Академический Документы
Профессиональный Документы
Культура Документы
Manual Therapy
journal homepage: www.elsevier.com/math
Original article
a r t i c l e i n f o a b s t r a c t
Article history: Measuring range of motion (ROM) is the first step of a physical examination and functional evaluation of
Received 7 September 2011 the shoulder joint. Digital inclinometers are available on the market; however, they are expensive, and
Received in revised form hence will not come into wide use. In this study, we present a new method for the shoulder ROM
9 January 2012
measurement using the inclinometer application on a smartphone. We hypothesized that the method
Accepted 16 February 2012
would demonstrate acceptable reliability and reproducibility. Three observers performed goniometric
and smartphone inclinometric measurements of various shoulder movements, including both active and
Keywords:
passive ROM for forward flexion, abduction, external rotation while the arms are at the sides, external
Range of motion
Clinometers
rotation at 90 abduction, and internal rotation at 90 abduction. Measurements were performed in the
Smartphone affected shoulders of 41 patients. All measurements were taken twice to assess the intra-observer
Shoulder reliability. Inter- and intra-observer reliabilities were evaluated using the intraclass correlation coeffi-
cient (ICC). Reliability between two measurements was also assessed in terms of the ICC. Both the
goniometric and inclinometric measurements showed satisfactory inter-observer reliability except for
internal rotation at 90 abduction for which the ICC value was <0.7 (range, 0.63e0.68). Intra-observer
reliability was excellent with an ICC value > 0.9, except for some movements. Within-day inclino-
metric measurements with a smartphone showed acceptable reliability compared to the classical
goniometric measurements of movements and the correlation between the two measurements was
fairly high. Considering convenience and cost-effectiveness, this new method could be widely used for
measuring the shoulder ROM, although the between-day reliability needs to be established first.
Ó 2012 Elsevier Ltd. All rights reserved.
1356-689X/$ e see front matter Ó 2012 Elsevier Ltd. All rights reserved.
doi:10.1016/j.math.2012.02.010
S.H. Shin et al. / Manual Therapy 17 (2012) 298e304 299
2. Methods
2.1. Subjects
2.2. Observers
2.3. Procedures
Statistical analyses were performed using the SPSS software In the ICC analysis, both goniometric and inclinometric
package version 18.0 (SPSS Inc., Chicago, IL, USA). Inter-observer measurements showed satisfactory inter-observer reliability,
reliability was evaluated by the ICC (2, 1), a two-way random except for the IR90 values, for which the ICC value was <0.7 (range,
effects model (observers and subjects are treated as random effects) 0.63e0.68, Table 1). Intra-observer reliability was excellent with an
with a single measure and absolute agreement for each movement. ICC value > 0.9 except for the goniometric measurement of active
Intra-observer reliability for each observer was assessed in terms of ROM for FF by observer A, the smartphone inclinometric
the ICC (3, 1), a two-way fixed effects model (observers are treated measurement of active ROM for IR90 by observer B, and the
as fixed and subjects are treated as random effects) with a single goniometric measurement for passive ROM for IR90 by observer C
measure and absolute agreement for each movement. In addition, (Table 2). However, the ICC values for these exceptions were still
the reliability of the goniometric and inclinometric measurements acceptable (range, 0.79e0.89).
were analyzed by the ICC (3, 1), a two-way fixed effects model with The SEM and MDC90 values for the inter-observer reliability are
a single measure and absolute agreement for each movement and shown in Table 3, and the SEM and MDC90 values for the intra-
each observer. The ICC value can range from 0 to 1; values close to 1 observer reliability of observer A are shown in Table 4. The values
were much greater in the inter-observer reliability analysis than in
Fig. 4. Measurement of external rotation while the arms are at the sides: smartphone Fig. 6. Measurement of internal rotation at 90 abduction: smartphone (left) and
(left) and goniometer (right). goniometer (right).
S.H. Shin et al. / Manual Therapy 17 (2012) 298e304 301
Table 1 Table 3
Inter-observer reliability for both the smartphone inclinometric and goniometric SEM and MDC90 for the inter-observer reliability (first and second sessions).
measurements (first and second sessions).
