Академический Документы
Профессиональный Документы
Культура Документы
CHAPTER
Analytical Techniques
Julia C. Drees, Alan H. B. Wu 5
C H A P T E R O U T L I N E
130
14460_Ch05.qxd 12/4/08 8:06 PM Page 131
path, as in Figure 5-3B. Most of the light is transmitted, that, as concentration increases, % T decreases in a loga-
but a small amount is absorbed by the solvent and cuvet rithmic manner.
or is reflected away from the detector. The electrical Absorbance A is the amount of light absorbed. It can-
readout of the instrument is set arbitrarily at 100% T, not be measured directly by a spectrophotometer but
while the light is passing through a “blank” or reference. rather is mathematically derived from % T as follows:
The sample containing absorbing molecules to be meas-
ured is placed in the light path. The difference in amount %T I 100 (Eq. 5-2)
of light transmitted by the blank and that transmitted by I0
the sample is due only to the presence of the compound where I0 is incident light and I is transmitted light.
being measured. The % T measured by commercial spec- Absorbance is defined as follows:
trophotometers is the ratio of the sample transmitted
beam divided by the blank transmitted beam. A log(I/I0) log (100%) log %T
Equal thicknesses of an absorbing material will absorb 2 log %T (Eq. 5-3)
a constant fraction of the energy incident upon the lay-
ers. For example, in a tube containing layers of solution According to Beer’s law, absorbance is directly pro-
(Fig. 5-4A), the first layer transmits 70% of the light in- portional to concentration (Fig. 5-4D):
cident upon it. The second layer will, in turn, transmit
Abc (Eq. 5-4)
70% of the light incident upon it. Thus, 70% of 70%
(49%) is transmitted by the second layer. The third layer where molar absorptivity, the fraction of a specific
transmits 70% of 49%, or 34% of the original light. wavelength of light absorbed by a given type of molecule;
Continuing on, successive layers transmit 24% and 17%, b is the length of light path through the solution; and c
respectively. The % T values, when plotted on linear is the concentration of absorbing molecules.
graph paper, yield the curve shown in Figure 5-4B. Absorptivity depends on molecular structure and the
Considering each equal layer as many monomolecular way in which the absorbing molecules react with differ-
layers, we can translate layers of material to concentra- ent energies. For any particular molecular type, absorp-
tion. If semilog graph paper is used to plot the same fig- tivity changes as wavelength of radiation changes. The
ures, a straight line is obtained (Fig. 5-4C), indicating amount of light absorbed at a particular wavelength
FIGURE 5-4. (A) Percent of original incident light transmitted by equal layers of light-absorbing
solution. (B) Percent T versus concentration on linear graph paper. (C) Percent T versus concentration on
semilog graph paper. (D) A versus concentration on linear graph paper.
14460_Ch05.qxd 12/4/08 8:06 PM Page 133
Entrance Exit
slit Sample PM tube
Light slit cuvet
source Monochromator
Grating
A/D Display
depends on the molecular and ion types present and may Monochromators
vary with concentration, pH, or temperature. Isolation of individual wavelengths of light is an impor-
Because the path length and molar absorptivity are tant and necessary function of a monochromator. The
constant for a given wavelength, degree of wavelength isolation is a function of the type of
device used and the width of entrance and exit slits. The
A⬃C (Eq. 5-5)
bandpass of a monochromator defines the range of wave-
Unknown concentrations are determined from a cali- lengths transmitted and is calculated as width at more
bration curve that plots absorbance at a specific wave- than half the maximum transmittance (Fig. 5-6).
length versus concentration for standards of known Numerous devices are used for obtaining monochro-
concentration. For calibration curves that are linear and matic light. The least expensive are colored-glass filters.
have a zero y-intercept, unknown concentrations can be These filters usually pass a relatively wide band of radi-
determined from a single calibrator. Not all calibration ant energy and have a low transmittance of the selected
curves result in straight lines. Deviations from linearity wavelength. Although not precise, they are simple, inex-
are typically observed at high absorbances. The stray pensive, and useful.
light within an instrument will ultimately limit the max- Interference filters produce monochromatic light
imum absorbance that a spectrophotometer can achieve, based on the principle of constructive interference of
typically 2.0 absorbance units. waves. Two pieces of glass, each mirrored on one side,
are separated by a transparent spacer that is precisely
Spectrophotometric Instruments one-half the desired wavelength. Light waves enter one
side of the filter and are reflected at the second surface.
A spectrophotometer is used to measure the light trans-
Wavelengths that are twice the space between the two
mitted by a solution to determine the concentration of the
glass surfaces will reflect back and forth, reinforcing oth-
light-absorbing substance in the solution. Figure 5-5 illus-
ers of the same wavelengths, and finally passing on
trates the basic components of a single-beam spectropho-
through. Other wavelengths will cancel out because of
tometer, which are described in subsequent sections.
phase differences (destructive interference). Because
interference filters also transmit multiples of the desired
Components of a Spectrophotometer
Light Source
The most common source of light for work in the visible
and near-infrared region is the incandescent tungsten or
tungsten-iodide lamp. Only about 15% of radiant energy
emitted falls in the visible region, with most emitted as
near-infrared.1–3 Often, a heat-absorbing filter is inserted
between the lamp and sample to absorb the infrared
radiation.
The lamps most commonly used for ultraviolet (UV)
work are the deuterium-discharge lamp and the mer-
cury-arc lamp. Deuterium provides continuous emission
down to 165 nm. Low-pressure mercury lamps emit a
sharp-line spectrum, with both UV and visible lines.
Medium and high-pressure mercury lamps emit a con-
tinuum from UV to the mid visible region. The most im-
portant factors for a light source are range, spectral
distribution within the range, the source of radiant pro- FIGURE 5-6. Spectral transmittance of two monochromators with
duction, stability of the radiant energy, and temperature. band pass at half height of 5 nm and 20 nm.
14460_Ch05.qxd 12/4/08 8:06 PM Page 134
wavelengths, they require accessory filters to eliminate composed of a film of light-sensitive material, frequently
these harmonic wavelengths. Interference filters can be selenium, on a plate of iron. Over the light-sensitive ma-
constructed to pass a very narrow range of wavelengths terial is a thin, transparent layer of silver. When exposed
with good efficiency. to light, electrons in the light-sensitive material are ex-
The prism is another type of monochromator. A nar- cited and released to flow to the highly conductive silver.
row beam of light focused on a prism is refracted as it In comparison with the silver, a moderate resistance op-
enters the more dense glass. Short wavelengths are poses the electron flow toward the iron, forming a hypo-
refracted more than long wavelengths, resulting in dis- thetical barrier to flow in that direction. Consequently,
persion of white light into a continuous spectrum. The this cell generates its own electromotive force, which
prism can be rotated, allowing only the desired wave- can be measured. The produced current is proportional
length to pass through an exit slit. to incident radiation. Photocells require no external
Diffraction gratings are most commonly used as mo- voltage source but rely on internal electron transfer to
nochromators. A diffraction grating consists of many produce a current in an external circuit. Because of their
parallel grooves (15,000 or 30,000 per inch) etched onto low internal resistance, the output of electrical energy is
a polished surface. Diffraction, the separation of light not easily amplified. Consequently, this type of detector
into component wavelengths, is based on the principle is used mainly in filter photometers with a wide band-
that wavelengths bend as they pass a sharp corner. The pass, producing a fairly high level of illumination so
degree of bending depends on the wavelength. As the that there is no need to amplify the signal. The photocell
wavelengths move past the corners, wave fronts are is inexpensive and durable; however, it is temperature
formed. Those that are in phase reinforce one another, sensitive and nonlinear at very low and very high levels
whereas those not in phase cancel out and disappear. of illumination.
