KMC650 Troubleshooting (X-ray)

TROUBLESHOOTING FOR KMC650(X-RAY) 0


IST 2014

Table of Contents

Unit 1. Tube(Monotank)-----------------------------------------------------------------------------------------------------------------------------------------
2

Unit 2. Drive Board(Monotank)----------------------------------------------------------------------------------------------------------------------------------


5

Unit 3. PWR Board------------------------------------------------------------------------------------------------------------------


9

Unit 4. Main Control Board------------------------------------------------------------------------------------------------------------------


11

TROUBLESHOOTING FOR KMC650(X-RAY) 1


IST 2014

Unit 1. Tube(Monotank)

WARING
You MUST switch off the Unit before replacing the part and ONLY an approved GEMSS Medical technician is qualified to carry out any intervention
on the Unit while it is switched on.

Spcification : Toshiba DF-151SBR

Special Tools : Oscilloscope, wrench set, DMM, Screwdriver set

Definition :
This device emits X-ray by creating electron beam which moves fast and collides with the material.

Common Service Cases :


1. Faulty X-ray generation
2. Damaged device

TROUBLESHOOTING FOR KMC650(X-RAY) 2


IST 2014

How to check :
TP9(IP) - mA Measurement
1. kV, mA Feedback
Scope Value
- Small: 250mV 1mA
- Large: 100mV 10mA

TP8(EP) - kVp Measurement

Scope Value
1V DC 25KV

GND

TROUBLESHOOTING FOR KMC650(X-RAY) 3


IST 2014

2. Tube Filament

SMALL FILAMENT ON (N/F, LARGE FILAMENT ON


M/F) (S/F, BOOST, RAD

CS-CS (about 0.3~0.5Ω)


CL-CL(about
Yellow Cable
0.3~0.5Ω)
(Small Filament)
White Cable

1) Check Filament Preheat status.


2) Measure Resistance of CS-CS, CL-CL (Filament in Contact : 0.3 ~ 0.5 Ω)

TROUBLESHOOTING FOR KMC650(X-RAY) 4


IST 2014

Unit 2. Drive Board(Monotank)


WARING
You MUST switch off the Unit before replacing the part and ONLY an approved GEMSS Medical technician is qualified to carry out any intervention
on the Unit while it is switched on.

Spcification : Drive Board

Special Tools : wrench set, DMM, Screwdriver set

Definition :
Drive Board is a negative circuit that changes direct current (DC) to alternating current (AC)

Common Service Cases :


1. X-ray is not generating properly
2. Poor cable connection

TROUBLESHOOTING FOR KMC650(X-RAY) 5


IST 2014

How to check :
1. IGBT terminal short circuit test
1) Must remove the input power(DC220V) cable from Drive board before checking

IGBT location (Q1,Q2,Q3,Q4)

2) Check, GATE - EMITTER & GATE - COLLECTOR as bellow picture


3) Check, 4 IGBT following the second step

*Test Point Description


Normal DMAGED
1. GATE-EMITTER: Open Short
2. GATE-COLLECTOR: Open Short

TROUBLESHOOTING FOR KMC650(X-RAY) 6


IST 2014

4) Measure resistance of EMITTER-COLLECTOR (Normal resistance: about 4.9 Ω)

5) Measure resistance of 4 IGBT following the 4th step

TROUBLESHOOTING FOR KMC650(X-RAY) 7


IST 2014

2. Drive Board voltage check point

*Test Point Description

1. Drive Board operating voltage:

+15VDC, -15VDC, +24VDC

2. X-ray generating voltage:

+220VDC

TROUBLESHOOTING FOR KMC650(X-RAY) 8


IST 2014

Unit 3. PWR Board


WARING
You MUST switch off the Unit before replacing the part and ONLY an approved GEMSS Medical technician is qualified to carry out any intervention
on the Unit while it is switched on.

Spcification : Input 110VAC / Output 220VDC

Special Tools : wrench set, DMM, Screwdriver set

Definition :
PWR Board is a electronic power supply circuit that changes alternating current (AC) to direct current (DC)

Common Service Cases :


1. X-ray is not generating properly
2. Poor cable connection

TROUBLESHOOTING FOR KMC650(X-RAY) 9


IST 2014

How to check :
1. PWR Board check point

*Test Point Description

1. Input voltage:

110VAC

2. Check the Fuse condition

3. Output voltage

1) Drive Board operating voltage:

+15VDC, -15VDC, +24VDC

2) X-ray generating voltage:

+220VDC

TROUBLESHOOTING FOR KMC650(X-RAY) 10


IST 2014

Unit 4. Main Control Board


WARING
You MUST switch off the Unit before replacing the part and ONLY an approved GEMSS Medical technician is qualified to carry out any intervention
on the Unit while it is switched on.

Spcification : Main Control Board(4 Layer)

Special Tools : DMM, Screwdriver set

Definition :
Main controller is a control board. It is responsible for varies electrical control by combining control switch, relay, electric contractor,
and controller.

Common Service Cases :


1. Poor communication between PC
2. Bad Main Control Board
3. Bad cable

TROUBLESHOOTING FOR KMC650(X-RAY) 11


IST 2014

How to check :
1. Measure voltage of TP2, TP4 to check the kV and mA signal data

* Test Point Description

Test Point Comment

1. TP4: mA data(0~5 VDC)

2. TP2: KVp data(0~5 VDC)

TROUBLESHOOTING FOR KMC650(X-RAY) 12