Вы находитесь на странице: 1из 7

Medical Engineering and Physics 38 (2016) 326–332

Contents lists available at ScienceDirect

Medical Engineering and Physics


journal homepage: www.elsevier.com/locate/medengphy

Design and evaluation of an actuated knee implant for postoperative


ligament imbalance correction✩
A. Collo a,∗, S. Almouahed b, P. Poignet a, C. Hamitouche b, E. Stindel b
a
LIRMM, CNRS UMR 5506 UM2, 161 rue Ada, Montpellier 34095, France
b
LaTIM, INSERM-UBO UMR 1101, CHRU Morvan, 2 av. Foch, Brest 29609, France

a r t i c l e i n f o a b s t r a c t

Article history: In Total Knee Arthroplasty (TKA), the collateral ligament tensioning stage cannot be standardised for all
Received 10 October 2015 patients and relies heavily on the surgeon’s experience and perception. Intraoperative inaccuracies are
Accepted 6 January 2016
practically unavoidable and may give rise to severe postoperative complications, leading to the need for
revision surgery already a few years after primary TKA. This work proposes a novel instrumented tibial
Keywords: component able to detect collateral ligament laxity conditions right after primary TKA and, if needed,
Instrumented knee prosthesis to compensate for them in the postoperative period. A miniaturised actuation system, designed to be
Knee ligament imbalance embedded in the tibial baseplate, was initially evaluated by means of 3D simulations and then fabricated
Actuation system as a full-scale prototype. Stability and force sensors tests carried out on a knee simulator allowed to
Miniaturised mechanical device
assess the effectiveness of the proposed design under normal working conditions and provided valuable
insights for future work and improvements.
© 2016 IPEM. Published by Elsevier Ltd. All rights reserved.

1. Introduction inevitably face with the intraoperative inaccuracies introduced by


the replacement of the employed sensing devices with the perma-
A key achievement of Total Knee Arthroplasty (TKA) lies in in- nent prosthetic components.
traoperatively setting up proper tension conditions of the two knee In this context, over the last decade, the development of instru-
collateral ligaments [1]. This stage, which directly determines the mented knee implants has been accepted as a way forward. For
postoperative Knee Joint (KJ) function and the implant lifespan, instance, dedicated telemetry systems and microelectronics com-
cannot be standardised and is typically carried out by evaluating ponents embedded in the prosthesis can provide unique intra-
variable and user-dependent loads [2]. Over the last decades, var- and postoperative KJ kinematics data [2,6,7]. Besides improving the
ious assistive tools have been developed to help surgeons during quality of the surgical procedure itself, such in vivo data made
this crucial TKA ligament tensioning stage. available to surgeons have turned out to be very helpful for clini-
A few examples of instrumented knee distractors can be found cians and prosthesis designers as well.
in the literature [3,4]. By controlling miniaturised mechanical However, no assistive tool has been developed to specifically
structures, surgeons are able to monitor important intraoperative address the problem of postoperative ligament imbalance. The cru-
parameters such as tibiofemoral gaps and loads. Another type of cial knee stabilisation process highly depends on the surgeon’s
assistive tools is represented by instrumented tibial baseplates: experience and perception [8,9]. The morphofunctional evolution
the use of embedded sensors, such as strain gauges [5], allows that the prosthetic KJ normally undergoes during the rehabili-
surgeons to perform fine-tuning adjustments in ligament tension tation period and in its aftermath may alter the collateral liga-
during the component positioning process. Although, the optimal ment tension values set during surgery. Some studies reported that
balance conditions that surgeons can determine during surgery some degree of postoperative collateral ligament laxity may be ac-
cepted [10] and even spontaneously corrected after TKA [11]. How-
ever, suboptimal balance conditions generally lead to postoperative
✩ complications and, in the worst-case scenario, to revision surgery
The authors received no specific funding for this manuscript. No conflict of in-
terest, financial or other, exists to any authors listed in this manuscript. (whose number is steadily increasing [12]). This suggests there-

