Вы находитесь на странице: 1из 10

ARTICLE IN PRESS

Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438

Improvement of high-voltage performance of acceleration


tubes by cleaning the walls with a high-pressure water jet
S. Takeuchi*, T. Nakanoya, H. Kabumoto, T. Yoshida
Japan Atomic Energy Research Institute, Tokai Research Establishment, 2-4 Shirakata-shirane, Tokai, Naka, Ibaraki 319-1195, Japan

Received 25 January 2003; received in revised form 17 June 2003; accepted 26 June 2003

Abstract

We cleaned electrostatic accelerator tubes by applying a high-pressure water jet and examined their high-voltage
performances at 1 and 3 MV. The cleaning was very effective in reducing discharge activities at their rated voltages. We
did some experimental investigations with the tubes and their ceramic insulators. We found that removal of
microparticles loosely bound on the vacuum-side ceramic surfaces had an important effect in eliminating the discharge
activities.
r 2003 Elsevier B.V. All rights reserved.

PACS: 52.75.Di; 52.80.Vp; 81.05.Mh; 81.65.Cf

Keywords: Electrostatic acceleration tubes; High-voltage performance; High-pressure water-jet rinsing; Loosely bound particles;
Ceramic insulator surfaces

1. Introduction shown in Fig. 1(b), to improve the high-voltage


performance toward 20 MV. The new 1 MV tube
At JAERI-Tokai, a folded-type tandem accel- modules will consist of two 300 mm long tubes
erator 20UR Pelletron (National Electrostatic with 21 13 mm acceleration gaps [2]. Replacing
Corporation) in which 20 1 MV column modules heatable aperture assemblies with additional accel-
are vertically stacked has been running well since eration gaps has been proved to be effective to
1982 [1]. The operational highest terminal voltage improve the high-voltage performance by Ass-
was 18 MV in the beginning and in the vicinity of mann et al. [3] and at the ORNL 25URC tandem
16 MV in recent years. The 1 MV tube modules are accelerator [4,5].
composed of three sections of a 173 mm long tube It is required for the tube replacement program
with 11 13 mm acceleration gaps and a 30 mm to resume stable operation for experiments as soon
thick heatable aperture assembly, as shown in as possible. However, high-voltage tube condition-
Fig. 1(a). We decided to replace the acceleration ing is generally very time consuming for a large
tubes with 300 mm long compressed geometry tubes accelerator system. Our idea to solve this problem
is to carry out cleaning of the tubes before
*Corresponding author.
installation. High Pressure Water Jet Rinsing
E-mail address: takeuchi@tandem.tokai.jaeri.go.jp (HWJR) has become a popular cleaning
(S. Takeuchi). method for superconducting accelerating cavities.
0168-9002/$ - see front matter r 2003 Elsevier B.V. All rights reserved.
doi:10.1016/S0168-9002(03)02080-1
ARTICLE IN PRESS

430 S. Takeuchi et al. / Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438

Fig. 1. High gradient acceleration tubes; (a) three short-tubes (old standard tubes) in a 1 MV module in the present 20UR tandem
accelerator at JAERI-Tokai and (b) two long tubes (so called compressed geometry tubes) for the replacement.

Removing pollution from the cavity surfaces is Our method of tube cleaning is described in
important to eliminate the electron field emission Section 2, investigations about ceramic surfaces in
that causes degradation in cavity quality factors at Section 3, the result of the 1 MV test of short tubes
high fields [6–10]. With respect to the electrostatic in Section 4 and the result of the 3 MV test of long
acceleration tubes composed of titanium electro- tubes in Section 5. The alumina ceramics used in
des and alumina ceramic insulators, the high- the short and long tubes had different origins. The
voltage insulation performance is predominated by physical properties we experimentally obtained
vacuum side ceramic surfaces. Pollution, dirt or from the ceramics are compared in Section 6.
loosely bound ceramic grains could be a source of
discharge activities and need to be removed from
the surfaces and tube interior. We, therefore, 2. Cleaning methods
planned to adopt the HWJR to the new 21 gap
long tubes to remove such sources of discharge High-pressure air jets were used to make water
activities, expecting a substantial reduction of jets as is illustrated in Fig. 2. From the top of the
lengthy high-voltage tube conditioning time and spraying rod, two jet streams of mist came out
an improvement of the high-voltage performance. toward the ceramic walls and electrodes. The
We carried out the following experiments. We nozzle holes were 2 mm in diameter. The spread
examined the cleaning effect of the HWJR on of the jets were about 10 mm in diameter at the
accelerator tube inside walls. As a feasibility study, acceleration tube wall. The air was supplied from
we did a 1 MV high-voltage test with 11 gap short commercially available bottles of air with a
tubes by putting them in a 1 MV module of the capacity of 470 MPa l and passed through 40 C
tandem accelerator. For 21 gap long tubes heaters and a 0.3 mm filter. The air was fed at a
obtained for the tube replacement, we carried out pressure of about 1 MPa and at a flow rate of
a 3 MV high-voltage test at NEC. about 0.6 MPa l/s. Estimated jet speed was
ARTICLE IN PRESS