Modality Movement First session Second session
Modality Movement First session Second session
SEM MDC90 SEM MDC90
ICC 95% CI ICC 95% CI Smartphone FF a 9.99 23 9.56 22
Smartphone FF a 0.83 0.73e0.90 0.84 0.74e0.90 inclinometer FF p 10.32 24 10.09 24
inclinometer FF p 0.73 0.59e0.83 0.74 0.61e0.84 ABD a 13.84 32 13.21 31
ABD a 0.78 0.66e0.87 0.79 0.68e0.87 ABD p 14.15 33 13.83 32
ABD p 0.70 0.56e0.81 0.72 0.58e0.83 ERS a 9.39 22 9.67 23
ERS a 0.77 0.64e0.86 0.76 0.63e0.86 ERS p 9.56 22 9.48 22
ERS p 0.78 0.66e0.87 0.78 0.66e0.87 ER90 a 7.80 18 7.76 18
ER90 a 0.87 0.79e0.92 0.87 0.80e0.93 ER90 p 7.34 17 7.15 17
ER90 p 0.89 0.82e0.93 0.90 0.83e0.94 IR90 a 10.29 24 10.62 25
IR90 a 0.67 0.51e0.79 0.66 0.49e0.79 IR90 p 11.54 27 10.50 25
IR90 p 0.63 0.47e0.77 0.68 0.50e0.81
Goniometer FF a 12.03 28 8.80 21
Goniometer FF a 0.77 0.64e0.86 0.86 0.77e0.92 FF p 7.53 18 6.30 15
FF p 0.84 0.75e0.91 0.89 0.82e0.93 ABD a 11.85 28 10.01 23
ABD a 0.85 0.76e0.91 0.89 0.82e0.94 ABD p 12.77 30 11.14 26
ABD p 0.78 0.65e0.87 0.83 0.72e0.90 ERS a 8.35 19 8.32 19
ERS a 0.81 0.70e0.89 0.80 0.69e0.88 ERS p 8.98 21 8.62 20
ERS p 0.79 0.67e0.88 0.80 0.68e0.88 ER90 a 6.33 15 7.49 17
ER90 a 0.91 0.85e0.95 0.87 0.77e0.93 ER90 p 6.83 16 7.71 18
ER90 p 0.90 0.84e0.94 0.88 0.79e0.93 IR90 a 10.94 26 11.35 26
IR90 a 0.67 0.43e0.82 0.67 0.33e0.83 IR90 p 11.70 27 11.43 27
IR90 p 0.64 0.41e0.79 0.68 0.31e0.85
SEM, standard error of measurement; MDC90, minimal detectable change at the 90%
ICC, intraclass correlation coefficient; CI, confidence interval; FF, forward flexion; ABD, confidence level; FF, forward flexion; ABD, abduction; ERS, external rotation while
abduction; ERS, external rotation while the arms are at the sides; ER90, external the arms are at the sides; ER90, external rotation at 90 abduction; IR90, internal
rotation at 90 abduction; IR90, internal rotation at 90 abduction; a, active; p, passive. rotation at 90 abduction; a, active; p, passive.
intra-observer reliability analysis. In the intra-observer reliability measurements for most of the shoulder movements in all the
analysis of observer A, the values for smartphone inclinometric observers (Table 5). The Bland and Altman’s 95% LOA ranged from
measurement were better (smaller) than those for goniometric 10 to 40 , showing greater values for the FF and ABD measurements
measurement, suggesting that the measurement error was smaller of observer A (Table 6). The PCC analysis showed strong positive
with the use of the smartphone inclinometer. However, the values correlation between the two methods of measurement, with coef-
for active ROM for FF were much greater than those for other ficient values ranging from 0.79 to 0.97 for each movement (Table 6).
movements, in both the smartphone inclinometric and goniometric
measurements.
4. Discussion
3.2. Agreement between smartphone inclinometric and goniometric In this study, we presented a new method for measuring the
measurements shoulder ROM using the inclinometer application on a smartphone.
As seen in the results, the reliability the measurements obtained
In the ICC analysis, the inclinometric measurements using this new method was similar to the classical double-armed
showed acceptable reliability when compared to the goniometric goniometric measurements and was even better with respect to
Table 2
Intra-observer reliability for the smartphone inclinometric and goniometric measurements.
