Вы находитесь на странице: 1из 5

Journal of Chromatography B, 998 (2015) 40–44

Contents lists available at ScienceDirect

Journal of Chromatography B
journal homepage: www.elsevier.com/locate/chromb

Validation of a multi-analyte HPLC-DAD method for determination of


uric acid, creatinine, homovanillic acid, niacinamide, hippuric acid,
indole-3-acetic acid and 2-methylhippuric acid in human urine
Daniela Remane a , Soeren Grunwald b,1 , Henrike Hoeke c , Andrea Mueller d ,
Stefan Roeder e , Martin von Bergen b,d,f , Dirk K. Wissenbach b,∗
a
Institute of Forensic Medicine, University Hospital Jena, Jena, Germany
b
Department of Metabolomics, Helmholtz Centre for Environmental Research – UFZ, Leipzig, Germany
c
Department of Pharmaceutical and Medicinal Chemistry, Institute of Pharmacy, University of Leipzig, Leipzig, Germany
d
Department of Proteomics, Helmholtz Centre for Environmental Research – UFZ, Leipzig, Germany
e
Department of Environmental Immunology, Helmholtz Centre for Environmental Research – UFZ, Leipzig, Germany
f
Department of Biotechnology, Chemistry and Environmental Engineering Aalborg University, Aalborg, Denmark

a r t i c l e i n f o a b s t r a c t

Article history: During the last decades exposure sciences and epidemiological studies attracts more attention to unravel
Received 4 September 2014 the mechanisms for the development of chronic diseases. According to this an existing HPLC-DAD method
Received in revised form 15 June 2015 for determination of creatinine in urine samples was expended for seven analytes and validated. Creati-
Accepted 17 June 2015
nine, uric acid, homovanillic acid, niacinamide, hippuric acid, indole-3-acetic acid, and 2-methylhippuric
Available online 24 June 2015
acid were separated by gradient elution (formate buffer/methanol) using an Eclipse Plus C18 Rapid
Resolution column (4.6 mm × 100 mm).
Keywords:
No interfering signals were detected in mobile phase. After injection of blank urine samples signals for
HPLC-DAD
Validation the endogenous compounds but no interferences were detected. All analytes were linear in the selected
Creatinine calibration range and a non weighted calibration model was chosen. Bias, intra-day and inter-day pre-
Uric acid cision for all analytes were below 20% for quality control (QC) low and below 10% for QC medium and
Hippuric acid high. The limits of quantification in mobile phase were in line with reported reference values but had to
Urine be adjusted in urine for homovanillic acid (45 mg/L), niacinamide 58.5(mg/L), and indole-3-acetic acid
(63 mg/L). Comparison of creatinine data obtained by the existing method with those of the developed
method showing differences from −120 mg/L to +110 mg/L with a mean of differences of 29.0 mg/L for 50
authentic urine samples. Analyzing 50 authentic urine samples, uric acid, creatinine, hippuric acid, and
2-methylhippuric acid were detected in (nearly) all samples. However, homovanillic acid was detected
in 40%, niacinamide in 4% and indole-3-acetic acid was never detected within the selected samples.
© 2015 Elsevier B.V. All rights reserved.

1. Introduction and the development of chronic diseases such as allergy [1,2]. The
Exposome is defined as the sum of all exposures that an individ-
During the last decades exposure research attracts more atten- ual is subjected to [3]. Its impact on human health is assessed in
tion as several studies found relationships between the “Exposome” epidemiological cohort studies.
According to this a prospective mother–child cohort study,
“Lifestyle and environmental factors and their influence on New-
born Allergy risk” (LiNA), was initiated 2005 by the Helmholtz
Abbreviations: CREA, creatinine; UA, uric acid; HVA, homovanillic acid; NA, Center for Environmental Research (Leipzig, Germany) and by the
niacinamide; HA, hippuric acid; IAA, indole-3-acetic acid; 2-MHA, 2-methylhippuric St. Georg Hospital in Leipzig, Germany [4,5]. In addition to, e.g.,
acid; QC, quality control.
environmental monitoring, bio analysis of urine and blood is con-
∗ Corresponding author. Fax: +49 341 235 451211.
ducted for every mother–child pair focusing on I. immunological
E-mail address: dirk.wissenbach@ufz.de (D.K. Wissenbach).
1
Current address: IDT Biologika GmbH, Dessau-Rosslau, Germany. parameters; II. external exposures such as pollutants, endocrine

