Adiabatic calorimetry has been the which the heat is released (kinetics).
cornerstone of chemical process In both cases there can be a
safety for the last 30 years. More significant increase in the sample
recently, adiabatic calorimetry has pressure and the combination of
been widely used in measuring the the temperature and pressure rise
potential of thermal runaway of can lead to an explosion. Therefore
Li-ion cells. These cases are similar in adiabatic calorimeters are generally
that they require measuring both designed to be much more robust
the amount of heat released than many of other types of
(thermodynamics) and the rate at calorimeters.
APTAC® 264
ARC® 244
2
When looking at a thermal runaway, NETZSCH has taken adiabatic taking the best technology and
whether it is within a battery cell or calorimetry to new levels. The making it better. With the
part of a chemical process in a plant, instruments described herein are the patented VariPhi™ system, not
the reaction produces heat which best in their class. Building continu- only can adiabatic test be done
increases the temperature of the ously on the original system more accurately but it also opens
reaction and further increases the developed by Dow in the late 1970’s up new areas of calorimetry not
rate. There can come a point when and commercialized by Columbia possible with the original ARC®
the rate of heat release from the Scientific Industries and then ADL/ units. The systems have been
reaction exceeds the rate at which the TIAX (acquired by NETZSCH in 2009), enhanced to do more and answer
heat can be lost to the surrounding NETZSCH continues its heritage of more questions.
environment- the point at which
thermal runaway inevitable.
Obviously the worst case is when
there is little or no ability for the heat
to be lost to the environment. Any
heat, even small amounts of heat,
cause the temperature to rise and
accelerates the reaction producing
even more heat. This is the adiabatic
condition and it is this precise
condition that adiabatic calorimeters
can safely measure in the small scale.
It is this same condition that can be
found inside large processing vessels
as well as inside a Li-ion pack running
inside a laptop computer.
ARC® 254
3
Understanding Adiabatic Reaction Calorimetry
Method
The energy release from chemical For decades accelerating rate calorim- Sample Container
reactions (decomposition etc.) is a eters have been widely used by
point of focus in chemical research, researchers in this field, offering the
battery research and other industries. capability to measure temperatures,
When energy is generated by a enthalpy changes and pressure
thermally induced chemical reaction changes quantitatively.
and the heat transfer to the outside is
smaller than the generated amount, The use of adiabatic systems has the
runaway reactions can occur. In the advantage that no heat loss is allowed
worst case this can cause catastrophic from the sample; the behavior in real
effects (explosions). Adiabatic calorim- large scale chemical reactors can
eters are ideal tools for analyzing such therefore be simulated (worst case
questions as they simulate the worst scenario).
case scenario with no heat exchange
with the surroundings. Adiabatic System: No heat in – no heat out
Top
Tube Heater
Technique
Magnetic Containment
Stirring Vessel
Bottom
4
Measurement Results
Thermal runaway reaction of a 20% solution of di‐tert. butyl peroxide (DTBP) in toluene.
5
Accelerating Rate Calorimeter ARC® 244 –
The Original DOW Designed for Basic ARC® Testing
ARC® 244
6
Operating Modes
··
runaway reactions
Iso-Fixed / Iso-Track for studying
storage conditions/auto-catalytic
··
reactions (iso-aging technique)
Ramp mode for fast screening of
unknown samples
7
Accelerating Rate Calorimeter ARC® 254 –
Advanced Technology for Specialized Testing
ARC® 254
8
Operating Modes
··
runaway reactions of chemicals or
··
battery samples Scanning and isothermal modes
Iso-Fixed / Iso-Track for studying allow detection of exothermic and
storage/conditions/auto-catalytic endothermic effects; similar to DSC
·· ··
reactions (iso-aging technique) method
Ramp mode for screening unknown In‐situ battery cycling using
samples isothermal calorimeter much
preferred over adiabatic
9
Automatic Pressure Tracking Adiabatic Calorimeter 264 –
Premier Solution for Big Sample/High Pressure Testing
APTAC® 264
10
Operating Modes
··
runaway reactions (no pressure balancing)
··
Iso-Fixed/Iso-Track for studying
storage/conditions/auto-catalytic Scanning and isothermal modes
··
reactions (iso-aging technique) allow detection of exothermic and
Ramp mode for screening unknown endothermic effects; similar to DSC
··
samples method
Fire-exposure mode for simulation
of thermodynamic and kinetic
parameters of the reaction in a
single test
Detection of smallest
Highest sensitivity
caloric effects even at lowest Extended range of applications
(0.002 K/min)
onset temperature
11
VariPhi™: A Patented Solution for Low Phi Operation
and Reduced Testing Time
Theory
Sample holders absorb some of the the mass of the stored material. Since
Thermocouples
energy from the reaction. How much this is not always true it is important to
heat is absorbed depends upon the be able to run multiple tests at varying
mass and heat capacity of the sample thermal inertia. The VariPhi™ gets its
container. The ratio between the name because the user can quickly
product of the sample mass and heat and easily run low and high thermal
capacity to the product of the inertia (˝Phi˝) tests in the same
Sample container mass and heat capacity is calorimeter using a small and inher-
Heater
what is known as thermal inertia. ently safer sample size.
