LCR-800
USER MANUAL
GW INSTEK PART NO. 82CR-81900MK1
Safety Guidelines
CAUTION
Do not place any heavy object on the LCR-800.
Avoid severe impact or rough handling that
leads to damaging the LCR-800.
This chapter contains important safety instructions Do not discharge static electricity to the LCR-
that you must follow when operating or storing 800.
the LCR-800. Read the following before any Do not block or obstruct the cooling fan vent
operation to insure your safety and to keep the opening.
LCR-800 in the best possible condition.
Do not perform measurement at circuits directly
connected to Mains (Note below).
Safety Symbols Do not disassemble the LCR-800 unless you are
These safety symbols may appear in this manual or on the LCR-800. qualified as service personnel.
(Measurement categories) EN 61010-1:2001 specifies the
measurement categories and their requirements as follows. LCR-800
Warning: Identifies conditions or practices that falls under category I.
WARNING could result in injury or loss of life. Measurement category IV is for measurement performed at the
source of low-voltage installation.
Caution: Identifies conditions or practices that Measurement category III is for measurement performed in the
CAUTION could result in damage to the LCR-800 or to other building installation.
properties. Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
DANGER High Voltage Measurement category I is for measurements performed on
circuits not directly connected to Mains.
5 6
SAFETY INSTRUCTIONS LCR-800 User Manual
To ensure fire protection, replace the fuse only Storage Location: Indoor
with the specified type and rating. environment
Relative Humidity: < 85%
Disconnect the power cord before fuse Temperature: −20°C to 60°C
replacement.
Make sure the cause of fuse blowout is fixed Disposal Do not dispose this instrument as unsorted
before fuse replacement. municipal waste. Please use a separate collection
facility or contact the supplier from which this
Cleaning LCR-800 Disconnect the power cord before cleaning. instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
Use a soft cloth dampened in a solution of mild
reduce environmental impact.
detergent and water. Do not spray any liquid.
Do not use chemical or cleaner containing harsh
material such as benzene, toluene, xylene, and
acetone.
7 8
SAFETY INSTRUCTIONS LCR-800 User Manual
GETTING STARTED
When using the LCR-800 in the United Kingdom, make sure the
power cord meets the following safety instructions.
9 10
GETTING STARTED LCR-800 User Manual
11 12
GETTING STARTED LCR-800 User Manual
13 14
GETTING STARTED LCR-800 User Manual
O_B 13
O_B 12
O_B 11
O_B 10
O_B 9
O_B 8
O_B 7
O_B 6
O_B 5
O_B 4
O_B 3
O_B 2
1
IN_
IN_
IN_
IN_
IN_
IN_
IN_
IN_
IN_
IN_
IN_
IN_
IN_
O_B
IG
_LO
_HI
O_P J
O_I OM
D
VCC
O_P GND
ER
ER
I_E CK
O_S X
_RE
NDE
GN
_TR
_LO
_OV
_OV
O_E
O_P
I_K
O_S
Turns the power on or off.
Power Switch LCR-829, LCR-827, LCR-826 LCR-821
On
Contrast control The LCD contrast control
Off
15 16
GETTING STARTED LCR-800 User Manual
17 18
GETTING STARTED LCR-800 User Manual
Counter- Clockwise:
clockwise: dark
Fixture Connection
bright
Fixture structure
19 20
GETTING STARTED LCR-800 User Manual
H side
Fixture connection
21 22
GETTING STARTED LCR-800 User Manual
Zeroing
Zeroing calibration
Power + +
Background Open and short circuit calibration (zeroing) should
LCD
23 24
GETTING STARTED LCR-800 User Manual
Warning If the test failed, ensure your cables or test fixtures are
open and not shorted. Ensure R.H is OFF. After
Component Measuring Guidelines
inspection try again.
Short circuit The short test will calibrate the short circuit Background For measuring Impedance, Capacitance,
impedance of the cables or test fixtures. This is Inductance, and Resistance, series or parallel
used for low impedance measurements equivalent circuit models are available. Usually a
component manufacturer will specify how a
4. Short the cables or test fixtures component should be measured and at what
using a short thick copper wire frequency. If not, use the guidelines below. Select
if necessary. the equivalent circuit and frequency according to
the component value. For more information about
5. Press R/L offset in the offset menu. equivalent circuit models and theory see page 124.
25 26
GETTING STARTED LCR-800 User Manual
27 28
BASIC MEASUREMENT LCR-800 User Manual
BASIC MEASUREMENT
In general, two measurement items, primary and secondary, are
combined in a single measurement. The following table shows the
available combinations. Details of the measurement modes and the
circuit theory and formula can be found in the appendix, page 124.
29 30
BASIC MEASUREMENT LCR-800 User Manual
31 32
BASIC MEASUREMENT LCR-800 User Manual
Panel operation 1. From the main menu, press the Set Bias voltage
MODE menu key to cycle between
the different modes.
*Only the LCR-821 can select L/R and Z/θ Background Voltage bias can be set internally or externally. An
Note measurement modes. internal voltage bias of 2 volts is normally applied
to a DUT. External voltage bias is able to accept 0
to 30 volts with a maximum current of 200mA. For
Select Equivalent Circuit Type external bias voltage connections see page 22.
When measuring a DUT, please allow 1 second to
stabilize a DUT after a bias voltage is applied. In
Background Series or Parallel equivalent circuits can be general a bias voltage should only be applied to
selected. Not all measurement modes can be used capacitors. If a bias voltage is applied to devices
with both series and parallel equivalent circuits. with low impedance, inaccurate measurements
For details about circuit types see the circuit theory will occur.
chapter on page 124.
