Академический Документы
Профессиональный Документы
Культура Документы
1.1 INTRODUCTION
This service guide is intended to provide the service technician with functional descriptions, troubleshooting
tips, replacement procedures, and part numbers for the ATAC 8000 chemistry analyzer. Circuit diagrams are
included only to aid in understanding functions. Component level repairs are not recommended. Electronic
repairs should be made by circuit card replacement. This service guide will not cover MAC computer
servicing nor will it provide operating instructions (see Operator’s Manual).
1.3.2 PRECAUTIONS
1) Do not eat, drink, smoke, or chew gum in a laboratory environment.
2) Use safety equipment in the proper manner.
3) Keep hands away from mouth and eyes. Also, keep pens, pencils, tools, etc. away from mouth.
1.3.3 DECONTAMINATION
Use a commercial disinfectant according to instructions on the label or prepare a solution of 9 parts water and
1 part common household chlorine bleach.
1) Clean the surface first with a detergent cleaning solution until all visible contamination is removed.
2) Wipe all surfaces with bleach solution and let stand for 10 minutes.
3) Wipe with water to remove chlorine residue.
1) CAUTION: The power switch does not remove all power from the analyzer. The refrigeration circuitry is
independent of the switch and is live so long as the analyzer is connected via power cord to a live
electrical outlet. The analyzer must be unplugged to remove all power.
2) CAUTION: The connectors are NOT keyed to prevent errors. They may or may not already be labeled
with connector numbers (CNxx). Several connectors can physically be connected incorrectly. This may
result in damage to the analyzer. Be certain of all connections BEFORE unplugging any cables.
3) CAUTION: Do not attempt to move the syringes, arm, reagent tray, serum tray, or washer by hand
against the force of the motor. This may damage the belt, the double pinion pulley, and/or the motor.
Remove power to the motors by turning off the analyzer.
NOTE: There is a way to remove power to all motors without turning off the analyzer. This is useful for
service personnel. One can purposely cause a RESET error. Do this by removing the serum tray and
performing a “Reset ATAC 8000”. (To return power to the motor, replace the serum tray and perform
another “Reset ATAC 8000”.)
4) CAUTION: The lamp gets extremely hot. Allow at least 30 seconds after removal of power before
touching the lamp. Never change the lamp with power applied.
5) CAUTION: Never touch the lamp with bare fingers. Oil from fingers will discolor the lamp, causing
premature failure.
6) CAUTION: Do not look directly at the lamp when it is on. The lamp is extremely bright and will cause
damage to vision.
This is a standard PC or “IBM” type keyboard. Earlier versions of the ATAC 8000 required the AT
type connector. Current ATAC 8000s use the PS/2 mini connector.
The ATAC 8000 Operational Program coordinates all operations of the ATAC 8000 analyzer. It is
stored on the hard drive and boots at start-up. Updated programs can be loaded via the floppy drive.
Disk 1 has the Installation Program and DOS functions. It allows for formatting the HD and loading
the Operational Program on the hard drive of the ATAC 8000 analyzer.
These programs are updated occasionally to improve performance. It is important to operate with the
most current version. This is printed on the floppy disks and can be viewed on the service monitor.
NOTE: The lamp is turned on by the program (INHL signal), during start-up. It does NOT come on
immediately upon turning on the power switch. It remains constantly on as long as the program is
running.
2.5.1 DILUTER(Fig.14)
The Diluter Assembly (P/N BT340) functions to draw and dispense reagent, sample, and wash water. It uses
two identical drive mechanisms with different sized syringes. The 2.5ml “Large” syringe is located on the
right. It is primarily used for moving ATAC Final Wash. However, it may be used for reagent/ sample handling
when total volume is greater than 475ul. The 500ul “small” syringe is located on the left. It is the primary
reagent/sample syringe. By virtue of its smaller volume, it provides a more precise measurement of sample
and reagent.
There is a syringe mount assembly with a rigid Teflon tube between syringes. This tube should NEVER be
bent, as it will be ruined. There is generally no reason to remove or replace this unless it has been bent.
There are two versions of syringes and syringe mounting hardware, the “threaded” type and the “unthreaded”
type. The threaded type is on older versions of BT340. The syringe glass barrel is connected with a luer fitting
at the top and the threaded metal plunger rod is connected with a knurled nut at the bottom. The newer
versions of BT340 use an unthreaded metal plunger rod and a clamp/thumbscrew at the bottom. Ignore the
syringe part numbers of Figure 14, they are incorrect. Part numbers are as follows:
Threaded type 2.5ml- P/N 230-001, threaded type 500ul- P/N 229-001
Unthreaded type 2.5ml -P/N 301-001, unthreaded type 500ul -P/N 300-001.
There is a retrofit kit P/N 662.0151A to change from threaded to unthreaded mounting.
There are 2-way pinch valves located on either side of the syringes. They are different. The one on the right,
near the large syringe, is normally closed (N.C., P/N 330.9065). It controls the inlet of ATAC Final Wash to the
syringes. The pinch valve on the left, near the small syringe, is normally open (N.O., P/N 330.9037). It
controls the inlet and outlet of reagent, sample, and ATAC Final Wash to and from the sample needle.
Each drive mechanism consists of a stepper motor, a 2-belted pulley system, carriage, and connector board
with home position sensor. There is a Fluid Distribution manifold on the right side. It connects to the ATAC
Final Wash Supply at the bottom-right port and the vacuum supply from the rear, out of view. The manifold
routes Final Wash to the syringes, Washer Module, and Needle Washer (through the Peristaltic Pump
Cartridge). It also routes vacuum to the Needle Washer. There are fittings for connection of the Heated Sample
Tube and vacuum waste tube. On the rear of the Diluter assembly, there is a Waste Manifold with Vacuum
switch. The manifold provides connection to the Vacuum Pump Assembly. The Vacuum Switch is a sensor to
provide a “low vacuum pressure” alarm. It is normally adjusted to give an alarm if the vacuum pressure is less
than -0.1bar (100 mbar). Normal pressure is -0.2 to –0.4 bar.
The Diluter assembly is controlled by PCB 8261. The actions of the syringes and pinch valves are coordinated
for various functions. When drawing wash from the Final Wash supply, the N.C. pinch valve opens, the N.O.
valve closes, and the large or small syringe move downward. When the syringe reaches the bottom of its stroke
the valve return to normal and the syringe drives upward to push the wash out of the needle. In a prime or
needle wash cycle, the peristaltic pump (located on the ISE panel) turns when the small syringe is moving
upward. When drawing reagent or sample into the needle, the valves remain in their normal conditions. The
reagent and sample never reach the syringes because the maximum volume is 500ul. That is just enough to
reach the end of the black portion of the Heated Sample Tube on the Arm.
