This manual is made for trained Topcon service


engineers, wrong use of this manual can cause
malfunction of the system.

© TOPCON EUROPE MEDICAL BV


CV-5000 Series Connection Manual Version 1.04

Purpose
This manual is written to support Topcon representatives during the installation of Topcon
stand alone system like CV-5000(PC) phoropter with the KB-50 Dial controller, the CV-5000
Lite with the KB-10 Dial controller, and a Chart device like ACP-8, MC-3 or CC-100 series, and
all other common possible connections with peripheral Topcon instruments and IMAGEnet i-
base.

Overview
The most current connections between CV-5000(pc) and peripheral instruments as well as
IMAGEnet i-base and/or Color Mapping Software is described in this manual.
In this manual connections are made with the latest Topcon instruments. If older
instruments need to be connected, first try the settings as they are written in this manual
for the current models. If you do not succeed, contact Topcon Europe for further support.

!
Please read this Connection Manual with care before you
connect Topcon CV-5000(PC) to TOPCON peripheral instruments,
IMAGEnet i-base software and/or Color Mapping Software.

This manual is written by:


David van der Klooster (TEM)
Assistant Product Manager
dkl@topcon.eu

2
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Optional accessories for all configurations


All optional accessories like cables needed for connections between all Topcon instruments
mentioned in this manual are described with item numbers and can be ordered with Topcon
Europe Medical BV.

Questions / Information
If you have questions after reading this manual or if you can not solve the problem concerning
connections after consulting this manual you can contact Product Support at Topcon Europe
Medical BV. For additional information you can also check the manual of the separate instruments.
These instrument manuals can be found on the Topcon Europe Medical FTP site.
All available software for Topcon instruments can also be found on the Topcon Europe Medical FTP
site.

Note:
Be aware that if you consulting product support at Topcon Europe Medical that you provide the
Product Support managers always with:

1) Software version of the Topcon instrument


2) Firmware version of the Topcon instrument
3) Serial number of the Topcon Instrument
!
To obtain this information see chapter 1.0 Software and Firmware versions.
The serial number is always found on the instrument itself.
Also provide the Topcon Europe Medical product support managers with the right formulated
question, with relevant and detailed information about the problem. Simply the remark “it doesn’t
work” is not enough.

Topcon Europe Medical BV


Office: Essebaan 11 P.O. Box 145
2908 LJ Capelle aan den IJssel 2900 AC Capelle aan den IJssel
The Netherlands The Netherlands
0031-(0)10-4585077
0031-(0)10-2844940
www.topcon.eu

Note:
Be aware that some selections on pictures used in this manual are
example selections. Always read the text provided with the pictures
to make the right selections.
!
3
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Escape Clauses
 TOPCON shall not take any responsibility for damage due to fire, earthquakes, actions by
third persons and other accidents, or the negligence and misuse of the user, and use
under unusual conditions.

 TOPCON shall not take any responsibility for damage derived from the use or
unavailability of this connection manual, such as loss of business profits and suspension
of business.

 TOPCON shall not take any responsibility for damage caused by usage other than that
described in this connection manual.

 TOPCON shall not take any responsibility for the results of diagnosis while using
equipment which is connected by the use of this connection manual.

Definitions
Definitions used
in this manual Consist of

CV-5000 Compact power supply with controllers: KB-50, KB-10

CV-5000PC Compact power supply with controllers: CV-5000PC

CV-5000(PC) Compact power supply with controllers: KB-50, KB-10, CV-5000PC

Main Chapter
Sub Chapter
Sub sub Chapter
Referring to Chapter

Please connect all instruments and make all settings in exactly the same order that are
described in all separate chapters in this Manual.

If you work this way, soon you will discover the logic of the setup of this manual.

Success.

4
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Contents
PURPOSE……………………………………………………………………………………………………..2

OVERVIEW……………………………………………………………………………………………………2

OPTIONAL ACCESSORIES FOR ALL CONFIGURATIONS…………………………………………...3

QUESTIONS / INFORMATION……………………………………………………………………………...3

ESCAPE CLAUSES………………………………………………………………………………………….4

DEFINITIONS………………………………………………………………………………………………….4

CONTENTS……………………………………………………………………………………………………5

1 General Information

1.0 SOFTWARE AND FIRMWARE VERSIONS………………………………………………………..16


1.0.1 Checking software versions on the KB-50 ......................................................................................16
1.0.2 Checking software versions on the KB-10 ......................................................................................17
1.0.3 Checking software versions on the CV-5000PC .............................................................................17

1.1 CABLE DESCRIPTIONS……………………………………………………………………………...18


1.1.1 Cables needed for the CV-5000(PC) stand alone setup.................................................................18
1.1.2 Cables needed for the CV-5000 Lite stand alone setup .................................................................19
1.1.3 Cable item 1830057 Length 5 meter............................................................................................20
1.1.4 Cable item 1830008 Length 5 meter............................................................................................21
1.1.5 Cable item 1830095 Length 0,6 meter.........................................................................................22

1.2 CABLE LENGTH AND CABLE INSTALLATION………………………………………………….22


1.2.1 Cable length ....................................................................................................................................22
1.2.2 Cable installation .............................................................................................................................22

1.3 INTERCHANGEABILITY OF CV-3000 AND CV-5000 COMPONENTS………………………...23

1.4 CV-5000 COMPACT POWER SUPPLY DESCRIPTION………………………………………….24

1.5 OVERVIEW DIP-SWITCH SETTINGS COMPACT POWER SUPPLY…………………………..25

1.6 OVERVIEW DIP-SWITCH SETTINGS KB-10………………………………………………………27

5
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

2 CV-5000 Stand Alone

2.0 PHOROPTER  COMPACT POWER SUPPLY………………………………………………….28


2.0.1 Cable connection.............................................................................................................................28
2.0.2 Compact power supply dip-switch settings .....................................................................................28

2.1 KB-50  COMPACT POWER SUPPLY…………………………………………………………..29


2.1.1 Cable connection.............................................................................................................................29
2.1.2 KB-50 settings .................................................................................................................................29
2.1.3 Compact power supply dip-switch settings .....................................................................................29

2.2 KB-10  COMPACT POWER SUPPLY…………………………………………………………..30


2.2.1 Cable connection.............................................................................................................................30
2.2.2 KB-10 settings .................................................................................................................................30
2.2.3 Compact power supply dip-switch settings .....................................................................................30

2.3 CV-5000PC  COMPACT POWER SUPPLY…………………………………………………….31


2.3.1 Purpose of setup .............................................................................................................................31
2.3.2 Cable connection.............................................................................................................................31
2.3.3 Compact power supply dip-switch settings .....................................................................................31
2.3.4 CV-5000PC software settings (COMM port) ...................................................................................32

3 Refraction Setup

3.0 ACP-8 MC  CV-5000 RS-232C CABLE CONNECTION……………………………………..33


3.0.1 Cable connection.............................................................................................................................33
3.0.2 Compact power supply dip-switch settings .....................................................................................33
3.0.3 KB-10 settings .................................................................................................................................33
3.0.3.1 Dip-switch setting ..........................................................................................................................33
3.0.3.2 Chart (device) setting.....................................................................................................................34
3.0.3.3 Chart Pattern setting.......................................................................................................................34
3.0.4 KB-50 settings .................................................................................................................................35
3.0.4.1 Chart Connect setting.....................................................................................................................35
3.0.4.2 Chart (Far) setting ..........................................................................................................................35
3.0.4.3 Chart (Near) setting........................................................................................................................35
3.0.5 CV-5000PC settings........................................................................................................................36
3.0.5.1 Chart Connect setting.....................................................................................................................36
3.0.5.2 Chart (Far) setting ..........................................................................................................................36
3.0.5.3 Chart (Near) setting........................................................................................................................36

3.1 ACP-8 E,ACP-8 EM  CV-5000 INFRA-RED LIGHT EMITTER……………………………..37


3.1.1 Light Emitter connection..................................................................................................................37
3.1.2 Compact power supply dip-switch settings (infra-red channel).......................................................37
3.1.3 Infra-red channel table for the ACP-8 Remote Controller ...............................................................37
3.1.4 KB-10 settings .................................................................................................................................37
3.1.4.1 Dip-switch setting ..........................................................................................................................37
3.1.4.2 Chart (device) setting.....................................................................................................................38
3.1.4.3 Chart Pattern setting.......................................................................................................................38
3.1.5 KB-50 settings .................................................................................................................................39
3.1.5.1 Chart Connect setting.....................................................................................................................39
3.1.5.2 Chart (Far) setting ..........................................................................................................................39
3.1.5.3 Chart (Near) setting........................................................................................................................39
3.1.6 CV-5000PC settings........................................................................................................................40

6
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.1.6.1 Chart Connect setting.....................................................................................................................40


3.1.6.2 Chart (Far) setting ..........................................................................................................................40
3.1.6.3 Chart (Near) setting........................................................................................................................40

3.2 MC-3  CV-5000 RS-232C CABLE CONNECTION…………………………………………...41


3.2.1 Cable connection.............................................................................................................................41
3.2.2 Compact power supply dip-switch settings .....................................................................................41
3.2.3 KB-10 settings .................................................................................................................................41
3.2.3.1 Dip-switch setting ..........................................................................................................................41
3.2.3.2 Chart (device) setting.....................................................................................................................42
3.2.3.3 Chart Pattern setting.......................................................................................................................42
3.2.4 KB-50 settings .................................................................................................................................43
3.2.4.1 Chart Connect setting.....................................................................................................................43
3.2.4.2 Chart (Far) setting ..........................................................................................................................43
3.2.4.3 Chart (Near) setting........................................................................................................................43
3.2.5 CV-5000PC settings........................................................................................................................44
3.2.5.1 Chart Connect setting.....................................................................................................................44
3.2.5.2 Chart (Far) setting ..........................................................................................................................44
3.2.5.3 Chart (Near) setting........................................................................................................................44
3.2.6 MC-3 distance and Alignment emitter distance settings .................................................................45
3.2.7 MC-3 Baud rate dip-switch settings ................................................................................................45

3.3 MC-3  CV-5000 INFRA-RED LIGHT EMITTER……………………………………………….46


3.3.1 Light Emitter connection..................................................................................................................46
3.3.2 Compact power supply dip-switch settings (infra-red channel).......................................................46
3.3.3 KB-10 settings .................................................................................................................................46
3.3.3.1 Dip-switch setting ..........................................................................................................................46
3.3.3.2 Chart (device) setting.....................................................................................................................47
3.3.3.3 Chart Pattern setting.......................................................................................................................47
3.3.4 KB-50 settings .................................................................................................................................48
3.3.4.1 Chart Connect setting.....................................................................................................................48
3.3.4.2 Chart (Far) setting ..........................................................................................................................48
3.3.4.3 Chart (Near) setting........................................................................................................................48
3.3.5 CV-5000PC settings........................................................................................................................49
3.3.5.1 Chart Connect setting.....................................................................................................................49
3.3.5.2 Chart (Far) setting ..........................................................................................................................49
3.3.5.3 Chart (Near) setting........................................................................................................................49
3.3.6 MC-3 Distance and Alignment Emitter distance settings ................................................................50
3.3.7 MC-3 Other dip-switch settings (for LR-1)......................................................................................50
3.3.8 Infra-red channel table for remote controllers, alignment emitter and MC-3 ..................................51

3.4 CC-100 SERIES  CV-5000 RS-232C CABLE CONNECTION………………………………52


3.4.1 Cable connection.............................................................................................................................52
3.4.2 Compact power supply dip-switch settings .....................................................................................52
3.4.3 KB-10 settings .................................................................................................................................52
3.4.3.1 Dip-switch setting ..........................................................................................................................52
3.4.3.2 Chart (device) setting.....................................................................................................................53
3.4.3.3 Chart Pattern setting.......................................................................................................................53
3.4.4 KB-50 settings .................................................................................................................................54
3.4.4.1 Chart Connect setting.....................................................................................................................54
3.4.4.2 Chart (Far) setting ..........................................................................................................................54
3.4.4.3 Chart (Near) setting........................................................................................................................54
3.4.5 CV-5000PC settings........................................................................................................................55
3.4.5.1 Chart Connect setting.....................................................................................................................55
3.4.5.2 Chart (Far) setting ..........................................................................................................................55
3.4.5.3 Chart (Near) setting........................................................................................................................55
3.4.6 CC-100 Series Chart DLL installation in KB-50 & CV-5000PC ......................................................56
7
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.4.6.1 Purpose ..........................................................................................................................................56


3.4.6.2 Software conditions .......................................................................................................................56
3.4.6.3 Location Chart DLL files...............................................................................................................56
3.4.6.4 Installation of Chart DLL files in CV-5000PC ..............................................................................56
3.4.6.5 Installation of Chart DLL files in KB-50 .......................................................................................56
3.4.7 Placing cover sheet CC-100 Series on remote controller ...............................................................58
3.4.8 Infra-red channel table for the Remote Controller...........................................................................58
3.4.9 CC-100 Series settings ...................................................................................................................59
3.4.9.1 Menu Settings overview ................................................................................................................59
3.4.9.2 Menu Settings selections (serial device menu & IR devices menu) ............................................60
3.4.9.3 Distance setting..............................................................................................................................60
3.4.9.4 Setting serial (RS-232C) connection for use with KB-50 / CV-5000PC .......................................61
3.4.9.5 Setting serial (RS-232C) connection for use with KB-10..............................................................61
3.4.9.6 Setting CC-100 series for use with Remote Controller..................................................................62

3.5 CC-100 SERIES  CV-5000 INFRA-RED LIGHT EMITTER………………………………….63


3.5.1 Light Emitter connection..................................................................................................................63
3.5.2 Compact power supply dip-switch settings (infra-red channel).......................................................63
3.5.3 KB-10 settings .................................................................................................................................63
3.5.3.1 Dip-switch setting ..........................................................................................................................63
3.5.3.2 Chart (device) setting.....................................................................................................................64
3.5.3.3 Chart Pattern setting.......................................................................................................................64
3.5.4 KB-50 settings .................................................................................................................................65
3.5.4.1 Chart Connect setting.....................................................................................................................65
3.5.4.2 Chart (Far) setting ..........................................................................................................................65
3.5.4.3 Chart (Near) setting........................................................................................................................65
3.5.5 CV-5000PC settings........................................................................................................................66
3.5.5.1 Chart Connect setting.....................................................................................................................66
3.5.5.2 Chart (Far) setting ..........................................................................................................................66
3.5.5.3 Chart (Near) setting........................................................................................................................66
3.5.6 CC-100 Series Chart DLL installation in KB-50 & CV-5000PC ......................................................67
3.5.6.1 Purpose ..........................................................................................................................................67
3.5.6.2 Software conditions .......................................................................................................................67
3.5.6.3 Location Chart DLL files...............................................................................................................67
3.5.6.4 Installation of Chart DLL files in CV-5000PC ..............................................................................67
3.5.6.5 Installation of Chart DLL files in KB-50 .......................................................................................67
3.5.7 Placing cover sheet CC-100 Series on remote controller ...............................................................69
3.5.8 Infra-red channel table for the Remote Controller...........................................................................69
3.5.9 CC-100 Series settings ...................................................................................................................70
3.5.9.1 Menu Settings overview ................................................................................................................70
3.5.9.2 Menu Setting selections (serial device menu & IR devices menu) ...............................................71
3.5.9.3 Distance setting..............................................................................................................................71
3.5.9.4 Setting Infra-Red connection for use with KB-50 / CV-5000PC...................................................72
3.5.9.5 Setting Infra-Red connection for use with KB-10 .........................................................................72
3.5.9.6 Setting CC-100 series for use with Remote Controller..................................................................73

4 Refraction Setup with External Instruments

4.0 COMPACT POWER SUPPLY  COMPACT POWER SUPPLY………………………………74


4.0.1 Cable connection.............................................................................................................................74
4.0.2 Compact power supply dip-switch settings .....................................................................................74
4.0.2.1 2 Compact power supply units.......................................................................................................74
4.0.2.2 3 Compact power supply units.......................................................................................................75
4.0.2.3 4 Compact power supply units.......................................................................................................75

8
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.1 CL-200/2800 OR EZ-200  CV-5000(PC)………………………………………………………...76


4.1.1 Cable connection.............................................................................................................................76
4.1.2 Compact power supply dip-switch settings .....................................................................................76
4.1.3 KB-10 settings .................................................................................................................................77
4.1.4 KB-50 settings .................................................................................................................................77
4.1.5 CV-5000PC settings........................................................................................................................78
4.1.6 CL-200/2800 settings ......................................................................................................................79
4.1.6.1 RS-232C setting .............................................................................................................................79
4.1.6.2 Format setting ................................................................................................................................79
4.1.7 EZ-200 settings ...............................................................................................................................80
4.1.7.1 RS-232C setting .............................................................................................................................80
4.1.7.2 Format setting ................................................................................................................................80

4.2 RM/KR-8800/8900  CV-5000(PC)………………………………………………………………..81


4.2.1 Cable connection.............................................................................................................................81
4.2.2 Compact power supply dip-switch settings .....................................................................................81
4.2.3 KB-10 settings .................................................................................................................................82
4.2.4 KB-50 settings .................................................................................................................................82
4.2.5 CV-5000PC settings........................................................................................................................83
4.2.6 RM-8800/8900 settings ...................................................................................................................84
4.2.6.1 Format setting ................................................................................................................................84
4.2.6.2 Baud-Rate setting...........................................................................................................................84
4.2.6.3 Output data setting .........................................................................................................................84
4.2.7 KR-8800/8900 settings....................................................................................................................85
4.2.7.1 Format setting ................................................................................................................................85
4.2.7.2 Baud-Rate setting...........................................................................................................................85
4.2.7.3 Output data type.............................................................................................................................86
4.2.7.4 Output data setting .........................................................................................................................86

4.2A KR-7000P  CV-5000(PC)………………………………………………………………………..87


4.2A.1 Cable connection ..........................................................................................................................87
4.2A.2 Compact power supply dip-switch settings...................................................................................87
4.2A.3 KB-10 settings...............................................................................................................................88
4.2A.4 KB-50 settings...............................................................................................................................88
4.2A.5 CV-5000PC settings .....................................................................................................................89
4.2A.6 KR-7000P settings ........................................................................................................................90
4.2A.6.1 PRINT OUT settings...................................................................................................................90
4.2A.6.2 ONLINE settings ........................................................................................................................90
4.2A.6.3 Dipswitch settings .......................................................................................................................90

4.3 TRK-1P  CV-5000(PC)…………………………………………………………………………….91


4.3.1 Cable connection.............................................................................................................................91
4.3.2 Compact power supply dip-switch settings .....................................................................................91
4.3.3 KB-10 settings .................................................................................................................................92
4.3.4 KB-50 settings .................................................................................................................................92
4.3.5 CV-5000PC settings........................................................................................................................93
4.3.6 TRK-1P settings ..............................................................................................................................93
4.3.6.1 Output data type.............................................................................................................................93
4.3.6.2 Data Form setting...........................................................................................................................94
4.3.6.3 Data Output Port ............................................................................................................................94
4.3.6.4 Baud Rate setting ...........................................................................................................................94

9
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5 Refraction Setup with i-base

5.0 CV-5000 (KB-50/10)  I-BASE………………………………………………………………….95


5.0.1 Cable connection.............................................................................................................................95
5.0.2 Compact power supply dip-switch settings .....................................................................................95
5.0.3 KB-10 settings .................................................................................................................................96
5.0.4 KB-50 settings .................................................................................................................................96
5.0.5 i-base settings (for Compact power supply)....................................................................................97
5.0.5.1 Plug-in installation .........................................................................................................................97
5.0.5.2 Serial port setup .............................................................................................................................99
5.0.5.3 Advanced settings ........................................................................................................................100
5.0.6 Decimal symbol setting in Windows for i-base..............................................................................101

5.1 CV-5000PC  I-BASE…………………………………………………………………………..103


5.1.1 CV-5000PC  Compact Power Supply……………………………………...See Chapter 2.3
5.1.2 CV-5000PC settings......................................................................................................................103
5.1.2.1 Import/Export setup .....................................................................................................................103
5.1.2.2 File Format Version .....................................................................................................................104
5.1.2.3 RS-232C Baud Rate.....................................................................................................................104
5.1.3 i-base settings (for CV-5000PC) ...................................................................................................105
5.1.3.1 Plug-in installation .......................................................................................................................105
5.1.3.2 CV Win setup...............................................................................................................................107
5.1.4 Decimal symbol setting in Windows for i-base..............................................................................108

6 Refraction Setup with External Instruments and i-base

6.0 CL-200/2800 OR EZ-200  CV-5000(PC)  I-BASE……………………………………...109


6.0.1 CL-200/2800 or EZ-200  CV-5000(PC)..................................................See Chapter 4.1
6.0.2 CV-5000  i-base..................................................................................See Chapter 5.0
6.0.3 CV-5000PC  i-base……………………………..…………………….…...See Chapter 5.1

6.1 CL-200/2800 OR EZ-200  I-BASE  CV-5000…………………………………………...110


6.1.1 CL-200/2800 or EZ-200  i-base ...............................................................................................110
6.1.2 Cable connection...........................................................................................................................110
6.1.3 CL-200/2800 settings ....................................................................................................................111
6.1.3.1 RS-232C setting ...........................................................................................................................111
6.1.3.2 Format setting ..............................................................................................................................111
6.1.4 EZ-200 settings .............................................................................................................................112
6.1.4.1 RS-232C setting ...........................................................................................................................112
6.1.4.2 Format setting ..............................................................................................................................112
6.1.5 i-base settings (for CL-200/2800 or EZ-200) ................................................................................113
6.1.5.1 Plug-in installation .......................................................................................................................113
6.1.5.2 Serial port setup ...........................................................................................................................115
6.1.5.3 Advanced settings ........................................................................................................................116
6.1.6 i-base  CV-5000………………………………………………………....…See Chapter 5.0