When an external voltage is applied, constant voltage
Measurement Series Parallel Note mode (C.V.ON) must be enabled, page 39.
type
Capacitance (C)
Inductance (L)
Impedance (Z)
Resistance (R)
33 34
BASIC MEASUREMENT LCR-800 User Manual
Panel operation 1. Press the 7/Bias key on the number The LCR-821 can provide 504 different frequencies
pad to cycle from internal to with a 5 digit resolution including decimal places.
external bias. The bottom of the Any frequency can be keyed from the number pad,
screen will display internal or and the closest available frequency (of 504) will be
external bias. selected automatically. The LCR-818/829 has 503
different frequencies and the LCR-817/827 and
LCR-816/826 have 489 and 245, respectively.
35 36
BASIC MEASUREMENT LCR-800 User Manual
1.1kHz 1V
Background Along with frequency, voltage can be set. Make 3. Press (F5) EXIT to exit the Setting
sure the appropriate voltage is selected, according menu.
to the component characteristics.
Range 5mV ~ 1.275V (5mV steps) <200kHz
100mV ~ 1.275 (5mV steps) @200kHz Set PPM for D/Q measurements
37 38
BASIC MEASUREMENT LCR-800 User Manual
C.V ON / OFF is toggled each time the 1/C.V Panel operation 1. From the main menu, press MENU, followed
button is pressed. by SETTING and AVGE.
39 40
BASIC MEASUREMENT LCR-800 User Manual
41 42
BASIC MEASUREMENT LCR-800 User Manual
43 44
BASIC MEASUREMENT LCR-800 User Manual
45 46
LCR-800 User Manual
BIN FUNCTIONS
The Handler interface is used to sort components
3. When the status bar has completed, into different bins. The handler menu compares
the calibration settings are recalled. results from a number of different user defined limits.
Component sorting can be accomplished in either
4. Press EXIT to exit to the main manual or automatic mode. For more information on
menu. using the handler interface to sort components please
see page 114.
If the function keys are not active after calibration
CAUTION settings have been recalled, DO NOT turn off the Handler Menu Overview .............................................50
Bin Functions
instrument. Wait a few minutes and try again. Handler Menu .............................................................52
Mode Setting ...............................................................52
Circuit Setting ..............................................................53
Speed Setting ...............................................................53
Display Setting .............................................................54
Frequency Setting ........................................................54
Select/Run Auto/Manu Sorting ..................................55
Voltage Setting.............................................................56
Bias Setting ..................................................................57
Constant Voltage Setting.............................................58
Delay Setting ................................................................59
Average Setting ............................................................60
47 48
BIN FUNCTIONS LCR-800 User Manual
Mode Setting The mode setting area shows basic settings for the
current bin mode.
SET BIN Configures the Bin settings
MODE Measurement mode
SPEED Measurement speed
BIN SUM Displays the Bin test results
CIRCUIT Selects between serial and parallel
circuits
DISPLAY Selects what measurement unit is
displayed.
Parameter BIN, VALUE, OFF
49 50
BIN FUNCTIONS LCR-800 User Manual
Parameter 1,2,3,4
Delay Delay between each measurement Handler Menu
Parameter 0~99999 ms
MANU/ Selects between automatic and manual Background Before Bin Sorting, the measurement settings must
AUTO mode be configured.
51 52
BIN FUNCTIONS LCR-800 User Manual
Background Use Circuit setting to change the equivalent circuit. Background Use the Display setting to change the measurement
results as values or bins.
Panel Operation 1. Use the arrow menu keys (F1/F2)
to move the cursor to CIRCUIT. Panel Operation 1. Use the arrow menu keys (F1/F2)
to move the cursor to Display.
Background Use the Speed setting to change the measurement Frequency Setting
speed.
Background Set the testing frequency.
Panel Operation 1. Use the arrow menu keys (F1/F2)
to move the cursor to SPEED.
Panel Operation 1. Use the arrow menu keys (F1/F2)
to move the cursor to F (Frequency)
53 54
BIN FUNCTIONS LCR-800 User Manual
Panel operation 1. Hold the START key for a few Voltage Setting
seconds to toggle from automatic or
manual bin sorting. Background Set the testing voltage.
55 56
BIN FUNCTIONS LCR-800 User Manual
Background Set internal or external bias voltage. Background The range can be selected from 1 to 4. Different
ranges should be used for different components
Panel Operation 1. Use the arrow menu keys (F1/F2) and component values and to ensure accurate
to move the cursor to INT.B or readings.
EXT.B. Component
Inductor Capacitor Resistor
Range1 1~16mH/f 1.6~25uF/f 6.25~100Ω
Range2 16~256mH/f 100~1600nF/f 0.1~1.6kΩ
Range3 256~4100mH/f 6.4~100nF/f 1.6~25.6kΩ
2. Press INT (F3) to use internal Range4* 4.1~65H/f 400~6400pF/f 25.6~410kΩ
biasing. f = test frequency in kHz
*
This range is not used above 20 kHz
3. Press EXT (F4) to use external
biasing. Panel Operation 1. Use the arrow menu keys (F1/F2)
to move the cursor to RANGE.
4. Use the arrow menu keys to
highlight OFF/ON.
57 58
BIN FUNCTIONS LCR-800 User Manual
3. Press OFF (F4) to turn constant Background The average function chooses how many averages
voltage off. (1-255) are used for each measurement.
Delay Setting Panel Operation 1. Use the arrow menu keys (F1/F2)
to move the cursor to AVERAGE.
