  2.516
to  0.253 to  0.779 to  1.386
HCO3 act (mmol/L) 23 (3) 23 (3) 0.4 0.63 23 (3) 0.8 0.22 23 (3) 1.2 0.03∗, 0.014
to 0.524
BEvt (mmol/L)  2.0 (3.1)  2.1 (3.1)  7.8 0.08  2.1 (3.1)  8.7 0.16  2.2 (3.0)  11.3 0.06
tHb (g/L) 103 (15) 102 (15) 0.3 0.64 103 (15) 0.04  0.9 102 (16) 0.4 0.58
Hct (g/dL) 30 (5) 30 (4) 0.2 0.98 30 (4)  0.03  0.9 30 (5) 0.4 0.79
O2SAT (%) 97.0 (1.9) 97.3 (1.7)  0.3 0.01∗,  0.4 97.5 (1.7)  0.5  0.001∗,  0.698 97.7 (1.7)  0.7  0.001∗,  0.927
55 to  0.038 to  0.249 to  0.428
COHb (%) 0.51 (0.31) 0.46 (0.29) 8.7 0.10 0.43 (0.27) 14.8 0.004∗, 0.018 0.41 (0.28) 19.8  0.001∗, 0.039
to 0.131 to 0.161
MetHb (%) 0.73 (0.20) 0.74 (0.20)  2.1 0.63 0.77 (0.20)  5.2 0.07 0.79 (0.19)  8.9 0.002∗,  0.112
to  0.018
Na (mmol/L) 138 (4) 138 (4) 0.2  0.001∗, 0.162 138 (4) 0.3  0.001∗, 0.283 138 (4) 0.6  0.001∗, 0.601
to 0.466 to 0.554 to 0.934
K (mmol/L) 4.2 (0.5) 4.1 (0.5) 0.2 0.54 4.1 (0.5) 0.2 0.54 4.1 (0.5) 0.8 0.002∗, 0.010
to 0.060
Ca2 (7.4) (mmol/L) 1.08 (0.06) 1.10 (0.10)  1.3 0.54 1.09 (0.06)  0.9  0.001∗,  0.013 1.10 (0.06)  1.6  0.001∗,  0.022
to  0.006 to  0.012
Lactate (mmol/L) 1.49 (0.54) 1.50 (0.54)  0.6 0.21 1.51 (0.54)  1.4 0.003∗,  0.038 1.52 (0.54)  1.8 0.002∗,  0.045
to  0.006 to  0.007

pO2, Oxygen tension; pCO2, Carbon dioxide tension; tHb, Total haemoglobin; metHb or FmetHb, Fraction of methemoglobin in total hemoglobin; COHB or FCOHb, Fraction of carboxyhemoglobin
in total hemoglobin; Hct, Hematocrit; HCO3 act, Actual concentration of bicarbonate; K, Potassium; Na, Sodium; Ca2, Ionized calcium; BEvt, Base excess in blood; O2Sat, Oxygen saturation.
Scand J Clin Lab Invest Downloaded from informahealthcare.com by CDL-UC Santa Cruz on 10/26/14
For personal use only.

Table II. Changes of venous blood gas, electrolyte and lactate values induced by different syringe sample volumes. Data are presented as mean with standard deviation (SD), differences between the
mean values (in absolute values and bias %) and statistical significances (a pair wise comparison; a p value  0.05 was considered statistically significant and is marked with asterisks∗). The 95% confidence
interval (95%CI) was presented, when a p value was  0.05.

Difference between Difference between Difference between


3.0 mL and 1.8 mL 3.0 mL and 1.5 mL 1.0 mL 3.0 mL and 1.0 mL
3.0 mL (n 56) 1.8 mL (n 56) 1.5 mL (n 56) (n 56)
Venous Sample Mean (SD) Mean (SD) Bias % p-value Mean (SD) Bias % p-value Mean (SD) Bias % p-value

