Академический Документы
Профессиональный Документы
Культура Документы
ENVIRONMENTAL ENGINEERING
(BACHELOR OF TECHNOLOGY)
PREPARED BY
JAYPEE UNIVERSITY OF
ENGINEERING & TECHNOLOGY
A.B. ROAD RAGHOGARH-GUNA (M.P.)-473226 (INDIA)
CONTENTS
Ex No Experiment Name Page No
NAME OF EXPERIMENT: pH
Theory:
pH value denotes hydrogen ion concentration in the liquid and it is the measure of acidity or
alkalinity of the liquid. According to the law of mass action, in any liquid
= constant = 10-14
Principle:
Apparatus:
1. pH-strips (papers),
2. pH meter,
3. thermometer
Reagents:
1) Distilled water
2) Standard buffer solutions: Standard buffer solutions having pH values of 4.0 and 9.2 are
readily available. Otherwise, they can be prepared easily by dissolving 1 pH tablet of
each buffer in distilled water and make up to 100 ml gives a standard solution of pH 4.0
and 9.2.
1. Using pH-paper:
(i) Dip a wide-range (0-14) pH-strip (paper) into the solution whose pH to be found. The
color of the litmus paper changes to thick red for highly acidic waters to dark green for
highly alkaline waters and to any other color depending on the pH of the solution.
Compare the color of paper with the standard colours supplied. Note down the pH by
noting down the number corresponding to the matching color.
(ii) Select a suitable narrow-range pH-strip and compare the color by using the same
method with wide-range pH-strips. Note down the pH by comparing the changed color
with standard colours. Narrow-range pH-strips are normally available in the following
pH ranges: 2 to 4.5; 4 to 6.5; 6 to 9.5; 8.5 to 10.5; 10 to 12.
2. Using pH-meter:
(i) Switch on the pH meter for 15 minutes.
(ii) After washing and wiping the pH electrode and the temperature probe dip it in a
solution of pH 4.0 buffer. Change knob from standby to pH.
(iii) With the CAL knob set the pH value to 4.0
(iv) With a pH 9.2 buffer, set the pH value to 9.2 using the SLOPE knob.
(v) Repeat steps 2-3 till the pH meter is standardized with respect to both pH 4.0 and 9.2
(vi)Take the pH values of the different water samples with the pH meter.
SIGNIFICANCE:
2. Corrosion of water mains is the main the main problem associated with acidic
waters. Acidic/alkaline waters cannot be used for construction purposes also. If pH is
less, algae die, fish cannot reproduce and it causes acidity, corrosion, irritation of
mucous membranes, tuberculosis and other health problems in humans.
Observation Table:
Precautions:
Principle:
When light is passed through a sample having suspended particles some of the light is scattered
by the particles. The scattering of the light is generally proportional to the turbidity. The
turbidity of sample is thus measured from the amount of light scattered by the sample taking a
reference with standard turbidity suspension.
Apparatus
Reagents:
Procedure:
1. Switch on Nephelometeric turbidity meter and wait for few minutes till it warms up.
2. Set the instrument at 100 on the scale with a 40 NTU standard suspension. In this case
every division on the scale will be equal to 0.4 NTU turbidity.
3. Shake thoroughly the sample and keep it for sometime to eliminate the air bubbles.
4. Take sample in Nephelometer sample tube and put the sample in sample chamber and
find out the value on the scale.
5. Dilute the sample with turbidity free water and again read the turbidity.
Environmental significance:
1. Turbidity is objectionable because of aesthetic and engineering considerations. The
colloidal material exerts turbidity provides adsorption sites for chemical that may be
harmful. In natural waters, turbidity may impart a brown or other color to water and be
Observation Table:
Precautions:
Aim:
To determine the Conductivity of the given water samples.
Principle:
The electrical conductivity is a total parameter for dissolved, dissociated substances. Its value
depends on the concentration and degrees of dissociation of the ions as well as the
temperature and migration velocity of the ions in the electric field.
Procedure:
Significance:
1. Dissolved minerals, gases and organics may produce aesthetically displeasing colours,
taste and odours to water.
2. EC measurements are the basis to evaluate the performance of the processes of
desalination, demineralization, distillation and many process plants.
3. salinity and total dissolved solids can be estimated very quickly from conductivity.
4. If conductivity of a water sample is more than 2000µmhos/cm i.e. if TDS exceeds 2100
mg/1, it cannot be disposed off without treatment.
5. If TDS is more, water cannot be used for drinking as well as construction purposes. TDS
affects strength and solidity of concrete and palatability of food cooked. It also causes
gastro intestinal irritation.
Precautions:
1. The electrode must be dipped into the sample carefully lest it is broken.
2. If the ECL falls outside the range of the instrument, dilute the sample, with distilled
water and then estimate the EC.
3. Keep the electrode immersed in distilled water always and at the end of each
measurement wash it thoroughly with distilled water. Organic material coating, if any,
can be removed with alcohol or acetone followed by distilled water.
Aim:
To determine acidity (base capacity) of the given water samples.
Principle:
The mineral acids present in the sample which are contributing mineral acidity can be
calculated by titrating or neutralizing samples with strong base NaOH to pH 4.3. The CO2 and
bicarbonates (carbonic acid) present and contribute CO2 acidity in the sample can be
neutralized completely by continuing the titration to p H 8.2
Apparatus:
1. Burette,
2. pipettes,
3. conical flask.
Reagent Required:
Procedure:
1. Pipette out 100 ml of the given water sample into a conical flask.
2. Add 1 drop of 0.1N sodium thiosulfate solution to destroy any residual chlorine.
3. Add 2 drops of methyl orange indicator. The sample turns pink.
4. Titrate against 0.02N standard sodium hydroxide solution until pink color changes to
yellow.
5. Note down the volume of the NaOH added (V1).
6. Take another conical flask containing 100 ml of water sample, add 2 drops of
phenolphthalein.
7. Proceed with titration until the sample turns pink.
8. Note down the total volume of NaOH added (V2).
Observation table:
Model Calculations:
Mineral acidity due to mineral acids (as CaCO3) (mg/l) = V1 X 1000/ml. of sample taken
CO2 acidity due to CO2 (as CaCO3) (mg/l) = V2 X 1000/ml. of sample taken
Precautions:
1. Cleaned apparatus & glassware should be used for the preparation.
2. All the reagents should be prepared freshly and standardized.
3. To avoid loss of CO2, titration should be carried out quickly & vigorous shaking should be
avoided.
Alkalinity can be obtained by neutralizing OH-, CO3--- and HCO3- with standard H2SO4. Titration
to pH 8.3 or decolourization of phenolphthalein indicator will show complete neutralization of
OH- and ½ of CO3--, while to pH 4.4 or sharp change from yellow to pink of methyl orange
indicator will indicate total alkalinity i.e. OH-, CO3---, and HCO3-.
Apparatus:
Reagents:
Procedure:
Significance:
1. Highly alkaline water are usually unpalatable and consumer acceptance decreases.
Water having an alkalinity of less than 200 mg/l as CaCO3 is desirable for drinking.
2. Large amount of alkalinity imparts a bitter taste to water.
Model Calculation:
Precautions:
1. The glass apparatus should be perfectly cleaned before the start of the experiment.
2. Since phenolphthalein and methyl orange indicators are used in determination of
alkalinity the end point of the titration should be observed carefully.