NOTE: The Syringe Brace (P/N 02708) pictured in figure 14 is no longer used.
NOTE: There was another version (BT340A, Fig. 30) of Diluter assembly that is NOT used on the ATAC
8000. It is identified by a different motor and pulley system using only one belt per drive assembly.
NOTE: There is a shipping retainer bracket to prevent damage to the rubber feet of the motor. The hex
screws of the bracket must be removed for normal operation to allow quiet, vibration-free operation. The
screws should be saved and taped to the housing after removal, in case the unit must be returned for
service. NEVER ship a Vacuum Pump Assembly without first draining the tank and installing the retainer
screws.
There are two versions of this circuit card in the ATAC 8000.
8130C is the DATA SIDE CPU. It is identified by the absence of chips in IC sockets U6, U8, and U9 that
would normally make up the Serial I/O port. It has an ATAC 8000 version EPROM.
8130E is the DISPENSE SIDE CPU. It has a Serial I/O chip Z84C40 in IC socket U6. IC sockets U8 and U9
have a PCB 8277 jumper board between them. It has the same version EPROM as 8130C.
Version 8130D is NOT used in the ATAC 8000. It has an ATAC 6000 version EPROM.
The PCB 8200 controls the temperature of the reagent tray. Basically, the circuit acts as a pulse width
modulator, controlling the current to the eight refrigeration Peltier modules. The mosfet IRF 540 is switched
on and off by the output of op-amp LM358, which functions as an oscillator. The duty cycle varies linearly
with the voltage change on the thermistor NTC2. The resistance (and voltage drop) of the NTC2 decreases
with higher temperatures, providing feedback which causes the Peltier modules to cool the reagent tray.
Adjustment of TR1 controls the temperature setting. Adjust to 4-10 degrees Celsius. AC power is supplied to
this circuit through transformer T2, independent of the Main ON/OFF switch. Thus, refrigeration is provided
so long as the instrument is plugged into an AC outlet.
Previous to version D this board contained circuitry for heating and sensing reagents. Redesign of this part, in
conjunction with changes to PCB 8299, have made it merely a connecting board for the needle to the PCB
8299.
The PCB 8205 Back Plane provides the interconnection of the various circuit cards in the rack via two
separate data buses and a power bus. The data buses are divided into “DATA SIDE” and “DISPENSE SIDE”.
The DATA SIDE receives the cards 8285, 8130C, and 8264 (L, B). The DISPENSE SIDE receives the cards
8261, 8130E, and 8264(R, A, T, S).
This circuit board provides connection of the cuvette Peltier modules (CN51) and NTC1 (CN52) to the PCB
8240 power supply. A two color LED indicates heating (red) and cooling (green). Connector CN50 provides
connection of the cuvette wheel Hall sensor with the motor drive circuit on PCB 8264 (L, B). Pin 1 of CN 50
connects theNTC2 and trimmer pot TR1 with the PCB 8285 Multifunction card. This provides a display of
cuvette temperature on the MAC screen.
This circuit board functions as the RS232 serial interface between the barcode scanner and the AT 486. Power
(+5V) to the MAX232 chip and the scanner module is provided through connector CN46.
The test tube sensor circuit determines the presence of test tubes or sample cups on the serum tray. There are
two sensors; one for the inner (STAT & STD) positions and one for the outer (PATIENT) positions. The light
from LED D2 is reflected to phototransistor OPTO1 when a test tube is present. The signal on pin 2 of the op-
amp is compared to the reference voltage on pin 3. When a tube is present the output goes high (indicated by a
red LED). The signal is routed to the AT 486 by way of PCB 8296 and PCB 8261.
8231 Wash Water Sensor PCB (This PCB is part of assembly 662.0611)
This circuit functions similar to 8210 except that the light from the LED (D1) passes through translucent
tubing rather than being reflected. The tubing passes more light to the photo-sensor (PHD1) when filled with
wash water, less light when empty. The output signal is high when the tubing is empty. The signal is routed to
the AT 486 by way of the PCB 8296 and PCB 8285. Calibration for the reference voltage at pin 2 of the op-
amp is made by adjustment of TR1. The specification is +2V when the tubing is empty. There are external
indicators; red for absence, green for presence of liquid.
**This part has been replaced by PCB 8248 (assembly 662.0734 Photometer).
The basic function of the circuit is to measure intensity of light passing through the cuvette (sample). The
“white” light from the lamp is focused through the reading cuvette to a quartz prism, downward to a reflecting
mirror, through semi-reflecting mirrors, through interference filters of specific wavelengths, to the photo-
detector diodes. Each wavelength has is own photodiode and op-amp circuit. These are nearly identical, with
slight variations of feedback resistors to increase gain. The reverse current of the photodiode increases with
light intensity. This is amplified by the op-amp. The output of each op-amp is connected to one of the Y
inputs of both multiplexer chips. From here it is amplified again by another op-amp. The PAL 110 output
frequency increases proportionally to its input voltage. This signal exits the 8238 at CN 49, pins 1 and 3. It is
routed to a counter-timer on PCB 8285, which outputs a binary number (representing the light intensity) to the
data bus.
It should be noted that the reading is taken while the cuvette wheel is moving. In fact, when the cuvette wheel
stops it is always between cuvettes. This is when it performs a dark current “zero offset” of the amplifiers.
This part replaces PCB 8238 in newer instruments. Its function is the same. Changes were made because of
component availability. See ECN 029.
8240 Power Supply Lamp/Peltier PCB (part of assembly 662.0606 Power Supply, Complete)
T his circuit board receives 21Vac from transformer T1. This is converted by bridge rectifier D1 and capacitor
C2 to provide Vpp, unregulated +26Vdc. The chipU1 is a switching regulator that provides very stable +12
Vdc for the lamp. When the instrument is switched on and during the initial phases of program loading, the
OPTO 1 signal “INHL”(inhibit lamp) keeps the mosfet Q1 and the chip U1 disabled. Therefore the Lamp 12
V and the Vpp 26V are not present. There is also a circuit for controlling the temperature of the cuvette wheel.
The op-amp U4b compares the NTC voltage signal (pin 6) with the reference voltage set by trimmer pot TR1
(pin 5). The output (pin 7) controls OPTO 2, which varies the output of U3 (pin 7). This in turn varies the V
supply of U2 (pin 2) and consequently controls the voltage to the Peltier modules. Op-amp U4a is used as a
polarity detector to switch the direction of current to the Peltier modules through U2. The trimmer TR1 is
adjusted (see Adjustment Procedures) to set the temperature inside the cuvettes at 37 degrees Celsius. The
ATAC 8000 program cannot change temperature to 30-degree range. Therefore the circuit involving OPTO 1
(pins 7&8) is not used.
