Академический Документы
Профессиональный Документы
Культура Документы
INTRODUCTION
&RULROLVVLJQDOSURFHVVLQJ
'63EDVHG
www.micromotion.com (TUTOR™)
1) Flow Profile: Establish if Coriolis is truly The AGA work focuses on two “sweet spots” for Coriolis:
insensitive to flow profile and swirl. Test matrix
includes “AGA-3 like” testing for single and 1. Medium to high pressure distribution metering
double elbows in and out of plane. Early results points. These city-gates or industry-gates often
from Micro Motion show there is no effect at least have high turndown requirements, and new
within 5d; one researcher (Dr. Umesh Karnik of installations must meet the recently revised,
TCPL/NOVA) found some profile dependence, more stringent AGA-3 straight run requirements,
as reported at the 4th International Fluid Flow adding additional cost to using traditional
Symposium (July, 1999 Denver, CO USA). Other measurement technologies. As well, Coriolis’
Coriolis vendors have indicated some small proven stability over time offers reduced
profile sensitivity, so this testing is critical to the maintenance and field proving/verification.
establishment of a technical note and/or
standard. 2. Wet gas measurement. If Coriolis is capable of
measuring wellhead or produced gas, prior to
0.5
processing, significant operational savings are
Mass Flow Rate Error, %
0.4
0.3
Baseline
possible.
0.2
0.1 Single El
0 Double El SIZING AND SELECTION
-0.1 Single El
-0.2 Selection of a Coriolis meter for gas applications is quite
Double El
-0.3
-0.4 straightforward but different than traditional technologies
-0.5 used on natural gas, such as orifice and turbine meters.
10000 100000 1000000 There are two reasons for this; one being that Coriolis is
Reynolds Number available in discrete sizes (like turbine or rotary); the
second being that the Coriolis meter can be sized for a
Results from a 1995 study (T. Patten; North Sea Flow much higher pressure drop than is the industry norm.
Workshop) using hot water. Within a few pipe diameters This can be useful as it increases useable turndown.
of the primary element, no effect of flow profile or swirl
was found. Coriolis can be installed upstream of a pressure regulator,
resulting in a smaller and less expensive primary (sensor)
and increased turndown.
1) Pressure loss,
PRESSURE DROP
METER ERROR VS. FLOW RATE Both pressure and temperature affect the meter vibration
characteristics, hence the magnitude of the sensed
Meter error versus flow rate is determined from a Coriolis force. In comparison to zero stability, these effects
performance curve, similar to the one shown below. The are small, but should be compensated for to achieve
error versus flow rate curve is based on the results of optimum meter performance. Most meter designs
laboratory calibrations. Most manufacturers state the compensate for temperature effect automatically, using
probable meter error as a percentage of rate, plus the zero the temperature of the tube(s), and for pressure effects
stability value. Third party meter error data are referenced either continuously using an external pressure transmitter,
in Appendix B. The error is typically expressed as: or by entering a fixed adjustment for the known pressure.
% Error = ± (0.50% ± (zero stability/flow rate) x 100) Other meter designs periodically check meter sensitivity
by applying a waveform reference force to the tube(s),
The base error value (0.50%) in equation (9) was chosen during field operation, and compare the system response
for illustrative purposes. The actual meter error can be to that achieved under reference flowing conditions. This
established from laboratory calibration. This should system will compensate for both pressure and
include the effects of laboratory uncertainty, linearity, temperature effects.
hysteresis, and repeatability.
Errors and compensation methods for pressure and
temperature effects should be stated in the meter
performance specifications, and included, if necessary,
Flow Error and Pressure Loss vs. Flow Rate, 3" Bending Mode Coriolis Meter
Gas Conditions: Gr = 0.59, Tf = 60 F
when establishing meter performance for sizing
3.00 600
considerations. Errors are estimated from data for each
Error limit
meter design and with the use of error propagation
2.00 Pressure loss-1000 psig 500 methods.
Pressure loss-500 psig
Pressure Loss, inches H2O
1.00 400
TURNDOWN RATIO
Flow Rate Error, %
0.00 300
Flow meter turndown ratio is the ratio of the acceptable
-1.00 200
maximum mass flow rate to the acceptable minimum
mass flow rate. The turndown ratio is application specific
-2.00 100 and dependent on gas conditions, allowable pressure
loss across the meter, and allowable meter error.
-3.00 0
0 200000 400000 600000 800000 1000000
Flow Rate, SCFH The maximum pressure loss (at maximum flow rate)
across the meter can be determined once the meter
Meter Error and Pressure Loss Versus Flow Rate
diameter, piping installation configuration, and maximum
allowable gas velocity are specified. Typically, the meter
selected is one line diameter smaller than the size of the
60
Fluid Swirl and Flow Profile Effects
40
The effect of fluid swirl and non-uniform velocity profiles
caused by upstream and downstream piping
20
configuration on meter performance differs from one
meter design to another. Flow conditioning and straight
0
upstream and downstream piping lengths may or may
0.5 1.0 1.5 2.0 2.5 3.0 not be required. Further testing is recommended to
Error at Minimum Flow, Percent evaluate these potential effects.
