Академический Документы
Профессиональный Документы
Культура Документы
1
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 0.1: (a) The in phase sinusoidal oscillations and (b) the electric component of radiation
2. Wave Parameters
The amplitude A of the sinusoidal wave is shown as the length of the electric
vector at a maximum in the wave.
The period p of an electromagnetic wave is the time in seconds required for the
passage of successive maxima or minima through a fixed point in space is called the
period, p, of the radiation.
In any medium containing matter, propagation of radiation is slowed by the
interaction between the electromagnetic field of the radiation and the bound electrons
in the atoms or molecules present.
Two other important wave parameters are frequency and wavelength.
• The frequency ν of an electromagnetic wave is the number of oscillations that
occur in one second. The unit of frequency is the Hertz (Hz) or S-1.
• The wavelength λ of an electromagnetic wave is the linear distance between
two successive maxima or minima. Its SI unit is metre (m). Other unit of nanomet
(nm, 1nm = 10-9 m) and Angstrong (Ao, 1 Ao = 10-10m) are also used for very short
wavelengths.
• The velocity C of all electromagnetic wave in vacuum is determined to be the
product of its wavelength λ and velocity ν and to be equal to 2.99792x108 m/s (in air
about 0.03% less):
C = λν = 3.00 x 108 m/s
Example: Calculate the frequency of an electromagnetic having the wavelength
of 325 cm.
Solution:
2
Bùi Xuân Vững Instrumental Analysis Chemistry
II.Electromagnetic spectrum
The electromagnetic spectrum of radiation includes a numerous range of
wavelength (and also energy) in which the portion of visible light detected by
naked eyes is very small compared with the other spectrum regions. It is noted that
methods based on electromagnetic spectrum use not only visible light but also
ultraviolet, infrared, and they are called spectrophotometry methods although
naked eyes are insensitive to the latter types.
The following table lists wavelength and frequency of important spectrum
regions for analytical purpose, and also names relevant spectrophotometric
methods. The last column lists types of quantum transition of nucleus, electron,
atom, and molecule that are the foundation of different electromagnetic spectrum
techniques.
Table 0.1: Application of electromagnetic radiation in analysis
Types of Wavelength Wavenumber Types of quantum
electromagnetic range range(cm-1) transition
spectrum techniques
Gamma ray emission 0.005 – 1,4 Ao - Nucleus
3
Bùi Xuân Vững Instrumental Analysis Chemistry
4
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 1.1: Electronic, vibrational and rotational energy levels corresponding to the ground
electronic state and the first excited electron state of diatomic molecules (for example CO 2)
Figure 1.2: UV-VIS molecular absorption spectrum of some organic compounds (continous
spectrum) and atomic absorption spectrum of Na (line spectrum).
6
Bùi Xuân Vững Instrumental Analysis Chemistry
If after passing throught the cell, its power of the beam is decreased to I 1 as a
result of absorption , the UV-VIS molecular absorption of the species in the cell can
be expressed by the two following terms:
Absorbance A = log(Io/I) or
I
x 100 ( )
Transmittance T = I o
7
Bùi Xuân Vững Instrumental Analysis Chemistry
close proximity of ions to the absorber alters the molar absorbility of the latter as a
result of electrostatic interactions. The effect is lessened by dilution.
Deviation from Beer’s law also arise because ε is dependant on the refractive
index of the medium. Thus, if concentration changes cause significant alterations in
the refractive index of a solution. However, this effect is rarely significant at
concentrations less than 0.01M.
b. Effects of chemical processes
Apparent deviations from Beer’s law arise when an analyte dissociates,
associates, or reacts with a solvcent to generate a product that has a different
absorption spectrum from that of the analyte.
A typical example is the dilution of a acid-base indicator leading to the shift in
the equilibrium:
HInd H+ + Ind-
As a result, if the solution is diluted doubly, the concentrations of Hind and Ind-
are not decreased a half respectively. Vì vậy nếu pha loãng gấp đôi thì nồng độ chất
màu Hind không tương ứng giảm đi một nửa. This leads to a deviation from Bee’s
law.
8
Bùi Xuân Vững Instrumental Analysis Chemistry
9
Bùi Xuân Vững Instrumental Analysis Chemistry
10
Bùi Xuân Vững Instrumental Analysis Chemistry
11
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 1.6: (a) Two types of chromators (a) grating monochromator (b) prism monochromator
d. Photo Detectors
A photo detector is a device that converts light intensity into electric signals
that can be subsequently ampflified, manipulated, and finally converted into numbers
that are related to the magnitude of the original signal.
There are many types of photo detectors in which two most common types
used for UV-VIS spectrometers are phototubes and photomultifliers.
Phototubes:
As can be seen in Figure 1.7a, a phototube consisits of a semicylindrical
cathode and a wire anode sealed inside a evacuated transparent envelope. The concave
surface of the cathode supports a layer of photosemissive material such as a alkali
metal or metal oxide that tends to emit electrons upon being irradiated. When a
potential is applied across the electrodes, the emitted photoelectrons flow to the wire
anode, producing a current that is readily amplified and displayed or recorded.
The number of electrons ejected from a photoemissive surface is directly
proportional to the radiant power of the beam striking that surface.
Photomultiflier tubes (PMT):
As shown in Figure 7b, the photomultiflier tube (PMT) is similar in
construction to the phototube but is significantly more sensitive. Electrons emitted
from the cathode are accelerated toward a dynode (labeled 1 in the figure) maintained
at a potential 90 V more positive than the cathode. When striking the dynode surface,
each accelerated photoelectron produces several additional electrons, all of which are
then accerelated to dynode 2, which is 90 V more positive than dynode 1. Here again,
electron amplitication occurs. By the time this process has been repeated at each of the
12
Bùi Xuân Vững Instrumental Analysis Chemistry
remaining dynodes, 106 to 107 electrons have been produced for each photon; these
electrons are finally collected at the anode. So the resulting current is then further
amplified electronically and measured.
(b)
Figure 1.7: (a) Diagram of a phototube and (b) a photomultiflier tube (PMT)
Figure 1.8a shows the design of a typical single beam spectrometer, and Figure
6b shows its optical diagram. It is simple, easy to maitain and inexpensive, so it is a
very common in use throughout the wolrd.
13
Bùi Xuân Vững Instrumental Analysis Chemistry
14
Bùi Xuân Vững Instrumental Analysis Chemistry
15
Bùi Xuân Vững Instrumental Analysis Chemistry
16
Bùi Xuân Vững Instrumental Analysis Chemistry
Problems
1.1) Calculate the energy of a mole of photons corresponding to a wavelength of
300 nm.
1-2)
1-3) Calculate the absorbance of an organic dye (C =7×10 −4mol L−1), knowing hat the
molar absorptivity = 650mol L−1cm−1and that the length of the optical path of
the cell used is 2×10−2m. What would happen to the absorbance if the cell used
was of double its present thickness?
1.4) A solution containing 6.23 ppm KMnO4, had a transmittance o[ 0.195 in a 1.00-
cm cell at 520 nm. Calculate the molar ahsorptivity of KMnO4, at 520 nm.
1.5) A solution containing 5.24 mg/100 mL of A (335 g/mol) has a transmittance of
55.2% in a 1.50 cm cell at 425 nm. Calculate the molar absorptivity of A at this
wavelength.
1.6) A 5.00 × 10–4 M solution of an analyte is placed in a sample cell that has a
pathlength of 1.00 cm. When measured at a wavelength of 490 nm, the
absorbance of the solution is found to be 0.338. What is the analyte’s molar
absorptivity at this wavelength?
1.7) A compound X is to be determined by UV-visible spectrophotometry. A
calibration curve is constructed from standard solutions of X with the
following. results: 0.50 ppm, A = 0.24; 1.5 ppm, A = 0.36; 2.5 ppm, A = 0.44;
3.5 ppm, A = 0.59; 4.5 ppm, A = 0.70. A solution of unknown X concentration
had an absorbance of A = 0.50. Find the slope and intercept of the calibration
17
Bùi Xuân Vững Instrumental Analysis Chemistry
18
Bùi Xuân Vững Instrumental Analysis Chemistry
tablet.
19
Bùi Xuân Vững Instrumental Analysis Chemistry
Under normal conditions, atoms existing in matter are at the stable state, which
has the minimum energy, called fundermental state of matter. These atoms neither
absorb nor emit energy. However, if the matter is evaporated to transfer the atoms to
free atomic gaseous state, and these atoms are provided suitable energy, they are
excited such that their outermost electrons will be transferred to other orbitals having
higher energy:
Ao + ΔE → A*
(Ground) (Excited)
Here, Ao and A* are ground and excited electronic states of the atom respectively;
DE is the provided energy that is smaller than the first ionization energy of the atom.
20
Bùi Xuân Vững Instrumental Analysis Chemistry
b is also constant, and this value is dependant on the nature of each atom, each
spectrum line, and the concentration of each analyzing element. The value of b always
lies in the following region:
0<b≤1
Here, for every spectrum lines, we always have a specific value of the
concentration C0 of analyte in sample at which the value of b is equal to 1, and we
always have:
- If the concentration of the analyst is smaller than the value of C 0, the value of
b is always equal to 1.
- Otherwise, if the above concentration is higher than C 0, the value of b
gradually decreases from 1 to zero with the increase in the concentration of the
analyte.
Formula (2.1) is a fundermental equation of the quantitative analysis method
based on atomic emission spectroscopy, and the relationship between emission
Iλ
C
B
0 Ca Co C
Figure 2.4: Relationship between the intensity Iλ of spectrum line and concentration C
As can be seen in the plot, the curve has two sections. Section AB corresponding
to the concentration range of Ca-C0 is linear. Here, the relationship between Il and C
is linear (like equation y=ax). Otherwise, section BC is not linear. The concentration
C0 is called the limit concentration of linear region. The more sensitive the spectrum
22
Bùi Xuân Vững Instrumental Analysis Chemistry
line, the greater the value of C0 and vice versa. In analytical practical, section AB is
always used due to easier construction of calibration curve and results derived from
the curve having higher precision.
Apart from quatitative analysis, atomic emission spectroscopy is a super
qualitative analysis method in detecting the presence of many elements, especially
metals, in different samples. That is due to the fact that each element has some distinct
spectrum lines that other elements do not have.
2.2. Effect of temperature on atomic spectrum
2.2.1. The Boltzmann Distribution
Consider an atom with energy Eo at ground state. If this atom absorb energy from
the medium, it transfer to excited state with energy E*. In general, an atom (or
molecule) may have more than one state at a given energy. The number of states at
each energy is called the degeneracy. We will call the degeneracies go and g*. The
Boltzmann distribution describes the relative populations of different states at thermal
equilibrium. At a given temperature, if equilibrium exists, the relative population
(N*/No) of any two states is
N ∗¿ g∗¿ − ΔE/ kT
= e
No go
¿
¿
The fraction of atoms in the excited state is still less than 0.02%, but that fraction
has increased by 100(1.74 - 1.67) /1.67 = 4%.