6.2 CL-200/2800 OR EZ-200  I-BASE  CV-5000PC……………………………………...118


6.2.1 CL-200/2800 or EZ-200  i-base……….……………………………..……..See Chapter 6.1.1
6.2.2 i-base  CV-5000PC………………………………………………..…...….See Chapter 5.1

6.3 RM/KR-8800/8900  CV-5000(PC)  I-BASE……………………………………………..118


6.3.1 RM/KR-8800/8900  CV-5000(PC) ...........................................................................................118
6.3.1.1 Output data type...........................................................................................................................118

10
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.3.2 CV-5000  i-base………………………………………………..…..……...See Chapter 5.0


6.3.3 CV-5000PC  i-base..............................................................................See Chapter 5.1

6.4 RM/KR-8800/8900  I-BASE  CV-5000…………………………………………………..119


6.4.1 RM/KR-8800/8900  i-base .......................................................................................................119
6.4.2 Cable connection...........................................................................................................................119
6.4.3 RM-8800/8900 settings .................................................................................................................120
6.4.3.1 Format setting ..............................................................................................................................120
6.4.3.2 Baud-Rate setting.........................................................................................................................120
6.4.3.3 Output data setting .......................................................................................................................120
6.4.4 KR-8800/8900 settings..................................................................................................................121
6.4.4.1 Format setting ..............................................................................................................................121
6.4.4.2 Baud-Rate setting.........................................................................................................................121
6.4.4.3 Output data type...........................................................................................................................122
6.4.4.4 Output data setting .......................................................................................................................122
6.4.5 i-base settings (for RM/KR-8800/8900).........................................................................................123
6.4.5.1 Plug-in installation .......................................................................................................................123
6.4.5.2 Serial port setup ...........................................................................................................................125
6.4.5.3 Advanced settings ........................................................................................................................126
6.4.6 i-base  CV-5000..................................................................................See Chapter 5.0

6.5 RM/KR-8800/8900  I-BASE  CV-5000PC……………………………………………….128


6.5.1 RM/KR-8800/8900  i-base………………………….………….……………See Chapter 6.4.1
6.5.2 i-base  CV-5000PC……………………………..…………………….…...See Chapter 5.1

6.6 TRK-1P  CV-5000(PC)  I-BASE………………………………………………………….128


6.6.1 TRK-1P  CV-5000(PC) ............................................................................................................128
6.6.1.1 Output data type...........................................................................................................................128
6.6.2 CV-5000  i-base………………………………….………………………...See Chapter 5.0
6.6.3 CV-5000PC  i-base…………………………………………………………See Chapter 5.1

6.7 TRK-1P  I-BASE  CV-5000……………………………………………………………….129


6.7.1 TRK-1P  i-base ........................................................................................................................129
6.7.2 Cable connection...........................................................................................................................129
6.7.3 TRK-1P settings ............................................................................................................................129
6.7.3.1 Output data type...........................................................................................................................129
6.7.3.2 Data Form setting.........................................................................................................................130
6.7.3.3 Data Output Port ..........................................................................................................................130
6.7.3.4 Baud Rate setting .........................................................................................................................130
6.7.4 i-base settings (for TRK-1P)..........................................................................................................131
6.7.4.1 Plug-in installation .......................................................................................................................131
6.7.4.2 Serial port setup ...........................................................................................................................133
6.7.4.3 Advanced settings ........................................................................................................................134
6.7.5 i-base  CV-5000……………………………….………………….………..See Chapter 5.0

6.8 TRK-1P  I-BASE  CV-5000PC……………………………………………………………136


6.8.1 TRK-1P  i-base……………………………………..……………………......See Chapter 6.7.1
6.8.2 i-base  CV-5000PC…………………………………………….……….….See Chapter 5.1

11
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7 Refraction Setup with CL/EZ to External Instruments and i-base

7.0 CL/EZ  RM/KR  CV-5000(PC)  I-BASE……………………………………………..137


7.0.1 CL/EZ  RM/KR  CV-5000(PC)............................................................................................137
7.0.2 Cable connection...........................................................................................................................137
7.0.3 Compact power supply dip-switch settings ...................................................................................138
7.0.4 KB-10 settings ...............................................................................................................................138
7.0.5 KB-50 settings ...............................................................................................................................139
7.0.6 CV-5000PC settings......................................................................................................................139
7.0.7 CL-200/2800 settings ....................................................................................................................140
7.0.7.1 RS-232C setting ...........................................................................................................................140
7.0.7.2 Format setting ..............................................................................................................................140
7.0.8 EZ-200 settings .............................................................................................................................141
7.0.8.1 RS-232C setting ...........................................................................................................................141
7.0.8.2 Format setting ..............................................................................................................................141
7.0.9 RM-8800/8900 settings .................................................................................................................142
7.0.9.1 CL input setting ...........................................................................................................................142
7.0.9.2 Format setting ..............................................................................................................................142
7.0.9.3 Baud-Rate setting.........................................................................................................................142
7.0.9.4 Output data setting .......................................................................................................................143
7.0.10 KR-8800/8900 settings................................................................................................................143
7.0.10.1 CL input setting .........................................................................................................................143
7.0.10.2 Format setting ............................................................................................................................143
7.0.10.3 Baud-Rate setting.......................................................................................................................144
7.0.10.4 Output data type.........................................................................................................................144
7.0.10.5 Output data setting .....................................................................................................................144
7.0.11 CV-5000  i-base………….……………………………………….………See Chapter 5.0
7.0.12 CV-5000PC  i-base………….………….………………………….…….See Chapter 5.1

7.1 CL/EZ  RM/KR  I-BASE  CV-5000(PC)……………………………………………..145


7.1.1 CL/EZ  RM/KR  i-base........................................................................................................145
7.1.2 Cable connection...........................................................................................................................145
7.1.3 CL-200/2800 settings ....................................................................................................................146
7.1.3.1 RS-232C setting ...........................................................................................................................146
7.1.3.2 Format setting ..............................................................................................................................146
7.1.4 EZ-200 settings .............................................................................................................................147
7.1.4.1 RS-232C setting ...........................................................................................................................147
7.1.4.2 Format setting ..............................................................................................................................147
7.1.5 RM-8800/8900 settings .................................................................................................................148
7.1.5.1 CL input setting ...........................................................................................................................148
7.1.5.2 Format setting ..............................................................................................................................148
7.1.5.3 Baud-Rate setting.........................................................................................................................148
7.1.5.4 Output data setting .......................................................................................................................149
7.1.6 KR-8800/8900 settings..................................................................................................................149
7.1.6.1 CL input setting ...........................................................................................................................149
7.1.6.2 Format setting ..............................................................................................................................149
7.1.6.3 Baud-Rate setting.........................................................................................................................150
7.1.6.4 Output data type...........................................................................................................................150
7.1.6.5 Output data setting .......................................................................................................................150
7.1.7 i-base settings (for RM/KR-8800/8900).........................................................................................151
7.1.7.1 Plug-in installation .......................................................................................................................151
7.1.7.2 Serial port setup ...........................................................................................................................153
7.1.7.3 Advanced settings ........................................................................................................................154
7.1.8 i-base  CV-5000……………………………………………………...…….See Chapter 5.0
7.1.9 i-base  CV-5000PC……………………………………………………......See Chapter 5.1

12
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.2 CL/EZ  TRK-1P  CV-5000(PC)  I-BASE…………………………………………….156


7.2.1 CL/EZ  TRK-1P  CV-5000(PC)...........................................................................................156
7.2.2 Cable connection...........................................................................................................................156
7.2.3 Compact power supply dip-switch settings ...................................................................................157
7.2.4 KB-10 settings ...............................................................................................................................157
7.2.5 KB-50 settings ...............................................................................................................................158
7.2.6 CV-5000PC settings......................................................................................................................158
7.2.7 CL-200/2800 settings ....................................................................................................................159
7.2.7.1 RS-232C setting ...........................................................................................................................159
7.2.7.2 Format setting ..............................................................................................................................159
7.2.8 EZ-200 settings .............................................................................................................................160
7.2.8.1 RS-232C setting ...........................................................................................................................160
7.2.8.2 Format setting ..............................................................................................................................160
7.2.9 TRK-1P settings ............................................................................................................................161
7.2.9.1 Output data type...........................................................................................................................161
7.2.9.2 Data Form setting.........................................................................................................................161
7.2.9.3 Data Output Port ..........................................................................................................................162
7.2.9.4 Baud Rate setting .........................................................................................................................162
7.2.9.5 CL input setting ...........................................................................................................................162
7.2.10 CV-5000  i-base……………….………………………………….………See Chapter 5.0
7.2.11 CV-5000PC  i-base…........................................................................See Chapter 5.1

7.3 CL/EZ  TRK-1P  I-BASE  CV-5000(PC)…………………………………………….163


7.3.1 CL/EZ  TRK-1P  i-base ......................................................................................................163
7.3.2 Cable connection...........................................................................................................................163
7.3.3 CL-200/2800 settings ....................................................................................................................164
7.3.3.1 RS-232C setting ...........................................................................................................................164
7.3.3.2 Format setting ..............................................................................................................................164
7.3.4 EZ-200 settings .............................................................................................................................165
7.3.4.1 RS-232C setting ...........................................................................................................................165
7.3.4.2 Format setting ..............................................................................................................................165
7.3.5 TRK-1P settings ............................................................................................................................166
7.3.5.1 Output data type...........................................................................................................................166
7.3.5.2 Data Form setting.........................................................................................................................166
7.3.5.3 Data Output Port ..........................................................................................................................166
7.3.5.4 Baud Rate setting .........................................................................................................................167
7.3.5.5 CL input setting ...........................................................................................................................167
7.3.6 i-base settings (for TRK-1P)..........................................................................................................168
7.3.6.1 Plug-in installation .......................................................................................................................168
7.3.6.2 Serial port setup ...........................................................................................................................170
7.3.6.3 Advanced settings ........................................................................................................................171
7.3.7 i-base  CV-5000……………………………………………………………See Chapter 5.0
7.3.8 i-base  CV-5000PC………………………………………….…………….See Chapter 5.1

8 Refraction Setup with Colormap 32 and i-base

8.0 KR-8100P  COLORMAP 32  CV-5000…………………………………………………….173


8.0.1 Cable connection...........................................................................................................................173
8.0.2 Compact power supply dip-switch settings ...................................................................................174
8.0.3 KB-10 settings ...............................................................................................................................175
8.0.4 KB-50 settings ...............................................................................................................................175
8.0.5 KR-8100P settings ........................................................................................................................176
8.0.5.1 ON-LINE settings ........................................................................................................................176
8.0.5.2 Output data setting .......................................................................................................................176
8.0.5.3 PLACIDO settings .......................................................................................................................177
8.0.6 Colormap 32 (KRP32) settings ...................................................................................................178
8.0.6.1 KR COM Port setting...................................................................................................................178
13
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0.6.2 Setup calibration data...................................................................................................................178


8.0.6.3 CV COM port setting...................................................................................................................179
8.0.6.4 Select instrument..........................................................................................................................179

8.0A KR-7000P  COLORMAP 32  CV-5000…………………………………………………..180


8.0A.1 Cable connection ........................................................................................................................180
8.0A.2 Compact power supply dip-switch settings.................................................................................181
8.0A.3 KB-10 settings.............................................................................................................................182
8.0A.4 KB-50 settings.............................................................................................................................182
8.0A.5 KR-7000P settings ......................................................................................................................183
8.0A.5.1 PRINT OUT settings.................................................................................................................183
8.0A.5.2 ONLINE settings ......................................................................................................................183
8.0A.5.3 PLACIDO settings ....................................................................................................................184
8.0A.5.4 Dipswitch settings .....................................................................................................................184
8.0A.5.5 Print wheel settings ...................................................................................................................185
8.0A.6 Colormap 32 (KRP32) settings .................................................................................................186
8.0A.6.1 KR COM Port setting................................................................................................................186
8.0A.6.2 Setup calibration data................................................................................................................186
8.0A.6.3 CV COM port setting................................................................................................................187
8.0A.6.4 Select instrument.......................................................................................................................187

8.1 KR-8100P  COLORMAP 32  CV-5000PC…………………………………………………188


8.1.1 CV-5000PC  Compact Power Supply……………………………………...See Chapter 2.3
8.1.2 Cable connection...........................................................................................................................188
8.1.3 Compact power supply dip-switch settings ...................................................................................188
8.1.4 CV-5000PC settings......................................................................................................................189
8.1.4.1 Setup CV-5000PC........................................................................................................................189
8.1.4.2 Installation check .........................................................................................................................189
8.1.4.3 Change RM (AR) Connect...........................................................................................................190
8.1.5 KR-8100P settings ........................................................................................................................191
8.1.5.1 ON-LINE settings ........................................................................................................................191
8.1.5.2 Output data setting .......................................................................................................................191
8.1.5.3 PLACIDO settings .......................................................................................................................192
8.1.6 Colormap 32 (KRP32) settings ...................................................................................................192
8.1.6.1 KR COM Port setting...................................................................................................................192
8.1.6.2 Setup calibration data...................................................................................................................193
8.1.6.3 Select instrument..........................................................................................................................193

8.2 KR-8100P  COLORMAP 32  CV-5000  I-BASE……………………………………194


8.2.1 Purpose of setup ...........................................................................................................................194
8.2.2 Cable connection...........................................................................................................................195
8.2.3 Compact power supply dip-switch settings ...................................................................................196
8.2.3.1 Dip-switch settings for cable using Colormap 32 software .........................................................196
8.2.3.2 Dip-switch settings for cable using i-base software.....................................................................196
8.2.3.3 Dip-switch settings bank 1...........................................................................................................196
8.2.4 KB-10 settings ...............................................................................................................................197
8.2.5 KB-50 settings ...............................................................................................................................198
8.2.6 KR-8100P settings ........................................................................................................................199
8.2.6.1 ON-LINE settings ........................................................................................................................199
8.2.6.2 Output data setting .......................................................................................................................199
8.2.6.3 PLACIDO settings .......................................................................................................................200
8.2.7 Colormap 32 (KRP32) settings ...................................................................................................201
8.2.7.1 KR COM Port setting...................................................................................................................201
8.2.7.2 Setup calibration data...................................................................................................................201
8.2.7.3 CV COM port setting...................................................................................................................202
8.2.7.4 Select instrument..........................................................................................................................202
8.2.8 i-base settings for CV-5000 (Compact power supply connection) ...............................................203
14
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.8.1 Plug-in installation .......................................................................................................................203


8.2.8.2 Serial port setup ...........................................................................................................................205
8.2.8.3 Advanced settings ........................................................................................................................206
8.2.9 i-base settings for Colormap 32 (KRP32) connection.................................................................208
8.2.9.1 Plug-in installation .......................................................................................................................208
8.2.9.2 KR8000P setup ............................................................................................................................210
8.2.10 Decimal symbol setting in Windows for i-base............................................................................212

8.3 KR-8100P  COLORMAP 32  I-BASE  CV-5000……………………………………214


8.3.1 Purpose of setup ...........................................................................................................................214
8.3.2 Cable connection...........................................................................................................................215
8.3.3 Compact power supply dip-switch settings ...................................................................................216
8.3.4 KB-10 settings ...............................................................................................................................217
8.3.5 KB-50 settings ...............................................................................................................................217
8.3.6 KR-8100P settings ........................................................................................................................218
8.3.6.1 ON-LINE settings ........................................................................................................................218
8.3.6.2 Output data setting .......................................................................................................................218
8.3.6.3 PLACIDO settings .......................................................................................................................219
8.3.7 Colormap 32 (KRP32) settings ...................................................................................................219
8.3.7.1 KR COM Port setting...................................................................................................................219
8.3.7.2 Setup calibration data...................................................................................................................220
8.3.7.3 Select instrument..........................................................................................................................220
8.3.8 i-base settings for CV-5000 (Compact power supply connection) ...............................................221
8.3.8.1 Plug-in installation .......................................................................................................................221
8.3.8.2 Serial port setup ...........................................................................................................................223
8.3.8.3 Advanced settings ........................................................................................................................224
8.3.9 i-base settings for Colormap 32 (KRP32) connection.................................................................226
8.3.9.1 Plug-in installation .......................................................................................................................226
8.3.9.2 KR8000P setup ............................................................................................................................228
8.3.10 Decimal symbol setting in Windows for i-base............................................................................230

8.4 KR-8100P  COLORMAP 32  CV-5000PC  I-BASE………………………………..232


8.4.1 Purpose of setup ...........................................................................................................................232
8.4.2 CV-5000PC  Compact Power Supply………………………………….…..See Chapter 2.3
8.4.3 Cable connection...........................................................................................................................233
8.4.4 CV-5000PC settings......................................................................................................................234
8.4.4.1 Setup CV-5000PC........................................................................................................................234
8.4.4.2 Installation check .........................................................................................................................234
8.4.4.3 Change RM (AR) Connect...........................................................................................................235
8.4.4.4 Import/Export setup .....................................................................................................................236
8.4.4.5 File Format Version .....................................................................................................................236
8.4.4.6 RS-232C Baud Rate.....................................................................................................................236
8.4.5 KR-8100P settings ........................................................................................................................237
8.4.5.1 ON-LINE settings ........................................................................................................................237
8.4.5.2 Output data setting .......................................................................................................................237
8.4.5.3 PLACIDO settings .......................................................................................................................238
8.4.6 Colormap 32 (KRP32) settings ...................................................................................................238
8.4.6.1 KR COM Port setting...................................................................................................................238
8.4.6.2 Setup calibration data...................................................................................................................239
8.4.6.3 Select instrument..........................................................................................................................239
8.4.7 i-base settings for Colormap 32 (KRP32) connection.................................................................240
8.4.7.1 Plug-in installation .......................................................................................................................240
8.4.7.2 KR8000P setup ............................................................................................................................242
8.4.8 i-base settings for CV-5000PC connection ..................................................................................244
8.4.8.1 Plug-in installation .......................................................................................................................244
8.4.8.2 CV Win setup...............................................................................................................................246
8.4.9 Decimal symbol setting in Windows for i-base..............................................................................247

15
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.0 Software and Firmware versions

Check the dealer support site for the latest released firmware for related Topcon instruments.
It is very important that ALL components of the CV-5000 installations are upgraded to the latest
firmware versions available, as compatibility between the components is only guaranteed after
the upgrade with the latest firmware.

Note:
Software versions of the: Compact power supply unit
Phoropter head
KB-50 Software
KB-10 Software
CV-5000PC Software
Can be displayed on the KB-50, KB-10 and CV-5000PC

Minimum Software versions requirements for this manual are:

Compact power supply 1.16.50


Phoropter head 2.11
KB-50 Software 1.21.14
KB-10 Software 5.11
CV-5000PC Software 1.21.14

1.0.1 Checking software versions on the KB-50

(Power up the Power


supply unit), Wait till
refraction screen is seen.
Hold the shift button down
while pressing the menu
button. Press on the touch
screen in the right corner
on the Topcon logo.

All installed software


versions are now
displayed.

note:
To exit this screen press
on the touch screen on
the blue KB-50 logo.

16
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.0.2 Checking software versions on the KB-10

(Power up the Power


supply unit), Wait till
refraction screen is seen.
Press in the upper left
corner of the KB-10 on the
Yellow button.

For a moment all installed


software versions are now
displayed.

1.0.3 Checking software versions on the CV-5000PC

Start up the CV-5000PC


software. Click on Help and
then on version information.

All installed software


versions are now displayed
on a popup screen.

note:
To exit this popup
screen click on the
popup screen itself. (for
example on the
CV-5000PC logo)

17
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.1 Cable descriptions


All cables mentioned in this manual are described with Topcon item numbers or part numbers.

1.1.1 Cables needed for the CV-5000(PC) stand alone setup

Connection Topcon cable item Packaged


(spare part item)

Power supply unit  Phoropter head 8 meter 4322157800 In box CPS

3 meter 4322170620 In box KB-50


Power supply unit  KB-50 Dial controller
5 meter 1830115* order Topcon

3 meter 4321986100 Attached KB-10

Power supply unit  KB-10 Dial controller Extension cable


3 meter 1830110 order Topcon
(Not to be used for KB-50)

Power supply unit KB port  PC


With use of CV-5000 PC software Curled cable 432150003 order Topcon
Instead off the use off KB-10 or KB-50

*)
IMPORTANT NOTE:

When using the 5 meter replacement


cable (item 1830115) you have to mount
the Ferrite Coils from the original KB-50
cable on to this replacement cable. This
to prevent malfunction of the KB-50.

Any cable longer than 5 meters


between Compact power supply
and KB-50 will give power
problems and results in Green,
Black or White screens on the KB-
50.

18
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.1.2 Cables needed for the CV-5000 Lite stand alone setup

Connection Topcon cable item Packaged


(spare part item)

Power supply unit  Phoropter head 8 meter 4322157800 In box CPS

3 meter 4321986100 Attached KB-10

Power supply unit  KB-10 Dial controller Extension cable


3 meter 1830110 order Topcon
(Not to be used for KB-50)

19
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.1.3 Cable item 1830057 Length 5 meter

For connecting CL/EZ/KR/RM/CV/ACP-8/MC-3 to CV


CL/EZ to KR/RM

D-sub 8 PIN male hooked D-sub 8 PIN male hooked


Instrument side Instrument side

Instrument side Instrument side

PIN layout solders side:

20
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.1.4 Cable item 1830008 Length 5 meter

For connecting CL/EZ/KR/RM/CV/TRK to PC


CC to CV

D-sub 9 PIN female D-sub 8 PIN male hooked


PC side Instrument side

PC side Instrument side

PIN layout solders side:

21
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.1.5 Cable item 1830095 Length 0,6 meter

For extension from D-sub 9 PIN female to PC USB

D-sub 9 PIN male USB

This cable is usually used to extend cable 1830008 to convert a D-sub 9 PIN female to USB.
With this cable Topcon instruments can also be connected to the PC USB port.