http://dx.doi.org/10.1016/j.jchromb.2015.06.021
1570-0232/© 2015 Elsevier B.V. All rights reserved.
D. Remane et al. / J. Chromatogr. B 998 (2015) 40–44 41

Table 1
Covered analytes (sorted by retention time), calibration (Cal) level concentrations [mg/L], quality control sample (QC) concentrations [mg/L], observed bias [%], intra-day
precision (intra-day) and inter-day precision (inter-day), and limit of quantification (LOQ) [mg/L] in MP A and urine.

Analytes Calibration level [mg/L] QC low QC medium QC high LOQ

1 2 3 4 5 6 7 c [mg/L] Bias RSD c [mg/L] Bias RSD c [mg/L] Bias RSD MP A urine
[%] [%] [%] [mg/L] [mg/L]

Intra Inter Intra Inter Intra Inter


day [%] day [%] day [%] day [%] day [%] day [%]

UA 100 200 300 450 650 850 1000 150 −5.5 1.8 5.5 500 −2.5 0.3 10.7 900 −0.7 1.3 3.2 25 25
CREA 100 500 750 1000 1250 1500 2000 600 1.0 4.7 5.8 1100 −0.1 5.4 5.4 1700 −0.1 2.8 3.9 100 100
HVA 2.5 5.0 10 25 50 75 100 15 14.9 4.6 17.2 60 1.1 3.2 6.1 90 −2.0 2.4 7.0 2.5 45
NAA 5.0 10 15 25 65 80 100 20 −1.5 5.1 9.0 50 3.6 4.2 4.9 90 1.0 3.0 4.7 5.0 58.5
HA 50 75 100 250 500 750 1000 200 0.4 3.1 4.3 400 0.5 2.5 2.4 900 0.3 1.3 2.3 50 50
IAA 10 25 40 55 70 85 100 20 12.8 3.7 16.9 50 9.3 8.0 10.2 80 6.0 3.0 3.4 10 63
2-MHA 20 30 40 50 70 85 100 25 13.7 8.0 10.4 60 3.7 6.8 8.6 80 4.5 1.7 4.0 20 22.5

UA = uric acid; CREA = creatinine; HVA = homovanillic acid; NAA = niacinamide; HA = hippuric acid; IAA = indole-3-acetic acid; 2-MHA = 2-methylhippuric acid.

