Schedule

Issue date: 2 November 2017


Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 1 of 30

LABORATORY LOCATION: METCAL TECHNOLOGIES (M) SDN. BHD.


(PERMANENT LABORATORY) NO. 36, CANGKAT BUKIT BELAH
11920 BAYAN LEPAS
PULAU PINANG
MALAYSIA

FIELDS OF CALIBRATION: DIMENSIONAL, MASS, FORCE, TORQUE,


TEMPERATURE, PRESSURE, VOLUME,
ELECRTRICAL, TIME AND FREQUENCY

This laboratory has demonstrated its technical competence to operate in accordance with MS
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

ISO/IEC 17025:2005 (ISO/IEC 17025:2005).

This laboratory’s fulfillment of the requirements of ISO/IEC 17025 means the laboratory
meets both the technical competence requirements and management system requirements
that are necessary for it to consistently deliver technically valid test results and calibrations.
The management system requirements in ISO/IEC 17025 are written in language relevant to
laboratory operations and operate generally in accordance with the principles of ISO 9001
(see Joint ISO-ILAC-IAF Communiqué dated April 2017).

* The expanded uncertainties are based on an estimated confidence probability of


approximately 95% and have a coverage factor of k=2 unless stated otherwise.

SCOPE OF CALIBRATION: DIMENSIONAL

Calibration and
Instrument
Measurement
Calibrated/
Range Capability Remarks
Measurement
Expressed as an
Parameter
Uncertainty(±)*
Feeler Gauge 0.01 mm to 3 mm 0.78 µm Direct
measurement with
micrometer based
on JIS B 7524:1992

External Micrometer 0 mm to 25 mm 0.44 µm Comparison with


25 mm to 100 mm 0.82 µm gauge block based
100 mm to 250 mm 2.3 µm on JIS B
250 mm to 500 mm 4.7 µm 7520:1981

Vernier, Dial & Digital 0 mm to 600 mm 6 µm Comparison with


Caliper caliper checker /
gauge block based
on JIS B 7507:1993

Dial Gauge / Digital 0 mm to 50 mm 1.5 µm Comparison with


Indicator gauge tester based
on JIS B 7503:2011

Dial Test Indicator 0 mm to 1.6 mm 1.5 µm Comparison with


gauge tester based
on JIS B 7533-1990

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 2 of 30

SCOPE OF CALIBRATION: DIMENSIONAL

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Vernier, Dial & Digital 0 mm to 100 mm (0.2 + 4.5 L) µm Comparison with
Height Gauge 100 mm to 600 mm (-0.3 + 9.5 L) µm caliper checker /
‘L’ in metre gauge block
based on JIS B
7517:1993
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

Gauge Block 0.5 mm to 100 mm (0.11 + 3L) µm Comparison


with master
L in m gauge block
based on ISO
3650 / JIS B
7506:1997

Dial & Digital 0 to 20 mm 2.5 µm Comparison


Thickness Gauge with gauge
block based on
JIS B
7503:1997

Plain Pin Gauge 0.10 mm to 25 mm 0.26 µm Direct


(diameter only) measurement
with micrometer
based on JIS B
7420:1997

Depth 0 mm to 300 mm 6 µm Comparison


Micrometer/Gauge with gauge
block based on
JIS B
7544:1994

Signatories:

1. Kenny Ching Ah Hoo


2. Eunice Wong Phaik Gaik
3. Ching Kar Chien
4. Chew Siou Lian

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 3 of 30

SCOPE OF CALIBRATION: DIMENSIONAL

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Height Master Up to 300 mm 2.3 µm Comparison with
gauge block
based on AS
3779:1990

Riser Block Up to 300 mm 2.3 µm Comparison with


Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

gauge block
based on AS
3779:1990

Electrical Comparator ± 5 µm 0.10 µm Comparison with


(Mu-checker) ± 15 µm 0.16 µm gauge block
± 50 µm 0.31 µm based on JIS B
± 150 µm 1.2 µm 7536:1982
± 500 µm 2.9 µm
± 1500 µm 12 µm

Parallel Thread Plug Major diameter 0.62 µm Three wire


Gauge 1 mm to 25 mm method based
on JIS B
Pitch diameter 1.7 µm 0261:1992
0.25 mm to 6 mm

Pitch 3.4 µm
0.25mm to 3 mm

Signatories:

1. Eunice Wong Phaik Gaik (Except for Scope Parallel Thread Plug Gauge)
2. Ching Kar Chien
3. Chew Siou Lian

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 4 of 30

SCOPE OF CALIBRATION: DIMENSIONAL

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Plain Ring Gauge 20 mm to 100 mm 0.4 µm + 12 L µm Length
‘L’ in metre Comparator
JIS B 7420

Glass Scale 0 mm to 200 mm 6.2 µm Measuring


System and
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

Reference
Scale on JIS B
7541:2001

Steel Ruler 0 mm to 300 mm 0.068 mm Measuring


Projector
JIS B 7516:2005

Angle Protractor 0 Degree to 60 Degree 0.13 Degree Sine Bar, Slip


Gauges,
Comparator
VDI/WDE/DGQ
2618

Signatories:

1. Chew Siou Lian


2. Ching Kar Chien

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 5 of 30

SCOPE OF CALIBRATION: DIMENSIONAL

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Profile Projector x, y-axis: 3 µm
(Measuring Accuracy 0 to 300 mm Comparison with
of Individual Axis) standard scale
based on JIS B
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

7184:1999
Measuring Scope x, y-axis: 2 µm Comparison with
(Measuring Accuracy 0 to 300 mm standard scale
of Individual Axis) based on JIS
7184:1999