59 60
BIN FUNCTIONS LCR-800 User Manual
61 62
BIN FUNCTIONS LCR-800 User Manual
4. Press SET (F3). Panel operation 1. Move the cursor to TOT_BIN in the
Bin menu.
5. The Bin menu appears.
Sort Type
Set Nominal Value
2. Press F3 to switch from primary or 2. Use the number pad to enter a nominal value
secondary sorting. for the current sort bin.
R↔Q, C↔D, C↔R, L↔Q
63 64
BIN FUNCTIONS LCR-800 User Manual
For
example:
20 Ω. 2. Use the number pad to enter the maximum
percentage value for the current sort bin.
For
Set Max/Min Absolute Limit example:
10%.
Background The maximum and minimum absolute limits of the
3. Repeat the above procedure for -%.
current bin can be set. The limit units depend on
the measurement type, see Sort Type, page 63.
Panel operation 1. Move the cursor to MAX to set the Set Max/Min Secondary Measurement Limits
absolute maximum limit.
65 66
BIN FUNCTIONS LCR-800 User Manual
Background All the bin settings can be cleared for all the bins. Bin Summary Menu Overview
Ω Ω F1
4. Press F2(YES ->) to confirm the
clear or press F1(NO->) to cancel.
Or
Bin Results
67 68
BIN FUNCTIONS LCR-800 User Manual
69 70
BIN FUNCTIONS LCR-800 User Manual
71 72
RS232 REMOTE LCR-800 User Manual
Background RS232 must first be enabled on the LCR-800 before Background LCR-Viewer mimics the LCR-800 series front panel
trying to connect with a PC. and operates in a similar manner.
Panel operation 1. From the main menu, press MENU, SORT Menu Bar Virtual Panel Meter Time
AND RS232.
RS232 INTERFACE: ON ON
OFF
OFF
73 74
RS232 REMOTE LCR-800 User Manual
75 76
RS232 REMOTE LCR-800 User Manual
77 78
RS232 REMOTE LCR-800 User Manual
4. To run a measurement in
manual mode, click the start
button.
5. To stop measuring in
Auto mode, right click the
start button and click the
AUTO/MANU pop-up
button.
79 80
RS232 REMOTE LCR-800 User Manual
View Data
View Help
For more information on the way the files store test
results see page 78. Background The Help menu is to view the software version and
copyright information
1. Ensure the LCR-800 is set to Page 80
manual (single) measurement
Operation 1. Ensure the LCR-800 is set to Page 80
mode.
manual (single) measurement
mode.
Operation 2. To view the test result data, go to
the Dataresult menu.
2. Go to the HelpAbout menu.
81 82
RS232 REMOTE LCR-800 User Manual
Terminal Connection
Note All file menus (File, Option, Data, Help) are restricted
in Auto mode. To change to manual mode see page 80
to change to Manual mode remotely.
83 84
RS232 REMOTE LCR-800 User Manual
6. Connect to the terminal program Disconnection 10. To disconnect remote control send
with the following configuration the following command with
settings: ^END^M or ^J^M as the terminal
COM port (as per PC) character.
Terminal command: COMU:OFF.
Baud rate- 38400
LCR Return: COMU:OFF.
Data bits- 8
Stop bit- 1
Parity- none
Flow control- none
85 86
PROGRAMMING LCR-800 User Manual
PROGRAMMING
Below are examples of ASCII and hexadecimal
commands
ASCII SORT:NOMV +32.0000< ^END^M> or <^J^M>
Hex 53 4F 52 54 3A 4E 4F 4D 56 20 2B 33 32 2E 30 30 30 30
Command overview lists all the LCR-800 0A 0D
commands and command queries. The command
syntax section shows you the basic rules you have
to apply when using commands. Command format 1: command header
2: single space
Command Syntax 3: parameter
4: message terminator
Command There are a number of different instrument Parameter Type Description Example
Background commands and queries. A command sends
instructions or data to the LCR meter and a query <string> Character string SLOW
receives data or status information from the LCR <NR1> Integers 0, 1, 2, 3
meter. Measurements are automatically sent when
<variable> number data 0.1, 3.14, 8.5
a measurement is made in manual or automatic
mode. Message <NL^END> New line or ASCII line feed
terminators Or character (HEX 0A)
Command Types
<NL^J>
Command Two or more commands
separated by a colon (:) <CR^M> Carry return character (Hex 0D)
with/without a parameter Input | Output The format of the input and output values differ
Example MEMO:STOR 100.<^END^M> value differences somewhat when dealing with positive values. The
total amount of characters used for each variable
Query A query is a compound
depends on the command/query.
command followed by a question
mark (?). A parameter (data) is Number ASCII HEX
returned. Input 1.0000 +1.0000 2B 31 2E 30
Example SORT:NOMV?<^END^M> 30 30 30
Measurement Returns measurement data. Can Output 1.0000 sp1.0000 20 31 2E 30
be manually or automatically 30 30 30
updated. Input -1.0000 -1.0000 2D 31 2E 30
Example MAIN:PRIM 32.705<^END> 30 30 30
87 88
PROGRAMMING LCR-800 User Manual
89 90
PROGRAMMING LCR-800 User Manual
<string> Display
VALU Unit Value
SPEED Command/Query DELP Delta %
The speed command sets the measurement speed of the instrument. DELT Delta
The faster the measurement speed the lower the accuracy. This Example MAIN:DISP:VALU<^END^M>
command also queries the current measurement speed.