This circuit board receives 18Vac from transformer T1. Bridge rectifier D1 and capacitor C2 convert this to
DC. The switching regulators U2 and U3 provide +5V, +12V, and -12V.
*There is a circuit designed to sense low input voltage that would automatically switch to battery, interrupt the
AT 486, store data and shut off the analyzer. U1 is a constant current source to charge the battery. Zener diode
D3 is a threshold detector that sends the interrupt (OFFLINE) to the AT486. The 486 will return the signal
INH BATT after storing data. This shuts off all DC power to the analyzer. Problems with this circuit include
power failure due to bad battery and premature activation of the low power shut down function
This plug-in circuit card has two identical circuits for controlling and driving the two diluter motors.
Each circuit includes two L6202 chips (1. 5A, D-MOS Full Bridge Driver) and an L297 (Stepper Motor
Controller) for controlling the drivers. There are two drivers per motor because there are two phases. The six
center pins of the L6202 chips are soldered to common ground for heat sinking.
The resistor on pin 1 (SENSE) of the L6202 chip limits the current through the motor windings. The value of
this resistor differs depending on the type of motor. The Sanyo Denki type 103-546-5540 motor requires 1.5
ohm, ¼ watt. There were instruments, serial # 62560 through 72187, which used Vexta model PX244M-02A
motors in single-belt diluter mechanism (BT340A). The sense resistors for these motors were 0.68 ohm, ½
watt. The Vexta/single-belt version of diluter is discontinued and should be replaced by the Sanyo/two-belt
diluter assembly (BT340). It is important to replace the 0.68-ohm resistors with 1.5-ohm resistors when
replacing diluter assemblies. Also, the service technician should always check the value of these resistors on
spare parts and all instruments as the possibility exists that service replacement parts had 0.68-ohm resistors.
The 82C55 I/O chip controls signals to/from the data bus. Signals to the data bus include Vacuum Alarm,
Waste Alarm, Test Tube (external), Test Tube (internal), diluter reset flags (F1, F2) and Liquid Sense of needle
(PC4, PC5). Signals from the data bus are drive signals to the L297s and the ULN2003. The ULN 2003
drives the two pinch valves (S1, S2) of the diluter and the motor (S1A) of the peristaltic pump (needle
cleaner). The S2A output is not used. U2 (74HC138) is a decoder for addressing U1 (82C55).
There are two versions of this plug-in circuit card, part number 668.8264(R, A, T, S) and 668.8264A(L, B).
They are not interchangeable.
The RATS card controls the stepping motors for the Reagent tray, Aspirator (needle up/down), Tower (arm
left/right), and Serum tray. The motor drive circuit is the same as that of the PCB 8261. Here also, the SENSE
resistors on pin 1 of the L6202s must be matched to the motor. The needle up/down motor is a Sanyo 103-
546-5540 that requires a 1.5-ohm, ¼ watt. The larger motors (Sanyo 103-770-1640) of the Reagent, Serum
and Tower require 0.5 or 0.68 ohm, ½ watt. Signals from the motor home position sensors are routed through
U16.
The LB card uses the same circuitry but only two of the four circuits are used. These have SENSE resistors of
1.5-ohm, ¼ watt to drive the cuvette wheel motor (L) and the barcode reader motor (B). Early examples of
this card had components for all 4 driver circuits. Now these cards are manufactured without components in
two of the circuits.
This circuit board plugs in between PCB 8278 and PCB 8279. This was added to fix a problem of over-
voltage of the op-amps on 8279 when the electrodes are empty.
This circuit board is located inside the anodized aluminum housing at the top left of the ISE module. It
connects to PCB 8278. Its purpose is to monitor sample and baseline fluid flow. The presence of fluid, air
bubbles, or absence of fluid is detected by two yellow LEDs and two corresponding photo-detectors. The 552
Micro-controller on PCB 8278 monitors signals from the detectors to control the pump and pinch valve. Two
red LEDs on the front side indicate to the operator the absence (lit) or presence (out) of fluid in the sensor.
The white button activates the pump motor and pinch valve to draw from the ISE bowl. The black button
actuates the pump only, to draw baseline solution.
This circuit board controls all functions of the ISE except sampling, which is controlled by the 8130E
Dispense side CPU card. It incorporates a 552 Micro-controller, an EPROM, a RAM, an AD7714 (22-bit A/D
converter), and a TA8435 (stepper motor driver).
The EPROM contains a program for communication between the 552 chip and the RS 232 port on the 8130E.
The RAM is loaded with process programs via the ATAC 8000 Operational Program floppy disk.
The AD7714 chip is used in differential input mode. The conversion result is equivalent to the difference
between the REF+ analog input (+2.5V virtual ground reference) and the selected analog input AIN1-AIN6
(K, Na, CL, CO2, REF). The TA8435 chip drives the peristaltic pump motor and the pinch valve is driven by
theULN2003. +12V and –12V to the circuit is provided through CN76 via a 10-pin flat cable from the 8130E
CPU card (CN13). This cable also carries RS 232 communication. The jumpers JP1 & JP2 are in place. The
+12V is also used by U10 to provide the circuit with regulated +5V. The chip U7 (555 timer) and related
circuitry is no longer used.
This circuit board is mounted on the front of the ISE module inside of a plastic housing. It connects to the
PCB 8278 through PCB 8270. The name is deceiving as the op-amps are configured as unity followers, which
have a non-inverted gain of 1. They are used as isolation to pass the electrode signals (in mV) to the A/D
converter without loading the circuit. The virtual ground reference is held at +2.5V by the circuit of D1 and
R1. This offset reference allows for the reading of both positive and negative signals generated by the
electrodes.
This circuit card is plugged into the AT 486. It uses an 82C55 I/O chip to control data between the AT486 and
the two PCB 8130 CPU cards. Data flows from the AT 486 to the 8130E (Dispense CPU), then to the 8130C
(Data CPU), then back to the AT486. . U2 is an address decoder for warm-up reset and U5, which clocks out
various power supply inhibit signals. The signal OFFLINE goes through R1 to IRQ2. In revision E there is a
jumper JP4 that allows for connection to IRQ5. The OFFLINE signal should be disabled by removing R1 or
JP4.