Corresponding author. fore to consider the possibility to compensate for intraoperative
E-mail addresses: andrea.collo@lirmm.fr, a.collo@gmail.com (A. Collo), shaban.
inaccuracies in the postoperative period; this can be achieved by
almouahed@telecom-bretagne.eu (S. Almouahed), poignet@lirmm.fr (P. Poignet),
chafiaa.hamitouche@telecom-bretagne.eu (C. Hamitouche), eric.stindel@univ-brest.fr means of a miniaturised actuation system embedded in the knee
(E. Stindel). prosthesis.

http://dx.doi.org/10.1016/j.medengphy.2016.01.002
1350-4533/© 2016 IPEM. Published by Elsevier Ltd. All rights reserved.
A. Collo et al. / Medical Engineering and Physics 38 (2016) 326–332 327

Fig. 1. A lax collateral ligament (left) can be properly re-tensioned by lifting the
tibial tray up on the corresponding side (right).

The aim of this work was to propose a novel instrumented knee


implant able to detect collateral ligament laxity conditions follow-
ing primary TKA and, if needed, to restore optimal balance condi-
tions before the occurrence of more severe complications. In the
long term, the proposed device is intended to strongly reduce the
need for revision surgery.
Fig. 2. The proposed Wedge-Leadscrew system.

2. Materials
KJ function. If needed, further refinements can be carried out until
2.1. Implant description optimal balance conditions are restored.
Only a continuous accurate monitoring of the evolution of the
The instrumented tibial component presented in this work fur- prosthetic KJ stability combined to the use of the proposed implant
ther develops the smart knee implant previously proposed by our would allow to restore optimal balance conditions after primary
research team [13], which includes two sub-parts: TKA, without the need for revision surgery. To the author’s knowl-
– for the evaluation of collateral ligament laxity conditions, four edge, no other ongoing study copes with the problem of postoper-
PiezoCeramics (PC) elements embedded in the tibial baseplate ative knee ligament imbalance using the same approach proposed
are employed as both force sensors and energy harvesters [14]. in this work.
The collected data are meant to be processed and wirelessly
transmitted to the clinician’s computer by means of a self- 2.3. Actuation system design
powered telemetry system [15]. This allows to define a proper
correcting action for restoring optimal balance conditions; For sake of clarity, the description and the figures reported in
– for the detected imbalance correction, the miniaturised actua- the following refer to the actuation of only one side of the tibial
tion system proposed by Collo et al. [16] is embedded in the platform (Fig. 2). The reader is invited to keep in mind that the
tibial baseplate. In order to re-tension any detected lax collat- instrumented tibial component actuation is achieved by means of
eral ligament, the height and inclination of the tibial tray can two identical systems, one per each side of the tibial baseplate.
be adjusted on the corresponding side (Fig. 1). Such correct- More details are provided elsewhere [16].
ing action aims to reproduce the local adjustments carried out The miniaturised actuation system is based on the translation
by means of spacer blocks during revision TKA [17] and tibial of a wedge whose shape is custom-designed (30 mm wide, 10 mm
osteotomy [18]. long, 4 mm high). The wedge is guided inside a rail grooved in the
tibial baseplate (Fig. 2). Via a threaded hole, the wedge is coupled
2.2. Clinical needs to a leadscrew (ISO M2x0.4 profile: 2 mm pitch diameter, 0.4 mm
pitch) which can only rotate without translating. As in a screw-nut
It is very important to detect any postoperative instability as configuration, any leadscrew revolution produces a corresponding
early as possible; in the aftermath of the rehabilitation period, pe- wedge translation according to the thread pitch; herein, such as-
riodic follow-up visits allow the surgeon to monitor the prosthetic sembly will be addressed as a Wedge-Leadscrew (WL) system.
KJ postoperative function. Any collateral ligament laxity condition A mobile tibial tray is shaped so as to be positioned from above
can be detected after analysing the data collected by four PC ele- onto the baseplate and fit the two WL systems. As depicted in
ments embedded in the tibial baseplate [13]. Accordingly, the sur- Fig. 2, without any actuation the tray is fully contained inside the
geon defines an appropriate correcting action in terms of tibial tray baseplate in a configuration that corresponds to that of conven-
lateral uplift (4 mm maximum, in order not to impair the prosthe- tional tibial platforms. In its starting position, the wedge is aligned
sis alignment to the lower limb mechanical axis [18]). to the center of the baseplate (null translation). As the leadscrew
A progressive fine-tuning correction process is proposed by rotates, the wedge translates laterally, towards the baseplate outer
means of successive stages: actuation is performed a first time and border. By doing this, the wedge slides under the mobile tray and
the tibial tray is locked in the new raised position. Right after, lifts it upwards. The greater the wedge translation, the greater the
the PC elements data are processed again to check the updated mobile tray lateral uplift.
328 A. Collo et al. / Medical Engineering and Physics 38 (2016) 326–332

Fig. 3. The proposed instrumented tibial component fabricated as a full-scale stainless steel prototype.