Set the display to Value
Syntax MAIN:SPEE:<string><^END^M>or<^J^M> Query Syntax MAIN:DISP?<^END^M>or<^J^M>
Parameter Return String
<string> Speed <string> Display
SLOW Slow MAIN:DISP:VALU<^END> Value
MEDI Medium MAIN:DISP:DELP<^END> Delta %
FAST Fast MAIN:DISP:DELT<^END> Delta
Example MAIN:SPEE:SLOW<^END^M> Query Example MAIN:DISP?<^END^M>
Set the measurement speed to slow. MAIN:DISP:VALU<^END>
Query Syntax MAIN:SPEE?<^END^M>or<^J^M> Currently the display is set at value.
Return String
<string> Speed MODE Command/Query
MAIN:SPEE:SLOW<^END> Slow The mode command sets the measurement mode of the LCR-800.
MAIN:SPEE:MEDI<^END> Medium Syntax MAIN:MODE:<string><^END^M>or<^J^M>
MAIN:SPEE:FAST<^END> Fast Parameter
Query Example MAIN:SPEE?<^END^M> <string> Primary Measurement Secondary Measurement
MAIN:SPEE:MEDI<^END>
RQ Resistance Quality factor
Medium measuring speed is returned.
CD Capacitance Dissipation factor
CR Capacitance Resistance
DISPLAY Command/Query
LQ Inductance Quality factor
The display command sets the displayed measurement as a value or LR* Inductance Resistance
as an offset from a nominal value (Delta or Delta%) ZQ* Impedance Angle
Syntax MAIN:DISP:<string><^END^M>or<^J^M> *For the LCR-821 only
Parameter
91 92
PROGRAMMING LCR-800 User Manual
MAIN:CIRC:PARA<^END> Parallel
Example MAIN:MODE:RQ<^END^M> Query Example MAIN:CIRC?<^END^M>
Sets the mode to R/Q (Resistance/Quality factor) MAIN:CIRC:PARA<^END>
Returns a parallel equivalent circuit as the current
Query Syntax MAIN:MODE?<^END^M>or<^J^M>
setting.
Return String
<string> Current measurement FREQUENCY Command/Query
mode
MAIN:MODE:RQ<^END> R/Q Set or queries the test frequency.
MAIN:MODE:CD<^END> C/D Syntax MAIN:FREQ <variable><^END^M>or<^J^M>
MAIN:MODE:LQ<^END> L/Q Parameter
MAIN:MODE:LR<^END> L/R <variable> Frequency (kHz)
MAIN:MODE:ZQ<^END> Z/Q 0.01200~100.000 (7 characters, including a 12 Hz~100kHz
decimal)
Query Example MAIN:MODE?<^END^M>
MAIN:MODE:RQ<^END> Example MAIN:FREQ 0.01200<^END^M>
Returns the current measurement mode as R/Q Sets the frequency to 12Hz (0.012 kHz)
Query Syntax MAIN:FREQ?<^END^M>or<^J^M>
CIRCUIT Command/Query Return String
The mode command sets the equivalent circuit to series or parallel. <string> Frequency
MAIN:FREQ < variable ><^END> Returns the test frequency
Syntax MAIN:CIRC:<string><^END^M>or<^J^M>
(<variable >=0.01200~100.000) in kHz.
Parameter
Query Example MAIN:FREQ?<^END^M>
<string> Equivalent Circuit MAIN:FREQ 0.01200<^END>
SERI Series Returns the current test frequency in kHz (12 Hz).
PARA Parallel
Example MAIN:CIRC:SERI<^END^M> VOLTAGE Command/Query
Sets the equivalent circuit to series
Set or queries the test signal voltage.
Query Syntax MAIN:CIRC?<^END^M>or<^J^M>
Syntax MAIN:VOLT < variable ><^END^M>or<^J^M>
Return String
Parameter
<string> Equivalent circuit
< variable > Test signal voltage
MAIN:CIRC:SERI<^END> Series
93 94
PROGRAMMING LCR-800 User Manual
0.005~1.275 (5 characters, including a 5mV~1.275 Returns Auto mode as the current measurement
decimal) mode.
Example MAIN:VOLT 0.005<^END^M>
Sets the test signal voltage to 5mV. START Command
Query Syntax MAIN:VOLT?<^END^M>or<^J^M>
Starts a measurement in manual mode.
Return String
Syntax MAIN:STAR<^END^M>or<^J^M>
<string> Voltage
Example MAIN:STAR<^END^M>
MAIN:VOLT :< variable ><^END> Returns the test voltage.
Starts the measurement
(<variable >= 0.005~1.275)
Query Example MAIN:VOLT?<^END^M>
MAIN:VOLT 0.005<^END> RANGE HOLD Command/Query
Returns the test voltage (5mV)
Turns range hold on or off or queries the range hold status.
Syntax MAIN:R.H.:<string><^END^M>or<^J^M>
AUTO/MANU Command/Query
Parameter
Sets automatic or manual measurement mode. <String> Range hold
Syntax MAIN:TRIG:<string><^END^M>or<^J^M> OFF. Off
Parameter ON.. On
<String> Test mode Example MAIN:R.H.:OFF.<^END^M>
AUTO Automatic mode Turn range hold off
MANU Manual mode Query Syntax MAIN:R.H.?<^END^M>or<^J^M>
Example MAIN:TRIG:MANU<^END^M> Return String
Sets the measuring mode to manual <string> Range Hold status
Query Syntax MAIN:TRIG?<^END^M>or<^J^M> MAIN:R.H.:OFF.<^END> Range hold is off
Return String MAIN:R.H.:ON..<^END> Range hold is on
<string> Voltage Query Example MAIN:R.H.?<^END^M>
MAIN:TRIG:AUTO<^END> Returns automatic mode MAIN:R.H.:ON..<^END>
Returns the Range Hold status (On)
MAIN:TRIG:MANU<^END> Returns manual mode
Query Example MAIN:TRIG?<^END^M>
MAIN:TRIG:AUTO<^END>
95 96
PROGRAMMING LCR-800 User Manual
97 98
PROGRAMMING LCR-800 User Manual
Example MAIN:EXTB:OFF.<^END^M>
Turn External Bias off. OPEN Command
Query Syntax MAIN:EXTB?<^END^M>or<^J^M> This command will perform an open circuit calibration. A return
Return String string will indicate if the calibration was successful or not.