This is a plug-in circuit card that has several functions. Like the PCB 8261, it has a motor driver circuit and an
82C55 I/O chip. In this case, the card drives the cuvette washer motor and controls the Water Alarm, washing
station Reset (home flag), S1 three-way pinch valve, and S2 serum tray LED signals. It also controls filter
selection of the photometer and U2 completes the analog/digital conversion of the photometer signal and Temp
Check (NTC2 cuvette temp readout).
This circuit board, mounted in the rear of the analyzer, provides connection of various signals to/from PCB
8261, 8264 (RATS), and 8285. Basically, it changes individual wires to flat ribbon cables.
This is a circuit board located inside of the Arm. It controls liquid sensing of the needle and the heating of the
sample tubing. In stand-by condition the op-amp U2 automatically zeroes with the base frequency of the
oscillator circuit U1. When the needle descends to contact the liquid, the change in capacitance of the circuit
causes a frequency variation that attenuates the DC voltage at D2 and D3. The unbalance produced by the
change on pin 5 of U2 generates a positive peak voltage at pin 1 (output) of U2. The green LED D8 is lit
during the length of the unbalance. The sensitivity is adjusted using trimmer pot P1. Just as the needle exits
the liquid, the sudden return to the base frequency generates a negative output at pin 1 of U2. Both signals are
sent to PCB 8261 through D7. **Earlier versions of this PCB had a resistor R9 in place of the 5.1V zener D7.
It is very important that the D7 is in place or a false negative signal will be sent to the 8261.
U3 is a constant current generator to power the heated sample tubing. Opto1 is the home position sensor of the
needle up/down motor.
Revision G of 8299 includes ESD protection circuitry that must be used in combination with 8201 rev. F.
2) Open the wash solution can drawer (front right of the instrument case). See FIG. 1 & FIG. 7.
3) Disconnect wash solution delivery tube female luer fitting from the male luer lock connector of the
wash solution can or the external wash solution delivery tube.
4) Remove the wash solution can.
5) Remove the drawer by slightly lifting to move past the stoppers.
6) Remove the dust covers from the cuvettes wheel and serum plate (FIG.8).
7) Loosen and remove four (4) screws that secure the instrument case, and then carefully slide out the
case. Place the case in a protected place and put the screws in a parts tray (FIG.7).
2) Disconnect blue waste tubing (rear panel) by pushing collar of the quick connect and pull tubing to
disconnect (FIG.2).
3) Remove three (3) screws from top edge of rear cover and put them in a parts tray. The cover will tilt
back. Then it can be lifted to remove.
CAUTION:
Failure to push the washer down into a cuvette at this point will cause severe damage in the next step. The
washer must be down to prevent the worm drive rod from being bent when the chassis is opened. Once this is
bent, it must be replaced.
4) Unscrew two (2) screws located at the front left-hand corner of the PVC base and gently swing open
the "L" shaped chassis from the base (FIG.9). On later ATAC 8000s there may be an additional screw
at the rear right corner near the hinge.
3.2.1 FDISK
This is only done when replacing the hard drive, or mini-computer bd., or if the Format procedure fails to
correct a problem.
1) Connect a VGA monitor and a keyboard to the analyzer.
2) With analyzer turned OFF, place ATAC 8000 Operational Program Disk 1 into the floppy drive of the
ATAC 8000.
3) Turn on the analyzer. It will boot to the Installation Program screen.
4) Press the F9 key. This will give you a DOS prompt. Type fdisk to begin the on-screen procedure.
5) Delete the primary partition and then create the primary partition. Enter yes to maximum size. When
asked for Volume label, use the exact word that is shown on-screen under the Volume column.
Typically, it is “FASTER” or very often, it is blank. If it is blank, don’t type any volume name.
6) When done it will ask if you want to reboot, say yes.
7) It will boot to the Installation Program screen again.
8) You must now perform FORMAT, followed by Start Installation. This may be done at this point by
pressing the appropriate function key and following on-screen instructions.
9) Exit and reboot with disk removed to start-up analyzer.
NOTE: After formatting the hard drive, certain data is lost. You MUST perform Home Position
Alignments, Liquid Volume Calibration, CFC procedure, and Zero Photometer in the SERVICE CHECK
function of the MAC computer. Do them in that order.
3.2.2 FORMAT
Formatting the hard drive of the analyzer may be done from either the MAC computer in SERVICE CHECK,
or from the analyzer using a keyboard and monitor.
NOTE: After formatting the hard drive, certain data is lost. You MUST perform Home Position
Alignments, Liquid Volume Calibration, CFC procedure, and Zero Photometer in the SERVICE CHECK
function of the MAC computer. Do them in that order.
From analyzer;
1) Connect a service monitor and a keyboard to the analyzer.
2) With analyzer turned OFF, place ATAC 8000 Operational Program Disk 1 into the floppy drive of the
ATAC 8000.
3) Turn on the analyzer. It will boot to the Installation Program screen.
4) Press appropriate function key to begin formatting. Enter yes to “are you sure?”. Screen will count
down percentage complete.
5) When complete, you must install the program by pressing the appropriate function key. Follow the
on-screen instructions.
6) Exit and reboot with disk removed to start-up analyzer.
NOTE: After formatting the hard drive, certain data is lost. You MUST perform Home Position
Alignments, Liquid Volume Calibration, CFC procedure, and Zero Photometer in the SERVICE CHECK
function of the MAC computer. Do them in that order.
From analyzer;
1) Connect a service monitor and a keyboard to the analyzer.
2) With analyzer turned OFF, place ATAC 8000 Operational Program Disk 1 into the floppy drive of the
ATAC 8000.
3) Turn on the analyzer. It will boot to the Installation Program screen. Press appropriate function key to
begin installation. Follow on-screen instructions.
4) Remove Op Disk 1 and re-boot the analyzer.
I/O port configurations (serial ports Com1/2F8, Com2/3F8) - directs the ATAC 8000/MAC
interface data to RS232 port on back of analyzer and the barcode reader data to the internal serial
port.
Boot sequence (A;C) or (A:,C:) - causes analyzer to boot from floppy disk if present in A: drive.
Floppy seek at boot (enabled) -causes analyzer to boot from floppy disk if present in A: drive.
Keyboard (absent) or (not installed) – allows analyzer to run without the service keyboard attached.
Drive A: (1.44MB, 3.5”) – identifies the type of floppy drive that is present
NOTE: The ATAC 8000 will not communicate with the MAC computer while in
DIAGNOSTICS. You may leave the MAC running but do not use the MAC keyboard, it will
result in “Timeout Error” and possibly lock up the MAC.
Select the desired function according to the menu displayed on the service monitor.
You must select filters by pressing the Alt key and letter F key simultaneously, before any other
functions will work. Typically, use 340nm and 700nm.