pH 7.33 (0.05) 7.33 (0.05)  0.02 0.54 7.33 (0.05)  0.01 0.99 7.33 (0.05)  0.01 0.99
pCO2 (kPa) 6.3 (0.6) 6.3 (0.6) 0.3 0.85 6.2 (0.6) 0.7 0.27 6.2 (0.6) 1.3 0.001∗, 0.026
to 0.130
pO2 (kPa) 4.5 (0.6) 4.7 (0.7)  2.3  0.001∗,  0.166 4.7 (0.6)  3.0  0.001∗,  0.192 4.8 (0.7)  5.2  0.001∗,  0.329
to  0.046 to  0.079 to  0.143
HCO3 act 24 (3) 24 (3) 0.04  0.9 24 (3) 0–7 0.39 24 (3) 1.2 0.003∗, 0.078
(mmol/L) to 0.511
BEvt  1.6 (3.2)  1.6 (3.2) 1.3  0.9  1.7 (3.0)  8.6 0.58  1.8 (3.2)  14.8 0.02∗, 0.021
(mmol/L) to 0.443
tHb (g/L) 101 (14) 101 (14) 0.2 0.99 100 (14) 0.6 0.47 100 (14) 0.6 0.36

Hct (g/dl) 30 (4) 30 (4) 0.5 0.71 30 (4) 0.8 0.32 29 (4) 0.9 0.10

O2SAT (%) 60.3 (8.6) 61.8 (9.1)  2.4  0.001∗,  2.238 62.2 (8.8)  3.1  0.001∗, 2.670 63.5 (8.9)  5.3  0.001∗,  4.253
to  0.676 to  1.076 to  2.083
COHb (%) 0.55 (0.30) 0.53 (0.28) 4.3 0.76 0.51 (0.30) 8.5 0.29 0.48 (0.29) 13.5 0.01∗, 0.011
to 0.139
MetHb (%) 0.74 (0.19) 0.76 (0.16)  2.6 0.75 0.78 (0.17)  5.6 0.08 0.78 (0.17)  6.0 0.11

Na (mmol/L) 139 (3) 139 (3) 0.1 0.03∗, 0.013 138 (3) 0.3  0.001∗, 0.266 138 (3) 0.6  0.001∗, 0.631
to 0.291 to 0.606 to 1.008
K (mmol/L) 4.2 (0.6) 4.2 (0.6) 0.1 0.99 4.2 (0.6) 0.6 0.31 4.2 (0.6) 1.0 0.001∗, 0.013
to 0.071
Ca2(7.4) 1.06 (0.05) 1.07 (0.05)  0.6  0.001∗,  0.011 1.06 (0.05)  0.4 0.51 1.07 (0.05)  1.2  0.001∗,  0.020
(mmol/L) to  0.003 to  0.006
Lactate 1.42 (0.43) 1.44 (0.43)  1.2 0.01∗,  0.030 1.44 (0.43)  1.6 0.01∗,  0.040 1.44 (0.42)  1.7 0.002∗,  0.040
(mmol/L) to  0.004 to  0.005 to  0.007

pO2, Oxygen tension; pCO2, Carbon dioxide tension; tHb, Total haemoglobin; metHb or FmetHb, Fraction of methemoglobin in total hemoglobin; COHB or FCOHb, Fraction of carboxyhemoglobin
in total hemoglobin; Hct, Hematocrit; HCO3–act, Actual concentration of bicarbonate; K, Potassium; Na, Sodium; Ca2, Ionized calcium; BEvt, Base excess in blood; O2Sat, Oxygen saturation.
Sample volume needed for blood gas parameters
589
590 P. Hedberg et al.

Table III. The effect of storage of 1 mL samples at  4C and 23  2C with different time points. Stability of blood gas, electrolyte and
lactate values. The results were compared to the 0-5 min results. Data are presented as mean of bias % with standard deviation (SD).