ICC, intraclass correlation coefficient; CI, confidence interval; FF, forward flexion; ABD, abduction; ERS, external rotation while the arms are at the sides; ER90, external rotation
at 90 abduction; IR90, internal rotation at 90 abduction; a, active; p, passive.
302 S.H. Shin et al. / Manual Therapy 17 (2012) 298e304
Table 4 Table 6
SEM and MDC90 for the intra-observer reliability of observer A. Bland and Altman’s 95% LOA and PCC between smartphone inclinometric and
goniometric measurements.
Movement Smartphone Goniometer
Movement Observer A Observer B Observer C
SEM MDC90 SEM MDC90
FF a 2.69 12 6.28 27 LOA PCC LOA PCC LOA PCC
FF p 2.30 2 5.36 4 FF a 29 0.84 15 0.95 18 0.93
ABD a 6.30 3 14.70 7 FF p 28 0.79 14 0.92 18 0.91
ABD p 4.51 3 10.52 6 ABD a 30 0.88 14 0.97 24 0.92
ERS a 3.26 3 7.61 6 ABD p 40 0.80 13 0.97 23 0.92
ERS p 2.79 2 6.51 4 ERS a 17 0.90 12 0.96 15 0.93
ER90 a 3.01 2 7.01 5 ERS p 18 0.89 16 0.91 17 0.91
ER90 p 3.04 3 7.09 8 ER90 a 14 0.95 13 0.95 10 0.97
IR90 a 1.86 3 4.33 6 ER90 p 13 0.96 14 0.95 11 0.97
IR90 p 3.18 3 7.42 7 IR90 a 14 0.93 15 0.90 14 0.93
IR90 p 14 0.92 11 0.95 22 0.84
SEM, standard error of measurement; MDC90, minimal detectable change at the 90%
confidence level; FF, forward flexion; ABD, abduction; ERS, external rotation while LOA, limits of agreement; PCC, Pearson’s correlation coefficient; FF, forward flexion;
the arms are at the sides; ER90, external rotation at 90 abduction; IR90, internal ABD, abduction; ERS, external rotation while the arms are at the sides; ER90,
rotation at 90 abduction; a, active; p, passive. external rotation at 90 abduction; IR90, internal rotation at 90 abduction; a, active;
p, passive.
the smaller measurement error. When compared to the previous being informed about the protocol. This also led to a greater SEM
studies on inclinometric measurements of shoulder ROM, the and MDC90 for active FF measured by observer A than those for
results of our study showed a slightly superior or similar reliability other movements (Table 4). The reason for lower inter-observer
based on the ICC analysis (Table 7) (Hoving et al., 2002; de Winter reliability and wider confidence intervals for intra-observer vari-
et al., 2004; Tveita et al., 2008; Mullaney et al., 2010; Cadogan et al., ability of IR90 measurement may be due to the varying amount of
2011; Kolber et al., 2011). scapular motion control, which is not restricted physiologically or
Previous reliability results vary based on the direction of due to the examination table as is the case when ER90 is measured.
movement (Riddle et al., 1987; Hoving et al., 2002; de Winter et al., Although we had consensus regarding controlling scapular motion
2004; Tveita et al., 2008; Mullaney et al., 2010; Cadogan et al., 2011; and the measurements were performed in the supine position, it is
Kolber et al., 2011). Despite the overall acceptable reliability, our difficult to completely eliminate inter-observer and even intra-
study also demonstrated variable reliabilities for some movements: observer variability.
the ICC values for inter-observer reliability of the IR measurement The PCC analysis showed strong positive correlation between
were lower and had wider confidence intervals (Table 1); the ICC the two methods of measurement including FF, ABD, and IR
values for intra-observer reliability of the FF and IR measurements measurements. This suggests that the two methods are consistent
of some observers were lower and had wider confidence intervals with each other despite the difference in the absolute values of
(Table 2). While comparing smartphone inclinometric and gonio- measurements, and therefore the new smartphone inclinometric
metric measurements, the ICC values for FF had wider confidence measurements can be as useful as the classical goniometric
intervals than those for other movements (Table 5); the Bland and measurements if the method is used consistently.