This results in complete spectra. Gratings with very fine A phototube (Fig. 5-7) is similar to a barrier-layer
line rulings produce a widely dispersed spectrum. They cell in that it has photosensitive material that gives off
produce linear spectra, called orders, in both directions electrons when light energy strikes it. It differs in that
from the entrance slit. Because the multiple spectra have an outside voltage is required for operation. Phototubes
a tendency to cause stray light problems, accessory filters contain a negatively charged cathode and a positively
are used. charged anode enclosed in a glass case. The cathode is
composed of a material (e.g., rubidium or lithium) that
Sample Cell acts as a resistor in the dark but emits electrons when
The next component of the basic spectrophotometer is exposed to light. The emitted electrons jump over to
the sample cell or cuvet, which may be round or square. the positively charged anode, where they are collected
The light path must be kept constant to have absorbance and return through an external, measurable circuit. The
proportional to concentration. This is easily checked by cathode usually has a large surface area. Varying the
preparing a colored solution to read midscale when using cathode material changes the wavelength at which the
the wavelength of maximum absorption. Fill each cuvet phototube gives its highest response. The photocurrent
to be tested, take readings, and save those that match is linear, with the intensity of the light striking the cath-
within an acceptable tolerance (e.g., 0.25% T). Because ode as long as voltage between the cathode and anode
it is difficult to manufacture round tubes with uniform
diameters, they should be etched to indicate the position
for use. Cuvets are sold in matched sets. Square cuvets
have plane-parallel optical surfaces and a constant light
path. They have an advantage over round cuvets in that
there is less error from the lens effect, orientation in the
spectrophotometer, and refraction. Cuvets with scratch-
ed optical surfaces scatter light and should be discarded.
Inexpensive glass cuvets can be used for applications in
the visible range, but they absorb light in the UV region.
Quartz cuvets must, therefore, be used for applications
requiring UV radiation.
Photodetectors
The purpose of the detector is to convert the transmit-
ted radiant energy into an equivalent amount of electri-
cal energy. The least expensive of the devices is known
as a barrier-layer cell, or photocell. The photocell is FIGURE 5-7. Phototube drawing and schematic.
14460_Ch05.qxd 12/4/08 8:06 PM Page 135
Grating
Dynode
chain
Incident
light
Shield
Photocathode Cell
Anode
Slit
Photodiode
remains constant. A vacuum within the tubes avoids array
scattering of the photoelectrons by collision with gas
molecules.
The third major type of light detector is the photomul-
FIGURE 5-9. Photodiode array spectrophotometer illustrating the
tiplier (PM) tube, which detects and amplifies radiant en- placement of the sample cuvet before the monochromator.
ergy. As shown in Figure 5-8, incident light strikes the
coated cathode, emitting electrons. The electrons are at-
tracted to a series of anodes, known as dynodes, each hav- sensitive as PM tubes because of the lack of internal am-
ing a successively higher positive voltage. These dynodes plification, their excellent linearity (6–7 decades of radi-
are of a material that gives off many secondary electrons ant power), speed, and small size make them useful in
when hit by single electrons. Initial electron emission at applications where light levels are adequate.4 Photodiode
the cathode triggers a multiple cascade of electrons within array (PDA) detectors are available in integrated circuits
the PM tube itself. Because of this amplification, the PM containing 256 to 2,048 photodiodes in a linear arrange-
tube is 200 times more sensitive than the phototube. PM ment. A linear array is shown in Figure 5-9. Each photo-
tubes are used in instruments designed to be extremely diode responds to a specific wavelength, and as a result,
sensitive to very low light levels and light flashes of very a complete UV/visible spectrum can be obtained in less
short duration. The accumulation of electrons striking than 1 second. Resolution is 1 to 2 nm and depends on
the anode produces a current signal, measured in am- the number of discrete elements. In spectrophotometers
peres, that is proportional to the initial intensity of the using PDA detectors, the grating is positioned after the
light. The analog signal is converted first to a voltage sample cuvet and disperses the transmitted radiation
and then to a digital signal through the use of an analog- onto the PDA detector (Fig. 5-9).
to-digital (A/D) converter. Digital signals are processed For single-beam spectrophotometers, the absorbance
electronically to produce absorbance readings. reading from the sample must be blanked using an ap-
In a photodiode, absorption of radiant energy by a propriate reference solution that does not contain the
reverse-biased pn-junction diode (pn, positive-negative) compound of interest. Double-beam spectrophotome-
produces a photocurrent that is proportional to the inci- ters permit automatic correction of sample and refer-
dent radiant power. Although photodiodes are not as ence absorbance, as shown in Figure 5-10. Because the
Entrance Exit
slit Sample PM tube
Light slit
cuvet
source Monochromator
Grating
A/D Display
Beam
splitters
Reference
cuvet
FIGURE 5-10. Double-beam spectrophotometer.
14460_Ch05.qxd 12/4/08 8:06 PM Page 136
intensities of light sources vary as a function of wave- light path and higher-order spectra produced by diffract-
length, double-beam spectrophotometers are necessary ion gratings. The major effect is absorbance error, espe-
when the absorption spectrum for a sample is to be ob- cially in the high-absorbance range. Stray light is
tained. Computerized, continuous zeroing, single-beam detected by using cutoff filters, which eliminate all radi-
spectrophotometers have replaced most double-beam ation at wavelengths beyond the one of interest. To
spectrophotometers. check for stray light in the near-UV region, for example,
insert a filter that does not transmit in the region of 200
Spectrophotometer Quality Assurance nm to 400 nm. If the instrument reading is greater than
Performing at least the following checks should validate 0% T, stray light is present. Certain liquids, such as
instrument function: wavelength accuracy, stray light, NiSO4, NaNO2, and acetone, absorb strongly at short
and linearity. Wavelength accuracy means that the wave- wavelengths and can be used in the same way to detect
length indicated on the control dial is the actual wave- stray light in the UV range.
length of light passed by the monochromator. It is most Linearity is demonstrated when a change in concen-
commonly checked using standard absorbing solutions tration results in a straight-line calibration curve, as
or filters with absorbance maxima of known wavelength. discussed under Beer’s law. Colored solutions may be
Didymium or holmium oxide in glass is stable and fre- carefully diluted and used to check linearity, using the
quently used as filters. The filter is placed in the light wavelength of maximal absorbance for that color. Sealed
path and the wavelength control is set at the wavelength sets of different colors and concentrations are available
at which maximal absorbance is expected. The wave- commercially. They should be labeled with expected ab-
length control is then rotated in either direction to locate sorbance for a given bandpass instrument. Less than ex-
the actual wavelength that has maximal absorbance. If pected absorbance is an indication of stray light or of a
these two wavelengths do not match, the optics must be bandpass that is wider than specified. Sets of neutral-
adjusted to calibrate the monochromator correctly. density filters to check linearity over a range of wave-
Some instruments with narrow bandpass use a mer- lengths are also commercially available.
cury-vapor lamp to verify wavelength accuracy. The mer- A routine system should be devised for each instrument
cury lamp is substituted for the usual light source, and to check and record each parameter. The probable cause
the spectrum is scanned to locate mercury emission of a problem and the maintenance required to eliminate it
lines. The wavelength indicated on the control is com- are generally described in the instrument’s manual.
pared with known mercury emission peaks to determine
Atomic Absorption Spectrophotometer
the accuracy of the wavelength indicator control.