23 2C  4C


n 5 15 min 30 min 60 min 90 min 15 min 30 min 60 min 90 min

pH 0.1 (0.1) 0.1 (0.1) 0.3(0.2) 0.6 (0.1) 0.1 (0.1)  0.1 (0.2) 0.3 (0.1) 0.2 (0.2)
pCO2 (kPa)  1.8 (1.4)  1.6 (5.1)  6.6 (3.7)  10.8 (3.0)  1.1 (2.2) 1.9 (4.6)  5.1 (1.5)  4.4 (2.3)
pO2 (kPa)  0.1 (3.4)  0.9 (0.7)  2.2 (8.0)  0.7 (6.6)  1.2 (6.4)  2.8 (5.8)  3.8 (8.6)  2.9 (5.7)
HCO3 act  0.7 (1.7)  0.4 (4.7)  0.7 (0.9)  0.1 (1.1)  0.1 (1.9) 0.9 (2.5)  0.4 (1.4)  0.6 (1.3)
(mmol/L)
BEvt  23.5  117.0 7.3 154.7  15.2  6.6 (11.6)  7.1 (13.6)  3.4
(mmol/L) (102.4) (271.0) (15.5) (224.7) (17.7) (16.2)
tHb (g/L) 0.6 (4.0) 3.5 (5.4) 1.4 (4.0)  0.1 (2.1)  1.7 (2.1)  1.7 (2.8)  1.5 (3.3)  1.0 (2.0)
Hct (g/dl)  0.4 (6.0) 2.2 (6.4) 1.2 (5.4)  0.8 (3.2)  1.5 (3.1)  0.7 (2.9)  1.5 (3.1)  0.7 (1.6)
O2SAT (%)  0.1 (3.0)  0.8 (0.8)  0.4 (1.6) 0.6 (1.4) 0.4 (4.8)  3.7 (6.9)  0.7 (1.1)  1.1 (5.0)
COHb (%)  90.0  120.0  85.0  105.0  13.3  66.7  51.7 5.0 (71.1)
(175.5) (164.3) (151.7) (94.2) (29.8) (188.2) (101.4)
Scand J Clin Lab Invest Downloaded from informahealthcare.com by CDL-UC Santa Cruz on 10/26/14

MetHb (%)  1.8  7.3 28.2 27.5 (18.9)  5.0 (25.9) 14.7 (24.3) 8.5 (35.6) 20.0 (27.7)
(43.3) (54.4) (28.1)
Na 0.7 (0.9) 0.6 (0.6) 0.7 (0.7) 0.6 (0.6)  0.1 (0.2)  0.1 (0.4) 0.2 (0.3) 0.2 (0.2)
(mmol/L)
K (mmol/L) 1.1 (0.8) 3.5 (3.5) 2.7 (0.6) 3.4 (0.7) 0.2 (0.3)  0.4 (0.7)  0.8 (0.9)  1.0 (1.3)
Ca2 (7.4)  0.2 (0.8) 1.4 (2.8) 0.9 (1.1) 1.6 (1.2)  0.2 (0.8) 0.5 (1.4) 0.6 (0.8) 1.3 (0.8)
(mmol/L)
Lactate  4.0 (2.5)  13.6  34.5  51.7 (21.8)  3.7 (2.3)  6.1 (9.1)  13.7  20.5
(mmol/L) (7.5) (14.8) (11.9) (12.3)

pO2, Oxygen tension; pCO2, Carbon dioxide tension; tHb, Total haemoglobin; metHb or FmetHb, Fraction of methemoglobin in total
hemoglobin; COHB or FCOHb, Fraction of carboxyhemoglobin in total hemoglobin; Hct, Hematocrit; HCO3 act, Actual concentration
of bicarbonate; K, Potassium; Na, Sodium; Ca2, Ionized calcium; BEvt, Base excess in blood; O2, Oxygen saturation.
For personal use only.