disruptors, and food ingredients III. the metabolome. Using these 2.2. HPLC-DAD Analysis
approaches it was shown that the development of chronic diseases
are strongly associated with prenatal and early lifetime exposures A Prominence modular HPLC-system (Shimadzu, Jena,
[4–7]. Germany) consisting of two binary pumps, an autosampler
In this study an existing HPLC-DAD-method for the analysis of (tray temperature: 20 ◦ C), a column oven (set to 40 ◦ C), and a diode
urine creatinine (CREA) concentration should be replaced by a fully array detector (detection wavelength from 200 to 400 nm) was
validated new HPLC-DAD-method. In addition to CREA, the new used in this study. Elution was performed on a ZORBAX Eclipse
method should also cover uric acid (UA), homovanillic acid (HVA), Plus C18 Rapid Resolution 4.6 mm × 100 mm (Agilent, Waldbronn,
niacinamide (NA), hippuric acid (HA), indole-3-acetic acid (IAA) as Germany) at a total flow rate of 1.0 mL/min. The gradient was as
well as 2-methylhippuric acid (2-MHA). While urine CREA con- follows 0.00–1.00 min 0% mobile phase (MP) B; 1.00–8.00 min–40%
centration is used for normalization of urine bio analysis results, MP B; 8.00–10.00 min–100% MP B; 10.00–15.00 min holding at
the other analytes showed associations either to immune response 100% MP B; 15.00–16.50 min–0% MP B; 16.50–25.00 min equilibra-
(UA, NA), to external exposures such as toluene (HA) or xylene (2- tion at 0% MP B. MP A consisted of 50 mM ammonium formate and
MHA), to the gut microbiome (IAA, HA), to nutrition (UA,NA,HA) or 5 mM octanesulfonic acid in 99% water and 1% methanol (adjusted
the endogenous response to occurring mental stress (HVA) [8–16]. to pH 6.55 with formic acid). MP B consisted of 50 mM ammonium
Even if not every single compound is necessarily associated to a sin- formate and 5 mM octanesulfonic acid in 100% methanol (adjusted
gle exposure (e.g., HA as biomarker for toluene, but also for phenolic to pH 6.55 with formic acid). The injection volume was 10 ␮L;
food compounds) they all provide metabolomic marker for certain after each injection the needle was washed using 200 ␮L of an
lifestyles (including nutrition, mental status, gut microbiome, and acetonitrile/water (1:1; v:v) mixture. An inhouse reference library
also exposures). So far only a few methods covering some of the was created by injecting each compound (c = 1000 mg/L) and stor-
above mentioned analytes in urine were described. For example ing the obtained spectra after background subtraction. Qualitative
Zuo et al. described a HPLC-UV method for CREA and UA, while data were obtained by reference library matching. All quantitative
Shirao et al. analyzed CREA, HVA and UA using miscellar electroki- analyses were achieved by determination of the area under the
netic chromatography [17,18]. Antunes et al. validated a HPLC-UV curve and calibration models using extracted chromatograms
method for CREA and UA in dried urine spots [11]. A liquid chro- (233 nm wavelength).
matography mass spectrometry method for CREA, UA, HA, and IAA
was developed by Gu et al. [19]. To reduce sample amount and
2.3. Preparation of solutions, calibration and quality control
work load the new method was supposed to cover all the above
samples
mentioned analytes by using a simple and cost effective sample
preparation and analysis technique.
With the exception of HA stock solutions of all analytes
(2000 mg/L for UA, CREA, 1000 mg/L for HVA, NA, IAA, 2-MHA) were
prepared in MP A, and stored at 8 ◦ C until use. Stock solution of
2. Materials and methods HA (1000 mg/L) was freshly prepared every day in 0.3 M potassium
hydroxide solution according to Kim et al. [20]. Working solutions
2.1. Chemicals, reagents, and standards were prepared by dilution of the stock solution with the respec-
tive amount of solvent. Calibration samples were prepared at seven
Formic acid, octanesulfonic acid, phosphoric acid, potassium different calibration levels (Cal. 1–7) (Table 1) by dilution of the
hydroxide, and IAA (each analytical grade), methanol and acetoni- appropriate amount of working solution with MP A. Three internal
trile (LC–MS grade) were obtained by Merck (Darmstadt, Germany). quality control (QC) samples (low, med, high, concentrations given
Ammonium formate, UA, CREA, HVA, NA, HA, and 2-MHA (each ana- in Table 1) were prepared by dilution of the appropriate amount of
lytical grade) were obtained from Fluka (Taufkirchen, Germany). a separately prepared working solution with MP A.
Ultrapure water was produced by a MilliporeQ (MerckMillipore,
Darmstadt, Germany) water purification system. External quality 2.4. Sample origin and sample preparation
control (QC) samples for CREA and UA (KU 01/12 A and B) were
obtained from Referenzinstitut für Bioanalytik (Bonn, Germany). Human blank urine samples (n = 6) were aliquoted and stored
CREA concentrations were 1020.0 and 1080.0 mg/L, UA concentra- at −20 ◦ C until usage. The calibration, internal QC samples (50 ␮L
tions were 209.0 and 254.0 mg/L for KUA 1/12 A and B, respectively. solution), urine samples (50 ␮L urine), and external QC samples
42 D. Remane et al. / J. Chromatogr. B 998 (2015) 40–44

(50 ␮L urine) were worked up by dilution with 450 ␮L MP A and


directly analyzed by the HPLC-system.