Stray light refers to any wavelengths outside the band The atomic absorption spectrophotometer is used to
transmitted by the monochromator. The most common measure concentration by detecting absorption of elec-
causes of stray light are reflection of light from scratches tromagnetic radiation by atoms rather than by molecules.
on optical surfaces or from dust particles anywhere in the The basic components are shown in Figure 5-11. The
usual light source, known as a hollow-cathode lamp, measuring circuit is tuned to the modulated frequency.
consists of an evacuated gas-tight chamber containing an Interference from the constant flame emission is elec-
anode, a cylindrical cathode, and an inert gas, such as tronically eliminated by accepting only the pulsed sig-
helium or argon. When voltage is applied, the filler gas is nal from the hollow cathode.
ionized. Ions attracted to the cathode collide with the The monochromator is used to isolate the desired
metal, knock atoms off, and cause the metal atoms to be emission line from other lamp emission lines. In addi-
excited. When they return to the ground state, light en- tion, it serves to protect the photodetector from excessive
ergy is emitted that is characteristic of the metal in the light emanating from flame emissions. A PM tube is the
cathode. Generally, a separate lamp is required for each usual light detector.
metal (e.g., a copper hollow-cathode lamp is used to Flameless atomic absorption requires an instrument
measure Cu). modification that uses an electric furnace to break chem-
Electrodeless discharge lamps are a relatively new ical bonds (electrothermal atomization). A tiny graphite
light source for atomic absorption spectrophotometers. A cylinder holds the sample, either liquid or solid. An elec-
bulb is filled with argon and the element to be tested. A tric current passes through the cylinder walls, evaporates
radiofrequency generator around the bulb supplies the the solvent, ashes the sample and, finally, heats the unit
energy to excite the element, causing a characteristic to incandescence to atomize the sample. This instru-
emission spectrum of the element. ment, like the spectrophotometer, is used to determine
The analyzed sample must contain the reduced metal the amount of light absorbed. Again, Beer’s law is used
in the atomic vaporized state. Commonly, this is done for calculating concentration. A major problem is that
by using the heat of a flame to break the chemical bonds background correction is considerably more necessary
and form free, unexcited atoms. The flame is the sam- and critical for electrothermal techniques than for flame-
ple cell in this instrument, rather than a cuvet. There based atomic absorption methods. Currently, the most
are various designs; however, the most common burner common approach uses a deuterium lamp as a secondary
is the premix long-path burner. The sample, in solu- source and measures the difference between the two ab-
tion, is aspirated as a spray into a chamber, where it is sorbance signals. However, there has also been extensive
mixed with air and fuel. This mixture passes through development of background correction techniques based
baffles, where large drops fall and are drained off. Only on the Zeeman effect.1 The presence of an intense static
fine droplets reach the flame. The burner is a long, nar- magnetic field will cause the wavelength of the emitted
row slit, to permit a longer path length for absorption radiation to split into several components. This shift in
of incident radiation. Light from the hollow-cathode wavelength is the Zeeman effect.
lamp passes through the sample of ground-state atoms Atomic absorption spectrophotometry is sensitive and
in the flame. The amount of light absorbed is propor- precise. It is routinely used to measure concentration of
tional to the concentration. When a ground-state atom trace metals that are not easily excited. It is generally
absorbs light energy, an excited atom is produced. The more sensitive than flame emission because the vast ma-
excited atom then returns to the ground state, emitting jority of atoms produced in the usual propane or air-
light of the same energy as it absorbed. The flame sam- acetylene flame remain in the ground state available for
ple thus contains a dynamic population of ground-state light absorption. It is accurate, precise, and specific. One
and excited atoms, both absorbing and emitting radiant disadvantage, however, is the inability of the flame to
energy. The emitted energy from the flame will go in all dissociate samples into free atoms. For example, phos-
directions, and it will be a steady emission. Because the phate may interfere with calcium analysis by formation
purpose of the instrument is to measure the amount of of calcium phosphate. This may be overcome by adding
light absorbed, the light detector must be able to distin- cations that compete with calcium for phosphate.
guish between the light beam emitted by the hollow- Routinely, lanthanum or strontium is added to samples
cathode lamp and that emitted by excited atoms in the to form stable complexes with phosphate. Another possi-
flame. To do this, the hollow-cathode light beam is ble problem is the ionization of atoms following dissoci-
modulated by inserting a mechanical rotating chopper ation by the flame, which can be decreased by reducing
between the light and the flame or by pulsing the elec- the flame temperature. Matrix interference, due to the
tric supply to the lamp. Because the light beam being enhancement of light absorption by atoms in organic sol-
absorbed enters the sample in pulses, the transmitted vents or formation of solid droplets as the solvent evap-
light also will be in pulses. There will be less light in orates in the flame, can be another source of error. This
the transmitted pulses because part of it will be ab- interference may be overcome by pretreatment of the
sorbed. There are, therefore, two light signals from the sample by extraction.5
flame—an alternating signal from the hollow-cathode Recently, inductively coupled plasma (ICP) has been
lamp and a direct signal from the flame emission. The used to increase sensitivity for atomic emission. The
14460_Ch05.qxd 12/4/08 8:06 PM Page 138
Fluorescence concentration measurements are related prepared to demonstrate that the concentration used falls
to molar absorptivity of the compound, intensity of the in a linear range.
incident radiation, quantum efficiency of the energy In fluorescence polarization, radiant energy is polar-
emitted per quantum absorbed, and length of the light ized in a single plane. When the sample (fluorophor) is
path. In dilute solutions with instrument parameters excited, it emits polarized light along the same plane as
held constant, fluorescence is directly proportional to the incident light if the fluorophor is attached to a large
concentration. Generally, a linear response will be ob- molecule. In contrast, a small molecule emits depolar-
tained until the concentration of the fluorescent species ized light because it will rotate out of the plane of polar-
is so high that the sample begins to absorb significant ization during its excitation lifetime. This technique is
amounts of excitation light. A curve demonstrating non- widely used for the detection of therapeutic and abused
linearity as concentration increases is shown in Figure drugs. In the procedure, the sample analyte is allowed to
5-14. The solution must absorb less than 5% of the ex- compete with a fluorophor-labeled analyte for a limited
citing radiation for a linear response to occur.6 As with antibody to the analyte. The lower the concentration of
all quantitative measurements, a standard curve must be the sample analyte, the higher is the macromolecular
Concentrated solution:
Dilute solution:
Intensity
trophotometry: specificity and sensitivity. Fluorometry
increases specificity by selecting the optimal wavelength
for both absorption and fluorescence, rather than just the
absorption wavelength seen with spectrophotometry.