2.2.3. Effect of temperature on emission and absorption
As can be seen from the above calculation, at 2600 K there are more than
99.98% of sodium atoms are in their ground state. Varying the temperature by 10 K
23
Bùi Xuân Vững Instrumental Analysis Chemistry
hardly affects the ground-state population and would not noticeably affect the signal
in atomic absorption. Emission intensity is proportional to the population of the
excited state. Because the excited-state population changes by 4% when the
temperature rises 10 K, emission intensity rises by 4%. Almost all atomic emission is
carried out with an inductively coupled plasma, whose temperature is more stable than
that of a flame.
2.3. Measurement principle and main components of an atomic emission
spectrometer
2.3.1. Measurement principle
Based on the appearance of atomic emission spectrum as mentioned above, for
applying the emission spectroscopy to analyze elements, we need to perform the
following steps:
1. Find suitable conditions to vaporize completely the sample.
2. The gaseous-state sample is completely atomized to form ground–state
atoms of the analyzing element.
3. Excite these atoms to generate emission spectrum in such a way that the
exciting efficiency is high, stable and reproducible.
4. Collect the whole emission beam of the sample, disperse this beam to obtain
emission spectrum of the sample/
5. Based on this spectrum to analyze qualitatively and quantitatively.
These five steps have to be performed for an analytical procedure using AES to
determine an element in a specific sample in which each step is standardized, and the
most suitable conditions are chosen for each step.
II.3.2. Main components
According to the general principle, an atomic emission spectrometer has to
consist of three main components as follows:
1. Power source: it is a divice or a component to perform steps of 1-3,
vaporization, atomization and exciting to form emission spectrum.
2. Spectrophotometer: it is a component containing monochromator and
photomultipliertube that has duties of collecting, dispersing the emission beam into
discrete spectrum lines of each element and measuring the intensities of these
spectrum lines.
3. Readout device: it displays the results of qualitative and quantitative
analysis.
These above components are indispensable for an atomic emission
spectrometer. In the present, in order to programme measuring processes, increase
24
Bùi Xuân Vững Instrumental Analysis Chemistry
25
Bùi Xuân Vững Instrumental Analysis Chemistry
Due to low flame temperature, flames have been only used as an exciting energy
source to analyse alkali metals (Li, Na, K, Rb, Cs) and some alkaline earth metals (Ca,
Mg) with the sensitivity of 1 - 10 µg/ml. These metals are determined using a method
called Flame Spectrophotometry, but in reality it is based on the principle of atomic
emission spectroscopy. Here, it should be noted that the spectrum excitation by flames
is subjected to the considerable effect of sample matrix.
26
Bùi Xuân Vững Instrumental Analysis Chemistry
Temperature Ionization
Electrodes Current (A)
(0C) potential (eV)
1. pressed 10 5800 (1) 11.25
27
Bùi Xuân Vững Instrumental Analysis Chemistry
Graphite
2. Fe metal 10 4400 (2) 7.86
3. Zn metal 10 5200 (3) 9.40
4. Al metal 10 3800 (4) 5.90
5. Cr metal 10 4000 (5) 6.76
2.4.3. Spark
This is also a thermal energy source produced by the electric discharge between
two electrodes having small electrical current (100 - 1000 mA) but very high potential
(20 - 30 kV). The discharge is periodically performed with from 50 to 500 times per
second. The electric spark temperature is about 4000 - 7000 0C. Because the spark
does not make the electrodes so hot, the evaporation of sample is difficult and slow.
Consequently, the analytical sensitivity is not so high (10 - 100 µg), but the stability is
better than arc. Therefore using spark leads to long excitation time. The spark
temperature depends mainly upon current density and is affected by the structure of
material used to make the electrodes.
The higher thermal stability of the material, the less effect on the temperatue of
spark. Because the electrodes are not subjected to be red-hot, spark excitation is very
suitable to analyze samples of metal, alloy and solution rather than ore, soil, salt,
oxide, silicate.
2.4.4. Inductively coupled plasma (ICP) source
Plasma is a conducting gaseous
misture containing a significant
concentration of cations and electrons.
In the argon plasma employed for
emission analyses, argon ion and
electrons are the principle conducting
species although cations from the
sample also contribute. Argon ion, once
formed in a plasma, are capable of
absorbing sufficient power from an
external source to maintain the
temperature at up to 10000K.
28
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 2.8: (a) Photograph and (b) temperature profile of a typical inductively coupled plasma
The aerosol of a liquid sample is generated by two ways: using a conventional
nebulizer with the carrier gas of argon and using ultrasonic nebulizer (being used
nowadays).
29
Bùi Xuân Vững Instrumental Analysis Chemistry
30
Bùi Xuân Vững Instrumental Analysis Chemistry
ICP has been using popularly because of the following good points:
- It has very high analytical sensitivity, about 10 – 0.1 ng;
- This method is capable of analyzing thermally stable elements as well as
sample having thermally stable matrix;
- The results have small error and good reproducibility;
- Linear range is wide;
- The rate of analyzing sample is fast (50 - 120 samples/h);
- The effect of sample matrix is insignificant.
Due to these above excellences, ICP excitation is mainly used in atomic
emission spectroscopy compared with other source excitations. Especially, ICP is a
very suitable source excitation to analyze rare-earth metals.
2.5. Monochromator and detector in ICP-AES
In Figure 2.11, the most common form of a monochromator (a Rowland circle)
and detector (photomultiplier; PMT) is shown:
The Rowland system utilizes a concave Echellette-style grating monochromator
to separate the various emission lines and simultaneously focus individual
wavelengths on to a series of slits, with each slit aligned to allow a specific
wavelength of radiation to pass to a detector. The standard detector, a photomultiplier
tube (PMT), was discussed in Section 1.3.2. Some systems use multiple PMTs at
fixed locations to monitor each wavelength simultaneously (Figure 2.11) whereas
other systems use a single PMT and move it to different locations to detect each
wavelength. Data from these detectors are processed by a computer because multiple
wavelengths are measured in an ICP-AES system at the same time.
31
Bùi Xuân Vững Instrumental Analysis Chemistry
32
Bùi Xuân Vững Instrumental Analysis Chemistry
33
Bùi Xuân Vững Instrumental Analysis Chemistry
34
Bùi Xuân Vững Instrumental Analysis Chemistry
From the data, calculate the lead concentration (mg/L) in the two sample
solutions, A and B.
35
Bùi Xuân Vững Instrumental Analysis Chemistry
36
Bùi Xuân Vững Instrumental Analysis Chemistry
37
Bùi Xuân Vững Instrumental Analysis Chemistry
38
Bùi Xuân Vững Instrumental Analysis Chemistry
39
Bùi Xuân Vững Instrumental Analysis Chemistry
analyzing element. These atoms would absorb some specific light and generate its
absportion spectrum. Here, parts of the initial intensity of the beam absorbed by
the atoms of analyzing element depend upon the concentration of these atoms in
the absorption medium. The emission beam of light of the analyzing element is
provided by a source as hollow cathode lamp (HCL).
3. Collimating, dispersing and choosing a disctint spectrum line of the
analyzing element to measure
The measured intensity is a absorption signal of the absorption spectrum
line. In a determinate limit of concentration, the ratio of the measured intensity
before and after passing through the absorption medium is linearly proportional to
the concentration C of the analyzing element in sample as presented in Equation
(3.8).
These above processes are principle of AAS measurement. So in order to
conduct an AAS measurement, an AAS spectrometer has to consist of the
following main components:
Component 1: A resonance emission source of the analyzing element
generates an emission beam of light to irradiate absorption medium containing a
vapor of the unbouned atoms of the analyzing element. For this purpose, a hollow
cathode lamp (HCL) is used.
A hollow-cathode lamp, in Figure 3.2, is filled with Ne (or Ar) at a pressure
of ≈ 1-5 Torr. The cathode is made of the element whose emission lines we want.
When are a potential of about 500V applied between the anode and the cathode,
gas is ionized and positive ions are accelerated toward the cathode. After
ionization occurs, the lamp is maintained at a constant current of 2–30 mA by a
lower voltage. Cations strike the cathode with enough energy to “sputter” metal
atoms from the cathode into the gas phase. Gaseous atoms excited by collisions
with high-energy electrons emit photons. This atomic radiation has the same
frequency absorbed by analyte in the flame or furnace. Atoms in the lamp are
cooler than atoms in a
40
Bùi Xuân Vững Instrumental Analysis Chemistry
Digital display
or computer
41
Bùi Xuân Vững Instrumental Analysis Chemistry
42
Bùi Xuân Vững Instrumental Analysis Chemistry
to 2500oC for 3 s to clean out any remaining residue. The furnace is purged with
Ar or N2 during each step except atomization to remove volatile material. Gas
flow is halted during atomization to avoid blowing analyte out of the furnace.
When developing a method for a new kind of sample, it is important to record the
signal as a function of time, because signals are also observed from smoke during
charring and from the glow of the red-hot oven in the last part of atomization. A
skilled operator must interpret which signal is due to analyte so that the right peak
is integrated.
43
Bùi Xuân Vững Instrumental Analysis Chemistry
44
Bùi Xuân Vững Instrumental Analysis Chemistry
3-3) Quigley and Vernon report results for the determination of trace metals in sea
water using a graphite furnace atomic absorption spectrophotometer.
Calibration was achieved by the method of standard additions. The trace
metals were first separated from their complex, high-salt matrix by
coprecipitating with Fe3+. In a typical analysis a 5.00-mL portion of 2000-
ppm Fe3+ was added to 1.00 L of sea water. The pH was adjusted to 9 using
NH4OH, and the precipitate of Fe(OH) 3 allowed to stand overnight. After
isolating and rinsing the precipitate, the Fe(OH) 3 and coprecipitated metals
were dissolved in 2 mL of concentrated HNO 3 and diluted to volume in a 50-
mL volumetric flask. To analyze for Mn 2+, a 1.00-mL sample of this solution
was diluted to 100 mL in a volumetric flask. The following samples were
injected into the graphite furnace and analyzed
Report the parts per billion of Mn2+ in the sample of sea water.
45
Bùi Xuân Vững Instrumental Analysis Chemistry
3-4) An AAS method is employed for the determination of lead (Pb) in a sample
of adulterated paprika by the introduction of lead oxide (of the same colour).
An electrothermal atomic absorption instrument that provides a background
correction based upon the Zeeman effect is used. 0.01 g of the paprika
powder is placed in the tube of the graphite furnace. The determination of the
area peak absorbance was made at λ=283.3nm first in the absence and later
in the presence of a magnetic field. The value of the peak absorption
following background correction was 1220 (arbitrary units). Under the same
conditions, 0.01mL of a solution of 10 g/L Pb led to a value of 1000 in the
same units.
Calculate the % mass of lead in the sample of paprika under study.
3-5) Free cyanide in aqueous solution can be determined indirectly by atomic
absorption on the basis of its ability to dissolve silver as it passes through a
porous silver membrane filter at pH 12.
4Ag(s) + 8CN+ 2H2O + O2 → 4Ag(CN)2 + 4OH-
A series of silver standards gave a linear calibration curve in flame atomic
absorption with a slope of 807 meter units per ppm Ag in the standard. (The
meter units are arbitrary numbers proportional to absorbance, and ppm refers
to µg Ag/mL.) An unknown cyanide solution passed through the silver
membrane gave a meter reading of 198 units. Find the molarity of in the
unknown.