2.5. Validation experiments

Validation was performed according to international and


national guidelines with respect to selectivity, carry over, limit of
quantification (LOQ), linearity, accuracy, precision, freeze thaw and
processed sample stability [21,22].
Selectivity for exogenous compounds (HA and 2-MHA) was
tested with six different blank urine samples. Selectivity for
endogenous compounds (CREA, UA, HVA, NA, IAA) was tested by
six fold injection of MP A. To test for carry over effects, MP A was
directly analyzed after the injection of Cal. 7 in three independent
runs. The LOQ was set at the lowest calibrator concentration, if
this could reliably be quantified (RSD and bias <20%) and the sig-
nal to noise ratio was above 10:1. A seven point calibration model,
covering characterized concentrations was selected [15]. Linearity Fig. 1. A: Extracted chromatogram (233 nm wavelength) of blank MP A showing no
and calibration model was tested by analyzing six replicates of Cal. interferences. B: Extracted chromatogram (233 nm wavelength) of QC level low (in
1–7. For accuracy and precision, freshly prepared QC samples (low, MP A) showing UA (1), CREA (2), HVA (3), NAA (4), HA (5), IAA (6), and 2-MHA (7).

med, high) were analyzed in duplicate on each of eight days. Accu-


racy was calculated in terms of bias as the percent deviation of the the endogenous compounds but no other signals were detected.
mean calculated concentration from the corresponding theoretical Hence, the method was shown to be selective. An example chro-
concentration [21]. Intra-day precision (intra-day) and inter-day matogram of spiked MP A (QC level low) is given in Fig. 1B. The LOQ
precision (inter-day) were calculated as relative standard deviation was determined for all analytes at the lowest calibrator concentra-
(RSD), according to Peters et al. [21].. For QC low, the acceptance tion (see Table 1) in MP A. No carry over effects were observed,
criterion was set to 20% CV and RSD, respectively; for QC med and which shows that the needle flushing is sufficient in this method.
high to 15% CV and RSD, respectively. To estimate benchtop stabil- All analytes were linear in the selected calibration range and a non
ity of processed sample under the conditions of analysis, a sample weighted calibration model was chosen. The analysis of internal
of Cal 4 was prepared in an adequate amount (500 ␮L) and placed QC samples showed acceptable results for all analytes at all moni-
into the autosampler. This sample was analyzed in triplicate on tored concentrations (see Table 1). A minimal decrease (5%) of the
eight days and resealed after the daily measurement. Freeze–thaw peak areas during the eight day lasting bench top stability tests was
stability was tested with a spiked urine sample (UA 225 mg/L, only monitored for CREA. However, using this kind of auto sam-
CREA 900 mg/L, HVA 22.5 mg/L, NAA 22.5 mg/L, HA 225 mg/L, IAA pler, the maximal amount of vials was 80. As the sample to sample
49.5 mg/L, and 2-MHA 45 mg/L) after one and two freeze (23 h at analysis time was 25 min, benchtop stability must only be guaran-
−20 ◦ C) and thaw (1 h at room temperature) cycles. Stabilities were teed for 80 × 25 min = 33.3 h. Variations of the peak areas after two
tested by comparing absolute peak areas of each analyte. Analytes freeze/thaw cycles were between 5.4% (2-MHA) and 9.1% (HA) with
were stable, if the average result of the stability samples was within no significant tendency except for HA, where a decrease of 8.9% was
90–110% of the corresponding average result of the control samples detected. Nevertheless, as all stability parameters were within the
and the 90% confidence interval of the stability samples was within acceptance criteria, stability was shown for all analytes.
80–120% of the corresponding average value of the control samples. According to literature sufficient LOQ concentrations were
obtained in MP A as given in Table 1 [15]. Unfortunately these could
2.6. Proof of applicability not be reproduced in authentic urine matrix. A reliable identifica-
tion using library matching, which is in addition to the signal to
External QC samples (for UA and CREA, KU 01/12 sample noise ratio, a prerequisite of a detection and quantification was only
A and B) were used for routine application. Additional three possible for UA, CREA, HA, and 2-MHA. The obtained LOQs for HA
different internal QC samples (low, medium, high concentra- and CREA were similar as or double than those reported by Antunes
tion) were prepared in MP A as described above and also used et al. also using an HPLC-DAD analyzer but three fold higher than
for this purpose. In order to proof the applicability, 50 LiNA those reported for CREA, UA; HA, and 3-IAA using a LC–MS method
cohort study urine samples were analyzed. Results for CREA were [11,19]. The spectra of HVS, IAA, and NSA were different to those
statistically compared to the results obtained with the estab- recorded in the library using MP A although the retention time was
lished method for analyzing CREA and to the literature data
[15,23–27]. The participation on the LiNA study was voluntary and
approved by the Ethics Committee of the University of Leipzig
(file refs. 046-2006; 160-2008; 160b/2008; 144-10-31052010;
113-11-18042011; 206-12-02072012). All samples were handled
according to the institutional protocol and regulations.