Error at Minimum Flow Versus Flow Turndown Ratio Effects of Contaminants, i.e. Compressor Oil, Liquids
and Free Mists
pipe in which the meter is installed. This usually provides Testing has shown that liquid carried in a gas stream may
more accurate measurement at lower flow rates. not have the same adverse affect on performance as gas
However, the resulting permanent pressure loss for a carried in a liquid stream. However, the meter will measure
given flow rate is higher than if the meter diameter is the the mass flow rate of the total flow stream, including the
same as the pipe diameter. Because higher pressure gas liquid i.e., condensate, glycol and compressor oil. The
has a higher mass flow rate for the same velocity, higher allowable liquid fraction will depend on the application
pressures will produce higher flow turndowns for the and sensor geometry. Care should be taken to remove
same meter arrangement. liquid slugs before measuring the gas flow.
A family of curves can be generated showing flow Vibration and Fluid Pulsation
turndown for different gas pressures. The figure above
presents example values of this relationship for a During product development, extensive analysis and
pressure loss of 1% of pipe pressure at the meter inlet. testing have resulted in meter designs that are inherently
stable under a wide range of “normal” plant vibration
INSTALLATION & MAINTENANCE conditions.
Proper mounting of the sensor is required. Consideration Other than the vibrating sensor element(s), Coriolis meters
should be given to the support of the sensor, the have no moving parts, requiring minimal maintenance.
alignment of the inlet and outlet flanges with the sensor.
A spool piece should be used in place of the meter to There are three common types of field checks which
align pipe-work during the construction phase. include meter zero, sensor checks, and transmitter
checks.
Piping should follow typical industry piping codes. Meter
performance, specifically zero stability, can be affected COMMON FIELD CHECKS
by axial, bending, and torsion stresses from pressure,
weight and thermal effects. Utilizing properly aligned Meter Zero
pipe-work and properly designed supports can minimize
these stresses and associated loads. Should be checked periodically and adjusted if it does
not meet the manufacturer’s specifications. If the meter
Mount the Coriolis transmitter so that it may be easily zero check shows degradation, then the following check
accessed to attach communications equipment, to view should be performed. Physical internal inspection may
displays, and to use keypads. Coriolis meters are be needed.
configured in two basic ways – the transmitter mounted
to the sensor or the transmitter mounted remotely. Sensor Checks
Secondary Element (Transmitter) Coriolis meters can be sized for very low-pressure drop
(100” H2O), but can also be installed upstream of the
Diagnostic LED(s) and display may be provided to pressure regulator for increased useable turndown. For
indicate operating status of the primary and secondary instance, in one application for custody transfer of
elements. See the manufacturer’s documentation for nitrogen, a 50-psid drop (2000” H2O) was taken through
detailed description of secondary element diagnostic and the primary element, and the pressure regulator adjusted
trouble shooting procedures. In addition, refer to API accordingly. This allowed the use of a 1” primary element
Manual of Petroleum Measurement Standard (MPMS) instead of a 3” element, and a 40:1 useable turndown
Chapter 21.1 for security issues. (better than 2% accuracy at minimum flow). The only
other alternative was two to three parallel orifice runs at
Density Checks much greater installed cost ($8k vs. $20k).
As of this writing, operating density measured by the Test/Production Separators: The application shown
meter should not be used to convert mass flow rate to below is a “before and after” scenario. Coriolis meters
volume flow rate. However, it is useful as a diagnostic on both the liquid (oil/water) and gas streams streamlined
tool to monitor changes in meter performance or the separator design, saving over $100k in design,
operating conditions. engineering, and fabrication. As well, numerous parallel
orifice runs were eliminated by the superior turndown of
Checking and Adjusting Meter Zero the Coriolis meter.
Improper zeroing will result in measurement error. In order Saudi Aramco Separator Gas: Saudi Aramco uses a
to adjust the zero of the meter there must be no flow number of Coriolis meters on both the liquid and gas
through the flow sensor, and the sensor must be filled side. This application is of particular note because
with gas at process conditions. The meter zero must be the gas stream is wet with entrained condensate.
established at process conditions of temperature, Measurement of this stream is within a few percent over
pressure and density. Even though the stream is not a wide range of conditions, greatly enhancing separator
flowing, the flow meter may indicate a small amount of operation.
flow, either positive or negative. Causes for the zero error
are usually related to the differences between the Fuel Control: A major US vendor of gas turbines designs
calibration conditions and the actual installation, which a high-efficiency, low emissions offering. This design
include but are not limited to the following: utilizes a trio of Coriolis meters to measure the natural
gas burned in each of three combustion zones (fuel
• Differences between the calibration media
density and the gas density
• Differences in temperature
• Differing mounting conditions
The meter should read a mass flow rate that is less than
the manufacturer’s zero stability specification under the
no-flow condition.
Although this methodology requires that the meter be Energy Metering: Coriolis meters have the potential to
removed from service, it defines very accurately the in- be excellent energy meters, with or without on-line
situ performance of the meter. Since steps 1 & 2 establish chromatography. Three basic levels of energy metering
the meter traceability between water and gas, verifying exist, as follows:
water performance in step 3 automatically validates the
meter in-situ (gas) performance. After some experience, 1. Small lines, where volumes do not justify on-
it is likely that the period to repeat step 3 would be line GC. Coriolis meter alone can be used to
lengthened from every year to every two or three years. measure energy content typically within 1-3%.
2. Medium lines, where volume justifies an on-line
A variation of this proving methodology is to use a Coriolis GC. Single or parallel Coriolis meters can be tied
meter as a master meter. By establishing the traceability to the GC, per the sketch below:
between water and gas measurement on the master Mass/time
meter, it can be used to prove other meters (of any type). Core Processor
Flow
Computer
Energy flow
(BTU/hr, etc)
“Puck”
Figure 3 shows a 1” Coriolis meter being used as an in-
E/mass
situ master to prove an in-line turbine meter. Since in GC