46
Bùi Xuân Vững Instrumental Analysis Chemistry
occurring on the interface between electrode and analyte solution, and the
electrochemical properties of solution containing the cell.
These methods have developed for a lomg time, and especially in recent decades.
They are not only used in qualitative and quantitative analysis, but as a means to
investigate the theory of electrochemical processes and chemical reactions of
inorganic and organic substances.
47
Bùi Xuân Vững Instrumental Analysis Chemistry
Analytical methods that are based upon potential measurements are termed
potentiometric methods. Potentialmetric methods are performed in a cell containing
analyte solution and two electrodes immersed in the solution including:
- 1 reference electrode has its potential constant (Eref = const)
- 1 indicator electrode indicates the potential of solution.
These two electrodes are connected to a potentiometer. The third component of
a potentiometric cell is a salt bridge that prevents the components of the analyte
solution from mixing with those of the reference electrode. The potential of the cell,
Ecell, is given by the equaton
Here, Eind is the potential of indicator electrode;E ref is the potential of reference
electrode.
Because Eref is constant in during shows a schematic diagram of a cell for
potential measurement, the potential of potentiometric analysis.
the cell only depends upon the potential
of indicator electrode that contains the
information about the concentration of
the analyte in the measured solution. So
a potentiometric analysis involves
measuring a cell potential, correcting
this potential for the reference
potential, and correcting the anayte
concentration from the indicator
electrode potential.
It is noted that measuring
potential must be conducted in the
condition of electric current of zero or Figure 4.1: A schematic diagram of a
constant (I = 0, or = const). Figure 4.1 cell for potentiometric analysis
48
Bùi Xuân Vững Instrumental Analysis Chemistry
49
Bùi Xuân Vững Instrumental Analysis Chemistry
50
Bùi Xuân Vững Instrumental Analysis Chemistry
Table 4.1 shows the effect of the concentration of potassium chloride on the
potential of both the calomel electrode and silver/silver chloride electrode. In general,
the larger concentration and the higher temperature, the more decreased the potential
of electrode.
4.3.2. Indicator electrodes
An indicator electrode is one that there are such electrochemical processes
occurring on its surface as oxidation, reducrion, electron exchange, precipitation, and
so on. An ideal indicator electrode responds rapidly and reproducibly to changes in the
concentration of an analyte ion or group of analyte ions. Indicator electodes are of
three types: metallic, membrane, and ion-selective electrodes.
4.3.2.1. Metallic indicator electrodes
a) Electrodes of the first kind
An electrode of the first kind is a pure metal electrode that is in direct
equilibrium with its cation in the solution. This kind of electrode is used to keep track
of or determine the concentration of metal ions in solutions. For instance, the
equilibrium between metal Cu and its ion Cu2+ is:
where Eind is the electrode potential of copper electrode and a Cu2+ is the activity
2+
of Cu ion.
b) Electrodes of the second kind
Metals not only serve as indicator electrodes for their own cations but also
respond to the activities of anions that form sparingly soluble precipitates or stable
complexes with such cations. For example, in silver/silver chloride electrode, the
electrode potential correlates to reproducibly with the activity of chloride ion in a
solution saturated with silver chloride. Here, the electrode reaction can be written as
AgCl (s) + 1e- Ago(s) + Cl- có EoAgCl = 0.222 V
The Nernst expression for this process is
51
Bùi Xuân Vững Instrumental Analysis Chemistry
Eind = E oAgCl -
0.0592
1
(
log a Cl- )
This equation shows that the potential of a silver electrode is proportional to
the activity of chloride ion. Thus, in a solution saturated with silver chloride, a silver
electrod can serve as an indicator electrode of the second kind of chloride ion.
Another electrode of the second kind is mercury/mercury-EDTA complex
electrode. For example, when a small amount of HgY2- is added to a solution
containing Y4- (EDTA ion), the half-reaction at a mercury cathode is:
HgY2 + 2e- 2Hg(l) + Y4- có Eo = 0.21V
0.0592 a 4-
Eind = 0.21 - log Y
2 a HgY2-
for which
The stability constant for HgY2- is very large (6.3x1023), so the
concentration of the complex remains essentially constant over a large range of
Y4- concentration. The Nernst equation can be written as
0.0592
Eind = K - log a Y 4-
2
0.0592 1
K = 0.21 - log
2 a HgY 2-
Here,
This electrode is used to titrate EDTA.
c) Inert metallic electrodes for redox system
This kind of electrode is made from inert metal such as platinum, gold,
palladium responding to the potential of a redox system with which it is in contact.
For example, the equilibrium of a Ce4+ / Ce3+ system on platinum electrode is
Ce4+ + 1e- Ce3+ Eo = 1.32V
And the potential of the platinum electrode is
0.059 a 4+
E ind = E o + log Ce
n a Ce3+
52
Bùi Xuân Vững Instrumental Analysis Chemistry
two dozen other ions with high sensitivity. Such electrodes are also called ion sective
electrodes. In this section we will discuss some of them such as glass electrodes,
liquid membrand electrodes, crystalline-membrand electrodes. It is important to note
that there are fundermental differences both in design and in principle between
membrand electrodes and metal electrodes.
a) The glass electrode for pH measurements
The glass electrode used to measure pH is the most common ionselective
electrode. Figure shows diagram of a glass combination electrode for measuring pH
and a line diagram of this cell.
Figure 4.4: A glass combination electrode and a line diagram of this electrode
The pH-sensitive part of the electrode is the thin glass bulb or cone at the bottom
of the electrodes. The sample reference electrode at the left of the line diagram is the
Ag/AgCl electrode at the right of the combination electrode. The internal reference
electrode at the right side of the line diagram is the straight Ag/AgCl electrode at the
center of the electrode in Figure 4.4. The two reference electrodes measure the electric
potential difference across the glass membrane. The salt bridge in the line diagram is
the porous ceramic plug at the bottom-right side of the combination electrode
53
Bùi Xuân Vững Instrumental Analysis Chemistry
connecting analyte solution to the reference electrode. Two silver wires coated with
AgCl are visible inside the electrode.
- The composition and structure of glass membranes
The property and operation of glass membrane is a key factor for a glass pH
electrode. The composition of a glass membrane depends on the manufacturer. For
example, one of glass membranes, which has been widely used, consists of
aproximately 22% Na2O, 6% CaO and 72% SiO 2. This membrane shows excellent
specificity toward hydrogen ions up to a pH of about 9. Other glass formulations are
now in use in which sodium and calcium ions are replayced to various degrees by
barium and lithium ions. These membrane have superior selectivity and lifetime.
Figure 4.5 shows the irregular structure of the silicate lattice in glass. Negatively
charged oxygen atoms in glass can bind to cations of suitable size. Monovalent
cations, such as sodium and lithium can move sluggishly through the silicate lattice,
and are responsible for electric conduction within the membrane.
Both membrane surfaces must be hydrated before a glass electrode with
function as a pH electrode. Metal ions in these hydrated gel regions of the membrane
diffuse out of the glass and into solution. H+ can diffuse into the membrane to replace
the metal ions.
The reaction in which H+ replaces cations in the glass is an ion-exchange
equilibrium. A pH electrode responds selectively to H+ because H+ is the only ion that
binds significantly to the hydrated gel layer. The ion-exchange reaction can be written
as
H+(aq) + Na+(glass membrane) Na+ (aq) + H+(surface membrane)
The potential difference between inner and outer silver-silver chloride electrodes
in Figure 4.3 depends on the chloride concentration in each electrode compartment
and on the potential difference across the glass membrane. Because the concentration
of Cl- is fixed in each compartment and because the concentration of hydrogen ions is
fixed on the inside of the glass membrane, the only variable is the pH of analyte
solution outside the glass membrane. So the potential difference E between two
reference electrodes can be expressed as a Nernst equation
Eb = constant + 0.05916loga1 = constant – 0.05916pH.
Here, a1 is the activity of H+ in the outer solution (analyte solution).
This equation states that the voltage of the ideal pH electrode changes by 59.16
mV for every pH-unit change of analyte activity at 25°C.
The ion- exchange process can be depicted in Figure 4.5,
54
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 4.5: Ion exchange at the interface between membrane and solution
reagent dissolves glass and exposes fresh surface. Wash the electrode with water and
try calibrating it again. Avoid contact with ammonium bifluoride, NH 4HF2, which
produces a painful HF burn.
4.4. Potentiometric titration
A potentiometric titration involves measurement of the potential of a suitable
indicator electrode as a function of titrand volume. Potentiometric titrations provide
data that are more reliable than data from titrations that use chemical indicator and are
particularly useful with colored or turbid solution and for detecting the precence of
unsuspected species. Such titrations are also readily automated. Manual
potentiometric titrations of this kind, on the other hand, suffer from the disadvantage
of being more time-consuming than those involved indicators.
A typical apparatus for performing a manual potentiometric titration
between the analyte of Fe2+ and the titrant of Ce4+ is illustrated in Figure 4.6 . The cell
consists of two electrodes in which one is reference electrode and the other is indicator
electrode that could be inert metal such as Pt, Ag, Pd or glassy carbon or glass
membrand electrode.
Figure 4.6: The diagram of apparatus for potentiometric titration of Fe2+ with standard Ce4+
and the titration curve of 100.0 mL of 0.050M Fe 2+ in 1M HClO4 with 0.100M Ce4+ trong dung dịch
HClO4 1M.
Here, in the titration process the titrant is dropped from the buret to the beaker
containing the analyte leading to the change of its concentration. This change of
concentration results in the change of potential of the indicator electrode, especially
56
Bùi Xuân Vững Instrumental Analysis Chemistry
the abrupt change of potential around the equivalence point of the titration. This is the
signal that is used to detect the end point of the titration.
The potentiometric titration is applied for neutratization, precipitation, complex,
and redox reactions.
Problems
4-1) (a) Write the half-reactions for the silver-silver chloride and calomel reference
electrodes.
(b) Predict the voltage for the following cell.
4-2) Convert the following potentials. The Ag AgCl and calomel reference electrodes
are saturated with KCl.
(a) 0.523 V versus S.H.E = ? versus Ag│AgCl
(b) -0.111 V versus Ag │AgCl = ? versus S.H.E.
(c) -0.222 V versus S.C.E. = ? versus S.H.E.
(d) 0.023 V versus Ag │AgCl = ? versus S.C.E.
(e) -0.023 V versus S.C.E = ? versus Ag │AgCl.
4-3) For a silver-silver chloride electrode, the following potentials are observed:
E° = 0.222 V E(saturated KCl) = 0.197 V
Predict the value of E for a calomel electrode saturated with KCl, given that for
the calomel electrode is 0.268 V.
4-4) A cell was prepared by dipping a Cu wire and a saturated calomel electrode into
0.10 M CuSO4 solution. The Cu wire was attached to the positive terminal of a
potentiometer and the calomel electrode was attached to the negative terminal.
(a) Write a half-reaction for the Cu electrode.
(b) Write the Nernst equation for the Cu electrode.
(c) Calculate the cell voltage.
4-5) Explain why a silver electrode can be an indicator electrode for Ag + and for
halides.