3. Results and discussion

As endogenous urinary metabolites are unavoidable substances


in blank urine, selectivity was firstly tested in MP A. As most
of the compounds provided an absorbance maximum of approx. Fig. 2. Extracted chromatogram (233 nm wavelength) of an authentic urine sample
showing UA (1), CREA (2), HVA (3), NAA (4) and, HA (5). CREA concentration was
233 nm, this wavelength was used for all analytes. No interfer-
676.7 mg/L. Creatinine normalized concentrations were 586.8 (UA), 140.4 (HVA),
ing signals were detected at the corresponding analyte retention 137.4 (NAA), and 663.6 (HA) mg/L/g/L. For IAA (6) the concentration was below
times (Fig. 1A). After injection of blank urine samples signals for LOQ, while 2-MHA (7) was not detected.
D. Remane et al. / J. Chromatogr. B 998 (2015) 40–44 43

Fig. 3. Bland–Altman plot/comparison of quantitative CREA results (50 authentic urine samples) obtained by the presented and the established method.

Table 2
Quantification results from n = 50 LiNA urine samples with frequency of quantification [%], concentration range, median, and mean after creatinine normalization [mg/L/g/L]
and literature data of healthy male (m) and female (f) Caucasians for all analytes.

Analytes Frequency of Concentration range Median concentration Mean concentration Literature data
quantification [%] [mg/L/g/L] [mg/L/g/L] [mg/L/g/L]

Min Max Concentration [mg/L/g/L] Reference

UA 100 106.5 2278.6 541.4 597.5 310.9 (m/f) [24]


CREA 100 240 1935 761.7 835.0 1400.0 (m/f) [25]
HVA 40 16.9 327.5 85.9 104 9.0 (m/f) [26]
NAA 4 26.7 47.1 36.9 36.9
HA 100 79.2 2983.5 599.3 696.0 328.0 (f); 363.1 (m/f) [25,27]
IAA 0 0 0 0 0 5.3 (m/f) [25]
2-MHA 98 8.4 69.2 23.5 28.0 22.0 (f) [24]

UA = uric acid; CREA = creatinine; HVA = homovanillic acid; NAA = niacinamide; HA = hippuric acid; IAA = indole-3-acetic acid; 2-MHA = 2-methylhippuric acid.