EXPERIMENT NO.: 3(a)
Department of Civil Engineering, JUET Guna (M.P.): 13
NAME OF EXPERIMENT : CHLORIDES
Aim:
To estimate the content of chlorides in the given water samples.
Principle:
The Mohr method for the determination of chloride in water is based upon the fact that in
solution containing chloride and chromate, silver reacts with all the chloride and precipitates
before the reaction with chromate begins. The appearance of the brick-red colour of the silver
chromate precipitate is the end-point of the titration.
Apparatus:
Reagents:
Procedure:
7. Note the volume of silver nitrate added for distilled water (B).
1. Analysis of a water sample is necessary to know whether the sample may be considered as
suitable for drinking, construction, hydro-testing and some industrial process or not.
2. Water containing chloride in excess of 250 mg/l is considered to be undesirable for drinking
purposes.
3. Chlorides are also corrosive and impart permanent hardness to water. Waters with chloride
in excess of 2000 mg/l are not recommended for many construction purposes also.
Observation table:
1.
2.
3.
Model Calculations:
Where
A = ml of AgNO3 required for sample
Precautions:
1. The whole apparatus should be washed with distilled water before the start of the
experiment.
2. The reaction mixture should be briskly shaken during the titration.
3. The end point of the reaction should be carefully observed.
4. The volume of the indicator should be same in all the titrations.
5. The pH of the sample should be adjusted to 7-8 range by adding acidic/basic solution to
the sample solution.
Aim:
To determine the fluoride in the given water sample.
Principle:
The test is based on the fact that fluoride ion combines with zirconium ion to form a stable
complex ion, ZrF6 and this results in bleaching the reddish colour of zirconium and alizarin
combination. The decrease in intensity of colour is directly proportional to Fluoride
concentration.
Reagents:
Procedure:
1. The sample should be free from chlorine, if chlorine present, it shall be dechlorinated
with a slight excess of sodium thiosulphate solution before use.
2. Take 0, 1, 2, 6, 8, 14 ml of standard sodium fluoride solution in six Nessler tubes.
3. Add 5 ml of acid zirconium reagent in each Nessler tube.
4. Similarly add 5 ml of acid zirconium reagent into the Nessler tubes containing 100 ml of
sample.
5. Mix thoroughly and compare the colours after standing for one hour exactly.
Significance:
1. During the formation of permanent teeth, fluoride combines with enamel of the teeth to
form strong, hard and durable teeth. As such many municipal water supplies are
supplemented with fluoride by addition of fluoride compounds. If fluoride is more than
1.5 mg/I it leads to discoloration of teeth called ‘’mottling’’. The white patches formed
later become yellow and turn brown or black. Fluoride in excess of 5 ppm, causes bone
Fluorosis and skeletal abnormalities.
2 Fluoride of water is an important in determining the suitability of water for domestic
and industrial use.
3. The size and design of defluoridation units depend on the level of fluoride present in
water. Excess Fluoride can be removed by defluoridation with activated charcoal or
Nalgonda technic using alum and lime.
Aim:
To determine the total hardness, calcium and magnesium of the given water sample.
Theory:
Hardness in water is defined as soap consuming capacity of water. This is mainly due to
presence of calcium and magnesium ions in water. The hardness of water is taken as a measure
of Ca++ and Mg++ contents. The hardness is temporary and permanent. When water is boiled,
bicarbonates are decomposed to form carbonate ions and CO2 is driven off. Thus total hardness
is eliminated. The permanent hardness is not removed by boiling.
Hardness is total hardness of water which is determined by complex metric method using
EDTA. EDTA forms a complex with Ca++ and Mg++. The complexes exist at pH 8 to 10. The indicator
used is Eriochrome Black T. The color change with Eriochrome Black T is in the presence of Mg ++.
If Mg++ ions are not present in solution, they must be added.
Reagents:
1. Buffer solution
2. Standard EDTA Solution 0.01 M
3. Eriochrome Black T indicator
4. Standard calcium solution
5. Murexide indicator
6. Sodium Hydroxide 2N
Procedure:
A. Total hardness:
(i) Take 100 ml of given water sample in a conical flask.
(ii) Add 2 ml of buffer solution.
(iii) Add 2 drops of Eriochrome Black T indicator. The color becomes wine red.
(iv) Titrate with standard EDTA solution till the color changes to blue.
(v) Note down the volume of EDTA solution used (A) ml.
(vi) Run a reagent blank if buffer is not checked properly. Note the volume of EDTA
required for blank (B) ml.
Calculations:
(A - B) x 1000
Total Hardness (mg/l) as CaCO3 = ----------------
ml of sample
Calculations:
A1 x 1000
Calcium Hardness (mg/l) as CaCO3 = ----------------
ml of sample
C. Magnesium hardness:
Significance:
1. Absolutely soft waters are tasteless (e.g. distilled water). On the other hand, hardness upto
600 mg/l can be relished if got acclimatized to.
2. Moderately hard water is preferred to soft water for drinking and irrigation purposes.
3. Hard waters form scales in hot water pipes and boilers.
4. Soft waters affects the human cardiovascular system and causes heart-attacks.
Calcium
hardness
Magnesium hardness
1. The glass wares namely burette, pipette, beakers should be rinsed only with the distilled
water.
2. All the solutions should be checked with care before used
3. Distilled water should be checked with care before use.
4. The same amount of indicator must be added each time.
5. The reaction mixture should be briskly shaken during the titration.
6. The end point should be observed carefully.
7. pH 10 should be maintained during the titration.
Aim:
Principle:
Total solids are determined as the residue left after evaporation and drying of the unfiltered
sample.
Apparatus:
Procedure:
1. A clean porcelain dish is ignited in a muffle furnace and after partial cooling in the air; it is
cooled in a desiccator and weighed.
2. A 100 ml of well mixed sample (graduated cylinder is rinsed to ensure transfer of all
suspended matter) is placed in the dish and evaporated at 1000C on water bath, followed by
drying in oven at 1030C for 1 hour.
3. Dry to a constant weight at 1030C, cool in a desiccator and weigh.
Significance:
1. Total solids determination is used to assess the suitability of potential supply of water
for various uses.
Observation Table
(A-B) x 1000
Total solids (mg/l) = --------------------
V
Aim:
Principle:
Total dissolved solids are determined as the residue left after evaporation and drying of the
filtered sample.
Apparatus:
Procedure:
1. A clean porcelain dish is ignited in a muffle furnace and after partial cooling in the air; it
is cooled in a desiccator and weighed.
2. A 100 ml of filtered sample is placed in the dish and evaporated at 100 0C on water bath,
followed by drying in oven at 1030C for 1 hour.
3. Dry to a constant weight at 1030C, cool in a desiccator and weigh.
Significance:
1. Many dissolved substances are undesirable in water. Dissolved minerals, gases and
organic constituents may produce aesthetically displeasing colour, taste and odour.
2. Estimation of total dissolved solids is useful to determine whether the water is suitable
for drinking purposes, agriculture and industrial processes or not.
3. High concentration of dissolved solids about 3000 mg/I may also produce distress in
livestock. In industries, the use of water with high amount to dissolved solids may lead
to scaling in boilers, corrosion and degraded quality of the product.
4. Water with higher solids content often has a laxative effect.
5. Refer to experiment on electrical conductivity (EC) also.
Calculations:
(A - B) x 1000
Total dissolved solids (mg/l) = --------------------
V
Where,
A = Final weight of the dish in mg
B = Initial weight of the dish in mg
V = Volume of sample taken in ml
Aim:
Principle:
Total suspended solids are determined as the residue left on gooch crucible or a glass fiber
after drying in oven.