1.2 Cable length and cable installation

1.2.1 Cable length

All cables used between Topcon peripheral instruments are guaranteed to a maximum length of 5
meters (16,40 feet), except for the CV-power supply cable with 8 meter. If it is necessary to use
cables longer then 5 meters (16,40 feet) you have to use line drivers to overcome the problem of
to much resistance in the cable due to the long length.
Use always original Topcon cables.

1.2.2 Cable installation

When installing cables always be sure that you give the cable enough length for movement,
and not install the cables to tight. Never install cables to close to heat sources and close to
transformers, like light transformers on the ceiling. Keep all cables away from electromagnetic fields.
Follow the instructions of the manuals of Topcon instruments.

IMPORTANT NOTE:

When using the 5 meter replacement cable


(item 1830115) you have to mount the Ferrite
Coils from the original KB-50 cable on to this
replacement cable. This to prevent malfunction
of the KB-50.

22
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.3 Interchangeability of CV-3000 and CV-5000 components


Below the overview of interchangeability between CV-5000 and CV-3000 components.

From the graph you can see that CV-3000 components are NOT interchangeable with CV-5000
components. The reason for that is the difference in Voltage. 5V for the CV-3000 and 12V for the
CV-5000 !

Phoropter controller overview

Phoropter head KB-50 KB-10 CV-5000PC

CV-5000(XP) V V V
CV-5000 Light X V X

23
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.4 CV-5000 Compact Power Supply description

1) Printer Unit
2) Port for connecting KB-50 or KB-10 Dial controllers
3) ON/OFF Power Switch
4) Dip Switch selector lid
5) USB 1
6) USB 2
7) Comport (COM 1)
8) CL port (COM 2)
9) RM port (COM 3)
10) MC (serial) port (COM 4)
11) ACP/MC (infrared ray communication)port
12) Port for connecting CV-5000 (Lite) phoropterhead
13) Fuse holder
14) 3P Inlet

24
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.5 Overview dip-switch settings Compact Power Supply

DIP-Switch bank 1(DSW1)


1:OFF
2:OFF OLD format
3:OFF
1:ON
2:OFF NEW format
1 3:OFF
Setting RS232C
2 format
3 1:OFF
2:ON STD1 format
3:OFF
1:ON
2:ON Not in use
3:OFF
4:OFF Baud rate 2400
4 Baud rate !
4:ON Baud rate 9600
5:OFF
Canal 1 ACP/MC/CC-100 series DIP-Switch
6:OFF
5:ON
Setting Canal 2 ACP/MC/CC-100 series DIP-Switch
5 ACP/MC/CC-100(P) 6:OFF
6 Infra-red signal 5:OFF
communication Canal 3 ACP/MC/CC-100 series DIP-Switch
6:ON
5:ON
Canal 4 ACP/MC/CC-100 series DIP-Switch
6:ON

7 Unavailable

8 Unavailable

25
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Note: Always reboot the power supply after changing dip-switch settings

DIP-Switch bank 2(DSW2)


1:OFF
Connect to RM / KR / CL
2:OFF
1
1:ON Connect to PC
2:OFF (Transfer data communication)
Setting COM 1 port
1:OFF Connect to MC / ACP / CC100 series
2:ON with serial communication
2
1:ON Connect to several instruments
(When connect to a second compact power
2:ON supply or more)
3:OFF
Connect to RM / KR / CL
4:OFF
3
3:ON Connect to PC
4:OFF (Transfer data communication)
Setting COM 2 port
3:OFF Connect to MC / ACP /CC100 series
4:ON with serial communication
4
3:ON Connect to several instruments
(When connect to a second compact power
4:ON supply or more)
5:OFF
Connect to RM / KR / CL
6:OFF
5
5:ON Connect to PC
6:OFF (Transfer data communication)
Setting COM 3 port
5:OFF Connect to MC / ACP /CC100 series
6:ON with serial communication
6
5:ON Connect to several instruments
(When connect to a second compact power
6:ON supply or more)
7:OFF
Connect to RM / KR / CL
8:OFF
7
7:ON Connect to PC
8:OFF (Transfer data communication)
Setting COM 4 port
7:OFF Connect to MC / ACP / CC100 series
8:ON with serial communication
8
7:ON Connect to several instruments
(When connect to a second compact power
8:ON supply or more)

26
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

1.6 Overview dip-switch settings KB-10

Dip-switch bank 1 (DSW1)


1: OFF Jackson Cross
1 Settings CC/CT 1: ON TOPCON Cross
2 Unused
3 Unused
4 Unused for European market
5 Unused for European market
6 Unused
7 Unused
8: OFF Demo mode is prohibited
8 Setting Demo mode 8: ON Demo mode is permitted

Dip-switch bank 2 (DSW2)


1: OFF
2: OFF Not specified (Invalid)
1
1: ON
2: OFF Not specified (Invalid)
Setting the output format for PC
1: OFF
2: ON STD 1 Format
2 1: ON
2: ON Not specified (Invalid)
3: OFF Data output is prohibited
3 Setting data output function for PC 3: ON Data output is permitted
4: OFF 2400
4 Setting baud rate for data to PC 4: ON 9600
5: OFF All the data blocks are outputted
5 Setting type off data output to PC 5: ON Only the stored data is outputted
6: OFF Printing ON
6 Setting ON /OFF for printing 6: ON Printing OFF
7: OFF Sending times: 4
7 Setting the data sending times for PC-2 7: ON Sending times: 6
Setting the type of 8: OFF Wireless type
8 MC-3 / ACP / CC-100 series 8: ON Serial type

27
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

2.0 Phoropter  Compact Power Supply

Cable 8 meter
item 432210091

CV-5000 and CV-5000 Lite Compact power supply unit

2.0.1 Cable connection

Connect the cable (item 4321455400) from the phoropter head to connection number 12 off the
compact power supply unit. (see chapter 1.4 power supply unit description)

2.0.2 Compact power supply dip-switch settings

No special dip-switch settings needed for communication with the CV-5000 (Lite) phoropter head.

28
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

2.1 KB-50  Compact Power Supply

Cable 3 meter
item 4322170620

Cable 5 meter
item 1830115

KB-50 Dial controller Back side


or
Front side

Compact power supply unit

2.1.1 Cable connection

Connect the cable on connection number 2 on the back or the front side of the power supply unit.
Choose the 3 meter cable (item 4322170620) or the optional 5 meter cable (item 1830115, order
Topcon) (see chapter 1.4 power supply unit description).

2.1.2 KB-50 settings

No dip-switch settings available. All settings are made within the software off the KB-50
No special settings are necessary for communication with CV-5000 power supply unit.

2.1.3 Compact power supply dip-switch settings

No special dip-switch settings necessary for communication with the KB-50 Dial controller.

29
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

2.2 KB-10  Compact Power Supply

Cable attached to KB-10


3 meter
Item 4321986100

3 meter Extension cable


3 meter
Item 1830110
Back side
or
KB-10 Dial controller Front side

Note:
Used in combination with the CV-5000 phoropter,
NOT with the CV-5000 LITE phoropter version.

Compact power supply unit

2.2.1 Cable connection

Connect the attached cable of the KB-10 to connection number 2 on the back or the front side of
the power supply unit. (see chapter 1.4 power supply unit description). Because the cable is
attached to the KB-10 you only can extend the cable with 3 meter cable item 1830110 (option,
order Topcon)

2.2.2 KB-10 settings

No special (dip-switch) settings are necessary for communication with CV-5000 power supply
unit.

2.2.3 Compact power supply dip-switch settings

No special dip-switch settings necessary for communication with the KB-10 Dial controller.

30
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

2.3 CV-5000PC  Compact Power Supply

2.3.1 Purpose of setup

This connection is needed for the use off the CV-5000PC software as a controller on a personal
computer, instead of the use of the KB-10 or KB-50 controller.

This only works with the CV-5000 and not with the CV-5000 Lite version !

Personal Computer (Topcon Shuttle) Compact power supply (back side)

Option 2) item 1830095 + 432150003 (curled cable)

Option 1) item 432150003 (curled cable)

Note: You can also use the KB-port on the front of the Power supply unit.

2.3.2 Cable connection

Option 1) Connect the curled cable (item 432150003) from connection nr 2 KB-50 port (see
chapter 1.4 power supply unit description) on the back or the front side of the compact power
supply unit to a serial port of the Personal Computer.

Option 2) Connect the curled cable (item 432150003) on connection nr. 2 (see chapter 1.4 power
supply unit description) on the back or the front side of the compact power supply unit, and extend
this cable with a Serial to USB converter cable (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper serial to USB converter driver on the PC)

2.3.3 Compact power supply dip-switch settings

No special (dip-switch) settings are necessary for communication with CV-5000 PC software

31
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

2.3.4 CV-5000PC software settings (COMM port)

After installation of the CV-5000PC software on the personal computer and the connection of all
cables, you can run the software by clicking on the CV-5000PC icon on your desktop.
After clicking on the icon and run the program for the first time the window
“Comm. Port No. for CV is not selected.” Pops up. Click on yes.

The following window will appear :

In this example COMM


port 1 is selected.
Find out in Windows
Device Manager which
COMM port on the PC
is used for the Power
supply and select the
appropriate COMM port
in this menu.

Select the right Comm port on which you connected the curled cable to the PC.

NOTE:
After the initial installation you can always select another COMM port if necessary by going to
“Tools” and next to “Settings”.

32
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.0 ACP-8 MC  CV-5000 RS-232C Cable connection

Cable 5 meter
Item 1830057

ACP-8 MC Compact power supply unit


3.0.1 Cable connection

Connect the cable (item 1830057, order Topcon) from ACP-8 MC to the Compact power supply
unit on COMM port 1,2,3 or 4 (see chapter 1.4 power supply unit description).

3.0.2 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : ON

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : ON

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : ON

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : ON

3.0.3 KB-10 settings

3.0.3.1 Dip-switch setting

Dip-switch bank 2 (DSW2) 8 : ON (serial type)

33
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.0.3.2 Chart (device) setting

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select
with the dial “SETTING” and push
the SET button. Go with the F2
button to page P.5 Now select with
the dial the appropriate chart
(device), in this case ACP-8.
Confirm with the SET button. Now
select with the dial the appropriate
Chart Pattern used by the ACP-8.
Confirm OK by pressing the F4
button, then press SET to restart the
KB-10.

Dial and SET button

3.0.3.3 Chart Pattern setting

Follow the steps described above in


3.0.3.2

(selection in picture is example)

34
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.0.4 KB-50 settings

3.0.4.1 Chart Connect setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
Connect, “CPS On Line”

3.0.4.2 Chart (Far) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart (Far)
the appropriate chart used by the
ACP-8.

(selection in picture is example)

3.0.4.3 Chart (Near) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
(Near) the appropriate chart used for
near vision tests.

(selection in picture is example)

35
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.0.5 CV-5000PC settings

3.0.5.1 Chart Connect setting

(Startup CV-5000PC software),


select Tools, then select Setting…
select chart / connect in the menu.
Now select Chart Connect, “CPS On
Line”

Note: COMM 1 to COMM 8 are not


in use.

CV Connect:
See chapter 2.3.4

3.0.5.2 Chart (Far) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Far) the appropriate
chart used by the ACP-8.

(selection in picture is example)

3.0.5.3 Chart (Near) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Near) the appropriate
chart used for near vision tests.

(selection in picture is example)

36
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.1 ACP-8 E,ACP-8 EM  CV-5000 Infra-red Light Emitter

Infra-
red
light

ACP-8 E, ACP-8 EM Light Emitter Compact power supply unit


item 424120005
3.1.1 Light Emitter connection
Connect the 3 PIN connector of the light emitter (item 424120005, order Topcon) to the Compact
power supply unit on connection 11 (see chapter 1.4 power supply unit description).

3.1.2 Compact power supply dip-switch settings (infra-red channel)


Must be set accordingly to the infra-red channel used by the ACP-8 device !
Using Infra-red channel 1 Dip-switch bank 1 (DSW1) 5 : OFF 6 : OFF
Using Infra-red channel 2 Dip-switch bank 1 (DSW1) 5 : ON 6 : OFF
Using Infra-red channel 3 Dip-switch bank 1 (DSW1) 5 : OFF 6 : ON
Using Infra-red channel 4 Dip-switch bank 1 (DSW1) 5 : ON 6 : ON

3.1.3 Infra-red channel table for the ACP-8 Remote Controller


channel ACP-8 Controller Remove the battery cover.
S1 S2
1 H H
2 L H
3 H L
4 L L

This remote controller only


works with ACP-8 E and
ACP-8 EM
In this picture channel 1 is selected. (example)

3.1.4 KB-10 settings


3.1.4.1 Dip-switch setting
Dip-switch bank 2 (DSW2) 8 : OFF (wireless type)

37
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.1.4.2 Chart (device) setting

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select
with the dial “SETTING” and push
the SET button. Go with the F2
button to page P.5 Now select with
the dial the appropriate chart
(device), in this case ACP-8.
Confirm with the SET button. Now
select with the dial the appropriate
Chart Pattern used by the ACP-8.
Confirm OK by pressing the F4
button, then press SET to restart the
KB-10.

Dial and SET button

3.1.4.3 Chart Pattern setting

Follow the steps described above in


3.1.4.2

(selection in picture is example)

38
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.1.5 KB-50 settings

3.1.5.1 Chart Connect setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
Connect, “Off Line (infrared)”

3.1.5.2 Chart (Far) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart (Far)
the appropriate chart used by the
ACP-8.

(selection in picture is example)

3.1.5.3 Chart (Near) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
(Near) the appropriate chart used for
near vision tests.

(selection in picture is example)

39
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.1.6 CV-5000PC settings

3.1.6.1 Chart Connect setting

(Startup CV-5000PC software),


select Tools, then select Setting…
select chart / connect in the menu.
Now select Chart Connect,
“Off Line (Infrared)”

Note: COMM 1 to COMM 8 are not


in use.

CV Connect:
See chapter 2.3.4

3.1.6.2 Chart (Far) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Far) the appropriate
chart used by the
ACP-8.

(selection in picture is example)

3.1.6.3 Chart (Near) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Near) the appropriate
chart used for near vision tests.

(selection in picture is example)

40
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.2 MC-3  CV-5000 RS-232C Cable connection

Cable 5 meter
item 1830057

MC-3 Compact power supply unit

3.2.1 Cable connection

Connect the cable (item 1830057, order Topcon) from MC-3 to the Compact power supply unit on
COMM port 1,2,3 or 4 (see chapter 1.4 power supply unit description).

3.2.2 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : ON

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : ON

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : ON

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : ON

3.2.3 KB-10 settings

3.2.3.1 Dip-switch setting

Dip-switch bank 2 (DSW2) 8 : ON (serial type)

41
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.2.3.2 Chart (device) setting

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select
with the dial “SETTING” and push
the SET button. Go with the F2
button to page P.5 Now select with
the dial the appropriate chart
(device), in this case MC-3.
Confirm with the SET button. Now
select with the dial the appropriate
Chart Pattern used by the MC-3.
Confirm OK by pressing the F4
button, then press SET to restart the
KB-10.

Dial and SET button

3.2.3.3 Chart Pattern setting

Follow the steps described above in


3.2.3.2

(selection in picture is example)

42
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.2.4 KB-50 settings

3.2.4.1 Chart Connect setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
Connect, “CPS On Line”

3.2.4.2 Chart (Far) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart (Far)
the appropriate chart used by the
MC-3.

(selection in picture is example)

3.2.4.3 Chart (Near) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart (Near)
the appropriate chart used for near
vision tests.

(selection in picture is example)

43
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.2.5 CV-5000PC settings

3.2.5.1 Chart Connect setting

(Startup CV-5000PC software),


select Tools, then select Settings…
select chart / connect in the menu.
Now select Chart Connect, “CPS On
Line”

Note: COMM 1 to COMM 8 are not


in use.

CV Connect:
See chapter 2.3.4

3.2.5.2 Chart (Far) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Far) the appropriate
chart used by the MC-3.

(selection in picture is example)

3.2.5.3 Chart (Near) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Near) the appropriate
chart used for near vision tests.

(selection in picture is example)

44
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.2.6 MC-3 distance and Alignment emitter distance settings


Alignment emitter MC-3 Body
Set distance Slide switch S3 3 4
0,70 meter N ON OFF
0,90 meter N OFF ON
1,10 meter F OFF OFF
1,65 meter (infinite) F ON ON

Alignment emitter MC-3 Body

3.2.7 MC-3 Baud rate dip-switch settings


ON 9600
8 Setting baud rate for RS-232C
OFF 2400 (standard)

LEFT = OFF
RIGHT = ON

Note:

The baud rate must set accordingly to the baud


rate selection of the Compact power supply unit.
(see chapter 2.4)

45
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.3 MC-3  CV-5000 Infra-red Light Emitter

Infra-red
light

MC-3 Light Emitter Compact power supply unit


Item 424120005

3.3.1 Light Emitter connection

Connect the 3 PIN connector of the light emitter (item 424120005, order Topcon) to the Compact
power supply unit on connection 11 (see chapter 1.4 power supply unit description).

3.3.2 Compact power supply dip-switch settings (infra-red channel)

Using Infra-red channel 1 Dip-switch bank 1 (DSW1) 5 : OFF 6 : OFF

Using Infra-red channel 2 Dip-switch bank 1 (DSW1) 5 : ON 6 : OFF

Using Infra-red channel 3 Dip-switch bank 1 (DSW1) 5 : OFF 6 : ON

Using Infra-red channel 4 Dip-switch bank 1 (DSW1) 5 : ON 6 : ON

3.3.3 KB-10 settings

3.3.3.1 Dip-switch setting

Dip-switch bank 2 (DSW2) 8 : OFF (wireless type)

46
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.3.3.2 Chart (device) setting

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select
with the dial “SETTING” and push
the SET button. Go with the F2
button to page P.5 Now select with
the dial the appropriate chart
(device), in this case MC-3.
Confirm with the SET button. Now
select with the dial the appropriate
Chart Pattern used by the MC-3.
Confirm OK by pressing the F4
button, then press SET to restart the
KB-10.

Dial and SET button

3.3.3.3 Chart Pattern setting

Follow the steps described above in


3.3.3.2

(selection in picture is example)

47
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.3.4 KB-50 settings

3.3.4.1 Chart Connect setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
Connect, “Off Line (infrared)”

3.3.4.2 Chart (Far) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart (Far)
the appropriate chart used by the
MC-3.

(selection in picture is example)

3.3.4.3 Chart (Near) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
(Near) the appropriate chart used for
near vision tests.

(selection in picture is example)

48
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.3.5 CV-5000PC settings

3.3.5.1 Chart Connect setting

(Startup CV-5000PC software),


select Tools, then select Settings…
select chart / connect in the menu.
Now select Chart Connect, “Off Line
(Infrared)”

Note: COMM 1 to COMM 8 are not


in use.

CV Connect:
See chapter 2.3.4

3.3.5.2 Chart (Far) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Far) the appropriate
chart used by the MC-3.

(selection in picture is example)

3.3.5.3 Chart (Near) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Near) the appropriate
chart used for near vision tests.

(selection in picture is example)

49
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.3.6 MC-3 Distance and Alignment Emitter distance settings


Set distance Alignment emitter MC-3 Body
Slide switch S3 3 4
0,70 meter N ON OFF
0,90 meter N OFF ON
1,10 meter (standard) F OFF OFF
1,65 meter (infinite) F ON ON

Alignment emitter MC-3 Body

3.3.7 MC-3 Other dip-switch settings (for LR-1)


ON Permitted
Setting Infra-red both-direction communication
7 (for the use of the LR-1 controller)
OFF Prohibited
(standard)

LEFT = OFF
RIGHT = ON

50
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.3.8 Infra-red channel table for remote controllers, alignment emitter and MC-3
channel MC-3 Controller LR-1 Controller Alignment emitter MC-3 Body
S1 S2 S1 S2 S1 S2 5 6
1 H H OFF OFF H H OFF OFF
2 L H OFF ON L H ON OFF
3 H L ON OFF H L OFF ON
4 L L ON ON L L ON ON

Channel 1 is standard

Alignment emitter
MC-3 Controller

MC-3 Body

LEFT = OFF
RIGHT = ON

Must be set accordingly to


light emitter channel setting of
the Compact power supply unit
(see chapter 3.8.2)

51
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.4 CC-100 Series  CV-5000 RS-232C Cable connection

Cable 5 meter
Item 1830008

CC-100 Series Compact power supply unit


3.4.1 Cable connection

Connect the cable (item 1830008, order Topcon) from CC-100 series to the Compact power
supply unit on COMM port 1,2,3 or 4 (see chapter 1.4 power supply unit description).