identical. This might be explained by co-elution of broad matrix negligible difference of approximately 3% may be explained by dif-
peaks leading to a high background, which cannot be countervailed ferent calibrations, calibration models, weighting or dilution biases.
by using automatic background compensation algorithms. A library Only a few times NAA (4%) and no IAA (0%) were quantified
search against reference spectra obtained in matrix was not suitable in the 50 LiNA samples as indicated in Table 2. Even if urine LOQ’s
to compensate this problem. Nevertheless, unacceptable quantifi- seemed unsuitable to cover normal urine concentrations those ana-
cation results for six different urine specimen spiked at the LOQ lytes should be included in the described method. If high urine
were obtained for HVA, NAA, and 3-IAA and therefore, LOQs were concentrations were observable for those compounds this infor-
adopted (Table 1). mation can be used for stratification in statistical analyses. HVA
was quantified in 40% of the 50 LiNA urine samples. Interestingly
the median and mean concentrations were approx. 10 times higher
in comparison to the reference values described by Guneral and
3.1. Proof of applicability
Bachmann [26]. In comparison to literature data higher values were
observed for UA, HVA, and HA. For CREA lower concentrations were
External QC samples (for UA and CREA, KU 01/12 sample A and
determined as depicted in Table 2. As reference and obtained data
B) were used during the validation process and for testing routine
were normalized to CREA, this could be one explanation for higher
application. Bias and precision results for both QC samples were
values. Obtained 2-MHA concentrations were in the same range as
within the acceptable range of ±15%. Therefore, calibration in MP
described by Inoue et al. [24]. For NAA no reference data for healthy
A was shown approvable for the analysis of UA and CREA.
Caucasian male or females (>18 years old) was available.
50 authentic LiNA urine samples were analyzed with the
described method. A representative chromatogram for an authen-
tic urine samples is given in Fig. 2. Quantitative CREA results were Summary
compared with those obtained using the method described by
Rolle-Kampczyk et al. [23]. As shown in Fig. 3 low differences (rang- The analytical portfolio for analysis of cohort study urine sam-
ing from −120 mg/L to +110 mg/L) were observed. These differ- ples was successfully extended by additional six analytes. All
ences were not influenced by the absolute concentration of CREA. compounds showed sufficient validation parameters according to
However, the new method (mean concentration = 942.4 mg/L) pro- recent guidelines. Routine application was tested for 50 authentic
vided 29.0 mg/L lower mean CREA concentrations – given in the urine samples. Comparison of the new method with the established
Bland–Altman plot (mean of differences, X-axis)– in comparison method showed a high level of agreement for quantitative CREA
to the established method (mean concentration = 971.4 mg/L). This results.
44 D. Remane et al. / J. Chromatogr. B 998 (2015) 40–44

