Вы находитесь на странице: 1из 32

NEW High-Accuracy Digital Contact Sensor

GT2 Series

GT2
Air Push Type

Air push models


for stable operation
with easy mounting.

[World’s First!]

Revolutionary technology enables high-accuracy measurement


The new GT2 Series provides even
higher “error free” accuracy
Adopting the world's first, ground-breaking scale-shot system, the new GT2 Series provides high-accuracy, without speed errors.
KEYENCE's long history of innovation and cutting edge technologies come together in the new GT2 Series.

Formed through a prism


LED parallel light
illumination
Free from deformation
Detachable Cable
Quartz glass scale

Approx. 6 µs shutter speed


CMOS sensor

2
Advantages of the conventional
scale method (pulse counting)
High-accuracy throughout the entire measurement range
Good temperature characteristics

Advantages of the conventional


transformer method
No tracking errors
Absolute position detection

KEYENCE introduces new technology World’s first

Absolute
Transmitter value scale

CMOS sensor

The world’s first, high-speed scale shot* system using a CMOS sensor that scans the
absolute value glass scale incorporating different patterns depending on the position

* Shutter speed: Approx. 6 µs, Sampling interval: 1 ms

Highest accuracy in its class


Absolute system

Resolution Accuracy

0.1 μm 0.004 Mil 1 μm 0.04 Mil

3
Tough and Rugged
[ Maintenance cost reduction ]

Long life
Detecting durability: 20 million times
Linear ball bearings are used for the
spindle to ensure a long service life.
* GT2-H12 (L/K/KL)
Strong
rubber

Linear ball
bearing used

Water resistant cable


IP67 enclosure rating
Even the connector meets the IP67
enclosure rating based on the IEC/JIS
standards
* Except for the low stress type

Connector
type

Strength against bending


Flexible free-cut robot cable
The sensor head cable uses a robot cable
The cable can be cut
that withstands continuous bending. 90° anywhere.
Load (W): 250 g
Bend radius: 50 mm 1.97"
Bending rate: 30 times/minute
(1 time includes left to right to
original position)
Free cut: can be cut to length R

Free-cut

Withstands bending
10 million times without W
breaking (Typical usage)

4
Easy Sensor Head Installation and Setup
[ Reduction in the number of installation/setup steps ]

Easy installation with a special bracket


Conventional GT2
method
Mounting made easy
simply by drilling a
10 mm 0.39" hole.

A slotted jig must be


prepared by the user.

Large indicators are convenient when setting up several units.

Flashing
lights
The user can
easily identify an
operating sensor
head.

Simple Tolerance Setting


[ Fewer steps for setup ]

First, Enter the tolerance value based on


set the reference. the master target to complete setup.

Example: Enter 0.010.0


A 0.01 mm
0.0004"
tolerance range
is automatically
set.

Press the PRESET


button to set to 0.

5
Easy Amplifier Installation and Setup
[ Fewer steps for installation/setup ]

Select the best display unit according to the application


Compact, panel Large display
DIN-rail mount Pulse output
mount panel mount

GT2-71N/71P GT2-75N/75P GT2-100N/100P GT2-71D


For mounting in a nearby For outputting position
Easy mounting anywhere Easy-to-see large display
enclosure data as pulses

Note: For the DIN-rail mounting type, a connector type and an analog output type are also available.

Features of the large display type

A single unit can connect


Easy operation with large display and large buttons several sensor heads
(11 heads max.).
Double the size of Power supply
a standard display Standard I/O (20-pin MIL
terminal block connector, male)
[Character height:
14 mm 0.55"]
Large display

Easy preset operation W: 159.5 mm


Large button 6.28"
H: 79.8 mm
Expansion board I/O
(30-pin MIL
Communication unit
connection port
3.14" connector, male)

The selected Easy selection with the Installing the expansion board
ID number dedicated button
lights in red. Sensor head
selection button
Easy viewing of each individual head’s OK/NG
and connection status. (OK/NG status is indicated
by red or green, connection status is indicated by
a blinking LED.) The expansion board allows for connection
of up to 11 sensor heads.

Features of the pulse output (line driver output) type

Outputs the spindle position data Pulse frequency selection


[10/50/100/500 kHz]

The maximum pulse frequency


can be selected according to
the input device. The absolute
detection ensures accurate
position data output to an
external device without tracking
Increment/decrement direction
errors even when a slow pulse
Pulse output
amplifier unit
Resolution frequency is selected.
GT2-71D Pulse width

All position data can be output together.


The absolute detection allows the output of all pulses from the origin to the measurement point
without origin return at power-up. Moreover, the operation is free from tracking errors.

6
Total cost reduction
[ Simplified configuration eliminates the need for analog I/O cards and tedious PLC programming ]

Analog I/O card not necessary


No additional equipment required
Forget about the tedious PLC programming that existing LVDT’s required, the GT2 Series does everything automatically. In
addition, the user can save money by eliminating the analog I/O card from the budget.

Conventional Method
Difficult and time consuming way:

Done

Display Analog card PLC programming

GT2 Solution
Easy way:

Done

Easy way:

Done

Push-button setup
HH, HI, GO, LO, LL discrete outputs
Featuring 5 digital comparator outputs (HH, HI, GO, LO and LL), the GT2 Series can be set up in minutes, not hours.
Push-button calibration will leave the user wondering why they hadn’t switched to the GT2 Series sooner.

HH
HI
GO
LO
LL

Quickly select from up to 4 sets of HH/HI/LO/LL limits


Simple product changeover
Using channel select inputs on the GT2 Series, users can easily toggle between up to 4 different sets of HH/HI/LO/LL limits,
skipping the extra PLC programming.

7
Detection of Jam Conditions
[ Current condition check ]

Sensor defect Equipment status check displayed


detection by a bar indicator

Maintenance Maintenance
needed Normal needed

HI Limit Safe LO Limit


detection
Fully Jammed
extended

The sensor outputs in the event of a jam when In actual operation, it is difficult to read a displayed value when it changes
the sensor head is not fully extended. quickly. With the bar indicator the equipment status can be easily read
without reading a displayed value.

Calculations made simple


[ Fewer steps for setup ]
When several amplifiers are connected, the GT2 Series can do automatic calculation simply by selecting a calculation mode.

Thickness [Shaft diameter measurement]


Positioning the target between two sensor heads will display the outer diameter.

Main unit Expansion unit Calculation result

Degree of [Flatness check]


flatness
By subtracting the minimum value from the maximum value the flatness will be displayed.
Main Expansion Expansion Expansion
unit unit unit unit
Maximum Minimum
value value Calculation result

Reference [Determining pin height]


difference
The height difference of a reference value can be displayed by subtracting the value of the main
unit (reference) from the value of the Expansion unit.

Expansion unit Main unit Calculation result

Versatile detection modes support all applications


Single head Multiple heads/when additional amplifier units are installed (Application modes)
Degree of Reference
STD P-H B-H P-P Maximum Minimum flatness Average difference Twist Warpage Thickness

Standard Peak hold Bottom hold Peak to peak

* For details regarding the calculation modes, refer to P.30.

8
Air push type extends the spindle and measures.
[ Easy to set up and saves space. ]

The air push type is part of the high-accuracy digital sensor GT2 Series.
A mechanism to move the sensor head is not necessary since it is possible to measure with a fixed sensor head,
saving space and reducing labor.

Jigs and mechanisms to


move the sensor are no
longer necessary.
Standard type

Linear guide

Dividing jig

Stopper

Air push type

Measuring range

Measuring range 12 mm
Measuring range
12 mm 0.47"

Measuring range 32 mm
1.26"
0.47"

GT2-A12
GT2-A12K
High-accuracy type
50 mm
1.97"
GT2-A32 General purpose type
Accuracy: 1 µm 0.04 Mil (p-p) All jigs unnecessary.

Accuracy: 3 µm 0.12 Mil (p-p) Accuracy: 2 µm 0.08 Mil (p-p)


A complex jig is not necessary since there is no sensor
GT2-A50 head movement. In addition, errors in accuracy due to
Accuracy: 3.5 µm 0.14 Mil (p-p) jigs have been eliminated.

Installation is quick and easy. Point 1 Point 2


Spindle compressed in its usual state Measuring force is unaffected
A dividing jig is by the amount of air pressure.
Air is supplied.
unnecessary.
GT2 Conventional
Dedicated ø4 mm spindle is
installation ø0.16" usually
metal fittings. outside
diameter stretched
Only needs air tube.
a ø10 mm
ø0.39" damage
hole. *1

*1 For the 32 mm 1.26" and 50 mm 1.97" type, open a ø14 mm ø0.55" hole. With the GT2, the spindle moves to the origin when it is
not being supplied with air. Even if there is a problem or a
condition arises where air is not being supplied, it is safe. Due to the pressure free construction
where air pressure is not directly applied to
measurement parts, it is also safe to use
with easily damaged workpieces.

9
Versatile detection modes support
all applications

Single head Multiple heads

Maximum Minimum Degree of Reference


STD P-H B-H P-P flatness Average difference Twist Warpage Thickness

Standard Peak hold Bottom hold Peak to peak

Measurement Packing insertion


of defects in a check
press fit
Confirms that
targets are
properly seated.

Board warpage Shaft diameter


detection measurement
Confirms if
the workpiece
warpage meets
the acceptable
limits.

Detecting twist Flatness


measurement of an
Judges if the engine block
workpiece
twisting falls
within acceptable
limits.