The leadscrew is held in place by specific supports and is rigidly material was Titanium Vanadium Alloy (Ti-6Al-4V), which repre-
fixed to the shaft of a rotary stepper micromotor (entirely embed- sents a conventional choice for knee prostheses [21].
ded in the tibial baseplate). Once detected the lax collateral liga- The system was considered in a configuration of full actua-
ment, only the microactuator on the corresponding side needs to tion on one side of the tibial baseplate (10 mm wedge transla-
be powered in order to drive the related WL system. tion, 3.6 mm mobile tibial tray lateral uplift). A compression load
L = 2600 N was defined perpendicularly to the mobile tibial tray
2.4. System efficiency upper surface, simulating the peak tibiofemoral force acting inside
the prosthetic KJ during normal gait [20].
The severe constraints imposed by a biocompatible environ- Under the aforementioned settings, static analysis simulations
ment were met by means of smart design choices: the combined were run on the actuated implant. The following two measures
use of a wedge and a screw resulted in an efficient mechanism were taken into account:
able to develop great output forces in face of reduced input ef- – the distribution of Von Mises Stress (VMS), a scalar value (unit:
forts. Efficiency was deemed a key design aspect. The miniaturised MPa) commonly used to predict yielding of materials under any
actuation system was needed to be sufficiently small in order to be user-defined loading condition [22];
embedded in the tibial component and, at the same time, robust – the deformation of the mobile tibial tray (unit: mm) under the
enough to face with the strong KJ cyclic efforts [19,20]. The me- effect of the compression load L.
chanical model of the proposed actuation system was accurately
studied by means of static force analysis. Different design expedi- A buckling analysis was carried out to estimate the maximum
ents were discussed for the minimization of friction losses. static load bearable by the mobile tibial tray before deformation. In
Due to the presence of the embedded actuation system, the tib- order to consider the worst-case postoperative scenario, the load L
ial baseplate is 9 mm thick. According to the wedge translation, the was then simulated as an intermittent tibiofemoral load repeatedly
mobile tibial tray lateral uplift varies in the range 0–3.6 mm and exerted onto the mobile tibial tray (1 Hz frequency). The number
linearly depends on the number of leadscrew revolutions (driven of repeated loads (cycles) that the implant was able to bear before
by the microactuator). Two miniature thrust ball bearings prop- failure was studied by means of a fatigue analysis.
erly fixed at the two leadscrew ends, together with the WL system
threaded profile, provide a robust passive locking system. Thus, the 3.2. Stability tests
mobile tibial tray can be rigidly blocked in the desired actuated po-
sition without the participation of the microactuator. The compo- Considering the extremely complex nature of KJ kinematics, it
nent shape was defined so as to allow only the mobile tray lateral was decided to carry out dynamic analyses by means of a knee
uplift: no undesired rotation with respect to the baseplate is likely simulator. The employed knee simulator (ADL-Force 5 Simulator,
to take place. AMTI Force and Motion) features a wide range of motion and can
As reported by Lahuec et al. [15], the energy harvested by the accurately replicate the KJ kinematics and physiological environ-
four PC elements can represent a power supply only for ultra-low ment associated with daily life activities.
power consumption devices (such as the aforementioned telemetry A full-scale prototype of the proposed tibial component was
system). Therefore, the actuation system was assumed to be pow- fabricated using stainless steel (Fig. 3). Such choice was deemed
ered and wirelessly controlled respectively by means of inductive appropriate for the validation of the proposed actuation system
coupling and radiofrequency techniques. All the microelectronics working principle. Upon satisfactory experimental outcomes, the
components necessary for the microactuator power supply, control fabrication of a titanium alloy prototype was reckoned with as the
and data transmission were assumed to be properly sealed inside next necessary step (currently ongoing).
the hollow tibial stem (available volume: 50 cm3 [16]). The design Mechanical tests were conducted to evaluate the stability of
of an integrated low-power consumption system is currently under the mobile tibial tray under different different physiological load-
development. ing conditions. Fifteen realistic walking scenarios were simu-
lated by progressively modifying the tibial component orienta-
tion on the knee simulator: various Internal-External Rotation (IER)
3. Methods
and Anterior-Posterior Displacement (APD) configurations were se-
lected as per ISO 14243 [20] in order to impair the homogeneous
3.1. 3D simulations
contact among the prosthetic components and replicate subopti-
mal balance conditions. The stability test procedure consisted in
The 3D model of the proposed tibial implant was developed un-
the following steps.
der PTC Creo design software. All the components were defined as
rigid bodies and modeled by means of three-dimensional tetrahe- (1) Before starting any test, the knee simulator was configured
dral meshes for finite element analysis. The selected component following the instructions provided by the manufacturer.
A. Collo et al. / Medical Engineering and Physics 38 (2016) 326–332 329