analysis is going to be delayed, the samples should hemoglobin derivative parameter values using different
be drawn and stored in glass [10]. According to the sample volumes remain still unclear. It may be due
studies of Pretto and Rochford in 1994 [11] for to the analyser used; the plunger of the syringe is very
the blood gas samples with high arterial pO2 should close to the sample needle of the analyser when 1 mL
be analysed immediately for accurate results and if a sample volume is used. Also, the possibility of lower
longer storage time is required, then the use of glass numbers of blood cells in the syringe may cause
syringes and adequate specimen cooling may be variations of blood gases. The stability study showed
required. According to Wiwanitkit in 2006 glass that pO2 is stabile in 1 mL samples at room
syringes are better than plastic for preserving temperature and at  4C for 90 minutes thus indi-
arterial blood gas for oxygen partial pressure deter- cating that the reason for differences was not the
mination: an explanation based on nanomaterial stability. The stability study also showed that the dif-
composition [12]. In the 100% oxygen test, the ferences in the hemoglobin derivate parameters may
glass (not plastic) syringes have been recommended be due to the very low levels of these parameters.
to be used. The syringe should be cooled immedi- The precision performance of each parameter has
ately, even when the sample is analyzed as soon as also to be considered when finalizing the results. Our
possible [13]. In the present study the main analyses experience with the Rapidlab 865 analysers are that
were done without delay thus there was no reason the between run precisions of the parameters with
for use of glass syringes. Additionally, in the stability different level of concentrations and values differ. For
study no substantial changes were observed in high example the precision for pO2 is typically 4–17%
pO2 values (bias% 3.8). (CV) or even higher depending of the level of
Excess sodium heparin solution as an anti- RapidQC complete control. The highest CV% is
coagulant, the type of heparin or dilution of blood usually observed for low level of pO2. The CV%s for
samples can significantly change the measured COHb are normally 15, 12.5 and 5% at the levels of
values [7,14–19]. According to the manufacturer 3, 5 and 20%, respectively. Among the metabolites,
Rapidlyte syringes used in this study are preheparinized lactate has the highest CV% at the very low
with dry heparin covering the whole syringe inner to concentration level. The CV% can reach up to 16%
avoid the possible dilution of the sample, and avoid too at the level of under 1.0 mmol/L, or even higher at
high an anticoagulant/blood volume ratio. Thus, the lower concentration values. The CV%s for MetHb
results of our study can not be explained by the pos- are usually 15, 12.5 and 5% at the levels 2.4, 7 and
sible effect of the ratio of anticoagulant/blood volume. 16%. The typical CV%s for all the other parameters
The reason for the differences in the pO2, BEvt and are 4.5% and for pH 0.17. The CV%s is
Sample volume needed for blood gas parameters 591

presumably higher in the very low levels of COHb References


and MetHb. The very high CV% may have had an [1] Rambaldi M, Baranzoni MT, Coppolecchia P, Moschello JN,
affect the results of this study. The precisions of the Novaco F. Blood gas and patient safety: considerations
calculated parameters (Bevt, HCO3act and O2Sat) based on experience developed in accordance with the
have not been followed in our laboratory. One Risk Management perspective. Clin Chem Lab Med.
weakness of our study was the constricted sample 2007;45:774–80.
material. During the study the samples with patho- [2] Batchelder PB, Raley D. Maximizing the laboratory setting
logical hemoglobin derivative (MetHb and COHb) for testing devices and understanding statistional output
and base excess values were not available. Thus in pulse oximetry. Anesth Analg. 2007;105(6 Suppl.):S85–94.
[3] Risch A, Biedler A, Mertzlufft F. Preanalytical errors on the
the possible 10% difference in these parameters at
determination of arterial O(2)-partial pressure and their
the pathological level remains uncertain. The SD impact on the AaDO(2). Anaesthesist. 2000;49:29–33.
for these three analytes was also very large in the [4] NCCLS. Blood gas and pH analysis and related measurements.
stability experiment. Further studies are needed to NCCLS document C46-A4 Wayne, PA. NCCLS, 2001.
clarify what are the exact relative differences at [5] NCCLS. Procedures for the collection of arterial blood speci-
higher values. mens. NCCLS document H11-A4. Wayne, PA. NCCLS, 2004.
Scand J Clin Lab Invest Downloaded from informahealthcare.com by CDL-UC Santa Cruz on 10/26/14

[6] Biswas CK, Ramos JM, Agroyannis B, Kerr DNS. Blood gas
Ricós et al. [20] have published the databases on
analysis: effect of air bubbles in syringe and delay in
biological variation and derived from these data the
estimation. Br Med J. 1982;284:923–7.
desirable specifications for total error, imprecision [7] Wuillemin WA, Gerber AU. Sources of errors in the pre-
and bias has been calculated [21]. Among the param- analytical phase of blood gas analysis. Scweiz Rundsch
eters in this list the desirable biases for hemoglobin, Med Prax. 1995;84:200–3.
lactate, pCO2 and pH are 1.8, 8.0, 1.8, 1.0%, respec- [8] Mahoney JJ, Harvey JA, Wong RJ, Van Kessel AL. Changes
tively. Thus, the present study indicates that the in oxygen measurement when whole blood is stored
biases with all sample volumes tested for these four in iced plastic syringe or glass syringes. Clin Chem.
parameters are lower than these thresholds. 1991;37:1244–8.
[9] Wu EY, Baranzanji KW, Johnson RL. Source of error on
From a clinical point of view the errors in arterial
A-aDO2 calculated from blood stored in plastic and
pO2 in 1.5 and 1.0 mL samples are critical. In bor-
For personal use only.