Fluorometry is approximately 1,000 times more sen-
sitive than most spectrophotometric methods.6 One rea-
son is because emitted radiation is measured directly; it
can be increased simply by increasing the intensity of the
exciting radiant energy. In addition, fluorescence meas- Time
ures the amount of light intensity present over a zero FIGURE 5-15. Representative intensity-versus-time curve for a
transient chemiluminescence signal.
background. In absorbance, however, the quantity of ab-
sorbed light is measured indirectly as the difference be-
tween the transmitted beams. At low concentrations, the Advantages of chemiluminescence assays include sub-
small difference between 100% T and the transmitted picomolar detection limits, speed (with flash-type reac-
beam is difficult to measure accurately and precisely, tions, light is only measured for 10 seconds), ease of use
limiting the sensitivity. (most assays are one-step procedures), and simple in-
The biggest disadvantage is that fluorescence is very strumentation.7 The main disadvantage is that impurities
sensitive to environmental changes. Changes in pH affect can cause a background signal that degrades sensitivity
availability of electrons, and temperature changes the and specificity.
probability of loss of energy by collision rather than fluo-
rescence. Contaminating chemicals or a change of solvents
may change the structure. UV light used for excitation can Turbidity and Nephelometry
cause photochemical changes. Any decrease in fluores- Turbidimetric measurements are made with a spec-
cence resulting from any of these possibilities is known as trophotometer to determine concentration of particu-
quenching. Because so many factors may change the in- late matter in a sample. The amount of light blocked by
tensity or spectra of fluorescence, extreme care is manda- a suspension of particles depends not only on concen-
tory in analytic technique and instrument maintenance. tration but also on size. Because particles tend to aggre-
gate and settle out of suspension, sample handling
Chemiluminescence becomes critical. Instrument operation is the same as
In chemiluminescence reactions, part of the chemical en- for any spectrophotometer.
ergy generated produces excited intermediates that decay Nephelometry is similar, except that light scattered by
to a ground state with the emission of photons.7 The the small particles is measured at an angle to the beam
emitted radiation is measured with a PM tube, and the incident on the cuvet. Figure 5-16 demonstrates two
signal is related to analyte concentration. Chemilumi- possible optical arrangements for a nephelometer. Light
nescence is different than fluorescence in that no excita- scattering depends on wavelength and particle size. For
tion radiation is required and no monochromators are macromolecules with a size close to or larger than the
needed because the chemiluminescence arises from one
species. Most important, chemiluminescence reactions
are oxidation reactions of luminol, acridinium esters, Cuvet
and dioxetanes characterized by a rapid increase in in-
Detector,
tensity of emitted light followed by a gradual decay. spectrophotometer
Usually, the signal is taken as the integral of the entire turbidometry
Light
peak. Enhanced chemiluminescence techniques increase source
Detector, nephelometer
the chemiluminescence efficiency by including an en- forward light scatter
hancer system in the reaction of a chemiluminescent
Detector
agent with an enzyme. The time course for the light in- nephelometer
tensity is much longer (60 minutes) than that for con- 90° light scatter
ventional chemiluminescent reactions, which last for FIGURE 5-16. Nephelometer versus spectrophotometer—optical
about 30 seconds (Fig. 5-15). arrangements.
14460_Ch05.qxd 12/4/08 8:06 PM Page 141
e
Salt bridge
Anode Cathode
Oxidation Reduction
2Ag° 2Ag 2E O2 2e 2OH
Field effect
transistor Macroelectrode Gas-Sensing Electrodes
FIGURE 5-20. Other examples of ion-selective electrodes. Gas electrodes are similar to pH glass electrodes but are
designed to detect specific gases (e.g., CO2 and NH3)
in solutions and are usually separated from the solution
wrapped around the indicator electrode. The outer glass by a thin, gas-permeable hydrophobic membrane. Figure
envelope is filled with KCl and has a tiny pore near the tip 5-21 shows a schematic illustration of the pCO2 elec-
of the liquid junction. The solution to be measured must trode. The membrane in contact with the solution is per-
completely cover the glass tip. Examples of other ISEs are meable only to CO2, which diffuses into a thin film of
shown in Figure 5-20. The reference electrode, electrom- sodium bicarbonate solution. The pH of the bicarbonate
eter, and calibration system described for pH measure- solution is changed as follows:
ments are applicable to all ISEs.
There are three major ISE types: inert-metal elec- CO2 H2O ↔ H2CO3 ↔ H HCO3 (Eq. 5-7)
trodes in contact with a redox couple, metal electrodes The change in pH of the HCO3 is detected by a pH
that participate in a redox reaction, and membrane elec- electrode. The pCO2 electrode is widely used in clinical
trodes. The membrane can be solid material (e.g., glass), laboratories as a component of instruments for measur-
liquid (e.g., ion-exchange electrodes), or special mem- ing serum electrolytes and blood gases.
brane (e.g., compound electrodes), such as gas-sensing In the NH3 gas electrode, the bicarbonate solution is
and enzyme electrodes. replaced by ammonium chloride solution, and the mem-
The standard hydrogen electrode is an example of an brane is permeable only to NH3 gas. As in the pCO2 elec-
inert-metal electrode. The Ag/AgCl electrode is an example trode, NH3 changes the pH of NH4Cl as follows:
of the second type. The electrode process AgCl e →
Ag Cl produces an electrical potential proportional to NH3 H2O ↔ NH4 OH (Eq. 5-8)
chloride ion (Cl) activity. When Cl is held constant, the The amount of OHproduced varies linearly with the log
electrode is used as a reference electrode. The electrode in of the partial pressure of NH3 in the sample.
contact with varying Cl concentrations is used as an in-
dicator electrode to measure Cl concentration.
The H-sensitive gel layer of the glass pH electrode is
considered a membrane. A change in the glass formu-
lation makes the membrane more sensitive to sodium
ions (Na) than to H, creating a sodium ISE. Other
solid-state membranes consist of either a single crystal or
fine crystals immobilized in an inert matrix such as si-
licone rubber. Conduction depends on a vacancy defect
mechanism, and the crystals are formulated to be se-
lective for a particular size, shape, and change—for
example, F-selective electrodes of LaF, Cl-sensitive
electrodes with AgCl crystals, and AgBr electrodes for the
detection of Br.
The calcium ISE is a liquid-membrane electrode. An
ion-selective carrier, such as dioctyphenyl phosphate
dissolved in an inert water-insoluble solvent, diffuses FIGURE 5-21. The pCO2 electrode.
14460_Ch05.qxd 12/4/08 8:06 PM Page 145
Detector
Capillary
Buffer Buffer
() ()
FIGURE 5-23. Densitometer—basic components.
Sample
FIGURE 5-25. Differential solute migration superimposed on electro-osmotic flow in capillary zone
electrophoresis. (Reprinted with permission from Heiger DN. High-performance capillary electrophoresis.
