The radiographic system described in this manual refers to a “wall installation”.

“de Götzen S.r.l.” reserves itself the right to modify the production and the manual without notice on their
designated distribution.

“de Götzen S.r.l.” shall not be liable for the incorrect use of the information contained in this
manual.

Copies, even excerpts, of this manual are allowed only for personal use and with written consent of
the manufacturer.

Edition: 01/2004

Drawn up by A. Colombo Approved by M. de Götzen

Manual Revision Date Description


1.0 01/2004 English Version

2 xgenus ac/dc
SUMMARY

INTRODUCTION page 2
SUMMARY page 3
PRELIMINARY INFORMATION page 4
INFORMATION FOR THE INSTALLER page 5

1. “xgenus ac/dc” RADIOGRAPHIC SYSTEMS page 6


2. PICTOGRAMS USED page 7
3. OVERALL DIMENSIONS page 9
4. INSTALLATION SPECIFICATIONS page 13
5. INSTALLATION page 15
5.1 UNPACKING page 15
5.2 ASSEMBLING THE WALL PLATE page 16
5.3 ASSEMBLING THE BRACKET/HORIZONTAL ARM page 18
5.4 ASSEMBLING THE POSITIONER ARM page 20
5.5 CONNECTION TO THE TERMINAL BOARD page 21
5.6 ASSEMBLING THE RADIOGRAPHIC UNIT page 23
5.7 BALANCING THE POSITIONER ARM page 26
5.8 ASSEMBLING THE “xgenus® ac/dc” X-RAY CONTROL page 29
5.9 X-RAY CONTROL ELECTRICAL CONNECTION page 31
6. X-RAY CONTROL PANEL page 33
7. SYSTEM CONFIGURATION page 35
8. CHANGING THE CONFIGURATION page 38
9. START UP page 40
10. CHECKING THE INSTALLATION page 41
11. CHECKING THE EXPOSURE FACTORS page 43
12. DIAGNOSTIC page 46
13. TUBE HEAD CALIBRATION page 47
14. ERROR MESSAGES page 48
15. REPLACEMENT OF FUSES page 49
16. MAINTENANCE page 50
17. CLEANING THE OUTER SURFACES page 51
18. IF A REPAIR SHOULD BECOME NECESSARY page 51
19. ATTACHMENTS page 52

xgenus ac/dc 3
PRELIMINARY INFORMATION

Before beginning to use the “xgenus ac/dc” radiographic systems, it is advisable to carefully read
and follow the instructions contained herein, so as to obtain the best possible performance.

Always pay close attention to the CAUTION, WARNING, and PLEASE NOTE messages when
operating the system.

LEGEND

CAUTION
The word CAUTION identifies those situations that might compromise the operator's personal
safety or cause injuries to others.

WARNING
The word WARNING identifies those occurrences which might compromise the xgenus® ac/dc
x-ray systems’ performance.

PLEASE NOTE
The words PLEASE NOTE serves to draw your attention to special indications to facilitate
maintenance or make important information more clear.

4 xgenus ac/dc
INFORMATION FOR THE INSTALLER

CAUTION
The installer is responsible for the installation, calibration and testing with regard to
system safety and efficient operation.

For safe and reliable installation of the “xgenus ac/dc” radiographic systems, it is advisable to:

 Check that the voltage mentioned on the rating plate matches the line voltage.

 Install the “xgenus ac/dc” radiographic systems according to the procedures described
in this manual.

 Provide the user with any information regarding the use of the “xgenus ac/dc”
radiographic systems according to that which is provided in the operators manual.

 Certify the work done by completing the “Declaration of Conformity” form.

 Return the completed warranty certificate to: TG Group.


Only when the completed warranty certificate is received is the warranty valid.

xgenus ac/dc 5
1. “xgenus ac/dc” RADIOGRAPHIC SYSTEMS

The “xgenus ac/dc” radiographic systems (Fig. 1) consists of:

 TUBE HEAD – XGG / TUBEHEAD – XGDC


 CONE
 POSITIONER ARM (Scissor)
 CONTROL – XGT
 WALL BOX
 HORIZONTAL ARM

OPTIONAL ACCESSORIES :

− long 12” (31cm) cone


− short 8” (20cm) cone
− long cone 12” (31cm) with a rectangular section sized 44x35mm
- remote control button: xgenus ECB



 Fig. 1






xgenus® dc only

6 xgenus ac/dc
2. PICTOGRAMS USED


The identification tags on the tube heads XGG/XGDC , on the control XGT and on the cone indicate the
model name, the serial number, the manufacturing date and the symbols of the main technical
characteristics.

ID TAG OF THE
“xgenus ac”
TUBE HEAD XGG

ID TAG OF THE
“xgenus dc”
TUBE HEAD XGDC

ID TAG OF THE
“xgenus”
CONTROL XGT

ID TAG OF THE SHORT 8” CONE: BLD20 - Ø 60mm (20 cm SSD)


ID TAG OF THE LONG 12” CONE: BLD31 - Ø 60mm (31 cm SSD)
ID TAG OF THE RECTANGULAR CONE: BLD31SQ - 44x35mm (31 cm SSD)

GRADUATED SCALE TAG ON TUBE HEADS


“xgenus® dc” only

xgenus ac/dc 7
SYMBOLS:

This symbol guarantees that the radiographic system complies with the regulations
contained in the European Directive EEC 93/42 regarding Medical Devices

The degree of protection against direct and indirect electric contacts is B type

Symbol indicating danger due to “ionizing radiation”

Refer to Manual's instructions

N “NEUTRAL" power supply wire

L “LIVE" power supply wire

Ground cable

8 xgenus ac/dc
3. OVERALL DIMENSIONS

Fig. 1A, 1B, 1C are the overall dimensions to accommodate varying installation situations:

 SHORT ARM (optional) length 41cm - 16.2”

 STANDARD ARM length 82,5cm - 32.5”

 LONG ARM (optional) length 110cm - 43.5”

SHORT ARM 41cm (16.2”) Fig. 1A

STANDARD ARM 82.5cm (32.5”) Fig. 1B

LONG ARM 110cm (43.5”) Fig. 1C

xgenus ac/dc 9
Fig. 2, 3A, 3B, 3C e 4 show the typical dimensions of the radiographic system:

STANDARD ARM 82.5cm (32.5”) ONE INCH = 2.54 cm


Fig. 2

10 xgenus ac/dc
STANDARD ARM 82.5cm (32.5”) Fig. 3A

Fig. 3B Fig. 3C
SHORT ARM 41cm (16.2”) LONG ARM 110cm (43.5”)
ONE INCH = 2.54cm

xgenus ac/dc 11
ONE INCH = 2.54 cm Fig. 4

12 xgenus ac/dc
4. INSTALLATION SPECIFICATIONS

WARNING
Prior to installing the radiographic system the Office Owner must be sure that the operatory and
its’ electrical services comply with all governing regulations and required installation specifications.
Otherwise one must bring them all to compliance.

1. OPERATORY REQUIREMENTS

a. The operatory must be of suitable width to accommodate the sizes and overall
dimensions provided (refer to section 3 “OVERALL DIMENSIONS”).
Check that no obstacles are present while positioning the radiographic system.
b. The operatory must not be pressurized or constitute an explosion hazard.
c. While operating, the ambient temperature must be within +5°C and +40°C
d. The storage temperature must be within -15°C and +50°C
e. The relative humidity must be within 25% and 75%

2. REQUIREMENTS OF THE SUPPORTING WALL

a. The supporting wall for the radiographic system must be able to withstand 200Kg sheer
stress at every bolt.
b. The subject wall material and construction must be checked. If required, consult a
contractor to determine strength and stability.
c. If there is any doubt as to the strength or stability of the wall to support the x-ray,
reinforce by providing either a steel or double plywood plate. Refer to the Pre-installation
drawings.