Surface Plate 600 x 600 mm 1.3 µm Planekator,


1200 x 1200mm 1.5 µm micro-indicator
1800 x 1800mm 1.7 µm based on JIS B
7513:1992

Vernier, Dial & Digital 0 mm to 600 mm (1.3 + 4.7L) µm Comparison with


Height Gauge / Linear ‘L’ in metre gauge blocks,
Height height master
based on
JIS B 7517:1993

Signatories:

1. Kenny Ching Ah Hoo


2. Eunice Wong Phaik Gaik
3. Ching Kar Chien
4. Chew Siou Lian

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 6 of 30

SCOPE OF CALIBRATION: MASS

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
1 mg 0.009 mg Direct
Standard Weights 2 mg 0.009 mg Comparison
5 mg 0.009 mg using the ABBA
10 mg 0.009 mg or AB1 …BnA
20 mg 0.010 mg weighing
50 mg 0.010 mg sequence.
100 mg 0.010 mg
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

200 mg 0.011 mg
500 mg 0.011 mg
1g 0.012 mg
2g 0.013 mg
5g 0.015 mg
10 g 0.018 mg
20 g 0.018 mg
50 g 0.046 mg
100 g 0.12 mg
200 g 0.14 mg
500 g 0.89 mg
1 kg 0.96 mg
2 kg 1.5 mg
5 kg 4.7 mg
10 kg 0.012 g
20 kg 0.018 g
50 kg 9.3 g

Signatories:

1. Ching Kar Chien


2. Chew Siou Lian

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 7 of 30

SCOPE OF CALIBRATION: MASS

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*

Weighing Balance Up to 50 g 0.039 mg Standard weight


(Analytical Balance, Up to 250 g 0.27 mg based on ASTM
Scale, Electronic type, Up to 5000 g 0.019 g E898-88:2013
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

platform or pallet) Up to 10 kg 0.65 g


Up to 20 kg 2.6 g
Up to 150 kg 0.049 kg
Up to 300 kg 0.067 kg
Up to 500 kg 0.12 kg
Up to 1000 kg 0.45 kg
Up to 3000 kg 1.6 kg

Signatories:

1. Kenny Ching Ah Hoo


2. Eunice Wong Phaik Gaik
3. Ching Kar Chien
4. Chew Chii Tsong

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 8 of 30

SCOPE OF CALIBRATION: FORCE

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Durometer Hardness Spring force for
Type A, B & O 0.65 durometer point Dead weights
Type C, D & DO 0.65 durometer point method

Signatory:
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

1. Ching Kar Chien

SCOPE OF CALIBRATION: FORCE

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Force Gauge / Push- 0 N to 10 N ± 0.0011 N Dead weights
Pull Gauge 10 N to 50 N ± 0.011 N method
50 N to 500 N ± 0.080 N
500 N to 1000 N ± 0.80 N

Signatories:

1. Kenny Ching Ah Hoo


2. Eunice Wong Phaik Gaik
3. Ching Kar Chien
4. Chew Chii Tsong

SCOPE OF CALIBRATION: FORCE

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Hardness Testing 30 to 60 HRC 0.25 HRC Standard
Machine (HRC) hardness block

Signatories:

1. Kenny Ching Ah Hoo


2. Eunice Wong Phaik Gaik
3. Ching Kar Chien

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 9 of 30

SCOPE OF CALIBRATION: PRESSURE

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Pressure Indicating Comparison
Devices (Pressure with Reference
Gauge, Vacuum Pressure
Gauge, Manometer, Calibrator
Magnehelic)

Manometer -5.8 to 5.8 psi 0.00078 psi


Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

Vacuum -12 to 0 psi 0.022 psi

Pneumatic/Hydraulic 0 to 300 psi 0.049 psi

Hydraulic 300 to 10000 psi 4.8 psi

Signatories:

1. Ching Kar Chien


2. Chew Siou Lian
3. Chew Chii Tsong

SCOPE OF CALIBRATION: PRESSURE

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Pressure Indicating Comparison
Devices (Pressure with Reference
Gauge, Vacuum Pressure
Gauge, Manometer, Calibrator
Magnehelic)

Manometer -5.8 to 5.8 psi 0.00078 psi

Vacuum -12 to 0 psi 0.022 psi

Pneumatic/Hydraulic 0 to 300 psi 0.049 psi

Hydraulic 300 to 10000 psi 4.8 psi

Signatories:

1. Ching Kar Chien


2. Chew Siou Lian
3. Chew Chii Tsong
SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)
LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 10 of 30

SCOPE OF CALIBRATION: TORQUE

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Torque Measurement 0 N.m to 1 N.m 0.0012 N.m Calibration using
Devices Above 1 N.m to 10 N.m 0.020 N.m deadweights and
Above 10 N.m to 40 N.m 0.042 N.m known radius of
torque wheel.

