“de Götzen S.r.l.” reserves itself the right to modify the production and the manual without notice on their
designated distribution.
“de Götzen S.r.l.” shall not be liable for the incorrect use of the information contained in this
manual.
Copies, even excerpts, of this manual are allowed only for personal use and with written consent of
the manufacturer.
Edition: 01/2004
2 xgenus ac/dc
SUMMARY
INTRODUCTION page 2
SUMMARY page 3
PRELIMINARY INFORMATION page 4
INFORMATION FOR THE INSTALLER page 5
xgenus ac/dc 3
PRELIMINARY INFORMATION
Before beginning to use the “xgenus ac/dc” radiographic systems, it is advisable to carefully read
and follow the instructions contained herein, so as to obtain the best possible performance.
Always pay close attention to the CAUTION, WARNING, and PLEASE NOTE messages when
operating the system.
LEGEND
CAUTION
The word CAUTION identifies those situations that might compromise the operator's personal
safety or cause injuries to others.
WARNING
The word WARNING identifies those occurrences which might compromise the xgenus® ac/dc
x-ray systems’ performance.
PLEASE NOTE
The words PLEASE NOTE serves to draw your attention to special indications to facilitate
maintenance or make important information more clear.
4 xgenus ac/dc
INFORMATION FOR THE INSTALLER
CAUTION
The installer is responsible for the installation, calibration and testing with regard to
system safety and efficient operation.
For safe and reliable installation of the “xgenus ac/dc” radiographic systems, it is advisable to:
Check that the voltage mentioned on the rating plate matches the line voltage.
Install the “xgenus ac/dc” radiographic systems according to the procedures described
in this manual.
Provide the user with any information regarding the use of the “xgenus ac/dc”
radiographic systems according to that which is provided in the operators manual.
xgenus ac/dc 5
1. “xgenus ac/dc” RADIOGRAPHIC SYSTEMS
CONE
POSITIONER ARM (Scissor)
CONTROL – XGT
WALL BOX
HORIZONTAL ARM
OPTIONAL ACCESSORIES :
Fig. 1
∗
xgenus® dc only
6 xgenus ac/dc
2. PICTOGRAMS USED
∗
The identification tags on the tube heads XGG/XGDC , on the control XGT and on the cone indicate the
model name, the serial number, the manufacturing date and the symbols of the main technical
characteristics.
ID TAG OF THE
“xgenus ac”
TUBE HEAD XGG
ID TAG OF THE
“xgenus dc”
TUBE HEAD XGDC
ID TAG OF THE
“xgenus”
CONTROL XGT
∗
“xgenus® dc” only
xgenus ac/dc 7
SYMBOLS:
This symbol guarantees that the radiographic system complies with the regulations
contained in the European Directive EEC 93/42 regarding Medical Devices
The degree of protection against direct and indirect electric contacts is B type
Ground cable
8 xgenus ac/dc
3. OVERALL DIMENSIONS
Fig. 1A, 1B, 1C are the overall dimensions to accommodate varying installation situations:
xgenus ac/dc 9
Fig. 2, 3A, 3B, 3C e 4 show the typical dimensions of the radiographic system:
10 xgenus ac/dc
STANDARD ARM 82.5cm (32.5”) Fig. 3A
Fig. 3B Fig. 3C
SHORT ARM 41cm (16.2”) LONG ARM 110cm (43.5”)
ONE INCH = 2.54cm
xgenus ac/dc 11
ONE INCH = 2.54 cm Fig. 4
12 xgenus ac/dc
4. INSTALLATION SPECIFICATIONS
WARNING
Prior to installing the radiographic system the Office Owner must be sure that the operatory and
its’ electrical services comply with all governing regulations and required installation specifications.
Otherwise one must bring them all to compliance.
1. OPERATORY REQUIREMENTS
a. The operatory must be of suitable width to accommodate the sizes and overall
dimensions provided (refer to section 3 “OVERALL DIMENSIONS”).