Prior to this function, an F10 mABS reading must be done. Normally the F10 would be done with
water (which has zero absorbence) in all cuvettes. Press the F7 key. The previous ABS values will be
zeroed. To view, press F10 again.
Zeroing would be required if one desired to use the F10 function to read true aborbence.
One use for this would be to check the stability of the photometer. Use the Alt+F function to choose
filters. Use the Alt+H function to fill all cuvettes with water. Then press F10. Ignore the readout and
press F7 to zero. Then press F10 again. The readout should indicate 0-5 for each cuvette. Continue
pressing F10 for several minutes and observe that the readings remain 0-5. Repeat the process using
two different filters until the stability of all filters has been checked.
Jumping results on one or more filters might indicate an unstable light source, an unstable
photometer, or a change in the absorbence of the liquid in the cuvettes.
This function, in conjunction with Alt+F, Alt+S, and F7, may be used to do manual tests to determine
absorbences.
NOTE: Observe the two-color led on PCB8206. Red indicates 8240 is trying to heat, green indicates
trying to cool, LED out indicates temperature is at the set level or the 8240 has no output. Under normal
operation (after at least 15 minutes of power on) the LED should slowly cycle between dimly red and out.
NOTE: This is an adjustment to the displayed temperature on the SERVICE CHECK screen. IT DOES
NOT ALTER THE ACTUAL TEMPERATURE OF THE CUVETTES.
NOTE: This adjustment requires at least an hour to stabilize, make several readings over the course of the
day. Be wary of readings that remain at 2 for long periods as a sudden drop in room temperature may
cause condensation in the chamber to freeze.
a) Principle of operation
The photoelectric sensor has been designed to detect the presence or absence of liquid through the translucent
Final Wash tubing. The sensor uses an LED as the light source and photodiode as the sensing element. This
minimizes the effect of operational environment and outside light on the operation of the sensor and thereby
improves reliability. The translucent plastic tubing conveniently fits into the appropriate channel on the sensor
without restricting flow. The LED light beam refracts through the tubing and the photodiode senses the
presence or absence of liquid by variations in transparency. The red/green LEDs on the front of sensor provide
the user with a visual indication of the liquid status. The presence of liquid is signaled by a green LED
indicator and the absence by a red LED. The sensor communicates via digital output the liquid status alarms to
the computer. The sensitivity adjustment trimmer pot, located on the rear of the sensor, permits detection
threshold optimization of the translucent tubing. This sensitivity adjustment is factory set, however a
readjustment may be required in case of tube replacement. Experience has shown that a tube other than Elan’s
is generally not recommended, as adjustment may not be possible.
NOTE:
A defective sensor that causes continuous or intermittent “Wash Bottle Empty” errors can be “tricked” into
displaying a green LED permanently by pulling the tubing out of the sensor. This must only be used as a
temporary fix and the OPERATOR must ensure that the bottle has adequate wash at all times. The instrument
will not be able to detect air bubbles or an empty bottle, causing erroneous results.
3.6.1 LOW PRESSURE ALARM (Fig. 13, 14, 26, 30, 45)
The sensor for this alarm is the Vacuum Switch, located on the back of the Diluter Panel. This does not adjust
the pressure; it adjusts the alarm threshold. The pressure can be adjusted by the Vacuum Limit valve and the
Pressure Relief valve; both located on the Waste Lid Manifold.
1) With the system operating, observe the pressure reading on the gauge. It should normally be between –0.2
and –0.4 bar.
2) Lower the pressure reading very slowly by gently pressing the ball on the Vacuum Limit valve. Observe
the pressure fall and note the reading when the ATAC 8000 gives a “Low Pressure Error”. It should alarm
at –0.1 bar.
3) Adjust the Vacuum Switch for proper alarm setting.
4.1 REPLACEMENTS
NOTE: The latest version of power supply (662.0606P) has plugs internal to the case to facilitate
replacement in the latest ATAC 8000s. In analyzers with serial numbers beginning with 92xxx, open the
power supply case. If the circuit boards are “plug in”, then replace them. Otherwise, proceed below.
2) Remove top cover, deck cover, and open the chassis. (See 3-1.1, 3-1.2, 3-1.4)
CAUTION:
Failure to push the washer down into a cuvette at this point will cause severe damage. The washer must be
down to prevent the worm drive rod from being bent when the chassis is opened. Once this is bent, it must be
replaced.
3) Trace all cables and familiarize yourself with the connections. Label them with tape if needed.
CAUTION:
The connectors are NOT keyed to prevent errors. Several connectors can physically be connected
incorrectly. This may result in damage to the analyzer. Be certain of all connections BEFORE
unplugging any cables.
CAUTION:
Failure to push the washer down into a cuvette at this point will cause severe damage. The washer must be
down to prevent the worm drive rod from being bent when the chassis is closed. Once this is bent, it must be
replaced.
NOTE: The cuvette temperature MUST BE ADJUSTED after replacement of the power supply because
the control is part of PCB 8240. Perform 3-4.1
6) With the plunger pushed inside the glass barrel, screw on the new syringe. Be careful not to cross-
thread the luer fitting.
7) Pull the plunger down and into the plunger clamp.
8) On the old style clamp, tighten the thumbnut. No adjustment is necessary.
9) On the new style clamp, push the clamp and plunger upward as far as possible.
10) To adjust for 1mm clearance, pull the plunger down while holding the clamp in the up-most position.
Tighten the thumbscrew.
NOTE: This adjustment ensures that the diluter motor can reach its HOME position before the plunger
hits the top, but minimizes the dead space.
9) With the cuvette wheel in place, wet each cuvette with reagent grade water and push it down into the
empty hole. When all are in place, remove the cuvette wheel again and dry the cuvettes with a lint-
free tissue.
10) Replace the cuvette wheel once more. Insert the 4 hex screws and lightly tighten until resistance is
just felt. Now tighten in a crosswise pattern until all are snug.
CAUTION:
Failure to push the washer down into a cuvette at this point will cause severe damage. The washer must be
down to prevent the worm drive rod from being bent when the chassis is opened. Once this is bent, it must be
replaced.
2) Unplug cable from connector CN51 on board PCB 8206 (located on the cuvettes system base plate,
Fig. 3). Connect ohmmeter to Peltier modules connector CN51 on PCB 8206 and verify impedance of
20-40 ohms. Unplug NTC1 connector CN52 on PCB 8206. Check NTC1 at CN52 of PCB 8206 for
5K ohms @ 20C.
3) Determine from readings whether to proceed.