3. REQUIREMENTS OF THE ELECTRIC SYSTEM

a. The electric system must comply with all applicable regulatory bodies.
b. The electric system must be able to supply the power-voltage and current required by
the Manufacturer's rating plate of the radiographic system as per Chart 1

Model  xgenus® ac xgenus® dc

115V 115V
Input voltage of the radiographic systems
-5% +10% ±15%
CHART 1

NOMINAL LINE VOLTAGE 115V 115V


MINIMUM LINE VOLTAGE 109 Vac 98 Vac
MAXIMUM LINE VOLTAGE 126.5 Vac 132 Vac
FREQUENCY 50/60 Hz 50/60 Hz
POWER DRAWN DURING EXPOSURE
800 VA 1.4 kVA
(MAX. TIME = 3.9 SEC)


PLEASE NOTE
the communication cables (C11, C12 – C21, C22) between the timer and the tube head must be two-pole, twisted
and shielded with ≥0,25mm2 (ex.: Belden 9501 type).


xgenus® dc only

xgenus ac/dc 13
4. REQUIREMENTS OF THE ELECTRIC LINE

a. The electric line must be "single-phase" type.


b. Separate circuit of 15 amps to be supplied.
c. Wire size according to AWG.

PLEASE NOTE
For longer lines, the conductor gauge must be increased in proportion.

Model  xgenus® ac xgenus® dc

115V 115V
Line voltage
-5% +10% ±15%
MAXIMUM LINE VOLTAGE DROP 1.2% 3%
MINIMUM LINE APPARENT RESISTANCE 0.2Ω 0.2Ω

5. ELECTRICAL CONNECTIONS

WARNING
Prior to installing the radiographic system, it is advisable that the electrical supply complies with
the recommended specifications in this manual.

X-ray control electrical connections

CAUTION
According to relevant and practical standards, the x-ray control must be installed in a
position allowing the operator to see and have full control of the radiographic
exposure.

On the x-ray control installation wall, suitable runs for the following electrical cables must be
provided, according to the electrical installation diagram (refer to section 18
“ATTACHMENTS”):

a. X-ray control electrical cables


(refer to section 5.9 “CONTROL ELECTRIC CONNECTIONS”)

b. Cables connecting the control and the radiographic unit


(refer to section 5.9 “ CONTROL ELECTRIC CONNECTIONS ”)

Radiographic unit electric connections

To the wall plate location on the installation wall, a suitable run for the cable connecting the
x-ray control and the radiographic unit must be provided
(refer to section 5.5 “CONNECTION TO THE POWER SUPPLY TERMINAL BOARD”)

14 xgenus ac/dc
5. INSTALLATION

CAUTION
The “xgenus ac/dc” radiographic systems must be installed by professionally trained
technicians, who must be able to certify their work.

CAUTION
Prior to proceeding with the installation, check that all requirements are met.

5.1 UNPACKING

The components of the “xgenus ac/dc” radiographic systems are delivered by the factory securely
packaged within a specifically designed box, as shown in the following sketch (Fig. 5):

DOCUMENTS: INSTRUCTIONS MANUAL, INSTALLATION & MAINTENANCE MANUAL, ANNUAL


D MAINTENANCE INSTRUCTIONS AND WARRANTY
 RADIOGRAPHIC UNIT PACKAGING

 X-RAY CONTROL PACKAGING + EXTERNAL LAMP (IF PROVIDED)

 POSITIONER ARM PACKAGING

 HORIZONTAL ARM AND WALL BOX PACKAGING

PLEASE NOTE
Prior to installation, please check all components for shipping damage. Advise the carrier.

PLEASE NOTE
The carton materials and the polystyrene foam can be completely recycled or disposed of by
authorized recycling companies.

PLEASE NOTE
It is advisable to store the original packaging to return the goods for repairs if required.

xgenus ac/dc 15
5.2 ASSEMBLING THE WALL PLATE

ASSEMBLY INSTRUCTIONS (Fig. 6)

1. Take the wall plate out of the packaging (refer to Fig. 5) Use the drilling template provided

2. Position the drilling template  on the radiographic system installation wall, at the required
height (130 cm from the base is the suggested height)
3. Fix the template  to the wall with adhesive tape
4. Check the holes for vertical and horizontal alignment with the floor, using a plumb line
5. Mark the plate fixing holes
6. Mark the holes for the electric cables connecting the x-ray control to the radiographic unit

PLEASE NOTE
To prevent any flaking of the white paint coat and to control the center distances between the
holes, it is advisable to start drilling with a tip Ø7, increasing this measure gradually.

7. Drill the plate fixing holes


8. If required, drill the holes for the electric cables connecting the x-ray control to the
radiographic unit
9. Remove the template  and insert the suitable anchor screws, according to the wall
characteristics 
10. Unscrew the screw  and remove the plug  from the wall plate
11. Withdraw the sliding cover 
12. Attach the plate  to the wall and insert the screws  with the relevant washers, then
tighten alternately
13. Check that the plate is securely fixed to the wall and is plumb and level.

PLEASE NOTE
If the wall is not perfectly leveled, put a suitable shim between the wall and the plate, so as to
prevent any possible warping of the mounting plate.

16 xgenus ac/dc


 





Fig. 6

xgenus ac/dc 17
5.3 ASSEMBLING THE BRACKET/HORIZONTAL ARM

PLEASE NOTE
The brackets come in the following lengths: 41cm (16.2”) - 82.5cm (32.5”) - 110cm (43.5”)
The 82.5 cm and 110 cm brackets are provided with a stop key  (Fig. 7A and 7B) to prevent the
electric cable from twisting.

PLEASE NOTE
Generally, the stop key is installed so that the equipment position at rest is on the right side of a
possible viewer when standing in front of the wall plate (Fig. 7A).
Should the position at rest be on the left side, the stop key must be rotated by 180° (Fig. 7B).

Fig. 7A Fig. 7B

WALL

 

18 xgenus ac/dc
After fixing the plate to the wall, assemble the bracket according to the following procedure (Fig. 8):

1. Take the bracket out of the packaging (refer to Fig. 5)


2. Insert the bracket pin  into the wall plate  (upwards)
3. Insert the supporting rest 

Fig. 8

PLEASE NOTE
Prevent any foreign material (drywall, dust, cement, etc. ) from settling on the pin seat.
The pin must slide freely in its seat. If required, thoroughly clean and lubricate with grease
“Molikote D".

PLEASE NOTE
Carefully check the bracket is parallel to the floor by using a water level.

xgenus ac/dc 19
5.4 ASSEMBLING THE POSITIONER ARM

ASSEMBLY INSTRUCTIONS (Fig. 9)

1. Take out the positioner arm from the packaging (refer to Fig. 5)
2. Remove the arm and cap  by unscrewing the fixing screw 
3. Slide the arm guard slat 
4. Insert the arm cable  through the washer  and then the positioner pin 








Fig. 9

5. If required, clean the pin and the washer and lubricate with grease “Molikote D".
6. Insert the electric cable into the bracket housing 
7. Assemble the guard slat
Insert the cable into the arm and push it until it reaches the pin outlet near the supply
8.
terminal board

20 xgenus ac/dc
5.5 CONNECTION TO THE TERMINAL BOARD

CAUTION
For electrical safety, it is essential that the ground conductors be correctly connected.

WARNING
The electric cable length is suitable for 82.5cm long arms.
DOT NOT tamper with the terminals arrangement when 41cm long arms are used.
The excess cable must be housed in the housing provided.
The 110cm long arms are equipped with a special extension cable.

OPERATING INSTRUCTIONS (Fig. 10)

1. Remove the terminal board cover  by removing the screw 

Fig. 10

xgenus ac/dc 21
2. Proceed to the electric connection as shown in Fig. 11

PLEASE NOTE
The “xgenus® dc” unit, differently from “xgenus® ac”, is equipped with 2 communication cables.

Fig. 11

TERMINAL BOARD CONNECTION DIAGRAM

 BROWN L (line)
 YELLOW GREEN T (ground)
 BLUE N (neutral)

 BLACK

COMMUNICATION

 RED

COMMUNICATION

 YELLOW GREEN T (ground)


 YELLOW GREEN T (ground)

3. Connect the positioner arm shielded cable to the grounding potential 


4. Connect the wall plate to the grounding potential 

5. Clamp the cables with the stress relief clamps provided


6. Reassemble the terminal board insulating cover.


xgenus® dc only

22 xgenus ac/dc
5.6 ASSEMBLING THE RADIOGRAPHIC UNIT

ASSEMBLY INSTRUCTIONS

1. Take out the radiographic unit from the packaging (refer to Fig. 5)
2. Check that all the rating data match the power supply voltage
3. Remove both guards from the positioner arm by loosening the relevant screws (Fig. 12)

Fig. 12

4. Using a screw driver tip, separate the front coupling device (Fig. 13A)
5. Remove both guards (Fig. 13B)

Fig. 13A Fig. 13B

xgenus ac/dc 23
6. Insert the arm unit pin into the x-ray head (Fig. 14A) and insert the supporting retainer “C” clip
(Fig. 14B)

Fig. 14A Fig. 14B

7. Check that during insertion, the pin of the rotation prevention device correctly fits the seat located
on the x-ray head (Fig. 15)

Fig. 15

24 xgenus ac/dc
8. Couple the arm and x-ray head connectors and make sure that they are seated (Fig. 16)

Fig. 16

xgenus ac/dc 25
5.7 BALANCING THE POSITIONING ARM

CAUTION
The arm must be adjusted only when the radiographic unit is fully assembled and the x-
ray head is mounted.