France: Hewlett-Packard, 1992.)
in any chromatographic technique are the mobile phase This results in the partitioning of the solute molecules
(gas or liquid), which carries the complex mixture (sam- into two separate phases.
ple); the stationary phase (solid or liquid), through which The ratio of the concentration of the solute in the two
the mobile phase flows; the column holding the station- liquids is known as the partition coefficient:
ary phase; and the separated components (eluate).
solute in stationary phase
K (Eq. 5-11)
solute in mobile phase
Modes of Separation
Modern partition chromatography uses pseudo liquid
Adsorption
stationary phases that are chemically bonded to the
Adsorption chromatography, also known as liquid-solid
support or high-molecular-weight polymers that are in-
chromatography, is based on the competition between
soluble in the mobile phase.15 Partition systems are con-
the sample and the mobile phase for adsorptive sites on
sidered normal phase when the mobile solvent is less
the solid stationary phase. There is an equilibrium of
polar than the stationary solvent and reverse phase when
solute molecules being adsorbed to the solid surface and
the mobile solvent is more polar.
desorbed and dissolved in the mobile phase. The mole-
Partition chromatography is applicable to any sub-
cules that are most the soluble in the mobile phase, move
stance that may be distributed between two liquid
fastest; the least soluble, move slowest. Thus, a mixture
phases. Because ionic compounds are generally soluble
is typically separated into classes according to polar
only in water, partition chromatography works best with
functional groups. The stationary phase can be either
nonionic compounds.
acidic polar (e.g., silica gel), basic polar (e.g., alumina),
or nonpolar (e.g., charcoal). The mobile phase can be a
single solvent or a mixture of two or more solvents, de- Steric Exclusion
pending on the analytes to be desorbed. Liquid-solid Steric exclusion, a variation of liquid-solid chromatogra-
chromatography is not widely used in clinical laborato- phy, is used to separate solute molecules on the basis of
ries because of technical problems with the preparation size and shape. The chromatographic column is packed
of a stationary phase that has homogeneous distribution with porous material, as shown in Figure 5-26. A sample
of absorption sites. containing different-sized molecules moves down the
column dissolved in the mobile solvent. Small molecules
Partition enter the pores in the packing and are momentarily
Partition chromatography is also referred to as liquid- trapped. Large molecules are excluded from the small
liquid chromatography. Separation of solute is based on pores and so move quickly between the particles.
relative solubility in an organic (nonpolar) solvent and Intermediate-sized molecules are partially restricted
an aqueous (polar) solvent. In its simplest form, partition from entering the pores and, therefore, move through the
(extraction) is performed in a separatory funnel. Mole- column at an intermediate rate that is between those of
cules containing polar and nonpolar groups in an aque- the large and small molecules.
ous solution are added to an immiscible organic solvent. Early methods used hydrophilic beads of cross-linked
After vigorous shaking, the two phases are allowed to dextran, polyacrylamide, or agarose, which formed a gel
separate. Polar molecules remain in the aqueous solvent; when soaked in water. This method was termed gel
nonpolar molecules are extracted in the organic solvent. filtration. A similar separation process using hydrophobic
14460_Ch05.qxd 12/4/08 8:06 PM Page 149
contains A and C, because the Rf values are the same. solvent systems have resulted in the technique of high-
This ratio is valid only for separations run under identi- performance thin-layer chromatography (HPTLC).16
cal conditions. Because Rf values may overlap for some Absorbance of each developed spot is measured using a
components, further identifying information is obtained densitometer, and the concentration is calculated by
by spraying different stains on the dried plate and com- comparison with a reference standard chromatographed
paring colors of the standards. under identical conditions.
TLC is most commonly used as a semiquantitative
screening test. Technique refinement has resulted in
High-Performance Liquid Chromatography
the development of semiautomated equipment and the
ability to quantitate separated compounds. For exam- Modern liquid chromatography uses pressure for fast sep-
ple, sample applicators apply precise amounts of arations, controlled temperature, in-line detectors, and
sample extracts in concise areas. Plates prepared with gradient elution techniques.17,18 Figure 5-29 illustrates the
uniform sorbent thickness, finer particles, and new basic components.
FIGURE 5-29. HPLC basic components. (Reprinted with permission from Bender GT. Chemical instrumentation: a labora-
tory manual based on clinical chemistry. Philadelphia, Pa.: WB Saunders, 1972.)
14460_Ch05.qxd 12/4/08 8:06 PM Page 151
FIGURE 5-30. Chromatograms. (A) Isocratic ion-exchange separation mobile phase contains
0.055 mol/L NaNO3. (B) Gradient elution mobile phase gradient from 0.01 to 0.1 mol/L NaNO3 at
2% per minute. (C) Gradient elution—5% per minute. (Reprinted with permission from Horváth C.
High performance liquid chromatography, advances and perspectives. New York, N.Y.: Academic
Press, 1980.)
(e.g., octane sulfonic acid) can result in better retention chromatography (GLC), with a nonvolatile liquid sta-
of negatively charged compounds onto the column. tionary phase. GLC is commonly used in clinical labora-
The late-eluting compounds may have long retention tories. Figure 5-31 illustrates the basic components of a
times, producing broad bands resulting in decreased GC system. The setup is similar to HPLC, except that the
sensitivity. In some cases, certain components of a sam- mobile phase is a gas and samples are partitioned be-
ple may have such a great affinity for the stationary tween a gaseous mobile phase and a liquid stationary
phase that they do not elute at all. Gradient elution is phase. The carrier gas can be nitrogen, helium, or argon.
an HPLC technique that can be used to overcome this The selection of a carrier gas is determined by the detec-
problem. The composition of the mobile phase is var- tor used in the instrument. The instrument can be oper-
ied to provide a continual increase in the solvent ated at a constant temperature or programmed to run at
strength of the mobile phase entering the column (Fig. different temperatures if a sample has components with
5-30B). The same gradient elution can be performed different volatilities. This is analogous to gradient elution
with a faster change in concentration of the mobile described for HPLC.
phase (Fig. 5-30C). The sample, which is injected through a septum,
must be injected as a gas or the temperature of the in-
jection port must be above the boiling point of the com-
Gas Chromatography
ponents so that they vaporize upon injection. Sample
Gas chromatography is used to separate mixtures of vapor is swept through the column partially as a gas
compounds that are volatile or can be made volatile.19 and partially dissolved in the liquid phase. Volatile
Gas chromatography may be gas-solid chromatogra- compounds that are present mainly in the gas phase
phy (GSC), with a solid stationary phase, or gas-liquid will have a low partition coefficient and will move
14460_Ch05.qxd 12/4/08 8:06 PM Page 153
Sample
inlet
Data
Vacuum pumps
FIGURE 5-33. The components of a mass spectrometer. In this case, the ionization source pic-
tured is electrospray ionization and the mass analyzer is a quadrupole.
proportional to the concentration of the ions is formed Sample Introduction and Ionization
and fed to the recorder.