Fig. 4. The positioning of the two ultra-thin contact piezoresistive force sensors onto the mobile tibial tray upper surface.

(2) The instrumented tibial component was manually set in its internal rotation and 2 mm anterior translation, so as to impair
starting position (null actuation: 0 mm mobile tibial tray lat- the full contact normally experienced at the PE insert-femoral
eral uplift). component interface. In order to allow for a more immediate in-
(3) The orientation of the implant on the knee simulator was set terpretation of the tibiofemoral load distribution, the unit chosen
in order to have full contact between the PolyEthylene (PE) for displaying the data collected by the two force sensors was the
insert and the femoral component (fixed onto the simulator solicitation percentage with respect to their total measurement
condyle holder): such position was saved as Zero Position range (0–4448 N). Herein, such values will be addressed as M%
(ZP) and considered as a reference configuration. and L% respectively for the medial and the lateral force sensors.
(4) 500 successive gait cycles were simulated (2600 N peak To compensate for the simulated imbalance condition, the pro-
tibiofemoral load, 1 s walking cycle [20]). gressive correction procedure defined in Section 2.2 was followed.
(5) IER was set to 15 degrees internal rotation; 500 gait cycles The criterion for optimal restoration of balance conditions con-
were reproduced. sisted in visually matching the recorded graphical representations
(6) IER was set to 15 degrees external rotation; 500 gait cycles of M% and L% . Due to the constraints imposed by the experimen-
were replicated. tal settings, each successive system actuation step (manually per-
(7) The tibial component was oriented again in ZP and subjected formed) was alternated to a series of 50 simulated gait cycles, nec-
to 500 gait cycles. essary to update the data collected by the force sensors.
(8) APD was set to 2 mm anterior translation; 500 gait cycles
were simulated.
4. Results
(9) APD was set to 2 mm posterior translation; 500 gait cycles
were reproduced.
4.1. 3D simulations
(10) The mobile tibial tray lateral uplift was manually set to
1.8 mm and steps (3) to (9) were repeated.
As expected, the first static analysis revealed a VMS concentra-
(11) The mobile tibial tray lateral uplift was manually set to
tion in the regions where the mobile tibial tray thickness was re-
3.6 mm (full actuation condition) and steps (3) to (9) were
duced by the different cuts made to fit the presence of the two WL
repeated.
systems. In particular, higher VMS occurred along the sharp edges
of the Medio-Lateral (ML) and Antero-Posterior (AP) cuts necessary
3.3. Force sensors tests to host the microactuators and the wedges (red-coloured in Fig. 5).
The selected Titanium Vanadium Alloy offers a tensile yield
The implementation of any sensor in the prototype should be strength of 1100 MPa [25]. Such value, being nine times greater
defined very accurately in order not to bias the test dynamics than the maximum VMS computed by the static analysis (121 MPa,
replicated by the knee simulator. Such requirement was met by Table 1), was considered to give a very reliable margin of safety.
positioning two Tekscan FlexiForce B201 sensing devices onto the The maximum material deformation was estimated to be on the
mobile tibial tray upper surface. Symmetrically aligned with re- order of half a micrometre. Therefore, it was concluded that the
spect to the tray geometric center (Fig. 4), the two contact piezore- risk of failure of the mobile tibial tray was not a major concern.
sistive force sensors were individually calibrated and snug fit the However, the stress distribution was further improved by means of
thin spacing between the tray and the PE insert. two design expedients applied locally:
The goal of this experimental stage was to evaluate the varia-
tions in the distribution of tibiofemoral loads on the mobile tibial – the quantity of volume removed by the cuts was minimised by
tray under different actuation system configurations. No standards locally increasing the mobile tibial tray thickness where possi-
exist in terms of knee collateral ligament length and insertion sites, ble. The medial and lateral edges of the tray were thickened by
which strongly vary among individuals [11,23,24]. Therefore, the 1 mm (green-coloured in Fig. 5) while still fitting the presence
proposed actuation system is not intended to tension the collat- of the leadscrews. This allowed to strengthen the tray borders
eral ligaments to specific tension values (which are not accessible), and reduce the risk of deformation;
but rather to restore homogeneous load distribution conditions be- – the ML and AP sharp edges of the mobile tibial tray bottom sur-
tween the two mobile tibial tray compartments [13]. face were smoothed by means of a round fillet (pink-coloured
Any simulated laxity condition results in an uneven solicitation in Fig. 5). On each side, due to the presence of the micromo-
of the two tibial tray compartments. On the prototype, this was tor and the wedge, the sharp edges were rounded off by 3 and
verified by evaluating the force sensors data. The laxity of the lat- 1 mm respectively for the ML and AP cuts. The opportunity to
eral collateral ligament was reproduced on the knee simulator by round off other minor edges was also considered, but turned
taking into account the IER and APD values already considered for out not to give any significant improvement in the distribution
the stability tests. The tibial component was oriented at 15 degrees of the VMS experienced by the implant.
330 A. Collo et al. / Medical Engineering and Physics 38 (2016) 326–332