CC-100 Series D-sub 9 pin for RS-232C

3.4.2 Compact power supply dip-switch settings


Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : ON
Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : ON
Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : ON
Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : ON

3.4.3 KB-10 settings

3.4.3.1 Dip-switch setting

Dip-switch bank 2 (DSW2) 8 : ON (serial type)

52
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.4.3.2 Chart (device) setting Note: CC-100 series selection option is not
available in KB-10. Select the ACP-8 mode! The
(Power up the Power supply unit), CC-100 Series will behave like a ACP-8 device. CC-
Wait until refraction screen is seen. 100 Series must set accordingly!
Hold the SFT button down while
(see chapter 3.4.9.5)
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.5 Now select with the dial the
appropriate chart (device), in this
case ACP-8. Confirm with the SET
button. Now select with the dial the
appropriate Chart Pattern you want
to use. Confirm OK by pressing the F4
button, then press SET to restart the
KB-10.

Dial and SET button

3.4.3.3 Chart Pattern setting

Follow the steps described above in


3.4.3.2

(selection in picture is example)

53
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

NOTE: First install the CC-100 Series Chart DLL files (See chapter 3.4.6.5)

3.4.4 KB-50 settings

3.4.4.1 Chart Connect setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
Connect, “CPS On Line”

3.4.4.2 Chart (Far) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart (Far)
the CC-100P (RC), the CC-100 (RC)
mode or CC-100XP (RC) mode.

Note: you can also select the ACP-8


mode, if you want the CC-100 Series
device behaving like a ACP-8 device.
You have to set the CC-100 Series
device accordingly!
(selection in picture is example)

3.4.4.3 Chart (Near) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
(Near) the appropriate chart used for
near vision tests.

(selection in picture is example)

54
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

NOTE: First install the CC-100 Series Chart DLL files (See chapter 3.4.6.4)
3.4.5 CV-5000PC settings

3.4.5.1 Chart Connect setting

(Startup CV-5000PC software),


select Tools, then select Setting…
select chart / connect in the menu.
Now select Chart Connect, “CPS On
Line”

Note: COMM 1 to COMM 8 are not


in use.

CV Connect:
See chapter 2.3.4

3.4.5.2 Chart (Far) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Far) the CC-100P (RC),
the CC-100 (RC) mode or CC-100XP
(RC) mode.
Note: you can also select the ACP-8
mode, if you want the CC-100 Series
device behaving like a ACP-8 device.
You have to set the CC-100 Series
device accordingly!

(selection in picture is example)

3.4.5.3 Chart (Near) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Near) the appropriate
chart used for near vision tests.

(selection in picture is example)

55
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.4.6 CC-100 Series Chart DLL installation in KB-50 & CV-5000PC

3.4.6.1 Purpose

In order to operate the CC-100 Series by KB-50 or CV-5000PC dedicated DLL

! files need to be installed into KB-50 or CV-5000PC.

As of half 2009, these dedicated DLL files will be standard installed on KB-50 and
CV-5000PC software. (and manually installation will not be necessary anymore)

3.4.6.2 Software conditions


KB-50 from version 1.15 and up CPS from version 1.16.50
CV-5000PC from version 1.15 and up
RC.DLL from version 3.0 and up
CC-100(P) from version 2.2.9 and up
CC-100XP from version 3.2.0 and up

3.4.6.3 Location Chart DLL files

All CC-100 series DLL files and Emulation DLL files needed for installing are available on our
Topcon FTP site.

3.4.6.4 Installation of Chart DLL files in CV-5000PC

1) Copy the respective RC.DLL files into the following folder:


C:\Program Files\TOPCON\CV-5000PC\ImageDLL\Far

2) Copy the “Emulation” folder containing the respective


RC_Emulation.DLL files into the following folder:
C:\Program Files\TOPCON\CV-5000PC\ImageDLL\Far

3.4.6.5 Installation of Chart DLL files in KB-50

1) Copy the respective RC.DLL files onto a USB memory stick If there is not enough
2) Start KB-50 by power up the Compact power supply unit space to install the
3) Wait to the normal refraction screen is seen respective RC-DLL files
4) Remove the cover on the back of the KB-50 in to the KB-50 first
5) Insert the memory stick in one of the USB ports
delete other DLL files
6) Hold button down while pressing button that you do not use!!
7) Press on touch screen YES and wait
8) Found New Hardware Wizard pops up Press CANCEL For example delete
9) Press on touch screen START KB-50 some MC-3 DLL files
10) Press on touch screen EXPLORER out of the folder “Far”
(see nr 11).
Step 11) till 21) see next Page

56
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

11) Open folder Far by pressing KB50EXE (R:), 17) Press on Dismount
press on ImageDLL, press on Far

18) Press on Commit File

12) Press on the location of the memory stick


(in this example Removable Disk (D:) )
13) Copy by dragging the respective DLL files
(CC100_RC.dll or CC100P_RC.dll or
CC100XP_RC.dll) into the folder
KB50EXE (R:) \ ImageDLL \ Far
14) Copy by dragging the “Emulation” folder
containing the respective
RC_Emulation.DLL files into the folder 19) Press on OK (when CommitFile Success)
KB50EXE (R:) \ ImageDLL \ Far when not success start from 17) again.

20) Press on Quit


15) Press on folder SDI

21) Reboot KB-50 By Power up the


16) Press on EWFTool.exe Compact power supply unit with the
ON/OFF switch.

57
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.4.7 Placing cover sheet CC-100 Series on remote controller

Place the CC-100 Series cover sheet on the


remote controller. (used by the ACP-8).
(Cover sheet enclosed in box CC-100 Series)

Note: Order the remote controller with


Masking possibilities !
(item: 424143000A)

3.4.8 Infra-red channel table for the Remote Controller


channel CC-100 Series / Remove the battery cover.
ACP-8 Controller
S1 S2
1 H H
2 L H
3 H L
4 L L

Make sure to use a different


channel then the channel
used by the light emitter In this picture channel 1 is selected. (example)

58
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.4.9 CC-100 Series settings

3.4.9.1 Menu Settings overview

CC-100(P)(XP) Chart A, Chart B1, Chart C1, Chart F

KB-50 KB-50 RC

Advice: Leave at least one channel (preferable channel 1) always in CC-100 Series
mode so you can always enter the settings menu of the CC-100 Series with
the IR remote controller.

Note: On each of the 4 IR channels you can make 1 of the combinations of the table
above. You can use 2 controllers at the same time. Example: IR remote controller
on IR channel 1 and KB-50 on IR channel 2.

59
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.4.9.2 Menu Settings selections (serial device menu & IR devices menu)

Mode CC-100 Series Functionality CC-100 Series Controller

CC-100 Series Full options of CC-100 Series IR remote controller with CC-100 Series
(not in serial device sheet
menu)

KB-10
Chart A, B1, C1, F CC-100 Series will have the IR remote controller with ACP-8 sheet**
functionality like ACP-8 KB-50 in ACP-8 mode**
CV-5000PC in ACP-8 mode**

KB-50 not in use -

KB-50 RC Full options of CC-100 Series KB-50 in CC-100 Series mode*


CV-5000PC in CC-100 Series mode*

OFF Channel not in use -

*) Dedicated DLL files must be installed ! (see chapter 3.4.4)


**) Not advised

3.4.9.3 Distance setting

Power up the CC-100 Series device with


the ON/OFF switch. Press on the menu
button of the remote controller. Press on
the arrow (down) button of the remote
controller and select Distance. Select the
right distance by using the left and right
arrow buttons of the infra-red remote
controller.

60
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.4.9.4 Setting serial (RS-232C) connection for use with KB-50 / CV-5000PC

Power up the CC-100 Series device with


the ON/OFF switch. Press on the menu
button of the remote controller. Press on
the arrow (down) button of the remote
controller and select Serial Device. Now
select the KB-50 RC mode with the
arrow buttons.

Note:
If you select Chart A, B1, C1 or F the
CC-100 Series will behave like a ACP-8
device. KB-50 or CV-5000PC must be set
also in ACP-8 mode.

3.4.9.5 Setting serial (RS-232C) connection for use with KB-10

Power up the CC-100 Series device with


the ON/OFF switch. Press on the menu
button of the remote controller. Press on
the arrow (down) button of the remote
controller and select Serial Device. Now
select Chart A, B1, C1 or F with the
arrow buttons.

Note:
CC-100 Series will behave like a ACP-8
device.

61
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.4.9.6 Setting CC-100 series for use with Remote Controller

Power up the CC-100 Series device with


the ON/OFF switch. Press on the menu
button of the remote controller. Press on
the arrow (down) button of the remote
controller and select I / R Devices. Press
on the arrow right button of the remote
controller.
Choose the Device (read channel)
accordingly with the infra-red remote
controller channel setting. (see chapter
3.4.8)
Now select CC-100, CC-100P or CC-
100XP

Advice:
To choose an infra-red channel, use
the letter combination. For example HH
or HL accordingly to the infra-red
remote controller channel setting.
Read: Device=Channel
(see chapter 3.5.8)

62
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.5 CC-100 Series  CV-5000 Infra-red Light Emitter

Infra-red
light

CC-100 Series Light Emitter Compact power supply unit


Item 424120005
3.5.1 Light Emitter connection

Connect the 3 PIN connector of the light emitter (item 424120005, order Topcon) to the Compact
power supply unit on connection 11 (see chapter 1.4 power supply unit description).

3.5.2 Compact power supply dip-switch settings (infra-red channel)

Using Infra-red channel 1 Dip-switch bank 1 (DSW1) 5 : OFF 6 : OFF

Using Infra-red channel 2 Dip-switch bank 1 (DSW1) 5 : ON 6 : OFF

Using Infra-red channel 3 Dip-switch bank 1 (DSW1) 5 : OFF 6 : ON

Using Infra-red channel 4 Dip-switch bank 1 (DSW1) 5 : ON 6 : ON

3.5.3 KB-10 settings

3.5.3.1 Dip-switch setting

Dip-switch bank 2 (DSW2) 8 : OFF (wireless type)

63
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.5.3.2 Chart (device) setting Note: CC-100 series selection option is not
available in KB-10. Select the ACP-8 mode! The
(Power up the Power supply unit), CC-100 Series will behave like a ACP-8 device. CC-
Wait until refraction screen is seen. 100 Series must set accordingly! (see chapter
Hold the SFT button down while
3.5.9.5)
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.5 Now select with the dial the
appropriate chart (device), in this
case ACP-8. Confirm with the SET
button. Now select with the dial the
appropriate Chart Pattern you want
to use. Confirm OK by pressing the F4
button, then press SET to restart the
KB-10.

Dial and SET button

3.5.3.3 Chart Pattern setting

Follow the steps described above in


3.5.3.2

(selection in picture is example)

64
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

NOTE: First install the CC-100 Series Chart DLL files (See chapter 3.5.6.5)

3.5.4 KB-50 settings

3.5.4.1 Chart Connect setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
Connect, “Off Line (infrared)”

3.5.4.2 Chart (Far) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart (Far)
the CC-100P (RC), the CC-100 (RC)
mode or CC-100XP (RC) mode.

Note: you can also select the ACP-8


mode, if you want the CC-100 Series
device behaving like a ACP-8 device.
You have to set the CC-100 Series
device accordingly!
(selection in picture is example)

3.5.4.3 Chart (Near) setting

(Power up the Power supply unit),


Wait till refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Chart in the menu. Select Chart
(Near) the appropriate chart used for
near vision tests.

(selection in picture is example)

65
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

NOTE: First install the CC-100 Series Chart DLL files (See chapter 3.5.6.4)
3.5.5 CV-5000PC settings

3.5.5.1 Chart Connect setting

(Startup CV-5000PC software),


select Tools, then select Setting…
select chart / connect in the menu.
Now select Chart Connect, “Off Line
(Infrared)”

Note: COMM 1 to COMM 8 are not


in use.

CV Connect:
See chapter 2.3.4

3.5.5.2 Chart (Far) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Far) the CC-100P (RC),
the CC-100 (RC) mode or CC-100XP
(RC) mode.
Note: you can also select the ACP-8
mode, if you want the CC-100 Series
device behaving like a ACP-8 device.
You have to set the CC-100 Series
device accordingly!

(selection in picture is example)

3.5.5.3 Chart (Near) setting

(Startup CV-5000PC software),


Select Tools, then select Setting…
select chart / connect in the menu.
Select Chart (Near) the appropriate
chart used for near vision tests.

(selection in picture is example)

66
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.5.6 CC-100 Series Chart DLL installation in KB-50 & CV-5000PC

3.5.6.1 Purpose

In order to operate the CC-100 Series by KB-50 or CV-5000PC dedicated DLL

! files need to be installed into KB-50 or CV-5000PC.

As of half 2009, these dedicated DLL files will be standard installed on KB-50 and
CV-5000PC software. (and manually installation will not be necessary anymore)

3.5.6.2 Software conditions


KB-50 from version 1.15 and up CPS from version 1.16.50
CV-5000PC from version 1.15 and up
RC.DLL from version 3.0 and up
CC-100(P) from version 2.2.9 and up
CC-100XP from version 3.2.0 and up

3.5.6.3 Location Chart DLL files

All CC-100 series DLL files and Emulation DLL files needed for installing are available on our
Topcon FTP site.

3.5.6.4 Installation of Chart DLL files in CV-5000PC

1) Copy the respective RC.DLL files into the following folder:


C:\Program Files\TOPCON\CV-5000PC\ImageDLL\Far

2) Copy the “Emulation” folder containing the respective


RC_Emulation.DLL files into the following folder:
C:\Program Files\TOPCON\CV-5000PC\ImageDLL\Far

3.5.6.5 Installation of Chart DLL files in KB-50

1) Copy the respective RC.DLL files onto a USB memory stick If there is not enough
2) Start KB-50 by power up the Compact power supply unit space to install the
3) Wait to the normal refraction screen is seen respective RC-DLL files
4) Remove the cover on the back of the KB-50 in to the KB-50 first
5) Insert the memory stick in one of the USB ports
delete other DLL files
6) Hold button down while pressing button that you do not use!!
7) Press on touch screen YES and wait
8) Found New Hardware Wizard pops up Press CANCEL For example delete
9) Press on touch screen START KB-50 some MC-3 DLL files
10) Press on touch screen EXPLORER out of the folder “Far”
(see nr 11).
Step 11) till 21) see next Page

67
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

11) Open folder Far by pressing KB50EXE (R:), 17) Press on Dismount
press on ImageDLL, press on Far

18) Press on Commit File

12) Press on the location of the memory stick


(in this example Removable Disk (D:) )
13) Copy by dragging the respective DLL files
(CC100_RC.dll or CC100P_RC.dll or
CC100XP_RC.dll) into the folder
KB50EXE (R:) \ ImageDLL \ Far
14) Copy by dragging the “Emulation” folder
containing the respective
RC_Emulation.DLL files into the folder 19) Press on OK (when CommitFile Success)
KB50EXE (R:) \ ImageDLL \ Far when not success start from 17) again.

20) Press on Quit


15) Press on folder SDI

21) Reboot KB-50 By Power up the


16) Press on EWFTool.exe Compact power supply unit with the
ON/OFF switch.

68
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.5.7 Placing cover sheet CC-100 Series on remote controller

Place the CC-100 Series cover sheet on the


remote controller. (used by the ACP-8).
(Cover sheet enclosed in box CC-100 Series)

Note: Order the remote controller with


Masking possibilities!
(item: 424143000A)

3.5.8 Infra-red channel table for the Remote Controller


channel CC-100 Series / Remove the battery cover.
ACP-8 Controller
S1 S2
1 H H
2 L H
3 H L
4 L L

Make sure to use a different


channel then the channel
used by the light emitter In this picture channel 1 is selected. (example)

69
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.5.9 CC-100 Series settings

3.5.9.1 Menu Settings overview

CC-100(P)(XP) Chart A, Chart B1, Chart C1, Chart F

KB-50 KB-50 RC

Advice: Leave at least one channel (preferable channel 1) always in CC-100 Series
mode so you can always enter the settings menu of the CC-100 Series with
the IR remote controller.

Note: On each of the 4 IR channels you can make 1 of the combinations of the table
above. You can use 2 controllers at the same time. Example: IR remote controller
on IR channel 1 and KB-50 on IR channel 2.

70
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.5.9.2 Menu Setting selections (serial device menu & IR devices menu)

Mode CC-100 Series Functionality CC-100 Series Controller

CC-100 Series Full options of CC-100 Series IR remote controller with CC-100 Series
(not in serial device sheet
menu)

KB-10
Chart A, B1, C1, F CC-100 Series will have the IR remote controller with ACP-8 sheet**
functionality like ACP-8 KB-50 in ACP-8 mode**
CV-5000PC in ACP-8 mode**

KB-50 not in use -

KB-50 RC Full options of CC-100 Series KB-50 in CC-100 Series mode*


CV-5000PC in CC-100 Series mode*

OFF Channel not in use -

*) Dedicated DLL files must be installed ! (see chapter 3.5.4)


**) Not advised

3.5.9.3 Distance setting

Power up the CC-100 Series device with


the ON/OFF switch. Press on the menu
button of the remote controller. Press on
the arrow (down) button of the remote
controller and select Distance. Select the
right distance by using the left and right
arrow buttons of the infra-red remote
controller.

71
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.5.9.4 Setting Infra-Red connection for use with KB-50 / CV-5000PC

Power up the CC-100 Series device with


the ON/OFF switch. Press on the menu
button of the remote controller. Press on
the arrow (down) button of the remote
controller and select I / R Devices. Press
on the arrow right button of the remote
controller. Choose the Device (read
channel) accordingly with the infra-red light
emitter channel setting on the CV power
supply. (see chapter 3.5.2)
Now select KB-50 RC

Note:
If you select Chart A, B1, C or F the CC-
100 Series will behave like a ACP-8
device. KB-50 or CV-5000PC must be set
also in ACP-8 mode.

3.5.9.5 Setting Infra-Red connection for use with KB-10

Power up the CC-100 Series device with


the ON/OFF switch. Press on the menu
button of the remote controller. Press on
the arrow (down) button of the remote
controller and select I / R Devices. Press
on the arrow right button of the remote
controller. Choose the Device (read
channel) accordingly with the infra-red light
emitter channel setting on the CV power
supply. (see chapter 3.5.2)
Now select Chart A, B1, C1 or F with the
arrow buttons.

Note:
CC-100 Series will behave like a ACP-8
device.

72
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

3.5.9.6 Setting CC-100 series for use with Remote Controller

Power up the CC-100 Series device with


the ON/OFF switch. Press on the menu
button of the remote controller. Press on
the arrow (down) button of the remote
controller and select I / R Devices. Press
on the arrow right button of the remote
controller.
Choose the Device (read channel)
accordingly with the infra-red remote
controller channel setting. (see chapter
3.5.8)
Now select CC-100, CC-100P or CC-
100XP

Advice:
To choose an infra-red channel, use the
letter combination. For example HH or
HL accordingly to the infra-red remote
controller channel setting.
Read: Device=Channel
(see chapter 3.5.8)

73
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.0 Compact power supply  Compact power supply

Example setup:

KR/RM
series

cable 5 meter
item 1830057

Compact power supply unit 1 Compact power supply unit 2

KB-50 KB-10 CV-5000 KB-50 KB-10 CV-5000


(CV-5000PC) (CV-5000PC)

4.0.1 Cable connection


Connect the cable (item 1830057, order Topcon) from power supply unit 1 on COMM port 1,2,3
or 4, to power supply unit 2 on COMM port 1,2,3 or 4. (see chapter 1.4 power supply unit
description). Possible to connect up to maximum of 4 power supply units.

4.0.2 Compact power supply dip-switch settings

4.0.2.1 2
Compact power supply units
Using COM port 1 on CPSU 1 Dip-switch bank 2 (DSW2) 1 : ON 2 : ON
Using COM port 2 on CPSU 1 Dip-switch bank 2 (DSW2) 3 : ON 4 : ON
Using COM port 3 on CPSU 1 Dip-switch bank 2 (DSW2) 5 : ON 6 : ON
Using COM port 4 on CPSU 1 Dip-switch bank 2 (DSW2) 7 : ON 8 : ON

Using COM port 1 on CPSU 2 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF


Using COM port 2 on CPSU 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF
Using COM port 3 on CPSU 2 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF
Using COM port 4 on CPSU 2 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF

74
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.0.2.2 3
Compact power supply units
Using COM port 1 on CPSU 1 Dip-switch bank 2 (DSW2) 1 : ON 2 : ON
Using COM port 2 on CPSU 1 Dip-switch bank 2 (DSW2) 3 : ON 4 : ON
Using COM port 3 on CPSU 1 Dip-switch bank 2 (DSW2) 5 : ON 6 : ON
Using COM port 4 on CPSU 1 Dip-switch bank 2 (DSW2) 7 : ON 8 : ON

Using COM port 1 on CPSU 2 Dip-switch bank 2 (DSW2) 1 : ON 2 : ON


Using COM port 2 on CPSU 2 Dip-switch bank 2 (DSW2) 3 : ON 4 : ON
Using COM port 3 on CPSU 2 Dip-switch bank 2 (DSW2) 5 : ON 6 : ON
Using COM port 4 on CPSU 2 Dip-switch bank 2 (DSW2) 7 : ON 8 : ON

Using COM port 1 on CPSU 3 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF


Using COM port 2 on CPSU 3 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF
Using COM port 3 on CPSU 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF
Using COM port 4 on CPSU 3 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF

4.0.2.3 4
Compact power supply units
Using COM port 1 on CPSU 1 Dip-switch bank 2 (DSW2) 1 : ON 2 : ON
Using COM port 2 on CPSU 1 Dip-switch bank 2 (DSW2) 3 : ON 4 : ON
Using COM port 3 on CPSU 1 Dip-switch bank 2 (DSW2) 5 : ON 6 : ON
Using COM port 4 on CPSU 1 Dip-switch bank 2 (DSW2) 7 : ON 8 : ON

Using COM port 1 on CPSU 2 Dip-switch bank 2 (DSW2) 1 : ON 2 : ON


Using COM port 2 on CPSU 2 Dip-switch bank 2 (DSW2) 3 : ON 4 : ON
Using COM port 3 on CPSU 2 Dip-switch bank 2 (DSW2) 5 : ON 6 : ON
Using COM port 4 on CPSU 2 Dip-switch bank 2 (DSW2) 7 : ON 8 : ON

Using COM port 1 on CPSU 3 Dip-switch bank 2 (DSW2) 1 : ON 2 : ON


Using COM port 2 on CPSU 3 Dip-switch bank 2 (DSW2) 3 : ON 4 : ON
Using COM port 3 on CPSU 3 Dip-switch bank 2 (DSW2) 5 : ON 6 : ON
Using COM port 4 on CPSU 3 Dip-switch bank 2 (DSW2) 7 : ON 8 : ON

Using COM port 1 on CPSU 4 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF


Using COM port 2 on CPSU 4 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF
Using COM port 3 on CPSU 4 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF
Using COM port 4 on CPSU 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF

Dip-switch settings for ALL compact power supply’s


RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)
Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

75
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.1 CL-200/2800 or EZ-200  CV-5000(PC)

Cable 5 meter
Item 1830057

EZ-200 CL-2800 CL-200 Compact Power Supply

EZ-200 from software version 1.42 (version 1.30 does not work with format NEW)
CL-200 from software version 1.30
CL-2800 from software version 2.10

4.1.1 Cable connection

Connect the cable (item 1830057, order Topcon) from EZ-200 or CL-200/2800 (port OUT) to the
Compact power supply unit on COMM port 1,2,3 or 4 (see chapter 1.4 power supply unit
description).

