Академический Документы
Профессиональный Документы
Культура Документы
Construction:
The fixed coil is divided into two sections to give a more uniform field near the
centre . The fixed coils are usually wound with heavy wire carrying the main
current . the wire is stranded where necessary to reduce eddy current losses in
conductors. The coils are usually varnished and backed to form a solid assembly .
The fixed coil is air cored and is supported on a former made of preferably
ceramic as metallic former would weaken the field of the fixed coil due to eddy
currents when use with a.c. A single element instrument has one moving coil .
the moving coil is wound either as a self-sustaining coil or else on a non-metallic
former . The moving coil is mounted on a aluminum spindle carrying pointer ,
which can move over the graduated scale. The controlling torque is provided by
the two hair springs , acting also as leads to the moving coil . Light and rigid
construction is used for the moving coil .Fixed and moving coils are connected in
series and the ends are connected to the instrument terminals. Damping is often
employed by air friction type mechanism consisting of air piston moving in
enclosed chamber or by pair of aluminum vanes , attached to the spindle at the
bottom. These vanes move in sector shaped chambers . eddy current damping
can not be used in these instruments as the operating field is very weak due to
use of air core coils and any introduction of a permanent magnet required for
eddy current damping would distort the operating magnetic field . The field
produced by the fixed coils is somewhat weaker than in other type of
instruments . It is nearly 0.0005 to 0.006 Wb/m . It is necessary to shield the
electrodynamometer type instrument from the effect of stray magnetic fields .
Thus the coil assembly is shielded by enclosing them in a casing of high
permeability alloy to protect the instrument from the external magnetic field .
1
Torque equation:
[ and ]
2
=
Therefore total electrical energy input=change in energy stored+ mechanical
energy produced
Mechanical energy=
Since the self-inductances of the coils are constant and therefore and are
both equal to zero.
Mechanical energy=
Suppose instantaneous torque= and is the change in deflection , then
Mechanical energy=work done=
Thus or
Shape of scale :
And
It is clear from the above expression that instrument does not have pure square
law response . In electrodynamometer instruments , as usually constructed ,
varies from about to 135 , i.e. from to from the position of zero
mutual inductance and as changes in value of M over this range are not large i.e.
is practically constant over this range , the instrument shows almost a square
law response .
3
Operation with D.C:
In Ammeter , the fixed and moving coils are series so that they can carry the
4
same current
i.e. and
5
Electrodynameter type Voltmeter:
In voltmeter configuration , the two coils are connected in series along with a
high value non-inductive resistance . The deflecting torque But
, where
impedance of the series combination of two coils., V= unknown voltage to be
measured. The deflection
The main source of error in dynamometer type instruments are due to a) low
torque/weight ratio b) frequency c) eddy current external magnetic field d)
temperature change.
The magnetic field produced by air core coil is essentially small and therefore ,
the flux linkage per amp in the moving coil are also small;( Only 3 to 4 % of
flux linkage per A in the PMMC instrument) . Therefore the deflecting torque
is low . Thus to produce a reasonable deflecting torque , the m.m.f of the
moving coil must be increased . The m.m.f of the moving coil can be
increased by increasing the current through the coil or by increasing the no. of
6
turns . The current through the moving coil can not be increased beyond
200mA as otherwise the spring would be heated too much . On the other
hand an increase the no. of turns of the moving coil, will give rise a heavy
moving system and , therefore , the friction losses in electrodynamometer
type instrument are larger than in other type .
b) Frequency
In order to reduce frequency error , the ratio of currents in fixed and moving
coils should be independent of frequency . This requires the time constants
of two circuits should be the same .
Let = resistance , inductance and impedance of fixed coil circuit.
=resistance , inductance and impedance of the moving coil circuit.
Then
Thus the ratio of currents with A.C. is the same as with D.C. , if the time
constants ( ratio ) of fixed and moving coil circuits are same . Therefore ,
the frequency error in shunted ammeters can be eliminated by having equal
time constants for both fixed and moving coil circuits . Also the phase
difference between the two currents is zero.
7
c) Eddy current
8
e) Temperature changes