4-6) A 10.0-mL solution of 0.050 0 M AgNO 3 was titrated with 0.025 0 M NaBr in the
cell
S.C.E. ║ titration solution │ Ag(s)
Find the cell voltage for 0.1, 10.0, 20.0, and 30.0 mL of titrant.
57
Bùi Xuân Vững Instrumental Analysis Chemistry
58
Bùi Xuân Vững Instrumental Analysis Chemistry
achievement for which he was awarded the Nobel Prize in Chemistry in 1959. The
detection limit of this method lies within 10-3 - 10-5 Mol/L.
Due to the existence of capacitor current, many species can not be determined at
concentration smaller than 10-5 Mol/l leading to the appearance of square wave
polarography and pulse polarography. These later methods minimize significantly
capacitor current when measuring Faradic current so that they have higher sensitivity.
However, the detection limit gains n.10-6 Mol/L for the majority of substances, and
n.10-7 Mol/L for some of them.
Together with the development of other analytical methods, the stripping
voltamperometry methods were introduced. These methods are based upon the
combination between the electrolysis for concentration and polarography to increase
sensitivity. Stripping is the most sensitive voltammetric technique, because analyte is
concentrated from a dilute solution. The longer the period of concentration, the more
sensitive is the analysis. The detection limit for many metallic ions in aqueous
solution can be lowered to 10-12 M.
5.2. Voltametric measurements
In order to perform a voltametric
measurement for analyzing substances,
we have to dissolve analyte in a
suitable solvent (usually in water) to
obtain a sample solution that is then
added a large excess of inert supporting
electrolyte, for example KCl. The
resulting solution is measured using a
voltammeter making use of a three-
electrode potentiostat, such as that Figure 5.1: Diagram of voltammeter making
shown in Figure 5.1. use of a three-electrode potentiostat
59
Bùi Xuân Vững Instrumental Analysis Chemistry
+ Voltmeter and current meter are used to measure the applied potential and the
Faradaic current when a reducing or oxidation reaction taking place at the woking
electrode.
A graph of current versus working electrode potential for a measuring sample
being oxidized or reduced at the working electrode is called voltammogram.
5.3. Current in Voltammetry
5.3.1. Faradaic current
When we oxidize an analyte at the working electrode by applying a suitable
potential, the resulting electrons pass through the potentiostat to the auxiliary
electrode, reducing the solvent or some other component of the solution matrix. If we
reduce the analyte at the working electrode, the current flows from the auxiliary
electrode to the cathode. In either case, the current from redox reactions at the
working electrode and the auxiliary electrodes is called a faradaic current.
The magnitude of Faradaic current is affected by three modes of mass transport
including diffusion, convection and migration.
Diffusion occurs whenever the concentration of an ion or molecule at the surface
of the electrode is different from that in bulk solution. The region of solution over
which diffusion occurs is the diffusion layer.
Convection occurs when we mechanically mix the solution, carrying reactants
toward the electrode and removing products from the electrode. The most common
form of convection is stirring the solution with a stir bar. Other methods that have
been used include rotating the electrode and incorporating the electrode into a flow-
cell.
The final mode of mass transport is migration, which occurs when a charged
particle in solution is attracted to or repelled from an electrode that carries a surface
charge. If the electrode carries a positive charge, for example, an anion will move
toward the electrode and a cation will move toward the bulk solution. Unlike diffusion
and convection, migration only affects the mass transport of charged particles.
Among the three described phenomena only diffusion can be related to
the concentration of the electro-active species, called also depolarizer.
The movement of material to and from the electrode surface is a complex
function of all three modes of mass transport. In the limit where diffusion is the only
significant form of mass transport, the current in a voltammetric cell is equal to
i=nFAD(Cbulk−Cx=0)/δ
Where n the number of electrons in the redox reaction, F is Faraday’s constant,
A is the area of the electrode, D is the diffusion coefficient for the species reacting at
60
Bùi Xuân Vững Instrumental Analysis Chemistry
the electrode, Cbulk and Cx=0 are its concentrations in bulk solution and at the electrode
surface, and δ is the thickness of the diffusion layer.
For the above equation to be valid, convection and migration must not interfere
with the formation of a diffusion layer.
We can eliminate migration by adding a high concentration of an inert
supporting electrolyte. Because ions of similar charge are equally attracted to or
repelled from the surface of the electrode, each has an equal probability of undergoing
migration. A large excess of an inert electrolyte ensures that few reactants or products
experience migration.
It is easy to eliminate convection by not stirring the solution.
5.3.2. Charging Current (capacitive current)
61
Bùi Xuân Vững Instrumental Analysis Chemistry
62
Bùi Xuân Vững Instrumental Analysis Chemistry
63
Bùi Xuân Vững Instrumental Analysis Chemistry
however, linearly changing potential is applied, with a rate slow enough that the
change of potential throughout the drop-life time is about a few millivolts. The current
is measured at the end of the drop life.
Oxygen removal
Dissolved oxygen undergoes a two-step irreversible reduction at the dropping
electrode; the H202 produced in the first step is reduced to H20 in the second. The
two waves of equal size result, the first with a half-wave potential at about -0.14 V
and the second at about -0.9 V vs. S.C.E. The two half-reactions are somewhat slow.
As a consequence, the waves are drawn out over a considerable potential range. While
these polarographic waves are convenient for the determination of oxygen in
solutions, the presence of this element often interferes with the accurate determination
of other species. Thus, gas accomplishes this end. A stream of the same gas, usually
nitrogen, is passed over the surface during the analysis to prevent re-absorption.
5.3.1.2. Basis of quantitative and qualitative polarographic analysis
A polarogram can be plotted between the current flowing through the
polarographic cell against the increasing potential of the dropping electrode. A typical
polarogram is depicted in Figure 5.3.
The virtually flat upper portion of the polarographic wave is called its plateau
and the total current which flows through the cell at a potential on the plateau is called
the limiting current of the substance responsible for the wave.
The difference between the limiting current and residual is called the diffusion
current or wave height (id) of that substance and is a function of the concentration of
the electroactive material. The potential at which the current is one-half of the
diffusion current is called half-wave potential designated as E 1/2. The half-wave
potential of a reducible substance is independent of concentration and is the
characteristic of the nature of the reacting material.
This is the essential basis of quantitative and qualitative polarographic analysis.
A schematic diagram of a simple polarographic set up is shown in Fig. 5.2.
64
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 5.3: Typical polarogram obtained with the dropping mercury electrode.
Table 5.1: The half-wave potential of some inorganic and organic species
65
Bùi Xuân Vững Instrumental Analysis Chemistry
The basis for the quantitative polarographic analysis is the existence of the linear
proportionality between the diffusion current and the concentration of the depolarizer
as given by Ilkovic equation.
Id = 607n . D1/2 . m 2/3 . t1/ 6 . C x
Where, Id = Average diffusion current, µA
n = Number of Faradays per mol involved in the electrode reaction
D = Diffusion coefficient of the electroactive material, (cm2/sec)
C = Concentration of the electroactive material, (millimoles/litre).
66
Bùi Xuân Vững Instrumental Analysis Chemistry
67
Bùi Xuân Vững Instrumental Analysis Chemistry
negative than about -1.8 Volts vs. S.C.E, visible hydrogen evolution occurs in acid
solutions and the usual supporting electrolytes commence to discharge.
5.3.1.4. Limitations of DC polarography
Ever since the development of DC polarography a great deal of efforts have been
undertaken to compensate several of the following three main limitations:
a) The drawback of direct current (DC) polarography as regards sensitivity is the
current required to change the potential of the dropping mercury electrode (DME) to
the required value: at a depolarizer concentration of 10 -5 mol.1-1, this time-averaged
charging current is comparable to the Faradaic current.
b) Lack of resolution of the sigmoid shaped waves (DC, NPP, etc.
polarographies). In order to sufficiently distinguish two polarographic waves so as to
be able to measure their corresponding half-wave potentials and limiting current
values, it is necessary that the former half-wave potentials should differ in more than
200 mV. All subsequent polarographic techniques, in which the obtained response is in
the form of a peak (derivative polarography, square-wave polarography, AC
polarography, differential polarography, differential pulse polarography, etc.) present a
higher resolution, being a difference of only about 50 mV sufficient to be able to
discriminate two substances presenting close half-wave potentials.
5.4. Pulse Polarography
In pulse methods the procedures are based on the application of pulse changes of
potential and the current response is measured at a suitable time relative to the time of
the pulse. The concept includes three methods: normal pulse polarography (NPP),
differential pulse polarography (DPP) and square-wave polarography (SWP). Pulse
techniques improve detection limits since they benefit from the different variation of
diffusion and capacitive current intensities with time: when carrying out
measurements at the pulse end, the capacitive current is practicably negligible, being
the value of the Faradaic current still significant.
5.4.1. Normal pulse polarography (NPP)
In normal pulse polarography (NPP) the mercury-drop electrode is held for
most of its duration at a constant potential Ein, at which no electrochemical reaction
takes place under given experimental conditions. The potential of interest E p is applied
in the last stage of the drop life, for a length of time t p (of the order of few
milliseconds). The values of Ein and tp are kept constant throughout the recording of
the polarogram and Ep is changed from drop to drop.
68
Bùi Xuân Vững Instrumental Analysis Chemistry
(a)
Figure 5.4: (a) Voltage profile of normal pulse polarography (NPP) (b) A typical NPP
polarogram
NPP keeps the DME at a constant potential before the start of the Faradaic current;
then, almost at the end of the drop life, a voltage pulse is applied, the amplitude of
which gradually increases from drop the drop. The pulse duration is about 200 ms.
The current is usually measured shortly and before the pulse end (about 20 ms after
the pulse edge) and is recorded versus the pulse amplitude. The current-potential
relationship for a reversible process is
Dx
Ii = n.F.S. . Cx
pt p
or: Ii = k1.Cx
Where S: The surface area of the dropping mercury electrode at the sampling
time (cm2)
F: Faraday’s constant
n: The electron number exchanged in the reaction at the eletrode
Dx: The diffusion coefficient of analyte
tp: The time length of a pulse (sec)
Cx: The concentration of analyte in the measuring solution
Ii : limitting current intensity
This mode is about seven times more sensitive than classical DC polarography.
The shape of the curve is that of a normal DC polarogram. The detection limit of NPP
is about 2·10-7 M.
5.4.2. Differential Pulse Polarography
This technique is comparable to normal pulse voltammetry in that the potential
is also scanned with a series of pulses. However, it differs from NPP because each
potential pulse is fixed, of small amplitude (10 to 100 mV), and is superimposed on a
slowly changing base potential. Current is measured at two points for each pulse, the
69
Bùi Xuân Vững Instrumental Analysis Chemistry
first point (1) just before the application of the pulse and the second (2) at the end of
the pulse. These sampling points are selected to allow for the decay of the nonfaradaic
(charging) current. The difference between current measurements at these points for
each pulse is determined and plotted against the base potential.
From analytical point of view, the sensitivity of DPP is even better than that of
NPP. The potential sequence on a single mercury drop and the potential sequence used
for recording an entire differential pulse polarogram are given in Fig.5.5. The current
is sampled twice during a drop life-time: (i) at t1, just before the pulse about 17 ms,
and (ii) at τ, just before the drop fall about 17 ms. The polarogram represents the
current difference as function of the base potential. The curve is
peaked shaped (Figure 5.6).