4.1.2 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

76
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.1.3 KB-10 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.4 Now select CL CONNECT
“ONLINE” Confirm OK by pressing the
F4 button, then press SET to restart the
KB-10.

Dial and SET button

4.1.4 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Data in the menu. Select CL (RX)
Connect “Online” Press OK, press
EXIT and press YES to restart the
KB-50.

77
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.1.5 CV-5000PC settings

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Data in the menu. Select CL
(RX) Connect “Online”. Click OK,
click EXIT and click YES to restart
CV-5000PC.

(see also chapter 2.3.4 for CV-5000pc


COMM port setting)

78
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.1.6 CL-200/2800 settings

Note: When single lens mode is selected on the CL-200 / CL-2800, data will not be
send to KB-50, KB-10 or CV-5000PC. Although CL-200 will display “RS-232C
data out Success”, if pressed on the print button to send data.

4.1.6.1 RS-232C setting

Power up the CL-200 / CL-2800


while holding down the 2nd button from
the right to enter the INITIAL MENU.
Select RS-232C “NORMAL” Thereafter
select EXIT and press the return button
(right down corner) to restart the
CL-200 / CL-2800.

4.1.6.2 Format setting

Power up the CL-200 / CL-2800.


Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “NEW FORMAT” Thereafter
select EXIT and press the return button
(right down corner) to exit the menu.

79
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.1.7 EZ-200 settings

NOTE: When pressing the print button on the EZ-200 to send data, there will be no
message displayed that data transfer is complete or successful. However the
data will be normally send successful.

4.1.7.1 RS-232C setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “OUTPUT” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

4.1.7.2 Format setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-FORMAT “NEW (CL)” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

80
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2 RM/KR-8800/8900  CV-5000(PC)

Cable 5 meter
Item 1830057

RM/KR-8800/8900 Compact power supply

RM-8800 from software version 1.25


KR-8800 from software version
RM-8900 from software version 1.00
KR-8900 from software version 1.00

4.2.1 Cable connection

Connect the cable (item 1830057, order Topcon) from RM/KR-88/8900 (port OUT) to the
Compact power supply unit on COMM port 1,2,3 or 4 (see chapter 1.4 power supply unit
description).

4.2.2 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

81
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2.3 KB-10 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.4 Now select RM/KR CONNECT
“ONLINE” Confirm OK by pressing
the F4 button, then press SET to
restart the KB-10.

Dial and SET button

4.2.4 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Data in the menu. Select RM (AR)
Connect “Online” Press OK, press
EXIT and press YES to restart the
KB-50.

82
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2.5 CV-5000PC settings

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Data in the menu. Select RM
(AR) Connect “Online”. Click OK,
click EXIT and click YES to restart
CV-5000PC.

(see also chapter 2.3.4 for CV-5000pc


COMM port setting)

83
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2.6 RM-8800/8900 settings

4.2.6.1 Format setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select DATA
FORMAT “NEW”. Select EXIT to
confirm.

4.2.6.2 Baud-Rate setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select BAUD-
RATE “2400”. Select EXIT to confirm.

4.2.6.3 Output data setting

Power up the RM-8800/8900. Press


the joystick button and wait until the
normal measuring screen is seen.
Press the menu button to enter the
SETTING MENU. Select OUTPUT
DATA “YES”. Select EXIT with the
joystick button to exit the Setting menu.

84
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2.7 KR-8800/8900 settings

4.2.7.1 Format setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select DATA
FORMAT “NEW”. Select EXIT to
confirm.

4.2.7.2 Baud-Rate setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select BAUD-
RATE “2400”. Select EXIT to confirm.

85
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2.7.3 Output data type

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select
OUTPUT DATA TYPE “REF”. Select
EXIT to confirm.

4.2.7.4 Output data setting

Power up the KR-8800/8900. Press the


joystick button and wait until the normal
measuring screen is seen. Press the
menu button to enter the SETTING
MENU. Select OUTPUT DATA “YES”.
Select EXIT with the joystick button to
exit the Setting menu.

86
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2A KR-7000P  CV-5000(PC)

Cable 5 meter
Item 1830057

KR-7000P Compact power supply

ALWAYS UPDATE THE LATEST AVAILABLE SOFTWARE VERSION ON THE KR-


7000P.

4.2A.1 Cable connection

Connect the cable (item 1830057, order Topcon) from KR-7000P (port OUT) to the Compact
power supply unit on COMM port 1,2,3 or 4 (see chapter 1.4 power supply unit description).

4.2A.2 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

87
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2A.3 KB-10 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.4 Now select RM/KR CONNECT
“ONLINE” Confirm OK by pressing
the F4 button, then press SET to
restart the KB-10.

Dial and SET button

4.2A.4 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Data in the menu. Select RM (AR)
Connect “Online” Press OK, press
EXIT and press YES to restart the
KB-50.

88
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2A.5 CV-5000PC settings

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Data in the menu. Select RM
(AR) Connect “Online”. Click OK,
click EXIT and click YES to restart
CV-5000PC.

(see also chapter 2.3.4 for CV-5000pc


COMM port setting)

89
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.2A.6 KR-7000P settings

4.2A.6.1 PRINT OUT settings

Power up the KR-7000P while holding


down the Contact-Glasses button to
enter the INITIAL MENU. Select PRINT
OUT to enter the PRINT OUT menu.
Make ALL selections as shown in
the screenshot

4.2A.6.2 ONLINE settings

Power up the KR-7000P while holding


down the Contact-Glasses button to
enter the INITIAL MENU. Select
ONLINE to enter the ONLINE menu.
Make ALL selections as shown in
the screenshot

4.2A.6.3 Dipswitch settings

Open the cover. Select Dipswitch


nr 8 ON, rest OFF.
Set the rotation switch to 0.

90
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.3 TRK-1P  CV-5000(PC)

Cable 5 meter
item 1830057

TRK-1P Compact power supply

NOTE: In this setup only Objective refraction data will be send to Compact power supply.
When i-base is also connected only Objective refraction data and Kerato data
can be send from TRK-1P to i-base through Compact power supply. Tono and
Pachy data (actually any data) can not be send to i-base through Compact power
supply if you select data out “ALL” on TRK-1P, because TRK-1P uses then
automatically STD3 format which Compact power supply does not support (yet,
November 2008)

4.3.1 Cable connection

Connect the cable (item 1830057, order Topcon) from TRK-1P (port OUT) to the Compact power
supply unit on COMM port 1,2,3 or 4 (see chapter 1.4 power supply unit description).

4.3.2 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

91
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.3.3 KB-10 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.4 Now select RM/KR CONNECT
“ONLINE” Confirm OK by pressing
the F4 button, then press SET to
restart the KB-10.

Dial and SET button

4.3.4 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Data in the menu. Select RM (AR)
Connect “Online” Press OK, press
EXIT and press YES to restart the
KB-50.

92
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.3.5 CV-5000PC settings

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Data in the menu. Select RM
(AR) Connect “Online”. Click OK,
click EXIT and click YES to restart
CV-5000PC.

(see also chapter 2.3.4 for CV-5000pc


COMM port setting)

4.3.6 TRK-1P settings

4.3.6.1 Output data type

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select OUTPUT DATA TYPE “REF”.

NOTE:
Only REF, KRT and R/K works with
CV-5000 Compact power supply.

TONO, PACHO, T/P and ALL will not


work in combination with CV-5000
Compact power supply. TRK-1P uses
for these settings automatically STD3
format, (no other format available)
which is not supported (yet) by CV-
5000 Compact power supply.

TRK-1P directly to i-base will work fine


in all settings !

93
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

4.3.6.2 Data Form setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select DATA FORM “NEW”.

NOTE: This selection can be


automatically changing if you
select an other “output data type”.
(example: if you change output
data type from “REF” to “ALL” the
data form will change automatically
to “STD3”)

4.3.6.3 Data Output Port

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select DATA OUTPUT PORT


“RS232C”.

4.3.6.4 Baud Rate setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select BAUD-RATE “2400”.

94
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.0 CV-5000 (KB-50/10)  i-base

Compact power supply Personal Computer (Topcon Shuttle)

Option 1) item 1830008 Option 2) item 1830095


+ item 1830008

5.0.1 Cable connection

Option 1) Connect the cable (item 1830008, order Topcon) from COMM port 1,2,3 or 4 (see
chapter 1.4 power supply unit description) of the CV-5000 Compact power supply to a serial port
of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect the cable (item 1830008, order Topcon) on COMM port 1,2,3 or 4 (see
chapter 1.4 power supply unit description) of the CV-5000 Compact power supply, and extend this
cable with a Serial to USB converter cable (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

5.0.2 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : ON 2 : OFF

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : ON 4 : OFF

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : ON 6 : OFF

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : ON 8 : OFF

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

95
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.0.3 KB-10 settings

Dip-switch settings Dip-switch bank 2 (DSW2)

Dip-switch bank 2 (DSW2) 1 : OFF

Note1: After changing dip-switch 2 : ON


settings restart the KB-10 by
switching the ON/OFF button 3 : ON
on the Compact Power Supply.
4 : OFF (2400)
Note2: After power up the KB-10 just
use the Dial button once to make 5 : OFF
the connection Bi-directional, in
order to receive data from i-base
by “Set Phoropter (Head)”

Dial button

5.0.4 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Other in the menu.

1) Select Print Switch “Export


only” or “Print and Export”.

2) Select RS-232C Baud Rate


“2400bps”. Press OK, press EXIT
and press YES to restart the
KB-50.

96
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.0.5 i-base settings (for Compact power supply)

5.0.5.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

97
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “Topcon Serial Refraction


Instruments” Plug-in.
Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“TCSerial” plug-in in Acquisitions of
i-base.

Note:
When the “TCSerial” Plug-in is not
active yet, right click on the plug-in to
ENABLE

98
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.0.5.2 Serial port setup

Double-Click on the TC Serial


plug-in.

1) Select the proper Com port on


which the CV-5000 Compact
Power Supply is connected to the
Personal Computer.

1a) (Find out in Windows Device


Manager which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

2) Enable this Comm port.

3) Select the Pre-setting for use with the selected com port in the picture is an example
CV compact PS KB-50 / KB-10

4) Enable (advised)

When you want to receive ALL


other data besides Subjective
data, like Lensmeter data, Final
data and Objective data in i-base
send by KB-50/KB-10.

Disable

When you want to receive ONLY


Subjective data in i-base send by
KB-50/KB-10. No other data like
Lensmeter data Final data and
Objective data will be received in
i-base.

Disable (obligatory)

If NOT disabled NO data at all will


be received in i-base from KB-50/
KB-10. (selection in picture is example)

5) Important: click on OK to confirm


all settings.

99
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.0.5.3 Advanced settings

Do NOT change these Pre-settings.

Unless you think is necessary and


know what you are doing !

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

Do NOT change these Pre-settings.

KB-50/KB-10 and Compact power


supply and i-base must set to the
same baud rate of 2400 !

Note:
You can select baud rate 9600 only if
necessary. But then KB-50/KB-10
and Compact Power supply must
also be set to 9600 ! (All 3 of them)

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

100
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the TCSerial plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

5.0.6 Decimal symbol setting in Windows for i-base

Open Windows Start menu.

Click on settings,
Click on Control Panel,
Click on Regional and
Language Options.

101
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Click on “Customize”

Select as an Decimal symbol a


“Comma” (not a dot).

102
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.1 CV-5000PC  i-base

5.1.1 CV-5000PC  Compact Power Supply…………………...See Chapter 2.3

5.1.2 CV-5000PC settings

5.1.2.1 Import/Export setup

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Other in the menu. Select
Page 2.
Select “Export only” or “Print and
Export”. Click OK, click EXIT and
click YES to restart CV-5000PC.

103
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.1.2.2 File Format Version

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Other in the menu. Select
Page 2.
Select “Version 1.00 (KB-1D Ver.
1.40 Compatible”. Click OK, click
EXIT and click YES to restart CV-
5000PC.

5.1.2.3 RS-232C Baud Rate

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Other in the menu. Select
Page 2.
Select “2400 bps”. Click OK, click
EXIT and click YES to restart CV-
5000PC.

104
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.1.3 i-base settings (for CV-5000PC)

5.1.3.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

105
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “CV for Windows” Plug-in.


Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“CVWin” plug-in in Acquisitions of
i-base.

Note:
When the “CVWin” Plug-in is not
active yet, right click on the plug-in to
ENABLE

106
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.1.3.2 CV Win setup

IMPORTANT: First install CV-


5000PC Software on the PC
before you do the CV Win setup.

Double-Click on the CVWin plug-


in.
Select the Pre-setting “CV-5000PC
Software”. Make exactly the same
settings as shown in the picture.

Make sure that the Program Path is


exactly the same as shown in the
picture.

Make sure that the Import and Export


path is exactly the same as shown
below in red.

C:\Documents and Settings\All Users\Application Data\TOPCON\CV-5000PC\DATA\ImpExp

When finishing making all selections


right-click on the CVWin plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

107
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

5.1.4 Decimal symbol setting in Windows for i-base

Open Windows Start menu.

Click on settings,
Click on Control Panel,
Click on Regional and
Language Options.

Click on “Customize”

108
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Select as an Decimal symbol a


“Comma” (not a dot).

6.0 CL-200/2800 or EZ-200  CV-5000(PC)  i-base

6.0.1 CL-200/2800 or EZ-200  CV-5000(PC)........................See Chapter 4.1

6.0.2 CV-5000  i-base......................................................See Chapter 5.0

6.0.3 CV-5000PC  i-base……………………………..………...See Chapter 5.1

109
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.1 CL-200/2800 or EZ-200  i-base  CV-5000

6.1.1 CL-200/2800 or EZ-200  i-base

EZ-200 CL-2800 CL-200 Personal Computer (Topcon Shuttle)

Option 1) item 1830008 Option 2) item 1830095


+ item 1830008

6.1.2 Cable connection

Option 1) Connect the cable (item 1830008, order Topcon) from EZ-200 or CL-200/2800
(port OUT) to a serial port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect the cable (item 1830008, order Topcon) to EZ-200 or CL-200/2800
(port OUT) and extend this cable with a Serial to USB converter cable (Item 1830095, order
Topcon) to a USB port of the Personal Computer. (Do not forget to install the proper Serial to
USB converter driver on the PC)

110
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.1.3 CL-200/2800 settings

Note: When single lens mode is selected on the CL-200 / CL-2800, data will not be
send to KB-50, KB-10 or CV-5000PC. Although CL-200 will display “RS-232C
data out Success”, if pressed on the print button to send data.

6.1.3.1 RS-232C setting

Power up the CL-200 / CL-2800


while holding down the 2nd button from
the right to enter the INITIAL MENU.
Select RS-232C “NORMAL” Thereafter
select EXIT and press the return button
(right down corner) to restart the
CL-200 / CL-2800.

6.1.3.2 Format setting

Power up the CL-200 / CL-2800.


Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “NEW FORMAT” Thereafter
select EXIT and press the return button
(right down corner) to exit the menu.

111
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.1.4 EZ-200 settings

NOTE: When pressing the print button on the EZ-200 to send data, there will be no
message displayed that data transfer is complete or successful. However the
data will be normally send successful.

6.1.4.1 RS-232C setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “OUTPUT” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

6.1.4.2 Format setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-FORMAT “NEW (CL)” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

112
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.1.5 i-base settings (for CL-200/2800 or EZ-200)

6.1.5.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

113
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “Topcon Serial Refraction


Instruments” Plug-in.
Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“TCSerial” plug-in in Acquisitions of
i-base.

Note:
When the “TCSerial” Plug-in is not
active yet, right click on the plug-in
and ENABLE

114
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.1.5.2 Serial port setup

Double-Click on the TC Serial


plug-in.

1) Select the proper Com port on


which the CL-200/2800 or EZ-200
is connected to the PC.

1a) (Find out in Windows Device


Manager which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

2) Enable this Comm port.

3) Select the Pre-setting for use


with Lensmeters.

4) Enable (obligatory)

5) Important: click on OK to confirm


all settings.

(selection in picture is example)

115
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.1.5.3 Advanced settings

Do NOT change these Pre-settings.

Unless you think is necessary and


know what you are doing !

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

Do NOT change these Pre-settings.

CL-200/2800, EZ-200 and i-base


must set to the same baud rate of
2400 !

Note:
CL-200/2800 and EZ-200 only
operates at baud rate of 2400 bps.
(new format)

116
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the TCSerial plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

6.1.6 i-base  CV-5000……………………………………………See Chapter 5.0

117
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.2 CL-200/2800 or EZ-200  i-base  CV-5000PC

6.2.1 CL-200/2800 or EZ-200  i-base……….………………..See Chapter 6.1.1

6.2.2 i-base  CV-5000PC……………………………………….See Chapter 5.1

6.3 RM/KR-8800/8900  CV-5000(PC)  i-base

6.3.1 RM/KR-8800/8900  CV-5000(PC)

See Chapter 4.2 ( + 6.3.1.1 below )

6.3.1.1 Output data type

This replace the settings of 4.2.7.3

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select
OUTPUT DATA TYPE “ALL”. Select
EXIT to confirm.

(in this setup KRT data is going


through the memory of the compact
power supply to i-base)

6.3.2 CV-5000  i-base…………………………………………...See Chapter 5.0

6.3.3 CV-5000PC  i-base.................................................See Chapter 5.1

118
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.4 RM/KR-8800/8900  i-base  CV-5000

6.4.1 RM/KR-8800/8900  i-base

RM/KR-8800/8900 Personal Computer (Topcon Shuttle)

Option 1) item 1830008 Option 2) item 1830095


+ item 1830008

6.4.2 Cable connection

Option 1) Connect the cable (item 1830008, order Topcon) from RM/KR-8800/8900
(port OUT) to a serial port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect the cable (item 1830008, order Topcon) to RM/KR-8800/8900


(port OUT) and extend this cable with a Serial to USB converter cable (Item 1830095, order
Topcon) to a USB port of the Personal Computer. (Do not forget to install the proper Serial to
USB converter driver on the PC)

119
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.4.3 RM-8800/8900 settings

6.4.3.1 Format setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select DATA
FORMAT “NEW”. Select EXIT to
confirm.

6.4.3.2 Baud-Rate setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select BAUD-
RATE “2400”. Select EXIT to confirm.

6.4.3.3 Output data setting

Power up the RM-8800/8900. Press


the joystick button and wait until the
normal measuring screen is seen.
Press the menu button to enter the
SETTING MENU. Select OUTPUT
DATA “YES”. Select EXIT with the
joystick button to exit the Setting menu.

120
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.4.4 KR-8800/8900 settings

6.4.4.1 Format setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select DATA
FORMAT “NEW”. Select EXIT to
confirm.

6.4.4.2 Baud-Rate setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select BAUD-
RATE “2400”. Select EXIT to confirm.

121
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.4.4.3 Output data type

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select
OUTPUT DATA TYPE “ALL”. Select
EXIT to confirm.

6.4.4.4 Output data setting

Power up the KR-8800/8900. Press the


joystick button and wait until the normal
measuring screen is seen. Press the
menu button to enter the SETTING
MENU. Select OUTPUT DATA “YES”.
Select EXIT with the joystick button to
exit the Setting menu.

122
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.4.5 i-base settings (for RM/KR-8800/8900)

6.4.5.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

123
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “Topcon Serial Refraction


Instruments” Plug-in.
Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“TCSerial” plug-in in Acquisitions of
i-base.

Note:
When the “TCSerial” Plug-in is not
active yet, right click on the plug-in
and ENABLE

124
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.4.5.2 Serial port setup

Double-Click on the TC Serial


plug-in.

1) Select the proper Com port on


which the RM/KR-8800/8900 is
connected to the PC.

1a) (Find out in Windows Device


Manager which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

2) Enable this Comm port.

3) Select the Pre-setting for use


with Refractometers

4) Disable (obligatory)

5) Important: click on OK to confirm


all settings.

(selection in picture is example)

125
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.4.5.3 Advanced settings

Do NOT change these Pre-settings.