Fig. 5. The mobile tibial tray bottom surface before (left) and after (right) the optimisation study (A, P, M, L respectively stand for Anterior, Posterior, Medial, Lateral). (For
interpretation of the references to colour in this figure legend, the reader is referred to the web version of this article.)

Table 1
The main optimisation steps.

Measure 1st Static an. Thickening 1 mm fillet 2 mm fillet 3 mm fillet

Max VMS [MPa] 121 98 71 81 46


Max deformation [mm] 4.2 × 10−4 5.2 × 10−4 3.6 × 10−4 2.8 × 10−4 1.7 × 10−4

The static analysis carried out on the optimised 3D model


showed a more uniform stress distribution of the load L = 2600 N
onto the mobile tibial tray. The computed values of the maximum
VMS and deformation were both reduced by a factor of 2.5, with
the estimated maximum VMS value being almost 24 times lower
than the limit represented by the Titanium Vanadium Alloy tensile
yield strength. The static analyses results and the main optimisa-
tion steps are summarised in Table 1.
The mobile tibial tray in its fully uplifted position was esti-
mated to be capable of withstanding a maximum static load of
1.976 · 106 N (uniformly distributed on its surface) before expe-
riencing any permanent deformation.
According to the fatigue analysis results, the minimum lifespan
of the implant (in fully actuated configuration) was estimated as
Fig. 6. M% and L% data at the simulated imbalance condition, shown only for the
1.3 · 1018 cycles. The number of gait cycles performed by an active
last 10 simulated gait cycles.
patient during 15 years after primary TKA was estimated by Al-
mouahed [26] to be 4.7 · 107 ; with respect to such reference value,
thus, the instrumented knee implant proposed in this work should together with the support point on the baseplate non-actuated
offer a 2.7 · 1010 times longer lifespan. It was concluded that no side were definitively validated as sufficient conditions for the tray
mechanical failure of the mobile tibial tray was likely to occur un- stability.
der normal conditions of use.
4.3. Force sensors tests
4.2. Stability tests
With the actuation system in its starting position (null wedge
During the experimental stage the prototype underwent a total translation, null mobile tibial tray lateral uplift), the implant was
of 9 × 103 gait cycles equally distributed among three distinct subjected to 50 successive gait cycles (2600 N peak tibiofemoral
actuation cases (null, intermediate and full mobile tibial tray lat- load, 1 s walking cycle [20]). As expected, the simulated laxity of
eral uplift). In each of the fifteen studied cases, no uncontrolled the lateral collateral ligament resulted in an overload of the tibial
movement of the actuated implant took place. As anticipated in platform medial compartment. Fig. 6 reveals the non-homogeneous
Section 2.4, the corresponding profiles of the mobile tibial tray and solicitation of the two force sensors.
the baseplate were chosen so as to prevent any undesired rela- Four successive manual actuations (10, 15, 18 and 20 total com-
tive movement between the two components. This design choice plete leadscrew revolutions) were necessary to obtain matching
proved to be really effective: the tray did not experience any unde- M% and L% data. Fig. 7 depicts the evolution of the sensed forces
sired vibration or unexpected oscillation throughout multiple gait throughout the whole actuation process (for sake of clarity, only
cycles simulated in different configurations. the last 5 simulated gait cycles of each step are shown).
In addition to this, the full KJ motions experienced by the im- Table 2 summarizes the results of the force sensors tests. LS
plant did not produce any change in the wedge position. This con- stands for the number of complete leadscrew revolutions. The
firmed the soundness of the design considerations made by Collo terms M% and L% denote respectively the medial and lateral force
et al. [16] about the passive locking system. Contextually, the con- sensors data (M% and L% ) averaged over the 50 gait cycles sim-
tact line between the mobile tibial tray and the actuated wedge ulated after each actuation step. The term xCoP is a qualitative
A. Collo et al. / Medical Engineering and Physics 38 (2016) 326–332 331