S. Takeuchi et al. / Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438 431

air. Then the tubes were dried and baked similarly


as the previous short tubes.
We paid special attention to the jet power. In the
tubes, 0.5 mm thick inner electrodes are attached
with small joints. Therefore, the jet power should
be low enough so as not to damage the thin
electrodes.

3. Some investigations on high-pressure water jet


rinsing and ultrasonic cleaning

We investigated the cleaning effect of the HWJR


by spraying an air jet in an acceleration tube and
counting particles in the air from the tube. The
measurement was done with the jet spray machine
(Fig. 2) without feeding water. The flow rate of
sprayed air was measured by an air flow meter. A
particle counter took in a part of the sprayed air at
a flow rate of 0.5 l/min. The spraying rod and
acceleration tube were moved just as in the case of
HWJR. The measurement was done with increas-
Fig. 2. Method of cleaning acceleration tubes with high- ing the air pressure in steps of 0.5 MPa. At each
pressure water jets. pressure increment, the counting rate exponen-
tially decreased after several minutes and became
small enough to stop counting in 10 min. More
250–300 m/s. De-ionized water of 15–18 MO cm particles came out with increasing air pressure
was supplied from a water purifier. The water flow because of blasting power of the air jet itself. Two
rate was 0.2–0.4 l/min. In order to rinse the inside short tubes, which had been unused and aged
thoroughly, the nozzles were tilted by 715 , the about 15 years, were used for the measurements.
spraying rod was moved up and down at a speed One of them was examined without any treatment
of 15 cm/min, and the tube was rotated on a and the other after applying HWJR. The results
turntable at a speed of 10 rev/min. The jet spray are shown in Fig. 3 as a function of air pressure.
machine and a tube drying heater were put in a Much fewer particles came out from the tube
clean booth in order to prevent dust from rinsed with HWJR than the untreated tube. This
accumulating on the wet surfaces after the rinse. indicates that the original wall surfaces were full of
Six 11 gap short tubes were rinsed for about loosely bound particles and were effectively
12 min spending one bottle of air. After the cleaned by the HWJR.
HWJR, the tubes were dried at 50 C for a few In addition, surfaces of ceramic samples were
hours in the clean booth and baked at about observed through microscopes. By taking pictures
200 C for 2 weeks with pumping by a turbo- of a site before and after the HWJR by a laser
molecular pump. The tubes were unused ones in microscope, it was confirmed that many tiny
storage for about 15 years. particles and dirt were removed, as is seen from
Six 21 gap long tubes were cleaned more Fig. 4. However, ceramic grain removals were not
thoroughly, i.e. their inside walls were washed in found in the observation. The ceramic insulators
a 600 W ultrasonic wave bath for 30 min with for the samples used in the experiments were
water and some neutral detergent and rinsed with obtained from NEC, which were the same as the
HWJR for about 25 min spending two bottles of ceramics used for our short tubes.
ARTICLE IN PRESS

