Академический Документы
Профессиональный Документы
Культура Документы
Rotating
Anode CIRCLEX
X-ray tube
Assembly 0.3/0.8P324&0.6/1.2P324DK-85
GENERAL
The Shimadzu CIRCLEX Focal spot combination of
0.3/0.8P324DK-85 & 0.3mm/0.8mm (for over-table) or
0.6/1.2P324DK-85, Rotating Anode 0.6mm/1.2mm (for C-arm table
X-ray tube assemblies are rated to Cvision & over-table) is available.
150kV and feature a 100mm (4inch)
target of Rhenium-tungsten faced CONFIGURATION
Molybdenum with a target angle 12˚. The CIRCLEX 0.3/0.8P324DK-85&
Anode heat content is 280kJ {400kHU} 0.6/1.2P324DK-85, Rotating Anode
and the maximum anode heat X-ray tube assemblies are composed
dissipation rate is 1600W {2250HU/s}. as follows.
The anticrystalization agent is mixed
in tungsten, protecting the surface of
the anode from becoming rough and CIRCLEX CIRCLEX
cracked, so that reduction of effective Model 0.3/0.8P324DK-85 0.6/1.2P324DK-85
X-ray dose is minimized over a long (Nominal focal spot; 0.3/0.8mm) (Nominal focal spot; 0.6/1.2mm)
time of use. The target disk coated Supporter B-80
with high heat emission material Fan F-3
enables stable X-ray output.
The CIRCLEX 0.3/0.8P324DK-85 &
0.6/1.2P324DK-85 are designed for
gastro-intestinal examination,
high-speed serialography.
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
SPECIFICATIONS
CIRCLEX CIRCLEX
0.3/0.8P324DK-85 0.6/1.2P324DK-85
X-ray tube Max. Anode stationary 200W {280HU/s} 355W {500HU/s} 355W {500HU/s} 425W {600HU/s}
Fluoroscopic load (10min.) Anode rotating 355W {500HU/s} 425W {600HU/s} 425W {600HU/s} 425W {600HU/s}
Leakage radiation in hour from the X-ray tube assembly and collimator
Is less than 0.87mGy at a distance of 1 meter from the focal spot.
Leakage radiation
However, leakage radiation in an hour from the collimator is less than
0.304mGy
The thermal switch actuates to break the high voltage generating circuit
Safety circuit
when the temperature inside the tube housing reaches 80℃.
*1 Refers to the maximum value usable in tube current regulation, and its limit in constant use is less than 4.6A for
focus of 0.3mm, less than 5.2A for focus of 0.8mm, and less than 5.2A for both of 0.6mm focus and 1.2mm focus.
*2 Combination with the optional SA-40, SA-41, SA-42, SA-50 and SA-60, starter.
2
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
HEATING AND COOLING CHARTS
X-ray Tube
X-RAY TUBE ASSEMBLY assembly
HEATING AND COOLING CURVE
1600
470W
X-ray tube assembly heat content (kHU)
1400 600W
1200
300W
1000
800
600
Cooling
冷却
400
200
0
0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150
Time (min)
3
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
RADIOGRAPHIC RATING CHARTS
4
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
EMISSION CHARTS
The charts show the emission characteristic under three phase full wave operation.
In the case of single phase full wave operation, mA values are decreased by approximately 30% for same filament current.
0 2 0
3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5 5.2 5.4
Filament Current (A) Filament Current (A)
5
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
HIGH SPEED SERIALOGRAPHIC RATING CHARTS
N.B. The load per exposure must be less than 70% or 80% of the maximum rating
given by the radiographic rating charts.
DIMENSIONS
245 (9.6) Max. 20 (0.8)
145.2 (5.7)
φ46 (1.8)
φ107 (4.2)
Focal Track
φ84 (3.3)
φ136 (5.4)
6
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
DIMENSIONS (cont.)
92 (3.6)
(2.0)
52
focal
100 (3.9)
spot
57 ± 1
(2.2)
464 (18.3) Reference axis
240 (9.4)
161 (6.3) 140 (5.5)
4-M6 12 deep U63.5
P.C.D. 92 20 Mount
φ184 (7.2)
φ158 (6.2)
15 (0.6)
148 (5.8)
UNIT : mm (inch)
UNIT : mm (inch)
7
CIRCLEX 0.3/0.8P324&0.6/1.2P324DK-85
Remarks
* Every value in this Product Date Sheet is a standard value, and it may vary a little from the actual at each site.
* The appearances and specifications are subject to change for reasons of improvement without notice.
* Certain configurations may not be available pending regulatory clearance. Contact your Shimadzu
representative for information on specific configurations.