For Nernstian processes Ox + ne = Red and C*red = 0, the faradaic component
of the current at the peak Ii is
D ( s - 1) x C
Ii ( max ) = n.F.S . x
p.t p ( s + 1) x
k2
Where:
�nFΔE �
σ = exp � x �
�RT 2 �
DE : The magnitute of applied pulse
As a result, The height of the peak, Ii(max), is proportional to the
concentration of the electroactive species.
Ii ( max ) = k 2 .Cx
This equation is a base for the quantitative analysis of DPP method to
determine a lot of both inorganic and organic electroactive species with the sensitivity
up 10-8 Mol/l.
70
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 5.5 : Comparison of the voltage profile between NPP (left) and DPP (right)
NHCH3 NHCH3
N N
E1/2 = -0.265 V
+ 2H+ + 2e-
Cl N+ N
Cl
O-
C6H5 C6H5
H NHCH3 NHCH3
N N
E1/2 = -1.120 V
NH + 2H+ + 2e-
Cl Cl NH
C6H5 C6H5
Figure 5.6: Comparison of direct current (D.C) and different puled polarograph (DPP) of 1.2
x10-4 M chlodiazepoxide trong 3 ml of 0.05M H2SO4. Modulation amplitude = 50 mV.
72
Bùi Xuân Vững Instrumental Analysis Chemistry
Two types of mercury electrode are commonly used, the Hanging Mercury
Drop Electrode (HMDE) and the Thin Mercury Film Electrode (TMFE).
HMDE
Figure 5.8: Schematic diagram of a HDME (left) and a voltammetric instrument with HDME (right)
*TMFE
73
Bùi Xuân Vững Instrumental Analysis Chemistry
mercury(II) salts in solution. CSV has been used for the detection of inorganic anion
such as halide ions, selenide and sulfide ions, and oxyanions such as [MoO4]2- and
[VO3]2-. Some organic compounds (e.g., thiols and nucleic acid bases also form
insoluble mercury(II) salts, and are therefore suitable for CSV.
5.5.4. Adsorptive stripping voltammetry
In the three previous techniques, preconcentration step for AdSV, no charge is
transfered, and accumulation is achieved via the inherent tendency of particular
molecules to adsorb to a mercury surface (e.g., due to hydrophobicity) ( however, it
should be noted that there is a potential dependence for the preconcentration step, and
the optimum preconcentration potential needs to be determined experimentally). This
technique has been promoted extensively for organic molecules (e.g., chlopromazine,
dopamine, bilirubin, triazine- containing pesticides and polychlorinated biphenyls. It
should be noted that the organic molecules listed are pharmaceuticals and
agrochemicals. The ability ot detect trace concentratons of such molecules has become
increasingly important in recent years, due to the increased efficacy of many
pharmaceuticals and the increased knowledge of the toxic effects of trace amounts of
many agrochemicals.
AdSV can also be used for detection of transition metal cations, through the
formation of insoluble complexes (e.g., dimethylglyoxime complexes of nickel(II) and
cobalt(II).
The stripping step in AdSV can be anodic or cathodic, so the direction of the
potential scan should be specified. However, this has led to some confusion in the
literature, since AdSV with a cathodic stripping step has sometimes been reffered to as
CSV. It is important to distinguish whether the preconcentration step is electrolytic
(oxidation of mercury followed by a mercury salt film formation in CSV) or non-
electrolytic (adsorption on the electrode surface in AdSV).
5.6. Amperometric titrations
5.6.1. Hydrodynamic voltammetry
Liner-scan voltammetry in which the solution or the indicator electrode is kept in
motion is called hydrodynamic voltammetry.
Hydrodynamic voltammetry is performed in several ways. One method involves
stirring the solution vigorously while it is in contact with a fixed microelectrode.
Alternatively, the microelectrode is rotated at a constant high speed in the solution.
As described in section , during an electrolysis, reactant is carried to the surface
of an electrode by three mechanism: (1) migration under the influence of an electric
field, (2) convection resulting from stirring or vibration, and (3) diffusion due to
concentration differences between the film of liquid at the electrode surface and the
75
Bùi Xuân Vững Instrumental Analysis Chemistry
bulk of the solution. In voltammetry, every effort is made to minimize the effect of
migration by introducing an excess of an inactive supporting electrolyte When the
concentration of supporting electrolyte exceeds that of the analyte by 50 to 100-fold,
the fraction of the total current carried by the analyte approaches zero.
As a result, the current at any point in the electrolysis is determined by a
combination of (1) the rate of mass transfort of the analyte A to the edge of the Nernst
diffusion layer by convection and (2) the rate of transport of A from the outer edge of
the diffusion layer to the electrode surface. Convection maitains a constant supply of
A at the outer edge of the diffusion layer, however. Thus, a steady state current that is
determined by the applied potential.
5.6.2. Amperometric titrations
Hydrodynamic voltammetry can be employed to estimate the equivalence point
of titrations, provided at least one of the participants or products of the reaction
involved is oxidized or reduced at a microelectrode. Here, the current at some fixed
potential in the limiting current region is measured as a function of the reagent
volume.
Figure 5.10b shows some typical forms of Amperometric titration curves.
Curves (a) represents a titration in which the analyte reacts at the electrode while the
reagent does not. Curve (b) is typical of a titration in which the reagent reacts at the
microelectrode and the analyte does not. Curve (c) corresponds to a titration in which
both the analyte and the titrant reacts at the microelectrode.
There are two types of amperometric electrode systems. One uses a single
polarizable microelectrode that is often a rotating platimum electrode coupled to a
reference electrode; the other uses a pair of identical solid state microelectrodes
immersed in a stirring solution.
A dropping mercury electrode may also be used, in which case the solution is not
stirred.
Figure 5.10a shows a typical cell arrangement for amperometric titrations with a
rotating platinum electrode.
Amperometric titrations with one indicator electrode have been confined to
titrations in which a precipitate or a stable complex is the product.
Precipitating reagents include silver nitrate for halide ions, lead nitrate for sulfate
ion, and several organic reagents such as 8-hydroxyquinoline dimethylglyoxime, and
cupferron for various metallic ions that are reducible at microelectrode.
Several metal ions have also been determined by titration with standard solutions
of EDTA.
76
Bùi Xuân Vững Instrumental Analysis Chemistry
The exception just noted involves titrations of some organic compounds such as
phenols, aromatic amine and olefins,; hydrazine; arsenic (III) and antimony(III) with
bromine.
A standard solution of potassium bromate is added slowly from buret to acidic
solution of the analyte that also contains an excess of potassium bromide. Bromine is
formed by the reaction
BrO3- + Br - + 6H + ��
� 3Br2 + 3H 2O
Before the equivalent point, the bromine reacts completly with the analyte, so no
current is observed. After this point a rapid increase in current takes place due to
electrochemical reduction at the microelectrode, which is applied a suitable potential
for the reduction of bromine,
Br2 + 2e- 2Br-
The use of a pair of identical metallic microelectrode to establish the equivalent
point in amperometric titrations offers the advantages of simplicity of equiment and
not having to prepare and maintain a reference electrode.
(a) (b)
77
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 5.10: (a) Experimental set up for amperometric titrations with a rotating patinum electrode;
(b) Typical amperometric titration curves.
5.7. Coulometric titration
5.7.1. Principle of the method
Coulometric titration is a method of titration in which the titrating agent is
produced in a solution by electrolysis and the required amount of the agent is
determined by measuring the number of coulombs used in preparing it. A constant
current source called an amperostat is used to maintain a constant current during the
electrolysis.
For example, consider the coulometric titration of iron(II) at a platinum anod.
The titrating agent Ce4+ is generated by the electrolysis of Ce 3+ that is introduced at the
outset an excess amount:
Ce3+ →Ce4+ + e-
With stirring , the cerium(IV) is rapidly transported from the surface of the
electrode to the bulk of the solution, where it oxidizes an equivalent amount of
iron(II):
Ce4+ + Fe2+ →Ce3+ + Fe3+
If we have a means of determining when the reaction between Ce 4+ and Fe2+ is
complete, the concentration of Fe 2+ in the solution can be calculated based on the
number of coulombs used in preparing Ce 4+. Here, the number of coulombs (Q)
resulting from a constant current of I amperes operated for t seconds is
Q = I.t
5.7.2. Determining the end point
In general, indicators in volumetric titrations for detection of the end points can
be used in coulometric titrations as well.Thus, for the titration of iron(II) with
cerium(IV) as just described, an oxidation-reduction indicator such as 1.10-
phenantroline can be used.
For many specific coulometric titrations, the end point detection can be
performed by using a pair of identical platinum microelectrodes immersed in a stirring
solution.
Apparatus for coulometric titrations
Figure 5.11a presents shows a diagram of a coulometric titration apparatus that
includes a constant-current source, a titration vessel, a switch, an electric timer and a
current meter.
Cell for coulometric titrations
78
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 5.11: (a) Diagram of a coulometric titration (b) A coulometric titration cell
As can be seen from Figure 5.11B, a typicall coulometric titration cell consists of
a generator electrode at which the titrating reagent is generated and a counter
electrode to complete the circuit. The generator electrode is a platinum rectangle (or a
wire coil) having a relatively large surface area to minimize elects of polarization. If
the reaction products from the counter electrode interfere the titration, the counter
electrode is isolated from the reaction medium by a sintered glass disk. For example,
hydrogen gas formed at the counter electrode reacts rapidly with most oxidizing
reagents generated at the generator electrode. This interference leads to a positive
systematic error unless the gas is formed in a separate compartment.
5.7.3. Advantages of coulometric titrations
Compared with conventional titrations, coulometric titrations have some
advantages as follow:
- Eliminating the problems associated with the preparation,
standarization, and storage of standard solution.
- Titrating small amounts of analyte with high accuracy.
- A single constant current source provides titrating reagents for
precipitation, neutralization, comlex formation, or oxidation/reduction
titrations.
- Being easily automated.
Karl Fischer Titration of H2O
The Karl Fischer titration, which measures traces of water in transformer oil,
solvents, foods, polymers, and other substances, is developed by Karl Fischer, a
79
Bùi Xuân Vững Instrumental Analysis Chemistry
80
Bùi Xuân Vững Instrumental Analysis Chemistry
A bipotentiometric measurement is the most common way to detect the end point
of a Karl Fischer titration. The detector circuit maintains a constant current (usually 5
or 10 µA ) between the two detector electrodes at the right in Figure 5.12 while
81
Bùi Xuân Vững Instrumental Analysis Chemistry
measuring the voltage needed to sustain the current. Prior to the equivalence point, the
solution contains I-, but little I2 (which is consumed in Reaction 17-19 as fast as it is
generated). To maintain a current of 10 µA, the cathode potential must be negative
enough to reduce some component of the solvent system. At the equivalence point,
excess I2 suddenly appears and current can be carried at very low voltage by Reactions
A and B in Demonstration 17-2. The abrupt voltage drop marks the end point.
Anode: I3- + 2e- → 3I- Cathode: 3I- → I3- + 2e-
Problems
5-1) Explain the difference between charging current and faradaic current.