Fig. 7. Evolution of M% and L% data during the actuation process, shown only for the last 5 simulated gait cycles.

Table 2
Actuation process data.

LS M% [%] L% [%] xCoP [mm]

0 16.00 6.00 10.23


10 13.50 10.50 2.81
15 16.00 13.00 2.33
18 13.00 11.50 1.38
20 11.00 10.50 0.52

evaluation of the tibiofemoral loads Center of Pressure (CoP) po-


sition. Due to the experimental settings, the CoP lies on the line
connecting the two force sensors. With respect to the simple ref-
erence system illustrated in Fig. 4, the CoP coordinate xCoP (unit:
mm) was computed from the sensors location on the tibial tray
Fig. 8. The evolution of IoB throughout the actuation process.
using Eq. (1):
M% (+22.5 ) + L% (−22.5 )
xCoP = (1) defined as the Index of Balance (IoB) of the two force sensors av-
M% + L% erage data. As depicted in Fig. 8, throughout the whole actuation
The key achievement of this experimental stage consisted in process IoB irregularly converged to the value 1, which represents
assessing the effectiveness of the actuation procedure proposed the ideal condition of identical solicitation of both tibial platform
in Section 2.2. Initially, the simulated laxity of the lateral collat- compartments (L% = M% ). In more general terms, the existence of
eral ligament resulted in an overload of the tibial platform medial a relationship between IoB and the mobile tibial tray lateral uplift
compartment (M% = 16%) with respect to the lateral one (L% = was deemed of great interest.
6%). Throughout the four actuation steps, the progressive mobile
tibial tray lateral uplift effectively allowed to compensate for the 5. Discussion
simulated ligament imbalance condition. By re-establishing proper
full contact conditions at the PE insert-femoral component inter- 5.1. Conclusion
face, the simulated KJ dynamics affected more homogeneously the
whole PE insert surface, without overloading only one tibial tray The idea of embedding a miniaturised actuation system inside
compartment (M% = 11% and L% = 10.50%). a knee prosthesis requires a multidisciplinary research work that
The data collected by the two force sensors further confirmed must be supported by experimental data. The creation of an accu-
the results of the stability tests. At each actuation step, the differ- rate 3D model of the proposed implant allowed to carry out sim-
ence between M% and L% remained fairly constant throughout the ulated static force analyses, to perform useful design optimisations
simulated gait cycles. This proved that the mobile tibial tray pro- and to evaluate the component robustness properties. In order
vided a stable support in all the actuated positions. to test the actuation system performances under normal working
The mobile tibial tray lateral uplift varied from 0 to 2.88 mm; conditions, a full-scale prototype of the instrumented tibial compo-
consequently, the CoP gradually translated from the tibial tray me- nent was fabricated and tested on a knee simulator. Stability and
dial compartment (10.23 mm) towards the platform geometric cen- force sensors tests were defined specifically to evaluate the actu-
ter (0.52 mm). Considering the variations in xCoP , the greatest dis- ation system working principle. The effectiveness of the proposed
tance was covered after the first 10 complete leadscrew revolutions design choices was assessed under physiological KJ conditions. The
with a translation from 10.23 mm to 2.81 mm. The remaining ac- promising results of the experimental validation stage globally met
tuation steps apparently had a smaller impact on the CoP posi- the expectations and provided positive indications for future stud-
tion. This aspect, as well as the total travelled distance (9.71 mm), ies and improvements.
should be considered as an objective of further investigation.
The criterion of considering knee balance conditions as opti- 5.2. Study limitations and perspectives
mally restored according to the visual matching of the two sensors
data limited any interpretation of the experimental outcomes to a The validation of the actuation system working principle un-
qualitative level. In these terms, the ratio between L% and M% was der realistic KJ walking conditions was deemed a crucial step
332 A. Collo et al. / Medical Engineering and Physics 38 (2016) 326–332

