2 K/min cooling
rate under different
magnetic field. Now I will close this Cp and D vs T labview and open another.
Before I close it, I will show you how to use this.
Okay
Now it is running, labview will set H = 4 T and wait for H stable, then cool to 2 K
by 0.2 K/min
under 4 T.
Now it cooling and measuring. You need to manually turn off labview when T is near
2 K, sorry this is
automatically yet. probably you can improve it in the future. Let's stop the
labview now since we already
have 4 T data.
Follow my operation.
Please do it again.
You are using 25 Oe/sec to set to 0 T, this is too slow, since you are not
measuring, you just set to
a H, you can use 200/sec.
Ok, now you can set the labview. Set the final H to 9 T. T is 3.5 K.
check the H in multivu, make sure H is at 0 T.
In the labviw you can set speed of H to 100 Oe/sec, since the peak only shows up
near several T.
Then when you see the peak, quickly set to 25 Oe/se. In this case we can save time.
Labview is wating for Tempeatture stable. then it will ramp H and measure.
In addition to the plots, you can also check the numbers in the box. I will show
you.
Clear? yes.
Ok, let's wait for the peak.
Set to 25 Oe/sec.
Too late....
let's remeasure the peak...
Since labview and multivu are parallel, you can change the H as you wish at any
time. but you have re-start
labivew if you want to change H in labview. So we simply change the H in multivu.
Now H is at 3 T, set to 9 T with a speed 25 Oe/sec, the the peak finish, set back
to 100 Oe/sec.
ok.
Good. Changing H speed is just to save time. Now it only takes 15 min for one curve
from 0 T to 9 T.
For a new measurement, you know nothing about the peak or detailed features of Cp
and D, you can simply set 25 Oe/sec
or 50 Oe/sec to slowly scan the curve.
The current measurement, I can change H speed is becasue that I know all details of
the peak.
Ok, now is very simple, when H = 9 T, stop labview and warm up to 300 K, also set H
to 0 T.