5-2) Explain what is done in anodic stripping voltammetry. Why is stripping the most
sensitive polarographic technique?
5-3) Write the chemical reactions that show that 1 mol of I 2 is required for 1 mol of
H2O in a Karl Fischer titration.
5-4) A dropping mercury electrode is regulated such that one drop falls every 4
seconds (t = 4s). The mass of the Hg corresponding to 20 drops is of 0.16 g.
Calculate the average mass flow of Hg in mg/s. If the flow is proportional to the
height of mercury in the reservoir, what would happen to the lifetime of the
drops if the height of Hg was increased three-fold?
5-5) The amount of sulfur in aromatic monomers can be determined by differential
pulse polarography. Standard solutions are prepared for analysis by dissolving
1.000 mL of the purified monomer in 25.00 mL of an electrolytic solvent, adding
a known amount of S, deaerating, and measuring the peak current. The following
results were obtained for a set of calibration standards
82
Bùi Xuân Vững Instrumental Analysis Chemistry
83
Bùi Xuân Vững Instrumental Analysis Chemistry
84
Bùi Xuân Vững Instrumental Analysis Chemistry
85
Bùi Xuân Vững Instrumental Analysis Chemistry
The stationary phase in the form of fine powder is spread on glass or plastic or
aluminum sheets.
b- Paper Chromatography (PC):
A specific type of papers is used as stationary phase in the form of sheets.
Columnar or Column Chromatography (CC):
The stationary phase is held in to a tube made of glass or metal.
86
Bùi Xuân Vững Instrumental Analysis Chemistry
87
Bùi Xuân Vững Instrumental Analysis Chemistry
88
Bùi Xuân Vững Instrumental Analysis Chemistry
- Passing a sufficient mobile phase through the column causes the isolation of
the separated species of A and B (time t2), and then each of them passes out the end of
the column where it can be detected by a suitable detector (time t3 and t4).
Chromatograms
- When the signal of a detector responding to solute concentration and placed at
the end of the column is plotted as a function of time, a series of symmetric peaks is
obtained as shown in Figure 6.2 b.
- A chromatogram is a graph showing the detector response as a function of
elution time.
- A chromatogram, is useful for both qualitative analysis, based upon the
position of a peak on the time axis, and quantitative analysis, based upon the area
under a peak.
Detector
Figure 6.2: (a) the separation of components A and B by column elution chromatography (b)
The output of the signal detector at the various stages of elution shown in (a)
6.2.2. Chromatographic characteristics of solute
6.2.2.1. Retention properties
The chromatogram:
Solutes eluted from a chromatography column are observed with various
detectors described in later chapters. A chromatogram is a graph showing the
detector response as a function of elution time as presented in Figure 6.3.
89
Bùi Xuân Vững Instrumental Analysis Chemistry
Retention time, tr, for each component is the time needed after injection of the
mixture onto the column until that component reaches the detector.
Retention volume Vr is the volume of mobile phase required to elute a
particular solute from the column.
Retention volume: Vr = tr.uv
where uv is the volume flow rate (volume per unit time) of the mobile phase. The
retention volume Vr of a particular solute is constant over a range of flow rates.
Unretained mobile phase travels through the column in the minimum possible
time, designated tm. The adjusted retention time, t’r, for a solute is the additional time
required for solute to travel the length of the column beyond the time required by
unretained solvent:
t’r = tr - tm
In gas chromatography, tm is usually taken as the time needed for CH4 to travel
through the column (Figure ).
6.2.2.2. The capacity factor k’
For each peak in the chromatogram, the capacity factor, k’, is defined as
tr' tr - tm
k'= =
tm tm
The longer a component is retained by the column, the greater is the capacity factor. The capacity
factor in the above equation is equivalent to
90
Bùi Xuân Vững Instrumental Analysis Chemistry
where Cm is the concentration of solute in the stationary phase, Vm is the volume of the
stationary phase, Cs is the concentration of solute in the mobile phase, and Vs is the volume of the
mobile phase.
tr CV V
k'= = s s =K s
tm CmVm Vm
The quotient Cs/Cm is the ratio of concentrations of solute in the stationary and
mobile phases. If the column is run slowly enough to be near equilibrium, the quotient
Cs/Cm is the partition coefficient, K, introduced in connection with solvent extraction.
6.2.2.3. The selectivity factor (The relative retention)
tr' 2 k2' K 2
a= = =
tr' 1 k1' K1
91
Bùi Xuân Vững Instrumental Analysis Chemistry
(a)
Figure 6.4: (a) the broadening of a peak (b) The width w at the baseline and w1/2 at the half of the
peak
6.3.2. Diffusion
One main cause of band spreading is diffusion. The diffusion coefficient D
measures the rate at which a substance moves randomly from a region of high
concentration to a region of lower concentration.
The number of moles crossing each square meter per second, called the flux, J,
is proportional to the concentration gradient:
mol dc
J ( 2 )=−D
m .s dx
Where:
D is the diffusion coefficient,
c is the concentration of the solute,
x is the traveled distance.
Diffusion in liquids is times slower than diffusion in gases. Macromolecules
such as ribonuclease and albumin diffuse 10 to 100 times slower than small
molecules.
If solute begins its journey through a column in an infinitely sharp layer with m
moles per unit cross-sectional area of the column and spreads by diffusion as it
travels, then the Gaussian profile of the band is described by
m −x 2/( 4 Dt )
c= e
√ 4 π Dt
Where:
c is concentration , (mol/m3)
92
Bùi Xuân Vững Instrumental Analysis Chemistry
H = σ2/L
L×W
σ=
Where L is the column length, and 4t R so:
2
16 t R
2
= ƯW
N
93
Bùi Xuân Vững Instrumental Analysis Chemistry
2
5 . 54×t R
2
or N = ƯW 1/ 2
Where:
The larger the number of plates, or the smaller the plate height, the better the
separation efficiency of a column.
Over the last years, an enormous amount of theoretical and experimental effort
has been devoted to developing quantitative relationships that account for the effects
of experimental variables such as linear velocity of mobile phase, diffusion coefficient
in mobile phase and in stationary phase, capacity factor, diameter of packing particles,
thickness of liquid coating on stationary phase, on column efficiency in GC and LC.
The efficiency of most chromatographic columns can be approximated by the
Van Deemter equation:
B B
H = A+ + CS u x + CM u x = A + + Cu x
ux ux
Where:
H is the plate height in centimeters;
ux is the linear velocity of the mobile phase in centimeters per second;
94
Bùi Xuân Vững Instrumental Analysis Chemistry
Where:
o ΔtR and ΔVR are the subtraction of two retention times or two retention
volumes of two peaks, respectively,
o Wav is the average width at the base of two peaks.
95
Bùi Xuân Vững Instrumental Analysis Chemistry
96
Bùi Xuân Vững Instrumental Analysis Chemistry
A and B ( the overlap is about 0.3%), where as a resolution of 0.75 does not and a
resolution of 1 gives the overlap between the two peaks is about 4%.
The resolution for a given stationary phase can be improved by lengthening the
column, thus increasing the number of plates. However, this leads to an adverse
consequence of increasing the time required for the resolution.
6.6. Applications of chromatography
6.6.1. Qualitative analysis
Chromatography is widely used for recognizing the presence or absence of
components in mixture that contain a limited number of species whose indentities are
known. This can be carried out by comparing the retention time of the peak of
suspected species with that of its standard under the same chromatographic conditions,
and the comparison should be carried on columns having different stationary phases.
For example, as can be seen from Figure 6.7, peaks labeled 2,3,4,7 and 9 can be
identified as methyl, ethyl, n-propyl, n-butyl, and n-amyl alcohol, respectively, based
on their retention times.
Another method to identify a peak is to compare its retention time with that of an
authentic sample of the suspected compound. The most reliable way to compare
retention times is by co-chromatography, in which an authentic compound is added to
the unknown. If the added compound is identical with a component of the unknown,
then the relative area of that one peak will increase. Identification is tentative only
when carried out with one column, but it is firmer when carried out on several
columns with different stationary phases.
It is important to note that while a chromatogram may not lead to positive
identification of the species in a sample, it often provides sure evidence of the absence
of species.
Thus, failure of a sample to produce a peak at the same retention time as a
standard obtained under identical conditions is strong evidence that the compound in
question is absent (or present at a concentration below the detection limit of the
procedure).
6.6.2. Quantitative analysis
Quantitative column chromatographv is based on a comparison of either the
height or the area of the analyte peak with that of one or more standards. For planar
chromatography, the area covered by the separated species serves as the analytical
variable. If conditions are properly controlled, these variables vary linearly with
concentration.
6.6.2.1. Internal normalization method
This method, also called ‘normalized to 100 per cent’ is used for mixtures for
which each component is producing a peak on the chromatogram, in order to be able
to make a complete assessment of the sample concerned. The solvent, if any, is
typically ignored.
Supposing that it is required to find themass concentrations of three compounds
1,2,3 in a mixture (Figure 6.8). The analysis is again carried out in two steps.
Step 1: Calculation of the relative response factors
A standard solution containing the three compounds 1,2, and 3 at known
concentrations C1, C2 and C3 is prepared. The chromatogram corresponding to the
injection of a volume V of this standard solution shows three peaks of area A1, A2 and
A3. These areas will be related to the masses m1, m2 and m3 of the compounds in
volume V.
98
Bùi Xuân Vững Instrumental Analysis Chemistry
Given that mi=Ci.V, then the following expressions for K1/3 and K2/3 are obtained:
99
Bùi Xuân Vững Instrumental Analysis Chemistry
be identical. On the resulting chromatogram the area A ref of the corresponding peak is
measured.
The second chromatogram results from the injection of the same volume V of the
sample in solution, containing an unknown concentration of the compound to be
measured (conc. Cunk). The area of the corresponding peak is A unk. Since an identical
volume of both samples has been injected, the ratio of the areas is proportional to the
ratio of concen-
trations which depend upon the masses injected (mi = Ci. V).
The single point calibration method, as depicted in Figure 6.9, assumes that the
calibration line goes through the origin. Precision will be improved if the
concentrations of the reference solution and of the sample solution are similar.
Precision can be also improved if several injections of the sample and the
reference solutions are made, always using equal volumes. In a multilevel calibration,
equal volumes of a series of standard solutions are injected. This allows to get a
calibration curve of A = f (C), obtained by a regression method (linear least-square or
quadratic least-square). This leads to a more precise value for Cunk (Figure 6.10).
For trace analysis it is preferable to use a method that relies on the relative
response factor of each compound to be measured against a marker introduced as a
reference. This means that any imprecision concerning the injected volumes, the
principal constraint of the previous method, is compensated.
The areas of the peaks to be quantified are compared with that of an internal
standard (designated by IS), introduced at a known concentration within the sample
solution.
Supposing that a sample contains two compounds 1 and 2 to be measured and
that compound (IS) represents the additional compound for use as an internal standard
(Figure 6.11).