Guard
Heater
When the sample mass is large
compared to the container mass, the Formula for thermal inertia (Φ-factor):
thermal inertia approaches one. These
tests are often called “low phi” tests. mc x Cp,c
Φ = 1+ ––––––––
They are important because most ms x Cp,s
industrial scale processing or storage
conditions are low thermal inertia mc = mass of the container
Diagram of the VariPhi™ Cp,c = heat capacity of the container
conditions. The mass of a large storage ms = mass of the sample
container is very small compared to Cp,s = heat capacity of the sample
12
Standard Modes such as Heat-Wait- Scanning Mode Isothermal Mode
·· ··
Search, Iso-Fixed or Iso-Track
··
Reduces test time by 75% or more True isothermal calorimeter is
··
Compensation for heat lost to the without loss of sensitivity possible on chemical samples and
·· ··
sample container during test Sample heat capacity can be battery cells
Thermal inertia or Phi (Φ) factor can measured as a function of 3D battery sensors are available
·· ··
be defined temperature based on the VariPhi™ concept.
Low Phi testing with small samples Endothermic transition can be They allow the study of charge/
accurately measured discharge of battery inside
calorimeter
Fire-Exposure Mode
·· ····
available to meet different test
requirements. Tube-type vessels are Spherical vessels with a wall 1 ml to 130 ml for spherical vessels
available for running smaller APTAC® thickness between 0.4 mm and 5.1 0.1 ml to 9 ml for tube-type vessels
samples or energetic materials and for mm are made of glass, Hastelloy,
use with solids and pastes in the Inconel, stainless steel, tantalum
··
VariPhi™. For battery testing, specific and titanium Pressure
····
holders are designed for common Tube-type vessels with a wall
commercial sizes. Custom vessels can thickness of 0.4 mm and 0.7 mm Up to 1000 bar for spherical vessels
also be designed. are made of stainless steel and Up to 200 bar for tube-type vessels
titanium
13
Selecting the Best Calorimetry for Your Applications
Technical Features
Pressure range (Standard) 0 bar to 200 bar 0 bar to 200 bar 0 bar to 140 bar
Injection collection
Optional Optional Standard
vessel for reaction gas
Process Safety;
Process Safety; ERSD;
Applications Battery;
Comparison to legacy data Operate as ARC® 254
Energetic Material
* Sample volumes shown are typical and not limits. Sample vessels for the ARCs and APTAC are interchangeable.
The choice of sample vessels can influence test results. See product manual or spare parts list for more details.
14
Configurable Options to Match Testing Needs
VariPhiTM Heater
P2
Syringe
Sample Pump
VariPhiTM Heater P1
Sample
Vessel Syringe
Injection
Guard
Venting Sample
Heaters
Vessel
Stir Bar
Stirring Magnet
Simple
Vent
Graphite Nitrogen
Stirringe TCs
Bearing
Vacuum
Exhaust Magnadrive
Stirrer
High pressure
syringe pump P4
Sample
P2
Single shot
Drop-out Injection
TCs pot
P1
Sample
Vessel
VariPhiTM Heater
TCs
Nitrogen
Vacuum
The schematics show the Magnadrive Relief
different options available stirrer valve
to the product line. P3
Exhaust
15
Intelligent Software Solutions
Measurement Task
The measurement software for the Accelerating Rate Calorimeters and the
APTAC® offers easy setup of a measurement, selection of the measurement mode
and follow-up of the test progress. All raw data signals are accessible on the screen,
switches can be set with one mouse click and simple plots of the running measure-
ments can be generated. The software stores data in binary and ASCII formats,
making it easy to load results into other software packages for advanced analysis.
··
application
Intuitive Windows set-up of testing
methods with input checks to reduce
··
typos
Methods can be saved and recalled
··
for later use
On-line help, comprehensive and
easy to understand, available
··
throughout the software
Data is automatically saved to
··
hard-drive throughout the run
Intelligent Firmware monitors the
status and health of key operating
··
sensors and active components
Safety integrated with hardware and
··
firmware controls
Seamless integration of data files to
··
Proteus®
Analysis, NETZSCH Advanced
Software Tools or 3rd party software
Software screenshot for APTAC® 264
16
Evaluation Software
17
Typical Applications
Accelerating Rate Calorimeters by NETZSCH can be used for the thermal analysis
of solid or liquid chemicals or for gas/liquid, liquid/liquid, gas/solid, and liquid/
solid mixtures. They can also be used for process simulation of batch and semi-
batch reactions, fire exposures, emergency relief venting, and physical properties
measurement.
Energetic Materials
Analysis of the phase changes and decomposition behavior of ammonium nitrate using the
VariPhi™-option
Autocatalytic Behavior
18
Vented Tests
19
Typical Battery Applications
Effect of state-of-charge on the thermal stability of a prototype 18650 lithium ion batteries
Cell Heater Power and Cell Current versus Time for an Isothermal Battery Cycling Test on ARC® 254
Lithium-ion battery
20
Nail Penetration
Air cylinder
Exchangable needle
Battery
21
Expertise in Service
····
stands for comprehensive support and Hotline service
expert, reliable service, before and Preventive maintenance
··
after sale. Our qualified personnel Calibration service
from the technical service and On-site repairs with emergency
····
application departments are always service for NETZSCH components
available for consultation. Moving / exchange service
··
Technical information service
In special training programs tailored Spare parts assistance
for you and your employees, you will
learn to tap the full potential of your
instrument.
22
Our Expertise – Applications Laboratories
23
The NETZSCH Group is a mid-sized, family-owned German company engaging in
the manufacture of machinery and instrumentation with worldwide production,
sales, and service branches.
The three Business Units – Analyzing & Testing, Grinding & Dispersing and
Pumps & Systems – provide tailored solutions for highest-level needs. Over
3,300 employees at 210 sales and production centers in 35 countries across
the globe guarantee that expert service is never far from our customers.