<string> External Bias status Syntax OFFS:OPEN<^END^M>or<^J^M>
MAIN:EXTB:OFF.<^END> Off Return String
MAIN:EXTB:ON..<^END> On
<string> Open calibration attempt
Query Example MAIN:EXTB?<^END^M>
OPEN:OK<^END> Successful
MAIN:EXTB:ON..<^END>
OPEN:FAIL<^END> Failure
Returns the External Bias status (On).
Example OFFS:OPEN<^END^M>
OPEN:OK<^END>
PPM Command/Query Returns the open circuit calibration attempt
(Successful).
Turns PPM on or off for Dissipation (D) or Quality factor (Q)
measurements.
SHORT Command
Syntax MAIN:PPM.:<string><^END^M>or<^J^M>
Parameter This command will perform a closed (short) circuit calibration. A
return string will indicate if the calibration was successful or not.
<String> PPM
Syntax OFFS:SHOR<^END^M>or<^J^M>
OFF. Off
Return String
ON.. On
<string> Short calibration attempt
Example MAIN:PPM.:OFF.<^END^M>
SHOR:OK<^END> Successful
Turns PPM off.
SHOR:FAIL<^END> Failure
Query Syntax MAIN:PPM.?<^END^M>or<^J^M> OFFS:SHOR<^END^M>
Example
Return String SHOR:OK<^END>
Returns the closed circuit calibration attempt
<string> PPM status
(Successful).
MAIN:PPM.:OFF.<^END> Off
MAIN:PPM.:ON..<^END> On
Query Example MAIN:PPM.?<^END^M>
NOM.VAL Command/Query
MAIN:PPM.:ON..<^END>
Sets or queries the nominal value. The nominal value unit depends
Returns PPM status (On). on the measurement mode.
Syntax SORT:NOMV<variable><^END^M>or<^J^M>
99 100
PROGRAMMING LCR-800 User Manual
101 102
PROGRAMMING LCR-800 User Manual
Example STEP:RECA<^END^M>or<^J^M>
RECA:OK<^END>
AVERAGE Command/Query
Calibration was recalled successfully
Sets the average number from 1~255. The average number indicates
how many test samples are used to create an averaged test result.
BAUD RATE Command
Syntax STEP:AVER <variable><^END^M>or<^J^M>
Sets the baud rate of the RS232 connection.
Parameter
Syntax COMU:<value><^END^M>or<^J^M>
<variable> Average number
1.00~255. (integer values) 1~255 Parameter
Recalls the calibration settings from memory. A return string This query returns the model number of the LCR-800.
indicates if the command was successful. Query Syntax COMU:MONO<^END^M>or<^J^M>
Syntax STEP:RECA<^END^M>or<^J^M> Return String
Return String <string> Model number
<string> Recall calibration COMU:MONO:816.<^END> LCR-816
RECA:OK<^END> Successful COMU:MONO:817.<^END> LCR-817
103 104
PROGRAMMING LCR-800 User Manual
105 106
PROGRAMMING LCR-800 User Manual
107 108
PROGRAMMING LCR-800 User Manual
Secondary measurement result is returned along with the primary The secondary measurement result exceeds the range.
unit and secondary unit (C/R, L/R only). This measurement is the kΩ is the secondary unit and nF is the primary unit.
second measurement displayed after measurements have been
started.
INITIATION HAS FINISHED (Initiate) Command
Return Syntax MAIN:SECO <value><unit1><unit2><^END>
Initiates the RS232 connection. A string is returned when the
<value> Test result
initiation has been completed.