Apparatus
Procedure :
1. A clean gooch crucible is ignited in a muffle furnace and after partial cooling in the air,
cool in a desiccator and weigh (W1).
2. Pour 100 ml of well mix sample on gooch crucible or glass fiber filter which is kept on
filter flask and apply suction.
3. Wash the gooch crucible with 100 ml of distilled water to remove all soluble salts.
4. Carefully remove the glass fiber filter paper or gooch crucible and dry in an oven at
1050C for one hour.
5. Cool in a desiccator and weigh (W2).
6. Ignite gooch crucible in a muffle furnace at 6000C for 15-20 minutes.
7. Cool the crucible partially in air until most of heat has been dissipated and then in a
dessicator and record final weigh (W3).
Significance:
1. Suspended solids may be objectionable in water for several reasons. It is aesthetically
displeasing and provides adsorption sites for chemical and biological agents.
2. Suspended organic solids which are degraded anaerobically may release obnoxious
odours.
3. Biologically active (live) suspended solids may include disease causing organisms.
4. The suspended solids parameter is used to measure the quality of wastewater Influent
and effluent and is useful in the analysis of polluted water.
Calculations:
(W2-W1) x 1000
Total suspended solids (mg/l) = -----------------------
ml of sample taken
(W2-W3) x 1000
Suspended volatile solids (mg/l) = ------------------------
ml of sample taken
Alternative:
Total suspended solids, mg/l = [Total solids, mg/l - Total dissolved solids] mg/l
Aim:
To find out total settleable solids in the given water sample.
Principle:
The particles in suspension whose specific gravity greater than that of water will settle under
quiescent conditions.
Apparatus:
1. Imhoff cone
2. Holding device
Procedure:
1. Gently fill the imhoff cone with the thoroughly well mixed sample usually one liter and
allow it to settle.
2. After 45 minutes, gently rotate the cone between hands to ensure that all solids
adhering to the sides are loosened.
3. Allow the solids to settle for 15 minutes more, to make up for a total period of 1 hour.
4. Read the volume of the sludge which ahs settled in the apex.
5. Express the results in ml settleable solids per liter of sample per hour (ml/l/hour).
Significance:
ml of solids x 1000
Total settleable solids (ml/l) = ---------------------------
ml of sample
Precautions:
1. The imhoff cones must be cleaned with a strong soap and hot water using a brush.
2. Wetting the cone with water before use, helps in preventing adherence of the solids to
the sides.
3. The method is subjected to considerable in accuracy if the solids contain large
fragments.
4. The determination of total settleable solids should be carried out soon after sampling in
order to avoid errors through flocculation.
AIM:
To find the optimum amount of coagulant required to treat the turbid waters.
Principle:
Metal salts hydrolyse in presence of the natural alkalinity to form metal hydroxides. The
divalent cat ions can reduce the zeta potential, while the metal hydroxides are good absorbents
and hence remove the suspended particles by enmeshing them.
Apparatus:
Reagents:
Procedure:
Aim:
To find out the amount of residual chlorine available in the given water sample.
Principle:
Apparatus:
Chloroscope
Reagents required:
Orthotolidone reagent
Procedure:
1. Take the water sample under question into one of the cylinders of the comparator and
distilled water into the other.
2. Add 5 drops of orthotolidine solution to both the cylinders and put them in the
comparator.
3. The colour which matches in both the cylinders directly gives the ‘residual chlorine’.
Significance:
1. Active chlorine should be present in drinking water within the range 0.1 to 0.2 mg/I.
However, excessive chlorine content may give out bad odour and may change even
taste of waters. Further, chlorine is said to be carcinogenic. Hence, except during
epidemics- super- chlorination is not to be done.
2. Determination of chlorine residuals is used universally in disinfection practice to control
addition of chlorine and ensure effective disinfection.
3. Chlorine in excess of 0.2 mg/ I is allergic to many humans, irritates eyes, nose and throat,
may damage skin, bleach hair and clothing and corrode pipes.
Aim: To determine the optimum dose of bleaching powder and chlorine by Break Point
chlorination.
Apparatus:
1. Six beakers of I litre capacity
2. Stirrer
3. Chloroscope
Principle:
Chlorine combines with water to form hypochlorous and hydrochloric acid. Hypochlorous acid
dissociates to give the OCI‾ ion. Quantities of OCI‾ and HOCl depend on pH of the solution.
Hypochlorites also give the OCI‾ ions, HOCl rupture the cell membrane of microbes (disease
producing organism). These also react with the impurities like ammonia, oxidisable inorganic
matter likes ferrous ions, nitrites etc. to form chloramines and stable ions of the latter
respectively.
Procedure:
Significance:
1. The residual chlorine of 0.1 to 0.2 mg/l is maintained in water distribution main in order
to prevent further contamination and supply the wholesome water to the consumer.
2. This test is useful to assess the quantity of bleaching powder.
Aim:
Principle:
The principle involved in the determination of DO is to bring about the oxidation of potassium
iodide to iodine with the dissolved oxygen present in the water sample after adding MnSO4,
KOH & KI, and the basic manganic oxide formed acts as an oxygen carrier to enable the
dissolved oxygen to take part in the reaction.
The liberated iodine is titrated against standard sodium thio-sulfate (hypo) solution using starch
as indicator. The blue color disappears to give indication of dissolve oxygen present originally.
Theory:
Determination of Dissolved Oxygen (DO) is important for drinking water. It is the indication of
purity of water. DO is needed for living organisms to maintain their biological processes. If DO is
less than the required limit, (6 to 7 mg/l) it is the indication of pollution due to sewage or
industrial waste. DO test is used to control the amount of oxygen in boiler feed water to
prevent corrosion. DO test helps to assess raw water quality and to keep a check on stream
pollution.
Oxygen is poorly soluble in water. The solubility of DO decreases with increase in concentration
of salt at 1 atm pressure. The solubility of oxygen of air in fresh water with low solid
concentration varies from 14.5 mg/l at 0°C to about 7.5 mg/l at 30°C. Iodometric (Winkler’s
method) are used for determining DO in water.
Apparatus
Procedure:
1. Take a glass stopered bottle of 300 ml capacity and fill it completely with water sample.
2. Add 0.9 ml conc. H2SO4 and 0.2 ml (4drops) KMnO4 solution with the help of a pipette.
The tip of pipette should dip below the liquid surface. Stopper the bottle and mix the
contents of the bottle by inverting it a few times. If the permanganate color is not
discharged within 5 minutes, add additional amount of KMnO4.
3. Add 0.5 ml of potassium oxalate solution, stopper and mix well. Add additional amount
of oxalate solution if the permanganate color is not discharged in 10 minutes.
4. Now add 2 ml of MnSO4 solution followed by 3 ml of alkaline KI solution. Stopper and
shake and allow the precipitate to settle.
5. Now add 1 ml of conc. H2SO4 solution and mix until the precipitate is completely
dissolved.
6. Measure 102.2 ml of this solution with a measuring cylinder in to a conical flask and
titrate slowly against N/50 hypo solution.
7. When the color of the solution is very light yellowish add 2 ml of freshly prepared starch
solution and continue the titration to the disappearance of the blue color and note
down the volume of hypo used.
8. Repeat the titration to get at least three readings and note down the volume of hypo
used.