Signatories:
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

1. Kenny Ching Ah Hoo


2. Eunice Wong Phaik Gaik
3. Ching Kar Chien

SCOPE OF CALIBRATION: TORQUE

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Torque Tools Up to 1 N.m 0.0016 N.m Based on ISO
(Torque Above 1 N.m to 10 N.m 0.016 N.m 6789:2003
Wrench, Torque Above 10 N.m to 40 N.m 0.18 N.m
Driver, Fixed or Above 40 N.m to 150 N.m 0.48 N.m
Adjustable) Above 150 N.m to 600 N.m 1.4 N.m
Above 600 N.m to 1000 N.m 1.8 N.m
Above 1000 N.m to 1500 N.m 2.7 N.m

Signatories:

1. Ching Kar Chien


2. Chew Siou Lian
3. Chew Chii Tsong

SCOPE OF CALIBRATION: TORQUE

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Torque Tools (Torque 0 N.m to 1 N.m 0.0055 N.m Based on ISO
Wrench, Torque Driver Above 1 N.m to 5 N.m 0.048 N.m 6789:2017
Fixed or Adjustable)

Signatories:

1. Ching Kar Chien


2. Chew Chii Tsong
SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)
LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 11 of 30

SCOPE OF CALIBRATION: TEMPERATURE

Instrument Calibration and


Calibrated/ Measurement Capability
Range Remarks
Measurement Expressed as an
Parameter Uncertainty(±)*
Temperature Sensor -70 oC to -20 oC 0.30 oC Comparison with
-20 oC to 0 oC 0.041 oC PT 100 in Liquid
0 oC to 100 oC 0.022 oC Bath and
100 oC to 300 oC 0.100 oC Temperature
300 oC to 500 oC 0.510 oC Block Calibrator
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

Surface Sensor 0 oC to 50 oC 0.50 oC Comparison with


50 oC to 200 oC 0.55 oC reference
200 oC to 300 oC 0.71 oC surface sensor
on electronic
hotplate /
equalizing block
of dry bath

Temperature Block -20 oC to 300 oC 0.061 oC Comparison with


Calibrator 300 oC to 600 oC 0.12 oC PT 100 /
600 oC to 700 oC 1.0 oC Thermocouple
Type – K
EURAMET
cg-13 Version
3.0

Radiation / Infrared 0 oC to 50 oC 0.52 oC Comparison with


Thermometer 50 oC to 100 oC 1.3 oC PT 100 in
100 oC to 140 oC 1.4 oC blackbody
source
ASTM E
1256:2007

Signatories:

1. Ching Kar Chien


2. Chew Chii Tsong

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 12 of 30

SCOPE OF CALIBRATION: TEMPERATURE

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Themohygro Devices

A) Relative Humidity 20 %rh to 80 %rh 1.8 %rh Comparison with


80 %rh to 90 %rh 2.2 %rh Reference Standard
in
Temperature 0 °C to 60 oC 0.10 oC Temperature/Humidity
Chamber
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

B) Relative Humidity 11 %rh 1.1 %rh Fixed Point


(Fixed Point) 33 %rh 1.3 %rh Calibration in Salt
75 %rh 1.4 %rh Chamber
97 %rh 1.7 %rh

Temperature 25 oC 0.31 oC

Liquid – In – Glass 0 oC to 200 oC 0.10 oC Comparison with


Thermometer PT100 in a stirred
(Total Immersion) liquid bath

Signatories:

1. Ching Kar Chien


2. Chew Siou Lian
3. Chew Chii Tsong

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 13 of 30

SCOPE OF CALIBRATION: TEMPERATURE

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Temperature Electrical
Indicator simulation using
multi-product
K -200 to 0 oC 0.14 oC calibrator based
0 to 1370 oC 0.12 oC on ITS-90 Table
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

J -210 to 0 oC 0.12 oC
0 to 1200 oC 0.10 oC

T -200 to 0 oC 0.12 oC
0 to 400 oC 0.10 oC

E -200 to 0 oC 0.10 oC
0 to 1000 oC 0.10 oC

N -200 to 0 oC 0.16 oC
0 to 1300 oC 0.10 oC

B 250 to 600 oC 0.50 oC


600 to 1820 oC 0.15 oC

R -50 to 0 oC 0.35 oC
0 to 1760 oC 0.14 oC

S -50 to 0 oC 0.33 oC
0 to 1760 oC 0.16 oC

Pt 100 -200 to 850 oC 0.065 oC

Signatories:

1. Ching Kar Chien


2. Chew Chii Tsong

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 14 of 30

SCOPE OF CALIBRATION: TEMPERATURE

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Temperature Measurement
Simulator using multimeter
based on ITS-90
K -200 to 0 oC 0.19 oC Table
0 to 1370 oC 0.13 oC
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

J -210 to 0 oC 0.16 oC
0 to 1200 oC 0.11 oC

T -200 to 0 oC 0.18 oC
0 to 400 oC 0.10 oC

E -200 to 0 oC 0.13 oC
0 to 1000 oC 0.10 oC

N -200 to 0 oC 0.25 oC
0 to 1300 oC 0.11 oC

B 250 to 600 oC 0.93 oC


600 to 1820 oC 0.23 oC

R -50 to 0 oC 0.64 oC
0 to 1760 oC 0.21 oC

S -50 to 0 oC 0.60 oC
0 to 1760 oC 0.24 oC

Pt 100 -200 to 850 oC 0.059 oC

Signatories:

1. Ching Kar Chien


2. Chew Chii Tsong

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 15 of 30

SCOPE OF CALIBRATION: TEMPERATURE

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Temperature
Simulator

K -100 ˚C to 1300 ˚C 0.19 ˚C Measurement


Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

using
J -100 ˚C to 1200 ˚C 0.16 ˚C Temperature
Calibrator
T -100 ˚C to 400 ˚C 0.18 ˚C

E -100 ˚C to 950 ˚C 0.13 ˚C

N -200 ˚C to 1300 ˚C 0.25 ˚C

B 600 to 1800 ˚C 0.93 ˚C

R 0 to 1700 ˚C 0.64 ˚C

S 0 to 1700 ˚C 0.60 ˚C

Pt 100 -100 to 800 ˚C 0.059 ˚C

Temperature Electrical
Indicator simulation using
Temperature
K -100 to 1300 ˚C 0.14 ˚C Calibrator based
on
J -100 to 1200 ˚C 0.12 ˚C EURAMET/cg-
11/v.01
T -100 to 400 ˚C 0.12 ˚C