Direct infusion is commonly used to interface a GC or
MASS SPECTROMETRY LC with an MS; however, the challenge of introducing a
liquid sample from an LC column into an MS was a sig-
Definitive identification of samples eluting from GC or nificant barrier until recent technological advances in
HPLC columns is possible when an MS is used as a de- ionization techniques.
tector.20 The coupled techniques, GC/MS and LC/MS,
have powerful analytic capabilities with widespread Electron Ionization
clinical applications. The sample in an MS is first The most common form of ionization used in GC/MS is
volatilized and then ionized to form charged molecular electron ionization (EI). This method requires a source
ions and fragments that are separated according to their of electrons in the form of a filament to which an elec-
mass-to-charge (m/z) ratio; the sample is then meas- tric potential is applied, typically at 70 electron volts.21
ured by a detector, which gives the intensity of the The molecules in the source are bombarded with high-
ion current for each species. These steps take place in energy electrons, resulting in the formation of charged
the four basic components that are standard in all molecular ions and fragments. Molecules break down
MSs: the sample inlet, ionization source, mass analyzer, into characteristic fragments according to their molecu-
and ion detector (Fig. 5-33). Ultimately, molecule lar structure (Fig. 5-35). The ions formed and their rel-
identification is based on the formation of characteris- ative proportions are reproducible and can be used for
tic fragments. Figure 5-34 illustrates the mass spec- qualitative identification of the compound. Since most
trum of 9-carboxytetrahydrocannabinol, a metabolite instruments use the same 70 eV potential, the fragmen-
of marijuana. tation of molecules on different days and different
8000
6000
473
4000
488
2000
297 355 298 417
208 231 265 289 327 445
0
m/z– – > 150 200 250 300 350 400 450 500
182 82
Electrons
303
105
182
CH3 O
N C OCH3
303
OC
82
O 105
FIGURE 5-35. Electron bombardment breaks cocaine into frag-
ments, with number and size quantified. Unlike the illustrative glass
tumbler, the result of mass fragmentation of cocaine or other chemi-
cal compounds is both predictable and reproducible, especially with
electron ionization.
instruments is remarkably similar, allowing the com- in addition to small molecules and has become the most
parison of unknown spectra to spectra in a published common ionization source for LC/MS.
reference library.21 ESI involves passing the LC effluent through a capil-
lary to which a voltage has been applied. The energy is
Atmospheric Pressure Ionization transferred to the solvent droplets, which become
Unlike EI in GC/MS, most LC/MS ionization techniques charged.21 Evaporation of the solvent through heat and
are conducted at atmospheric pressure. As such, the ion gas causes the droplets to decrease in size, which in-
source of this type of instrument is not included in the creases the charge density on the surface. Eventually,
high vacuum region of the instrument. Two types of the Coulombic repulsion of like charges lead to the
ionization for LC/MS will be discussed here: electrospray ejection of ions from the droplet22 (Fig. 5-36). The in-
ionization (ESI) and atmospheric pressure chemical dividually charged molecules are drawn into the MS for
ionization (APCI), while matrix-assisted laser desorption mass analysis.
(MALDI) and surface-enhanced laser desorption (SELDI) ESI is adept at forming singly charged small mole-
ionization will be discussed later in the section on pro- cules, but larger molecules can also be ionized using
teomics. ESI and APCI also differ from EI in that they are this method. Larger molecules such as proteins be-
“soft” ionization techniques that leave the molecular ion come multiply charged in ESI, and since MSs measure
largely intact in the source. Many LC/MS techniques em- the m/z, even these large molecules can be observed
ploy technologies after the source, in the mass analyzer, to in an instrument with a relatively small mass range22
fragment molecules and generate the “fingerprint” spectra (Fig. 5-37).
used in identification. However, ionization techniques
used in LC/MS produce fragments and therefore mass Atmospheric Pressure Chemical Ionization
spectra that are somewhat less reproducible between in- Another important ionization source is APCI, which is
struments than EI used in GC/MS. This may prove to limit similar to ESI in that the liquid from LC is introduced di-
the utility of reference library spectra produced on other rectly into the ionization source. However, the droplets
instruments. are not charged and the source contains a heated vapor-
izer to allow rapid desolvation of the drops.22 A high
Electrospray Ionization voltage is applied to a corona discharge needle, which
Thanks to its wide mass range and high sensitivity, ESI can emits a cloud of electrons to ionize compounds after they
be applied to a wide range of biological macromolecules are converted to the gas phase.
14460_Ch05.qxd 12/4/08 8:06 PM Page 156
Cone
Spray needle tip (counterelectrode)
Multiply
charged droplet
Solvent “Coulombic”
evaporation explosion
Analyte Multiply
charged droplet Analyte
molecule
ions
+ve –ve
Power supply
FIGURE 5-36. Diagram of electrospray ionization, the most common ionization source for LC/MS.
Mass Analyzer to pass through the analyzer to the detector. All other
ions are deflected into the rods. The rods can be scanned
The actual measuring of the m/z occurs when the gas from low to high mass to allow ions of increasing mass to
phase ions pass into the mass analyzer. Two types of form stable sinusoidal orbits and traverse the filtering sec-
mass analyzers will be discussed here: the quadrupole tor. This technique will generate a full scan mass spec-
and ion trap, while time-of-flight (TOF) will be dis- trum. Alternatively, specific masses can be selected to
cussed later in the proteomics section. Both quadru- monitor a few target analytes. This technique is called se-
pole and ion trap mass analyzers contain rods or lected ion monitoring (SIM) and it allows for a longer
plates that are charged with varying AC and DC volt- dwell time (time spent monitoring a single ion) and
ages to form electric fields. These electric fields manip- therefore higher sensitivity.21 A full scan provides more
ulate the charged molecules to sort them according to information than SIM since ions not specifically selected
their m/z. in SIM are not detected. Therefore, a full scan would be
preferable for general unknown screening while SIM
Quadrupole analysis is more suitable for target compound analysis.
A diagram of a quadrupole MS is shown in Figure 5-38.
The quadrupole is the most common mass analyzer in use Ion Trap
today. The electric field on the two sets of diagonally op- The ion trap can be thought of as a modified quadrupole.
posed rods allows only ions of a single selected m/z value A linear ion trap employs a stopping potential on the end
3+
4+ 3334 2+
5+ 2501 5001 1+
2001 10,001
0 m/z 10,000
5+ 4+ 3+ 2+ 1+
IONS
Ion of selected m/z
(detected)
Unselected ion
(not detected)
Source slit
FIGURE 5-38. Single quadrupole mass spectrometer.
electrodes to confine ions along the two-dimensional (Fig. 5-39). Generally, each quadrupole has a separate
axis of the quadrupoles. In a three-dimensional ion trap, function. 22 Following an appropriate ionization
the four rods, instead of being arranged parallel to each method, the first quadrupole (Q1) is used to scan across
other, form a three-dimensional sphere in which ions are a preset m/z range and select an ion of interest. The sec-
“trapped.” In all ion traps, after a period of accumulation, ond quadrupole (Q2) functions as a collision cell. In a
the electric field adjusts to selectively destabilize the process called collision-induced dissociation (CID), the
trapped ions, which are mass-selectively ejected from the ions are accelerated to high kinetic energy and allowed
cavity to the detector based on their m/z.21 The unique to collide with neutral gas molecules (usually nitrogen,
feature of ion trap MSs is that they trap and store ions helium, or argon) to fragment the ions. The single ion
generated over time, effectively concentrating the ions of passed through the first analyzer is called the precursor
interest and yielding a greater sensitivity. (or parent) ion while the ions formed during fragmen-
tation of the precursor ions are called product (or
MS/MS daughter) ions. The third quadrupole (Q3) serves to an-
Tandem MSs (GC/MS/MS and LC/MS/MS) can be used alyze the product ions generated in Q2. This last
for greater selectivity and lower detection limits. A quadrupole can be set to scan all of the product ions
common form of MS/MS is to link three quadrupoles in to produce a full product ion scan, or to selectively
series; such an instrument is referred to as a triple quad allow one or more of these product ions through to the
GC / MS / MS
Tandem-in-Space
Tandem-in-Time
Ionization
Mass analysis
Dissociation
Mass analysis
Detection
Q1 Q2 Q3
B. SIM
IAS/MS
C. Product Ion Scan
IAS/MS
D. SRM
CID gas
FIGURE 5-40. Scanning modes used in a triple quadrupole mass spectrometer. (A) Full scan MS de-
tects all ions. (B) Selected ion monitoring (SIM) detects ions of one selected m/z. (C) Product ion scans
select ions of one m/z in Q1 to pass on to Q2, the collision cell, where the ion is fragmented. All ion
fragments are allowed to pass through to the detector. (D) Selected reaction monitoring (SRM) is simi-
lar to the product ion scan, but only fragments of one selected m/z are allowed to pass on to the de-
tector. Both C and D are examples of tandem mass spectrometry (MS/MS).