Significance:
1. It is necessary to know DO levels to assess quality of raw water and to keep a check on
stream pollution. If DO is less than 4 mg/l. aquatic ecosystems would be badly affected.
2. DO test is necessary for all aerobic biological wastewater treatment processes to control
the rate of aeration.
Calculations:
6.6 ml of the reagents have been added under such conditions that approximately equal
volume of the sample is displaced. This dilutes the sample and so a correction factor is
needed.
N 1 V1 = N2 V2
(Oxygen solution) (Hypo solution)
N1 × 100 = 1/50 × V
N1 (normality of the sample with respect to DO) = (1/100) × (V/50) = 8V/5000 g/l =
1.6 × V mg/l
Volume of original sample that will be equivalent to 102.2 ml of the prepared solution is
given by:
Aim:
Principle:
The BOD is an empirical biological test. This BOD test may be considered as wet oxidation
procedure in which the living organisms serve as the medium for oxidation of the organic
matter to carbon-dioxide and water.
a b 3 Bacteria a 3
CnHaObNc + n+ --- - --- - ---- cO2 nCO2 + --- - --- c
4 2 4 2 2
X H2O + cNH3
On the basis of the above relationship, it is possible to interpret BOD data in terms of organic
matter as well as the amount of oxygen used during its oxidation.
Interference:
Reagents:
1. Distilled water
Significance:
1. BOD test gives an idea of the biodegradability of any sample and strength of the waste.
2. Drinking water usually has a BOD of less than 1 mg/l and water is considered fairly upto
3 mg/l of BOD. When BOD value reaches 5 mg/l, the water is doubtful in purity.
Sl Vol. of Dilution Initial Final D.O. Initial D.O. Final D.O. 5 days BOD
No sample ratio D.O. of of sample of blank of blank at 200C
sample mg/l mg/l mg/l (mg/l)
mg/l
1.
2.
3.
4.
5.
Aim:
To find out Chemical Oxygen Demand (COD) of given wastewater sample.
Principle:
The organic matter present in sample gets oxidized completely by K 2Cr2O7 in the presence of
H2SO4 to produce CO2 and H2O. The excess K2Cr2O7 remaining after the reaction is titrated with
Fe (NH4)2 (SO4)2. The dichromate consumed gives the O2 required to oxidation of the organic
matter.
Apparatus:
1. Reflux apparatus
2. Hot plate/ heating mantle
3. Burette
Reagents:
1. Standard potassium dichromate 0.25 N
2. Sulphuric acid with reagent (Conc. H2SO4 + Ag2SO4)
3. Standard ferrous ammonium sulphate 0.1 N
4. Ferroin indicator
5. Mercuric sulphate
Procedure:
1. Place 0.4 gm of HgSO4 in the reflux flask.
2. Add 20 ml of sample (or an aliquot diluted to 20 ml).
3. 10 ml of more concentrated dichromate solution are placed into flask together with
glass beeds.
4. Add slowly 30 ml of H2SO4 containing Ag2SO4 and mix thoroughly.
5. Connect the flask to condenser. Mix the contents thoroughly before heating. Improper
mixing results in bumping and the sample may be blown out.
6. Reflux for a minimum period of 2 hours. Cool and wash down the condenser with
distilled water.
7. Dilute the sample to make up 150 ml and cool.
8. Titrate excess K2Cr2O7 with 0.1 N Fe(NH4)2 SO4 using ferroin indicator, Sharp colour
change from blue green to wine red indicates the end point.
9. Reflux the blank in the same manner using distilled water instead of sample.
1. COD test is widely used in the place of BOD in the operation of treatment facilities
because of the speed with which the results can be obtained.
2. The ratio of BOD to COD is useful to assess the amenability of waste for biological
treatment. Ratio of BOD to COD greater than or equal to 0.8 indicates that wastewaters
are highly amenable to the biological treatment.
Calculations:
(Aeration apparatus)
Aim:
To measure the rate of gas liquid mass transfer characteristics in a diffused bubble system and
to examine variables that influences the rate of mass transfer in Aeration set-up.
Introduction:
Mass transfer between gas and liquid phases is commonly encountered in the environment
sector. In aerobic metabolism, oxygen acts as an electron acceptor for the catabolism. Since the
activated sludge process is designed to be substrate limiting, metabolism sets the rate of
oxygen demand. Dissolved oxygen concentration is one of the most important control variables
in any aerobic reactor. The purpose of the aeration system in an aerobic system is to provide
and transfer oxygen to the liquid at such a rate that it never becomes the limiting factor.
Theory:
Since oxygen has a low solubility in aqueous solution, and biological systems often has a high
demand for oxygen, an understanding of how oxygen is transferred to the liquid phase hence to
cells in reactor is crucial for successful design and analysis. Maintaining the design
concentration while organisms are consuming oxygen requires that oxygen be transferred
continuously into the reaction liquid.
Where,
kL = Oxygen Transfer Co-efficient (cm/h)
a = Gas Liquid interfacial area (cm2/cm3)
kL a = Volumetric Oxygen Transfer Co-efficient
C* = Saturated Dissolved Oxygen concentration
C = Actual dissolved oxygen concentration in the broth
OTR = Oxygen transfer rate (mg O2/L-h)
Where,
C = dissolved oxygen concentration (mmol O2/L)
In case of static method where no bio mass is present, oxygen uptake rate is zero.
The saturation oxygen concentration in water (C*) varies with temperature. Values for C* are
given in the table bellow:-
Static method:
This experiment will focus on determining kLa by means of the static method. The static method
requires that no yeast be present in the broth. This drives our OUR (oxygen uptake rate) to zero
so that any measured changes in C are due to OTR (oxygen transfer rate).
Here the difference between the saturation dissolved oxygen concentration and the actual
concentration is the driving force behind any transfer of oxygen into solution.
Description:
The setup consists of an open water tank with variable speed stirrer. Air is supplied from air
pump provided to bottom of vessel. Flow of air is directly metered with the help of rotameter
and can be regulated by means of valve provided with rotameter. A heater with controller is
provided to maintain the desired temperature in the vessel. Three air diffusers of different type
are provided with the set up and are interchangeable.
1) Vessel
Material : Acrylic
OD : 300 mm
ID : 284 mm
Height : 380 mm
Data:
Aim:
To study the continuous softening or demineralization of water using two bed Ion-Exchange
Unit.
Introduction:
The most usual ion exchange material employed in water softening is a sulphonated styrene
based resin, supplied by the makers in the sodium form. This resin has a strong affinity for
calcium and magnesium ions, and will also remove ferrous ions after the more or less complete
removal of calcium and magnesium.
Softening can be carried out as a batch process by stirring a suspension of the resin in the
water for a period until equilibrium, or an acceptable level of hardness is reached.
Theory:
It is more convenient to operate a continuous flow process by passing the water slowly
downwards through a column of resin beads. The exchange reaction takes place rapidly enough
for the upper layers of the bed to approach exhaustion before the lower layers of the bed are
able to exchange ions. There is thus a zone of active exchange which moves down the column
until the resin at all depths becomes exhausted. The position at an intermediate stage can be
illustrated as below.
When the zone of active exchange reaches the bottom of the column, the emerging water
begins to show an increasing hardness. This is the breakthrough point, when it becomes
necessary to regenerate the resin with a strong sodium chloride solution.
A low cost, bench-mounted unit designed to demonstrate the use of ion-exchange resins for
either continuous water softening of demineralization. The equipment is designed to emulate
the industrial operation of such units, including monitoring ’break-through’ and regeneration
cycles.