Any 6 digit ASCII character including sp(+) or Secondary measurement
– characters and a decimal point. value Syntax COMU:OVER<^END^M>or<^J^M>
109 110
INTERACE LCR-800 User Manual
INTERACE
Configure RS-232 interface
111 112
INTERACE LCR-800 User Manual
PC LCR Meter
Connection
DB9 Pin Signal Signal DB9 Pin Handler interface
2 RxD TxD 3
Connection
3 TxD RxD 2 Connect the male DSUB 25 pin
cable to the Handler interface
4 DTR DSR, DCD 6,1 socket.
5 GND GND 5
Pin assignment HANDLER INTERFACE
6,1 DSR, DCD DTR 4
7 RTS CTS 8
13
12
O_ _11
10
O_ _9
O_ _8
7
O_ _6
O_ _5
3
O_ _2
1
N_
N_
N_
N_
N_
N_
N_
N
N
8 CTS RTS 7
BI
BI
BI
BI
BI
BI
BI
BI
BI
BI
BI
BI
BI
O_
O_
O_
O_
O_
O_
O_
M
O_ EJ
I_E CK
O_ G
O_ EX
O_ O
HI
ER
D
R
GN
E
VC
GN
L
RI
EO
P_
R
D
P_
OV
OV
O
_T
S_
IN
_L
P_
S_
O_
I_K
O_
O_
Signal Function
Pin1 /O_BIN_1 Go, Assigned BIN 1
Pin2 /O_BIN_2 Go, Assigned BIN 2
Pin3 /O_BIN_3 Go, Assigned BIN 3
Pin4 /O_BIN_4 Go, Assigned BIN 4
Pin5 /O_BIN_5 Go, Assigned BIN 5
Pin6 /O_BIN_6 Go, Assigned BIN 6
Pin7 /O_BIN_7 Go, Assigned BIN 7
Pin8 /O_BIN_8 Go, Assigned BIN 8
Pin9 /O_BIN_9 Go, Assigned BIN 9
Pin10 /O_BIN_10 Go, Assigned BIN 10
113 114
INTERACE LCR-800 User Manual
115 116
INTERACE LCR-800 User Manual
117 118
INTERACE LCR-800 User Manual
Timing Characteristics
Trigger Pulse Width T1 MIN MAX
5us ~
Measurement start delay T2 MIN MAX
time 140us ~
/O_EOM Delay Time After T3 MIN MAX
Data Output 5us ~
Calculation and binning time T4 MIN MAX
6ms ~
119 120
FAQ LCR-800 User Manual
FAQ
Q1. What is the correct procedure for Open/Short Zeroing when using
the LCR-06A test fixture?
A1. The LCR-06A test fixture is very sensitive and thus must be used
correctly.
For Open Zeroing, make sure that the test fixture wires do not
move and that there is nothing in close proximity to the test clips.
For Short Zeroing ensure the clips are properly shorted. See page
24 for details.
A2. There are two possible reasons that Short Zeroing can fail.
The test fixture has an open circuit between the wires and
terminal.
Some functions can impede the short test. Ensure Range Hold and
Internal/External Bias (R.H and INT.B/EXT.B) are disabled. See
pages 40, 39 & 34.
A3. Use the display contrast control on the rear panel to adjust the
contrast.
A4. Make sure the correct terminal characters are used. For example
use “CTRL J” ”CTRL M” as the <^J^M> message terminator in a
terminal session.
121 122
APPENDIX LCR-800 User Manual
CS CP 1 D 2
CP
CS
D=dissipation factor
1 D2
D=dissipation factor
2. Replace the fuse in the holder.
Inductance (L) Series diagram Parallel diagram
123 124
APPENDIX LCR-800 User Manual
RS
RP
RP RS 1 Q 2
1 Q2 Q=quality factor Note
Conductance is only shown for its relation to
Resistance, Conductance is not a measurable feature of
Q=quality factor the LCR-800 series.
DC Resistance Rdc
Formula V 1 I 1
R Z S jX GP YP jB
I G V R
j j
Z S j L Z S YP jC YP
C L
ZS R 2
X2 YS
GB
RX
G 2
B2
ZP
R 2
X2 YP G 2
B2
RS Z cos GP Y cos
125 126
APPENDIX LCR-800 User Manual
Background Capacitance measures the amount of electronic Background Reactance measures the imaginary part of
charge stored between two terminals. Impedance (Z) caused by capacitors or inductors.
Susceptance is the reciprocal of Reactance and
Type Series Capacitance CS Parallel Capacitance CP measures the imaginary part of Admittance (Y),
which is the reciprocal of Impedance.
Formula j YP G jC Reactance and Susceptance is only shown for their
ZS R Note
C GP
relation to impedance. Reactance and Susceptance are
Q C P RP D not measurable features of the LCR-800 series.
1 C P
Q
C S RS Type Series Reactance (XS) Parallel Susceptance (BP)
D C S RS
Formula 1 1
X Z sin B Y sin
B X
Inductance (L) Formula
ZS R 2
X2 YS
GB
127 128
APPENDIX LCR-800 User Manual
Impedance (Z) and Admittance (Y = 1/Z) Formula Quality factor (Q) and Dissipation factor (D) Formula
Background Impedance measures the total amount of Background Both Quality factor and its reciprocal, Dissipation
opposition between two terminals in an AC circuit. factor, are used for measuring the rate of energy
Admittance is the reciprocal of Impedance and dissipation relative to the measurement frequency.
measures how easily the electricity flows in an AC Low energy dissipation: high Q, low D
circuit.
High energy dissipation: low Q, high D
Admittance is only shown for its relation to
Note impedance. Admittance is not measurable with the
Type Quality factor (Q) Dissipation factor (D)
LCR-800 series.
RX
G 2
B2
ZP
R 2
X 2 YP G 2
B2
RS Z cos GP Y cos
X S Z sin BP Y sin
129 130
APPENDIX LCR-800 User Manual
Background The Angle (θ) measures the phase on which Primary Measurement Readout Error Formula
Impedance (Z), Admittance (Y), Quality factor (Q), C 2 counts±0.03%+0.02%[(1+Ka)# or (X/Ymax )# or
and Dissipation factor (D) are measured. (Ymin/X)# ] (1+│D│)(1+Kb+Kc)
R 2 counts±0.03%+0.02%[(1+Ka)# or (X/Ymax )# or
Type Angle (θ) (Ymin/X)# ] (1+│Q│)(1+Kb+Kc)
L 2 counts±0.03%+0.02%[(1+Ka)#or(X/Ymax)#or(Ymin/X)#
Formula Z S R jX YP G jB ] (1+ 1 /│Q│)(1+Kb+Kc)
|Z| Ze = Depends on whether the component is a
j j capacitor(C), resistor(R) or inductor(L):
R j L R G j C G
C L Circuit Formula for relevant circuit
C Ze = 2 counts±
1 1 1
D tan 90 0.03%+0.02%[(1+Ka)#or(X/Ymax)#or(Ymin/X)# ]
Q
tan 90 D
Q (1+│D│)(1+Kb+Kc)
R Ze = 2 counts±
RS Z cos GP Y cos 0.03%+0.02%[(1+Ka)#or(X/Ymax)#or(Ymin/X)# ]
(1+│Q│)(1+Kb+Kc)
X S Z sin BP Y sin L Ze=2 counts±
0.03%+0.02%[(1+Ka)#or(X/Ymax)#or(Ymin/X)#]
(1+ 1 /│Q│)(1+Kb+Kc)
131 132
APPENDIX LCR-800 User Manual
133 134
APPENDIX LCR-800 User Manual
Table3 Specifications
Y Range constant- Range Hold
Specification accuracy is only applicable when the LCR meter has
Component
Inductor Capacitor Resistor been warmed up for 30 minutes with an operating temperature of
Range Max Min Max Min Max Min 18°C ~28°C.