detector in a process called selected reaction monitor- These electrons are attracted to the next dynode where
ing (SRM). Various scanning modes commonly used more secondary electrons are emitted due to the higher
in a triple quad are shown in Figure 5-40. In some potential of subsequent dynodes. A cascade of electrons
triple quad instruments, the third quadrupole can also is formed by the end of the chain of dynodes, resulting
function as a linear ion trap to add further sensitivity in overall signal amplification on the order of 1 million
to MS/MS. or greater.22
Signal out
Dry node (x106)
Ions in
determination.23 GC/MS systems are widely used for A “shotgun” approach is often used in the discovery of
measuring drugs of abuse in urine toxicology confirma- new biochemical markers. The proteins from samples
tions. Drugs and metabolites must be extracted from body (e.g., serum, urine, tissue extract) from normal individu-
fluids and typically reacted with derivatizing reagents to als are compared with those derived from patients with
form compounds that are more volatile for the GC the disease being studied. Techniques such as two-
process. Computerized libraries and matching algorithms dimensional electrophoresis can be used to separate pro-
are available within the instrument to compare mass spec- teins into individual spots or bands. Proteins that only
tral results of an unknown substance obtained from a appear in either the normal or diseased specimens are
sample to the reference library. further studied. Computer programs are available that
Increasingly, LC/MS (including LC/MS/MS) technol- digitally compare gels to determine spots or areas that
ogy is taking its place alongside GC/MS in clinical labo- are different. When candidate proteins have been found,
ratories. LC offers a number of advantages over GC. the spots can be isolated and subjected to sophisticated
Typically, LC requires less extensive extraction proce- MS analysis to identify the protein and possibly any post-
dures and derivatization is rarely used, saving time and translational modifications that may have occurred.
expense. In addition, polar and heat-labile compounds Using this approach, the researcher does not have any
fare better in LC.21 However, the chromatography itself preconceptions or biases as to what directions or partic-
in LC can be somewhat less robust than in GC, resulting ular proteins to look for.
in wider peaks, more variable retention times and poten-
tially requiring more frequent maintenance. Another Two-Dimensional Electrophoresis
disadvantage of LC/MS is the less reproducible mass
spectra, as mentioned earlier. This electrophoresis assay combines two different elec-
Besides its use in toxicology, LC/MS also has great trophoresis dimensions to separate proteins from com-
potential for measuring low-level and mixed-polarity plex matrices such as serum or tissue. In the first
analytes such as vitamin D, testosterone, and immuno- dimension, proteins are resolved according to their
suppressant drugs due to its superior sensitivity and isoelectric points (pIs), using immobilized pH gradi-
specificity over immunoassays. In addition, LC/MS has ents. Commercial gradients are available in a variety
the advantage of being able to detect multiple analytes of pH ranges. In the second dimension, proteins are
(such as a panel of drugs or a series of metabolites) in separated according to their relative size (molecular
one run. LC/MS is free from the antibody interferences weight), using sodium dodecyl sulfate-polyacrylamide
seen in immunoassays, although LC/MS has its own type gel electrophoresis (SDS-PAGE). A schematic of this
of interference in the form of ion suppression. This effect procedure is shown in Figure 5-42. Gels can be run
is seen when a co-eluting chemical in the sample pre- under denaturing or nondenaturing conditions (e.g.,
vents a compound of interest from being ionized, thereby for the maintenance of enzyme activity) and visualized
reducing or eliminating its signal. LC/MS also requires by a variety of techniques, including the use of col-
highly skilled operators and is not nearly as automated as orimetric dyes (e.g., Coomassie blue or silver stain),
immunoassay instruments. radiographic, fluorometric, or chemiluminescence of
appropriately labeled polypeptides. These latter tech-
INSTRUMENTATION FOR PROTEOMICS niques are considerably more sensitive than the colori-
metric dyes.
The next generation of biomarkers for human dis-
eases will be discovered using techniques found within
MALDI-TOF and SELDI-TOF Mass
the research fields of genomics and proteomics.
Spectrometry
Genomics use the known sequences of the entire
human genome for determining the role of genetics in Matrix-assisted laser desorption ionization time-of-
certain human diseases. Proteomics is the investiga- flight mass spectrometry (MALDI-TOF) is used for the
tion of the protein products encoded by these genes. analysis of biomolecules, such peptides and proteins.
Protein expression is equal to and, in many cases, more Protein samples, such as those isolated from a two-
important for disease detection than genomics because dimensional electrophoretogram, are mixed with an ap-
these products determine what is currently occurring propriate matrix solvent and spotted onto a stainless
within a cell, rather than the genes, which indicate steel plate. The solvent is dried and the plate is intro-
what a cell might be capable of performing. Moreover, duced into the vacuum system of the MALDI-TOF ana-
many (posttranslational) changes can occur to the lyzer. As shown in Figure 5-43, a laser pulse irradiates
protein, as influenced by other proteins and enzymes the sample causing desorption and ionization of both
that cannot be easily predicted by knowledge at the the matrix and sample. Because the monitored mass
genomic level. spectral range is high (
500 daltons), the ionization
14460_Ch05.qxd 12/4/08 8:06 PM Page 160
Time-of-flight
Laser mass spectrometry
Analyte
Matrix
FIGURE 5-42. Hypothetical example of a two-dimensional elec- FIGURE 5-43. Sample desorption process prior to MALDI-TOF
trophoretogram from a patient with a disease (panel 1) compared analysis. (Diagram courtesy of Stanford Research Systems,
with a normal subject (panel 2). The patient exhibits a protein (oval ) Sunnyvale, Ca.)
that is not expressed in the normal subject. This protein might be a
potential marker for this disease. (Gels courtesy of Kendrick
Laboratories, Madison, Wis.)
of the low-molecular-weight matrix can be readily where is osmotic coefficient, n is number of dissociable
distinguished from high-molecular-weight peptides and particles (ions) per molecule in the solution, and C is
proteins and do not interfere with the assay of the pro- concentration in moles per kilogram of solvent.
tein. Ions from the sample are focused into the mass The osmotic coefficient is an experimentally derived
spectrum. The time required for a mass to reach the de- factor to correct for the fact that some of the molecules,
tector is a nonlinear function of the mass, with larger even in a highly disociated compound, exist as molecules
ions requiring more time than smaller ions. The molec- rather than ions.