Unless you think is necessary and


know what you are doing !

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

Do NOT change these Pre-settings.

RM/KR-8800/8900 and i-base must


set to the same baud rate of 2400 !

126
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the TCSerial plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

6.4.6 i-base  CV-5000......................................................See Chapter 5.0

127
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.5 RM/KR-8800/8900  i-base  CV-5000PC

6.5.1 RM/KR-8800/8900  i-base………………………….……See Chapter 6.4.1

6.5.2 i-base  CV-5000PC……………………………..………...See Chapter 5.1

6.6 TRK-1P  CV-5000(PC)  i-base

6.6.1 TRK-1P  CV-5000(PC)

See Chapter 4.3 ( + 6.6.1.1 below )

6.6.1.1 Output data type

This replace the settings of 4.3.6.1

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select OUTPUT DATA TYPE “R/K”.

NOTE:
Only REF, KRT and R/K works with
CV-5000 Compact power supply.

TONO, PACHO, T/P and ALL will not


work in combination with CV-5000
Compact power supply. TRK-1P uses
for these settings automatically STD3
format, (no other format available) (in this setup KRT data is going
which is not supported (yet) by CV- through the memory of the compact
5000 Compact power supply. power supply to i-base)
TRK-1P directly to i-base will work fine
in all settings !

6.6.2 CV-5000  i-base………………………………….………..See Chapter 5.0

6.6.3 CV-5000PC  i-base………………………………….……See Chapter 5.1

128
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.7 TRK-1P  i-base  CV-5000

6.7.1 TRK-1P  i-base

TRK-1P Personal Computer (Topcon Shuttle)

Option 1) item 1830008 Option 2) item 1830095


+ item 1830008

6.7.2 Cable connection

Option 1) Connect the cable (item 1830008, order Topcon) from TRK-1P (port OUT) to a
serial port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect the cable (item 1830008, order Topcon) to TRK-1P (port OUT) and extend
this cable with a Serial to USB converter cable (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

6.7.3 TRK-1P settings

6.7.3.1 Output data type

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select OUTPUT DATA TYPE “ALL”.

129
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.7.3.2 Data Form setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select DATA FORM “STD3”.

NOTE: This selection can be


automatically changing if you
select an other “output data type”.

6.7.3.3 Data Output Port

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select DATA OUTPUT PORT


“RS232C”.

6.7.3.4 Baud Rate setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select BAUD-RATE “9600”.

130
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.7.4 i-base settings (for TRK-1P)

6.7.4.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

131
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “Topcon Serial Refraction


Instruments” Plug-in.
Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“TCSerial” plug-in in Acquisitions of
i-base.

Note:
When the “TCSerial” Plug-in is not
active yet, right click on the plug-in
and ENABLE

132
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.7.4.2 Serial port setup

Double-Click on the TC Serial


plug-in.

1) Select the proper Com port on


which the TRK-1P is connected
to the PC.

1a) (Find out in Windows Device


Manager which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

2) Enable this Comm port.

3) Select the Pre-setting for use


with TRK-1P

4) Disable (obligatory)

5) Important: click on OK to confirm


all settings.

(selection in picture is example)

133
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.7.4.3 Advanced settings

Do NOT change these Pre-settings.

Unless you think is necessary and


know what you are doing !

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

Do NOT change these Pre-settings.

TRK-1P and i-base must set to the


same baud rate of 9600 !

134
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the TCSerial plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

6.7.5 i-base  CV-5000……………………………….…………..See Chapter 5.0

135
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

6.8 TRK-1P  i-base  CV-5000PC

6.8.1 TRK-1P  i-base……………………………………..….....See Chapter 6.7.1

6.8.2 i-base  CV-5000PC……………………………………….See Chapter 5.1

136
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.0 CL/EZ  RM/KR  CV-5000(PC)  i-base

7.0.1 CL/EZ  RM/KR  CV-5000(PC)

EZ-200
CL-200/2800 RM/KR-8800/8900 Compact Power Supply

out in out
Cable item 1830057 (5m) Cable item 1830057 (5m)

7.0.2 Cable connection


Connect the cable (item 1830057, order Topcon) from CL-200/2800 or EZ-200 (port OUT) to the
RM/KR-8800/8900 (port IN).
Connect the cable (item 1830057, order Topcon) from RM/KR-8800/8900 (port OUT) to the
Compact power supply unit on COMM port 1,2,3 or 4 (see chapter 1.4 power supply unit
description).

For cable connection CV-5000  i-base see Chapter 5.0.1

For cable connection CV-5000PC  i-base see Chapter 2.3

137
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.0.3 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

7.0.4 KB-10 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.4 Now select RM/KR CONNECT
“ONLINE” and select CL CONNECT
“ONLINE”. Confirm OK by pressing
the F4 button, then press SET to
restart the KB-10.

Dial and SET button

138
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.0.5 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Data in the menu. Select RM (AR)
Connect “Online” and select CL
(RX) Connect “Online”. Press OK,
press EXIT and press YES to restart
the KB-50.

7.0.6 CV-5000PC settings

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Data in the menu. Select RM
(AR) Connect “Online” and select CL
(RX) Connect “Online”. Click OK,
click EXIT and click YES to restart
CV-5000PC.

(see also chapter 2.3.4 for CV-5000pc


COMM port setting)

139
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.0.7 CL-200/2800 settings

Note: When single lens mode is selected on the CL-200 / CL-2800, data will not be
send to KB-50, KB-10 or CV-5000PC. Although CL-200 will display “RS-232C
data out Success”, if pressed on the print button to send data.

7.0.7.1 RS-232C setting

Power up the CL-200 / CL-2800


while holding down the 2nd button from
the right to enter the INITIAL MENU.
Select RS-232C “NORMAL” Thereafter
select EXIT and press the return button
(right down corner) to restart the
CL-200 / CL-2800.

7.0.7.2 Format setting

Power up the CL-200 / CL-2800.


Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “NEW FORMAT” Thereafter
select EXIT and press the return button
(right down corner) to exit the menu.

140
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.0.8 EZ-200 settings

NOTE: When pressing the print button on the EZ-200 to send data, there will be
no message displayed that data transfer is complete or successful.
However the data will be normally send successful.

7.0.8.1 RS-232C setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “OUTPUT” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

7.0.8.2 Format setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-FORMAT “NEW (CL)” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

141
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.0.9 RM-8800/8900 settings

7.0.9.1 CL input setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select CL
INPUT “NEW”. Select EXIT to confirm.

7.0.9.2 Format setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select DATA
FORMAT “NEW”. Select EXIT to
confirm.

7.0.9.3 Baud-Rate setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select BAUD-
RATE “2400”. Select EXIT to confirm.

142
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.0.9.4 Output data setting

Power up the RM-8800/8900. Press


the joystick button and wait until the
normal measuring screen is seen.
Press the menu button to enter the
SETTING MENU. Select OUTPUT
DATA “YES”. Select EXIT with the
joystick button to exit the Setting menu.

7.0.10 KR-8800/8900 settings


7.0.10.1 CL input setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select CL
INPUT “NEW”. Select EXIT to confirm.

7.0.10.2 Format setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select DATA
FORMAT “NEW”. Select EXIT to
confirm.

143
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.0.10.3 Baud-Rate setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select BAUD-
RATE “2400”. Select EXIT to confirm.

7.0.10.4 Output data type

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select
OUTPUT DATA TYPE “REF”. Select
EXIT to confirm.

If also CV-5000(PC)  i-base


Select OUTPUT DATA TYPE “R/K”.
Kerato data will than be stored in
CPS and send to i-base.

7.0.10.5 Output data setting

Power up the KR-8800/8900. Press the


joystick button and wait until the normal
measuring screen is seen. Press the
menu button to enter the SETTING
MENU. Select OUTPUT DATA “YES”.
Select EXIT with the joystick button to
exit the Setting menu.

7.0.11 CV-5000  i-base………….………………………………See Chapter 5.0

7.0.12 CV-5000PC  i-base………….………….……………….See Chapter 5.1


144
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.1 CL/EZ  RM/KR  i-base  CV-5000(PC)

7.1.1 CL/EZ  RM/KR  i-base

EZ-200 Personal Computer


CL-200/2800 RM/KR-8800/8900 (Topcon Shuttle) Compact Power Supply

out 1 in out 2 3 usb


5 4

7.1.2 Cable connection


CL/EZ to RM/KR
Connect cable 1 (item 1830057, order Topcon) from CL-200/2800 or EZ-200 (port OUT) to the
RM/KR-8800/8900 (port IN).
RM/KR to PC Shuttle
Option 1) Connect cable 2 (item 1830008, order Topcon) from RM/KR-8800/8900
(port OUT) to a serial port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 2 (item 1830008, order Topcon) to RM/KR-8800/8900


(port OUT) and extend this cable with a Serial to USB converter cable 3 (Item 1830095, order
Topcon) to a USB port of the Personal Computer. (Do not forget to install the proper Serial to
USB converter driver on the PC)

PC Shuttle to CPS
Option 1) Connect cable 4 (item 1830008, order Topcon) from COMM port 1,2,3 or 4 (see
chapter 1.4 power supply unit description) of the CV-5000 Compact power supply to a serial port
of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 4 (item 1830008, order Topcon) on COMM port 1,2,3 or 4 (see chapter
1.4 power supply unit description) of the CV-5000 Compact power supply, and extend this cable
with a Serial to USB converter cable 5 (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

For cable connection CV-5000PC  i-base see Chapter 2.3

145
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.1.3 CL-200/2800 settings

Note: When single lens mode is selected on the CL-200/2800, data will not be
send to KB-50, KB-10 or CV-5000PC. Although CL-200/2800 will display
“RS-232C data out Success”, if pressed on the print button to send data.

7.1.3.1 RS-232C setting

Power up the CL-200 / CL-2800


while holding down the 2nd button from
the right to enter the INITIAL MENU.
Select RS-232C “NORMAL” Thereafter
select EXIT and press the return button
(right down corner) to restart the
CL-200 / CL-2800.

7.1.3.2 Format setting

Power up the CL-200 / CL-2800.


Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “NEW FORMAT” Thereafter
select EXIT and press the return button
(right down corner) to exit the menu.

146
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.1.4 EZ-200 settings

NOTE: When pressing the print button on the EZ-200 to send data, there will be no
message displayed that data transfer is complete or successful. However the
data will be normally send successful.

7.1.4.1 RS-232C setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “OUTPUT” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

7.1.4.2 Format setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-FORMAT “NEW (CL)” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

147
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.1.5 RM-8800/8900 settings

7.1.5.1 CL input setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select CL
INPUT “NEW”. Select EXIT to confirm.

7.1.5.2 Format setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select DATA
FORMAT “NEW”. Select EXIT to
confirm.

7.1.5.3 Baud-Rate setting

Power up the RM-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select BAUD-
RATE “2400”. Select EXIT to confirm.

148
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.1.5.4 Output data setting

Power up the RM-8800/8900. Press


the joystick button and wait until the
normal measuring screen is seen.
Press the menu button to enter the
SETTING MENU. Select OUTPUT
DATA “YES”. Select EXIT with the
joystick button to exit the Setting menu.

7.1.6 KR-8800/8900 settings


7.1.6.1 CL input setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select CL
INPUT “NEW”. Select EXIT to confirm.

7.1.6.2 Format setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select DATA
FORMAT “NEW”. Select EXIT to
confirm.

149
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.1.6.3 Baud-Rate setting

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select BAUD-
RATE “2400”. Select EXIT to confirm.

7.1.6.4 Output data type

Power up the KR-8800/8900 while


holding down the menu button to enter
the INITIAL MENU. Select ON-LINE to
enter the on-line menu. Select
OUTPUT DATA TYPE “ALL”. Select
EXIT to confirm.

7.1.6.5 Output data setting

Power up the KR-8800/8900. Press the


joystick button and wait until the normal
measuring screen is seen. Press the
menu button to enter the SETTING
MENU. Select OUTPUT DATA “YES”.
Select EXIT with the joystick button to
exit the Setting menu.

150
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.1.7 i-base settings (for RM/KR-8800/8900)

7.1.7.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

151
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “Topcon Serial Refraction


Instruments” Plug-in.
Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“TCSerial” plug-in in Acquisitions of
i-base.

Note:
When the “TCSerial” Plug-in is not
active yet, right click on the plug-in
and ENABLE

152
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.1.7.2 Serial port setup

Double-Click on the TC Serial


plug-in.

1) Select the proper Com port on


which the RM/KR-8800/8900 is
connected to the PC.

1a) (Find out in Windows Device


Manager which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

2) Enable this Comm port.

3) Select the Pre-setting for use


with Refractometers

4) Disable (obligatory)

5) Important: click on OK to confirm


all settings.

(selection in picture is example)

153
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.1.7.3 Advanced settings

Do NOT change these Pre-settings.

Unless you think is necessary and


know what you are doing !

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

Do NOT change these Pre-settings.

RM/KR-8800/8900 and i-base must


set to the same baud rate of 2400 !

154
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the TCSerial plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

7.1.8 i-base  CV-5000……………………………………..…….See Chapter 5.0

7.1.9 i-base  CV-5000PC…………………………………….....See Chapter 5.1

155
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.2 CL/EZ  TRK-1P  CV-5000(PC)  i-base

7.2.1 CL/EZ  TRK-1P  CV-5000(PC)

EZ-200
CL-200/2800 TRK-1P Compact Power Supply

out in out
Cable item 1830057 (5m) Cable item 1830057 (5m)

NOTE: In this setup only Objective refraction data from TRK-1P together with CL/EZ
data will be send to Compact power supply. When i-base is also connected only
Objective refraction data and Kerato data together with CL/EZ data can be send
from TRK-1P and CL/EZ to i-base through Compact power supply. Tono and
Pachy data (actually any data) can not be send to i-base through Compact power
supply if you select data out “ALL” on the TRK-1P, because TRK-1P uses then
automatically STD3 format which Compact power supply does not support.

7.2.2 Cable connection

Connect the cable (item 1830057, order Topcon) from CL-200/2800 or EZ-200 (port OUT) to the
TRK-1P (port IN).
Connect the cable (item 1830057, order Topcon) from TRK-1P (port OUT) to the Compact power
supply unit on COMM port 1,2,3 or 4 (see chapter 1.4 power supply unit description).

For cable connection CV-5000  i-base see Chapter 5.0.1

For cable connection CV-5000PC  i-base see Chapter 2.3

156
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.2.3 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

7.2.4 KB-10 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.4 Now select RM/KR CONNECT
“ONLINE” and select CL CONNECT
“ONLINE”. Confirm OK by pressing
the F4 button, then press SET to
restart the KB-10.

Dial and SET button

157
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.2.5 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Data in the menu. Select RM (AR)
Connect “Online” and select CL
(RX) Connect “Online”. Press OK,
press EXIT and press YES to restart
the KB-50.

7.2.6 CV-5000PC settings

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Data in the menu. Select RM
(AR) Connect “Online” and select CL
(RX) Connect “Online”. Click OK,
click EXIT and click YES to restart
CV-5000PC.

(see also chapter 2.3.4 for CV-5000pc


COMM port setting)

158
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.2.7 CL-200/2800 settings

Note: When single lens mode is selected on the CL-200 / CL-2800, data will not be
send to KB-50, KB-10 or CV-5000PC. Although CL-200 will display “RS-232C
data out Success”, if pressed on the print button to send data.

7.2.7.1 RS-232C setting

Power up the CL-200 / CL-2800


while holding down the 2nd button from
the right to enter the INITIAL MENU.
Select RS-232C “NORMAL” Thereafter
select EXIT and press the return button
(right down corner) to restart the
CL-200 / CL-2800.

7.2.7.2 Format setting

Power up the CL-200 / CL-2800.


Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “NEW FORMAT” Thereafter
select EXIT and press the return button
(right down corner) to exit the menu.

159
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.2.8 EZ-200 settings

NOTE: When pressing the print button on the EZ-200 to send data, there will be no
message displayed that data transfer is complete or successful. However the
data will be normally send successful.

7.2.8.1 RS-232C setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “OUTPUT” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

7.2.8.2 Format setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-FORMAT “NEW (CL)” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

160
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.2.9 TRK-1P settings

7.2.9.1 Output data type

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select OUTPUT DATA TYPE “REF”.

If also CV-5000(PC)  i-base


Select OUTPUT DATA TYPE “R/K”.
Kerato data will than be stored in
CPS and send to i-base.

NOTE:
Only REF, KRT and R/K works with
CV-5000 Compact power supply.

TONO, PACHO, T/P and ALL will not


work in combination with CV-5000
Compact power supply. TRK-1P uses
for these settings automatically STD3
format, (no other format available)
which is not supported (yet) by CV-
5000 Compact power supply.

TRK-1P directly to i-base will work fine


in all settings !

7.2.9.2 Data Form setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select DATA FORM “NEW”.

NOTE: This selection can be


automatically changing if you
select an other “output data type”.
(example: if you change output
data type from “REF” to “ALL” the
data form will change automatically
to “STD3”)

161
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.2.9.3 Data Output Port

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select DATA OUTPUT PORT


“RS232C”.

7.2.9.4 Baud Rate setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select BAUD-RATE “2400”.

7.2.9.5 CL input setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select CL INPUT “NEW”.

7.2.10 CV-5000  i-base……………….…………………………See Chapter 5.0

7.2.11 CV-5000PC  i-base…............................................See Chapter 5.1

162
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.3 CL/EZ  TRK-1P  i-base  CV-5000(PC)

7.3.1 CL/EZ  TRK-1P  i-base

EZ-200 Personal Computer


CL-200/2800 TRK-1P (Topcon Shuttle) Compact Power Supply

out 1 in out 2 3 usb


5 4

7.3.2 Cable connection


CL/EZ to TRK-1P
Connect cable 1 (item 1830057, order Topcon) from CL-200/2800 or EZ-200 (port OUT) to the
TRK-1P (port IN).
TRK-1P to PC Shuttle
Option 1) Connect cable 2 (item 1830008, order Topcon) from TRK-1P (port OUT) to a serial
port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 2 (item 1830008, order Topcon) to TRK-1P (port OUT) and extend this
cable with a Serial to USB converter cable 3 (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

PC Shuttle to CPS
Option 1) Connect cable 4 (item 1830008, order Topcon) from COMM port 1,2,3 or 4 (see
chapter 1.4 power supply unit description) of the CV-5000 Compact power supply to a serial port
of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 4 (item 1830008, order Topcon) on COMM port 1,2,3 or 4 (see chapter
1.4 power supply unit description) of the CV-5000 Compact power supply, and extend this cable
with a Serial to USB converter cable 5 (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

For cable connection CV-5000PC  i-base see Chapter 2.3

163
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.3.3 CL-200/2800 settings

Note: When single lens mode is selected on the CL-200/2800, data will not be
send to KB-50, KB-10 or CV-5000PC. Although CL-200/2800 will display
“RS-232C data out Success”, if pressed on the print button to send data.

7.3.3.1 RS-232C setting

Power up the CL-200 / CL-2800


while holding down the 2nd button from
the right to enter the INITIAL MENU.
Select RS-232C “NORMAL” Thereafter
select EXIT and press the return button
(right down corner) to restart the
CL-200 / CL-2800.

7.3.3.2 Format setting

Power up the CL-200 / CL-2800.


Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “NEW FORMAT” Thereafter
select EXIT and press the return button
(right down corner) to exit the menu.

164
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.3.4 EZ-200 settings

NOTE: When pressing the print button on the EZ-200 to send data, there will be no
message displayed that data transfer is complete or successful. However the
data will be normally send successful.

7.3.4.1 RS-232C setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-232C “OUTPUT” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

7.3.4.2 Format setting

Power up the EZ-200. Wait until the


normal measuring screen is seen.
Press “mode” button then press “menu”
button to enter the user MENU. Select
RS-FORMAT “NEW (CL)” Thereafter
select EXIT and press the return button
(left down corner) to exit the menu.

165
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.3.5 TRK-1P settings

7.3.5.1 Output data type

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select OUTPUT DATA TYPE “ALL”.

7.3.5.2 Data Form setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select DATA FORM “STD3”.

NOTE: This selection can be


automatically changing if you
select an other “output data type”.

7.3.5.3 Data Output Port

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select DATA OUTPUT PORT


“RS232C”.

166
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.3.5.4 Baud Rate setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select BAUD-RATE “9600”.

7.3.5.5 CL input setting

Power up the TRK-1P while holding


down the 2nd button from the right to
enter the INITIAL MENU. Select ON-
LINE.

Select CL INPUT “NEW”.

167
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.3.6 i-base settings (for TRK-1P)

7.3.6.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

168
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “Topcon Serial Refraction


Instruments” Plug-in.
Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“TCSerial” plug-in in Acquisitions of
i-base.

Note:
When the “TCSerial” Plug-in is not
active yet, right click on the plug-in
and ENABLE

169
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.3.6.2 Serial port setup

Double-Click on the TC Serial


plug-in.

1) Select the proper Com port on


which the TRK-1P is connected
to the PC.

1a) (Find out in Windows Device


Manager which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

2) Enable this Comm port.

3) Select the Pre-setting for use


with TRK-1P

4) Disable (obligatory)

5) Important: click on OK to confirm


all settings.

(selection in picture is example)

170
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

7.3.6.3 Advanced settings

Do NOT change these Pre-settings.

Unless you think is necessary and


know what you are doing !

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

Do NOT change these Pre-settings.

TRK-1P and i-base must set to the


same baud rate of 9600 !