E -100 to 950 ˚C 0.10 ˚C

N -200 ˚C to 1300 ˚C 0.16 ˚C

B 600 to 1800 ˚C 0.50 ˚C

R 0 to 1700 ˚C 0.35 ˚C

S 0 to 1700 ˚C 0.33 ˚C

Pt 100 -100 to 800 ˚C 0.065 ˚C

Signatories:

1. Ching Kar Chien


2. Chew Chii Tsong

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 16 of 30

SCOPE OF CALIBRATION: TEMPERATURE

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Temperature -25 oC to 500 oC 0.58 oC Using
Enclosure temperature
sensor / logger
based on AS
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

2853-1986
Humidity 10 % RH to 98 % RH 2.5 % RH Using Dry bulb &
Enclosure Wet bulb sensor
10 oC to 95 oC 0.58 oC / logger based
on JTM K01-
1991

Temperature -10 oC to 0 oC 0.12 oC Comparison with


Sensor 0 oC to 100 oC 0.10 oC PT 100 in Liquid
100 oC to 300 oC 0.12 oC Bath and
300 oC to 500 oC 0.51 oC Temperature
Block Calibrator

Signatories:

1. Ching Kar Chien


2. Chew Siou Lian
3. Chew Chii Tsong

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 17 of 30

SCOPE OF CALIBRATION: VOLUME

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Volumetric 1 to 100 µl 0.17 µl Gravimetric
Apparatus Above 100 to 200 µl 0.24 µl Method
(To Deliver) Above 200 to 1000 µl 3.1 µl
Above 1000 to 2000 µl 6.5 µl
Piston-Operated Above 2000 to 5000 µl 7.2 µl
Pipette, Type A
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

One-Mark 1 to 10 ml 0.0066 ml
Pipette Above 10 to 100 ml 0.019 ml
Above 100 to 200 ml 0.029 ml
Above 200 to 500 ml 0.069 ml
Above 500 to 1000 ml 0.15 ml

Volumetric 1 to 10 ml 0.041 ml
Apparatus Above 10 to 100 ml 0.24 ml
(To Contain) Above 100 to 200 ml 0.35 ml
Above 200 to 500 ml 0.58 ml
Measuring Above 500 to 1000 ml 1.6 ml
Cylinder Flask

Signatories:

1. Chew Siou Lian


2. Ching Kar Chien

SCOPE OF CALIBRATION: VOLUME

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*

pH Meter 4.01 pH 0.03 pH Standard Buffer


7.01 pH 0.03 pH Solutions
10.01 pH 0.06 pH

Refractometer / 0 % to 50 % Brix ± 0.32 % Brix Sucrose Powder


Sucrose Mass (1.333 nD ~ 1.420 nD) with analytical
Fraction / balance
Refractive Index

Signatories:

1. Chew Siou Lian


2. Ching Kar Chien

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 18 of 30

SCOPE OF CALIBRATION: ELECTRICAL

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Measuring Generation using
Instrument calibrator model
0 V to 320 mV 5.7 μV/V + 1.1 μV Wavetek 9100
DC Voltage 0.32 V to 3.2 V 14 μV/V - 1.4 μV
3.2 V to 32 V 13 μV/V + 2.4 μV
32 V to 320 V 12 μV/V + 33 μV
320 V to 1050 V 16 μV/V - 1.5 mV
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

AC Voltage 0 V to 1050 V See Matrix A

DC Current 0 A to 320 µA 390 μA/A + 5.2 mA


0.32 mA to 3.2 mA 5.9 μA/A + 0.13 mA
3.2 mA to 32 mA 34 μA/A + 0.038 mA
32 mA to 320 mA 40 μA/A - 0.14 mA
0.32 A to 3.2 A 0.83 mA/A - 0.26 mA
3.2 A to 20 A 0.58 mA/A + 0.55 mA

AC Current 0 A to 20 A See Matrix B

Resistance 0 Ω to 40 Ω 0.055 mΩ/Ω + 0.18 mΩ


40 Ω to 400 Ω 0.017 mΩ/Ω + 1.7 mΩ
400 Ω to 4 kΩ 0.024 mΩ/Ω - 0.96 mΩ
4 kΩ to 40 kΩ 0.019 mΩ/Ω + 19 mΩ
40 kΩ to 400 kΩ 0.036 mΩ/Ω – 0.67 Ω
400 kΩ to 4 MΩ 0.14 mΩ/Ω - 42 Ω
4 MΩ to 40 MΩ 0.020 mΩ/Ω + 0.43 kΩ
40 MΩ to 400 MΩ 0.25 mΩ/Ω - 9.2 kΩ

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 19 of 30

Matrix A
AC Voltage Measurement
Frequency
Range Hz kHz
0 to 10 10 to 50 0.05 to 1 1 to 30 30 to 50 50 to 100
0.0007 + 0.0027 + 0.0024 + 0.084 0.00028 + 0.00093 + 0.21 0.00028 + 0.47
0 mV to 320 mV
0.12 0.084 0.21
0.34 + 0.006 0.24 + 0.0093 0.2 + 0.0092 0.66 + 1.6 – 0.29 1.5 + 0.00010
320 mV to 3.2 V
0.00010
3.2 V to 32 V 0.59 – 0.79 0.48 – 0.77 0.48 – 0.77 1.2 – 1.6 0.62 + 2.7 2.2 – 2.4
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