Altman’s 95% LOAs for FF and ABD measured by observer A were The result of SEM and MDC90 analyses for inter-observer reli-
greater than that measured by the other observers (Table 6). ability of both smartphone inclinometric and goniometric
The lower reliability and wider confidence intervals of the FF measurements was not as good as the result of ICC analysis,
and ABD measurements could be attributed to the variability of the showing greater values (range, 6.30 e14.15 for SEM; 15 e33 for
normal carrying angle of the elbow and the degree of uncontrolled MDC90) (Table 3) compared to those reported by Kolber et al.
elbow flexion during the measurement. The smartphone was fixed (range, 1.64 e4.27 for SEM; 4 e9 for MDC90) (Kolber et al., 2011).
to the patient’s forearm, while the reference for goniometric This also means that the absolute values for each ROM measured by
measurement of FF and ABD was the patient’s upper arm. There- the three observers generally varied more in our study than in
fore, there can be variable differences depending on the elbow previous studies, despite consensus on the measurement protocol.
angle, especially in old aged patients who tend to flex their elbow in However, SEM and MDC90 values for intra-observer reliability of
order to elevate the arm higher while measuring FF or ABD despite smartphone inclinometric measurements of observer A were much
better (range, 1.86 e6.30 for SEM; 2 e12 for MDC) (Table 4) than
those for inter-observer reliability of smartphone inclinometric
Table 5
ICCs between smartphone inclinometric and goniometric measurements. measurements, suggesting that consistent measurements could be
achieved using the smartphone inclinometer, if the measurements
Movement Observer A Observer B Observer C
were performed consistently by one observer or if a more strictly
ICC 95% CI ICC 95% CI ICC 95% CI standardized measurement protocol was used.
FF a 0.80 0.56e0.90 0.90 0.42e0.97 0.89 0.72e0.95 Once the reliability of a method for specific physical examina-
FF p 0.72 0.42e0.86 0.87 0.53e0.95 0.87 0.66e0.94 tion is proven, ease of use and cost-effectiveness could be the next
ABD a 0.88 0.79e0.94 0.97 0.94e0.98 0.91 0.94e0.95
ABD p 0.80 0.65e0.89 0.97 0.95e0.98 0.88 0.72e0.95
issue. Previous authors have already emphasized on the ease of
ERS a 0.89 0.81e0.94 0.93 0.77e0.97 0.92 0.86e0.96 measuring ROM with an inclinometer (de Winter et al., 2004;
ERS p 0.89 0.81e0.94 0.90 0.82e0.95 0.90 0.79e0.95 Cadogan et al., 2011; Kolber et al., 2011). Realignment of the
ER90 a 0.94 0.89e0.97 0.95 0.91e0.97 0.97 0.94e0.98 device during an inclinometric measurement is usually not
ER90 p 0.94 0.88e0.97 0.95 0.91e0.98 0.97 0.94e0.98
required, so it can be held with one hand, leaving the examiner’s
IR90 a 0.93 0.86e0.96 0.87 0.66e0.94 0.93 0.86e0.96
IR90 p 0.91 0.81e0.95 0.93 0.71e0.97 0.84 0.72e0.91 other hand free to stabilize the device. Moreover, if the inclinom-
eter is attached to the patient’s arm, as was the case in our study,
ICC, intraclass correlation coefficient; CI, confidence interval; FF, forward flexion;
ABD, abduction; ERS, external rotation while the arms are at the sides; ER90,
both the examiner’s hands are free. Therefore, the axis of the
external rotation at 90 abduction; IR90, internal rotation at 90 abduction; a, active; patient’s arm can be fixed with one hand, and the other hand can
p, passive. provide pressure to the arm while measuring the passive ROM. In
S.H. Shin et al. / Manual Therapy 17 (2012) 298e304 303
Table 7
Reliabilities and characteristics of previous studies on inclinometric measurements of the shoulder ROM.
Authors Publication Methods Cases No. of raters Range of motion Inter-observer Intra-observer reliability
year included reliability
Cadogan et al. 2011 Inclinometer 40 Patients 2 Active and passive 0.45e0.95 0.85e0.99 (within-day)
Kolber et al. 2011 Inclinometer 30 Volunteers 2 Active 0.58e0.95 0.83e0.94 (within-day)
Mullaney et al. 2010 Inclinometer and 20 Patients 2 Active 0.31e0.93a 0.91e0.99a (within-day)
goniometer
Tveita et al. 2008 Inclinometer 32 Patients 1 Active and passive 0.72e0.93 (between-day)
de Winter et al. 2004 Inclinometer 155 Patients 2 Passive 0.28e0.90
Hoving et al. 2002 Inclinometer 6 Volunteers 6 Active 0.11e0.80 0.32e0.91 (within-day)
a
Note that the data are of inclinometric measurements.