171
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the TCSerial plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

7.3.7 i-base  CV-5000……………………………………………See Chapter 5.0

7.3.8 i-base  CV-5000PC……………………………………….See Chapter 5.1

172
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0 KR-8100P  Colormap 32  CV-5000

KR-8100P PC (shuttle) with CM32 software Compact power supply unit

out 1 2 4 usb 3

NOTE: CL-200/2800 or EZ-200 connected to KR-8100P in this setup will unfortunately not work.
KRP32 software will not pass the CL/EZ data to CV-5000. The CL-200/2800 or EZ-200
can be directly connected to the Compact power supply.

8.0.1 Cable connection

KR-8100P to PC Shuttle
Option 1) Connect cable 1 (item 1830008, order Topcon) from KR-8100P (port OUT) to a serial
port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 1 (item 1830008, order Topcon) to KR-8100P (port OUT) and extend
this cable with a Serial to USB converter cable 2 (Item 1830095, order Topcon) to a USB port of
the Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the
Personal Computer).

PC Shuttle to CPS
Option 1) Connect cable 3 (item 1830008, order Topcon) from COMM port 1,2,3 or 4 (see
chapter 1.4 power supply unit description) of the CV-5000 Compact power supply to a serial port
of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 3 (item 1830008, order Topcon) on COMM port 1,2,3 or 4 (see chapter
1.4 power supply unit description) of the CV-5000 Compact power supply, and extend this cable
with a Serial to USB converter cable 4 (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

173
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0.2 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF *


Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF *
Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF *
Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF *

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

*) NOTE: The dip-switches should be both OFF for the Comm. port in use. Although the Compact
power supply is connected to a PC, the dip-switch setting of the Compact power supply
unit must be set accordingly like connected with a KR instrument. For the Compact
power supply it seems that the data is coming from a KR instrument.

174
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0.3 KB-10 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.4 Now select RM/KR CONNECT
“ONLINE” Confirm OK by pressing
the F4 button, then press SET to
restart the KB-10.

Dial and SET button

8.0.4 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Data in the menu. Select RM (AR)
Connect “Online” Press OK, press
EXIT and press YES to restart the
KB-50.

175
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0.5 KR-8100P settings

8.0.5.1 ON-LINE settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select ON-LINE to enter the on-line


menu.

Select CL INPUT NO
Select RS232C FORM EXT2
Select BAUD-RATE 38400
Select RS232C OUT TYPE ALL

(make same selection as in Picture)

Select EXIT to confirm.

8.0.5.2 Output data setting

Power up the KR-8100P(A). Press the


joystick button and wait until the normal
measuring screen is seen. Press the
menu button to enter the SETTING
MENU. Select RS232C “YES”.
Select EXIT with the joystick button to
exit the Setting menu.

176
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0.5.3 PLACIDO settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select PLACIDO SET to enter the


Placido set menu.

Select PC S/W SETUP and EXECUTE


to calibrate the KR-8100P(A) with the
KRP32 Software.

Select EXIT to confirm.

NOTE:
To calibrate the KR-8100P(A) with the
KRP32 Software, KRP32 software
must be installed, configured and
running.
(see chapter 8.0.6)

177
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0.6 Colormap 32 (KRP32) settings

8.0.6.1 KR COM Port setting

Install Color Mapping 32 software


(V1.25) on the PC (Shuttle).
Select Port, leftclick on “Select COM port
for KR”.

Select the Com port on which the KR-


8100P(A) is connected to the Personal
Computer.

(Find out in Windows Device Manager


which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

NOTE:
Be aware of conflicting Com Ports or
Com ports already in use by other
devices like in i-base.

(Selected COM port is example)

8.0.6.2 Setup calibration data

After installation of the KRP32 software


and when Com port selection for KR-
8100P is made, the following window
pops up.

For setup calibration data see chapter


8.0.5.3 Placido settings.

178
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0.6.3 CV COM port setting

Install Color Mapping 32 software


(V1.25) on the PC (Shuttle).
Select Port, leftclick on “Select COM port
for CV”.

Select the Com port on which the CV-


5000 Compact power supply is
connected to the Personal Computer.

(Find out in Windows Device Manager


which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

Select CV Format “NEW FORMAT”.

NOTE:
Be aware of conflicting Com Ports or
Com ports already in use by other
devices like in i-base.
(Selected COM port is example)

8.0.6.4 Select instrument

Select Port, leftclick on “Select


Instrument” and select “KP8000/38K”.

NOTE: (optional)
To be sure that all Com ports are
configured leftclick on “Reset COM port”.

179
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0A KR-7000P  Colormap 32  CV-5000

KR-7000P PC (shuttle) with CM32 software Compact power supply unit

out 1 2 4 usb 3

NOTE: CL-200/2800 or EZ-200 connected to KR-7000P in this setup will unfortunately not work.
KRP32 software will not pass the CL/EZ data to CV-5000. The CL-200/2800 or EZ-200
can be directly connected to the Compact power supply.

8.0A.1 Cable connection

KR-7000P to PC Shuttle
Option 1) Connect cable 1 (item 1830008, order Topcon) from KR-7000P (port OUT) to a serial
port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 1 (item 1830008, order Topcon) to KR-7000P (port OUT) and extend
this cable with a Serial to USB converter cable 2 (Item 1830095, order Topcon) to a USB port of
the Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the
Personal Computer).

PC Shuttle to CPS
Option 1) Connect cable 3 (item 1830008, order Topcon) from COMM port 1,2,3 or 4 (see
chapter 1.4 power supply unit description) of the CV-5000 Compact power supply to a serial port
of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 3 (item 1830008, order Topcon) on COMM port 1,2,3 or 4 (see chapter
1.4 power supply unit description) of the CV-5000 Compact power supply, and extend this cable
with a Serial to USB converter cable 4 (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

180
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0A.2 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF *


Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF *
Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF *
Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF *

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

*) NOTE: The dip-switches should be both OFF for the Comm. port in use. Although the Compact
power supply is connected to a PC, the dip-switch setting of the Compact power supply
unit must be set accordingly like connected with a KR instrument. For the Compact
power supply it seems that the data is coming from a KR instrument.

181
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0A.3 KB-10 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.4 Now select RM/KR CONNECT
“ONLINE” Confirm OK by pressing
the F4 button, then press SET to
restart the KB-10.

Dial and SET button

8.0A.4 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Data in the menu. Select RM (AR)
Connect “Online” Press OK, press
EXIT and press YES to restart the
KB-50.

182
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0A.5 KR-7000P settings

8.0A.5.1 PRINT OUT settings

Power up the KR-7000P while holding


down the Contact-Glasses button to
enter the INITIAL MENU.

Select PRINT OUT to enter the PRINT


OUT menu.

Select PRINT TYPE ALL


Select S.E. DATA YES
Select CL PRINT NO
Select AUTO PRINT YES
Select PRINT R/L DATA
Select REF GRAPHIC YES

(make same selection as in Picture)

Select NEXT to confirm.

8.0A.5.2 ONLINE settings

Power up the KR-7000P while holding


down the Contact-Glasses button to
enter the INITIAL MENU. Select
ONLINE to enter the ONLINE menu.
Make ALL selections as shown in
the screenshot

183
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0A.5.3 PLACIDO settings

Power up the KR-7000P while holding


down the Contact-Glasses button to
enter the INITIAL MENU.

Select PLACIDO SET to enter the


PLACIDO SET menu.

(make same selection as in Picture)

Select NEXT to confirm.

NOTE:
To calibrate the KR-7000P with the
KRP32 Software, KRP32 software
must be installed, configured and
running.
(see chapter 8.0A.6)

8.0A.5.4 Dipswitch settings

Open the cover. Select Dipswitch


nr 7 and 8 ON, rest OFF.
Set the rotation switch to 7.

184
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0A.5.5 Print wheel settings

In order to be able to send data, be


sure to set the print wheel as in the
picture !

When receiving this error message


please check the baud-rate.

185
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0A.6 Colormap 32 (KRP32) settings

8.0A.6.1 KR COM Port setting

Install Color Mapping 32 software


(V1.25) on the PC (Shuttle).
Select Port, leftclick on “Select COM port
for KR”.

Select the Com port on which the KR-


8100P(A) is connected to the Personal
Computer.

(Find out in Windows Device Manager


which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

NOTE:
Be aware of conflicting Com Ports or
Com ports already in use by other
devices like in i-base.

(Selected COM port is example)

8.0A.6.2 Setup calibration data

After installation of the KRP32 software


and when Com port selection for KR-
8100P is made, the following window
pops up.

For setup calibration data see chapter


8.0A.5.3 Placido settings.

186
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.0A.6.3 CV COM port setting

Install Color Mapping 32 software


(V1.25) on the PC (Shuttle).
Select Port, leftclick on “Select COM port
for CV”.

Select the Com port on which the CV-


5000 Compact power supply is
connected to the Personal Computer.

(Find out in Windows Device Manager


which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

Select CV Format “NEW FORMAT”.

NOTE:
Be aware of conflicting Com Ports or
Com ports already in use by other
devices like in i-base.
(Selected COM port is example)

8.0A.6.4 Select instrument

Select Port, leftclick on “Select


Instrument” and select “KR7000P”.

NOTE: (optional)
To be sure that all Com ports are
configured leftclick on “Reset COM port”.

187
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.1 KR-8100P  Colormap 32  CV-5000PC


KR-8100P PC (shuttle) with CM32 software CV-5000PC

out serial usb


Option 1) item 1830008 Option 2) item 1830095
+ item 1830008

KR-8100P(A) software version 1.34 and up


KRP 32 software version 1.25 and up
CPS unit software version 1.15 and up
CV-5000PC software version 1.21 and up

8.1.1 CV-5000PC  Compact Power Supply…………………...See Chapter 2.3

NOTE: CL-200/2800 or EZ-200 connected to KR-8100P in this setup will unfortunately not work.
KRP32 software will not pass the CL/EZ data to CV-5000. The CL-200/2800 or EZ-200
can be directly connected to the Compact power supply.

8.1.2 Cable connection

KR-8100P to PC Shuttle
Option 1) Connect the cable (item 1830008, order Topcon) from KR-8100P (port OUT) to a
serial port of the Personal Computer. (The serial port is usually COMM 1).
Option 2) Connect the cable (item 1830008, order Topcon) to KR-8100P (port OUT) and extend
this cable with a Serial to USB converter cable (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the
Personal Computer).

8.1.3 Compact power supply dip-switch settings

No special dip-switch settings needed for this setup

188
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.1.4 CV-5000PC settings

8.1.4.1 Setup CV-5000PC

Install the latest CV-5000PC software


available on the TOPCON FTP site.

(If CV-5000PC software is already


installed on the PC see chapter
8.1.4.3).

When execute the installation


program, the “setup TOPCON CV-
5000PC window pops up.

Enable “Ref/Krt data will be received


from ‘Color-Mapping Software’, (same
as in picture).

8.1.4.2 Installation check

Proper installation procedure for


receiving objective refraction data by
color mapping 32 software can be
checked.

(Startup CV-5000PC software),


select Tools, then select Setting…
select Data in the menu. Now check
RM (AR) Connect is showing “Color
Mapping Software”
This selection can not be changed in
this setup menu.

For changing this setting see chapter


8.1.4.3 Change RM (AR) Connect.

189
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When data is received in


CV-5000PC from Colormap
32 software it shows
sequence number, date and
time. (see picture)

8.1.4.3 Change RM (AR) Connect

When CV-5000PC is already


installed on the Personal
Computer, RM (AR) Connect
can be changed by Windows
Add or Remove Programs.

Select “TOPCON CV-


5000PC (version)” Change.

Modify TOPCONCV-5000PC
window pops up.
Enable or Disable “Ref/Krt
data will be received from
‘Color-Mapping Software’.

190
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.1.5 KR-8100P settings

8.1.5.1 ON-LINE settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select ON-LINE to enter the on-line


menu.

Select CL INPUT NO
Select RS232C FORM EXT2
Select BAUD-RATE 38400
Select RS232C OUT TYPE ALL

(make same selection as in Picture)

Select EXIT to confirm.

8.1.5.2 Output data setting

Power up the KR-8100P(A). Press the


joystick button and wait until the normal
measuring screen is seen. Press the
menu button to enter the SETTING
MENU. Select RS232C “YES”.
Select EXIT with the joystick button to
exit the Setting menu.

191
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.1.5.3 PLACIDO settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select PLACIDO SET to enter the


Placido set menu.

Select PC S/W SETUP and EXECUTE


to calibrate the KR-8100P(A) with the
KRP32 Software.

Select EXIT to confirm.

NOTE:
To calibrate the KR-8100P(A) with the
KRP32 Software, KRP32 software
must be installed, configured and
running.
(see chapter 8.1.6)

8.1.6 Colormap 32 (KRP32) settings

8.1.6.1 KR COM Port setting

Install Color Mapping 32 software


(V1.25) on the PC (Shuttle).
Select Port, leftclick on “Select COM port
for KR”.

192
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Select the Com port on which the KR-


8100P(A) is connected to the Personal
Computer.

(Find out in Windows Device Manager


which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

NOTE:
Be aware of conflicting Com Ports or
Com ports already in use by other
devices like in i-base.

(Selected COM port is example)

8.1.6.2 Setup calibration data

After installation of the KRP32 software


and when Com port selection for KR-
8100P is made, the following window
pops up.

For setup calibration data see chapter


8.1.5.3 Placido settings.

8.1.6.3 Select instrument

Select Port, leftclick on “Select


Instrument” and select “KP8000/38K”.

NOTE: (optional)
To be sure that all Com ports are
configured leftclick on “Reset COM port”.

193
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2 KR-8100P  Colormap 32  CV-5000  i-base


IMPORTANT NOTE: In this setup Objective refraction data will be send directly from
ColorMap 32 software to CV-5000.

8.2.1 Purpose of setup

This connection will see to it that Objective data is send from KR-8100P to CM32 and then
directly to CV-5000. Kerato data is send from KR-8100 to CM32 and then to i-base.

This setup (chapter 8.2) will prevent that Objective data first goes to i-base and then to CV-5000.
If you require Objective data first to i-base and then to CV-5000 an extra handling will be needed
by click in i-base on “set phoropter head”. However this (alternative) setup will be described in
chapter 8.3

Notice the double set of cables between PC and CV-5000(L) needed for this setup.

PC with: Kerato data KB-50/KB-10


KR-8100P Objective data CV-5000
(see 8.2.3.1)
CM32
Objective &
Kerato data Kerato data*

Bi-directional connection
Not possible i-Base
(see 8.2.3.2)
(see NOTE)
Subjective data
Possibility 2 Objective data**
(see chapter 6.1)
Final data**
Lensmeter data**
Kerato data **
Lensmeter CL/EZ Possibility 1
(see chapter 4.1)
*) Setting made in the KR8000P plug-in (in i-base) will see to it that Objective data from
CM32 to i-base is blocked, Only Kerato data is send to i-base (see 8.2.9.2 #3)
**) Setting made in the TCSerial plug-in (in i-base) will see to it that beside Subjective
data, also Objective, Final, Kerato and Lensmeter data (when available)will be
received from CV-5000 in i-Base. These data can also be blocked. However these
data will never be blocked when send from i-base to CV-5000. (see 8.2.8.2 #4)
Be aware that Kerato data can be received double in i-base in this setup

NOTE: CL-200/2800 or EZ-200 connected to KR-8100P in this setup will unfortunately not work.
CM32 software will not pass the CL/EZ data to CV-5000. The CL-200/2800 or EZ-200
can be directly connected to the Compact power supply (1) or directly to i-base (2).

194
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

KR-8100P PC (shuttle) with CM32 & i-Base Compact power supply unit

out 1 2 4 5 6 3

Notice the double set of cables (3 & 6) between PC and CV-5000 Compact power supply

8.2.2 Cable connection

KR-8100P to PC Shuttle
Option 1) Connect cable 1 (item 1830008, order Topcon) from KR-8100P (port OUT) to a serial
port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 1 (item 1830008, order Topcon) to KR-8100P (port OUT) and extend
this cable with a Serial to USB converter cable 2 (Item 1830095, order Topcon) to a USB port of
the Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the
Personal Computer).

PC Shuttle to CPS For data from CM32 to CV-5000


Option 1) Connect cable 3 (item 1830008, order Topcon) from COMM port 1,2,3 or 4 (see
chapter 1.4 power supply unit description) of the CV-5000 Compact power supply to a serial port
of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 3 (item 1830008, order Topcon) on COMM port 1,2,3 or 4 (see chapter
1.4 power supply unit description) of the CV-5000 Compact power supply, and extend this cable
with a Serial to USB converter cable 4 (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

PC Shuttle to CPS For data from CV-5000 to i-Base


Option 1) Connect cable 6 (item 1830008, order Topcon) from COMM port 1,2,3 or 4 (see
chapter 1.4 power supply unit description) of the CV-5000 Compact power supply to a serial port
of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 6 (item 1830008, order Topcon) on COMM port 1,2,3 or 4 (see chapter
1.4 power supply unit description) of the CV-5000 Compact power supply, and extend this cable
with a Serial to USB converter cable 5 (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

195
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.3 Compact power supply dip-switch settings

8.2.3.1 Dip-switch settings for cable using Colormap 32 software

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : OFF 2 : OFF *


Using COM port 2 Dip-switch bank 2 (DSW2) 3 : OFF 4 : OFF *
Using COM port 3 Dip-switch bank 2 (DSW2) 5 : OFF 6 : OFF *
Using COM port 4 Dip-switch bank 2 (DSW2) 7 : OFF 8 : OFF *

*) NOTE: The dip-switches should be both OFF for the Comm. port in use. Although the Compact
power supply is connected to a PC, the dip-switch setting of the Compact power supply
unit must be set accordingly like connected with a KR instrument. For the Compact
power supply it seems that the data is coming from a KR instrument.

8.2.3.2 Dip-switch settings for cable using i-base software

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : ON 2 : OFF

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : ON 4 : OFF

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : ON 6 : OFF

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : ON 8 : OFF

8.2.3.3 Dip-switch settings bank 1

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

196
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.4 KB-10 settings

Dip-switch settings Dip-switch bank 2 (DSW2)

Dip-switch bank 2 (DSW2) 1 : OFF

Note1: After changing dip-switch 2 : ON


settings restart the KB-10 by
switching the ON/OFF button 3 : ON
on the Compact Power Supply.
4 : OFF (2400)
Note2: After power up the KB-10 just
use the Dial button once to 5 : OFF
make the connection Bi-
directional, in order to receive
data from i-base by
“Set Phoropter (Head)”

Dial button

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the SFT button down while
pressing the PRG button. Select with
the dial “SETTING” and push the SET
button. Go with the F2 button to page
P.4 Now select RM/KR CONNECT
“ONLINE” Confirm OK by pressing
the F4 button, then press SET to
restart the KB-10.

Dial and SET button

197
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.5 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Other in the menu.

1) Select Print Switch “Export


only” or “Print and Export”.

2) Select RS-232C Baud Rate


“2400bps”. Press OK, press EXIT
and press YES to restart the
KB-50.

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Data in the menu. Select RM (AR)
Connect “Online” Press OK, press
EXIT and press YES to restart the
KB-50.

198
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.6 KR-8100P settings

8.2.6.1 ON-LINE settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select ON-LINE to enter the on-line


menu.

Select CL INPUT NO
Select RS232C FORM EXT2
Select BAUD-RATE 38400
Select RS232C OUT TYPE ALL

(make same selection as in Picture)

Select EXIT to confirm.

8.2.6.2 Output data setting

Power up the KR-8100P(A). Press the


joystick button and wait until the normal
measuring screen is seen. Press the
menu button to enter the SETTING
MENU. Select RS232C “YES”.
Select EXIT with the joystick button to
exit the Setting menu.

199
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.6.3 PLACIDO settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select PLACIDO SET to enter the


Placido set menu.

Select PC S/W SETUP and EXECUTE


to calibrate the KR-8100P(A) with the
KRP32 Software.

Select EXIT to confirm.

NOTE:
To calibrate the KR-8100P(A) with the
KRP32 Software, KRP32 software
must be installed, configured and
running.
(see chapter 8.2.7)

200
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.7 Colormap 32 (KRP32) settings

8.2.7.1 KR COM Port setting

Install Color Mapping 32 software


(V1.25) on the PC (Shuttle).
Select Port, leftclick on “Select COM port
for KR”.

Select the Com port on which the KR-


8100P(A) is connected to the Personal
Computer.

(Find out in Windows Device Manager


which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

NOTE:
Be aware of conflicting Com Ports or
Com ports already in use by other
devices like in i-base.

(Selected COM port is example)

8.2.7.2 Setup calibration data

After installation of the KRP32 software


and when Com port selection for KR-
8100P is made, the following window
pops up.

For setup calibration data see chapter


8.2.6.3 Placido settings.

201
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.7.3 CV COM port setting

Install Color Mapping 32 software


(V1.25) on the PC (Shuttle).
Select Port, leftclick on “Select COM port
for CV”.

Select the Com port on which the CV-


5000 Compact power supply is
connected to the Personal Computer.

(Find out in Windows Device Manager


which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

Select CV Format “NEW FORMAT”.

NOTE:
Be aware of conflicting Com Ports or
Com ports already in use by other
devices like in i-base.
(Selected COM port is example)

8.2.7.4 Select instrument

Select Port, leftclick on “Select


Instrument” and select “KP8000/38K”.

NOTE: (optional)
To be sure that all Com ports are
configured leftclick on “Reset COM port”.

202
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.8 i-base settings for CV-5000 (Compact power supply connection)

8.2.8.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

203
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “Topcon Serial Refraction


Instruments” Plug-in.
Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“TCSerial” plug-in in Acquisitions of
i-base.