10
Thickness Warp
measurement of measurement
a liquid crystal of a mechanical
substrate chassis

Parallel Volume
measurement of a measurement of a
wafer cam lift

Thermal Measurement of
expansion a transmission
measurement of a mount
machine tool

Thickness Detection of crank


measurement of a shaft runout
steel plate

11
How to select a GT2 Series sensor
Follow the steps below to select the GT2 Series sensor best suited to your operating environment
from the wide range of units and accessories.

step step step step step step step step


1 2 3 4 5 6 7 8
Sensor Sensor Mounting Display unit Display unit Contact Communication Communication
head head cable bracket (Amplifier options unit unit options
unit)

GT2 Series LINE-UP


step 1 Sensor head step 2 Sensor head step 5 Display unit
cable options
Standard Type Air push Type NEW

step 3 Mounting step 6 Contact


bracket

12 mm 0.47" range 12 mm 0.47" range


12 mm 0.47" range High-accuracy, 12 mm 0.47" range High-accuracy,
High-accuracy low stress sensor High-accuracy low stress sensor
sensor head head sensor head head
GT2-H12K GT2-H12KL GT2-A12K GT2-A12KL

step 4 Display unit step 7 Communication


unit

DIN-rail mount type


Amplifier unit BCD Output unit
GT2-71N / 71P DL-RB1A

12 mm 0.47" range 12 mm 0.47" range 12 mm 0.47" range 12 mm 0.47" range


Panel mount type RS-232C
General-purpose General-purpose, General-purpose General-purpose,
Compact amplifier Communication
sensor head low stress sensor sensor head low stress sensor
unit unit
GT2-H12 head GT2-A12 head
GT2-75N / 75P DL-RS1A
GT2-H12L GT2-A12L

Panel mount type step 8 Communication


Large display unit options
amplifier unit
GT2-100N / 100P

DIN-rail mount type


Pulse output
amplifier unit
32 mm 1.26" range 32 mm 1.26" range 32 mm 1.26" range GT2-71D
General-purpose General-purpose, General-purpose
sensor head low stress sensor sensor head
GT2-H32 head GT2-A32
GT2-H32L

50 mm 1.97" range 50 mm 1.97" range


General-purpose General-purpose
sensor head sensor head
GT2-H50 GT2-A50

12
How to select a GT2 Series
step

step First, select a sensor head. Required

1 The sensor head is determined by the measuring range, force, and accuracy.

Standard Type
Commonly used type

12 mm 0.47" range
For 20 µm 0.79 Mil or less When particularly high-
tolerance differentiation accuracy is not required

High-accuracy Type General-purpose Type

Resolution: 0.1 µm 0.004 Mil Resolution: 0.5 µm 0.02 Mil


Accuracy: 1 µm 0.04 Mil (P-P) Accuracy: 2 µm 0.08 Mil (P-P)

High-accuracy type sensor head General-purpose type sensor head


GT2-H12K GT2-H12
Low stress Low stress
High-accuracy / Low stress type type
General-purpose / Low stress type type
sensor head sensor head
GT2-H12KL GT2-H12L

For a wider measuring range For a larger target

32 mm 1.26" 50 mm 1.97"
range range
Resolution: 0.5 µm 0.02 Mil Resolution: 0.5 µm 0.02 Mil
Accuracy: 3 µm 0.12 Mil (P-P) Accuracy: 3.5 µm 0.14 Mil (P-P)

General-purpose type sensor head General-purpose type sensor head


GT2-H32 GT2-H50
General-purpose / Low stress type The low stress type is not available
sensor head Low stress for the 50 mm 1.97" range sensor
GT2-H32L type head.

What is the low stress type?


The low stress type The low stress type uses a low repulsion force spring to reduce
does not require a dust the pressure on the target. When using a target that can be easily
boot. deformed or damaged, select this type.

Measuring force*
Measuring Type of accuracy Measuring force Resolution Accuracy Downward Side Upward Model
range
mounting mounting mounting
Standard 1.0 N 0.9 N 0.8 N GT2-H12K
High-accuracy 0.1 µm 0.004 Mil 1 µm 0.04 Mil
Low stress 0.4 N 0.3 N 0.2 N GT2-H12KL
12 mm 0.47"
Standard 1.0 N 0.9 N 0.8 N GT2-H12
General-purpose 0.5 µm 0.02 Mil 2 µm 0.08 Mil
Low stress 0.4 N 0.3 N 0.2 N GT2-H12L
Standard 2.1 N 1.8 N 1.5 N GT2-H32
32 mm 1.26" 3 µm 0.12 Mil
General-purpose Low stress 0.5 µm 0.02 Mil 1.2 N 0.9 N 0.6 N GT2-H32L
50 mm 1.97" Standard 3.5 µm 0.14 Mil 3.2 N 2.8 N 2.4 N GT2-H50
* Value at center of measuring range.

Dust boot (optional accessory for replacement)


Appearance Applicable sensor head Model
GT2-H12
OP-84332
GT2-H12K
GT2-H32 OP-84459
GT2-H50 OP-84460

13
How to select a GT2 Series
step

Air Push Type NEW

Commonly used type

12 mm 0.47" range
For 20 µm 0.79 Mil or less When particularly high-
tolerance differentiation accuracy is not required

High-accuracy Type General-purpose Type

Resolution: 0.1 µm 0.004 Mil Resolution: 0.5 µm 0.02 Mil


Accuracy: 1 µm 0.04 Mil (P-P) Accuracy: 2 µm 0.08 Mil (P-P)

High-accuracy type sensor head General-purpose type sensor head


GT2-A12K GT2-A12
Low stress Low stress
High-accuracy / Low stress type type
General-purpose / Low stress type type
sensor head sensor head
GT2-A12KL GT2-A12L

For a wider measuring range For a larger target

32 mm 1.26" 50 mm 1.97"
range range
Resolution: 0.5 µm 0.02 Mil Resolution: 0.5 µm 0.02 Mil
Accuracy: 3 µm 0.12 Mil (P-P) Accuracy: 3.5 µm 0.14 Mil (P-P)

General-purpose type sensor head General-purpose type sensor head


GT2-A32 GT2-A50
The low stress type is not available The low stress type is not available
for the 32 mm 1.26" range sensor for the 50 mm 1.97" range sensor
head. head.

Measuring force*
Measuring Type of accuracy Measuring force Resolution Accuracy Downward Side Upward Model
range
mounting mounting mounting
Standard 1.2 N 1.1 N 1.0 N GT2-A12K
High-accuracy 0.1 µm 0.004 Mil 1 µm 0.04 Mil
Low stress 0.4 N 0.3 N 0.2 N GT2-A12KL
12 mm 0.47"
Standard 1.2 N 1.1 N 1.0 N GT2-A12
General-purpose 0.5 µm 0.02 Mil 2 µm 0.08 Mil
Low stress 0.4 N 0.3 N 0.2 N GT2-A12L
32 mm 1.26" Standard 3 µm 0.12 Mil 2.1 N 1.8 N 1.5 N GT2-A32
General-purpose 0.5 µm 0.02 Mil
50 mm 1.97" Standard 3.5 µm 0.14 Mil 3.2 N 2.8 N 2.4 N GT2-A50
* Value at center of measuring range.

Adjustable spindle speed


The speed controller (optional)
allows the spindle movement
speed to be adjusted.

Speed controller
OP-82133

14
step

step Select a sensor head cable Required

2 Select the sensor head cable according to installation conditions and the distance between the sensor head and the display unit (amplifier).
The L-shaped type can be used for the 12 mm 0.47" range sensor head (GT2-H12) only.

Cable connection

Commonly used type

Straight Type
Cable connection
For space-saving installation

L-shaped Type This connector does not rotate.


It is fixed in the direction as shown
in the photo (left).

Type Appearance Cable length Model Weight This connector is required if the cable is cut.
2 m 6.6' GT2-CH2M Approx. 80 g
5 m 16.4' GT2-CH5M Approx. 190 g The cable can be
Straight cut anywhere.
10 m 32.8' GT2-CH10M Approx. 360 g
20 m 65.6' GT2-CH20M Approx. 680 g
L-shaped 2 m 6.6' GT2-CHL2M Approx. 80 g
Applicable to 5 m 16.4' GT2-CHL5M Approx. 190 g
12 mm 0.47" 10 m 32.8' GT2-CHL10M Approx. 360 g
range type only 20 m 65.6' GT2-CHL20M Approx. 680 g Connector used to connect to a display
unit (2 pcs.) OP-84338
The cable can be freely cut. If the cable is cut, a connector (OP-84338) is required to connect to the display
unit (amplifier).

step Select mounting brackets as required.


3 Select a mounting bracket depending on the sensor head mounting method.

For GT2 12 mm 0.47" range models


Commonly used type
Mounting Bracket A
OP-76874

Vibration resistant, Reinforced


holding force
Mounting Bracket C After drilling a ø10 mm ø0.39" hole, insert the
OP-84396 sensor head through the hole, and fasten it.

For side mounting

Mounting Bracket B
Horizontally
mounts the
OP-76875 sensor head.

For GT2 32 mm 1.26" and 50 mm 1.97" range models


Vibration resistant, Reinforced After drilling a ø14 mm ø0.55" hole, insert the
holding force sensor head through the hole and fasten it.
* The mounting method is the same as mounting brackets A
and C.
Mounting Bracket D OP-84327 * When using the GT2-H32L with its contact probe directing
upward, use the mounting holes in the sensor housing.

15
How to select a GT2 Series
step

step Next, select a display unit (amplifier). Required

4 After selecting a display unit depending on the mounting method, select optional accessories as required.

Easy installation in any place


Connecting additional expansion units

DIN-rail Mount Type Use the connector on the side of the unit.

Main unit
Expansion
unit

Commonly used type Easy installation/replacement


Loose wire type Connector type
Connector

To connect an additional unit, purchase the optional


end unit (OP-26751).

For each main unit, up to 14 expansion units


can be connected.
When several units are connected, a power
supply cable is not required for the expansion
GT2-71N/71P GT2-71CN/71CP units. The GT2 Series can be connected
with the GT Series. However, the maximum
number of connectable units is 10 units
(including main unit).
Unit expansion is not possible for the
Model GT2-71D.
Type
NPN output PNP output Line driver output
Main unit GT2-71N GT2-71P
Standard –
Loose wire Expansion unit GT2-72N GT2-72P
Pulse output – – GT2-71D
Standard GT2-71CN GT2-71CP The analog output type amplifier unit provides
Main unit
Connector Analog GT2-71MCN GT2-71MCP – three discrete outputs.
Standard Expansion unit GT2-72CN GT2-72CP
When the connector type amplifier unit is used, an optional socket cable should be purchased
separately (See the next page.).

Installation on easy-to-see places Connecting additional main and expansion


units [Compact type only]
Connect using the cable supplied with the expansion unit.
Panel Mount Type Expansion
To connect additional cable
units, place the main
Commonly used type Easy-to-see display for easier operation unit in the top position
and connect the
Compact type Large display type expansion units below
(vertically).

* For horizontal installation, the optional expansion cable


(OP-35361, 300 mm 11.81") is required. For details, refer to
the description in the option list.