WARNING
To prevent damage to the internal mechanism while performing adjustments and the balancing
tests, the adjustment wrench/key must not be left in place.

WARNING
The wrench/key provided must remain with the unit for future adjustments.

PLEASE NOTE
To reach the adjustment screw X the arm “A” must be put in vertical position.
To reach the adjustment screw Y the arm “B” must be put in horizontal position.
The adjustment key provided can be inserted only under the above conditions (Fig. 17)

arm “B”

Fig. 17

arm “A”

Fig. 17

26 xgenus ac/dc
Procedure to be followed to correctly balance the positioner arm (refer to Fig. 17):

1. BALANCING THE ARM “A”

PLEASE NOTE
The positioner type arm is supplied with arm “A” already factory adjusted.
The Arm “B” is supplied unloaded (no tension) for safety reasons.

2. BALANCING THE ARM “B”

For correct balancing, proceed as follows:

a. ARM “A” VERTICAL


b. ARM “B” HORIZONTAL
c. INSERT THE ADJUSTMENT WRENCH/KEY IN “Y”
d. LOAD THE SPRING BY APPLYING 22 TURNS WITH THE WRENCH IN A CLOCKWISE
DIRECTION.
e. WITHDRAW THE WRENCH/KEY

3. CHECKING THE BALANCE

To check the balance proceed as follows:

BRING THE ARM “B” TO SEVERAL DIFFERENT POSITIONS

if the head and arm do not hold the position you have selected:

a. BRING THE ARM “B” BACK TO THE HORIZONTAL POSITION


b. INSERT THE ADJUSTMENTWRENCH IN “Y”
c. ROTATE THE ADJUSTMENT WRENCH BY HALF TURN:
− CLOCKWISE IF IT TENDS TO DRIFT DOWN
− COUNTERCLOCKWISE IF IT DRIFTS UP
d. WITHDRAW THE WRENCH

PLEASE NOTE
Repeat the test and adjustment until the arm ”A” is steady and stable in all positions,
even when the arm “B” is completely extended.

4. READJUSTMENT OF ARM “A”

If the head and arm do not hold the position you have selected:

a. BRING THE ARM “A” TO THE VERTICAL POSITION AGAIN


b. INSERT THE ADJUSTMENT WRENCH IN “X”
c. TURN THE ADJUSTMENT WRENCH BY HALF TURN:
− CLOCKWISE IF IT TENDS TO COME DOWN
− COUNTERCLOCKWISE IF IT DRIFTS UP
d. WITHDRAW THE WRENCH

xgenus ac/dc 27
PLEASE NOTE
Repeat the test and adjustment until the arm “A” is steady and stable in all positions,
even when the arm “B” is completely extended.

5. UPON COMPLETION OF BALANCING:

a. Insert the movable finish cover (A) between the arm guard and metal frame (Fig. 18)

Fig. 18

b. Insert the pins  of one guard  into the relevant seats; then position it and check that the
finish piece  is coupled to the guards (Fig. 19)

 Fig.19

28 xgenus ac/dc
5.8 ASSEMBLING “xgenus® ac/dc” X-RAY CONTROL

CAUTION
Check that the cable runs are arranged in the x-ray control installation wall; check the
compliance of the power supply with the installation specifications
(refer to § 4 “INSTALLATION SPECIFICATIONS”).

CAUTION
Check that the rating data match the power supply voltage.

ASSEMBLY INSTRUCTIONS (Fig. 20)

1. Take the x-ray control out of the packaging (refer to Fig. 5)


2. Take out the drilling template 
3. Confirm the vertical and horizontal alignment. Mark the mounting holes on the wall using
the drilling template.
4. Drill using a Ø3mm tip, then drill again with a Ø6mm tip to prevent any flaking of the white
coat
5. Insert the three screw anchors provided into the holes 
6. Open the x-ray control by unscrewing and removing the three screws 

Fig. 20

xgenus ac/dc 29
7. Remove the 26-pole connector from its seat to release both control guards (Fig. 21)

CONNECTOR 26 pole

Fig. 21

8. Insert the electric feed cables into the hole (Fig. 22)

POWER SUPPLY CABLE


INLET HOLE

INLET HOLE
FOR THE RADIOGRAPHIC UNIT(S)
AND WARNING LAMP(S) CABLE

Fig. 22

9. Insert the connection cables coming from the radiographic units into the rectangular slot
10. Insert the cables of the external signal lamps (IF INSTALLED) (Fig. 22)
11. Insert the cables of the remote control button (IF INSTALLED) (Fig. 22)
12. Attach the x-ray control base to the wall, matching the three anchor screws with the holes
13. Tighten the screws (refer to Fig. 20 ) with the relevant washers (refer to Fig. 20 ) into
the screw anchors

PLEASE NOTE
If the wall is uneven, use appropriate shims to prevent the x-ray control from warping.

PLEASE NOTE
Keep the x-ray control clean. Protect the board from dust, dirt or wire particles.

30 xgenus ac/dc
5.9 X-RAY CONTROL ELECTRICAL CONNECTION

CAUTION
Before proceeding to connections, the power supply must be turned off.

CAUTION
If the equipment is mounted on metal wall or studs, these must be connected to the
grounding circuit of the system.

WARNING
While carrying out electrical connection procedures, always respect the polarity PHASE – NEUTRAL.

WARNING
While stripping the cable, pay attention to the small copper wires that may fall on the printed circuit
board and cause short circuits or malfunctions.

OPERATING INSTRUCTIONS (Fig. 23) (refer to §19 “ATTACHMENTS”)

1. Connect the power supply cable to the terminal board 


2. Insert the three mains cables into the rack
3. Fix them with the cable clamp 
4. Connect the cables coming from the operatory unit# 1 to the terminals XRAY1

4a. Connect the communication cables from the tube head 1 to the terminals C11 and C12

5. Connect the YELLOW-GREEN grounding cable to the metal "Grounding plate” 

6. Connect the cables coming from the operatory unit# 2 to the terminals XRAY2

6a. Connect the communication cables from the tube head 2 to the terminals C21 and C22

7. Connect the YELLOW-GREEN grounding cable to the metal “Grounding plate” 

8. Fasten the cables in the cable clamp 


9. Connect the remote control button (if installed) (refer to § 19 “ATTACHMENTS”)
10. Check the configuration on the dip-switches (refer to section 7 “SYSTEM CONFIGURATION”)
11. Reconnect the 26-pole connector
12. Close the x-ray control with the three screws
13. Mount the plate guard (refer to Fig. 6 )
14. Reconnect the power supply


xgenus® dc only

xgenus ac/dc 31
 

1 2 3A-3B 4A-4B 5 6 7

Fig. 23

The electronic PC card terminal boards include the follow:

1 KEY1 RADIOGRAPHIC UNIT N° 1 CONTROL BUTTON WITH LED


2 KEY2 RADIOGRAPHIC UNIT N° 2 CONTROL BUTTON (OPTIONAL) WITH LED
3A S12/S11 RADIOGRAPHIC UNIT N° 1 SIGNAL LAMP OUTSIDE THE OPERATORY
3B S22/S21 RADIOGRAPHIC UNIT N° 2 SIGNAL LAMP OUTSIDE THE OPERATORY
4A∗ C12/C11 RS232 COMMUNICATION FOR DC TUBE HEAD 1
4B∗ C22/C21 RS232 COMMUNICATION FOR DC TUBE HEAD 2
5 XRAY1 RADIOGRAPHIC UNIT N° 1 POWER SUPPLY
6 XRAY2 RADIOGRAPHIC UNIT N° 2 POWER SUPPLY
7 LINE X-RAY CONTROL POWER SUPPLY