32 V to 320 V 0.34 + 7.2 0.35 + 3.4 0.35 + 3.4 0.11 +32


320 V to 1050 V 0.63 - 84 0.63 - 84 0.98 - 195
The expanded uncertainties given in above table are expressed in mV/V + mV

Matrix B
AC Current Measurement

Frequency Frequency
Range
(Hz ) ( kHz )
0 to 10 0.01 to 3 3 to 10 10 to 30
0 to 320 µA 0.11 mA/A 0.096 mA/A 0.068 mA/A 0.089 mA/A
+ 285 mA + 285 mA + 285 mA + 285 mA
320 µA to 3.2 mA 0.16 mA/A 0.27 mA/A 0.22 mA/A 0.22 mA/A
+ 0.27 μA + 0.23 μA + 0.24 μA + 0.25 μA
3.2 mA to 32 mA 0.39 mA/A 0.40 mA/A 0.40 mA/A 0.40 mA/A
- 0.49 μA - 0.18 μA - 0.35 μA - 0.35 μA
32 mA to 320 mA 0.95 mA/A 0.95 mA/A 0.95 mA/A 0.95 mA/A
- 18 μA - 18 μA - 18 μA - 18 μA
0.32 A to 3.2 A 0.73 mA/A 0.69 mA/A 0.69 mA/A -
- 0.053 mA + 0.067 mA + 0.067 mA
3.2 A to 20 A - 0.35 mA/A - -

Signatories:

1. Kenny Ching Ah Hoo


2. Gabriel Ching Kar Ee
3. Ching Kar Chien

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 20 of 30

SCOPE OF CALIBRATION: ELECTRICAL

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Measuring Generation
Instruments using Calibrator
model Wavetek
Frequency 0.5 Hz to 320 Hz 29 ppm 9100
320 Hz to 3.2 kHz 29 ppm
3.2 kHz to 32 kHz 29 ppm
32 kHz to 3.2 MHz 29 ppm
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

3.2 MHz to 10 MHz 29 ppm

Capacitance 0.5 to 4 nF 4.9 mF/F


4 nF to 40 nF 3.7 mF/F
40 nF to 400 nF 2.4 mF/F
400 nF to 4 µF 2.7 mF/F
4 µF to 40 µF 3.7 mF/F
40 µF to 400 µF 3.9 mF/F
400 µF to 4 mF 3.9 mF/F
4 mF to 40 mF 5.1 mF/F

Direct
Resistance 10 mΩ to 90 mΩ 8.7 mΩ/Ω measurement
100 mΩ to 900 mΩ 1.2 mΩ/Ω with TIME
1 Ω to 9 Ω 0.24 mΩ/Ω Decade
10 Ω to 90 Ω 0.17 mΩ/Ω Resistance Box
100 Ω to 900 Ω 0.16 mΩ/Ω
1 kΩ to 9 kΩ 0.16 mΩ/Ω
10 kΩ to 90 kΩ 0.16 mΩ/Ω
100 kΩ to 900 kΩ 0.16 mΩ/Ω
1 MΩ to 9 MΩ 0.16 mΩ/Ω
10 MΩ to 90 MΩ 0.16 mΩ/Ω
100 MΩ to 199.99 MΩ 0.16 mΩ/Ω

Clamp Meter

DC Current (+/- polarities)


- 50 Turn Coil 1.6 A to 16 A 0.37 mA/A Generating using
16 A to 160 A 0.37 mA/A Calibrator model
160 A to 525 A 0.12 mA/A Wavetek 9100
525 A to 1000 A 0.58 mA/A with Current Coil

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 21 of 30

SCOPE OF CALIBRATION: ELECTRICAL

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Clamp Meter

AC Current 1.6 A to 16 A
- 50 Turn Coil 10 Hz to 100 Hz 1.3 mA/A Generating using
100 Hz to 440 Hz 3.2 mA/A Calibrator model
Wavetek 9100
16 A to 160 A with Current Coil
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

10 Hz to 100 Hz 1.4 mA/A


100 Hz to 440 Hz 3.4 mA/A

160 A to 1000 A
10 Hz to 100 Hz 1.3 mA/A
100 Hz to 440 Hz 3.2 mA/A

Generating
Instruments

DC Voltage 0 mV to 100 mV 36 nV/V + 1.7 µV


100 mV to 1 V 7.7 µV/V + 0.85 µV
1 V to 10 V 6.9 µV/V + 1.6 µV
10 V to 100 V 8.3 µV/V – 12 µV
100 V to 1000 V 9.6 µV/V – 140 µV

AC Voltage 0 to 1000 V See Matrix C

DC Current 0 µA to 100 µA 46 mA/A + 17 nA


100 µA to 1 mA - 43 µA/A + 67 nA Measurement
1 mA to 10 mA 4.9 µA/A + 19 nA using KeySight
10 mA to 100 mA 5.6 µA/A + 12 nA 3458A Digital
100 mA to 1 A 0.12 mA/A - 12 µA Multimeter

Measurement
1 A to 3 A 1.8 mA/A - 1.7 mA using Agilent
34401A
Multimeter

AC Current 0 to 3 A See Matrix D

Resistance 0 Ω to 10 Ω 42 µΩ/Ω + 0.32 mΩ


10 Ω to 100 Ω 15 µΩ/Ω + 0.18 mΩ
100 kΩ to 1 kΩ 13 µΩ/Ω + 0.44 mΩ
1 kΩ to 10 kΩ 13 µΩ/Ω - 0.15 mΩ
10 kΩ to 100 kΩ 15 µΩ/Ω - 21 mΩ
100 kΩ to 1 MΩ 26 µΩ/Ω - 1.1 Ω
1 MΩ to 10 MΩ 73 µΩ/Ω - 49 Ω
10 MΩ to 100 MΩ 360 µΩ/Ω - 2.9 kΩ

