Вы находитесь на странице: 1из 6

Long-term outcome of transcatheter secundum-type atrial septal defect closure using Amplatzer

septal occluders
Listen to the Editors Audio Summary
Jozef Masura, MD, PhD; Pavol Gavora, MD; Tomazz Podnar, MD, PhD
[+] Author Affiliations
J Am Coll Cardiol. February 15, 2005,45(4):505-507 doi:10.1016/j.jacc.2004.10.066
[+] More
ABSTRACT
Jump to Section [+]
Objectives The aim of this study was to assess long-term results of percutaneous closure of
secundum-type atrial septal defect (ASD II) using Amplatzer septal occluders (ASO).
Background Only immediate-, short-, and intermediate-term results of ASO implantation are
known so far.
Methods Between September 1995 and January 2000, 151 patients underwent a successful
percutaneous closure of ASD II in our institution. All were included in the present study and were
followed up until September 2004.
Results This group of patients was followed up from 56 to 108 months (median 78 months). The
mean stretched defect diameter was 15.9 ± 4.8 mm. There were no deaths or significant
complications during the study. At three years of follow-up, all defects were completely closed
and remained closed thereafter.
Conclusions Since the first human implantations in September 1995, the Amplatzer septal
occluder proved as a safe and effective device for percutaneous closure of ASD II.
ABBREVIATIONS AND ACRONYMS
Jump to Section [+]
ASD II
secundum-type atrial septal defect
ASO
Amplatzer septal occluder
TEE
transesophageal echocardiography
TTE
transthoracic echocardiography
Transcatheter closure of secundum-type atrial septal defect (ASD II) is an increasingly widespread
alternative to surgical closure (1). The Amplatzer septal occluder (ASO) is currently one of the
most frequently used devices for transcatheter closure of ASD II (2). Immediate-, short-, and
intermediate-term results of percutaneous closure of ASD II using ASO are promising (2–4).
However, long-term data after ASO implantation are lacking. Therefore, the purpose of the present
study was to evaluate a long-term outcome of transcatheter closure of ASD II using ASO in a
single institution.
METHODS
Jump to Section [+]
Patient population
From September 1995 to January 2000, 151 patients having isolated ASD II underwent
transcatheter closure using ASO. All 151 patients were followed up until September 2004. All
patients were included in our previous study analyzing morphological characteristics of isolated
ASD II (5). At the time of implantation, the mean patient age was 11.9 ± 11.6 years and weight
36.0 ± 20.9 kg. The study was part of a clinical trial approved by an authorized ethics committee.
An informed written consent has been obtained from all patients or their parents.
The occluder
The Amplatzer septal occluder and delivery system (AGA Medical, Golden Valley, Minnesota)
have been described in detail previously (2,6).
Selection of patients suitable for transcatheter closure
The selection of patients suitable for a transcatheter closure using ASO was based on the
measurement of the maximal defect diameter and morphological characteristics of the defect. In
fact, ASD II characteristics of all patients included in the present study were reported in detail
previously (5).
Preimplantation protocol
A physical examination, a standard 12-lead electrocardiogram (ECG), chest radiograph,
transthoracic echocardiography (TTE), and transesophageal echocardiography (TEE) were
performed in all patients.
Implantation procedure
The protocol for ASO implantation has been reported in detail previously (2).
Follow-up protocol
Immediately after the ASO release, a precise TEE examination was performed. The shape of the
occluder was evaluated. Thrombus formation on the device was sought. A detailed color Doppler
interrogation of the interatrial septum was performed to detect and quantify any residual shunts. A
color Doppler signal width <2 mm was considered as a small residual shunt, 2 to 4 mm as a
moderate shunt, and >4 mm as a significant residual shunt. Relationships between the occluder
and both atrioventricular valves were evaluated. Drainage of the caval veins, right pulmonary
veins, and the coronary sinus were evaluated for obstruction.
At 24-h follow-up, ECG, chest radiograph, and TTE were performed. Both chest radiograph and
TTE allowed evaluation of the ASO shape. Thrombi on both discs of the ASO were searched for
using TTE. Residual shunts were sought and quantified using the same color Doppler criteria as
during TEE examination. A relationship of both atrioventricular valves toward the occluder was
assessed. Drainage of the caval veins, right pulmonary veins, and coronary sinus were evaluated
for obstruction.
Thereafter, follow-up ECG and TTE were performed at 1 month, 3 months, 12 months, and then
annually after the implantation. The same TTE examination protocol was used throughout a
follow-up period as performed 24 h after the procedure. In addition, complications related to the
ASO implantation were noted at each follow-up visit. Aspirin, 5 mg/kg daily, and infective
endocarditis prophylaxis were recommended for 6 months after the procedure in all patients.
Statistics
The data are expressed as mean ± SD or as median and ranges as appropriate.
RESULTS
Jump to Section [+]
From September 1995 to January 2000, 154 patients were taken to the catheterization laboratory
with an intent-to-treat. Transcatheter closure was attempted in 151 patients, and implantation of
the occluder was successful in all of them.