Description:
The two vertical transparent tubes mounted on the backboard contain the respective cation
and anion resin. A manifold arrangement at the inlet and outlet to the tubes allows the flow
configuration to be changed to simulate the cycles involved in operation of a deionizer. Union
couplings permit the tubes to be removed from the manifolds and interchanged for
softening/demineralization experiments. Regenerant and test or wash solutions are contained
in independent sumps and delivered to the apparatus by pump. Effluent may be fed to a sump
tank and treated water collected in bottles for tests on hardness, conductivity or dissolved
Specifications:
Utilities Required:
Experimental Procedure:
Aim:
Apparatus:
1. Auto clave
2. Hot air oven
3. Incubator
4. Durham Vial
5. Fermentation tubes
6. Inoculation loops (3 mm diameter of platinum wire) and
7. All other material used in standard plate count test.
Reagents:
Procedure:
Shake the water sample thoroughly before making dilutions or before inoculation.
The completed test though not always necessary, is sometimes done to demonstrate with
certainly that organism showing positive results for the confirmed test are really members of
the coli form group.
1. Prepare endo or Eosin methylene blue (EMB) agar Petri plates. The number of the Petri
plates to be prepared is the same as that of tubes showing gas production in BGLB
medium.
2. Streak inoculums from the BGLB tubes on these plates in such a way that the colonies
after separation have a distance of 0.5 cm.
3. Incubate these plates at 370C ± 0.2 for 24 ± 2 hours.
4. Examine the plates for bacterial growth as colonies. Well isolated colonies with a dark
center (nucleated) are typical coli form colonies. They may have a metallic surface
sheen. The colonies that are pink or opaque and or not of coli form and are reported as
negative in the completed test.
5. Now inoculate as isolated coliform colony (avoid picking up a mixture) from each plate
into the tubes of Lauryl tryptose broth and record the gas production within 48 hours at
370C
Significance:
1. M.P.N. is the most probable number of coliform bacteria in 100ml of water. No water
sample should contain more than 10 MPN and the average of several values of MPN
should not be more than 1. It is a measure of the bacterial density of water and is a
measure of the population of pathogens in water. If MPN is more, the pathogens may
cause several water borne diseases such as typhoid, cholera, dysentery, fevers,
infectious hepatitis(jaundice), poliomyelitis, amoebeasis and gastro-entireties. Improper
disposal of domestic and municipal wastes pollutes water sources with MPN. MPN can
be removed by disinfection in water treatment plants. Strict precautions against back
syphonage and cross connections are required if amoebic cysts are not to be found in a
distribution system.
2. MPN determination is useful in the selection of water supplies for human needs.
3. It is useful to assess the degree of contamination of receiving waters by domestic
sewage.
4. It is useful to find efficiency of disinfectant and treatment units for bacterial removal.
Record the number of positives findings of coli form group organism (presumptive, confirmed
and completed) resulting from multiple portion decimal dilution planting as the combination of
positives and compute in terms of the MPN.
Table: MPN Index For Various Combinations Of Positive Results In Fermentation Tube Tests
Aim:
To determine the suspended particulate matter (SPM), Respirable particulate matter (RPM) in
ambient air.
Glass micro fiber filters are particularly suitable for determination of particulates because of
their high retention efficiency combined with low pressure drop, high resistance to blocking
and low affinity for moisture. Three types of filters-GF/A. EPM-1000, and EPM -2000 are
available for Respirable Dust Sampler. The what- man GF/A glass fiber filter has been in use for
more than 20 years and can be used safely where trace element analysis is not required. What
man EPM-1000 or EPM-2000 spectra quality glass fiber filters containing very low levels of
contaminants are recommended for use where additional chemical analysis is anticipated.
Each filter should be visually inspected using a light table. Remove loose fibers with a
soft brush. Discard or return to the supplier the filters with pinholes and other defects such as
tears, creases or lumps.
Identification:
Assign a serial number to each filter. Stamp this number on two diagonally opposite corners –
one stamp on each side of the filter. Apply gentle pressure to avoid damage to the filter. EPM -
2000 comes numbered directly from the manufacturer.
Equilibration:
Equilibrate the filter in the conditioning environment for 24 hours before weighing it to
minimize errors in the weight; longer periods of equilibration will not affect accuracy. The
conditioning environment should average between 20 and 250C and not than ±30C with a
relative humidity (RH) less than 50 % without varying more than ±5 % RH. A convenient working
RH is 40 %.
Weighing:
Clean filters are usually processed in lots-that is, served at one time. Clean filters must not be
folded or creased prior to weighing or use. Before weighing the first filter, check the balance by
weighing a standard class-S weight of between 3 and 5 g. Record the actual and measured
weights the date and the operators initials.
Sampling Procedure:
Sampling is usually done at 1.5 m height. Ensure that the filter is parallel to the ground. To
obtain a representative sample, the sampler should not be placed under a tree near a wall or
other obstruction that would prevent free air flow from the ambient atmosphere.
Filter Installation:
2. Center the filter with the left side up on the wire screen so that the packet forms an
airtight seal on the outer edge (1 cm) of the tilter when the faceplate is in position when
aligned correctly, the edges of the tilter will be parables both to the edges of the Screen
behind end to the faceplate gasket above it. Poorly aligned alters show uneven white
borders around the filter.
3. Tighten the four knobs just enough to prevent leakage when the filter is aligned and the
dome cover is in place. Excessive tightening may cause the filter to stick or permanently
damage the gasket.
Operational Checks:
Before connecting the instruments to the mains, see that all switches of instrument should be
in %% OFF position ‘‘. Now connect the instrument to the mains. Switch on the power with the
help of voltage stabilizer switch. The stabilizer voltage Output Should be in the range 230 a:15
VAC. Set the timer by removing the red needle to full scale i-e 24 hours and take back the
desired sampling time. Move the black needle to the extreme and clockwise or set to desired
sampling time. Rotate the flow controller knob. Now the blower will be on and it will draw ' the
sample through the filter and impinger tubes. Make flow rate measurement after a warm - up
time of at least 5 minutes. '
Refer to the Calibration Graph and adjust the speed of the Blower motor to obtain the ∆P on
the manometer corresponding to the required flow rate. Let this flow-rate be Qi. ∆P is addition
of both side deviations in manometer in 'mm. However the actual flow through the alter paper
is expected to be reduced because of deposition of dust. Note down ∆P manometer again, just
before the end of the sampling and read corresponding flow rate from the calibration graph.
Let this reduced final flow rate be Qf. After the sampling is over, note down the ' sampling time
as indicated on the time totalizes.
3. Check the fitter for signs of air leakage. Leakage may result from a worn faceplate
gasket or from an improperly installed gasket. If leakage signs are observed, void the
sample, determine the cause, and take corrosive actions before starting another
sampling period. Generally a gasket deteriorate slowly and thus the operator can decide
well in advance (by the increased fuzziness of the sample outline) to change the gasket
before a total failure occurs.
4. Inspect visually the gasket faces to see if glass fibers from the alter are being left behind
due to over tightening of the faceplate knob stud and consequent cutting of the tilter
along the gasket interface.
5. Check the exposed filter for physical damage that may have aired during or after
sampling. physical damage after sampling would not invalidate the samples if all pieces
of the tilter were put in the folder; however, sample losses due to leakages during the
sampling period or bosses of loose participates after sampling (e.g. loss when folding
the filter ) would invalidate the tilter, so mark much samples void before forwarding
them to tie laboratory.