WARNING
Check the value of the fuses F1-F2 (see schemes on § 19) which must be in compliance with the type
of tube head connected to the XRAY1 and XRAY2 terminal boards:

TUBE HEAD TYPE


FUSE xgenus® ac xgenus® dc
F1 (XRAY1) 8AF - 250V 12.5A - 250V
F2 (XRAY2) 8AF - 250V 12.5A - 250V


xgenus dc only

32 xgenus ac/dc
6. X-RAY CONTROL PANEL

The control panel is the user's interface, where the operator can:

with the keys

 select the radiographic unit operating parameters


 change the exposure times displayed
 store any new customized exposure times

with the light indicators “green Led”

 check the x-ray control operating conditions


 check the active function key
 check the selected radiographic unit parameter

xgenus ac/dc 33
control panel

MAIN ON/OFF SWITCH

LED INDICATOR

DISPLAY

X-RAY EXPOSURE
BUTTON
RADIOGRAPHIC
DISTANCE INDICATOR
(LONG CONE
RECOMMENDED-
MANDATORY OF DC
OPERATION)
KEY TO DECREASE
EXPOSURE TIME KEY TO INCREASE
EXPOSURE TIME

TUBEHEAD TYPE
TUBEHEAD
INDICATOR
SELECTION

SELECTION OF PATIENT’S
RADIOGRAPHIC SIZE
KV/VOLTAGE SELECTOR

MEMORY

MAXILLARY RADIOGRAPHIC CURRENT


UPPER TEETH SELECTOR

MANDIBULARY
LOWER TEETH OCCLUSAL EXAM

X-RAY EMISSION
SIGNAL

BITE-WING EXAM
DIGITAL RADIOGRAPHIC
DUTY CICLE PAUSE TECHNIQUE (HOLD FOR 3
INDICATOR SEC.-ALWAYS USE LONG
CONE TECHNIQUE)
MALFUNCTIONING
INDICATOR

FILM TYPE SELECTION

KEY SWITCH LOCK

34 xgenus ac/dc
7. SYSTEM CONFIGURATION
The “xgenus ac/dc” radiographic systems is factory configured for the following operative “standard
mode” configuration

on the control panel by pressing the RX button


Installation of 2 “xgenus” tube heads  Led 1 lights up,
by pressing again the button, Led 2 lights up

IF THE LONG 12” (31cm) CONE IS USED


A radiographic distance
SSD = 31cm  on the control panel Led 12” is lit
with long 12” cone

IF THE SHORT 8” (20cm) CONE IS USED


A radiographic distance
SSD = 20cm  on the control panel Led 8” is lit
with short 8” cone

with film type “D” 


on the control panel Led “D” is lit

CONTROL BUTTON The control includes a control button with long


to perform the exposure  flexible coil cord.

The installed radiographic units include the following features:

Led “AC” lit: “xgenus® ac”


tube head type
Led “DC” lit: “xgenus® dc”

the Led 70 kV or 60 kV
radiographic voltage
on the control panel must be lit

the Led 4mA or 8mA
radiographic current on the control panel must be lit
(ONLY AVAILABLE IN DC MODE)

The above configuration depends on the position of n° 8 mini-switches (dip-switch ) on the


x-ray control board.

DIP
PARAMETER ON OFF
SWITCH
1 RADIOGRAPHIC UNIT N° 1 INSTALLED NOT INSTALLED
2 TYPE OF RADIOGRAPHIC UNIT N° 1 DC AC
3 RADIOGRAPHIC UNIT N° 2 INSTALLED NOT INSTALLED
4 TYPE OF RADIOGRAPHIC UNIT N° 2 DC AC
5 CONTROL BUTTON N° 2 INSTALLED NOT INSTALLED
LONG (12”) SHORT (8”)
6 CONE
INSTALLED INSTALLED
7 NOT AVAILABLE NOT AVAILABLE
8 NOT AVAILABLE NOT AVAILABLE


xgenus® dc only

xgenus ac/dc 35
IF THE LONG 12” (31cm) CONE IS USED

ON OFF
1 ■ radiographic unit n° 1

2 ■ ■ AC (=off) / DC (=on)
3 ■ radiographic unit n° 2

4 ■ ■ AC (=off) / DC (=on)
5 ■ control button n° 2
6 ■ cone 12” (31cm)
7 ■ not available
8 ■ not available

IF THE SHORT 8” (20cm) CONE IS USED

ON OFF
1 ■ radiographic unit n° 1

2 ■ ■ AC (=off) / DC (=on)
3 ■ radiographic unit n° 2
4 ■∗ ■ AC (=off) / DC (=on)
5 ■ control button n° 2
6 ■ cone 8” (20cm)
7 ■ not available
8 ■ not available


xgenus® dc only

36 xgenus ac/dc
The configuration may be changed if:

POSSIBLE MODIFICATIONS HOW TO CARRY OUT THE MODIFICATIONS

− type “E” and “F” films are used

− a digital system is used By pressing the keys of the control panel



− radiographic current (4mA /8mA)  refer to USER’S MANUAL
∗ sect. 6 “ USE INSTRUCTIONS”
− radiographic voltage (60kV /70kV)

− the short 8” (20cm) cone is used


− the long 12” (31cm) cone is used by changing the dip-switch position
− one single radiographic unit is used  THIS OPERATION MUST BE CARRIED OUT BY THE AUTHORIZED
SERVICE PERSONAL ONLY
− n° 2 CONTROL BUTTONS are used

PLEASE NOTE
“xgenus® dc” tube head allows the following combinations for the radiographic voltage and current:

60kV/4mA – 60kV/8mA – 70kV/4mA – 70kV/8mA

“xgenus® ac” tube head allows only:


70kV/8mA


xgenus® dc only

xgenus ac/dc 37
8. CHANGING THE CONFIGURATION

CAUTION
This operation must be carried out by a qualified installer/technician.

WARNING
To install the changes you have made, turn off the x-ray control and then turn it on again.

To change the configuration, the dip-switch positions in the x-ray control must be changed:

DIP
PARAMETER ON OFF
SWITCH
1 RADIOGRAPHIC UNIT N° 1 INSTALLED NOT INSTALLED
2 TYPE OF RADIOGRAPHIC UNIT N° 1 DC AC
3 RADIOGRAPHIC UNIT N° 2 INSTALLED NOT INSTALLED
4 TYPE OF RADIOGRAPHIC UNIT N° 2 DC AC
5 CONTROL BUTTON N° 2 INSTALLED NOT INSTALLED
LONG (12”) SHORT (8”)
6 CONE
INSTALLED INSTALLED
7 NOT AVAILABLE NOT AVAILABLE
8 NOT AVAILABLE NOT AVAILABLE

1. CHANGING THE NUMBER OF RADIOGRAPHIC UNITS INSTALLED

To change the number of radiographic units installed, move the dip-switch n° 1 or n° 3

a. If the radiographic unit is connected to the RX1 terminal board, put the dip-switch n° 1
to the ON position, otherwise to the OFF position.

b. If the radiographic unit is connected to the RX2 terminal board, put the dip-switch n° 3
to the ON position, otherwise to the OFF position.

2. CHANGE THE TYPE OF TUBE HEAD INSTALLED

a. if the tube head connected to the XRAY1 is “ac” type, move the dip-switch n° 2 to the

OFF position; otherwise to the ON position.

b. if the tube head connected to the XRAY2 is “ac” type, move the dip-switch n° 4 to the

OFF position; otherwise to the ON position.

3. CHANGING THE CONE

To replace the 12” cone with the 8” cone, move the dip-switch n° 6 to the ON position.
Check that the led “8” lights up on the control panel.

PLEASE NOTE
After modification, the set exposure times are automatically changed.


xgenus® dc only

38 xgenus ac/dc
To replace the 8” cone with the 12” cone, move the dip-switch n° 6 to the ON position.
Check that the led “12” lights up on the control panel.

PLEASE NOTE
After modification, the set exposure times are automatically changed.

4. CHANGING THE AMOUNT OF THE REMOTE BUTTONS

To change the number of control buttons, move the dip-switch n° 5 to the ON position.