Range1 16mH/f 1mH/f 25uF/f 1.6uF/f 100Ω 6.25Ω
Range2 256mH/f 16mH/f 1600nF/f 100nF/f 1.6kΩ 0.1kΩ
Measurement Inductance (Ls/Lp)*, Capacitance (Cs/Cp), Resistance
Range3 4100mH/f 256mH/f 100nF/f 6.4nF/f 25.6kΩ 1.6kΩ
Parameters (Rs/Rp), Dissipation (D), Quality Factors (Q),
Range4* 65H/f 4.1H/f 6400pF/f 400pF/f 410kΩ 25.6kΩ Equivalent Series Resistance (ESR) and Equivalent
f= test frequency in kHz Parallel Resistance (EPR), Impedance (|Z|), Phase
* This range is not used above 20 kHz angle of Impedance [degree] (θ).
Measurement R/Q, C/D, C/R, L/Q, Z/ θ, L/R
Table4 Modes
Y Range constant- Auto Range Display Ranges Primary Inductance 0.00001mH ~ 99999H
Component Display (L)
Inductor Capacitor Resistor Capacitance 0.00001pF ~ 99999μF
Range Max Min Max Min Max Min (C)
Auto range 65H/f** 1mH/f 25uF/f 400pF/f ** 410kΩ** 6.25Ω** Resistance 0.00001Ω ~ 99999kΩ
**: Above 20kHz, Cmin = 6.4 nF/f, and Lmax = 4100mH/f (R)
f = test frequency in kHz. Absolute of 0.00001Ω ~ 99999kΩ
Impedance
(Z)
Secondary Dissipation 0.0001 ~ 9999
Display factor (D)+
Quality 0.0001 ~ 9999
factor (Q)**
Phase angle -180.00∘ ~ 180.00∘
of
Impedance
(degree)
Equivalent 0.0001Ω ~ 9999 kΩ
Series
Resistance
(ESR)+
Equivalent 0.0001Ω ~ 9999 kΩ
Parallel
Resistance
(EPR)+
135 136
APPENDIX LCR-800 User Manual
Dissipation 1 ppm ~ 9999 ppm Delta% DELTA% shows the percent deviation
factor (D)+ of the measured L,
in ppm C, R or Z value from a saved
Quality 1 ppm ~ 9999 ppm NOMINAL VALUE. The deviation is
factor (Q)** indicated.
in ppm Delta Delta is similar to the DELTA% except
DELTA % 0.00001% ~ 99999% that the deviation is shown in suitable
*s=series, p=parallel ESR=Rs units (ohms, henries, etc.)
** with L or R Measurement Speed LCR- LCR-826/827/829
+ with C Speed 816/817/819/821
Note: Only LCR-821 has Z/θ and L/R measurement Slow 896ms Please refer to the
parameters. Medium 286ms Handler timing
If any of these quantities is negative, the “-“ negative Fast 135ms diagram on page
indicator is displayed. 119 for details.
Accuracy LCR-821/819/817 Equivalent circuit Parallel L/R, L/Q, C/D, C/R, R/Q
R, L, C, Z 0.05%(Basic) Serial L/R, L/Q, C/D, C/R, R/Q, Z/θ
D, Q 0.0005(Basic) Trigger Auto/Manual
θ 0.03˚(Basic) Average 1-255
LCR-829/827/826/816 Battery 3V-DC lithium ion (*BR-2/3A) used for memory and
R, L, C, Z 0.10%(Basic) calibration data backup. (Recommended replacement
D, Q 0.001 every three years.
Please refer to the accuracy definition *The battery should only be replaced by a GW Instek
on page132 for details. approved service center.
Basic Accuracy 0.05% LCR-821/819/817 Memory 100 blocks of memory
0.1% LCR-829/827/826/816 Display 240X128 dot matrix C.C.F.L. back lit LCD (contrast
Test Frequency LCR-821 12Hz~200kHz (504 Steps) adjustable)
LCR-819/829 12Hz~100kHz (503 Steps) Test voltage LCR-817/819/827/829/821 LCR-816/826
LCR-817/827 12Hz~10kHz (489 Steps) 5mV~1.275V (5mV steps) 0.1V~1.275V (5mV steps)
LCR-816/826 100Hz~2kHz (245 Steps) Note: When the test frequency is at 200kHz, test
Measurement Value R/Q, C/D, C/R, L/Q, Z/ θ, L/R voltage must be greater than 100mV.
displays *The resolution of primary display (L, DC bias Internal 2V
C, R or Z) is five digits. External Up to 30VDC (200mA max), tolerable
*The resolution of secondary display up to 35VDC.