6. Check the appearance of the participates. Any changes from normal colour for example
may indicate new emission sources or construction activity in the area. Note any change
on the field data forms along with any obvious reason for the change. Place the folder
filter into a plastic bag for storage/shipping.
Analysis of Samples:
Gravimetric Analysis:
1. Weigh the exposed filters to the nearest milligram (mg) on the analytical balance within
30 sec removing them from the equilibration chamber.
2. Weigh the filter in the conditioning environment if practical, if not, be Sure that the
analytical balance is as close as possible to the conditioning chamber where it is
relatively free of air currents and where it is at or near the temperature of the chamber.
3. Record the weight in the laboratory data log and on the sampler field data form.
4. For the NRPM, the weighing papers with non - respirable fraction are weighed in the
same manner as the exposed filters.
V = ( Qi + Qf ) / 2) T------------------------------------ Eq. ( 1)
Where
SPM Calculation:
RPM Calculation:
1. Calculate the RPM concentration by using the Eq (2) as above, except substituting:
“RPM (µg/m3)'' in place of SPM (µg/m3)-
2. Calculate the (total) SPM Concentration by using Eq (2) again, except substituting:
Where
Wp = Weight of material that was collected on the pan including: the 'weighing paper, gm.
Wj = Initial weight of the weighing paper, gm.
Record all original calculation in the data log book.
Take the absorbing solution (as mentioned in the specific procedures) in each or the impinger
tubes. 'To adjust the individual flow rates through the impinger tubes connect inlet of one of
the impinger tubes to the Rota meter and outlet to (the of the four input connections of the
manifold. Adjust the required flow rate with the help of the respective restricted and then
lighten the screw to keep the flow rate constant. Let this initial flow rate be F1.
Follow the same procedure to adjust flows through the other impinger tubes. Measure the
individual flow rates again with the Rota meter just before the end of the experiment. Let that
flow rate be F2. Acer the sampling is over, note down the sampling time as indicated on the
localizer.
Where
F1 = Initial flow rate at start of sampling. lpm
F2' = Final flow rate at end of sampling. lpm
T = Sampling time period, min.
Aim:
Determination of Nitrogen Dioxide (NO2) in Ambient Air the suspended particulate matter in
ambient air
Nitrogen dioxide (NO2) is collected by bubbling air through a sodium hydroxide - sodium
arsenate solution to form a stable solution of sodium nitrite (NaNO2). The nitrate ion produced
during sampling is reacted it Phosphoric acid, aulphanilamide and N-1 (naphthyl)
ethylenediamine di-ydrochloride, to form an azo dye, and then determined colourimetrically.
The method is applicable to collection of 4 to 24 hours samples in the field and Subsequent
analysis in the laboratory.
The range of the analysis is 0.01 to 1.5 μg No2 / ml. A concentration 0.04 μg NO2/ml will
produce absorbency approximately 0.02 with 1 cm cells, and Beer's law is obeyed throughout
this range (0 to 0.1 absorbency unites). With 25 ml absorbing reagent and a sampling rate of 0.2
Lpm for 24 hours. The range of the method Is 3.5 to 180μg/m, (0.002 to 0.100 ppmv ) of
nitrogen dioxide.
Interference:
Nitric oxide (NO) Is a positive interferers. The presence NO can increase the NO2 response by 5
to 15 % of the NO2 sampled. The interference of sulphur dioxide (SO2) is eliminated by
converting it to sulphate ion with hydrogen peroxide before analysis.
The relative standard deviations for sampling NO2 concentrations of 78, 105 and 328 gamma
are 3, 4 and zo/orespedively. Collected samples are stable for at least 6 weeks.
Apparatus:
G = 100 x (1.500 / A)
Where
G =% amount of NaNO2. grams
1.500 = Gravimetric factor in converting No2 into NaNo2
A = Assay Percent
Analysis:
Replace any water lost by evaporation during sampling by adding distilled water tip 'to' the
calibration mark on the absorption tube. Pipette In . 1 ml of hydrogen peroxide solution, 10 ml
of Sulphanilamlde solution and 1..4 my NEO'A solution with thorough Su p mixing after the
addition of each reagent. Prepare a blank in the same manner using IQ my of unexposed
absolution ' reagent after IQ minute concur - development interval, measure the absorbable at
540 nm against the blank . Read μg No2 /mI from the calibration curve. Samples with an
absonance greater than 1.0 must be reanalyzed after ' diluting an aliquot (less than 10 ml) of
the colleted sample with unexposed absorbing reagent.
Calculate the concentration of nitrogen dioxide (NO2), using the following formula:
0.82 = empirical factor ' fir collection calculated by using Eq (3) as given efficiency
Calculate the concentration of nitrogen dioxide In units of ppm vol. by the formula: (NO 2) (
ppmv) = (NO2) (μg /m3) x 0.000532
Aim:
1.1.1 Sulfur dioxide, in an air sample, is absorbed into a solution of potassium or sodium
tetrachloromercuratell-tEzM) by aspirating ' a ' measured air sample through unabsorbed
vessel. This procedure results in the formation of monochlorosulfonatomerarate (11) complex
(3), which resists oxidation by the oxygen in the air ( 1,2,3), Ethylenediamine-tetraacetic acid di-
sodium salt (EDTA) Is added to this solution to complex heavy metals that catalyze the oxidation
of the colleted Sulfur dioxide (4.5). Once this monochlorosulfonatomercurate complex is
formed, it is stable to strong oxidants (egg ozone, oxides of nitrogen and hydrogen peroxide)
(2). After sampling is completed, any ozone in the solution is allowed to decay (5). The liquid is
treated first with a solution of sulfuric acid to destroy the nitrite anion formed from the
absorption of oxides of nitrogen present in the atmosphere (6). It is then treated with solution
of formaldehyde and specially purified, acid- bleached pararosanline containing phosphoric acid
to control pH. Pararosanillne, formaldehyde and the biosulfite anion read to form the $
Intensely coloured Pararosanllinemethylsulfonic acid, which behaves as a two-color PH
indicators amid = 548 nm at ad 1.6 +.0.1, c = 47.7 x ' a 10 ). The E value assumes quantitative
production of the absorbing species. Re pH of the final solution is adjusted to 1.6 ± 0.1 by the
addition of a prescribed amount of 3M phosphoric acid to the pararosanlline reagent (5). This
technique is the basis of the reference method for the determination of sulfur dioxide by the
U.S. Environmental Protection Agency (7).
1.1.2 Two variations are given, differing only in the pH of the final solution. The variation
described above is designated Variation A and is the method of choice. It elves the highest
sensitivity. In Variation B, a larger quantity of phosphoric acid is added to yield a pH of 1.2±0.1
in the final solution. The wavelength of maximum absorbable under these conditions is 575 nm,
and the compound has a molar extinction of 37.0 x 103, assuming quantitative production of
absorbing Species.
Variation B is less sensitive, but has the advantage of a tower blank. It is PH dependent, and
may be more suitable with less expensive spectrophotometers.
2.1 Atmospheric sulfur dioxide concentrations are measurable by this technique in the range
from 10 ppbv to a few ppmv. Use smaller gas samples when analyzing higher concentrations (S
to 500 ppmv), found in special studies. A rapid redox reaction occurs between Hg (11) and the
sulfite ion if concentrations of th latter exceed 500 μg/ml (8). Collection efficiency falls off
2.3 Beer's Law is followed through the working range from 0.1 to 1.0 absorbance units (0 to 35
μg in 25 ML final solution).