Similarly, there are two steps for unknown calculation:
Step1: Calculation of the relative response factors
A solution containing compound 1 at known concentration C 1, compound 2 at
known concentration C2 and the internal standard IS at known concentration C IS is
prepared then injected onto the chromatograph. A1, A2, AIS will be the areas of the
elution peaks in the chromatogram due to the three compounds. If m 1, m2 and mIS
represent the real quantities introduced onto the column, of these three substances,
then three relations can be derived:
These ratios enable the calculation of the relative response factors of 1 and 2,
against IS and designated by K1/IS and K2/IS:
101
Bùi Xuân Vững Instrumental Analysis Chemistry
From the relative response factors calculated in the first experiment as well as
from the known concentration of the internal standard within the sample, C’ IS, this
leads to:
This method becomes even more precise if several injections of the solution and
of the sample are carried out. Often a same volume of an IS stock solution is spiked
with all standards and samples.
In conclusion, this general and reproducible method demands nevertheless a
good choice of internal standard, which should have the following characteristics:
• it must be stable, pure and not exist in the initial sample
• it must be measurable, giving an elution peak well resolved on the
chromatogram
• its retention time must be close to that (or those) of the solute(s) to be
quantified
• its concentration must be close to, or above that of the analytes to quantify to
gain a linear response from the detector
• it must not interfere and co-elute with a sample component.
Problems
102
Bùi Xuân Vững Instrumental Analysis Chemistry
103
Bùi Xuân Vững Instrumental Analysis Chemistry
104
Bùi Xuân Vững Instrumental Analysis Chemistry
105
Bùi Xuân Vững Instrumental Analysis Chemistry
106
Bùi Xuân Vững Instrumental Analysis Chemistry
107
Bùi Xuân Vững Instrumental Analysis Chemistry
Flow rates can be measured at the column head by using a soap-bubble meter.
Many modern computer-controlled gas chromatographs are equipped with electronic
flow meters that can be regulated to maintain the flow rate at the desired level.
Carrier gas
The nature of the carrier gas has no significant influence upon the values of the
partition coefficients K of the compounds between the stationary and mobile phases,
owing to an absence of interaction between the gas and solutes.
By contrast, the viscosity of the carrier gas and its flow rate have an effect on the
analytes’ dispersion in the stationary phase and on their diffusion in the mobile phase
(cf. Van Deemter’s equation), and by consequence upon the efficiency N and the
sensitivity of detection. Hydrogen is the carrier gas of choice. If this gas cannot be
used for safety reasons, helium may be substituted.
7.2.3. Sample Injection Systems
7.2.3.1. Sample introduction
The most common injection method is where a microsyringe is used (Figure 7.3)
to inject a very small quantity of sample in solution (e.g. 05L), through a rubber
septum into a flash vaporizer port at the head of the column.
108
Bùi Xuân Vững Instrumental Analysis Chemistry
7.3. Columns
There are two column types, which differ in their performance: packed columns
and capillary, also called open tubular, columns (Figure 7.6). For packed columns the
stationary phase is deposited or bonded by chemical reaction onto a porous support.
For capillary columns a thin layer of stationary phase is deposited onto, or bound to
the inner surface of the column.
109
Bùi Xuân Vững Instrumental Analysis Chemistry
In the past, the vast majority of gas chromatographic analyses used packed
columns. For most current applications, packed columns have been replaced by the
more efficient open tubular columns.
(b)
(a)
Figure 7.6: Typical capillary columns (a) and packed column (b)
Figure 7.7: The cross-section of (a) a packed column and (b) a capillary column
linker assures the cohesion of the grains. After calcinations these diatomaceous earths
are often designated by the name of Chromosorb®. Other synthetic materials have
been developed such as Spherosil®, composed of tiny silica beads. All of these
supports have a chemical reactivity comparable to silica gel because of the presence of
silanol groups.
Although the performance of packed columns is more modest than capillary
columns, they are still usually employed for many routine analyses. Easy to
manufacture and with a large choice of stationary phases available, they are not
however, well adapted to trace analyses.
7.3.2 Capillary columns (open tubular columns)
The vast majority of analyses use long, narrow open tubular columns (Figure
7.6a) made of fused silica (SiO2) and coated with polyimide (a plastic capable of
withstanding 350°C) for support and protection from atmospheric moisture. Open
tubular columns offer higher resolution, shorter analysis time, and greater sensitivity
than packed columns, but have less sample capacity.
The wall-coated column in Figure 7.7b and 7.8a features a 0.1- to 5 µm thick
film of stationary liquid phase on the inner wall of the column. A support-coated
column has solid particles coated with stationary liquid phase and attached to the inner
wall. In the porous-layer column in Figure, solid particles are the active stationary
phase. With their high surface area, support-coated columns can handle larger samples
than can wall-coated columns. The performance of support-coated columns is
intermediate between those of wall-coated columns and packed columns.
(b)
Figure 7.8: (a) A wall coated column (b) Types of tubular columns
Column inner diameters are typically 0.10 to 0.53 mm and lengths are 15 to 100
m, with 30 m being common. Narrow columns provide higher resolution than wider
columns (Figure ) but require higher operating pressure and have less sample capacity.
Diameters of 0.32 mm or greater tend to overload the gas-handling system of a mass
spectrometer, so the gas stream must be split and only a fraction is sent to the mass
spectrometer. The number of theoretical plates, N, on a column is proportional to the
111
Bùi Xuân Vững Instrumental Analysis Chemistry
112
Bùi Xuân Vững Instrumental Analysis Chemistry
Squalane (nonpolar)
HC-(CH2)3-CH-(CH2)3-CH-(CH2)4-CH-(CH2)3-CH-(CH2)3-CH
113
Bùi Xuân Vững Instrumental Analysis Chemistry
7.4. Detector
7.4.1. Thermal conductivity detector
In the past, thermal conductivity
Table 7.3: Thermal conductivity of some gases detectors were most common in gas
chromatography because they are
simple and universal: They respond to
all analytes. Unfortunately, thermal
conductivity is not sensitive enough to
detect minute quantities of analyte
eluted from open tubular columns
smaller than 0.53 mm in diameter.
Thermal conductivity detectors are still
used for 0.53-mm columns and for
packed columns.
114
Bùi Xuân Vững Instrumental Analysis Chemistry
115
Bùi Xuân Vững Instrumental Analysis Chemistry
(a) (b)
Figure 7.10: (a) the cross-section and (b) the diagram of FID
116
Bùi Xuân Vững Instrumental Analysis Chemistry
117
Bùi Xuân Vững Instrumental Analysis Chemistry
118
Bùi Xuân Vững Instrumental Analysis Chemistry
Problems
7-1) Why do capillary columns predominate in analytical GC?
7-2) What is temperature programming in GC? How does it gain an advantage over
single T separations?
7-3) What is the electron capture detector? Explain its basis for operation, why is N2
necessary? What types of species are detected with the ECD?
7-4) How does the capillary column configuration achieve its advantages over the
packed column setup in gas chromatography?
7- 5) What is an FID and how does it work?
What types of analytes does the FID respond to?
7-6) The analysis of the trihalomethanes CHCl 3, CHBr3, CHCl2Br, and CHClBr2 in
drinking water uses a packed column with a nonpolar stationary phase. Predict
the order in which these four trihalomethanes will elute.
7-7) A mixture of n-heptane, tetrahydrofuran, 2-butanone, and n-propanol elutes in
this order when using a polar stationary phase such as Carbowax. The elution
order is exactly the opposite when using a nonpolar stationary phase such as
polydimethyl siloxane. Explain the order of elution in each case.
7-8) Zhou and colleagues determined the %w/w H 2O in methanol by GC, using a
capillary column coated with a nonpolar stationary phase and a thermal
conductivity detector. A series of calibration standards gave the following
results.
119
Bùi Xuân Vững Instrumental Analysis Chemistry
(a) What is the %w/w H2O in a sample giving a peak height of 8.63?
(b) The %w/w H2O in a freeze-dried antibiotic is determined in the following
manner. A 0.175-g sample is placed in a vial along with 4.489 g of methanol.
Water in the vial extracts into the methanol. Analysis of the sample gave a peak
height of 13.66. What is the %w/w H2O in the antibiotic?
7-9) A GC‐FID analysis was conducted on a soil sample containing pollutant X. The
following separations were conducted:
tr(minutes)
peak area
Injection 1: 21.1 ppm Toluene Internal Standard 10.11 36,242
33.4 ppm X 14.82
45,997
Injection 2: 21.1 ppm Toluene Internal Standard 10.05
38,774
unknown concentration X 14.77
39,115
What is the concentration of X in the sample?
7-10) The amount of camphor in an analgesic ointment can be determined by GC
using the method of internal standards. A standard sample was prepared by
placing 45.2 mg of camphor and 2.00 mL of a 6.00 mg/mL internal standard
solution of terpene hydrate in a 25-mL volumetric flask and diluting to volume
with CCl4. When an approximately 2-µL sample of the standard was injected, the
FID signals for the two components were measured (in arbitrary units) as 67.3
for camphor and 19.8 for terpene hydrate. A 53.6-mg sample of an analgesic
ointment was prepared for analysis by placing it in a 50-mL Erlenmeyer flask
along with 10 mL of CCl4. After heating to 50 °C in a water bath, the sample was
cooled to below room temperature and filtered. The residue was washed with
two 5-mL portions of CCl4, and the combined filtrates were collected in a 25-mL
volumetric flask. After adding 2.00 mL of the internal standard solution, the
contents of the flask were diluted to volume with CCl4. Analysis of an
120
Bùi Xuân Vững Instrumental Analysis Chemistry
approximately 2-µL sample gave FID signals of 13.5 for the terpene hydrate and
24.9 for the camphor. Report the %w/w camphor in the analgesic ointment.
121
Bùi Xuân Vững Instrumental Analysis Chemistry
and using packings with particle diameters as small as 3 to 10 µm was developed. This
technology required sophisticated instruments operating at high pressures, which
contrasted markedly with the simple glass columns of classic gravity-flow liquid
chromatography. The name high-performance liquid chromatography (HPLC) was
originally used to distinguish these newer procedures from the original gravity-flow
methods. Today, virtually all LC is done using pressurized flow, and we use the
abbreviations LC and HPLC interchangeably.
LC is the most widely used of all of the analytical separation techniques. The
reasons for the popularity of the method are its sensitivity. its ready adaptability to
accurate quantitative determinations, its ease of automation, its suitability for
separating nonvolatile species or thermally fragile ones, and above all, its widespread
applicability to substances that are important to industry, to many fields of science,
and to the public.
In several basic types of chtomatography the mobile phase is a liquid. These
types are often classified by separation mechanism or by the type of stationary phase.
The varieties include: (1) partition chtomatography (2) adsorption, or liquid-solid
chtomatography (3) ion exchange, or ion chtomatography; (4) size exclution
chromatography; (5) affinity chtomatography, and (6) chiral chromatography; most of
this chapter deals with column applicatiors of these important types of
chromatography. The final section, however, presents a brief description of planar
liquid chromatography because this technique provides a simple and inexpensive way
of determining likely optimal conditions for column separations.
8.2. Main components of a HPLC
Figure 8.1 illustrates main components of a currently common HPLC
chromatography.
122
Bùi Xuân Vững Instrumental Analysis Chemistry
Figure 8.2: High-pressure piston pump for HPLC [Courtesy Hewlett-Packard Co., Palo Alto, CA.].