The mean maximal defect diameter measured by TEE was 12.9 ± 4.4 mm. The mean stretched
defect diameter was 15.9 ± 4.8 mm. Altogether, 152 occluders have been implanted. Two
occluders were implanted in a single patient having two widely separated defects. The mean size
of the implanted ASO was 16.1 ± 5.3.
Follow-up evaluation was complete at each of the intervals. There were no deaths, cardiac
perforations, device embolizations or malpositions, thrombus formations or thromboembolisms,
significant arrhythmias, infective endocarditis, or other morbidity associated with ASD II closure
during the entire follow-up period ranging from 56 to 108 months (median 78 months).
Time course of residual atrial shunt disappearance after ASO implantation is represented in Table
1.
Table 1Time Course of Residual Atrial Shunt Disappearance After Amplatzer Septal Occluder
Implantation in All 151 Patients in the Study
View table
The mitral valve and tricuspid valve were not encroached on by the occluder in any patient in the
study. The caval veins, right pulmonary veins, and coronary sinus drained freely in all patients
during the follow-up period.
Satisfactory images of implanted devices were obtained in all patients in the study. Deformation of
implanted occluder or occluder integrity problems were not detected during follow-up in any
patient.
DISCUSSION
Jump to Section [+]
The present study demonstrates an excellent outcome of percutaneous ASD II closure using ASO
during a follow-up period ranging up to 9 years.
Safety of ASD II closure using ASO
Complications of percutaneous ASD II closure using ASO reported in the literature were rare and
were early in the vast majority of patients (4,7). Exceptionally, reported complications were late
(7–10). No deaths or significant complications were experienced in the present study. Therefore,
this study confirms safety of percutaneous ASD II closure using ASO both early and during long-
term follow-up.
The cardiac perforation is a rare, life-threatening complication of transcatheter closure of ASD II
and was not experienced in the present study. A procedure-related cardiac perforation is an
avoidable complication of ASO implantation (7). In contrast, a cardiac perforation occurring six
months after ASO implantation causing hemodynamic collapse remained unexplained and is,
therefore, more worrisome (8). An aorta-to-right-atrial fistula is an additional rare complication
caused by erosion into the aorta by the right atrial disc of ASO (9). It was detected three months
after ASO implantation in a defect with a deficient superior anterior rim. A 26-mm device was
implanted in a defect with a stretch defect diameter of 26 mm. Selecting a device 2 to 4 mm larger
than the stretch defect diameter might prevent the development of fistula in this defect
morphology.
Device embolizations and malpositions are well known early complications of percutaneous ASD
II closure (7,11). An early device embolization or malposition are avoidable complications of
percutaneous ASD II closure using ASO and were not experienced in this study. Delayed ASO
embolization, occurring one week after implantation, was also reported (10). The defect was oval
in shape in the reported patient, and, therefore, undersized ASO was implanted.
The incidence of thrombus formation on ASD II closure devices, particularly on ASO, is low
(12,13). Thrombi develop on devices both early and late during follow-up. Neither thrombus
formation nor systemic thromboembolism were detected in our group of patients either
immediately after implantation or during follow-up. However, TEE was performed only
immediately after the device implantation in the present study, and, therefore, small, clinically
silent thrombi may not be detected in our group of patients during follow-up.
So far, arrhythmias were reported only early after ASO implantation (7,14). Similarly, we noted an
increase in supraventricular and ventricular ectopy immediately and 24-h after the procedure and
not later during follow-up.
Infective endocarditis on ASO was not experienced in our study. However, an infective
endocarditis on an ASO was reported in a single patient almost two months after implantation,
underscoring the need for antibiotic prophylaxis until a complete endothelialization occurs (6,15).
The function of both atrioventricular valves and drainage of the caval veins, right pulmonary
veins, and coronary sinus were undisturbed both immediately after ASO implantation and during
follow-up. Careful patient selection and precise evaluation after device positioning is necessary to
prevent a compromise of structures surrounding the device.
So far, device integrity problems or deformations of implanted occluders were not reported after
ASO implantation and were also not observed in this study. In contrast, a large ASO profile
obtained immediately after implantation decreases significantly during follow-up, resulting in a
much lower device profile (16).
The reported incidence of late events after ASO implantation is low. Therefore, a small sample
size may not be representative of the incidence of rare events; this can only be resolved by a large,
multi-centered registry.
Effectiveness of ASD II closure using ASO
The present study confirmed effectiveness of percutaneous ASD II closure using ASO. Our group
of 151 patients was selected from a group of 190 consecutive children having isolated ASD II (5).
Thus, 79.4% of patients with isolated ASD II underwent successful implantation of ASO.