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 22 of 30

SCOPE OF CALIBRATION: ELECTRICAL

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Generating
Instruments

Frequency 1 Hz to 40 Hz 566 ppm


40 Hz to 100 Hz 114 ppm
100 Hz to 100 kHz 114 ppm
100 kHz to 10 MHz 127 ppm
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

Capacitance 1 µF to 10 µF 40 µF/F + 150 nF Measuring using


10 µF to 100 µF 65 mF/F – 0.50 µF SANWA CD
800a Multimeter

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 23 of 30

AC Voltage:

Matrix C

Frequecny Hz kHz
Voltage
0 to 40 0.04 to 1 1 to 20 20 to 50 50 to 100
0 to 10 mV 2.9 mV/V 0.88 mV/V 0.55 mV/V 0.31 mV/V 1.1 mV/V
+ 19 μV + 3.7 μV + 7.4 μV + 15 μV + 15 μV
10 mV to 100 mV - 0.23 mV/V 0.12 mV/V 0.11 mV/V 0.80 mV/V 0.71 mV/V
+ 52 μV + 12 μV + 12 μV + 9.6 μV + 18 μV
100 mV to 1 V 109 μV/V 114 μV/V 102 μV/V 184 μV/V 185 μV/V
+ 18 μV + 12 μV + 12 μV + 71 μV + 70 μV
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

1 V to 10 V 76 μV/V 77 μV/V 69 μV/V 188 μV/V 190 μV/V


+ 50 μV + 49 μV + 45 μV + 67 μV + 66 μV
10 V to 100 V 79 μV/V 80 μV/V 82 μV/V 245 μV/V 245 μV/V
+ 19 μV + 19 μV - 81 μV - 500 μV - 488 μV
100 V to 1000 V 68 μV/V 68 μV/V - - -
+ 1.2 mV + 1.2 mV

AC Current:

Matrix D

Frequency Hz kHz Measuring


Voltage Instrument
0 to 50 0.05 to 1 1 to 5 5 to 30
0 A to 1 mA 0.26 mA/A 0.17 mA/A 0.21 mA/A 0.92 mA/A KeySight
+ 0.38 μA + 70 nA + 59 nA + 59 nA 3458A
1 mA to 10 mA 0.19 mA/A 0.23 mA/A 0.23 mA/A 0.15 mA/A KeySight
+ 0.45 μA + 9.9 nA + 45 nA + 0.84 nA 3458A
10 mA to 100 mA 0.21 mA/A 0.21 mA/A 0.21 mA/A 0.21 mA/A KeySight
+ 0.21 μA + 0.17 μA + 0.17 μA + 0.17 μA 3458A
100 mA to 1 A 0.46 mA/A 0.45 mA/A 0.45 mA/A - KeySight
- 25 μA - 25 μA - 25 μA 3458A
1 A to 3 A 1.6 mA/A 1.6 mA/A 1.6 mA/A - Agilent
- 1.1 mA - 1.1 mA - 1.1 mA 34401A

Signatories:

1. Gabriel Ching Kar Ee


2. Ching Kar Chien

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 24 of 30

SCOPE OF CALIBRATION: ELECTRICAL

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Measuring Generation using
Instruments Calibrator model
Wavetek 9100
DC Voltage 0 mV to 320 mV 5.7 μV/V + 1.1 μV
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

0.32 V to 3.2 V 14 μV/V - 1.4 μV


3.2 V to 32 V 13 μV/V + 2.4 μV
32 V to 320 V 12 μV/V + 33 μV
320 V to 1050 V 16 μV/V - 1.5 mV

AC Voltage 0 V to 1050 V See Matrix A

DC Current 0 µA to 320 µA 390 μA/A + 5.2 mA


0.32 mA to 3.2 mA 5.9 μA/A + 0.13 mA
3.2 mA to 32 mA 34 μA/A + 0.038 mA
32 mA to 320 mA 40 μA/A - 0.14 mA
0.32 A to 3.2 A 0.83 mA/A - 0.26 mA
3.2 A to 20 A 0.58 mA/A + 0.55 mA

AC Current 0 A to 20 A See Matrix B

Resistance 0 Ω to 40 Ω 0.055 mΩ/Ω + 0.18 mΩ


40 Ω to 400 Ω 0.017 mΩ/Ω + 1.7 mΩ
400 Ω to 4 kΩ 0.024 mΩ/Ω - 0.96 mΩ
4 kΩ to 40 kΩ 0.019 mΩ/Ω + 19 mΩ
40 kΩ to 400 kΩ 0.036 mΩ/Ω – 0.67 Ω
400 kΩ to 4 MΩ 0.14 mΩ/Ω - 42 Ω
4 MΩ to 40 MΩ 0.020 mΩ/Ω + 0.43 kΩ
40 MΩ to 400 MΩ 0.25 mΩ/Ω - 9.2 kΩ

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 25 of 30

SCOPE OF CALIBRATION: ELECTRICAL

SITE: CATEGORY I

Matrix A
AC Voltage Measurement

Frequency
Range Hz kHz
0 to 10 10 to 50 0.05 to 1 1 to 30 30 to 50 50 to 100
0.0007 + 0.0027 + 0.0024 + 0.084 0.00028 + 0.00093 + 0.21 0.00028 + 0.47
0 mV to 320 mV
0.12 0.084 0.21
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