4) Proceed with cuvette/washer/photometer assembly as follows.
5) Unplug CN60 (20-pin flat cable) and CN63 (2-pin cable to pinch valve) from the PCB 8158 located
on the rear of the washer module (FIG.3).
6) Push upwards the washer body and remove the four (4) hex screws and washers, which secure the
washer to the cuvettes system housing (FIG.3).
7) Make sure the cuvettes are empty of liquid. If not, use a pipette to remove all liquid from the cuvettes.
8) Unscrew the four (4) hex screws (located on the top of base plate, near the PCB 8206) and remove the
photometer from the underside of base plate (FIG.3 & 43). Leave it out of the way, there is no need to
unplug it from PCB 8285.
9) Unscrew four 10mm nuts and washers that secure the cuvette system base plate to the vibration
mounts (FIG.3).
10) Carefully remove the cuvettes system base plate, turn upside down, and place it gently on table or flat
surface.
11) Remove two (2) screws and washers securing two (2) clamps to hold together cooling fan supporting
bracket, heat sink, and Peltier modules on the underside of cuvettes system housing (FIG.3).
12) Carefully remove clamps, supporting bracket, and heat sink from the bottom of the cuvettes system
housing exposing the Peltier modules.
13) Unscrew four (4) screws (located at the bottom of base plate) and washers holding together cuvettes
system housing and the base plate (FIG.3).
14) Gently remove the base plate to bring into view Peltier modules assembly.
15) Measure each Peltier module's impedance across terminal leads to locate the damaged Peltier.
16) Unsolder the damaged Peltier leads.
17) Carefully remove the damaged Peltier. The removed Peltier is useless. Discard it.
CAUTION:
It is imperative to install the Peltier exactly as it was oriented. Observe the red and black leads, as shown in
Fig.3. Be absolutely sure that all Peltier modules are oriented as shown. OTHERWISE, IT WILL NOT WORK
CORRECTLY!
18) Thoroughly clean the Peltier modules, heat sink, and cuvettes housing bottom surfaces to ensure that
all burrs, grease, and dirt etc. have been removed as presence of such material will degrade thermal
performance.
19) Replace with new Peltier. Make sure that the new Peltier is of the same type: P/N 330.4601.
20) Coat the "cold" side of the Peltier modules as well as the corresponding locations on the cuvette
housing with a thin layer of conductive thermal grease and place the modules cold side down on the
housing. Gently push down on each module and apply a back and forth turning motion to squeeze out
excess thermal grease.
NOTE: Suggested thermal greases include: "Wakefield Engineering Type 120", "Dow Corning
Type 340", "General Electric Type G641", and "American Oil And Supply Type 300".
21) Coat the hot side of the modules as well as the appropriate heat sink surfaces with a thin layer of
thermal grease.
NOTE: Before fastening the base plate onto the housing, carefully run the Peltier modules’ leads
through the base plate grooves on both sides. Exercise extreme care to prevent damage to the parts.
Peltier leads are extremely fragile where they connect to the module.
CAUTION:
Failure to push the washer down into a cuvette at this point will cause severe damage. The washer must be
down to prevent the worm drive rod from being bent when the chassis is closed. Once this is bent, it must be
replaced.
32) Connect all the cable assemblies to their appropriate connectors on PCB8206, PCB 8158 (on the rear
of washer module), and PCB 8285 (photometer cable).
33) Connect the cuvettes drive motor, cooling fan, and lamp cables to their appropriate cable connectors.
Be sure connectors are firmly attached.
34) Gently swing down the "L" shaped chassis and lock into position with two (2) M4x6 screws.
35) Attach the ends of silicone tubing to both fluid inlet (bottom prong, rear slot of 3-way pinch valve)
and vent (upper prong, front slot of 3-way pinch valve) prongs on the piston sleeve in the washer
module (FIG.15).
36) Attach the end of washer suction/discharge tube (other end connected to the plastic elbow on the
diluter panel) to the uppermost duct on the washer-module (FIG.15).
37) Connect the heated sample tube to the diluter panel, and then attach the needle washer I/O tubing to
the appropriate ducts on the white manifold (FIG.15).
38) Slide the deck back into position in the instrument.
39) Reconnect the ISE bowl tubing
40) Replace serum plate.
41) Set the sampler arm to the central position by gently rotating counterclockwise.
42) Clean the dust covers carefully and replace them.
43) Carefully slide the case onto the analyzer ensuring that the wash solution delivery tube is properly
routed through the appropriate slot in the right side of the case, and then insert and tighten the case
screws.
44) Slide the wash solution drawer into its compartment.
45) Put the wash solution can back into the drawer, and reconnect the wash solution delivery tube to the
wash solution can or the external wash solution delivery tube.
CAUTION:
Failure to push the washer down into a cuvette at this point will cause severe damage. The washer must be
down to prevent the worm drive rod from being bent when the chassis is opened. Once this is bent, it must be
replaced.
3) Unscrew the four 2mm hex screws (located on the top of base plate, near the PCB 8206) and remove
the photometer from the underside of base plate (FIG.3 & 43). Unplug it from PCB 8285.
4) Replace the photometer. Hold it up in place and get all four hex screws started before tightening them
completely.
5) Plug the cable into CN 4 of PCB 8285.
6) Re-assemble covers and prepare analyzer for normal operation.
7) Photometer replacement requires that a new Cuvette Factor Calculation (CFC) be run. Do this from
the SERVICE CHECK Function menu.
NOTE: Before proceeding, measure reagent Peltier modules at CN91 on PCB 8200 (located behind rear
cover) to determine if service is needed. Disconnect CN91 and measure between the black and red wires
of the cable for about 20ohms. Very high or infinite resistance indicates one or more faulty Peltier
modules (330.4605), causing complete lack of cooling. Also disconnect CN90 and measure the cable
between the blue wires for a reading of 5K ohms @ 20C. A lower reading would indicate a faulty
Reagent NTC (662.0663), probably causing a freezing problem. The NTC can be replaced in step 8.
The service person should have checked and ruled out the refrigeration fuses and PCB 8200. At this
point, you should already know that a Peltier module is bad.
CAUTION:
Be sure to identify all leads for proper reconnection. Make ID marks if necessary.
5) Unsolder the refrigerator module black sheathed cable (red and black leads) from the PCB 8124
located between the two large cooling fans on the right side of the PVC base. See Figure 9.
6) Unsolder "reagents NTC Sensor assembly" (PART NO. 662.0663) blue leads from the PCB 8124.