432 S. Takeuchi et al. / Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438

Fig. 3. Count of particles in the air from acceleration tubes


during air jet spraying. The tubes were short tubes in different
surface conditions; one as it was, one rinsed with a high-
pressure water jet and one washed with ultrasonic wave.
Particles (>0.3 mm) were measured by using a particle counter
and a gas flow rate meter and with increasing air pressure in
discrete stages.
Fig. 4. Laser microscope images of a ceramic surface (a) before
and (b) after a high-pressure water jet rinse. Tiny points and
Ultrasonic wave cleaning was also effective. dark spots in white circles vanished after the rinse.
Counts of particles in the sprayed air from
short tube walls cleaned by ultrasonic wave are
plotted in Fig. 3. We examined the water used for module (4th module from the bottom) of the
ultrasonic wave cleaning of the long-tube walls by tandem accelerator in place of three existing low-
filtering it through a sheet of filter paper. energy tubes and three high-energy tubes. This
From the observation through a microscope, configuration corresponds to testing two sets of
the numbers were estimated on the order of 1 MV tube modules in parallel. For potential
1.5  106 particle/tube for the particles of 1 to grading, metal oxide on ceramic resisters rated at
10 mm, 1  105 particle/tube for 10–100 mm and 1.2 GO were put on each acceleration gap. The test
1  103 particle/tube for 100 mm or greater. Rela- was done during the conditioning for all the tubes
tively large particles mostly seemed to be metallic just after the replacement. The high tension was
fragments from titanium electrodes. From the supplied only to this 1 MV module by short
observation through a laser microscope, the clean- circuiting all the others. Fig. 5 shows the records
ing result was similar to the one seen in Fig. 4. of the voltage, low-energy side tube vacuum
pressure and X-ray monitor at the pressure vessel
wall. The voltage was smoothly increased up to
4. Result of the high-voltage test of short tubes for 1.07 MV, although several sparks occurred. The
1 MV generating voltmeter and capacitor pickoff signal
were quiet, indicating that continuous or frequent
The six short tubes cleaned with HWJR as discharge activities were not occurring in the tubes.
described in Section 2 were set in a 1 MV column In the records of the vacuum pressures, there were
ARTICLE IN PRESS

S. Takeuchi et al. / Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438 433

Fig. 5. Strip chart recorded during the first high-voltage tensioning for the six short tubes installed in the both low- and high-energy
sides at the 4th 1 MV module of the tandem accelerator. For space considerations, only the low-energy tube vacuum record is shown.

no continuous or frequent outgas spikes due to the metal oxide on ceramic resisters rated at 1.1 GO
discharges in the tubes and the vacuum recovery were used both on the tube and column structures.
was quick after the sparks. There were small The vessel was filled with SF6 gas at a pressure of
increases in the X-ray radiation with a few of the 0.57 MPa. The tube vacuum pressure was the mid
discharges which diminished quickly. The condi- 10 6 Pa range. An X-ray monitor was placed at
tioning time for the cleaned tubes was much 1 m from the pressure vessel wall. The high-voltage
shorter than that for the original tubes. tensioning was done three times over a period of 3
For comparison with tubes not cleaned by days. The strip chart records of the terminal
HWJR, presented in Fig. 6 is a typical old record voltage and tube vacuum pressure are shown in
of voltages during a conditioning done for a series Fig. 7.
of single 1 MV modules in 1980. Small and full In the first day, the voltage was brought up
sparks had happened much more frequently. It is carefully and smoothly increased to the rated
also seen from the records of the X-ray, tube voltage of 3.0 MV in 4 h. There were a few mild
vacuum and capacitor pickoff signal that there were discharges. The associated vacuum outgas recov-
more discharge activities than the sparks seen in the ered quickly. The column structure seemed to be
record of voltage. Much more conditioning was conditioning as much as the acceleration tubes.
required for them to run stably at the rated voltage. The second and third days were eventless, as there
were only a few mild sparks. The voltage was held
stable and calm with a natural small increase from
5. Result of the high-voltage test of long tubes for 3.0 to 3.1 MV during 9 h and from 3.1 to 3.2 MV
3 MV during 9 h, respectively. The very small voltage
fluctuations became even smaller after a minor
The six long tubes cleaned at JAERI as discharge on the third day. With respect to the
described in Section 2 were examined in a 3 MV X-ray radiation, there was only one momentary
testing machine at NEC. For potential grading, increase at 2.1 MV in the first day, which was a
ARTICLE IN PRESS

434 S. Takeuchi et al. / Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438

Fig. 6. Strip chart recorded during a high-voltage conditioning done in 1980 for the 5th 1 MV module of the tandem accelerator. For
space considerations, only the low-energy tube vacuum record is shown.