For the compact type amplifier unit, up to 14


expansion units can be connected to one
main unit without providing power supply
cables individually. The expansion connection
can include the GT Series, however, the
GT2-75N/75P GT2-100N/100P maximum number of connectable units is
W: 66 mm 2.60" H: 23.9 mm 0.94" W: 159.5 mm 6.28" H: 79.8 mm 3.14"
limited to 10 including the main unit.

Model
Type A single unit of the large display type can connect two sensor heads.
NPN output PNP output
Main unit GT2-75N GT2-75P When the optional expansion boards are used, up to 11 sensor
Compact heads can be connected. For the details of the expansion board, see
Expansion unit GT2-76N GT2-76P
Large display* GT2-100N GT2-100P the next page.
* When these amplifier units are not mounted to a panel, an optional
mounting bracket should be purchased separately (See the next page.).

16
step

step Select the optional accessories for the display unit as required.
5 Select the optional accessories according to the display unit selected in Step 4.

Optional accessories for the GT2-71/72


(DIN-rail mount)
Type Description Model

For installation without a DIN-rail


Mounting Bracket for DIN-rail The amplifier can be mounted without using a
OP-76877
mount Type Display unit DIN-rail.
(optional accessory for main unit) Amplifier mounting bracket
End
To connect an additional expansion unit, use End unit unit
Required for connecting additional units
the end units to secure the display units on
End unit both ends. When connecting additional units,
OP-26751
(Optional) be sure to use the end units. (2 pcs.)

Required when the connector type is used GT2-CA2M


Used to connect the connector type display unit. (2 m 6.6')
Socket cable
Since the amplifier unit does not include the socket cable, it must be
2 m 6.6'/10 m 32.8' purchased separately. GT2-CA10M
(Optional) (10 m 32.8')

Optional accessories for the GT2-75/76


(panel mount, compact)
Type Description Model

Standard equipment
The panel mounting bracket is included in the panel mount type amplifier.
Panel mounting bracket If the supplied bracket is lost or damaged, order a new bracket.
OP-84394
(Included in panel mount type amplifier)

If the standard cable length is not enough Extension cable used for panel mount type amplifier.
Use this cable if the standard 50 mm 1.97" cable is not long enough.
Expansion cable:
Up to 4 extension cables can be used at once. However, if an extension OP-35361
300 mm 11.81" cable is used, the number of expansion units that can be connected to a
(Optional) single main unit is limited to 4 units.

Optional accessories for the GT2-100


(panel mount, large display)
Type Description Model

Installing this expansion board into the GT2- No. of sensor No. of expansion GT2-E3N
To connect three or more sensor heads 100 allows you to connect more than two heads boards (NPN)
Expansion board sensor heads. Since one expansion board 1 or 2 None
3 to 5 1
For GT2-100N (Optional) allows connection of one to three sensor
6 to 8 2
For GT2-100P (Optional) heads, the maximum of 11 sensor heads can GT2-E3P
9 to 11 3
be connected in total. (PNP)

To install the GT2-100 on a workbench, Secure the


etc. When the amplifier unit is not mounted
to a panel, use this mounting bracket to amplifier unit OP-84331
Mounting bracket with M4 screws.
secure it.
(Optional)

OP-22185
I/O connector (20-pin)
A set of connector and contacts (for AWG 24 to 22).
20-pin MIL connector (for the amplifier unit, optional)
The cable should be prepared by the user.
30-pin MIL connector (for the expansion board, optional) OP-84456
(30-pin)

Contact for maintenance work (Optional) For AWG 24 to 22, Package of 200 OP-22186

Small diameter contact (Optional) For AWG 28 to 26, Package of 200 OP-30594

Special crimping tool (Optional) Use this tool to crimp the contact. OP-21734

17
How to select a GT2 Series
step

step Select a contact as needed.


6 The standard type and carbide type contacts are included in the GT2-H12 (L) / H32 (L) / H50 / A12 (L) / A32 / A50 and the GT2-H12K (L) /
A12K (L) respectively, as standard equipment. Other types of contacts are optionally available. Select optional contacts as required.

For standard measurement Made from a carbide alloy – For high-accuracy measurement For targets with curved surfaces

Standard Super-tough Flat Plate (Optional)


OP-77678 OP-77682 OP-77679

Standard equipment of general- Standard equipment of Ignore the contours of


purpose type sensor head high-accuracy type a target with a curved
GT2-H12 (L) / H32 (L) / H50 / sensor head GT2-H12K (L) / A12K (L) surface.
A12 (L) / A32 / A50

For moving targets To prevent damage to target Electrically insulated from the target

Roller (Optional) Fluorocarbon Resin Ceramic (Optional)


OP-77680 OP-81970
(Optional)
The roller smoothly OP-80228 Ceramic is used for the
detects moving ball of the contact tip.
targets. Made from material that is unlikely to
damage the target’s surface.

For measurement in narrow spaces For measurement at several points with small targets For 12.2 mm 0.48" spindle
extension

Needle (Optional) Offset (Optional) Spacer


OP-77684
(Optional)
OP-77681 OP-77683
When lifting a spindle by hand
The needle contact is Detecting points can be
made of super-tough added in narrow range. Lift lever (Optional)
tungsten alloy. OP-84397

step Select a communication unit as needed.


7 A communication unit is required to extract numerical data from the GT2 Series to an external device. With a single unit, data from up to 15
GT2 Series units can be extracted.

To extract BCD data

BCD Output
Reading
Unit DATA

Detection data can be read from a specified sensor Programmable controller


amplifier via an external device such as a PLC.
BCD output unit
DL-RB1A

For RS-232C communication

RS-232C Reading

Communication Programmable controller


Unit DATA
Writing

Sensor amplifier detection data, settings, and


calculation results from several GT units can be Reading
read and amplifier settings can be written via an RS-232C
communication unit
external device such as a PLC or PC. DL-RS1A PC

18
step

step Select the optional accessories for the communication unit as necessary.
8 Select the optional accessories according to the type of the communication unit selected in Step 7.

Optional accessories common to the DL-RB1A


and DL-RS1A
Type Description Model

To install the unit without using


a DIN-rail The communication unit can be mounted without using a DIN-rail. OP-60412
Mounting bracket

Although the DL Series is designed


DL
To connect the unit to the panel mount for DIN-rail mounting only, the Main unit
type display unit optional expansion cable Expansion unit
Expansion cable: (OP-35361, 300 mm 11.81") Expansion unit OP-35361
300 mm 11.81" enables communication with the
(Optional) panel mount type amplifier unit Expansion cable
(compact/large display). (300 mm 11.81")
OP-35361

Optional accessories for the DL-RB1A


Type Description Model

I/O connector
A set of connector and contacts (for AWG 24 to 22).
34-pin vertical MIL connector (Optional) OP-23139
The cable should be prepared by the user.
[40 contacts for AWG 24 to 22 included]

I/O connector
A set of connector and contacts (for AWG 24 to 22).
34-pin slanting MIL connector (Optional) OP-42224
The cable should be prepared by the user.
[40 contacts for AWG 24 to 22 and a cable tie included]

Special crimping tool Use this tool to crimp the contact. OP-21734
(Optional)

Contact for maintenance work For AWG 24 to 22, Package of 200 OP-22186
(Optional)

Small diameter contact For AWG 28 to 26, Package of 200 OP-30594


(Optional)

Optional accessories for the DL-RS1A


Type Description Model

D-sub 9-pin connector-to-loose wires cable This cable can be used to connect the DL-RS1A to a personal computer or
OP-81283
(5 m 16.4', optional) other serial device.

19
Specifications

❙ Sensor heads for the GT2 Standard Type


Model GT2-H12K GT2-H12KL GT2-H12 GT2-H12L GT2-H32 GT2-H32L GT2-H50
Standard / Low stress (L) type
Type Standard / Low stress (L) type (Sensor head for 12 mm 0.47")
(Sensor head for 32 mm 1.26"/50 mm 1.97" range)

Appearance

Detection system Quartz glass scale, CMOS image sensor projection system, Absolute type (without tracking error)
Measuring range 12 mm 0.47" 32 mm 1.26" 50 mm 1.97"
Resolution 0.1 µm 0.004 Mil 0.5 µm 0.02 Mil
Accuracy (20°C 68°F) 1 µm 0.04 Mil (p-p) 2 µm 0.08 Mil (p-p) 3 µm 0.12 Mil (p-p) 3.5 µm 0.14 Mil (p-p)
Downward mounting 1.0 N 0.4 N 1.0 N 0.4 N 2.1 N 1.2 N 3.2 N
Measuring
Side mounting 0.9 N 0.3 N 0.9 N 0.3 N 1.8 N 0.9 N 2.8 N
force *1
Upward mounting 0.8 N 0.2 N 0.8 N 0.2 N 1.5 N 0.6 N 2.4 N
Mechanical response 10 Hz 4 Hz 10 Hz 4 Hz 6 Hz 5 Hz 7 Hz
Probe Carbide ball ø3 Steel ball ø3
Operation indicator 2-color LED (red, green)
Enclosure rating IP67 – IP67 – IP67 – IP67
Environmental Ambient temperature -10 to +55°C 14 to 131°F
resistance Relative humidity 35 to 85% RH (No condensation)
Vibration 10 to 55 Hz, 1.5 mm 0.06" double amplitude, 2 hours in each of X, Y, and Z directions
Sensor head cable Optional (M8 connector)
Main body Main body cast: Zinc die-casting, Indicator: Polyarylate, Dust boot: NBR *2
Materials TYPE304 Stainless steel, super-tough
Contact TYPE304, 440C Stainless steel
tungsten alloy
Weight (excluding cable) Approx. 95 g Approx. 270 g Approx. 320 g
Accessories Refer to the instruction manual.

*1 Value at center of measuring range.


*2 A dust boot is not provided with the GT2-H12KL, the GT2-H12L or the GT2-H32L.