(D, Q, R with C, or R with Operation Indoor use
L) is four digits. Environment Altitude up to 2000M
*The resolution of secondary display Installation category II
(θ) is 2 digits after decimal place. Pollution degree 2
Operating temperature 10°C~50°C, <85% relative
humidity
Storage -20°C~60°C
Environment
137 138
APPENDIX LCR-800 User Manual
Power Consumption
400Hz
45 Watts maximum
EC Declaration of Conformity
Fuse Slow-blow 5X20 mm, We
3A/250V UL/CSA 5TT GOOD WILL INSTRUMENT CO., LTD.
GMD No. 7-1, Jhongsing Rd., Tucheng City, Taipei County 236, Taiwan
Dimensions 330mm (W) × 149mm (H) × 437mm (D) GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
Weight 5.5kg No. 69 Lushan Road, Suzhou New District Jiangsu, China.
declare that the below mentioned products:
LCR-817/819/827/829/816/826/821
are herewith confirmed to comply with the requirements set out in
the Council Directive on the Approximation of the Law of Member
States relating to Electromagnetic Compatibility (2004/108/EC) and
Low Voltage Equipment Directive (2006/95/EC).
◎ EMC
EN 61326-1 Electrical equipment for measurement, control and
EN 61326-2-1 laboratory use –– EMC requirements 2006
Conducted and Radiated Emissions Electrostatic Discharge
CISPR11: 2003+A1: 2004+A2: 2006 IEC 61000-4-2: 2001
Class A
Current Harmonic Radiated Immunity
EN 61000-3-2: 2006 IEC 61000-4-3: 2006+A1: 2007
Voltage Fluctuation Electrical Fast Transients
EN 61000-3-3: 1995+A1: 2001+A2: IEC 61000-4-4: 2004 +Corr.1: 2006
2005 +Corr.2: 2007
Surge Immunity
——————————————
IEC 61000-4-5: 2005
Conducted Susceptibility
—————————————— IEC 61000-4-6: 2003+A1: 2004+A2:
2006
Power Frequency Magnetic field
——————————————
IEC 61000-4-8: 1993+A1: 2000
Voltage Dips/ Interrupts
——————————————
IEC 61000-4-11: 2004
◎ Safety
Low Voltage Equipment Directive 2006/95/EC
Safety Requirements
IEC/EN 61010-1:2001
139 140
INDEX LCR-800 User Manual
INDEX
Dissipation accuracy definition .............. 132
accuracy definition .............. 132 overview ............................... 129
Dissipation factor Inductance
overview ............................... 130 accuracy definition .............. 132
EN61010 overview ............................... 127
Absolute limit settings series/parallel model .......... 124 measurement category ........... 6 series/parallel model .......... 124
set bin menu ........................... 65 Caution symbol .........................5 pollution degree ...................... 7 Interface ................................. 111
Admittance Circuit settings Environment LCR viewer
overview ............................... 129 handler menu ......................... 53 operation .................................. 7 file settings ............................. 78
Angle overview ..................... 131 Circuit theory and formula .. 124 storage ...................................... 8 LCR Viewer
Auto/Manual settings Cleaning the instrument...........7 Equivalent circuit model connection ........................ 75, 84
handler menu ......................... 55 Clear bin overview ............................ 124 overview ................................. 74
Automatic mode set bin menu ........................... 67 Equivalent circuit type ........... 33 remote measurement ............ 80
basic measurement ................ 44 Command Equivalent circuits view data .......................... 81, 82
Average setting combining ............................... 89 specification ......................... 138 Limit settings
basic measurement ................ 40 format ...................................... 88 Exit set bin menu ........................... 65
Average settings forms ....................................... 88 set bin menu........................... 67 Manual measurement
handler menu ......................... 60 input/output .......................... 88 External voltage bias basic measurement ................ 43
Bais voltage .............................. 34 list ............................................ 90 connection .............................. 22 Measurement
Basic accuracy parameter ................................ 88 Features.................................... 11 accuracy definition .............. 132
specification ......................... 137 syntax ...................................... 87 Fixture basic measurement ................ 29
Basic measurement ................. 29 terminators ............................. 88 connection .............................. 21 item and combination ........... 12
combination ........................... 30 types ........................................ 87 overview ................................. 20 Measurement displays
Bias settings Conductance Frequency and voltage factor specification ......................... 137
handler menu ......................... 57 overview ............................... 126 accuracy definition .............. 133 Measurement mode ............... 33
Bias voltage Constant voltage factor table1..................................... 134 Measurement modes
specification ......................... 138 accuracy definition .............. 133 table2..................................... 134 specification ......................... 138
Bin functions ............................ 48 constant voltage setting Frequency setting Measurement parameters
Bin menu overview ................. 50 basic measurement ................ 39 basic measurement................ 35 specification ......................... 136
Bin number Constant voltage settings Frequency settings Measurement speed ............... 31
set bin menu ........................... 64 handler menu ......................... 58 handler menu ........................ 54 specification ......................... 138
Bin summary............................ 70 Declaration of conformity .... 140 Front panel overview ............. 13 Measurement units................. 32
Bin summary overview .......... 68 Delay settings Fuse Measuring
Calibration handler menu ......................... 59 replacement.......................... 123 air-cored coils ......................... 28
open circuit............................. 24 Dimensions safety instruction ..................... 6 Capacitors............................... 27
short circuit ............................ 25 specification .......................... 139 Ground symbol ......................... 5 Component cases................... 27
zeroing .................................... 24 Display Handler interface Guidlines ................................ 26
Capacitance overview ................................. 30 configuration ....................... 114 Indcutors ................................ 26
accuracy definition .............. 132 specification .......................... 138 Handler Interface iron-core inductors ................ 28
overview ............................... 127 Display ranges signal overview ................... 116 Resistors.................................. 27
141 142
INDEX LCR-800 User Manual
143 144