Service Representatives,
The purpose of this email is to inform you of an addition to the WEAR Checklist (revision or
rev h). Specifically, a checkbox has been added regarding instructions for patients in response
to the “Call for service – Message Code 102”. ZOLL has determined that in certain cases, a
“Call for service – Message Code 102” could mean that the patient’s LifeVest may not be able to
deliver a lifesaving treatment shock if needed. It is very rare for a Message Code 102 to be
associated with a device issue that would prevent the device from being able to deliver a
treatment shock, and a Message Code 102 can be displayed for reasons that do not impact the
devices ability to deliver a treatment shock.
In response to this rare, but potentially serious situation, ZOLL has decided to modify the
instruction for patients to call ZOLL immediately if a Message Code 102 is displayed on their
device.
Patients are to be instructed to call ZOLL immediately at 1‐800‐543‐3267 for a replacement
LifeVest if a “Call for service ‐ Message Code 102” appears on the LifeVest screen during the
start‐up sequence. A replacement device will be provided within 24 hours
The patient should continue to use the LifeVest as prescribed by their physician until a
replacement device is received.
Key Facts:
Patients currently using the LifeVest should continue to do so as originally instructed
by their physician.
It is very rare for a Message Code 102 to be associated with a device issue that would
prevent the device from being able to deliver a treatment shock.
No physician action or communication is required.