❙ Sensor heads for the GT2 Air Push Type


Model NEW GT2-A12K NEW GT2-A12KL NEW GT2-A12 NEW GT2-A12L NEW GT2-A32 NEW GT2-A50
Standard type (Sensor head
Type Standard / Low stress (L) type (Sensor head for 12 mm 0.47")
for 32 mm 1.26"/50 mm 1.97" range)

Appearance

Detection system Quartz glass scale, CMOS image sensor projection system, Absolute type (without tracking error)
Measuring range 12 mm 0.47" 32 mm 1.26" 50 mm 1.97"
Resolution 0.1 µm 0.004 Mil 0.5 µm 0.02 Mil
Accuracy (20°C 68°F) *1 1 µm 0.04 Mil (p-p) 2 µm 0.08 Mil (p-p) 3 µm 0.12 Mil (p-p) 3.5 µm 0.14 Mil (p-p)
Downward mounting 1.2 N 0.4 N 1.2 N 0.4 N 2.1 N 3.2 N
Measuring
Side mounting 1.1 N 0.3 N 1.1 N 0.3 N 1.8 N 2.8 N
force *2
Upward mounting 1.0 N 0.2 N 1.0 N 0.2 N 1.5 N 2.4 N
Operation indicator 2-color LED (red, green)
Applied pressure range 0.25 to 0.5 MPa
Pressure resistance 1.0 MPa
Enclosure rating IP67 *3 – IP67 *3 – IP67 *3 IP67 *3
Environmental Ambient temperature 0 to 55°C 32 to 131°F
resistance Relative humidity 35 to 85% RH (No condensation)
Vibration *4 10 to 55 Hz, 1.5 mm 0.06" double amplitude, 2 hours in each of X, Y, and Z directions
Sensor head cable Optional (M8 connector)
Main body case: Zinc die-casting; Cylinder part: Aluminium alloy;
Main body
Air joint resin part: Polyacetal; Air joint metal part: Brass nickel plating; Indicator: Polyarylate
Materials Dust boot NBR – NBR – NBR NBR
TYPE304 Stainless steel, super-tough
Contact *5 TYPE304, 440C Stainless steel
tungsten alloy
Weight (excluding cable) Approx. 145 g Approx. 340 g Approx. 405 g
Accessories Refer to the instruction manual.
*1 Value when ambient temperature is 20°C 68°F
*2 Value at center of measuring range.
Please note that the measurement force changes depending on the installation state of the dust boots.
*3 Make sure the air tube is connected to the air exhaust joint and that no foreign materials enter inside from the joint.
*4 In the case where a mounting bracket D is used with GT2-A32 and GT2-A50, the double amplitude becomes 0.75 mm 0.03".
20 *5 The contact is included with the sensor.
❙ DIN-rail mount type amplifier unit (For the pulse output type, see page 19.)
NPN output GT2-71N GT2-72N GT2-71CN GT2-72CN GT2-71MCN
Model
PNP output GT2-71P GT2-72P GT2-71CP GT2-72CP GT2-71MCP
Mounting method DIN-rail mount
Type Standard loose wire type Standard connector type Analog output, connector type
Main unit / Expansion unit *1 Main unit Expansion unit Main unit Expansion unit Main unit

Appearance

Power supply voltage *1 10 to 30 VDC, including 10% ripple (P-P) 20 to 30 VDC, including 10% ripple (P-P)
Power Normal 2200 mW max. (30 V, 73.3 mA max.) 2700 mW max. (30 V, 90 mA max.)
consumption *2 Power saving (Eco) 1700 mW max. (30 V, 56.7 mA max.) 2200 mW max. (30 V, 73.3 mA max.)
Measured value display 6 + 1/2-digit 7-segment LED (red)
Display power
Other displays 2-color, 13-level bar LED display (red, green), indicators (red, green)
Display range -199.999.9 to 199.999.9
Display resolution 0.1 µm 0.004 Mil
Sampling rate 1000 times/sec.
Timing input Input time: 2 ms min.
Preset input
Control input
Reset input Input time: 20 ms min.
Bank input
NPN/PNP open collector, Applicable current: 50 mA,
Control HH / High / Go / Maximum applicable voltage (NPN: 40 V, PNP: 30 V)
output Low / LL Residual voltage at ON: 1 V max. * N.O./N.C switchable
(GT2-71MCN/71MCP: HH and LL are not available.)
4 to 20 mA, maximum load resistance: 350 Ω
Analog output –
Analog output range can be changed as desired.
Response time hsp (3 ms), 5 ms, 10 ms, 100 ms, 500 ms, 1 s
Ambient temperature -10 to +50°C 14 to 122°F (GT2-71MCN/71MCP: -10 to +45°C 14 to 113°F when expansion unit(s) is connected)
Environmental
Relative humidity 35 to 85% RH (No condensation)
resistance
Vibration 10 to 55 Hz, 1.5 mm 0.06" double amplitude, 2 hours in each of X, Y, and Z directions
Materials Main body case: Polycarbonate, Key top: Polyacetal, Front sheet: PET, Cable: PVC
Weight GT2-71N/71P/72N/72P: Approx. 140 g (including power cable), GT2-71CN/71CP/72CN/72CP/71MCN/71MCP: Approx. 70 g
Accessories GT2-71N/71P/71CN/71CP/71MCN/71MCP: Instruction manual, Cover seal, GT2-72N/72P/72CN/72CP: None

❙ Panel mount type amplifier unit


NPN output GT2-75N GT2-76N GT2-100N GT2-E3N
Model
PNP output GT2-75P GT2-76P GT2-100P GT2-E3P
Panel mount/Screw mounting using Installed into GT2-100N/100P
Mounting method Panel mount optional mounting bracket (up to 3 boards)
Main unit / Expansion unit * 1 2 channels provided as standard. Addition of expansion
Main unit Expansion unit Expansion unit (1 to 3 channels)
boards allows expansion to 11 channels max.

Appearance

Power supply voltage *1 10 to 30 VDC, including 10% ripple (P-P) 20 to 30 VDC, including 10% ripple (P-P) Supplied from GT2-100N/100P
Power Normal 2200 mW max. (30 V, 73.3 mA max.) 4500 mW max. (30 V, 150 mA max.) 4200 mW max. (30 V, 140 mA max.)
consumption *2 Power saving (Eco) 1700 mW max. (30 V, 56.7 mA max.) 3600 mW max. (30 V, 120 mA max.) 4000 mW max. (30 V, 133.3 mA max.)
Measured value display 6 + 1/2-digit 7-segment LED (red)
Display power
Other displays 2-color, 13-level bar LED display (red, green), indicators (red, green)

Display range -199.999.9 to 199.999.9
Display resolution 0.1 µm 0.004 Mil
Sampling rate 1000 times/sec.
Timing input Input time: 2 ms min.
Preset input
Control input
Reset input Input time: 20 ms min.
Bank input
NPN/PNP open collector, Applicable current: 50 mA,
Control HH / High / Go / Maximum applicable voltage (NPN: 40 V, PNP: 30 V)
output Low / LL Residual voltage at ON: 1 V max. * N.O./N.C switchable
(Applicable current decreases to 20 mA or lower when 2 or more sensor heads are connected to the GT2-100 Series.)
Response time hsp (3 ms), 5 ms, 10 ms, 100 ms, 500 ms, 1 s
Ambient temperature -10 to +50°C 14 to 122°F
Environmental
Relative humidity 35 to 85% RH (No condensation)
resistance
Vibration 10 to 55 Hz, 1.5 mm 0.06" double amplitude, 2 hours in each of X, Y, and Z directions (GT2-100N/P: 0.15 mm 0.01" double amplitude)
Materials Main body case: Polycarbonate, Key top: Polyacetal, Front sheet: PET, Cable: PVC (GT2-75N/75P/76N/76P only) –
Power supply: Terminal block,
Power supply / I/O cable 12-core connector connection (Connecting cable included) I/O: 30-pin connector (MIL standard)
I/O: 20-pin connector (MIL standard)
GT2-75N/75P/76N/76P: Approx. 140 g (including panel mounting bracket, protective front cover and power cable)
Weight
GT2-100N/100P: Approx. 380 g, GT2-E3N/E3P: Approx. 80 g
GT2-75N/75P: Panel mounting bracket, protective front cover, power cable, instruction manual, cover seal
Accessories GT2-76N/76P: Panel mounting bracket, protective front cover, power cable, expansion cable
GT2-100N/100P: Instruction manual, panel mounting bracket, GT2-E3N/E3P: None

*1 Precautions when the GT2-70 Series is used with an expansion unit (excluding GT2-71D)
Up to 15 amplifier units can be connected; main unit: 1 unit, expansion unit: 14 units. To use the DIN-rail mount type, be sure to mount the amplifier unit to a DIN-rail (using the metal bracket). When
mounting additional units, be sure to use the end units (OP-26751). When mounting additional units, depending on the number of units to be added, confirm the following specification limitations.
<Up to 8 units (including main unit)> <Up to 9 to 15 units (including main unit)>
• The power supply voltage is 20 to 30 VDC. • The power supply voltage is 20 to 30 VDC.
• The maximum applicable current for each output is 20 mA. • The maximum applicable current for each output is 10 mA (including DL-RB1A output current).
• The residual voltage at ON Is 1.5 V or less.
*2 For GT2-100 Series, when the maximum number of sensor heads are connected and all units are set to power saving mode
21
Specifications

❙ DIN-rail mount type pulse output amplifier unit


Model GT2-71D

Appearance

Mounting method DIN-rail mount


Power supply voltage 10 to 30 VDC, including 10% ripple (P-P)
Current consumption 1600 mW (30 V, 53.3 mA max.)
Power supply (green)/Alarm (red) indicator,
Indicator
pulse output indicator (green), input indicator (green)
Pulse resolution Selectable from 0.1 μm 0.004 Mil, 0.5 μm 0.02 Mil, 1 μm 0.04 Mil, and 10 μm 0.39 Mil (Factory setting: 0.5 μm 0.02 Mil)
Minimum phase difference Selectable from 0.5 μs, 2.5 μs, 5 μs, and 25 μs (Factory setting: 2.5 μs)
Control input Origin return Non-voltage input (Contact, non-contact), input time: 20 ms min.
Output signal 90° phase difference, differential square wave (EIA-422 compliant)
Output signal level +5V, øA, øA-, øB, øB-, øZ, øZ-: Line driver output
Ambient temperature -10 to +50°C 14 to 122°F
Environmental Relative humidity 35 to 85% RH (No condensation)
resistance 10 to 55 Hz, 1.5 mm 0.06" double amplitude,
Vibration
2 hours in each of X, Y, and Z directions
Material Main body case: Polycarbonate, Cable: PVC
Weight Approx. 110 g (including power supply cable)
Accessories Instruction manual, SW protection sticker