glass syringes. J Appl Physiol. 1997;82:196–202.


derline these errors can lead to misdiagnosis. The risk [10] Knowles T, Mullin R, Hunter J, Douce FH. Effects of syringe
for wrong clinical conclusion is high especially dur- material, sample storage time and temperature on
ing follow-up if the sample size varies. Measured blood gases and oxygen saturation in arterialized
10% differences between 3.0 mL and 1.5 mL samples human blood samples. Respir Care. 2006;51:732–6.
in COHb have no clinical significance in normal [11] Pretto JJ, Rochford PD. Effects of sample storage time,
situations, but in carbon monoxide intoxication they temperature and syringe type on blood gas tensions in samples
with high oxygen partial pressures. Thorax. 1994;49:610–2.
might be misleading.
[12] Wiwanitkit V. Glass syringes are better than plastic for
In conclusion, using Rapidlyte plastic syringes preserving arterial blood gas for oxygen partial pressure
with filter cap, the minimum sample volume when determination: an explanation based on nanomaterial
blood gases, electrolytes and lactate are all measured composition. In J Nanomedicine. 2006;1:223–4.
should be 1.8 mL using a 3 mL syringe. No substantial [13] Smeenk F, Janssen J, Arends B, Harff J, van den Bosch J,
Schönberger J, Postmus P. Effects of four different methods
changes were found when 1.8 mL sample results were
of sampling arterial blood and storage time on gas tensions
compared to 3 mL sample results. According to our and shunt calculation in the 100% oxygen test. Eur Respir
results, 1.0 and 1.5 mL sample volumes, especially J. 1997;10:910–3.
when accurate results of pO2 and arterial blood is [14] Toffaletti J, Ernst P, Hunt P, Abrams B. Dry electrolyte-
needed for patient’s diagnosis, are not recommended. balanced heparinised syringes evaluated for determining
ionized calcium and other electrolytes in whole blood.
Clin Chem. 1991;37:1730–3.
[15] Hopper K, Rezende Ml, Haskins SC. Assessment of the
Acknowledgements dilution of blood samples with sodium heparin on blood gas,
electrolyte, and lactate measurements in dogs. Am J Vet Res.
In memoriam: We owe this study and our warmest 2005;66:656–60.
thanks to Dr Päivi Kaukoranta for her invaluable [16] Börner U, Müller H, Höge R, Hempelmann G. The influence
collaboration during this work. of anticoagulant on acid-base status and blood-gas analysis.
Acta Anaesthesiol Scand. 1984;28:277–9.
[17] Goodwin NM, Schreiber MT. Effects of anticoagulants on acid-
base and blood gas estimations. Crit Care Med. 1979;7:473–4.
Declaration of interest: The authors report no
[18] Crawford AG. Liquid versus non-liquid heparinization of
conflicts of interest. The authors alone are responsible blood samples. Resp Ther. 1983;5:103–5.
for the content and writing of the paper. [19] Hutchison AS, Ralston SH, Dryburgh FJ, Small M,
Fogelmann I. Too much heparin: possible source of error in
blood gas analysis. Brit Med J. 1983;287:1131–2.
Abbreviations: [20] Ricós C, Alvarez V, Cava F, García-Lario JV, Hernández A,
Jiménez CV, Minchinela J, Perich C, Simón M. Current
CLSI: Clinical and Laboratory Standards Institute; databases on biological variation: pros, cons and progress.
NCCLS: National Committee for Clinical Laboratory Scand J Clin Lab Invest. 1999;59:491–500.
Standards. [21] http://westgard.com/biodatabase1.htm

Вам также может понравиться