mild one, well below 10 mSv/h. The X-ray level was 6. Investigations of differences between the two
normally in the ambient level of 0.02–0.05 mSv/h. ceramics
The test accelerator lost charge current was within
the range of 0.4–2.1 mA, which was given by the Although the alumina ceramics in our new long
difference between the charging current and drain tubes come from the same supplier as the older
current. short tubes, they may have been made from a
For comparison, given in Fig. 8 is a chart record different material source and caused a limitation in
of the high-voltage test done at NEC for six long- the high-voltage performance as is seen from Fig.
tubes before delivering to JAERI. These long- 8, according to NEC. We investigated the old and
tubes belonged to the same ceramic batch as the new ceramics, using ceramic samples and old short
tubes used for the test described above. There were and new long tubes.
continuous discharge activities and frequent full From an observation of the surfaces through a
sparks even in the vicinity of 2.9 MV. The lost microscope, the old ceramics contained many
charge current ranged from 5 to 10 mA. grains as large as 40 mm in diameter while the
It is noted that the cleaning had a very big effect new ceramics have small grains, as is seen from the
on improving the high-voltage performance, i.e., SEM images in Fig. 9. The mean grain sizes from
an ideally stable condition was obtained at the the observation were 25 and 14 mm, respectively.
rated voltage with very little high-voltage con- From an experiment of air jet spraying and
ditioning as a result of applying the HWJR to the particle counting done as described in Section 3,
tubes similar to those that suffered from stubborn the curves in Fig. 10 indicate that the new ceramics
discharge activities. in the long tubes had more particles coming off
ARTICLE IN PRESS

S. Takeuchi et al. / Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438 435

Fig. 7. Strip charts recorded during the first high-voltage tensioning for the six long-tubes cleaned with high-pressure water jet rinsing.
From the top, the charts are of the first, second and third days.

easily at a low pressure, 0.5 MPa, than the old 4 mol% for the new ceramics. The diffusion
ceramics in the short tubes. Such loosely bound coefficients estimated by diffusion equation were
microscopic particles on the surfaces are probably of the order of 4  10 4 and 7  10 6 cm2/s,
a source of discharge activities. respectively. These differences seem to be due to
From an experiment of outgas from a sample their grain sizes.
heated by a 850 C boat, it was found that the new In summary, we found that the new ceramics
ceramics contained much more gas and had a were composed of smaller grains and had a lot of
much smaller diffusion coefficient than the old microscopic grains loosely bound on the surfaces
ceramics. The outgas curves are shown in Fig. 11. compared to the old ceramics. These surface
The gas contents estimated from the curves were conditions limit the high-voltage performance of
roughly 0.6 mol% for the old ceramics and the long tubes without wall cleaning.
ARTICLE IN PRESS

436 S. Takeuchi et al. / Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438

Fig. 8. Strip chart recorded of voltage during the high-voltage tensioning test done for the six long tubes not cleaned with high-
pressure water jet rinsing.

Fig. 10. Count of particles in the air from original short and
long tubes during air jet spraying. Particles (>0.3 mm) were
measured by using a particle counter and a gas flow rate meter
and with increasing air pressure in discrete stages.

7. Discussion

Large tandem accelerators built many years ago


at JAERI, ORNL and other institutions did not
attain their design voltages expected from a linear
function of the number of acceleration tube
Fig. 9. Scanning electron microscope images of two different sections. ORNL applied a compressed geometry
ceramic samples: (a) of the old ceramics and (b) of the new acceleration tube to the 25URC tandem accel-
ceramics. erator to exceed its design value [5]. The
ARTICLE IN PRESS