PLEASE NOTE
After modification, each radiographic unit is controlled by the relevant button.

PLEASE NOTE
It’s possible to connect to the timer:
n° 2 “xgenus® ac”
n° 2 “xgenus® dc”
n° 1 “xgenus® ac” + n°1 “xgenus® dc”

xgenus ac/dc 39
9. START UP

CAUTION
For further information, see the USER'S MANUAL.

TURN ON THE X-RAY CONTROL to provide power to the system

a. Turn THE “KEY SWITCH” to position “I” (ON).

b. Set the “MAIN SWITCH” located on the upper part of the control to the
position “I” (ON).

 The green light turns on indicating that the system is powered


 The LED’s of the set radiographic parameters automatically light up
 The exposure time is shown on the display

c. THE RADIOGRAPHIC SYSTEM IS NOW READY FOR USE

ATTENTION
If an error is detected when the system is turned on, the anomaly is indicated as
follows:

• an intermittent beep sounds off

• The “MALFUNCTIONING INDICATOR” Led turns on intermittently


The error code (E ….) appears on the display
(refer to §14 “ERROR MESSAGES”)
• All CONTROL PANEL functions are inhibited

In this case, turn off the timer and then turn it back on again.
If the error should repeat itself, call for “Technical Assistance”.

PLEASE NOTE
The exposure time and radiographic parameters which appear on the display are the last that
were set before the x-ray control was turned off.

PLEASE NOTE
If installed outside the operatory, the RX signaling lamp corresponding to the selected tube head
turns on.

PLEASE NOTE
If the x-ray control remains unused and inactive for a few minutes, it switches to the stand-by
mode. Press any key of the CONTROL PANEL to restore it to the operative mode.
The previous settings will be retained.

40 xgenus ac/dc
10. CHECKING THE INSTALLATION

CAUTION
When assembly and all connections are completed, the installer must check the electrical
safety and function of the system for compliance.

OPERATING INSTRUCTIONS:

1. Checking the configuration

Check on the control panel that all LEDs corresponding to the required configuration are lit;
otherwise, change them.
(Refer to section. 7 “SYSTEM CONFIGURATION” and section.8 “CHANGING THE
CONFIGURATION”.)

2. Checking the x-ray control operation

a. Verify the correct operation of the control panel by selecting different exposure times
b. Verify the time on the display
c. When changing the selected radiographic unit, check that the corresponding external
RX signal lamps (if installed) turn on.

3. Checking the exposure: CAUTION! Radiation will be emitted.

a. Set an exposure time of 1 sec


b. Lift the "CONTROL BUTTON" on the x-ray control
c. Using the extensible cable of the “CONTROL BUTTON”, keep a safe distance (length
of the cable) or at least 2 meters (7 feet) from the radiographic unit, to check the
radiographic exposure.

d. Press the “X-RAY KEY” and keep it pressed until the acoustic signal (beep)

stops and the yellow “X-RAY OUTPUT SIGNAL” Led turns off.

PLEASE NOTE
If the “EXPOSURE BUTTON” is released early, the exposure is immediately
interrupted and the E12 error message appears on the display.

e.
At the end of the exposure, verify the operation of the Led “PAUSE INDICATOR”

f. Verify on the display the indication of the actual exposure time


(Refer to section 11.4 “EXPOSURE TIME (sec)”)

xgenus ac/dc 41
4. Checking the operation of the radiographic unit: CAUTION! Radiation will be
emitted

Carry out several exposures on the installed radiographic units installed and check for the
following:

a. There are no errors.


b. The LED (AC or DC) of the selected radiographic unit is lit.
c. The "CONTROL BUTTON" LED is lit over the duration of the acoustic signal.

5. Checking the power drawn by the radiographic system: CAUTION! Radiation will
be emitted.

To check the power drawn by the radiographic system, a meter must be used, with the
Ammeter in “AC”

a. Connect the instrument to the power supply line


(refer to section 5.9 “X-RAY CONTROL ELECTRICAL CONNECTION”)
b. Set an exposure time of approx. 3 sec. on the x-ray control.
∗ ®
b. If using an xgenus dc, set 70kV-8mA-3.2s on the x-ray control.
c. Carry out an exposure and read the current value on the instrument

PLEASE NOTE
The radiographic system complies with the requirements when:

the current drawn is ≤ 8A for line voltage of 115V



the current drawn is ≤ 12.5A for line voltage of 115V

Otherwise, check the electrical system or call for “Technical Assistance”.

6. Checking the electrical system

To check the electrical system, a meter must be used, with the Voltmeter in “AC” mode.

a. Connect the instrument to the terminals L and N on the x-ray control


b. Measure the line voltage (load voltage V0)
c. Connect the instrument to the terminals L and N of the wall plate terminal board
(refer to section 5.5 – fig.11 - “CONNECTION TO THE TERMINAL BOARD”)
d. Set an exposure time of approx. 3 sec on the x-ray control
∗ ®
d. If using an xgenus dc, set 70kV-8mA-3.2s on the x-ray control.
e. Release the exposure
f. Measure the line voltage during exposure (load voltage V1)

PLEASE NOTE
The electric system complies with the requirements when the result of the formula is
as follows:
∗ ∗∗
(V0 – V1 ) / V0 ≤ 0.03 or 0.012

Otherwise, the electric system must be adjusted


(refer to section 4: “INSTALLATION SPECIFICATIONS”)


xgenus® dc only

42 xgenus ac/dc
11. CHECKING THE EXPOSURE FACTORS

1. Radiographic voltage (kVp)

The radiographic high voltage is measured using a calibrated "non invasive" instrument with an initial
∗∗
delay set of 50ms . The kVp value is factory checked with a PMX-I D.

xgenus® ac xgenus® dc
Line Voltage Vn -5% / +10% Vn ±15%
Max Voltage drop 1.2% 3%
SET Nominal High Voltage 70 kV 60-70 kV
TECHNICAL Nominal current 8mA 4mA – 8 mA
FACTORS
Set exposure time 3.2 s 3.2 s
The radiographic VOLTAGE is: 70kVp ±15% for “xgenus® ac”
60kVp – 70kVp ±10% for “xgenus® dc”

2. Radiographic current (mA)

The radiographic current is measured by connecting a milliampmeter. The mA value is factory


checked with a digital multimeter.
xgenus® ac xgenus® dc
Line Voltage Vn -5% / +10% Vn ±15%
Max Voltage drop 1.2% 3%
SET Nominal High Voltage 70 kV 60-70 kV
TECHNICAL Nominal current 8mA 4mA – 8 mA
FACTORS
Set exposure time 3.2 s 3.2 s
®
The radiographic CURRENT is: 8mA ±15% for “xgenus ac”
4mA – 8mA ±10% for “xgenus® dc”

3. Dose (mGy)

The dose in air is measured with a "non invasive" instrument, by positioning the detector at a spot-
skin distance = 310mm (12”). The mGy are factory checked with the instrument SOLIDOSE.

xgenus® ac xgenus® dc
Line Voltage Vn -5% / +10% Vn ±15%
Max Voltage drop 1.2% 3%
Nominal High Voltage 70 kV 60-70 kV
Nominal current 8mA 4mA – 8 mA
SET Set exposure time 1s 1s
TECHNICAL ∗∗
Dose in air  3.3 mGy/s ±30% (with 12” cone)
FACTORS
6.5 mGy/s ±30% (with 8” cone)

Dose in air  (60kVp – 4mA)=2.2mGy/s±30% (with 12” cone)
(70kVp – 4mA)=3mGy/s±30% (with 12” cone)
(60kVp – 8mA)=4.5mGy/s±30% (with 12” cone)
(70kVp – 8mA)=5.5mGy/s±30% (with 12” cone)

∗∗
xgenus® ac only
∗∗
xgenus® ac only

xgenus® dc only equipped with a 12” cone. See §13 of the “USER’S MANUAL” for the dose in air with a 8” cone.

xgenus ac/dc 43
4. Exposure time (s)

The exposure time is measured with a “non invasive” instrument.