3.0 Inferences:
3.1 The effects of the principal potential interferences (oxides of nitrogen, ozone and transition
metals (e.g. iron, manganese, and chromium) have been minimized or eliminated. The
interferences by oxides of nitrogen are eliminated by sulfuric acid (5,6), the ozone by time delay
(4), and the transition metals by EOTA and phosphoric acid (4,5). At least 60 pg of V (V) in 1.0 ml
of absorbing reagent cause no interference.
5.0 Apparatus:
5.1 ABSORBER - Satisfactory absorbers are (a) the midget or standard fritted bubbterthe midget
impinger; (c) the Greenberg-smith impinger; and (d) the multiple jet bubbler (12).
5.2 AIR VOLUME MEASUREMENT- Re sample air volume must be measured within + 2%. See
Section 5 of Part I for details of air volume measurement. A critical orifice is recommended (13).
6.0 Reagents:
6.1 PURITY OF CHEMICALS - All chemicals must be ACS analytical reagent grade. The
pararosaniline dye should meet the specifications described in 6.9.
6.2 WATER - Water must conform to the ASTM Standard for Reference ' Reagent Water Water.
Type 11. This can be produced either by distillation or a cartidged deionization system.
(TCM), K2HgCI2- Dissolve 10.86 j mercuric chloride (CAUTION: Highly poisonous and corrosive.
If spilled on skin, flush off with water immediately): 5.96 g of potassium chloride and 0.066 g of
6.4 SULFAMIC ACID (0.6%)- Dissolve 0.6 g of sulfuric acid in 100 my Of water. This reagent can
be kept for 10 days if it is stored in a stoppered bottle.
6.5 BUFFER STOCK SOLUIRON (PH 4.7). In a 100 IML volumetric black, dissolve 13.61 g of
sodium acetate tri-hydrate in water. Add 5.7 ml of glacial acetic acid and dilute to volume with
water.
6.7 3M PHOSPHORIC ACID (H3PO4) to IL. Dilute 83 ml of 12.1M acid (36% HCl) to 1L.
6.8 BUTANOL. 1-Butanol is required for the purification Of the pararosaniline dye. The butanes
should be checked for oxidants that can consume SO:. Check it by shaking 20 ml of 1-butanol
with 5 IML of 20% aqueous Kl. A yellow color in the alcohol phase indicates tine presence of
oxidant and the butanes must be redistilled from silver oxide.
6.9 PURIFIED PARAROSANILINE 0.2 % (NOMINAL) STOCK SOLUTION Pararosaniline dye needed
to ' ' prepare this reagent must meet the Following performance specifications. The dye must
have a wavelength of maximum absorbency of 540 nm when essayed in 0.1M sodium acetate-
acetic acid buffer the assonance of the reagent blank (section 7.3), which is temperature
sensitive (0.o1sA.wc), should not exceed 0.170 absorbency unit at 22*c; and the reagent should
give a calibration curve with a slope of 0.746 z 0.04 assonance units (μg SO2, mL) for a 1-cm
cell. Prepare a solution of the dye; weigh 0.200 g and completely dissolve by shaking with 100
my of 1N HCl In a 100 ml glass stoppered graduated cylinder. If the pararosaniline dye is
obtained in purified form as ozone solution, proceed to 6.9.5. Specially purified dye, 0.2% in 1N
HCI solution, is available from Eastman Kodam, J.T. Baker or Harleco.
6 .9.1 Dye Purification Procedure. The pararosaniline dye (PRA) is purified by extraction of
impurities into 1-butanol. In a 250 ml separating funnel shake 100 ml each of 1-butanol and 1 N
HCI and allow the layers to separate. Collect each layer separately. Add 0.1 g of pararosaniline
to 50 ml of the butanol-saturated acid and allow it to stand for several minutes. Add this acid
pararosaniline solution to 50 ml of acid-saturated 1-butanol in a 12S my separation funnel. The
impurities violet coloured, will be transferred to the phase. Save the lower (aqueous) phase and
extract again with 20 ML of 1-butanol. Repeat this procedure three times using 10 ml portions
of 1-butanol. If a violet color still appears in the l-butanol after 5 extractions, filter the aqueous
phase through a glass wool plug into a 50 ml volumetric flask and bring to volume IN HCI. The
final solution is nominally 0.2% pararosaniline in 1N HCI saturated with l-butanol.
6.9.2 1.0 M Sodium Acetate - Acetic Acid Buffer. Dissolve 57.3 my glacial acetic acid and 136 g
sodium acetate dehydrate in water and make up to the mark in a 1-L volumetric flask. .
6.9.3 Assay Procedure. The concentration of PRA need be assayed only once for each lot of dye
in the following manner: Dilute 1 my of the stock reagent (0.2%) to the mark in a 100-mL
volumetric flask with distilled water. Transfer a 5-mL aliquot to a 50 my volumetric flask. Add 5
6. I 1.1 Stock iodine Solution, 0.1 N. Place 12.7 g of iodine in a 250 ml beaker, add 40 gm. of
potassium iodide and 25 my of water. Stir until all the solid is dissolved, then dilute to 1 L with
water. If desired, prestandardized solutions of iodine can be purchased from laboratory supply
dealers. (a) Working Iodine Solution, 0.01N: Prepare approximately 0.01N iodine solution by
diluting 50 ml of the stock solution to 500 ml with distilled water.
6.11.1 Starch Indicator Solution. Triturate 0.4 g of soluble starch and 0.002 g of mercuric iodide
(preservative) with a little water and add t the paste slowly to 220 ml of boiling water. Continue
boiling until clear; cool and transfer to a glass stoppered bottle. If the Indicator solution is only
going to be kept for few days, omit the mercuric iodide.
(X - Y) NZ
SO2 (μg/mI) = ----------------
S
Where:
7.0 Procedure:
7.1.1 COLLECTION OF SAMPLE. Place 10 my of 0.04M TCM (20mL for sampling of long duration)
absorbing solution in a fidget impinges or 75 to 100 mL in a larger absorber. The sampling
probe-to-absorber linkage will depend on the installation. This should be kept as short and
direct as possible using glass or Teflon tubing. Care must be taken to avoid condensation in the
sample Inlet, which can occur when warm humid air is brought In to an air conditioned location.
After the absorber, a trap, such as a flask tightly packed with glass wont, should be placed to
prevent corrosion damage to downstream components from aerosolized TCM droplets, which
are highly corrosive. Provision must be made for the flow volume measurement farther
downstream of the absorber. The duration of sampling will depend on the concentration of
SO2. With midget impingers, rates of 0.5 to 2.5 L/min are satisfactory; with large absorbers the
rate can be 5 to 15 L/min.
Rates of sampling within the above ranges generally have absorption efficiency of 98% or
greater. For best result, rates and sampling tinge should be chosen to absorb 0.5 to 30 μg/ml-
of SO2. Shield tile absorber from direct sunlight by covering with a suitable wrapping. If the
sample must be stored for more than a day before analysis, keep it at S C, if possible.
7.1.3 ANALYSIS. Acer collection, transfer the sample quantitatively to a 25ml volumetric flask;
use about 5 ill of water for rinsing. Liquors may be taken, at this point, if the concentration or
7.1.4 If the assonance of the sample solution ranges between 1.0 and 2.0, the sample can be
diluted 1: l with a million of the reagent blank and read within a few minutes. Solutions with
higher absorbency retire the diluted up to six-fold with the reagent blank in order to obtain of')
artists readings within 10% of the true assonance value.