S. Takeuchi et al. / Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438 437

loose microparticles contributed to carrying elec-


trons on the vacuum side surfaces and to emitting
electrons from triple junctions where ceramic
walls, electrodes and vacuum meet. We believe
that secondary electrons are easily emitted from
loose microparticles because they can be saturated
and heated with fewer electrons. Moreover, very
loose microparticles can be highly charged, pulled
out easily, and accelerated by the electric field.
When they collide with surfaces in the vacuum,
electrons, gas and additional microparticles are
emitted. This chain process may develop to a
heavier discharge activity.
There was a clear difference in ceramic grain
Fig. 11. Outgas curves from two different ceramic samples of sizes between the two ceramics. But, the present
the old and new ceramics heated on an 850 C boat in a vacuum
chamber pumped by a 500 l/s pump. Zigzag structures were due
work gives no conclusions for their contributions
to periodic short removals of the sample from the boat for to high-voltage stabilities. Both the cleaned short
background measurements of the vacuum. and long tubes have cleared their rated voltages of
1 and 3 MV, respectively, with a very high
stability. It would be interesting to know the
compressed geometry, which reduced electric field highest voltage one can go with the clean tubes. A
gradients along the ceramic surfaces, was useful to test at a voltage much higher than the rated
hinder discharge activities, although it did not voltages was, regrettably, not allowed by the
solve the problem that the high voltages were not present column conditions.
proportional to the number of tube sections. The
nonlinear effect must be essentially attributed to
the condition that 1 MV tube modules still had 8. Conclusions
significant discharge activities at the rated voltage
of 1 MV, as is seen from Fig. 6. In such a condition A HWJR with highly de-ionized water followed
for the tube modules linked by high impedance by de-gas processing was applied to the cleaning of
voltage dividers, a chain of discharges is easily acceleration tube vacuum side wall surfaces. As a
triggered by a surge originated in a tube module result of experimental investigations, it was found
and a full breakdown occurs below the design that there were millions of microparticles loosely
voltage. In an ideal condition that discharge bound on the ceramic surfaces which were easily
activities are much less in the tube modules, a removed by the cleaning. The result of the high-
chain of discharges is not likely to happen. This voltage test with 11 gap short tubes in a 1 MV
condition was realized with the cleaned long tubes module in the tandem accelerator proved that the
during the 3 MV high-voltage test. It arouses the HWJR is very useful to reduce the discharge
interest whether such a stable condition will be activities at the rated voltage. The result of the
presented at 20 MV in the upcoming high-voltage high-voltage test with 21 gap long tubes for 3 MV
test of the JAERI 20UR column with cleaned long modules confirmed that the HWJR is also effective
tubes. to a larger scale system composed of the long
The clear reduction of discharge activities was a tubes, although their new ceramics consisted of
result of removing loosely bound particles on tube small grains and originally held more microparti-
ceramic. In other words, loose microparticles cles on the wall surfaces compared to the old
dominated the discharge activities. The lost charge ceramics for the short tubes. The HWJR cleaning
current through the long tubes was significantly is, then, a promising method to drastically reduce
decreased by HWJR. These facts suggest that lengthy tube conditioning and to improve the
ARTICLE IN PRESS

438 S. Takeuchi et al. / Nuclear Instruments and Methods in Physics Research A 513 (2003) 429–438

high-voltage performance of a large acceleration [3] W. Assmann, G. Korschinek, H. Munzer,. Nucl. Instr. and
tube system. Meth. 220 (1984) 86.
[4] C.M. Jones, K.A. Erb, D.L. Haynes, J.T. Mitchell, N.F.
Ziegler, J.E. Raatz, R.D. Rathmell, Nucl. Instr. and Meth.
A 268 (1988) 361.
Acknowledgements
[5] M.J. Meigs, C.M. Jones, D.L. Hayes, R.C. Juras, N.F.
Ziegler, Proceedings of the Symposium on N.E. Accel-
We would like to thank Dr. J.A. Ferry and erator Personnel, World Scientific, Singapore, 1988,
Dr. G.A. Norton for kindly giving us ceramic pp. 23–36.
samples for investigations and for carrying out the [6] C.Z. Antoine, B. Bonin, J.M. Cavedon, C. Chaianelli, J.M.
Hisleur, B. Mahut, J.P. Poupeau, Proceedings of the Fifth
high-voltage test for us according to our request.
Workshop on RF Superconductivity, Hamburg, 1991,
pp. 456–462.
[7] S. Takeuchi, T. Ishii, B.J. Min, M. Shibata, Proceedings of
References the Fifth Workshop on RF Superconductivity, Hamburg,
1991, pp. 395–405.
[1] S. Takeuchi, S. Abe, S. Hanashima, K. Horie, N. Ishizaki, [8] P. Kneisel, B. Lewis, L. Turlington, Proceedings of the
S. Kanda, M. Matsuda, I. Ohuchi, H. Tayama, Sixth Workshop on RF Superconductivity, Newport
Y. Tsukihashi, T. Yoshida, AIP Conference Pro- News, 1993, pp. 628–636.
ceedings 473, Heavy Ion Accelerator Technology, 1998, [9] K. Saito, T. Higuchi, T. Suzuki, E. Kako, S. Noguchi, M.
pp. 152–167. Ono, Shishido, Proceedings of the Sixth Workshop on RF
[2] G.A. Norton, Proceedings of the Symposium on N.E. Superconductivity, Gif-sur-Yvette, 1995, pp. 379–383.
Accelerator Personnel, World Scientific, Singapore, 1987, [10] S. Takeuchi, T. Ishii, M. Matsuda, Y. Zhang, T. Yoshida,
pp. 235–245. Nucl. Instr. and Meth. A 382 (1996) 153.

Вам также может понравиться