Similarly, Fischer et al. (4) reported successful ASO implantation in 200 of 236 consecutive
patients (84.7%). An immediate complete closure was demonstrated by TEE in 120 of 151 patients
in the present study (79.4%) (Table 1). The following day, TTE proved a complete closure in 138
of 151 patients (91.3%). The closure rate increased steadily thereafter, and, at three years follow-
up, TTE proved a complete closure in all patients. Other authors also reported excellent
immediate-, short-, and intermediate-term closure rates (2–4). For comparison, at a median follow-
up of 2.3 years, Fischer et al. (4) reported a complete closure in 94% of patients with a trivial
residual shunt in the remaining patients.
Conclusions
The present study proved that percutaneous closure of ASD II using ASO is safe and effective
during a follow-up period of up to 9.0 years. A precise selection of suitable patients based on ASD
II size and morphology, a selection of ASO of appropriate size, attention to technical details
during implantation, and consideration for infective endocarditis prophylaxis and low-dose aspirin
during a six-month period after ASO implantation are crucial to prevent complications and to
achieve a high closure rate.
REFERENCES
Jump to Section [+]
1
Holzer R., Hijazi Z.M.; Interventional approach to congenital heart disease. Curr Opin Cardiol.
19 2004:84-90.
CrossRef | PubMed
2
Masura J., Gavora P., Formanek A., Hijazi Z.M.; Transcatheter closure of secundum atrial septal
defects using the new self-centering Amplatzer septal occluder: initial human experience. Cathet
Cardiovasc Diagn. 42 1997:388-393.
CrossRef | PubMed
3
Berger F., Ewert P., Björnstadt P.G.; Transcatheter closure as standard treatment for most
interatrial defects: experience in 200 patients treated with Amplatzer septal occluder. Cardiol
Young. 9 1999:468-473.
PubMed
4
Fischer G., Stieh J., Uebing A., Hoffmann U., Morf G., Kramer H.H.; Experience with
transcatheter closure of secundum atrial septal defects using the Amplatzer septal occluder: a
single center study in 236 consecutive patients. Heart. 89 2003:199-204.
CrossRef | PubMed
5
Podnar T., Martanovic P., Gavora P., Masura J.; Morphological variations of secundum-type
atrial septal defects: feasibility for percutaneous closure using Amplatzer septal occluders. Cathet
Cardiovasc Intervent. 53 2001:386-391.
CrossRef
6
Sharafuddin M.J.A., Xiaoping G., Titus J.L., Urness M., Cervera-Ceballos J.J., Amplatz K.;
Transvenous closure of secundum atrial septal defects. Preliminary results with a new self-
expanding nitinol prosthesis in a swine model. Circulation. 95 1997:2162-2168.
CrossRef | PubMed
7
Chessa M., Carminati M., Butera G.; Early and late complications associated with transcatheter
occlusion of secundum atrial septal defect. J Am Coll Cardiol. 39 2002:1061-1065.
CrossRef | PubMed
8
Preventza O., Sampath-Kumar S., Wasnick J., Gold J.P.; Late cardiac perforation following
transcatheter atrial septal defect closure. Ann Thorac Surg. 77 2004:1435-1437.
CrossRef | PubMed
9
Chun D.S., Turrentine M.W., Moustapha A., Hoyer M.H.; Development of aorta-to-right atrial
fistula following closure of secundum atrial septal defect using Amplatzer septal occluder. Cathet
Cardiovasc Intervent. 58 2003:246-251.
CrossRef
10
Verma P.K., Thingnam S.K., Sharma A., Taneja J.S., Varma J.S., Grover A.; Delayed
embolisation of Amplatzer septal occluder device: an unknown entity. A case report. Angiology.
54 2003:115-118.
CrossRef | PubMed
11
Levi D.S., Moore J.W.; Embolisation and retrieval of the Amplatzer septal occluder. Catheter
Cardiovasc Intervent. 61 2004:543-547.
CrossRef
12
Krumsdorf U., Ostermayer S., Billinger K.; Incidence and clinical course of thrombus formation
on atrial septal defect and patent foramen ovale devices in 1,000 consecutive patients. J Am Coll
Cardiol. 43 2004:302-309.
CrossRef | PubMed
13
Anzai H., Child J., Natterson B.; Incidence of thrombus formation on the CardioSEAL and the
Amplatzer interatrial closure devices. Am J Cardiol. 93 2004:426-431.
CrossRef | PubMed
14
Hill S.L., Berul C.I., Patel H.T.; Early ECG abnormalities associated with transcatheter closure
of atrial septal defects using the Amplatzer septal occluder. J Interv Card Electrophysiol. 4
2000:469-474.
CrossRef | PubMed
15
Bullock A.M., Menahem S., Wilkinson J.L.; Infective endocarditis on an occluder closing an
atrial septal defect. Cardiol Young. 9 1999:65-67.
CrossRef | PubMed
16
Cao Q.L., Du Z.D., Joseph A.; Immediate and six-month results of the profile of the Amplatzer
septal occluder as assessed by transesophageal echocardiography. Am J Cardiol. 88 2001:754-759.
CrossRef | PubMed

Вам также может понравиться