8.2 The total observances of the solutions are plotted (as ordinates) against the total
micrograms of SO2. A linear relationship is obtained. The intercept with the vertical axis of the
line best little: the points usually is within 0.02 absorbency units of ''the blank (zero standard)
reading. Under these conditions the plot need be determined only once to evaluate the
calibration factor (reciprocal of the slope of the line). More accurate values of slope and
interceipt can be obtained by linear regression. This calibration factor can be used for
calculating results provided there are no radical changes in temperature of pH. At least one
control sample is recommended per series of determinations to ensure the reliability of this
factor.
Permeation tubes containing liquefied sulfur dioxide are calibrated gravimetrically and .used to
prepare standard concentrations of sulfur dioxide in air (16-18). Sampling of these known
concentrations and subsequent analysis of collected samples give calibration curves that
simulate all of the operational conditions during the sampling and analytical procedure. Such
calibration curves include the important correction for collection efficiency at various
concentrations of sulfur dioxide. For details on the use of permeation tubes as calibration
sources see Section 3, Part 1.
Solutions of monochlorosulfonatomtxrcura|e (11) are Sable towards the loss of SO2 when
stored at 5 C for 30 days. At 25 C toeless of SO2 ranges from 1-5% per day (3, 19, 20). These
losses of SO2 follow a first order reaction; the rate constant is independent of concentration.
For storage of samples under conditions where SO2 loss may be important, e.g., without
refrigeration correction factors must be applied. Alternatively, an increase In the Cl
concentration will improve the stability of the monochtorosulfonatomercurate (11) species
(3.19). This modification leads to some difficulties in sampling and a loss of sampling efficiency.
Thermoelectrically cooled enclosures for housing the absorber vessels during or after sampling
are commercially available from RAC, Division of Andersen samplers, Inc. Atlanta, ' Georgia.
Calculations:
Compute the concentrations of sulphur dioxide (SO2), in units of μg/m3 in the sample by using
the following formula:
(so2) (μg/m3) = (A - Ao ) x B
----------------
V
Where
The concentration of SO2 in units of ppm (Volume) is given by (SO2) (ppmv) = (SO2) (μg/m3) x
0.000382,
Where 0.000382 = the volume (mI) of 1 μgSO2 at STP (25 deg C. 101.3 kpa).
Aim:
1. To plot a graph between the height of interface (z) and time (θ) for slurries having five
different concentration of CaCO3.
Introduction:
The suspension of dilute slurry by gravity settling into a clear liquid and slurry of higher solid
concentration is known as sedimentation. The mechanism of sedimentation observed during a
batch settling set is shown in figure.
As sedimentation continues, the height of each zone varies as indicated in the figure. Both A
and D grow larger at the expense of B and ultimately zone B & C disappear and all the solids
appear in a D. This is known as critical settling point. The same zones shall be present in a
continuous thickner.
Batch sedimentation tests are used to determine the settling characteristics of a slurry/sludge.
The analysis is further used for the design of continuous thickner.
Theory:
For any batch sedimentation experiment, on slurry of known concentration, the height of a
liquid-solid interface is obtained as a function of time. Slopes of this curve at any point of time
represent settling velocities of the suspension at that time and are characteristics of specific
solid concentration.
Utilities Required:
2. Stop watch
Experimental Procedure:
Prepare five samples of CaCO3 slurry in each cylinder. Mix the samples thoroughly before
starting the experiment so that homogeneous slurry is obtained in each case. Note the initial
height (z0). Start recording the time and interfacial height in each of the graduated cylinder
separately continue to take the record till height of interface stops any further. Note the final
height (z∞).
Plot all the curves z vs θ. Find the slope of initial settling rates (dz/dθ)θ=0 for each concentration
(C0) and plot on a log-log graph.
Aim:
To study the characteristics of a deep bed filter column using 50 to 150 mg/l kaolin suspension
in water.
Introduction:
Slow sand filters have been used to remove particles from drinking water since early 1800’s.
The particles that are captured on slow sand filters have been shown to significantly improve
filter performance.
Theory:
In slow sand filters with clean filter media, particles are initially removed by attaching to the
filter media. However, as the filter media begins to be covered with removed particles, particles
begin to attach to previously removed particles. If particle-particle interaction is more favorable
than particle-media interaction then particle removal efficiency increases as the media
becomes covered with particles. This improvement in filter performance with time is commonly
observed in slow sand filters and is referred to as filter “ripening”. Filter ripening often takes
several weeks to several months for new slow sand filters.
Filteration theory suggests that particle removal will be first order with respect to depth if filter
media is homogeneous. In equation from the relationship between particle concentration, C,
and depth can be expressed as:
Where, λ is the filter coefficient with units of [1/L]. Using appropriate limits
C0 0
dC/C = -λ dz ……………………………..(2)
C L
Where, L is the depth of the filter bed and C0 is the influent particle concentration and
integrating Eq (2) yields:
Turbidimeters measure the amount of light scatter caused by a suspension of particles. Because
absorption and scattering of light are influenced by both size and surface characteristics of the
suspended material, turbidity is not a direct quantitative measurement of the concentration of
suspended solids. In a turbidimeter the scattered light and the transmitted light is proportional
to the turbidity of the sample. The constant of proportionality is determined by measuring a
known standard.
Experimental Procedure:
1. Before the start of experiment, prepare a calibration curve using standard solutions of
kaolin in water ( 50 mg/L, 20 mg/L, 10 mg/L, 8 mg/L, 6 mg/L, 4 mg/L, 2 mg/L, 1 mg/L)
and measure the absorbance in terms of NTU in a turbid meter and plot the calibration
curve.
2. Prepare kaolin suspension in water in the range of 50 to 150 mg/L in the feed tank. Keep
the solution under constant stirring.
3. If the filter bed has been previously used, then using backwash mode use clean water to
wash the bed. Carefully observe the sand surface as you gradually increase the flow rate
from zero in backwash mode. Continue backwashing the filter to clean the sand until the
effluent turbidity is less than 0.5 NTU.
4. Obtain head loss (in cm) as a function of flow rate of pure clean water over a range of 1
to 25 m/hr using at least 5 data points. Use the rotameter to measure the flow rate..
5. Set the down flow rate to 5 m/hr.
Data:
Filteration Parameters
Parameter Symbol Value Units
Column Diameter D 10 cm
Column Area A 78.5 cm2
Column Length L column 120 cm
Media depth L 135 cm
Bulk Density of Media Bulk density 145 kg/m3
Mass of Media Sandmass 11 kg
Influent Clay Concentration C0 In the range of 50 to Mg/L
300
Data Analysis:
1. Plot head loss vs. flow rate for a clean bed and estimate the hydraulic conductivity of
the sand.
2. Plot head loss at different depths.
3. Plot the fraction of influent particles remaining in the effluent vs. time and vs. depth for
each run on a single graph.
4. Plot head loss as a function of time foe each run on a single graph.
5. Calculate the filter coefficient (Eq (3)) for the filter with and without the filter aid.
3. Mathur, RP “Water and wastewater testing (A laboratory Manual)” New Chand & Bros
Company, 2001.
5. Bailey, JE and Ollis, DF “Bio chemical Engineering Fundamentals” 2nd Ed, Mc Graw –Hill,
1986.
8. Deory et al. “Laboratory scale equipment for the determination of kLa in Bio-reactors –