Note:
When the “TCSerial” Plug-in is not
active yet, right click on the plug-in to
ENABLE

204
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.8.2 Serial port setup

Double-Click on the TCSerial plug-in.

1) Select the proper Com port on


which the CV-5000 Compact
Power Supply is connected to the
Personal Computer. Be aware of
the Comports already in use by
CM32. (see 8.2.7.1 & 8.2.7.3)

1a) (Find out in Windows Device


Manager which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

2) Enable this Comm port.

3) Select the Pre-setting for use with


CV compact PS KB-50 / KB-10
the selected com port in the picture is an example
4) Enable (obligatory)

When you want to receive ALL


other data besides Subjective
data, like Lensmeter data, Final
data and Objective data in i-base
send by KB-50/KB-10.

Disable

When you want to receive ONLY


Subjective data in i-base send by
KB-50/KB-10. No other data like
Lensmeter data Final data and
Objective data will be received in
i-base.

Disable (obligatory)

If NOT disabled NO data at all will


be received in i-base from KB-50/
KB-10.
(selection in picture is example)
5) Important: Click on OK to confirm
all settings.

205
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.8.3 Advanced settings

Do NOT change these Pre-settings.

Unless you think is necessary and


know what you are doing !

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

Do NOT change these Pre-settings.

KB-50/KB-10 and Compact power


supply and i-base must set to the
same baud rate of 2400 !

Note:
You can select baud rate 9600 only if
necessary. But then KB-50/KB-10
and Compact Power supply must
also be set to 9600 ! (All 3 of them)

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

206
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the TCSerial plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

207
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.9 i-base settings for Colormap 32 (KRP32) connection

8.2.9.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

208
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “KR8000P” Plug-in.


Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“KR8000P” plug-in in Acquisitions of
i-base.

Note:
When the “KR8000P” Plug-in is
not active yet, right click on the plug-
in to ENABLE

209
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.9.2 KR8000P setup

Double-Click on the KR8000P


plug-in.

1) Application path:
C:\Program Files\TOPCON\
KRP32\KRP32.EXE
(by standard installation)

Or Exactly the same path (folder)


where KRP32.EXE is installed.

2) File location:
C:\Program Files\TOPCON\
KRP32
(by standard installation)

Or Exactly the same path (folder)


where the permanently saved files
are located.

3) (obligatory)

By Disable the Objective data, the


Objective data from CM32 is not
received by i-base.
(because this Objective data is
directly send from CM32 to CV-
5000 in this setup).

4) Enable (obligatory)

By enable “Capture the


permanently saved files” i-base
will find Cornea Map data at the
File location(2), and showing the
Cornea Map in i-base by opening
a link. The Cornea map data from
CM32 will be stored at the File
location(2). (You have to save
this data file manually in CM32).

5) Important: Click on OK to confirm


all settings.

210
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the KR8000P plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

211
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.2.10 Decimal symbol setting in Windows for i-base

Open Windows Start menu.

Click on settings,
Click on Control Panel,
Click on Regional and
Language Options.

Click on “Customize”

212
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Select as an Decimal symbol a


“Comma” (not a dot).

213
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3 KR-8100P  Colormap 32  i-base  CV-5000

8.3.1 Purpose of setup

This connection will see to it that Objective and Kerato data is send from KR-8100P to CM32 and
then to i-base. From i-base this Objective data can be send back and forward to CV-5000.

This setup (chapter 8.3) will prevent that Objective data directly goes to CV-5000. If you require
that the Objective data directly goes from CM32 to CV-5000 then see this (alternative) setup
described in chapter 8.2

Schematic setup:

PC with: KB-50/KB10
KR-8100P CV-5000

Objective & CM32


Kerato data Objective & Kerato data*

Not possible i-Base Bi-directional connection


(see NOTE)
Subjective data
Possibility 2 Objective data**
(see chapter 6.1) Final data**
Lensmeter data**
Lensmeter CL/EZ Possibility 1
(see chapter 4.1)

*) Setting made in the KR8000P plug-in (in i-base) will see to it that Objective and
Kerato data is send from CM32 to i-base. (Objective not blocked) (see 8.3.9.2 #3)
**) Setting made in the TCSerial plug-in (in i-base) will see to it that beside Subjective
data, also Objective, Final and Lensmeter data (when available) will be received
from CV-5000 in i-Base. These data can also be blocked. However these
data will never be blocked when send from i-base to CV-5000. (see 8.3.8.2 #4)
Be aware that Objective data can be received double in i-base in this setup.

NOTE: CL-200/2800 or EZ-200 connected to KR-8100P in this setup will unfortunately not work.
CM32 software will not pass the CL/EZ data to CV-5000. The CL-200/2800 or EZ-200
can be directly connected to the Compact power supply (1) or directly to i-base (2).

214
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

KR-8100P PC (shuttle) with CM32 & i-Base Compact power supply unit

out 1 2 5 6

8.3.2 Cable connection

KR-8100P to PC Shuttle
Option 1) Connect cable 1 (item 1830008, order Topcon) from KR-8100P (port OUT) to a serial
port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 1 (item 1830008, order Topcon) to KR-8100P (port OUT) and extend
this cable with a Serial to USB converter cable 2 (Item 1830095, order Topcon) to a USB port of
the Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the
Personal Computer).

PC Shuttle to CPS For data from CV-5000 (KB-50/KB-10) to i-Base


Option 1) Connect cable 6 (item 1830008, order Topcon) from COMM port 1,2,3 or 4 (see
chapter 1.4 power supply unit description) of the CV-5000 Compact power supply to a serial port
of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 6 (item 1830008, order Topcon) on COMM port 1,2,3 or 4 (see chapter
1.4 power supply unit description) of the CV-5000 Compact power supply, and extend this cable
with a Serial to USB converter cable 5 (Item 1830095, order Topcon) to a USB port of the
Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the PC)

215
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.3 Compact power supply dip-switch settings

Using COM port 1 Dip-switch bank 2 (DSW2) 1 : ON 2 : OFF

Using COM port 2 Dip-switch bank 2 (DSW2) 3 : ON 4 : OFF

Using COM port 3 Dip-switch bank 2 (DSW2) 5 : ON 6 : OFF

Using COM port 4 Dip-switch bank 2 (DSW2) 7 : ON 8 : OFF

RS-232C Format Dip-switch bank 1 (DSW1) 1 : ON 2 : OFF 3 : OFF (NEW)

Baud rate Dip-switch bank 1 (DSW1) 4 : OFF (2400)

216
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.4 KB-10 settings

Dip-switch settings Dip-switch bank 2 (DSW2)

Dip-switch bank 2 (DSW2) 1 : OFF

Note1: After changing dip-switch 2 : ON


settings restart the KB-10 by
switching the ON/OFF button 3 : ON
on the Compact Power Supply.
4 : OFF (2400)
Note2: After power up the KB-10 just
use the Dial button once to make 5 : OFF
the connection Bi-directional, in
order to receive data from i-base
by “Set Phoropter (Head)”

Dial button

8.3.5 KB-50 settings

(Power up the Power supply unit),


Wait until refraction screen is seen.
Hold the shift button down while
pressing the menu button. Select
Other in the menu.

1) Select Print Switch “Export


only” or “Print and Export”.

2) Select RS-232C Baud Rate


“2400bps”. Press OK, press EXIT
and press YES to restart the
KB-50.

217
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.6 KR-8100P settings

8.3.6.1 ON-LINE settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select ON-LINE to enter the on-line


menu.

Select CL INPUT NO
Select RS232C FORM EXT2
Select BAUD-RATE 38400
Select RS232C OUT TYPE ALL

(make same selection as in Picture)

Select EXIT to confirm.

8.3.6.2 Output data setting

Power up the KR-8100P(A). Press the


joystick button and wait until the normal
measuring screen is seen. Press the
menu button to enter the SETTING
MENU. Select RS232C “YES”.
Select EXIT with the joystick button to
exit the Setting menu.

218
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.6.3 PLACIDO settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select PLACIDO SET to enter the


Placido set menu.

Select PC S/W SETUP and EXECUTE


to calibrate the KR-8100P(A) with the
KRP32 Software.

Select EXIT to confirm.

NOTE:
To calibrate the KR-8100P(A) with the
KRP32 Software, KRP32 software
must be installed, configured and
running.
(see chapter 8.3.7)

8.3.7 Colormap 32 (KRP32) settings

8.3.7.1 KR COM Port setting

Install Color Mapping 32 software


(V1.25) on the PC (Shuttle).
Select Port, leftclick on “Select COM port
for KR”.

219
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Select the Com port on which the KR-


8100P(A) is connected to the Personal
Computer.

(Find out in Windows Device Manager


which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

NOTE:
Be aware of conflicting Com Ports or
Com ports already in use by other
devices like in i-base.

(Selected COM port is example)

8.3.7.2 Setup calibration data

After installation of the KRP32 software


and when Com port selection for KR-
8100P is made, the following window
pops up.

For setup calibration data see chapter


8.3.6.3 Placido settings.

8.3.7.3 Select instrument

Select Port, leftclick on “Select


Instrument” and select “KP8000/38K”.

NOTE: (optional)
To be sure that all Com ports are
configured leftclick on “Reset COM port”.

220
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.8 i-base settings for CV-5000 (Compact power supply connection)

8.3.8.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

221
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “Topcon Serial Refraction


Instruments” Plug-in.
Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“TCSerial” plug-in in Acquisitions of
i-base.

Note:
When the “TCSerial” Plug-in is not
active yet, right click on the plug-in to
ENABLE

222
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.8.2 Serial port setup

Double-Click on the TCSerial plug-in.

1) Select the proper Com port on


which the CV-5000 Compact
Power Supply is connected to the
Personal Computer. Be aware of
the Comports already in use by
CM32. (see 8.2.7.1 & 8.2.7.3)

1a) (Find out in Windows Device


Manager which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

2) Enable this Comm port.

3) Select the Pre-setting for use with


CV compact PS KB-50 / KB-10 the selected com port in the picture is an example

4) Enable

When you want to receive ALL


other data besides Subjective
data, like Lensmeter data, Final
data and Objective data in i-base
send by KB-50/KB-10.

Disable

When you want to receive ONLY


Subjective data in i-base send by
KB-50/KB-10. No other data like
Lensmeter data Final data and
Objective data will be received in
i-base.

Disable (obligatory)

If NOT disabled NO data at all will


be received in i-base from KB-50/
KB-10.

5) Important: Click on OK to confirm (selection in picture is example)


all settings.

223
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.8.3 Advanced settings

Do NOT change these Pre-settings.

Unless you think is necessary and


know what you are doing !

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

Do NOT change these Pre-settings.

KB-50/KB-10 and Compact power


supply and i-base must set to the
same baud rate of 2400 !

Note:
You can select baud rate 9600 only if
necessary. But then KB-50/KB-10
and Compact Power supply must
also be set to 9600 ! (All 3 of them)

Every instrument has it’s own Pre-


setting in i-base Serial port setup.
It is advisable NOT to change them.

224
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the TCSerial plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

225
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.9 i-base settings for Colormap 32 (KRP32) connection

8.3.9.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

226
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “KR8000P” Plug-in.


Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“KR8000P” plug-in in Acquisitions of
i-base.

Note:
When the “KR8000P” Plug-in is
not active yet, right click on the plug-
in to ENABLE

227
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.9.2 KR8000P setup

Double-Click on the KR8000P


plug-in.

1) Application path:
C:\Program Files\TOPCON\
KRP32\KRP32.EXE
(by standard installation)

Or Exactly the same path (folder)


where KRP32.EXE is installed.

2) File location:
C:\Program Files\TOPCON\
KRP32
(by standard installation)

Or Exactly the same path (folder)


where the permanently saved files
are located.

3) Enable
By enable “Use data from
temporary received files”
Objective data is send to i-base
automatically. This (temporary)
data file is overwritten every time
CM32 is making this data file. If
you do not want to overwrite this
file, use “permanently saved
files”,but than the file must be
manually saved in order to
receive kerato data in i-base.

4) Enable (obligatory)

By enable “Capture the


permanently saved files” i-base
will find Cornea Map data at the
File location(2), and showing the
Cornea Map in i-base by opening
a link. The Cornea map data from
CM32 will be stored at the File
location(2). (You have to save
this data file manually in CM32).

5) Important: Click on OK to confirm


all settings.
228
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the KR8000P plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

229
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.3.10 Decimal symbol setting in Windows for i-base

Open Windows Start menu.

Click on settings,
Click on Control Panel,
Click on Regional and
Language Options.

Click on “Customize”

230
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Select as an Decimal symbol a


“Comma” (not a dot).

231
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4 KR-8100P  Colormap 32  CV-5000PC  i-base

8.4.1 Purpose of setup

This connection will see to it that Objective and Kerato data is send from KR-8100P to CM32 and
then directly to CV-5000PC. From CV-5000PC this Objective data can be send back and forward
to i-base.
This prevent an extra handling by start-up i-base and have to click the button “set phoropter head”
By sending data from KR8100P immediately Objective data is available in CV-5000PC.

Schematic setup:

PC with:
KR-8100P

CM32
Objective &
Kerato data
Not possible Kerato data*
(see NOTE)
i-Base

Possibility 2
(see chapter 6.1)

CV-5000PC
Objective &
Lensmeter CL/EZ Possibility 1 Kerato data**
(see chapter 4.1) CPS

*) Setting made in the KR8000P plug-in (in i-base) will see to it that Objective data from
CM32 to i-base is blocked, Only Kerato data is send to i-base (see 8.4.7.2 #3)

**) Setting made in the installation setup TOPCON CV-5000PC will see to it that
Objective data & Kerato data will be received directly in CV-5000PC from CM32.
(see 8.4.4.1 #3)

Be aware that Kerato data can be received double in i-base in this setup.

NOTE: CL-200/2800 or EZ-200 connected to KR-8100P in this setup will unfortunately not work.
CM32 software will not pass the CL/EZ data to CV-5000. The CL-200/2800 or EZ-200
can be directly connected to the Compact power supply (1) or directly to i-base (2).

232
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

KR-8100P PC (shuttle) with CM32 & i-Base

out 1 2

8.4.2 CV-5000PC  Compact Power Supply…………………..See Chapter 2.3

8.4.3 Cable connection

KR-8100P to PC Shuttle
Option 1) Connect cable 1 (item 1830008, order Topcon) from KR-8100P (port OUT) to a serial
port of the Personal Computer. (The serial port is usually COMM 1).

Option 2) Connect cable 1 (item 1830008, order Topcon) to KR-8100P (port OUT) and extend
this cable with a Serial to USB converter cable 2 (Item 1830095, order Topcon) to a USB port of
the Personal Computer. (Do not forget to install the proper Serial to USB converter driver on the
Personal Computer).

233
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4.4 CV-5000PC settings

8.4.4.1 Setup CV-5000PC

Install the latest CV-5000PC software


available on the TOPCON FTP site.

(If CV-5000PC software is already


installed on the PC see chapter
8.4.4.3).

When execute the installation


program, the “setup TOPCON CV-
5000PC window pops up.

Enable “Ref/Krt data will be received


from ‘Color-Mapping Software’, (same
as in picture).

8.4.4.2 Installation check

Proper installation procedure for


receiving objective refraction data by
color mapping 32 software can be
checked.

(Startup CV-5000PC software),


select Tools, then select Setting…
select Data in the menu. Now check
RM (AR) Connect is showing “Color
Mapping Software”
This selection can not be changed in
this setup menu.

For changing this setting see chapter


8.4.4.3 Change RM (AR) Connect.

234
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When data is received in


CV-5000PC from Colormap
32 software it shows
sequence number, date and
time. (see picture)

8.4.4.3 Change RM (AR) Connect

When CV-5000PC is already


installed on the Personal
Computer, RM (AR) Connect
can be changed by Windows
Add or Remove Programs.

Select “TOPCON CV-


5000PC (version)” Change.

Modify TOPCONCV-5000PC
window pops up.
Enable or Disable “Ref/Krt
data will be received from
‘Color-Mapping Software’.

235
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4.4.4 Import/Export setup

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Other in the menu. Select
Page 2.
Select “Export only” or “Print and
Export”. Click OK, click EXIT and
click YES to restart CV-5000PC.

8.4.4.5 File Format Version

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Other in the menu. Select
Page 2.
Select “Version 1.00 (KB-1D Ver.
1.40 Compatible”. Click OK, click
EXIT and click YES to restart CV-
5000PC.

8.4.4.6 RS-232C Baud Rate

(Startup CV-5000PC software),


Select Tools, then select Setting…
select Other in the menu. Select
Page 2.
Select “2400 bps”. Click OK, click
EXIT and click YES to restart CV-
5000PC.

236
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4.5 KR-8100P settings

8.4.5.1 ON-LINE settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select ON-LINE to enter the on-line


menu.

Select CL INPUT NO
Select RS232C FORM EXT2
Select BAUD-RATE 38400
Select RS232C OUT TYPE ALL

(make same selection as in Picture)

Select EXIT to confirm.

8.4.5.2 Output data setting

Power up the KR-8100P(A). Press the


joystick button and wait until the normal
measuring screen is seen. Press the
menu button to enter the SETTING
MENU. Select RS232C “YES”.
Select EXIT with the joystick button to
exit the Setting menu.

237
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4.5.3 PLACIDO settings

Power up the KR-8100P(A) while


holding down the menu button to enter
the INITIAL MENU.

(Navigate with the “joystick button”,


make selections with the “print button”)

Select PLACIDO SET to enter the


Placido set menu.

Select PC S/W SETUP and EXECUTE


to calibrate the KR-8100P(A) with the
KRP32 Software.

Select EXIT to confirm.

NOTE:
To calibrate the KR-8100P(A) with the
KRP32 Software, KRP32 software
must be installed, configured and
running.
(see chapter 8.4.6)

8.4.6 Colormap 32 (KRP32) settings

8.4.6.1 KR COM Port setting

Install Color Mapping 32 software


(V1.25) on the PC (Shuttle).
Select Port, leftclick on “Select COM port
for KR”.

238
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Select the Com port on which the KR-


8100P(A) is connected to the Personal
Computer.

(Find out in Windows Device Manager


which Com port is used.
Usually the serial port of the PC is
Com 1, however the selected com
port in the picture is an example)

NOTE:
Be aware of conflicting Com Ports or
Com ports already in use by other
devices like in i-base.

(Selected COM port is example)

8.4.6.2 Setup calibration data

After installation of the KRP32 software


and when Com port selection for KR-
8100P is made, the following window
pops up.

For setup calibration data see chapter


8.4.3.3 Placido settings.

8.4.6.3 Select instrument

Select Port, leftclick on “Select


Instrument” and select “KP8000/38K”.

NOTE: (optional)
To be sure that all Com ports are
configured leftclick on “Reset COM port”.

239
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4.7 i-base settings for Colormap 32 (KRP32) connection

8.4.7.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

240
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “KR8000P” Plug-in.


Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“KR8000P” plug-in in Acquisitions of
i-base.

Note:
When the “KR8000P” Plug-in is
not active yet, right click on the plug-
in to ENABLE

241
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4.7.2 KR8000P setup

Double-Click on the KR8000P


plug-in.

1) Application path:
C:\Program Files\TOPCON\
KRP32\KRP32.EXE
(by standard installation)

Or Exactly the same path (folder)


where KRP32.EXE is installed.

2) File location:
C:\Program Files\TOPCON\
KRP32
(by standard installation)

Or Exactly the same path (folder)


where the permanently saved files
are located.

3) (obligatory)

By Disable the Objective data, the


Objective data from CM32 is not
received by i-base.
(because this Objective data is
directly send from CM32 to CV-
5000 in this setup).

4) Enable (obligatory)

By enable “Capture the


permanently saved files” i-base
will find Cornea Map data at the
File location(2), and showing the
Cornea Map in i-base by opening
a link. The Cornea map data from
CM32 will be stored at the File
location(2). (You have to save
this data file manually in CM32).

5) Important: Click on OK to confirm


all settings.

242
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

When finishing making all selections


right-click on the KR8000P plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

243
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4.8 i-base settings for CV-5000PC connection

8.4.8.1 Plug-in installation

Go in windows to
Select IMAGEnet i-base and click on
Change/Remove.

Select Modify and click Next>

244
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Install “CV for Windows” Plug-in.


Click on Next>

Start i-base with the proper Key-tag


plugged in a free USB port of your
Personal Computer.

As a result you will see a active


“CVWin” plug-in in Acquisitions of
i-base.

Note:
When the “CVWin” Plug-in is not
active yet, right click on the plug-in to
ENABLE

245
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4.8.2 CV Win setup

IMPORTANT: First install CV-


5000PC Software on the PC
before you do the CV Win setup.

Double-Click on the CVWin plug-


in.
Select the Pre-setting “CV-5000PC
Software”. Make exactly the same
settings as shown in the picture.

Make sure that the Program Path is


exactly the same as shown in the
picture.

Make sure that the Import and Export


path is exactly the same as shown
below in red.

C:\Documents and Settings\All Users\Application Data\TOPCON\CV-5000PC\DATA\ImpExp

When finishing making all selections


right-click on the CVWin plug-in
and RESTART the plug-in.

With RESTART you activate all the


changes in selections of the plug-in.

246
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

8.4.9 Decimal symbol setting in Windows for i-base

Open Windows Start menu.

Click on settings,
Click on Control Panel,
Click on Regional and
Language Options.

Click on “Customize”

247
© TOPCON EUROPE MEDICAL BV
CV-5000 Series Connection Manual Version 1.04

Select as an Decimal symbol a


“Comma” (not a dot).

248
© TOPCON EUROPE MEDICAL BV