Nephros DSU-D – Dialysis Installation Instructions

PARTS LIST - Connect a quick connector to the free end of the old hose to
serve as the DSU-D filter inlet.
Filter Mounting
- Snap a DSU-D into the wall mount clamps.
- Place the filter and clamps vertically against the wall in the
DSU-D In-Line Filter position they will be mounted (see photo).

Wall Mount
Clamp (2)

Sample Port

Speedfit Fitting (2)


John Guest Check Valve

Quick
Connectors (2)

LOCATION
The DSU-D may be installed at the point of use (where the - Mark the hole locations of the mounting clamps.
dialysis machine is connected to the loop or an individual RO) - Remove the filter from the clamps and set it aside.
or at the beginning of the dialysis bicarbonate distribution loop. - Mount the clamps with their enclosed screws.
- Remove the DSU-D from its package and press it into the two
PROCEDURE clamps making sure the direction of flow and the label arrows
are the same.
Note: Prior to handling new filters it is recommended that one - Connect quick connector to new distribution line (filter outlet)
wash their hands and wear disposable gloves. tubing.
Point of Use Installation - Cut the distribution line tubing 6-12 inches from the quick
- Turn off water supply to the dialysis machine from the wall connector and install the sample port.
box or individual RO outlet. - Cut tubing again 6-12 inches from sample port and install the
- Disconnect water inlet hose from the back of the dialysis John Guest check valve with Speedfit fittings. Make sure the
machine and replace with a new hose to avoid potential back- direction of flow of the check valve is correct.
contamination of the filter from the old hose. (see photo) - Having the components installed in the correct order (see
photo) is necessary to avoid potential back-contamination.
- Connect the inlet and outlet quick connectors as per the
New Inlet diagram.
Hose
Sample
Port

Check
Valve

Manufacturer:
Nephros Inc.
41 Grand Ave
Point of Use (continued) River Edge, NJ 07661 USA
- Turn on the water supply and prime the DSU-D to drain via 201-343-5202
the sample port for several minutes to purge air out of the 50-3015 r.4
filter.
- Following installation it is recommended to disinfect the lines
downstream of the DSU-D as per standard clinic procedures.
- The DSU-D is compatible with typically used disinfectants.
See the filter IFU for details.
Bicarbonate Loop Installation
- Turn off the bicarbonate pump.
- Disconnect the supply line from the bicarbonate tank at the
inlet to the distribution loop and replace it with a new hose to
avoid potential back-contamination of the filter.
- Connect a quick connector to the free end of the line from the
tank outlet.
- Perform the Filter Mounting steps.
- Connect the inlet and outlet quick connectors as per the
diagram.
- Turn on the bicarbonate pump. Prime the DSU-D to drain via
the sample port for several minutes to purge air out of the
filter.
- Following installation it is recommended to disinfect the lines
downstream of the DSU-D as per standard clinic procedures.
- The DSU-D is compatible with typically used disinfectants.
See the filter IFU for details