∗∗
PLEASE NOTE
The “xgenus” x-ray control features the "self-compensating" technology.
The automatic correction or compensation of the exposure time offsets the effects of the
line voltage fluctuations on x-ray film quality.
Radiographic over/under exposure is eliminated.
The automatic correction system works as follows:
• A voltmeter located inside the timer constantly reads the mains voltage
while the operator selects the required exposure time.
• The last voltage measured (before the exposure) is sent to the
microprocessor. The selected exposure time is corrected to voltage by
applying a special algorithm.
• At the end of the exposure, during the pause time, the time-corrected value
is shown on the display of the x-ray control.
THE PAUSE TIME IS APPROX. 32 TIMES THE SET TIME.
The correct time is the actual duration of the exposure

The accuracy of the “actual time” is factory checked with instrument PMX-I.

xgenus® ac xgenus® dc
Line Voltage Vn -5% / +10% Vn ±15%
Max Voltage drop 1.2% 3%
SET Nominal High Voltage 70 kV 60-70 kV
TECHNICAL Nominal current 8mA 4mA – 8 mA
FACTORS
Set exposure time 0.2 s 1.6 s
®
The EXPOSURE TIME measured is: 0.2s ±10% for “xgenus ac”
1.6s ±10% for “xgenus® dc”

∗∗
5. Checking the exposure time (continued on p. 45: “Operating instructions”)

WARNING
While performing periodic quality checks, some differences between the set exposure time and
the measured exposure time may be detected.
The differences are due to the automatic time compensation that intervenes when the power
voltage differs from the nominal one.

The x-ray control is provided with two types of accuracy measurements for the exposure time:

A. Set exposure (is the exposure time set by the operator on the x-ray control).

The max error is +100% / - 100% when the voltage fluctuation is -5% / +10%.

B. After exposure: it’s the exposure duration (automatically adjusted). It’s calculated by the
microprocessor according to the line voltage, and it’s shown on the display.

∗∗
xgenus® ac only

44 xgenus ac/dc
OPERATING INSTRUCTIONS:

CAUTION
Radiations will be emitted.

To check the exposure times (sec) it is advisable to use calibrated “non invasive” instruments;
i.e.:
NERO by Victoreen
PMX by RTI
TRIAD QA by Keithley
a. Put the “X-ray detector” at the cone outlet; center the primary beam.
b. Turn on the x-ray control
c. Set one exposure time on the control.
d. Release the exposure.
e. At the end of the exposure, when the functional pause period starts, read the actual time of
exposure on the display.
f. To check the accuracy of the exposure time, use the chart below.

CONFORMITY CHECK CHART

At the end of the exposure

read the time value


1° = Actual exposure time
on the display

read the measure


2° = Emission time measure
on the instrument

3° subtract 1° from 2°

4° divide 3° by 2° = Absolute error

5° multiply 4° by 100 = Percentage error (%)

The accuracy is checked when the


by adopting the most favorable
% of error is ±10% o ±20msec
condition.

PLEASE NOTE
The exposure time complies with the requirements when:

((Measured time – Actual time) / Measured time)) x 100 ≤ ±10% or ±20 ms

Otherwise, please contact “Technical Support”.

xgenus ac/dc 45
12. DIAGNOSTIC

The “xgenus ac/dc” radiographic systems allows the installer to set or display some functional
parameters of the x-ray control.

1. To set the minimum exposure time:

a. Select the tube head


b. Turn off the x-ray control

c. Turn on the x-ray control, by keeping the key pressed


d. The message “SEC” is displayed for approx 1sec
e. The present value of the lower set limit is displayed

f. To change the value, press the keys and

g. To confirm, press the key


h. To exit this mode, turn the x-ray control off and then on again

WARNING
Both in digital and in conventional mode, the minimum exposure time MUST NOT BE LESS than
∗∗
0,08s

2. To visualize the parameters, the installer must:

a. Press simultaneously and keep pressed the following keys


(2) UPPER JAW MOLAR + (31) LOWER JAW MOLAR

b. Press the key associated to the parameter one wishes to see

KEY DISPLAYED PARAMETER Example U.M.


BITE-WING RADIOGRAPHIC SYSTEM
110 Volt
ANTERIOR VOLTAGE

BITE-WING
LINE VOLTAGE 105 Volt
POSTERIOR

UPPER MAXIMUM LINE VOLTAGE Volt


∗∗ 115
INCISOR VALUE DETECTED

LOWER MINIMUM LINE VOLTAGE


∗∗ 118 Volt
INCISOR VALUE DETECTED

OCCLUSAL SOFTWARE VERSION 1.5 ----

∗∗
xgenus® ac only

46 xgenus ac/dc

13. TUBE HEAD CALIBRATION∗

CAUTION
During this operation radiations will be emitted.
It is mandatory to carry out all the safety measures relevant to radioprotection.

INSTRUCTIONS

1. turn off the timer

2. turn on the timer by keeping the key pressed.

3. the message TUBE is displayed.

4. select the appropriate tube head to calibrate.

5. take an exposure by pressing the EXPOSURE BUTTON key and keep it

pressed until the acoustic signal (beeping) stops and the yellow led turns off.

6. Once the exposure has been taken, if the display does not show errors, the calibration has been
successfully done.


xgenus® dc only

xgenus ac/dc 47
14. ERROR MESSAGES

The following chart gives a list of error messages that may appear while the “xgenus ac/dc”
radiographic systems is working.

The chart also includes the reason for the error messages and what to do to solve them.

ERROR
CAUSE SOLUTION
MESSAGES
RX1 TUBEHEAD
CALL TECHNICAL SUPPORT
IS NOT CONNECTED OR IS OUT OF ORDER
RX2 TUBEHEAD CALL TECHNICAL SUPPORT
IS NOT CONNECTED OR IS OUT OF ORDER
CORRUPTED EEPROM DATA CALL TECHNICAL SUPPORT

EEPROM DATA NOT SAVED PROPERLY CALL TECHNICAL SUPPORT

LINE VOLTAGE VALUE NOT CALL TECHNICAL SUPPORT


WITHIN RANGE
THE X-RAY KEY
MAKE SURE IT IS NOT JAMMED OR DEFECTIVE
ALWAYS SEEMS TO BE PRESSED
ANOMALY IN THE CONTROL PANEL CALL TECHNICAL SUPPORT
THE EXPOSURE KEEP THE X-RAY KEY
HAS BEEN PREMATURELY INTERRUPTED PRESSED TILL THE END OF THE EXPOSURE
ANOMALY IN THE TRIAC/RELAY CALL TECHNICAL SUPPORT

ANOMALY IN THE ELECTRONIC CIRCUIT CALL TECHNICAL SUPPORT

ANOMALY IN THE CONTROL CIRCUIT CALL TECHNICAL SUPPORT

INCORRECT DIP-SWITCH CONFIGURATION CALL TECHNICAL SUPPORT

THE CONTROL BUTTON DOES NOT CORRESPOND CALL TECHNICAL SUPPORT


TO THE SELECTED TUBEHEAD
∗ THE TUBEHEAD CALL TECHNICAL SUPPORT
DOES NOT WORK PROPERLY
∗ THE TUBEHEAD CALL TECHNICAL SUPPORT
IS NOT IN THE CORRECT MODE
∗ THE TUBEHEAD REPEAT THE EXPOSURE
HAS NOT COMPLETED THE EXPOSURE CALL TECHNICAL SUPPORT
∗ THE TUBEHEAD CALL TECHNICAL SUPPORT
IS NOT CALIBRATED
∗ EEPROM DATA NOT SAVED PROPERLY CALL TECHNICAL SUPPORT

∗ CORRUPTED EEPROM DATA CALL TECHNICAL SUPPORT

∗ OVERVOLTAGE ERROR CALL TECHNICAL SUPPORT

∗ ANODE VOLTAGE OUT OF TOLERANCE CALL TECHNICAL SUPPORT

∗ ANODE CURRENT OUT OF TOLERANCE CALL TECHNICAL SUPPORT

∗ CONTROL CONNECTOR CALL TECHNICAL SUPPORT

∗ PROBLEM IN THE REFERENCE VOLTAGE CALL TECHNICAL SUPPORT

ALL FUNCTIONS ARE DISABLED


MAJOR ERROR
CALL TECHNICAL SUPPORT


xgenus® dc only

48 xgenus ac/dc
15. REPLACEMENT OF FUSES

The x-ray control electronic equipment is protected by 4 fuses located on the electronic P.C.B.
To replace them, proceed as follows (Fig. 25):
1. Turn off the power supply
2. Remove the front cover from the x-ray control  by unscrewing the fixing screws 

3. Identify the fuse to be replaced (Fig. 26)


4. Remove the plastic protection (Fig. 27)
5. Remove the fuse
6. ‡
Replace it with one of same type

TYPE OF TUBE HEAD CONNECTED xgenus® ac xgenus® dc


POWER VOLTAGE
OF THE RADIOGRAPHIC SYSTEM
115V (-5% /+10%) 115V (±15%)
Internal protection fuses F1 (XRAY1) 8AF – 250V 12.5A – 250V
(5x20) quick F2 (XRAY2) 8AF – 250V 12.5A – 250V
(see schemes on § 19) F3 – F4 12.5A – 250V 12.5A – 250V

7. Put the fuse protection back in place.

Fig. 26 Fig. 27
8. Replace the x-ray control cover.
9. Restore the power supply.


n° 2 spare fuses are supplied in a plastic bag inside the XGT control.

xgenus ac/dc 49
16. MAINTENANCE

To guarantee the safety of the radiographic system, the following checks must be carried out:

CAUTION
The owner is responsible for arranging the maintenance schedule.
The radiographic system must undergo maintenance to check correct operation every
12 months.

