Академический Документы
Профессиональный Документы
Культура Документы
Numeric/User 1 Enter
Numeric/User 1 Enter
ing temperature changes, the Agilent 3458A provides the TEMP? command (Ready to accept commands.)
Example 1: Making a measurement with any function after turning The shifted Menu Keys allow alphabetic access to all (found in the command Menu) to measure the internal instrument tempera-
on the Agilent 3458A. The ACV function serves as a typical example Agilent 3458A commands and command parameters. The com- ture. Refer to the User’s Guide for more details. ACAL keystroke sequence. FUNC ACV (blue) E
(three times)
of the immediate execute keys found in the Function/Range key mands and command parameters are scrolled from A to Z by using Auto
Cal
, 3 4 5 8 Enter (four times) (blue) ;
group. Connect the appropriate signal to the input terminals and the Menu Scroll keys ( and ) found in the Function/Range (The delimiter used to separate commands is " ; ".
press the keys shown. keys. Command parameters are accessed by using the Display/
Window keys and , also found in the Function/Range Configuration for Highly Accurate Measurements The delimiter used to separate parameters is " , ".)
keys, after the command is located. with the Agilent 3458A SETACV SYNC (blue) S (blue) ;
Command Key Group Key Sequence (You should perform an ACAL every 24 hours for the appropriate function after the RES .002 (blue) R . 0 0 2 (blue) ;
ACV Function/Range ACV Example 3: Making a very precise Vac measurement using the Syn- Agilent 3458A has been powered-on and in a stable temperature environment for a minimum LFILTER ON (blue) L Enter
(Any other function key in the Function/Range Key Group acts chronous ACV measurement function. (At power-on, the Agilent of four hours.)
3458A is set to the generally more useful Analog ACV measurement DEFKEY F1, " 1
similarly; the Agilent 3458A changes function, automatically
ranges, and begins measurements.) function if you select ACV, but you should use Synchronous ACV for Function Comment Key Sequence MATH NULL (blue) L
(four times)
very precise measurements of periodic waveforms.) (eight times) Enter
RESET Function/Range (blue) Reset ____________________________DCV____________________________ (MATH NULL displays the difference between the first and subse-
Command Key Group Key Sequence RESET Power-on state (VDC) quent measurements.)
(RESET returns the Agilent 3458A to the powered-on state.) (blue) Reset
RESET Function/Range (blue) Reset
NPLC 200 Integration Time (1) NPLC 2 0 0 Enter
Now Use It!
ACV Function/Range ACV
NDIG 8 8 1/2 Digits (blue) N 8 Enter (blue) f0 Enter
SETACV SYNC Function/Range (blue)
(blue) f1 Enter
Menu S ____________________________ACV____________________________
Example 2: Making a precision Vdc measurement on the 10 V range. (Note the MATH annunciator is turned on.)
Function/Range RESET Power-on state (blue) Reset
Numeric/User Enter Command Recall Buffer Operation
Command Key Group Key Sequence ACV Function Vac ACV
(The MORE INFO annunciator should be illuminated. The MORE The command recall buffer recalls previously entered commands. If you
RESET Function/Range (blue) Reset SETACV SYNC Synchronous AC (blue) S Enter
have performed the sequence in the User Definable Keys you can expect to
INFO annunciator indicates that some part of the measurement
DCV Function/Range DCV
configuration is in a non-standard state. To view this information RES .002 Max AC Reading Res. (blue) R . 0 0 2 Enter
have the results listed below.
10 V Range Function/Range or press . Your display should show SETACV SYNC. Now press LFILTER ON Internal LP Filter (2) (blue) L Enter
Command Key Group Key Sequence
(The and keys are pressed until the desired range is again; the display is returned to the measurement state.) __________________________OHMS____________________________ RECALL Function/Range (blue)
displayed: in this case, XX.XXXXX V DC (X is any number). RES .002 Function/Range (blue) RESET Power-on state (blue) Reset Numeric/User Recall
Leading zeros are always displayed for easy range identifica-
Menu R OHMF Function Four-Wire Ohms (blue) OHMF (MATH NULL should appear in the Display.)
tion.)
Function/Range OCOMP ON For R <= 100 kOhms (3) Offset
Comp 9
Enter
The MENU Command
Numeric/User . 0 0 2 Enter
NPLC 100 Menu NPLC
DELAY 1
(The RES command sets the % resolution of the measurement. A Adds 1 s delay (4) (blue) C 1 Enter To access the commands alphabetically, you use the Menu Keys. As the
Numeric/User 1 0 0 Enter
Agilent 3458A is shipped, only the most commonly used commands are
function change causes RES to default to RES .01 (0.01%) with NPLC 200 Integration Time NPLC 2 0 0 Enter
(Measurement will take about 3 seconds.) available in this "SHORT" menu. To access all the commands, you have to use
reading rate of 1 per 3 s.) 1. Longer integration times reduce measurement noise and increase measurement resolution. the Menu Key with the "FULL' parameter.
NDIG 8 Function/Range (blue) 2. Recommended for frequencies less than 50kHz.
LFILTER ON Function/Range (blue) Command Key Group Key Sequence
Menu N 3. OCOMP ON turns on the offset compensation feature of the Agilent 3458A. OCOMP ON min-
Menu L
Numeric/User 8 Enter
Function/Range imizes the effect of any thermally generated offset voltage from creating errors in resistance MENU FULL Function/Range (blue)
measurements. Numeric/User Menu
(Sets the display to 8 1/2 digits.) Numeric/User Enter
4. Adds 1 s settling delay to reduce effects of dielectric absorption errors of cabling and the Function/Range
(Suggested for inputs with frequencies below 50kHz.) Enter
resistor under test Numeric/User
*03458-90007*
03458-90007