our study, all three observers mentioned this convenience of using is possible and this can motivate the patients for an enthusiastic self
an inclinometer. Additionally, we expect that the reading error rehabilitation. Therefore, we are planning to evaluate the effective-
could be minimized further, because we used the figures displayed ness of home monitoring of shoulder ROM in our next study.
on the monitor, rather than the gradations on the goniometer. There are limitations to this study. Although we measured
Moreover, we did not have to set an imaginary vertical line, which shoulder movements in various directions, we could not measure
usually is required while performing a goniometric measurement, the internal rotation at the back of the shoulder or cross body
so the error can be further minimized. However, commercial adduction. These movements can be measured at the level of the
inclinometers have many disadvantages and the clinics must be spinous process using a tape. Therefore, this limitation is true for
equipped with expensive devices. In contrast, when the smart- our study and all other studies with inclinometric measurements.
phone is used as an inclinometer, there is no need to purchase The other limitation is that we claimed that the smartphone could
a special device. Besides, the application that we used in this study relieve the financial burden of buying an inclinometer; however, it
was free of cost (Android market for smartphones). There are many is assumed that the doctor has a smartphone, particularly the one
inclinometer programs for smartphones on the market and we with a gyro-sensor system. We confirmed that some other smart-
think other applications would provide similar convenience phones are equipped with gyro-sensors and have an inclinometer
compared to the one that we used. In addition, we usually do not application as well (e.g., iPhone 3G and 4G [Apple Computer Inc.,
need any additional room for a smartphone in our clinic. In this Cupertino, CA, USA] and Optimus 2X [LG Electronics, Seoul, Korea]),
study, we simply attached the smartphone to the patient’s forearm although the applications for some smartphones need to be
using a commercial armband designed for smartphones, which purchased (e.g., $ 0.99 for the iPhone) However, as far as we know,
allowed both the examiner’s hands to be free; however, a patient not all smartphones can be used as an inclinometer because of the
can also attach a smartphone to him/her self at home, so that there lack of a gyro-sensor system. Yet another limitation is the lack of
is no need to visit a particular place where the measuring device between-day reliability data. It is well known that studies in which
exists, as was the case in the previous studies (Mullaney et al., 2010; repeated tests are performed at short time intervals, as was the case
Cadogan et al., 2011; Kolber et al., 2011). in the present study, may yield very different results than those in
The observers who participated in this study had varying levels the studies in which repeated tests are performed at longer time
of experience; from a novice orthopedic resident to a fellowship- intervals (i.e., days or weeks) (Gajdosik & Bohannon, 1987).
trained shoulder surgeon. We speculated that the intra-observer However, there is a substantial difficulty in following up the
reliability would be inferior in the less experienced doctors; patients for the ROM measurement within a short time interval
however, this was not the case. We think that standardization and such as days, which is not absolutely necessary for the management
education regarding the measurement method in several prelimi- of the patients’ problems. Meanwhile, if the follow-up interval is
nary meetings minimized this problem. longer such as weeks or months, the ROM may vary to a greater
In our study, only symptomatic patients with the affected side extent according to the patients’ clinical course, than according to
were included for evaluation. We believe that this represents the the observational error. This generally limits the use of a measuring
actual clinical situation rather than including volunteers (Hoving device for comparing the effects of rehabilitation over time. For this
et al., 2002; Kolber et al., 2011). Some previous studies included reason, a further study should be performed to investigate whether
ROM of the unaffected side in symptomatic patients for the analysis the between-day reliability of the smartphone is acceptable, before
and calculation of reliability (de Winter et al., 2004; Tveita et al., its clinical use for documenting the change in the ROM over time.
2008; Mullaney et al., 2010). One study showed similar reliability
between both (Tveita et al., 2008); however, two other studies 5. Conclusions
showed inferior reliability in the unaffected side (de Winter et al.,
2004; Mullaney et al., 2010). The shoulder ROM measurements with a smartphone application
Due to multiplanar motion of the shoulder, researchers usually were comparably reliable to the conventional goniometric measure-
choose either a passive or active ROM for a reliability analysis ments, although the between-day reliability remains to be established.