1. Turn off the power


2. Release the spring tension of the arm “B” of the positioner using the wrench provided.
(Please see page 27)
3. Remove the radiographic generator.
(Refer to section 5.6)
4. Remove the wall plate cover.
(Refer to section 5.2)
5. Remove the terminal board cover and disconnect the cable
6. Remove the bracket plug and withdraw the guard slat.
(Refer to section 5.4)
7. Remove the positioner and the relevant cable from the bracket
8. Remove the bracket from the wall plate
(Refer to section 5.3)
9. Check the vertical alignment of the wall plate – adjust if required
10. Check the six screws on the wall plate – tighten if required
11. Clean off the old lubricating grease from the bracket shaft - should the bracket shaft be
damaged, install a new bracket
12. Clean the old lubricating grease off the bracket bushing - should the bracket bushing be
damaged, install a new bracket
13. Grease the bracket shaft (use grease MOLIKOTE D)
14. Lubricate the bracket bushing with lubrication grease (use grease MOLIKOTE D)
15. Install the bracket in the wall plate
(Refer to section 5.3)
16. Check the arm cable – should it be damaged, send the arm to the manufacturer for repairs
17. Check the arm covers
18. Replace if the guards are damaged
19. Clean the old grease off the shaft - should the shaft be damaged, send the arm to the
manufacturer/dealer for repairs.
(Refer to section 5.4)
20. Lubricate the arm shaft with lubricating grease (use grease MOLIKOTE D) and reposition it in
the bracket
21. Put again the arm cable in the bracket and the wall plate. Connect it to the terminal board
and re-attach the terminal board cover
(Refer to section 5.5)
22. Position the guard slat in the bracket
23. Position the bracket plug
24. Position the plate guard
25. Check the electric contact of the radiographic unit. If damaged, send the radiographic unit to
the manufacturer/dealer for repairs
26. Clean the old grease from the radiographic unit assembly shaft
27. Grease the assembly shaft of the radiographic unit with a thin layer of lubricating grease
(use grease MOLIKOTE D)
28. Position the radiographic unit again
29. Load the spring of the positioner arm “B” using the wrench provided
(Refer to 5.7)
30. Turn on the power and check the correct operation of the radiographic system.

50 xgenus ac/dc
17. CLEANING THE OUTER SURFACES

Use a soft cloth dampened with water and mild soap to clean the outer surfaces of the x-ray.

DO NOT permit water or any liquid to penetrate the machine or severe electrical component damage
may occur. Damage of this nature is not covered under warranty.

For protection against cross-infection, it is recommended that barrier protection be used and replaced
for each patient.

Germicides must be selected carefully, realizing that many are not suitable for use with such plastics
that are used in manufacturing. Damage of this nature is not covered under warranty.

18. IF A REPAIR SHOULD BECOME NECESSARY

In case of a malfunction, contact your local Technical Service Representative.


They will call TG Group for an RA Number (Return Authorization). This will authorize one the return of
the defective part to TG Group for replacement or repair.

USING THE ORIGINAL PACKAGING

Send all materials to:

TG Group, Inc.
3505 Laird Road, Unit # 7
Mississauga, Ontario L5L 5Y7
Canada

Tel. 1-877-557-4888
Fax 1-905-832-0732

Manufacturer:

de Götzen S.r.l.
Via Roma, 45
21057 OLGIATE OLONA VA - ITALY

Tel. +39 0331 376760 r.a.


Fax +39 0331 376763

E-mail: degotzen@degotzen.com

xgenus ac/dc 51
19. ATTACHMENTS

The manufacturer undertakes to supply, upon request, drawings, circuit diagrams, component parts
lists, instructions or other information needed by qualified technical personnel to perform repairs on
those parts of the “xgenus ac/dc” radiographic systems which may be repaired.

52 xgenus ac/dc
xgenus ac/dc 53
A B C

REF SPARE PARTS TUBEHEAD ‘xgenus® ac’ CODE Q.TY PCS


1 AC XG (1) FORK PLASTIC COVER 29700983 7
2 AC/DC XG (2) RING NUT SET FOR THE 29700980 5
FORK
3 AC/DC XG (3) FORK 29700777 3
4 AC/DC XG (4) ROTATION LIMITING 29700750 1
DEVICE
5 AC XG (5) CABLE + 3 POLES CONNECTOR 39700302 1
A LONG CONE 31cm (12”) Ø60 29700397 1
B SHORT CONE 20cm (8”) Ø60 29700396 1
C RECTANGULAR CONE 31cm (12”) 44x35 29700615 1

54 xgenus ac/dc
REF SPARE PARTS TUBEHEAD ‘xgenus® dc’ CODE Q.TY PCS
1 DC XG (1) FORK PLASTIC COVER 29700979 8
2 AC/DC XG(2) RING NUT SET FOR THE 29700980 5
FORK
3 AC/DC XG (3) FORK 29700777 3
4 AC/DC XG (4) ROTATION LIMITING 29700750 1
DEVICE
5 DC XG (5) WASHER 27500959 1
6 DC XG (6) AIR CARD 29700981 9
7 DC XG (7) AIR CARD COVER 29700982 11
8 DC XG (8) HEAD CABLE 39700385 1

xgenus ac/dc 55
REF SPARE PARTS ‘xgenus® ECB’ CODE Q.TY PCS
(OPTIONAL)
1 XG ECB (1) SUP & INF SHELL 29700978 4
2 MAIN BOARD WITH SWITCH AND LED 39700303 1
3 XG ECB (3) KEYLOCK SWITCH 35100298 1

56 xgenus ac/dc
REF SPARE PARTS xgenus® LIGHT CODE Q.TY PCS
(OPTIONAL)
1 XG LIGHT (1) SUP & INF SHELL 29700977 4
2 XG LIGHT (2) LAMP HOLDER 33400372 1
3 XG LIGHT (3) LAMP 230V 15W 33300352 1
3 XG LIGHT (3) LAMP 115V 15W 33300390 1
4 XG LIGHT (4) 230V BOARD 39200354 3
4 XG LIGHT (4) 115V BOARD 39200353 3

xgenus ac/dc 57
Ref SPARE PARTS TIMER ‘xgenus®’ ac/dc CODE Q.TY PCS
1 AC/DC XG TIMER (1) TEMPLATE 24200883 1
2 AC/DC XG TIMER (2) CONTROL BOARD 39200306 7
2 AC/DC XG TIMER (2) CONTROL BOARD US 39200307 7
3 AC/DC XG TIMER (3) XM COILED CABLE 32000325 1
4 AC/DC XG TIMER (4) XM DOUBLE COILED 32000359 1
CABLE
5 AC/DC XG TIMER (5) EXPOSURE BUTTON 29700974 4
HOUSING
6 AC/DC XG TIMER (6) EXPOSURE BUTTON 39700303 1
BOARD
7 AC/DC XG TIMER (7) LED BOARD 39200313 7
8 AC/DC XG TIMER (8) DEG KEYBOARD 39200308 1
8 AC/DC XG TIMER (8) SAT KEYBOARD 39200309 1
9 AC/DC XG TIMER (9) REAR COVER 29700975 1
10 AC/DC XG TIMER (10) DEG FRONT COVER 29700304 4
10 AC/DC XG TIMER (10) SAT FRONT COVER 29700976 4
11 AC/DC XG TIMER (11) KEYSWITCH LOCK 39700305 2
12 AC/DC XG TIMER (12) FILTER 36200382 3
13 GREEN BIPOLAR SWITCH 35000044 1
14 BIPOLAR SWITCH PROTECTION 33500228 1