8) Loosen and remove the sixteen screws that attach the insulated round reagent chamber (FIG.10 & 33)
to the PVC base, reagent cylindrical coolers and NTC sensor. Remove the chamber to expose the
belt/pulley drive system and conductive spacers. Screws used: 15 screws-M4x8, 1 screw-M3x8 for
NTC sensor.
9) Carefully lift the front of the instruments to stand the instrument on its back (protected by pads) in
slightly inclined position so that the base frame is accessible. Ensure that protruding parts on the back
panel do not touch the surface of the table. See figure 11.
10) Remove the fourteen (14) screws that attach the metallic base frame to the PVC base. Before
removing the base frame, disconnect the yellow-green ground wire from the grounding terminal on
the frame and set frame aside. Screws used: 14 screws-M6x12. Refer to Figure 11.
11) Remove the four (4) "reagent refrigeration module" mounting screws (located on the top right of PVC
base) which secure the module to the underside of PVC base. See FIG.11.
12) While holding both sides of the heat sink of refrigeration module, carefully pull forward the entire
module along with wires & cables from the PVC base. Exercise extreme care to prevent damage to
parts. Refer to Figure 11.
13) If the refrigeration module is difficult to remove, then get a large screwdriver or similar object and
slightly insert between the module heat sink and PVC base. Then apply lever action evenly on all
sides, just enough to ease it out of the PVC base. Afterwards remove the entire module from the PVC
base and place it gently on table or flat surface. Remove the black foam insulation pad.
14) Do not remove individual Peltier modules yet. Inspect the Peltier modules for green corrosion and
broken leads. Cut the leads and measure any suspected modules with an ohmmeter. A good Peltier
module will measure less than 20 ohms. If no module looks suspicious, you must isolate the bad
part(s) one at a time or you can measure 4 at once, then 2.
15) Once the bad part(s) is isolated, remove the 4 screws that hold the conductive spacer. See Figure 12.
16) Remove the bad part and discard it. Clean the surface of the heat sink and conductive spacer.
17) Refer to Figure 12 to ensure the proper placement of the leads.
NOTE: The red lead of one connects to the black lead of the next and the leads are never twisted. This is
imperative to ensure that the “HOT” side of the Peltier module is down on the heat sink and the “COLD”
side is up on the conductive spacer.
18) Coat both sides with heat sink compound. Suggested types include Dow Corning type 340, GE type
G641, and American Oil and Supply type300.
CAUTION:
The re-assembly of the system parts is not the same as disassembly, in reverse order. It is important to read the
steps carefully.
24) Hold and turn the refrigeration module heat sink, with foam insulation, as necessary to mate the
Peltier/conductive spacers with the corresponding holes at the rear of the PVC base.
25) Gently push forward with uniform light pressure until conductive spacers are inserted all the way
through the holes (FIG.11).
26) Secure the refrigeration module into the PVC base by screwing four (4) mounting screws from the
front of the PVC base into the heat sink but leave them slightly loose (FIG.11).
CAUTION:
These MUST be left slightly loose. Tightening these now will push the conductive spacers higher than the
PVC base, causing the reagent trays to rub on the chamber.
27) Position the metallic base frame near the bottom of PVC base and connect the grounding lead
(yellow-green) to the base frame. See Figure 11.
28) Reinstall the base frame into the PVC base using same 14 screws-M6x12. See Figure 11.
29) Gently lay the instrument back down on a flat surface of the table. See Figure 10.
30) Coat the top surfaces of the conductive spacers (exposed through holes on the PVC base) with an
even thin layer of heat sink compound. See Figure 10.
31) Put the NTC sensor (yellow and green-ground leads facing outwards) in its proper place (small hole)
on the bottom of the reagent chamber and retain loosely with M3x8 screw from the top.
32) Fit the reagent chamber into position over the conductive spacers and PVC base holes ensuring that
the protruding NTC sensor with leads is properly seated inside the appropriate slot in the PVC base.
Firmly secure the chamber onto the PVC base by using seven (7) M4x8 screws (inner circle of holes).
See Figure 10.
33) Fully screw eight (8) M4x8 screws into the outer circle of holes (coinciding with Peltier conductive
spacers) so that the conductive spacers make uniform contact with the bottom of chamber. There
should be no space between the chamber and the PVC base.
34) Slightly tighten the four (4) mounting screws (without forcing) securing the refrigeration module. Do
not over-tighten.
35) Re-solder the refrigeration module terminal leads (red and black) and the NTC sensor leads (blue) to
the correct terminal pins of the PCB 8124.
36) At this point, the analyzer can be plugged in to verify operation of the refrigeration. Check the main
power switch on the rear panel and ensure that it is in the "off" position prior to attaching the power
cord. Cooling will begin immediately upon plugging the power in. Feel the reagent chamber to verify
cooling. Unplug the analyzer and continue assembling the analyzer.
37) Replace five (5) reagent trays into their compartments and ensure proper engagement with the detent
balls.
38) Replace deck cover, top cover, and rear cover. Reconnect all tubing and cables. Prepare analyzer for
normal operation.
CAUTION:
In the N.C. (normally closed) “rear” slot of the 3- way pinch valve is the wash water supply (Fig.15-
#19) to the Wash Station. This tube is normally primed with Final Wash. It is imperative to
disconnect the Final Wash Bottle connection before removing this tube from the pinch valve.
Otherwise, gravity will cause it to drain the contents of the Final Wash Bottle into the top of the
cuvette wheel, flooding the photometer. BE CERTAIN TO RECONNECT THE WASH WATER
TUBING INTO TO THE REAR SLOT (N.C.) OF THE 3-WAY PINCH VALVE.
NOTE:
Remember to record this procedure in the “Services Performed” function of SERVICE
CHECK on the MAC. This will reset the cycle count.
1) Remove both ends of the peristaltic pump tubing from the diluter manifold (1).
2) Squeeze the locking catches (3) and remove the defective cartridge (2).
3) Place the new cartridge on the drive shaft (4) and gently press to snap-fit.
4) Carefully attach the pump tubing to both ducts in the diluter manifold. The pump is now ready for
operation.
NOTE: The service person can open the cartridge to inspect the tubing. Sometimes after long idle
periods, the tubing sticks together under the pressure of the rollers. The service person may decide to fix
the tubing rather than replace the cartridge. It is suggested that the service person practice opening a
cartridge, removing the tube, unsticking it and re-assembling the cartridge.
1) The analyzer and pump should be off. If not, make sure the Drain-Suction lever is in the suction position.
2) Remove the waste drain tube by pressing the collar of the quick-connect fitting and pulling out the tube.
3) Original equipment comes with a check valve that is larger than the replacement (330.9080). The smaller
one is easier to replace.
