Академический Документы
Профессиональный Документы
Культура Документы
These can be mitigated by subsequent processes and also This paper shall covers the defect encountered at pad
through device and circuit design. There are two etch which is shown in Fig. 3. The problem was detected
during wafer disposition due to backside helium leak
alarm. A burnt like material was observed to have
inhibited the wafer and exposing the top metal. The
wafer is believed to encounter plasma instability during
transition from Main Etch (ME) step to Over Etch (OE)
step. However, this is only detected during backside
helium leak alarm.
345
TABLE II
TABLE I
RESULT FROM TABLE I
DESIGN OF EXPERIMENT TO MINIMIZE
Std Run RF Chucking BHe
ARCING DEFECT Order Order Power Voltage Flow Result
Std Chucking BHe 5 1 0 1 1 0.9
Order Run Order RF Power Volt Flow 3 2 0 0 0 1.2
5 1 0 1 1 8 3 1 0 1 1.5
3 2 0 0 0 7 4 0 0 1 1.6
8 3 1 0 1 4 5 1 0 0 19.0
7 4 0 0 1 2 6 1 1 0 1.2
4 5 1 0 0 1 7 0 1 0 1.0
2 6 1 1 0 6 8 1 1 1 1.6
1 7 0 1 0
6 8 1 1 1
using visual image only. This is also to study the effect
The factors selected for analyses were: of polymer peeling which could drop onto the ESC which
1) RF Power could cause backside helium leak alarm flow fault to
2) Chucking voltage occur.
3) Backside Helium (BHe) flow
IV. RESULTS & DISCUSSION
The method for analysis is using the Design of A. Hardware Optimal Setting
Experiment Tool. These factors require high and low
values which were obtained from other baseline process The result for the experiment is shown in Table 2
and from vendors’ recommendation. The preferred above. The result is presented in such way that zero
setting for backside helium flow is the inner zone must would be the best result without any leak rate flow fault.
always be flowing higher than outer zone. The chucking This result represents the leak rate value that occurred
voltages proposed for the experiments were -500 volt (V) during the alarm. Run order 5 reported the highest leak
as high condition and -750V as low condition. The RF rate and upon verification, it is found that the wafer
power employed for these experiments were 1600 watt backside has white polymer sticking due to incoming
(W) as high value while 500 W as low value. The process.
backside helium setting for inner and outer were 16 torr
(T)/ 12T as high condition while 15 T/9 T as low This white polymer is later confirmed to be coming
condition. The design of experiment is tabulated in Table from the deposition chambers. This caused higher
1. The result is categorized as having a minimum backside helium leak rate since the helium flow is
backside helium leak rate flow fault alarm as 0 while obstructed by foreign materials.
getting maximum backside helium leak rate flow fault
alarm as 1 from these standard run. Comparing between backside helium (BHe) flow of
15/9 T and 16/12 T, pad dual profile was observed at the
B. Chamber Mix Run wafer edge as shown in Fig. 4. This is due to temperature
The final approach is to segregate these products into factor. The backside helium cooling is responsible for
two main groups for data collection. The period selected the temperature cooling efficiency across the wafer.
is during one chamber wet clean duration. Product with However, there are limitation due to thickness with
low RF power is dedicated inside one chamber and respect to the numbers of metal layers stack and the
product with low RF power is mixed run with high RF backside helium flow pressure. The current setting was
tested using 5 metal layers. In the case of more than 5
346
The chucking voltage also plays an important role in
controlling the wafer backside helium flow. Steady wafer
control on the Electro Static Chuck (ESC) was achieved
by having negative chucking. The negative chucking
voltage allows small gap for the backside helium to flow
through causing less helium leak rate alarm to occur. The
steady flow of BHe is obtained using 16/12 T and
negative chucking ESC.
B. Chamber Mix Run
The wet clean duration for one pad chamber is set to
150 RF hours. However, this is hard to achieve due to
arcing. ESC has to be verified in good working condition Fig. 8 Polymer blocking helium flow.
to prevent lift pin from melting and the only way to
perform this is to open the chamber for wet clean.
There is also impact observed from mix run chamber
After chamber dedication, high RF power chamber is
in term of polymer peeling and dropping onto the ESC.
able to reach the maximum lifetime which is around RF
This is prevented by having high RF power recipe and
148 RF hours, which normally could only be achieved
low RF power recipe being separated into different
around 80 RF hours due to backside helium leak flow
process chambers.
fault alarm. Fig.6 shows the image taken from the
Incoming process also contributed to backside helium
product mix run process kits during chamber dedication
flow obstruction with reference to Run Order 4, where
period. Polymer is observed to be peeling from the
white polymer is observed at the wafer backside causing
process kits. This is compared with uniform polymer
high leak rate flow fault alarm. This is prevented by
sticking on process kit chamber wall in Fig. 7 taken from
highlighting to Thin Film group to ensure the robot blade
the single product dedication for one wet clean duration
is free from particle to prevent polymer sticking to wafer
period. The polymer is observed to be uniformly
backside.
distributed on the chamber wall reducing the risk of
Backside helium flow plays an important role in
polymer peeling and blocking the backside helium flow
cooling the temperature across the wafer during
on the ESC as shown in Fig 8.
processing, however care to be taken for five metal layers
where the backside helium cooling will not be able to
V. CONCLUSIONS
remove the heat from the photo resist causing it to melt
The arcing defect detected at pad etch is presented. and creating the pad dual profile.
This occurred due to plasma instability caused by Currently, there is no arcing defect detected since
obstruction at backside helium flow. The design of recipe release and implementation. Due to this factor, all
experiment to find the optimal setting for backside the chambers are able to reach maximum wet clean cycle.
helium flow and chucking voltage were presented.
Recipe optimization is performed to ensure there is no ACKNOWLEDGMENT
equipment alarm that could lead to arcing defect. The The paper is not possible without the support and
arcing defect could be prevented by having low RF encouragement from the X-FAB Management Team. We
power, no magnetic field, shorter processing time and would like to take this opportunity to extend our
lower pressure based recipe. appreciation to the Defect Management, Yield
Enhancement, Quality, Reliability & Assurance and
Operation. We would like to express our humble
gratitude to Miss. Florence Chai from DM for her
technical advice and contributions for this paper.
REFERENCE
347