(Riddle et al., 1987; Hoving et al., 2002; de Winter et al., 2004; Considering their convenience and cost-effectiveness, smartphones
Mullaney et al., 2010; Kolber et al., 2011). However, some studies could be used widely in clinics. Another advantage of a smartphone is
have analyzed both the ranges of motions and showed similar that the patient can monitor their ROM during home-based physical
reliabilities (Tveita et al., 2008; Cadogan et al., 2011). Our study therapy for conservative or postoperative management.
measured both the ranges of motions and showed similar results as
those in previous studies. References
Another great advantage of measuring the ROM with the smart-
phone application is that the patient can monitor his/her ROM Atkinson G, Nevill AM. Statistical methods for assessing measurement error (reli-
without having to buy an expensive device. Home-based physical ability) in variables relevant to sports medicine. Sports Medicine 1998;26(4):
217e38.
therapy is important during the postoperative course of rehabilitation Bland JM, Altman DG. Statistical methods for assessing agreement between two
or the treatment course for a stiff shoulder. Immediate self-feedback methods of clinical measurement. Lancet 1986;1(8476):307e10.
304 S.H. Shin et al. / Manual Therapy 17 (2012) 298e304
Burdock EIFJL, Hardesty AS. A new view of interobserver agreement. Perspectives in Kolber MJ, Vega F, Widmayer K, Cheng MS. The reliability and minimal detectable
Psychology 1963;16:373e84. change of shoulder mobility measurements using a digital inclinometer.
Cadogan A, Laslett M, Hing W, McNair P, Williams M. Reliability of a new hand-held Physiotherapy Theory and Practice 2011;27(2):176e84.
dynamometer in measuring shoulder range of motion and strength. Manual Mullaney MJ, McHugh MP, Johnson CP, Tyler TF. Reliability of shoulder range of
Therapy 2011;16(1):97e101. motion comparing a goniometer to a digital level. Physiotherapy Theory and
de Winter AF, Heemskerk MA, Terwee CB, Jans MP, Deville W, van Practice 2010;26(5):327e33.
Schaardenburg DJ, et al. Inter-observer reproducibility of measurements of Portney LG, Watkins MP. Foundations of clinical research e applications to practice.
range of motion in patients with shoulder pain using a digital inclinometer. Connecticut: Appleton and Lange; 1993.
BMC Musculoskeletal Disorders 2004;5:18. Riddle DL, Rothstein JM, Lamb RL. Goniometric reliability in a clinical setting.
Gajdosik RL, Bohannon RW. Clinical measurement of range of motion. Review of goni- Shoulder measurements. Physical Therapy 1987;67(5):668e73.
ometry emphasizing reliability and validity. Physical Therapy 1987;67(12):1867e72. Scientific Advisory Committee of the Medical Outcome Trust. Assessing health
Green S, Buchbinder R, Forbes A, Bellamy N. A standardized protocol for measurement of status and quality-of-life instruments: attributes and review criteria. Quality of
range of movement of the shoulder using the Plurimeter-V inclinometer and Life Research: An International Journal of Quality of Life Aspects of Treatment,
assessment of its intrarater and interrater reliability. Arthritis Care and Research: The Care and Rehabilitation 2002;11(3):193e205.
Official Journal of the Arthritis Health Professions Association 1998;11(1):43e52. Tveita EK, Ekeberg OM, Juel NG, Bautz-Holter E. Range of shoulder motion in
Hayes K, Walton JR, Szomor ZR, Murrell GA. Reliability of five methods for assessing patients with adhesive capsulitis; intra-tester reproducibility is acceptable for
shoulder range of motion. The Australian Journal of Physiotherapy 2001;47(4):289e94. group comparisons. BMC Musculoskeletal Disorders 2008;9:49.
Hoving JL, Buchbinder R, Green S, Forbes A, Bellamy N, Brand C, et al. How reliably Valentine RE, Lewis JS. Intraobserver reliability of 4 physiologic movements of the
do rheumatologists measure shoulder movement? Annals of the Rheumatic shoulder in subjects with and without symptoms. Archives of Physical Medicine
Diseases 2002;61(7):612e6. and Rehabilitation 2006;87(9):1242e9.