124
Bùi Xuân Vững Instrumental Analysis Chemistry
The injection valve in Figure 8.3 has interchangeable sample loops, each of
which
holds a fixed volume. Loops of different sizes hold volumes that range from 2 to 1000
µL. In the load position, a syringe is used to wash and load the loop with fresh sample
at atmospheric pressure. High-pressure flow from the pump to the column passes
through the segment of valve at the lower left. When the valve is rotated 60°
counterclockwise, the content of the sample loop is injected into the column at high
pressure.
Figure 8.3: Injection valve for HPLC. Replaceable sample loop comes in various fixed-volume sizes.
The HPLC equipment in Figure 8.1 decreases the viscosity of the solvent,
uses steel or plastic columns that are 5– thereby reducing the required pressure
30 cm in length, with an inner diameter or permitting faster flow.
of 1–5 mm (Figure 8.4). Columns are The most common HPLC column
expensive and easily degraded by dust diameter is 4.6 mm. There is a trend
or particles in the sample or solvent and toward narrower columns (2 mm, 1
by irreversible adsorption of impurities mm, and capillary columns down to 25
from the sample or solvent. Therefore, µL) for several reasons. Narrow
the entrance to the main column is columns are more compatible with
protected by a short guard column mass spectrometers, which require low
containing the same stationary phase as solvent flow. Narrow columns require
the main column. Fine particles and less sample and produce less waste.
strongly adsorbed solutes are retained
in the guard column, which is
periodically replaced. Heating a
chromatography column usually
125
Bùi Xuân Vững Instrumental Analysis Chemistry
126
8.2.4.2. Photodiode Aray Detector
The system in Figure 8.6 is able to use a photodiode array to record the spectrum
of each solute as it is eluted. Dual-beam optical system uses grating polychromator, one
diode array for the sample spectrum, and another diode array for the reference spectrum.
By this means, it was possible to determine which compounds were in each peak.
In a photodiode array spectrophotometer (Figure 8.6), all wavelengths are recorded
simultaneously, allowing more rapid acquisition of the spectrum or higher signal-to-noise
ratio, or some combination of both.
Each diode receives a different wavelength, and all wavelengths are measured
simultaneously.
Figure 8.6: Photodiode array ultraviolet detector for HPLC. (a) Dual-beam optical system (b)
chromatogram of sample containing 0.2 ng of anthracene, with detection at 250 nm.
Full-scale absorbance is 0.001. (c) Spectrum of anthracene recorded as it emerged from
the column. [Courtesy Perkin-Elmer Corp., Norwalk, CT.]
Figure 8.8: Schematic structure of silica particle. [From R. E. Majors, LCGC May 1997, p. S8.]
8.3.2. Partition Chromatography (Liquid-Liquid chromatography)
The most widely used type of HPLC is partition chromatography, in which
the stationary phase is a second liquid that is immiscible with the liquid mobile
phase. In the past, most of the applications have been to nonionic, polar
compounds of low to moderate molecular mass (usually <3000).
The early forms of partition chromatography used liquid-liquid columns.
These have been replaced in modern LC systems by liquid bonded-phase columns.
In liquid-liquid chromatography, the liquid was held in place by physical
adsorption. In bonded-phase chromatography, on the other hand, it is attached by
chemical bonding, resulting in highly stable packings insoluble in the mobile
phase. Bonded-phase columns are also compatible with gradient elution
techniques.
Therefore, our discussion focuses exclusively on bonded-phase partition
chromatography.
Columns for Bonded-Phase Chromatography
The supports for the majority of bonded-phase packings for partition
chromatography are prepared from rigid silica, or silica-based, compositions.
These solids are formed as uniform, porous, mechanically sturdy particles
commonly having diameters of 1.5-10 µm with 3- and 5- µm particles being most
common. The surface of fully hydrolyzed silica (hydrolyzed by heating with 0.1 M
HCI for a day or two) is made up of chemically reactive silanol groups. Typical
silica surfaces contain about 81 µmol/m2 of OH groups and have surface areas of
100 to 300 m2/g.
The most useful bonded-phase coatings are siloxanes formed by reaction of
the hydrolyzed surface with an organochlorosilane. For example,
Mobile phases
Depending upon the type of stationary phase, the counter ions present in the mobile
phase derived from acids (perchloric, benzoic, phthalic, methane sulfonic), or bases (the
most popular for anion analyses are variants of sodium hydroxide and sodium
carbonate/bicarbonate).
Figure 8.11: Elution of lanthanide(III) ions from a cation-exchange resin. Eluent was varied from
20 to over 25 min.
Conductivity detectors
Besides the spectrophotometric detectors based on absorbance or fluorescence of
uv/visible radiation, and used when the mobile phase does not absorb appreciably,
another mode of detection exists based upon electrolyte conductivity.
Thus, at the outlet of the column, the conductance (the inverse of the
resistance) of the mobile phase is measured between two microelectrodes. The
measuring cell should be of a very small volume (approx. 2µL).
The difficulty is to recognize in the total signal the part due to ions or ionic
substances present in the sample. In order to do direct measurements, the ionic
charge of the mobile phase has to be as low as possible and the measuring cell
requires strict temperature control to within 0.01 oC because of the high
dependence of conductance on temperature (∼5%/).
Ion suppressors
The mobile phase contains ions that create a background conductivity, making it
difficult to measure the conductivity due only to the analyte ions as they exit the column.
To improve the signal to noise ratio, when using a conductivity detector, a device called a
suppressor, designed to selectively remove the mobile phase ions is placed after the
analytical column and before the detector. The principle consists to convert the mobile
phase ions to a neutral form or replacing them by others of higher conductivity (Figure
8.12).
.
Figure 8.12: Schematic illustrations of (a) suppressed-ion anion chromatography and (b)
suppressed-ion cation chromatography.
8.3.5. Molecular Exclusion Chromatography
In molecular exclusion chromatography (also called size exclusion or gel filtration
or gel permeation chromatography), molecules are separated according to size. Small
molecules penetrate the pores in the stationary phase, but large molecules do not. Because
small molecules must pass through an effectively larger volume, large molecules are
eluted first. This technique is widely used in biochemistry to purify macromolecules.
Size-exclusion. or gel, chromatography is a powerful technique that is particularly
applicable to highmolecular-mass species." Packings for size-exclusion chromatography
consist of small (-10 flm) silica or polymer particles containing a network of uniform
pores into which solute and solvent molecules can diffuse. While in the pores, molecules
are effectively trapped and removed from the flow of the mobile phase. The average
residence time in the pores depends on the effective size of the analyte molecules.
Molecules larger than the average pore size of the packing are excluded and thus suffer
essentially no retention; such species are the first to be eluted. Molecules having
diameters significantly smaller than the pores can penetrate or permeate throughout the
pore maze and are thus entrapped for the greatest time; these are last to be eluted.
Between these two extremes are intermediate size molecules whose average penetration
into the pores of the packing depends on their diameters.
Within this group, fractionation occurs, which is directly related to molecular size
and to some extent molecular shape. Note that size-exclusion separations differ from the
other procedures we have been considering in that no chemical or physical interaction
between an-alytes and the stationary phase are involved.
Column packing
Two types of packing for size-exclusion chromatography are encountered: polymer
bcads and silica-based particles, both of which have diameters of 5 to 10 µm.
Most early size-exclusion chromatography was carried out; n cross-linked styrene-
divinylbenzene copolymers similar in structure (except that the sulfonic acid groups arc
absent) to that shown in Figure 8.10. The pore size of these polymers is controlled by the
extent of cross-linking and hence the relative amount of divinylbenzene present during
manufacture. Polymeric packings having several different average pore sizes are thus
marketed. Originally, styrene-divinylbenzene gels were hydrophobic and thus could be
used only with nonaqueous mobile phases. Now, however, hydrophilic gels are available,
making possible the use of aqueous solvents for the separation of large, water-soluble
molecules such as sugars. These hydrophilic gels are sulfonated divinylbenzenes or
polyacrylamides (Figure 8.14). Chromatography based on the hydrophilic packings was
once called gel filtration, and techniques based on hydrophobic packings were termed gel
permeation. Today, both techniques are described as size-exclusion methods.
Figure 8.15: Developing chamber and TLC plate. Left, available in a variety of dimensions
According to the size of the plates (of 5×5to20×20 cm), the chambers are made of
glass and equipped with a tight fitting cover. Right, typical appearance after the TLC
plate is partially dry. A faint line can be observed at the location of the solvent upper
limit. This line is called the solvent front. Calculation of R f (cf. paragraph 5.4). A
substance that does not migrate from the sample origin has a R =0.
8-6) An HPLC analysis was conducted for caffeine on a sports drink. A 10.1 ppm
methanol IS standard was introduced both into the sample and a standardized
solution of 304 ppm of caffeine. The measured by a diode‐array detector at each
λmax for the absorptions for methanol and for caffeine are summarized in the table
below. What is the concentration of caffeine in that sports drink?
IS Caffeine
Sample 23141 52777
304 ppm Caffeine standard 28441 77313
8-7) The concentration of PAHs in soil can be determined by first extracting the PAHs
with methylene chloride. The extract is then diluted, if necessary, and the PAHs are
separated by HPLC using a UV/Vis or fluorescence detector. Calibration is achieved
using one or more external standards. In a typical analysis, a 2.013-g sample of
dried soil is extracted with 20.00 mL of methylene chloride. After filtering to
remove the soil, a 1-mL portion of the extract is removed and diluted to 10 mL with
acetonitrile. Injecting 5 µL of the diluted extract into an HPLC gives a signal of
0.217 (arbitrary units) for the PAH fluoranthene. When 5 µL of a 20.0-ppm
fluoranthene standard is analyzed using the same conditions, a signal of 0.258 is
measured. Report the parts per million of fluoranthene in the soil.
8.8) The amount of caffeine in an analgesic tablet was determined by HPLC using a
normal calibration curve. Standard solutions of caffeine were prepared and analyzed
using a 10-µL fixed-volume injection loop. Results for the standards are
summarized in the following table.
The sample was prepared by placing a single analgesic tablet in a small beaker and
adding 10 mL of methanol. After allowing the sample to dissolve, the contents of
the beaker, including the insoluble binder, were quantitatively transferred to a 25-
mL volumetric flask and diluted to volume with methanol. The sample was then
filtered, and a 1.00-mL aliquot was transferred to a 10-mL volumetric flask and
diluted to volume with methanol. When analyzed by HPLC, the signal for the
caffeine was found to be 21469. Report the number of milligrams of caffeine in the
analgesic tablet.
8-9) The concentrations of Cl–, NO3–, and SO42– may be determined by ion
chromatography. A 50-µL standard sample of 10.0-ppm Cl –, 2.00-ppm NO3–, and
5.00-ppm SO42– gave signals (in arbitrary units) of 59.3, 16.1, and 6.08, respectively.
A sample of effluent from a wastewater treatment plant was diluted tenfold, and a
50-µL portion gave signals of 44.2 for Cl–, 2.73 for NO3–, and 5.04 for SO42–. Report
the parts per million of each anion in the effluent sample.
8-10) A series of polyvinylpyridine standards of different molecular weight were
analyzed by size-exclusion chromatography, yielding the following results.