before proceeding with further experimentation. Indeed, the actu- [8] Winemaker MJ. Perfect balance in total knee arthroplasty: the elusive compro-
ation system fatigue and wear properties, as well as those of the mise. J Arthroplasty 2002;17(1):2–10.
[9] Schmidt F, Lampe F, Elfring R. Soft-tissue management in primary knee arthro-
PE insert, are a major concern and must be evaluated over millions plasty: common techniques, navigation and force-sensing devices. Z Orthop
of simulated gait cycles. The ongoing fabrication of a titanium al- Unfall 2007;145(5):599–607.
loy prototype would allow to collect more reliable data about the [10] Mayr HO, Reinhold M, Hube R, Von Roth P, Bernstein A, Suedkamp N, et al. Ro-
tational laxity and collateral ligament laxity following total knee arthroplasty
implant robustness properties. Contextually, it appears imperative with rotating platform. Int Orthop 2014;38(7):1379–86.
to define more clinically significant criteria for the assessment of [11] Sekiya H, Takatoku K, Takada H, Sasanuma H, Sugimoto N. Postoperative lat-
optimal balance conditions. eral ligamentous laxity diminishes with time after tka in the varus knee. Clin
Orthop Relat Res 2009;467(6):1582–6.
For biocompatibility reasons, all the mechanical and electron-
[12] Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision
ics components must not enter in contact with the physiological hip and knee arthroplasty in the united states from 2005 to 2030. J Bone Joint
fluids that bath the knee articulation. To this purpose, the use of Surg 2007;89(4):780–5.
[13] Almouahed S, Gouriou M, Hamitouche C, Stindel E, Roux C. Design and
an elastic micro-membrane could be envisaged. Properly fixed be-
evaluation of instrumented smart knee implant. IEEE Trans Biomed Eng
tween the mobile tibial tray lateral edges and the baseplate, the 2011;58(4):971–82.
membrane would create a sealed volume for all the embedded [14] Almouahed S, Gouriou M, Hamitouche C, Stindel E, Roux C. The use of piezoce-
components. Biocompatible casings may also be employed as a ramics as electrical energy harvesters within instrumented knee implant dur-
ing walking. IEEE ASME Trans Mechatron 2011;16(5):799–807.
precautionary measure. [15] Lahuec C, Almouahed S, Arzel M, Gupta D, Hamitouche C, Jezequel M, et al.
Upon implementation of the selected rotary stepper micromo- A self-powered telemetry system to estimate the postoperative instability of a
tor [27] on the prototype, it would be possible to drive the actua- knee implant. IEEE Trans Biomed Eng 2011;58(3):822–5.
[16] Collo A, Poignet P, Hamitouche C, Almouahed S, Stindel E. A miniaturised actu-
tion system under more realistic conditions, as well as to investi- ation system embedded in an instrumented knee implant for postoperative lig-
gate the wireless power supply and control issues. Contextually, a ament imbalance correction. In: Proceedings of the 36th international confer-
force sensor could be employed to evaluate the total force devel- ence on IEEE engineering in medicine and biology society (EMBC 2014); 2014.
p. 6211–14.Chicago, Illinois, USA
oped by the mobile tibial tray during actuation and to compare it [17] Insall JN. Total knee replacement. In: Insall JN, editor. Surgery of The Knee.