4) If it is a large valve there will be clear tubing attached directly to the valve. If it is a small valve it will
have black (or blue) tubing attached to the valve body.
5) If it is a small one, simply replace it by pulling off the old and pushing on the new in the proper direction.
6) If it is original, remove the short piece of black(or blue) 5.5mm x 8mm tubing from the tank side and
save.
7) Cut the clear 7mm x 10mm tubing off the sink drain side. Discard the old valve.
8) Cut the saved piece of black tubing in half with strong wire cutters. Make a clean 90 degree cut.
9) Place a piece on each end of the new valve.
10) Insert the pointed end black tube into the long clear drain tubing. Make sure of a tight connection or waste
will spray all over when the pump is drained.
11) Insert the flat end black tube fully into the Drain quick-connect fitting on the pump lid. Make sure of a
tight connection or waste will spray all over when the pump is drained.
12) Test by draining the tank. Observe proper check valve operation.
1) With power off, disconnect and remove the Waste Lid Manifold.
2) Remove the ball knob/lever by unscrewing it.
3) Remove the (4) M6x50mm hex screws.
4) Clean the valve cone, valve body, and wave spring.
5) Cover the contacting surfaces with a thin coating of Dow high-vacuum grease.
6) Clean and then coat all O-rings, including the large one on the bottom of the lid.
7) Fit the valve cone and body together, making sure that the hole for the lever is accessible.
8) Put the wave spring in place and carefully align the valve body so that the (4) screw holes align with the
(4) threaded holes in the lid. It only fits in one direction.
CAUTION:
Permanent damage to the lid will result from trying to screw into the unthreaded ports (the ones with the O-
rings) in the lid.
9) Start all screws, making sure that they are not cross-threaded.
10) Tighten them evenly in a crossing pattern until they compress the spring and just come tight against the o-
rings.
11) Reconnect the lid to the tank.
12) Verify that the tank will create and hold vacuum between -0.2 bar and –0.4 bar.
13) Verify that the tank will drain. (see Section 3-6)
6.1 BASICS
The best troubleshooting tools are a thorough knowledge of normal operation and the ability to quickly find
information in the appropriate manual.
The machine is plugged in but turned OFF. What sound do you hear?
Fans running on the right side?
What lights come on immediately when you flip the switch to On?
What is on the service screen? Did you hear a chirp as the RAM was checked?
Listen near the floppy drive. What kind of sound? Program booting from HD?
When did the Vacuum Pump turn on? Do you hear the hiss of the vacuum at the needle
washer?
That is the start-up cycle. Take time to observe other cycles such as a Zero Photometer, a CFC, the ISE
module, test calibration, control run, etc.
1. Identify the most obvious problem. Typically, it is the point where the operator noticed a problem. Do not
assume that the operator’s description of the problem is correct. Often it is over-simplified. For instance,
“My instrument is DEAD” rarely indicates an open fuse and the service person would probably observe
that there are indeed fans running and LEDs lit on the ISE panel.
2. Perform tests or functions to provide more information. Review results. What worked and what did not?
What are the differences between functions that work and those that did not? What do they have in
common?
For example:
Do end-point tests run well, while kinetic tests fail?
-End-point tests are calibrated; kinetic tests are not.
-Kinetic tests are very dependent on precise cuvette temperature; end-points are less so.
-Both require precise sample handling and cuvette washing.
3. Narrow the search to a block. Is the problem with power supply, program loading, motion of a particular
assembly, sample handling, ISE module only, temperature control, etc? Perhaps it is not an instrument
problem, but a reagent problem, MAC computer problem, or operator error.
4. Isolate the faulty component in the block and replace, repair, or resolve the problem.
5. Test thoroughly to confirm the resolution of the problem and assure normal operation of the analyzer.
1 beep heard during initialization. If last line displayed on service screen is “Mode Select
Function Code Complete” or “Error SIO Z80”_one of
CPU cards PCB 8130C or PCB8130E may be faulty or
PCB8284.
Fans running, refrigeration cold. At turn-on, LEDs on, “Error 100 or error101” on service monitor?
1 beep heard, vacuum pump on, lamp lights. Motors Program lost.
move. Zero photometer begins but movement stops. FDISK and/or Format HD. Reload software.
Lamp may or may not go out. No Scroll bar on MAC. Replace HD.
Replace mini-computer bd.
Fans running, refrigeration cold. At turn-on, LEDs on, Connect service keyboard and monitor. Press F5, does
1 beep heard, vacuum pump on, lamp lights. Motors analyzer reset? Can you do other functions on service
move. Zero photometer begins and appears to keyboard?
complete. No Scroll bar on MAC. Yes- Problem with RS232 communication between
ATAC 8000 and MAC.
Most likely-Check cable, LightningPCI, MP91, or
MAC modem port for problem.
Less likely- problem could be serial port on mini-
computer board
No- Problem similar to Error 101 problem above.
Fans running, refrigeration cold. At turn-on, LEDs on, Connect service monitor/kybd. Error is likely displayed
1 beep heard, vacuum pump on, lamp lights. Motors such as “Waste Full” or “Low Pressure”.
move. Constant error beeping heard. Possibly double error. Rectify error, press F5 to reset,
No Scroll bar on MAC. Reset attempt gives Timeout then try reset from MAC.
error.
7.5 Printer
1) There are currently 3 printers that will operate with the ATAC DMS software; Apple Imagewriter II,
Apple LaserWriter, and Epson Stylus Color 740.
2) The Epson printer driver software has been changed several times. The service person should verify the
latest version is installed when troubleshooting printer lock-up problems.
3) NOTES:
1-1 INTRODUCTION.......................................................................................................................1-1
3-5 Calibration Procedure for Wash Solution Bubble Detector (Fig. 28)...................................3-10
4-1 REPLACEMENTS.....................................................................................................................4-1
4-1.1 POWER SUPPLY.................................................................................................................4-1
4-1.2 LAMP................................................................................................................................... 4-2
4-1.3 MINI-COMPUTER BOARD “KIT”....................................................................................4-3
4-1.4 HARD DRIVE......................................................................................................................4-4
4-1.5 FLOPPY DRIVE..................................................................................................................4-4
4-1.6 SYRINGES(Fig. 22 & 27)....................................................................................................4-5
4-1.7 CUVETTES(Fig. 3 & 5).......................................................................................................4-5
4-1.8 CUVETTE PELTIER MODULES and NTC........................................................................4-6
4-1.9 PHOTOMETER....................................................................................................................4-8
4-1.10 REAGENT PELTIER MODULES.......................................................................................4-9
7-5 Printer..........................................................................................................................................7-3