ular weight of the proteins acquired by mass spectrum The four physical properties of a solution that change
is used to determine the identity of the sample and is with variations in the number of dissolved particles in
helpful in determining posttranslational modifications the solvent are osmotic pressure, vapor pressure, boiling
that may have occurred. For very large proteins, sam- point, and freezing point. Osmometers measure osmo-
ples can be pretreated with trypsin, which cleaves pep- lality indirectly by measuring one of these colligative
tide bonds between lysine and arginine, to produce properties, which change proportionally with osmotic
lower-molecular-weight fragments that can then be pressure. Osmometers in clinical use measure either
measured. The detection limit of this assay is about freezing-point depression or vapor-pressure depression;
1015 to 1018 moles. A modification of MALDI-TOF results are expressed in milliosmolal per kilogram
MS is surface-enhanced laser desorption ionization (mOsm/kg) units.
time-of-flight (SELDI-TOF) MS, in which proteins are
directly captured on a chromatographic biochip with- Freezing-Point Osmometer
out the need of sample preparation. Figure 5-44 illus- Figure 5-45 illustrates the basic components of a freez-
trates the SELDI-TOF process. ing-point osmometer. The sample in a small tube is
lowered into a chamber with cold refrigerant circulat-
OSMOMETRY ing from a cooling unit. A thermistor is immersed in the
sample. To measure temperature, a wire is used to gen-
An osmometer is used to measure the concentration of
tly stir the sample until it is cooled to several degrees
solute particles in a solution. The mathematic defini-
below its freezing point. It is possible to cool water to
tion is
as low as 40°C and still have liquid water, provided
Osmolality n C (Eq. 5-13) no crystals or particulate matter are present. This is
14460_Ch05.qxd 12/4/08 8:06 PM Page 161
1 2 3 4 5 6 7 8
Sample
1 2 3 4 5 6 7 8
Wash
EAM 1 2 3 4 5 6 7 8
7
SELDI
8 Reader
(PBSII)
Intensity
Raw
spectra
Gray scale
(gel) view
A Low Mass/ Charge High Work station
FIGURE 5-44. (A, B) Overview of the SELDI-TOF process. (Diagram courtesy of Ciphergen Biosystems, Fremont, Ca.) (continues)
14460_Ch05.qxd 12/4/08 8:06 PM Page 162
N-Laser
4. Analyze in a Protein Chip Reader
1
The proteins that are retained on the 2 Detector
array are detected on the Protein Chip 3
Reader. 4
5 TOF-MS
6
B
FIGURE 5-44. (Continued)
14460_Ch05.qxd 12/4/08 8:06 PM Page 163
Membrane Biodetection
barrier product
** Signal
* *
Display
processing
* * Target
analytes
Biodetection Transducer
agent
The most commonly used POCT devices used at bed- The next generation of POCT devices use biosen-
side, in physician offices, and at home are the fingerstick sors.22 A biosensor couples a specific biodetector, such
blood glucose monitors. The first-generation devices use a as an enzyme, antibody, or nucleic acid probe, to a
photometric approach, whereby glucose produces hydro- transducer for the direct measurement of a target ana-
gen peroxide (H2O2) with glucose oxidase immobilized lyte without the need to separate it from the matrix24
onto test strips. The H2O2 is coupled to peroxidase to pro- (Fig. 5-47). The field has exploded in recent years with
duce a color whose intensity is measured as a function of the development of microsilicon chip fabrication
concentration and measured using reflectance photome- because biosensors can be miniaturized and made avail-
try. A schematic of this technique is shown in Figure 5-46. able at low costs. An array of biosensors can be
These strips are measured for glucose concentration with- produced onto a single silicon wafer to produce a mul-
out the need to wipe the blood off the strips. tipanel of results, such as an electrolyte profile.
The strip technology in a POCT platform can also be Commercial POCT devices use electrochemical (e.g.,
used to measure proteins and enzymes, such as cardiac microion-selective electrodes) and optical biosensors
markers. The separation of analytes from the matrix is for the measurement of glucose, electrolytes, and arte-
accomplished by paper chromatography, in which spe- rial blood gases. With the immobilization of antibodies
cific antibodies immobilized onto the chromatographic and specific DNA sequences, biosensor probes will soon
surface capture the target analyte as it passes through. be available for detection of hormones, drugs and drugs
For qualitative analysis, detection is made by visual of abuse, and hard-to-culture bacteria and viruses such
means. Reflectance meters similar to those used for glu- as Chlamydia, tuberculosis, or human immunodefi-
cose are also available for quantitative measurements. ciency virus.22
REFERENCES
1. Christian GD, et al. Instrumental analysis. 2nd ed. Boston, Mass.: 5. Holland JF, et al. Mass spectrometry on the chromatographic time
Allyn and Bacon, 1986. scale: realistic expectations. Anal Chem 1983;55:997A.
2. Willard HH, et al. Instrumental methods of analysis. Belmont, 6. Guilbault GG. Practical fluorescence, theory, methods and tech-
Calif.: Wadsworth, 1981. niques. New York, N.Y.: Marcel Dekker, 1973.
3. Coiner D. Basic concepts in laboratory instrumentation. Bethesda, 7. Kricka LJ. Chemiluminescent and bioluminescent techniques. Clin
Md.: ASMT Education and Research Fund, 1975–1979. Chem 1991;37:1472.
4. Ingle JD, Crouch SR. Spectrochemical analysis. Upper Saddle 8. Wild D. The immunoassay handbook. London, UK: Macmillan
River, N.J.: Prentice-Hall, 1988. Press, 1994.
14460_Ch05.qxd 12/4/08 8:06 PM Page 165
9. Svelto O, et al. Principles of lasers. 2nd ed. New York, N.Y.: 17. Bender GT. Chemical instrumentation: a laboratory manual
Plenum Press, 1982. based on clinical chemistry. Philadelphia, Pa.: WB Saunders,
10. Coulter hematology analyzer: multidimensional leukocyte differ- 1972.
ential analysis, vol. 11 (1). Miami, Fla.: Beckman Coulter, 1989. 18. Horváth C. High performance liquid chromatography, advances
11. Weast RC. CRC handbook of chemistry and physics. 61st ed. and perspectives. New York, N.Y.: Academic Press, 1980.
Cleveland, Ohio: CRC Press, 1981. 19. Constantin E, et al. Mass spectrometry. New York, N.Y.: Ellis
12. Burtis CA. Tietz textbook of clinical chemistry. 2nd ed. Horwood, 1990.
Philadelphia, Pa.: WB Saunders, 1993. 20. Kebarle E, Liang T. From ions in solution to ions in the gas phase.
13. Heiger DN. High-performance capillary electrophoresis: an intro- Anal Chem 1993;65:972A.
duction. 2nd ed. Waldbronn, Germany: Hewlett-Packard, 1992. 21. Karasek FW, Clement RE. Basic gas chromatography: mass spec-
14. Monnig CA, Kennedy RT. Capillary electrophoresis. Anal Chem trometry. New York, N.Y.: Elsevier, 1988.
1994;66:280R. 22. Levine, B. Principles of forensic toxicology. Washington, D.C.:
15. Parris NA. Instrumental liquid chromatography: a practical man- AACC Press, 2006.
ual on high performance liquid chromatographic methods. New 23. Siuzdak, G. The expanding role of mass spectrometry in biotech-
York, N.Y.: Elsevier, 1976. nology. San Diego: MCC Press, 2006.
16. Jurk H. Thin-layer chromatography. reagents and detection meth- 24. Rosen A. Biosensors: where do we go from here? MLO Med Lab
ods, vol. 1a. Weinheim, Germany: Verlagsgesellschaft, 1990. Obs 1995;27:24.