to the computed theoretical value. Philadelphia, MS: Churchill Livingstone; 1984.
Finally, a fully functional instrumented tibial component in- [18] Wright JM, Crockett HC, Slawski DP, Madsen MW, Windsor RE. High tibial os-
teotomy. J Am Acad Orthop Surg 2005;13:279–89.
cluding both the PC elements and the actuation system sub-parts
[19] Mundermann A, Dyrby CO, D’Lima DD, Colwell CW, Andriacchi TP. In vivo knee
could be manufactured so as to evaluate the proposed implant as loading characteristics during activities of daily living as measured by an in-
a whole. strumented total knee replacement. J Orthop Res 2008;26(9):1167–72.
[20] ISO standard 14243, implants for surgery - wear of total knee-joint prosthe-
ses - part 1: Loading and displacement parameters for wear-testing machines
References
with load control and corresponding environmental conditions for test. Inter-
national Organisation for Standardization; 2009.
[1] Scuderi GR, Tria AJ. Knee arthroplasty handbook: techniques in total knee and [21] Vail TP, Lang JE. Surgical techniques and instrumentation in total knee arthro-
revision arthroplasty. New York, NY: Springer; 2006. plasty. In: Scott WN, editor. Surgery of the knee. Philadelphia, MS: Churchill
[2] Zimmermann F, Schwenninger C, Nolten U, Firmbach FP, Elfring R, Livingstone; 2006.
Radermacher K. A new approach to implant alignment and ligament balanc- [22] Ford H, Alexander JM. Advanced Mechanics of Materials. London, United King-
ing in total knee arthroplasty focussing on joint loads. Biomed Tech (Berl) dom: Longmans; 1963.
2012;57(4):283–91. [23] Bloemker KH, Guess TM, Maletsky L, Dodd K. Computational knee ligament
[3] Marmignon C, Leimnei A, Cinquin P. Robotized distraction device for knee re- modeling using experimentally determined zero-load lengths. Open Biomed
placement surgery. Int Congr 2004;1268:638–43. Eng J 2012;6:33–41.
[4] Marmignon C, Leimnei A, Lavallee S, Cinquin P. Automated hydraulic [24] Delport H, Labey L, DeCorte R, Innocenti B, VanderSloten J, Bellemans J. Collat-
tensor for total knee arthroplasty. Int J Med Robot Comput Assist Surg eral ligament strains during knee joint laxity evaluation before and after TKA.
2005;1(4):51–7. Clin Biomech 2013;28:777–82.
[5] Crottet D, Maeder T, Fritschy D, Bleuler H, Nolte LP, Pappas IP. Development of [25] Matweb. Titanium vanadium alloy properties. http://www.matweb.com/search/
a force amplitude- and location-sensing device designed to improve the liga- DataSheet; 2015. [Online; accessed 13-May-2015].
ment balancing procedure in TKA. IEEE Trans Biomed Eng 2005;52(9):1609–11. [26] Almouahed S. Study, implementation and evaluation of a prototype of self-
[6] Hardwick ME, Pulido PA, D’Lima DD, Colwell CW Jr. e-Knee: the electronic powered diagnostic knee implant, Brest: Telecom Bretagne; 2011. Ph.D. thesis.
knee prosthesis. Orthop Nurs 2006;25(5):326–9. [27] Faulhaber. Miniature drive systems - adm0620-2r-v2-05 and precision gear-
[7] Arami A, Simoncini M, Atasoy O, Hasenkamp W, Bertsch A, Meurville E, et al. head 06/1 k 1024:1. http://www.micromo.com/ecatalog/page245.html and
Instrumented knee prosthesis for force and kinematics measurements. IEEE http://www.micromo.com/ecatalog/page294.html; 2015. [Online; accessed 12-
Trans Autom Sci Eng 2013;10:615–24. May-2015].

Вам также может понравиться