❙ Communication unit (Common specifications) ❙ Communication unit (DL-RS1A Communication Specifications)


Model DL-RB1A DL-RS1A Model DL-RS1A

Appearance Appearance

20 to 30 VDC, including ripple, Communication method Full duplex


Power supply voltage Ripple (P-P): 10% max. Synchronization method Start-stop
(Supplied via connected sensor amplifier) Transmission code ASCII
Power consumption 27 mA max. 25 mA max. 2400/4800/9600/19200/38400 bps
Baud rate
Number of connectable sensor selectable (Factory-setting: 9600 bps)
Up to 15 units (including main unit)
amplifiers Date bit length 8-bit/7-bit selectable (Factory-setting: 8 bits)
Communication None/Even/Odd selectable
Parity check
indicator lamp (Factory- setting: None)
Alarm indicator lamp
(green x 2), Stop bit length 1 bit
(red),
Indicator Alarm indicator lamp Data reception: automatically recognizes CR
Power indicator lamp
(red), Data delimiter or CR+LF
(green)
Power indicator lamp Data transmission: Fixed to CR+LF
(green)
Ambient temperature -10 to +55°C 14 to 131°F
Relative humidity 35 to 85% RH (No condensation)
Environmental
resistance 10 to 55 Hz,
Vibration 1.5 mm 0.06" double amplitude,
2 hours in each of X, Y, and Z directions
Material Main body case: Polycarbonate
Weight Approx. 46 g Approx. 53 g
Instruction manual, End units (x 2),
Accessories Switch protection seal,
Expansion connector cover

❙ Pin assignment for the DL-RB1A (BCD)


Pin No. Signal name Description Pin No. Signal name Description 1 2
1 IDSEL 1 ID No. selection input 1 18 BCD DIGIT 4 (1) BCD 4th digit 1×103 3 4
2 IDSEL 2 ID No. selection input 2 19 BCD DIGIT 4 (2) BCD 4th digit 2×103 5 6
3 IDSEL 3 ID No. selection input 3 20 BCD DIGIT 4 (4) BCD 4th digit 4×103 7 8
4 IDSEL 4 ID No. selection input 4 21 BCD DIGIT 4 (8) BCD 4th digit 8×103 9 10
5 DRQ Data request input 22 BCD DIGIT 5 (1) BCD 5th digit 1×104 11 12
6 BCD DIGIT 1 (1) BCD 1st digit 1×100 23 BCD DIGIT 5 (2) BCD 5th digit 2×104 13 14
7 BCD DIGIT 1 (2) BCD 1st digit 2×100 24 BCD DIGIT 5 (4) BCD 5th digit 4×104 15 16
8 BCD DIGIT 1 (4) BCD 1st digit 4×100 25 BCD DIGIT 5 (8) BCD 5th digit 8×104 17 18
9 BCD DIGIT 1 (8) BCD 1st digit 8×100 26 BCD DIGIT 6 (1) BCD 6th digit 1×105 19 20
10 BCD DIGIT 2 (1) BCD 2nd digit 1×101 27 BCD DIGIT 6 (2) BCD 6th digit 2×105 21 22
11 BCD DIGIT 2 (2) BCD 2nd digit 2×101 28 BCD DIGIT 6 (4) BCD 6th digit 4×105 23 24
12 BCD DIGIT 2 (4) BCD 2nd digit 4×101 29 BCD DIGIT 6 (8) BCD 6th digit 8×105 25 26
13 BCD DIGIT 2 (8) BCD 2nd digit 8×101 30 BCD SIGN BCD data polarity sign 27 28
14 BCD DIGIT 3 (1) BCD 3rd digit 1×102 31 BCD STB Strobe output 29 30
15 BCD DIGIT 3 (2) BCD 3rd digit 2×102 32 ALARM Alarm output 31 32
16 BCD DIGIT 3 (4) BCD 3rd digit 4×102 33 COM Common 33 34
17 BCD DIGIT 3 (8) BCD 3rd digit 8×102 34 COM Common

22
Input/Output Circuits

I/O circuit (GT2-71N/72N/71CN/72CN/71MCN/75N/76N) External input circuit

Brown *1 Pink: Preset input


10 to 30 VDC *3
Purple: Bank A input
Orange (HH output) *2
Overcurrent protection circuit
Load 5 to 40 VDC Pink/purple: Timing input
Black (HI output) Yellow: Bank B input
Load 5 to 40 VDC
Red: Reset input
Main circuit

Gray (GO output)


Load 5 to 40 VDC

Main circuit
White (LO output)
Load 5 to 40 VDC
Green (LL output) *2 Blue *1
Load 5 to 40 VDC
1
Blue *
0V
(Short-circuit current: 1 mA max.)
Pink, purple, pink/purple, yellow, red *4
Input circuit External input

I/O circuit (GT2-71P/72P/71CP/72CP/71MCP/75P/76P) External input circuit


10 to 30 VDC
Brown *1 3
10 to 30 VDC * Brown *1
Orange (HH output) *2
Overcurrent protection circuit

Load

Main circuit
Black (HI output)
Gray (GO output) Load
Main circuit

White (LO output) Load Pink: Preset input


Purple: Bank A input
Green (LL output) *2 Load
Pink/purple: Timing input
Blue *1 Load
0V Yellow: Bank B input
(Short-circuit current: 2 mA max.)
Pink, purple, pink/purple, yellow, red *4 Red: Reset input
Input circuit External input

*1 Brown and blue are applicable only to main units (GT2-71N/71P/71CN/71CP/71MCN/71MCP/75N/75P). Not to expansion units (GT2-72N/72P/72CN/72CP/76N/76P).
The connector type expansion unit (GT2-72CN/72CP) is not connected to the internal circuit.
*2 The orange and green cables are used as analog output cables for the analog type amplifier unit (GT2-71MCN/71MCP).
For details, refer to the analog output circuit diagram.
*3 20 to 30 VDC when expansion unit is connected or for the analog type amplifier unit (GT2-71MCN/71MCP).
*4 For details on external input, refer to the external input circuit diagram.

Analog output circuit Pulse output amplifier unit (GT2-71D)


GT2-71MCN/71MCP I/O circuit External input circuit Recommended input device

Brown AM26LS32 line receiver or equivalent device


10 to 30 VDC
5V
Orange Black, white,
Analog output 33 Ω orange
(4 to 20 mA) Black: Phase A, White: Phase Pink 390 Ω
Main circuit

Main circuit

B, Orange: Phase Z
Main circuit

Purple, gray,
33 Ω green Purple: Reversed phase A
Gray: Reversed phase B
Green: Reversed phase Z Blue
Green *
0 V (for analog) Blue
0V
Pink * (Short-circuit current: 1 mA max.)
Input circuit Input

* The green and blue cables are common internally. * For details of the external input, refer to the diagram of the external input circuit.

Output circuit of the large display amplifier unit Input circuit of the large display amplifier unit
GT2-100N/GT2-E3N GT2-100N/GT2-E3N
(Pin Nos. 1 to 5, 11 to 15, 21 to 25) (Pin Nos. 6 to 10, 16 to 20, 26 to 30)
+* +5 V
20 to 30 VDC
Overcurrent protection circuit

1 to 5, 11 to 15, 21 to 25
Sensor main circuit

Load 5 to 40 VDC For details of the I/O connector


6 to 10, 16 to 20, 26 to 30
pin assignment and the expansion
Main circuit

board, refer to the instruction


-*
-* manual. The information is also
0V
(Short-circuit current: 1 mA max.)
available at the KEYENCE home
* The +/- terminals are provided in the GT2-100N only. * The - terminal is provided in the GT2-100N only. page.
They are not provided in the GT2-E3N. It is not provided in the GT2-E3N.

Output circuit of the large display amplifier unit Input circuit of the large display amplifier unit
GT2-100P/GT2-E3P GT2-100P/GT2-E3P
(Pin Nos. 1 to 5, 11 to 15, 21 to 25) (Pin Nos. 6 to 10, 16 to 20, 26 to 30)
+* 20 to 30 VDC
20 to 30 VDC
+*
Overcurrent protection circuit
Sensor main circuit

Main circuit

1 to 5, 11 to 15, 21 to 25
6 to 10, 16 to 20, 26 to 30
Load
-*
0V (Short-circuit current: 2 mA max.)

* The +/- terminals are provided in the GT2-100P only. * The + terminal is provided in the GT2-100P only.
They are not provided in the GT2-E3P. It is not provided in the GT2-E3P.

23
Dimensions

Unit: mm inch

Sensor head (Standard) M8 cable connector


Cable connection Cable connection
1.10" 0.63"
GT2-H12K/GT2-H12 28 16 (Straight)
0.42"
10.7 (L-shaped) 0.80"
0.61" ø0.37" ø9.5 20.2
(0.49")
15.5 1.15 ø9.5 3.05
(12.5) 5.9 (10.1) 0.05" ø0.37" 0.12"
ø8 0.23" (0.40")
ø0.31" (6.4)(0.25") 0.05" 9
1.15 0.35"

(58)
(2.28")
95.3 92.3
66.7 73.1 3.75"
2.63" 2.88" 3.63"

8.7 0.34"
15.1
ø11.6 0.59"41.5 45.3
ø0.46" to to
29.0 32.8
1.63"1.78"
to to
1.14"1.29"
0.24" ø4 ø0.16"
6 3.8
ø8 ø0.31" h5 (+0 (22) 0.15" 6.4 (9.6)
-0.006 ) (0.87")
0.25" (0.38")
( +0
-0.0002")

Sensor head (Low stress) M8 cable connector


Cable connection Cable connection
1.10" 0.63"
GT2-H12KL/GT2-H12L 28 16 (Straight) 0.42"
10.7 (L-shaped) 0.80"
0.61" ø0.37" ø9.5 20.2
(0.49")
15.5 1.15 ø9.5 3.05
(12.5) 5.9 (10.1) 0.12"
ø8 0.23" 0.05" ø0.37"
(6.4)(0.25") (0.40")
ø0.31" 0.05" 9
1.15 0.35"

(58)
(2.28")
95.3 92.3
66.7 73.1 3.75"
2.63" 2.88" 3.63"