58 xgenus ac/dc
REF SPARE PARTS POSITINER ARM CODE QTY PCS
‘xgenus®’ ac/dc
1 AC XG B (1) CABLE + 3 POLES CONNECTOR 39700384 1
2 DC XG B (2) CABLE + 5 POLES CONNECTOR 39700381 1
3 AC/DC XG B (3) HEAD SIDE SHELLS 29700990 9
4 POSITIONER ARM CONNECTING ROD ANT 29700460 4
5 POSITIONER ARM TIE ROD HOLDER 29700458 6
6 AC/DC XG B (6) CAPS (set) 29700991 7
7 AC/DC XG B (7) EXT COVER 27500757 1
8 AC/DC XG B (8) INT COVER 27500758 1
9 AC/DC XG B (9) CENTRAL SHELLS 29700993 8
10 POSITIONER ARM LOADING WRENCH 22000543 1
11 AC/DC XG B (11) INT COVER 27500758 1
12 AC/DC B (12) EXT COVER 27500757 1
13 POSITIONER ARM CONNECTING ROD POST 29700992 7
14 AC/DC XG B SAFETY KEY 22000176 1
15 AC/DC B (15) BRACKET SIDE SHELLS 29700994 8

xgenus ac/dc 59
REF SPARE PARTS WALL PLATE 350 CODE Q.TY PCS
1 WALL BOX COVER (1) DEG 29700985 5
1 AC/ XG WALL BOX (1) COVER SAT 29700986 5
2 AC/DC XG B & WALL PLATE SAFETY KEY 22000176 1
3 WALL PLATE (3) TEMPLATE 24200984 1
4 WALL PLATE (4) INSERT + CAPS 29701036 3

WALL PLATE 350 DEG 29700383 COMPLETO


AC/ XG WALL PLATE 350 SAT 29700987 COMPLETO

60 xgenus ac/dc
REF SPARE PARTS WALL PLATE 350 TOP Q.TY PCS
CODE
1 WALL PLATE (1) COVER DEG 29700985 5
1 AC/ XG WALL PLATE (1) COVER SAT 29700986 5
2 AC/DC XG B & WALL PLATE SAFETY KEY 22000176 1
3 WALL PLATE (3) TEMPLATE TOP MOUNT 24201002 1
4 WALL PLATE (4) INSERT + CAPS 29701036 3
5 WALL PLATE (5) WASHER TOP MOUNT 22000597 1

WALL PLATE 350 TOP DEG 29700988 COMPLETO


AC/ XG WALL PLATE 350 TOP SAT 29700989 COMPLETO

xgenus ac/dc 61
Ref ARM SPARE PARTS CODE Q.TY PCS
BRACKET 400 29700380 COMPLETO
BRACKET 800 29700378 COMPLETO
BRACKET 1100 29700379 COMPLETO
BRACKET 1100 DC XG 29701019 CPMPLETO

Ref ARM SPARE PARTS CODE Q.TY PCS


CEILING BRACKET 29700382 COMPLETO

Ref ARM SPARE PARTS CODE Q.TY PCS


UNIT BRACKET 29700381 COMPLETO

Ref ARM SPARE PARTS CODE Q.TY PCS


BRACKET 400 TOP MOUNT 29700620 COMPLETO
BRACKET 800 TOP MOUNT 29700621 COMPLETO
BRACKET 1100 TOP MOUNT 29700622 COMPLETO
BRACKET 1100 TOP MOUNT DC XG 29701035 COMPLETO

62 xgenus ac/dc
Dichiarazione di Conformità
Declaration of Conformity Statement - Déclaration de Conformité
Konformitäts-Erklärung - Déclaration de Conformidad de la

Impresa
Company
Società
Firma
Empresa

dichiara che il dispositivo medico sistema radiografico per diagnosi intra-orale


declares that the medical device intraoral X-ray system
declare que le dispositif medical appareil pour radioagnostic dentaire
erklärt, daß das medizinisches Gerät Röntgendiagnose-Einheit
declara que el dispositivo medico rayos-X infraorales

modello
type of machine
modèle de machine xgenus®
bauart der maschine
modelo de la máquina

è conforme ai requisiti delle direttive


is in compliance with the directives
est conforme aux directives
ist den Richtlinien gemäß
comple las directives

dispositivi medici
medical devices
dispositives medicaux 93/42/CEE
medizinische geräte
dispositivos medicos

organismo notificato
notified body
organisme notizie 0434
verantwortliche gestaltung
organismo responsible

certificato n°
certificate No.
certificat n° 2002 – OSL – MDD – 0044
zertifikat Nr.
certificato n°

compatibilità elettromagnetica e successive modifiche


electromagnetic compatibility and subsequent amendments
compatibilità electromagnétique 89/336/CEE et modifications suivantes
elektromagnetische verordnung und folgenden Änderungen
contabilidad electromagnetica y modificaciones singuentes

bassa tensione
low voltage
basse tension 73/23/CEE
niederspannung
baja tension
Il dispositivo definito di classe IIb è conforme ai requisiti dell’allegato II della direttiva 93/42/CEE
This product, definited as a class IIb device, is conformed to the requirements of annex II of the council directive 93/42/ECC
Le produit ci-mentionné, defini comme dispositif de classe IIb, est conforme aux requises de l’annex II de la directive 93/42/EEC
Dieses Gerät, bestimmt als Klasse IIb, ist den richtilinien 93/42/EWG, annex II gemäß
El producto aqui mencionado esta definido como dispositivo de clase IIb y esta conformado a los requisitos de l’annex II de la directiva 93/42/CEE

data di emissione
issuing date
date de èmission 18.03.2002
ausstelldatum
fecha de emisio

xgenus ac/dc 63
Dichiarazione di Conformità
Declaration of Conformity Statement - Déclaration de Conformité
Konformitäts-Erklärung - Déclaration de Conformidad de la

Impresa
Company
Società
Firma
Empresa

dichiara che il dispositivo medico sistema radiografico per diagnosi intra-orale


declares that the medical device intraoral X-ray system
declare que le dispositif medical appareil pour radioagnostic dentaire
erklärt, daß das medizinisches Gerät Röntgendiagnose-Einheit
declara que el dispositivo medico rayos-X infraorales

modello
type of machine
modèle de machine xgenus® dc
bauart der maschine
modelo de la máquina

è conforme ai requisiti delle direttive


is in compliance with the directives
est conforme aux directives
ist den Richtlinien gemäß
comple las directives

dispositivi medici
medical devices
dispositives medicaux 93/42/CEE
medizinische geräte
dispositivos medicos

organismo notificato
notified body
organisme notizie 0434
verantwortliche gestaltung
organismo responsible

certificato n°
certificate No.
certificat n° 2002 – OSL – MDD – 0044
zertifikat Nr.
certificato n°

compatibilità elettromagnetica e successive modifiche


electromagnetic compatibility and subsequent amendments
compatibilità electromagnétique 89/336/CEE et modifications suivantes
elektromagnetische verordnung und folgenden Änderungen
contabilidad electromagnetica y modificaciones singuentes

bassa tensione
low voltage
basse tension 73/23/CEE
Niederspannung
baja tension
Il dispositivo definito di classe IIb è conforme ai requisiti dell’allegato II della direttiva 93/42/CEE
This product, definited as a class IIb device, is conformed to the requirements of annex II of the council directive 93/42/ECC
Le produit ci-mentionné, defini comme dispositif de classe IIb, est conforme aux requises de l’annex II de la directive 93/42/EEC
Dieses Gerät, bestimmt als Klasse IIb, ist den richtilinien 93/42/EWG, annex II gemäß
El producto aqui mencionado esta definido como dispositivo de clase IIb y esta conformado a los requisitos de l’annex II de la directiva 93/42/CEE

data di emissione
issuing date
date de èmission 25.06.2002
ausstelldatum
fecha de emisio

64 xgenus ac/dc