Acknowledgments [13] A.L. Kau, P.P. Ahern, N.W. Griffin, A.L. Goodman, J.I. Gordon, Nature 474 (2011)
327.
[14] M. Li, M. Wang, S.M. Donovan, Semin. Reprod. Med. 32 (2014) 74.
The authors thank their colleagues Sven Baumann, Ralph Fel- [15] D.S. Wishart, C. Knox, A.C. Guo, R. Eisner, N. Young, B. Gautam, D.D. Hau, N.
tens, Susanne Pfeiffer, and Brigitte Winkler for scientific discussion Psychogios, E. Dong, S. Bouatra, R. Mandal, I. Sinelnikov, J. Xia, L. Jia, J.A. Cruz,
and technical assistance. E. Lim, C.A. Sobsey, S. Shrivastava, P. Huang, P. Liu, L. Fang, J. Peng, R. Fradette,
D. Cheng, D. Tzur, M. Clements, A. Lewis, A. De Souza, A. Zuniga, M. Dawe, Y.
Xiong, D. Clive, R. Greiner, A. Nazyrova, R. Shaykhutdinov, L. Li, H.J. Vogel, I.
References Forsythe, Nucleic. Acids Res. 37 (2009) D603.
[16] R.W. Pero, Curr. Clin. Pharmacol. 5 (2010) 67.
[1] P.J. Lioy, S.M. Rappaport, Environ. Health Perspect. 119 (2011) A466. [17] Y. Zuo, C. Wang, J. Zhou, A. Sachdeva, V.C. Ruelos, Anal. Sci. 24 (2008) 1589.
[2] E.S. Epel, E.H. Blackburn, J. Lin, F.S. Dhabhar, N.E. Adler, J.D. Morrow, R.M. [18] M.K. Shirao, S. Suzuki, J. Kobayashi, H. Nakazawa, E. Mochizuki, J. Chromatogr.
Cawthon, Proc. Natl. Acad. Sci. U. S. A. 101 (2004) 17312. B Biomed. Sci. Appl. 693 (1997) 463.
[3] C.P. Wild, Int. J. Epidemiol. 41 (2012) 24. [19] L. Gu, X. Wang, Y. Zhang, Y. Jiang, H. Lu, K. Bi, X. Chen, J. Sep. Sci. 37 (2014)
[4] G. Herberth, M. Bauer, M. Gasch, D. Hinz, S. Röder, S. Olek, T. Kohajda, U. 1058.
Rolle-Kampczyk, M. von Bergen, U. Sack, M. Borte, I. Lehmann, Maternal and [20] K.M. Kim, G.N. Henderson, X. Ouyang, R.F. Frye, Y.Y. Sautin, D.I. Feig, R.J.
cord blood miR-223 expression associates with prenatal tobacco smoke Johnson, J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 877 (2009) 2032.
exposure and low regulatory T-cell numbers, J. Allergy Clin. Immunol. 133 (2) [21] F.T. Peters, M. Hartung, M. Herbold, G. Schmitt, T. Daldrup, F. Musshoff,
(2014) 543–550, e4. Toxichem Krimtech 76 (2009) 185.
[5] D. Hinz, M. Bauer, S. Roder, S. Olek, J. Huehn, U. Sack, M. Borte, J.C. Simon, I. [22] I.H.T. Guideline, Website: http://www.ich.org/fileadmin/Public Web Site/ICH
Lehmann, G. Herberth, LS Group, Allergy 67 (2012) 380. Products/Guidelines/Quality/Q2 R1/Step4/Q2 R1 Guideline.pdf (2010).
[6] K. Weisse, I. Lehmann, D. Heroux, T. Kohajda, G. Herberth, S. Roder, M. von [23] U.E. Rolle-Kampczyk, M. Rehwagen, U. Diez, M. Richter, O. Herbarth, M. Borte,
Bergen, M. Borte, J. Denburg, Clin. Exp. Allergy 42 (2012) 1337. Arch. Environ. Health 57 (2002) 326.
[7] K.M. Junge, F. Hornig, G. Herberth, S. Roder, T. Kohajda, U. Rolle-Kampczyk, M. [24] O. Inoue, K. Seiji, T. Suzuki, T. Watanabe, H. Nakatsuka, H. Satoh, M. Ikeda,
von Bergen, M. Borte, J.C. Simon, D. Heroux, J.A. Denburg, I. Lehmann, Clin. Bull. Environ. Contam. Toxicol. 47 (1991) 204.
Immunol. 152 (2014) 68. [25] S. Bouatra, F. Aziat, R. Mandal, A.C. Guo, M.R. Wilson, C. Knox, T.C. Bjorndahl,
[8] M.S. Gersch, R.J. Johnson, Nephrol. Dial. Transplant. 21 (2006) 3046. R. Krishnamurthy, F. Saleem, P. Liu, Z.T. Dame, J. Poelzer, J. Huynh, F.S. Yallou,
[9] G. Ng, E.M. Chau, Y. Shi, Arch. Immunol. Ther. Exp. (Warsz) 58 (2010) N. Psychogios, E. Dong, R. Bogumil, C. Roehring, D.S. Wishart, PLoS One 8
273. (2013) e73076.
[10] C.M. John, R. Ramasamy, G. Al Naqeeb, A.H. Al-Nuaimi, A. Adam, Curr. Med. [26] F. Guneral, C. Bachmann, Clin. Chem. 40 (1994) 862.
Chem. 19 (2012) 5181. [27] N.G. Psihogios, I.F. Gazi, M.S. Elisaf, K.I. Seferiadis, E.T. Bairaktari, NMR
[11] M.V. Antunes, C.G. Niederauer, R. Linden, Clin. Biochem. 46 (2013) 1276. Biomed. 21 (2008) 195.
[12] T. Sumiyoshi, T. Yotsutsuji, M. Kurachi, H. Itoh, K. Kurokawa, O. Saitoh,
Neuropsychopharmacology 19 (1998) 70.

Вам также может понравиться