Yesterday I finished the Cp-D vs T in a range from 5 K to 2 K by 0.

2 K/min cooling
rate under different
magnetic field. Now I will close this Cp and D vs T labview and open another.
Before I close it, I will show you how to use this.

Okay

First you save a .txt file.


Severval parameters are important, so you need to save them into the file name.
[1] frequency, [2] signal level, [3] measuring speed, [4] magnetic field, [5]
cooling rate
[6] number of average.

If I want to measure from 5 K to 2 K by cooling, first I will set the PPMS at 5 K,


then set all other parameters
in Labview including magnetic field and final tempeature 2 K.

Now it is running, labview will set H = 4 T and wait for H stable, then cool to 2 K
by 0.2 K/min
under 4 T.

Now it cooling and measuring. You need to manually turn off labview when T is near
2 K, sorry this is
automatically yet. probably you can improve it in the future. Let's stop the
labview now since we already
have 4 T data.

Now we will measure Cp_D vs H. We start at 2 K.

We will scan H from 0 T to 9 T by a speed of 25 Oe/second.


The time will be around 60 minutes.

Since at low H, there is no peak, we can manually speed up the measurment by


changing ramping rate.
I will manually change it to 100 Oe/Sec.
When the peak show up, I will change back to 25 Oe/sec.

I will go to offic now. In my impression, no peak at 2 K. So let's just use 100


Oe/sec.

There is a peak!!?? Let's scan the peak from 6 T to 5 T by 25 Oe/sec.

Wait, it will be 100 Hz.

Follow my operation.

Please do it again.

Set to 3.5 K. Set H to 0 T, so that you can scan from 0 T.

You are using 25 Oe/sec to set to 0 T, this is too slow, since you are not
measuring, you just set to
a H, you can use 200/sec.

Ok, now you can set the labview. Set the final H to 9 T. T is 3.5 K.
check the H in multivu, make sure H is at 0 T.

In the labviw you can set speed of H to 100 Oe/sec, since the peak only shows up
near several T.
Then when you see the peak, quickly set to 25 Oe/se. In this case we can save time.

check the H again. Ok, it is 0 T. you can run labviww no.w

Labview is wating for Tempeatture stable. then it will ramp H and measure.

In addition to the plots, you can also check the numbers in the box. I will show
you.

Clear? yes.
Ok, let's wait for the peak.

Set to 25 Oe/sec.

Not in the labview.

Too late....
let's remeasure the peak...

Since labview and multivu are parallel, you can change the H as you wish at any
time. but you have re-start
labivew if you want to change H in labview. So we simply change the H in multivu.

Now H is at 3 T, set to 9 T with a speed 25 Oe/sec, the the peak finish, set back
to 100 Oe/sec.
ok.

The peak is gone, so we can set to 100 Oe/sec now.

Good. Changing H speed is just to save time. Now it only takes 15 min for one curve
from 0 T to 9 T.
For a new measurement, you know nothing about the peak or detailed features of Cp
and D, you can simply set 25 Oe/sec
or 50 Oe/sec to slowly scan the curve.
The current measurement, I can change H speed is becasue that I know all details of
the peak.

Ok, now is very simple, when H = 9 T, stop labview and warm up to 300 K, also set H
to 0 T.