0.34 + 0.006 0.24 + 0.0093 0.2 + 0.0092 0.66 + 1.6 – 0.29 1.5 + 0.00010
320 mV to 3.2 V
0.00010
3.2 V to 32 V 0.59 – 0.79 0.48 – 0.77 0.48 – 0.77 1.2 – 1.6 0.62 + 2.7 2.2 – 2.4
32 V to 320 V 0.34 + 7.2 0.35 + 3.4 0.35 + 3.4 0.11 +32
320 V to 1050 V 0.63 - 84 0.63 - 84 0.98 - 195
The expanded uncertainties given in above table are expressed in mV/V + mV

Matrix B
AC Current Measurement

Frequency Frequency
Range
(Hz ) ( kHz )
0 to 10 0.01to 3 3 to 10 10 to 30
0 to 320 µA 0.11 mA/A 0.096 mA/A 0.068 mA/A 0.089 mA/A
+ 285 mA + 285 mA + 285 mA + 285 mA
320 µA to 3.2 mA 0.16 mA/A 0.27 mA/A 0.22 mA/A 0.22 mA/A
+ 0.27 μA + 0.23 μA + 0.24 μA + 0.25 μA
3.2 mA to 32 mA 0.39 mA/A 0.40 mA/A 0.40 mA/A 0.40 mA/A
- 0.49 μA - 0.18 μA - 0.35 μA - 0.35 μA
32 mA to 320 mA 0.95 mA/A 0.95 mA/A 0.95 mA/A 0.95 mA/A
- 18 μA - 18 μA - 18 μA - 18 μA
0.32 A to 3.2 A 0.73 mA/A 0.69 mA/A 0.69 mA/A -
- 0.053 mA + 0.067 mA + 0.067 mA
3.2 A to 20 A - 0.35 mA/A - -
The expanded uncertainties given in above table are expressed in mA/A

Signatories:

1. Kenny Ching Ah Hoo


2. Gabriel Ching Kar Ee
3. Ching Kar Chien

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 26 of 30

SCOPE OF CALIBRATION: ELECTRICAL

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Measuring Generation using
Instruments Calibrator model
Wavetek 9100
Frequency 0.5 Hz to 320 Hz 29 ppm
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

320 Hz to 3.2 kHz 29 ppm


3.2 kHz to 32 kHz 29 ppm
32 kHz to 3.2 MHz 29 ppm
3.2 MHz to 10 MHz 29 ppm

Capacitance 0.5 nF to 4 nF 4.9 mF/F


4 nF to 40 nF 3.7 mF/F
40 nF to 400 nF 2.4 mF/F
400 nF to 4 µF 2.7 mF/F
4 µF to 40 µF 3.7 mF/F
40 µF to 400 µF 3.9 mF/F
400 µF to 4 mF 3.9 mF/F
4 mF to 40 mF 5.1 mF/F

Direct
Resistance 10 mΩ to 90 mΩ 8.7 mΩ/Ω measurement
100 mΩ to 900 mΩ 1.2 mΩ/Ω with TIME
1 Ω to 9 Ω 0.24 mΩ/Ω Decade
10 Ω to 90 Ω 0.17 mΩ/Ω Resistance Box
100 Ω to 900 Ω 0.16 mΩ/Ω
1 kΩ to 9 kΩ 0.16 mΩ/Ω
10 kΩ to 90 kΩ 0.16 mΩ/Ω
100 kΩ to 900 kΩ 0.16 mΩ/Ω
1 MΩ to 9 MΩ 0.16 mΩ/Ω
10 MΩ to 90 MΩ 0.16 mΩ/Ω
100 MΩ to 199.99 MΩ 0.16 mΩ/Ω

Generating using
Clamp Meter Calibrator model
Wavetek 9100
DC Current (+/- polarities) with Current Coil
- 50 Turn Coil 1.6 to 16 A 0.37 mA/A
16 to 160 A 0.37 mA/A
160 to 525 A 0.12 mA/A
525 to 1000 A 0.58 mA/A

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 27 of 30

SCOPE OF CALIBRATION: ELECTRICAL

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*
Clamp Meter

AC Current 1.6 to 16 A
- 50 Turn Coil 10 to 100 Hz 1.3 mA/A Generating using
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

100 to 440 Hz 3.2 mA/A Calibrator model


Wavetek 9100
16 to 160 A with Current Coil
10 to 100 Hz 1.4 mA/A
100 to 440 Hz 3.4 mA/A

160 to 1000 A
10 to 100 Hz 1.3 mA/A
100 to 440 Hz 3.2 mA/A

Generating Measurement
Instruments Keysight 3458A
Digital
DC Voltage 0 to 100 mV 3.2 nV/V + 1.7 μV Multimeter
100 mV to 1 V 7.7 μV/V + 0.85 μV
1 V to 10 V 6.9 μV/V + 1.6 μV
10 V to 100 V 8.3 μV/V - 12 μV
100 V to 1000 V 9.6 μV/V - 140 μV

AC Voltage 0 to 1000 V See Matrix C

DC Current 0 to 10 mA 46 mA/A + 17 nA
100 μA to 1 mA - 43 μA /A + 67 nA
1 mA to 10 mA 4.9 μA /A + 19 nA
10 mA to 100 mA 5.6 μA/A + 12 nA
100 mA to 1 A 0.12 mA/A - 12 μA
1 A to 3 A 18 mA/A – 17 mA