8.7 0.34"
15.1
ø11.6 0.59"41.5 45.3
ø0.46" to to
29.0 32.8
1.63"1.78"
to to
1.14"1.29"
0.24" ø4 ø0.16"
6 3.8
ø8 ø0.31" h5 (+0 (22) 0.15" 6.4 (9.6)
-0.006 ) (0.87")
0.25" (0.38")
( +0
-0.0002")

1.48" M8 cable connector


37.6
0.78"
Sensor head (Standard) 0.99" 19.7 Cable connection
0.70" 25.1 (12.5) (0.49") ø9.5 20.35
GT2-H32 17.8 (19.8) (0.78") 8.15 (11.55) (Straight) ø0.37" 0.80"
0.32" (0.45")
ø8
ø0.31"
(6.4) (0.25")
3.4
0.13"

(53.3) 97.3
(2.10") 3.83" 119.5
116.4 4.70"
4.58" 0.38" 90.9
9.7 3.58"
0.91" (88.5)
23 (0.56") (3.48")
0.34" 8.7 (14.3) 0.3 0.01"
5.9 0.23" 27.9 1.10" 20 0.79"

17 0.67" 2.4
ø15
ø0.59" 0.09" 43.8 47.6
to to
2-ø4.3 30 77.8 81.6
ø0.17" 1.18" (17.8) 1.72" 1.87"
to to
0.15" (0.70") 3.06" 3.21"
3.8

0.48" ø0.20"
12.2 (25.4) (1.00") ø5
ø12 ø0.47" h6 ( +0
-0.011 ) 12.2
0.48"
(7.8) (0.31")
+0
( -0.0004" )

1.48" M8 cable connector


37.6
0.78"
Sensor head 0.99"
0.70" 25.1 (12.5) (0.49")
19.7
Cable connection ø9.5 20.35
(Low stress) 17.8 (19.8) (0.78") 8.15 (11.55) (Straight) ø0.37" 0.80"
0.32" (0.45")
GT2-H32L ø8
ø0.31"
(6.4) (0.25")
3.4
0.13"

(53.3) 97.3
(2.10") 3.83" 119.5
116.4 4.70"
4.58" 0.38" 90.9
9.7 3.58"
0.91" (88.5)
23 (0.56") (3.48")
0.34" 8.7 (14.3) 0.3 0.01"
5.9 0.23" 27.9 1.10" 20 0.79"

17 0.67" 2.4
ø15
ø0.59" 0.09" 43.8 47.6
to to
2-ø4.3 77.8 81.6
ø0.17" 30
1.18" (17.8) 1.72" 1.87"
to to
0.15" (0.70") 3.06" 3.21"
3.8

0.48" ø0.20"
12.2 (25.4) (1.00") ø5
ø12 ø0.47" h6 ( +0
-0.011)
12.2 (7.8) (0.31")
0.48"
( +0
-0.0004")

24
Unit: mm inch

Sensor head (Standard) 1.48" M8 cable 0.78"


Cable connection
37.6 connector 19.7
GT2-H50 0.99" (Straight) ø0.37" ø9.5 20.35 0.80"
25.1 (12.5) (0.49") 8.15 (11.55)
0.70" 0.32" (0.45")
17.8 (19.8) (0.78")
ø8 ø0.31"
(0.25")
(6.4)
3.4
0.13"

152.4 (53.3)
6.00" (2.10") 115.3
0.38" 4.54" 137.5
9.7 5.41"
108.9
4.29"
0.91"
23 (106.5)
(4.19")
0.34" 8.7 (14.3)
0.23" (0.56")
5.9 45.9 (0.3) (0.01")
1.80" 20 0.79"

17 0.67"
ø15
2-ø4.3 ø0.59" 2.4
ø0.17" 0.09"
30 48.6 52.4
1.18" to to
(17.8) 100.6 104.4
(0.70") 1.91" 2.06"
to to
3.96" 4.11"
0.15"
3.8

12.2 (25.4) ø5
0.48" (1.00") ø0.20"
ø12 ø0.47" h6 (+0 12.2 (7.8)
-0.011) 0.48" (0.30")
( +0
-0.0004")
32.5 One-touch air supply joint
1.28" (for 4 mm 0.16" outside diameter air tube.)
1.00" 25.5
28 16
Sensor head (Standard) 0.42" 10.75
1.10"
ø9.5 ø0.37" (0.40") (10.1)
0.63" Air exhaust joint
(for 2.5 mm 0.10" inside diameter air tube.)
GT2-A12/GT2-A12K 0.61" 15.5 (12.5) (0.49") 0.23" 5.9
0.05" 1.15 ø14 ø0.55"
0.23" 5.95 M8 connector

17.7
(0.22") 0.70"
(5.5) 97.35
0.04" 1.05 3.83" 13.1 0.52"
95.3
(58) 0.22" 12.5
0.06" 1.55 3.75"
(2.28") 66.7 0.88" 22.4 84.85 15 0.59"
0.99" 25.15 3.34" 18 0.71"
2.63" 39.8
0.12" 3.05
1.57"
0.34" 8.7
ø0.46" ø11.6 1.14" 29 32.8 1.29"
to to to to 8.75
0.59" 15.1 1.63"41.5 45.3 1.78" 0.34"

(22) (0.87")
0.24" 6 3.8 0.15" ø0.16" ø4
There are no dust boots on the low
ø8 ø0.31" h5 0.25" 6.4 (9.6) (0.38")
stress type
34.5
GT2-A12L/A12KL 1.36"
1.46"
37.6 37
1.48" One-touch air supply joint
0.49" 25.1 (12.5) (0.49") 30 (for 4 mm 0.16" outside diameter air tube.)
1.18"
Sensor head (Standard) 0.70" 17.8 (19.8) (0.78") 19.7
0.78" (11.55)
GT2-A32 0.80" 20.35 ø9.5 ø0.37" 0.32" 8.15
(0.45") Air exhaust joint ø14 ø0.55"
(for 2.5 mm 0.10" inside
diameter air tube.)

1.17" 29.75 15 0.59"


M8 connector
0.13" 3.4
1.06" 27
155.2
0.49" 12.5 6.11"
(53.3) 142.7
(2.10") (5.5) 5.62" 18 0.71"
119.5
88.5 4.70" (0.22") 28.1 1.11"
0.91" 23 3.48" 32.7 1.29"
(0.56") (14.3) 90.9 75.65
9.7 0.38" 3.58" 2.98" 15 0.59"
0.34" 8.7 18 0.71"
0.09" 0.01" 0.3
0.23" 5.9 27.9 2.4
1.10"
ø0.17"
2 x ø4.3 44.3
ø15 to 5.6
ø0.59" 77.8 48.1 1.89" 0.22"
17 1.74" to to
1.18" 30 to 81.6 3.21"
0.67" 3.06"

(0.70") (17.8) 3.8 0.15" ø0.20" ø5


0.48" 12.2 (25.4) (1.00") 0.48" 12.2 (7.8) (0.31")
ø12 ø0.47" h6 39
1.54"
1.46"
37.6 37
1.48" 30 One-touch air supply joint
Sensor head (Standard) 0.99" 25.1 (12.5) (0.49") 1.18" (for 4 mm 0.16" outside diameter air tube.)
0.70" 17.8 (19.8) (0.78") 19.7
GT2-A50 0.80" 20.35 ø9.5 ø0.37" 0.32" 8.15 0.78" (11.55) (0.45") ø14 ø0.55"

Air exhaust joint


(for 2.5 mm 0.10" inside 15 0.59"
1.17" 29.75
diameter air tube.)
1.06" 27

M8 connector
0.13" 3.4 178.7
7.04"
0.49" 12.5
(53.3) 191.2
(2.10") (5.5) 7.53" 18 0.71"
137.5 (0.22")
106.5 5.41" 28.1 1.11"
4.19" 108.9 32.7 1.29"
0.91" 23 4.29" 15 0.59"
9.7 0.38" 93.65
(0.56") (14.3) 3.69"
0.34" 8.7 45.89 0.01" 0.3 18 0.71"
0.23" 5.9 1.81" 0.09"
2.4
ø0.17"
2 x ø4.3 ø15 49.1 1.93" 5.6
ø0.59" to to 52.9 2.08" 0.22"
17 100.6 3.96" to to
0.67" 104.4 4.11"
1.18" 30

(0.70") (17.8) 3.8 0.15" ø0.20" ø5


0.48" 12.2 (25.4) (1.00") 0.48" 12.2 (7.8) (0.31")
ø12 ø0.47" h6 39
1.54"

25
Dimensions

Unit: mm inch

Sensor head cable (Straight)


GT2-CH2M/CH5M/CH10M/CH20M (Optional)
C
L
27.2 9.4 0.37" A B Sheath color Model L
ø11 ø0.43"
1.07" 1 1 Brown GT2-CH2M 2000 78.74"
A ø9.5 B 2 3 White GT2-CH5M 5000 196.85"
ø0.37" 3 4 Blue GT2-CH10M 10000 393.70"
ø0.19"
7.2 ø4.8, 4 x 0.22 mm2 20.6 0.81" C-C 4 2 Black GT2-CH20M 20000 787.40"
0.28" 10.4 0.41"
1 4 21.4
2 3 C
0.84"
4 2
3 1
A B

Sensor head cable (L-shaped)


GT2-CHL2M/CHL5M/CHL10M/CHL20M (Optional)
C
L
20.6 0.81" A B Sheath color Model L
23 ø0.43" ø11 9.4 0.37"
0.91" 1 1 Brown GT2-CHL2M 2000 78.74"
B 2 3 White GT2-CHL5M 5000 196.85"
0.95" 24.2 ø7 ø0.19" 3 4 Blue GT2-CHL10M 10000 393.70"
ø0.28" ø9 ø4.8, 2 C-C 4 2 Black GT2-CHL20M 20000 787.40"
ø0.35" 4 x 0.22 mm 10.4 0.41"
ø9 21.4
ø0.35" 1 4 0.84" C
A 2 3
4 2
3 1
A B

