Because emka TECHNOLOGIES has a policy of continuous product improvement, we reserve the right to change design and specifications without notice.
it50 is a robust and rugged isometric transducer for
measuring forces in the range 0-50g. force range 0 to 50 g
Its stainless steel body contains four solid-state gauges displacement 0.89 mm/50g
wired as a Wheatstone bridge. sensitivity@ Vcc=5V 0.1 mV/g
excitation voltage 10 V DC
It comes with a DIN or a LEMO connector.
full scale output 5mV
maximum accepted 100g (0.1N)
overload
In the emkaBATH setups combined error (linearity, <0.2%
hysteresis, repeatability,
The same it50 is used for setups with manual and temperature)
automated tissue tensioning.
temperature drift <0.03% / °C
The force must be measured at a 90° angle from the long temperature range -10 to 70°C
All contents Copyright © October 2015 emka TECHNOLOGIES SAS. All rights reserved.
axis (the hole on the transducer body must face upwards). input impedance 1000 Ω
The upper tissue hook is connected to the it50 transducer output impedance 1000 Ω
via a swivel, which permits rotatory movement. dimensions cylinder:
10mm, 80mm length
References
emkaBATH_IT_50 with connector for emkaBATH2 or 4
USB_IT_50 with connector for emka usbAMP
IT_50 with connector for emka analog amplifier
isolated-organ.emka.fr