AC Current 0 to 3 A See Matrix D

Resistance 0 to 10 Ω 42 nΩ/Ω + 0.32 mΩ


10 Ω to 100 Ω 15 μΩ/Ω + 0.18 mΩ
100 Ω to 1 kΩ 13 μΩ/Ω + 0.44 mΩ
1 kΩ to 10 kΩ 13 μΩ/Ω - 0.15 mΩ
10 kΩ to 100 kΩ 15 μΩ/Ω - 21 mΩ
100 kΩ to 1 MΩ 26 μΩ/Ω - 1.1 Ω
1 MΩ to 10 MΩ 73 μΩ/Ω - 49 Ω
10 MΩ to 100 MΩ 360 μΩ/Ω - 2.9 kΩ

Frequency 1 Hz to 40 Hz 566 ppm


40 Hz to 100 Hz 114 ppm
100 Hz to 100 kHz 114 ppm
100 kHz to 10 MHz 127 ppm

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 28 of 30

SCOPE OF CALIBRATION: ELECTRICAL

SITE: CATEGORY I

Generating Instruments

AC Voltage:

Matrix C

Frequecny Hz kHz
Voltage
0 to 40 0.04 to 1 1 to 20 20 to 50 50 to 100
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

0 to 10 mV 2.9 mV/V 0.88 mV/V 0.55 mV/V 0.31 mV/V 1.1 mV/V
+ 19 μV + 3.7 μV + 7.4 μV + 15 μV + 15 μV
10 mV to 100 mV - 0.23 mV/V 0.12 mV/V 0.11 mV/V 0.80 mV/V 0.71 mV/V
+ 52 μV + 12 μV + 12 μV + 9.6 μV + 18 μV
100 mV to 1 V 109 μV/V 114 μV/V 102 μV/V 184 μV/V 185 μV/V
+ 18 μV + 12 μV + 12 μV + 71 μV + 70 μV
1 V to 10 V 76 μV/V 77 μV/V 69 μV/V 188 μV/V 190 μV/V
+ 50 μV + 49 μV + 45 μV + 67 μV + 66 μV
10 V to 100 V 79 μV/V 80 μV/V 82 μV/V 245 μV/V 245 μV/V
+ 19 μV + 19 μV - 81 μV - 500 μV - 488 μV
100 V to 1000 V 68 μV/V 68 μV/V - - -
+ 1.2 mV + 1.2 mV
The expanded uncertainties given in above table are expressed in mV/V

AC Current:

Matrix D

Frequency Hz kHz Measuring


Voltage 0 to 50 0.05 to 1 1 to 5 5 to 30 Instrument
0 A to 1 mA 0.26 mA/A 0.17 mA/A 0.21 mA/A 0.92 mA/A KeySight
+ 0.38 μA + 70 nA + 59 nA + 59 nA 3458A
1 mA to 10 mA 0.19 mA/A 0.23 mA/A 0.23 mA/A 0.15 mA/A KeySight
+ 0.45 μA + 9.9 nA + 45 nA + 0.84 nA 3458A
10 mA to 100 mA 0.21 mA/A 0.21 mA/A 0.21 mA/A 0.21 mA/A KeySight
+ 0.21 μA + 0.17 μA + 0.17 μA + 0.17 μA 3458A
100 mA to 1 A 0.46 mA/A 0.45 mA/A 0.45 mA/A - KeySight
- 25 μA - 25 μA - 25 μA 3458A
1 A to 3 A 1.6 mA/A 1.6 mA/A 1.6 mA/A - Agilent
- 1.1 mA - 1.1 mA - 1.1 mA 34401A
The expanded uncertainties given in above table are expressed in mA/A

Signatories:

1. Gabriel Ching Kar Ee


2. Ching Kar Chien

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 29 of 30

SCOPE OF CALIBRATION: TIME AND FREQUENCY

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*

Stopwatch and Calibrated by


Timer 1 sec to 60 minutes 0.063 sec using
comparison
method
according to
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

NIST stopwatch
and timer
calibration.

Tachometer 0 to 2000 rpm 2.6 rpm Comparison with


(Non-contact) reference
2000 to 5000 rpm 7.1 rpm tachometer.

Signatories:

1. Gabriel Ching Kar Ee


2. Ching Kar Chien
3. Chew Chii Tsong

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA
Schedule
Issue date: 2 November 2017
Valid until: 15 October 2020

NO: SAMM 256 MS ISO/IEC 17025


(Issue 2, 2 November 2017 replacement
of SAMM 256 dated 28 September 2017)
Page: 30 of 30

SCOPE OF CALIBRATION: TIME AND FREQUENCY

SITE: CATEGORY I

Instrument Calibration and


Calibrated/ Measurement
Range Remarks
Measurement Capability Expressed
Parameter as an Uncertainty(±)*

Stopwatch and Calibrated by


Timer 0 to 60 sec 0.12 sec using
comparison
Scan this QR Code or visit www.jsm.gov.my/cab-directories for the current scope of accreditation

1 minute to 5 minute 0.12 sec method


according to
5 minute to 60 minute 0.34 sec NIST stopwatch
and timer
calibration.

Tachometer 0 to 2000 rpm 2.6 rpm Comparison with


(Non-contact) reference
2000 to 5000 rpm 7.1 rpm tachometer.

Signatories:

1. Gabriel Ching Kar Ee


2. Ching Kar Chien
3. Chew Chii Tsong

SKIM AKREDITASI MAKMAL MALAYSIA (SAMM)


LABORATORY ACCREDITATION SCHEME OF MALAYSIA