Sensor head Dust boot


mounting bracket A (Optional)
* The dust boot is factory-attached
OP-76874 0.47" 12 12 0.47" Mounting hole to the sensor head
R6.5
R0.26" (except for the low stress type).
ø8 ø0.31"
G6 (+0.014 Material: TYPE304
+0.005) ( +0.0002")
+0.0006"
ø10 ø0.39"
ø10 ø0.39" Stainless steel, NBR
g6 (-0.005 -0.0002"
-0.014) ( -0.0006") +0.001"
+0.005) ( +0.0002")
G8 (+0.027
C1.5
6 16.3 C0.5 to C1.0
0.24" 0.64" C0.02" to C0.04"
C0.06"
7.8
3.5 0.31"
Material: TYPE304 0.14"
Stainless 5.5 to 11.3
steel 0.22" to 0.44" 81
3.19"
58.2
2.29"
Sensor head 25.3
1.00"
mounting bracket C (Optional)
OP-84396 0.47" 12 12 0.47" Mounting hole ø7.6 ø0.30" ø11.6 ø0.46" ø11.6 ø0.46"
R6.5
R0.26"
ø8 ø0.31"
ø10 ø0.39" G6 (+0.014
+0.005 ) ( +0.0002" )
+0.0006"
g6 (-0.005 -0.0002"
-0.014) ( -0.0006")
+0.1
ø10 ø0.39" ( +0.005 ) For GT2-H12K/H12 For GT2-H32 For GT2-H50
6 C0.5 to C1.0
C1.5 0.24" 16.3 C0.02" to C0.04" ( +0.004"
+0.0002")
(Optional) (Optional) (Optional)
C0.06" 0.64"
7.8 OP-84332 OP-84459 OP-84460
5 0.31"
Material: TYPE304 0.20"
Stainless 5.5 to 11.3
steel 0.22" to 0.44"

Sensor head
mounting bracket B (Optional) 10.5 6.5
0.26"
0.41"
OP-76875 4-C2 C0.08"
R8
(0.01") 32 1.26" R0.31"
(0.2)
12.1 0.48"
ø4.2
16.3 11 0.25" 6.3 ø0.17"
0.64" 0.43"
15
R3.8 0.59" 3.5
R0.15" 0.14"

Material: TYPE304
Stainless steel, Zinc
ø8 ø0.31" G6 (+0.014) ( +0.0006")
+0.005 +0.0002"

Sensor head
mounting bracket D (Optional) 0.63" 0.63"
16 16 R0.33"
OP-84327 R8.5 Mounting hole
ø0.47"
ø12 G6 +0.017 ø14 ø0.55"
g6 (-0.006
) ø0.55" ( ) +0.006
-0.017
-0.0002" ø14 (+0.1"
+0.006")
( ) ( +0.004"
+0.0002")
-0.0007"
6 0.24" 16.3
C1.5 7.8 0.31" 0.64" C0.02" to C0.04"
C0.06" C0.5 to C1.0
5 0.20"

17 0.67"
Material: TYPE303
Stainless 5.5 to 11.3
17 17 0.22" to 0.44"
steel
0.67" 0.67"

26
Unit: mm inch

Contact

Standard 0.02" Super-tough 0.02" Flat plate (Optional) 0.02" Roller (Optional)
M2.5 × 0.45 5 0.20" M2.5 × 0.45 M2.5 × 0.45
5 0.20" 5 0.20" ø9 ø0.35"
OP-77678 OP-77682 OP-77679 OP-77680 0.02"
ø7.5 M2.5 × 0.45
ø0.30"
0.14"
3.5
3.8 0.15" SR1.5 0.06" 3.8 0.15" SR1.5 0.06" ø6 ø0.24"
4.5
0.18" ø7 0.37" 9.5
ø5 ø0.20" ø5 ø0.20" ø0.28"
Standard Standard 0.30" 7.5
equipment equipment ø9
of the general- of the high-accuracy ø0.35"
Material: TYPE304 4.5 0.18"
purpose sensor head Material: TYPE304 type sensor head Material: TYPE304
Stainless steel Material: TYPE304
GT2-H12 (L)/ Stainless steel GT2-H12K (L)/ Stainless steel
Super-tough Stainless steel
H32 (L)/H50/ TYPE440C A12K (L) Super-tough tungsten alloy
tungsten alloy TYPE440C
A12 (L)/A32/A50 Stainless steel
Stainless steel 4.4 4
0.17" 0.16" 9
0.35"
Fluorocarbon resin
(Optional) Ceramic (Optional) Needle (Optional) Offset (Optional) ø7.5 ø0.30"
0.02" 0.02" 0.02" 0.02"
OP-80228 0.20" M2.5 × 0.45 OP-81970 0.20"
5 M2.5 × 0.45 OP-77681 0.20" M2.5 × 0.45 OP-77683 ø6
M2.5 × 0.45
5 5 ø0.24"
8
0.31"
6.2 0.20" 5
9 0.24"
0.35" 3.5
3.8 SR1.5 0.06" 0.14"
SR1.5 0.06" 0.15" 21 20
ø5 ø0.20" ø5 ø0.20" 0.83" 0.79"

SR0.8 0.03" 10
C0.2 C0.008" 0.39"
Material: Fluorocarbon resin (PTFE) Material: TYPE304 Stainless steel, ø1.6 ø0.06"
Ceramic ø1.2 ø0.05"

ø4 ø0.16"
Material: TYPE304 Material: TYPE304
Stainless steel Stainless steel
Lift lever (Optional) 21.4 0.84" Spacer (Optional) M2.5 × 0.45 0.02" Super-tough tungsten alloy Super-tough tungsten alloy
OP-84397 19.2 0.76" OP-77684 5.5
3.6 8.3 0.33" 0.22"
0.14" 0.08"
12.2 2
0.09" 15.3 0.48"
2.3 0.60" M2.5 × 0.45 0.02"
Effective screw depth:
3.6 6.5 mm 0.26"
5.2 0.14"
4.7 0.8
0.20" 0.19" 0.03" ø4 ø0.16"
Material: TYPE304 Material: TYPE304
Stainless steel Stainless steel

Amplifier unit
DIN-rail mount type GT2-71N/71P/71MCN/71MCP/71CN/71CP GT2-72N/72P/72CN/72CP
* *
24 24
0.94" 14.6 0.94" 14.6
0.57" 0.57"
8.9 8.9
0.35" 0.35"

0.12" 135° max.


0.12"
3 3
(Maximum
when cover
is opened)
36.9 82.9 3.26" 41.8 36.9
41.8 1.45" 1.45"
1.65" 1.65"
12.4 17.4 12.4 17.4
0.49" 0.69" 0.49" 0.69"
15 0.59" min. 20 21.1 35.4 1.39" 15 0.59" min. 20 21.1 35.4 1.39"
0.79" 0.83" 0.79" 0.83"
75.8 2.98" 15 0.59" min. 75.8 2.98" 15 0.59" min.
(Maximum when cover
is opened) 132.6 5.22"
* Cable specifications
GT2-71N/71P: ø4.7 ø0.19", 12-core x Brown/Blue: 0.20 mm2, Black/White/Gray/Orange/Green/Pink/Purple/Yellow/Red/Pink purple: 0.15 mm2, Cable length: 2 m 6.6'
GT2-72N/72P: ø4.7 ø0.19", 10-core x Black/White/Gray/Orange/Green/Pink/Purple/Yellow/Red/Pink purple: 0.15 mm2, Cable length: 2 m 6.6'

GT2-71MCN/71MCP/71CN/71CP/72CN/72CP GT2-CA2M/CA10M
Connector Connection cable
(connector type/analog output type amplifier unit)
46 1.81" 44 1.73"

*
ø15 ø0.59" ø0.59" ø15

* Cable specifications
Cable length: 0.3 m, ø4.7
1.18', ø0.19" Outer diameter: ø4.7 mm ø0.19", Cable length: 2 m 6.6' (GT2-CA2M), 10 m 32.8' (GT2-CA10M),
12-core x Brown/Blue: 0.20 mm2, Black/White/Gray/Orange/Green/Pink/Purple/Yellow/Red/
Pink purple: 0.15 mm2

27
Dimensions

Unit: mm inch

Amplifier unit
Pulse output GT2-71D
*
0.94" 24
8.9 14.6 0.57"
0.35"

0.12" 135° max.


3
(Maximum
when cover
is opened)
36.9 82.9 3.26"
1.45"
0.49"
41.8
1.65" 12.4 17.4 0.69"

15 0.59" min. 20 21.1 35.4


0.79" 0.83" 1.39"
75.8 2.98" 15 0.59" min.
(Maximum when cover
is opened) 132.6 5.22"

* Cable specifications
Outer diameter: ø4.7 mm ø0.19", Cable length: 2 m 6.6',
9-core x Brown/Blue/Purple/Pink/Orange/Green/Gray/White/Black: 0.15 mm2

Mounting bracket for


DIN-rail mount type
amplifier (Optional) End unit (Optional) (2 pcs.) When several units are connected
OP-76877 OP-26751 L Number of
units L
0.24" 6
1 24 0.94"
5 15 2-ø3.4 2 48 1.89"
0.20" 0.59" ø0.13" 6 6 3 72 2.83"
DIN-rail mount 0.24" 0.24"
2-(4.4 × 3.4) 4 96 3.78"
(0.17" × 0.13") 5 120 4.72"
(22.6) 20.8 6 144 5.67"
(0.89") 0.82" 7 168 6.61"
24 22
0.94" 0.87" 8 192 7.56"
9.2 35.4 1.39" 53.8 53.8 9 216 8.50"
0.36" 2.12" 2.12" 10 240 9.45"
15 53.8 2.12"
0.59" 11 264 10.39"
28.6 12 288 11.34"
1.13"
Material: Polycarbonate, Stainless steel 13 312 12.28"
35 1.38" 10.1 10.1
0.40" 0.40" 14 336 13.23"
Material: SPCC 15 360 14.17"

Amplifier unit
Panel mount type GT2-75N/75P/76N/76P
44.7 1.76"
13.9 0.55"
13.4 0.53"

* Cable specifications (common to all amplifier units)


GT2-75N/75P: ø4.7 ø0.19", 12-core x Brown/Blue: 0.20 mm2, Black/White/Gray/
Orange/Green/Pink/Purple/Yellow/Red/Pink purple: 0.15 mm2,
52
2.05" Cable length: 2 m 6.6'
0.28" GT2-76N/76P: ø4.7 ø0.19", 10-core x Black/White/Gray/Orange/Green/Pink/
7 Purple/Yellow/Red/Pink purple: 0.15 mm2, Cable length: 2 m 6.6'

0.78"
0.06" 1.5 2.3 0.09" 19.8

23.9 20.9
0.94" 12.2 0.82"
* 0.48"

66 2.60" 8.5 4.1


0.33" 0.16"
Panel mounting bracket
(Accessory)
OP-84394 Panel cutout Panel mounting bracket
45 1.77" (73) (2.87")
11.9 0.47"
38.8
1.53"
Panel thickness
X = 24 x (A-1) + 21 1 to 6
X
0.94" 0.84" 0.04" to 0.24"
49 1.93" When A amplifier units are
closely mounted
24
0.94"
69.5 2.74" min.

Protective front cover 21+0.4


-0
0.84" +0.02"
-0
23.8
0.94"
45 1.77"
69 2.72"
11.7
0.46"
Material: (Mounting bracket) Polyacetal,
(Protective front cover) Polycarbonate

28
Unit: mm inch

Amplifier unit When the supplied mounting bracket is attached GT2-E3N/E3P Expansion board
Large display 0.04" to 0.24"
Panel thickness: 1 to 6 133 5.24"
GT2-100N/100P 159.5 6.28"

79.8 49.3 75.8 82.7


3.14" 1.94" 2.98" 3.26"

0.43" 11
0.48" 12.1 82 3.23" 0.05" 1.2
1.18" 85.2 3.35"
29.9
60 2.36"

15.1
1.57" 40 0.59"

97.2 Panel cutout 1.6


3.83" 0.06"
1.10" 28
(119.6)
(4.71")
Screw hole for mounting OP84331
Nominal diameter: 3 0.12", Tapping depth: 5 0.20" max. 76 2.99" +1 +0.04"
-0 -0
(139.4)
(5.49")

139.5 5.49" +1 +0.04"


-0 -0
138.8 5.46"
161.8 6.37" (Outer dimension
of the supplied mounting bracket)

When the optional mounting bracket (OP-84331) is used

84.4
3.32"

4.6 t=1.6 6.6


0.18" ø0.14" 0.06" 0.26"
4-ø3.5
1.34" 34

40
1.57" 40 1.57" 1.10" 28

Nominal diameter: 3 0.12", 4-ø5 ø0.20" 40.5 60 2.36"


fastened with four tapping screws (supplied) 1.59"
5.6 0.22" 5.6 150 5.91"
0.22" 160 6.30"
150 5.91"

Communication unit
BCD output type
DL-RB1A DIN-rail mount When the mounting bracket is attached (Optional)
OP-60412
22.5
0.89"

53.4 53.4
2.10" 2.10"
(57.8) 37.1 60.9 37.1
(2.28") 1.46" 2.40" 1.46"

34-pin MIL connector 21.1 35.4 31.4 15 2-ø3.4


0.83" 1.39" 1.24" 0.59" ø0.13"
70 2.76" 70 2.76"
2-(4.4 × 3.4)
(0.17" × 0.13")
22.5
0.89"
15
0.59"
Material: Stainless 28.5
Communication unit steel 1.12"
35.5
RS-232C communication type 1.40"
DL-RS1A DIN-rail mount When the mounting bracket is attached (Optional)
OP-60412
22.5
0.89"

(48.2) 51.3
(1.90") 37.2 2.02" 37.2
1.46" 1.46"
43.8 43.8
1.72" 1.72"

21.1 35.4 31.4 15 2-ø3.4


0.83" 1.39" 1.24" 0.59" ø0.13"
70 2.76" 70 2.76"
2-(4.4 × 3.4)
(0.17" × 0.13")
22.5
0.89"
15
CAD DATA Download Service
0.59"
Material: Stainless
28.5
1.12"
www.keyence.com/cad
steel 35.5
1.40"

29
Product configuration

Sensor head cable Display unit (amplifier unit)


→P.15/step 2 →P.16/step 4-5

Sensor head
→P.13/step 1

Communication unit
→P.18/step 7-8
Mounting bracket
→P.15/step 3

Contact
→P.18/step 6

Reference

❙ Calculation functions using expansion units


The GT2-70 Series* enables calculations such as the maximum and the About “Calculation mode”
minimum values of several detection points when several expansion units When “Calculation mode” is selected for “A1: Calculation mode”, the main
(up to 14 units) are connected to a main unit. Outline of each calculation unit outputs a calculation result, and output of the expansion unit is OFF.
function is as follows: (Except for a case where “C5: Difference from reference display” is
selected)
Number of connectable
No. Calculation
expansion units
Description
function Dedicated
Calculation mode Calculation mode
NOTE - The calculation mode can be selected for the main unit to which one or
C1 Maximum Displays the maximum value of the values of more expansion units are connected.
value the main unit and expansion unit(s).

Minimum Displays the minimum value of the values of


C2 value the main unit and expansion unit(s). About “Calculation dedicated mode”
2 to 14 units
Displays the difference between the maximum When “Calculation dedicated mode” is selected for “A1: Calculation
Degree of
C3 and the minimum value of the main unit and 1 to 14 units
flatness mode”, only the main unit outputs a calculation result, and the expansion
expansion unit(s).
Displays the resulting calculation of the unit outputs a judgment result based on each setting.
C4 Average average of the main unit and expansion unit(s).

Reference Displays the difference obtained by subtracting NOTE - The calculation dedicated mode can be selected for the main unit to which
C5 the display value of the main unit from each of Not selectable two or more expansion units are connected.
difference
the expansion unit(s).
- When the calculation dedicated mode is selected, “C5: Reference
Displays the degree of twist calculated from the Difference” cannot be selected for calculation method.
C6 Twist Only 3 units Only 4 units
values of four sensor heads.
- When the calculation dedicated mode is selected, a sensor head cannot
Displays the degree of warpage calculated be connected to the main unit. Attach the supplied “sensor head connector
C7 Warpage Only 2 units Only 3 units seal” to the sensor head connector of the main unit.
from the values three sensor heads.

C8 Thickness Displays the thickness calculated by two sensor Only 1 unit Only 2 units
heads "sandwiching" a target between them.

* Except GT2-71D.

❙ Bank function ❙ 5-output function (Except GT2-71MCN/71MCP)


With the Bank function, up to four patterns of HH, HIGH, LOW and LL This function enables differentiation of five output statuses (HH,
set values, preset values can be registered in advance. HIGH, GO, LOW and LL). HH and LL outputs can also be used for
Since the registered settings (four patterns) can be easily changed, this jam detection.
function is useful for detection of several targets.

❙ About accuracy of the high-accuracy type


When the high-accuracy type measures a distance between two
arbitrary points in the measuring range at an ambient temperature (20°C
(68°F)), the deviation from the actual value is 1 µm 0.04 Mil or less. For
example, when a measurement result is 11 mm 0.43", the actual value
is in a range of 10.999 to 11.001 mm 0.4330" to 0.4331". For several
measurements, variations in the measured value are 1 µm 0.04 Mil or
less.
30
General purpose digital contact sensor
GT Series

Features
[Easy setup]
Head installation is as easy as drilling
a hole.
[Easy programming]
Auto-tuning function enables simple,
sensor-like tuning.
[Smart operation]
Self-diagnostic function outputs an
alarm when the sensor head cable
becomes damaged

Setup
Amplifier units BCD output / RS-232C communication unit
Model Appearance Type Output type Model Appearance Type
GT-71A Main unit
NPN
GT-72A DIN rail Expansion unit
DL-RB1A BCD
GT-71AP mount Main unit
PNP
GT-72AP Expansion unit
GT-75A Main unit
NPN
GT-76A Panel Expansion unit
DL-RS1A RS-232C
GT-75AP mount Main unit
PNP
GT-76AP Expansion unit

Standard model Low Stress model Air Push model

Shortest in its class


Total length 90 mm 3.54"
Smallest length among the 10-mm
0.39" stroke type sensor heads

Using
low force
spring

10-mm 0.39" 22-mm 0.87"


10-mm 0.39" 22-mm 0.87" 10-mm 0.39" 22-mm 0.87" stroke air stroke air
stroke sensor stroke sensor stroke, low stress stroke, low stress push type push type
head head sensor head sensor head sensor head sensor head
GT-H10 GT-H22 GT-H10L GT-H22L GT-A10 GT-A22

31
http://www.measurecentral.com
NEW
Web Site Dedicated to
Measurement Solutions!
❙ TECHNICAL INFORMATION
❙ APPLICATION SEARCH FUNCTION
❙ FREE SOFTWARE DOWNLOAD

Laser Displacement (1D) LK-G Series

Measuring the thickness of a silicon Measuring the thickness of a glass


wafer plate

❙ Sampling rate of 50 kHz


❙ Linearity of ± 0.03% of F.S.
❙ Repeatability down to 0.01 µm 0.0004 Mil

Thickness measurement/loop Measuring the displacement of a


control of a rubber sheet rotating shaft

Specifications are subject to change without notice.

CALL
TOLL
TO CONTACT YOUR LOCAL OFFICE
www.keyence.com
FREE 1 - 8 8 8 - 5 3 9 - 3 6 2 3 Fax : 201-930-0099
KEYENCE CORPORATION OF AMERICA
Corporate Office 50 Tice Blvd., Woodcliff Lake, NJ 07677 Phone: 201-930-0100 Fax: 201-930-0099 E-mail: keyence@keyence.com
■ Regional offices CO Denver IN Indianapolis MI Detroit NJ Woodcliff Lake OH Cincinnati SC Greenville VA Richmond
AL Birmingham FL Tampa KS Kansas City MI Grand Rapids NY Rochester OH Cleveland TN Nashville WA Seattle
CA N.California GA Atlanta KY Louisville MN Minneapolis NC Charlotte OR Portland TN Knoxville
CA Los Angeles IL Chicago MA Boston MO St. Louis NC Raleigh PA Philadelphia TX Dallas

KEYENCE CANADA INC. KEYENCE MEXICO S.A. DE C.V.


Head Office Phone: 905-696-9970 Fax: 905-696-8340 E-mail: keyencecanada@keyence.com Phone: +52-81-8220-7900 Fax: +52-81-8220-9097
Montreal Phone: 514-694-4740 Fax: 514-694-3206 E-mail: keyencemexico@keyence.com
KA1-0049

Copyright (c) 2009 KEYENCE CORPORATION. All rights reserved. GT2-KA-C2-E 0069-4 611176 Printed in Japan

* 6 1 1 1 7 6 *

Вам также может понравиться