Check that no obstacles are present while positioning the radiographic system.
b. The operatory must not be pressurized or constitute an explosion hazard.
c. While operating, the ambient temperature must be within +5°C and +40°C
d. The storage temperature must be within -15°C and +50°C
e. The relative humidity must be within 25% and 75%
a. The supporting wall for the radiographic system must be able to withstand 200Kg sheer
stress at every bolt.
b. The subject wall material and construction must be checked. If required, consult a
contractor to determine strength and stability.
c. If there is any doubt as to the strength or stability of the wall to support the x-ray,
reinforce by providing either a steel or double plywood plate. Refer to the Pre-installation
drawings.
a. The electric system must comply with all applicable regulatory bodies.
b. The electric system must be able to supply the power-voltage and current required by
the Manufacturer's rating plate of the radiographic system as per Chart 1
115V 115V
Input voltage of the radiographic systems
-5% +10% ±15%
CHART 1
∗
PLEASE NOTE
the communication cables (C11, C12 – C21, C22) between the timer and the tube head must be two-pole, twisted
and shielded with ≥0,25mm2 (ex.: Belden 9501 type).
∗
xgenus® dc only
xgenus ac/dc 13
4. REQUIREMENTS OF THE ELECTRIC LINE
PLEASE NOTE
For longer lines, the conductor gauge must be increased in proportion.
115V 115V
Line voltage
-5% +10% ±15%
MAXIMUM LINE VOLTAGE DROP 1.2% 3%
MINIMUM LINE APPARENT RESISTANCE 0.2Ω 0.2Ω
5. ELECTRICAL CONNECTIONS
WARNING
Prior to installing the radiographic system, it is advisable that the electrical supply complies with
the recommended specifications in this manual.
CAUTION
According to relevant and practical standards, the x-ray control must be installed in a
position allowing the operator to see and have full control of the radiographic
exposure.
On the x-ray control installation wall, suitable runs for the following electrical cables must be
provided, according to the electrical installation diagram (refer to section 18
“ATTACHMENTS”):
To the wall plate location on the installation wall, a suitable run for the cable connecting the
x-ray control and the radiographic unit must be provided
(refer to section 5.5 “CONNECTION TO THE POWER SUPPLY TERMINAL BOARD”)
14 xgenus ac/dc
5. INSTALLATION
CAUTION
The “xgenus ac/dc” radiographic systems must be installed by professionally trained
technicians, who must be able to certify their work.
CAUTION
Prior to proceeding with the installation, check that all requirements are met.
5.1 UNPACKING
The components of the “xgenus ac/dc” radiographic systems are delivered by the factory securely
packaged within a specifically designed box, as shown in the following sketch (Fig. 5):
PLEASE NOTE
Prior to installation, please check all components for shipping damage. Advise the carrier.
PLEASE NOTE
The carton materials and the polystyrene foam can be completely recycled or disposed of by
authorized recycling companies.
PLEASE NOTE
It is advisable to store the original packaging to return the goods for repairs if required.
xgenus ac/dc 15
5.2 ASSEMBLING THE WALL PLATE
1. Take the wall plate out of the packaging (refer to Fig. 5) Use the drilling template provided
2. Position the drilling template on the radiographic system installation wall, at the required
height (130 cm from the base is the suggested height)
3. Fix the template to the wall with adhesive tape
4. Check the holes for vertical and horizontal alignment with the floor, using a plumb line
5. Mark the plate fixing holes
6. Mark the holes for the electric cables connecting the x-ray control to the radiographic unit
PLEASE NOTE
To prevent any flaking of the white paint coat and to control the center distances between the
holes, it is advisable to start drilling with a tip Ø7, increasing this measure gradually.
PLEASE NOTE
If the wall is not perfectly leveled, put a suitable shim between the wall and the plate, so as to
prevent any possible warping of the mounting plate.
16 xgenus ac/dc
Fig. 6
xgenus ac/dc 17
5.3 ASSEMBLING THE BRACKET/HORIZONTAL ARM
PLEASE NOTE
The brackets come in the following lengths: 41cm (16.2”) - 82.5cm (32.5”) - 110cm (43.5”)
The 82.5 cm and 110 cm brackets are provided with a stop key (Fig. 7A and 7B) to prevent the
electric cable from twisting.
PLEASE NOTE
Generally, the stop key is installed so that the equipment position at rest is on the right side of a
possible viewer when standing in front of the wall plate (Fig. 7A).
Should the position at rest be on the left side, the stop key must be rotated by 180° (Fig. 7B).
Fig. 7A Fig. 7B
WALL
18 xgenus ac/dc
After fixing the plate to the wall, assemble the bracket according to the following procedure (Fig. 8):
Fig. 8
PLEASE NOTE
Prevent any foreign material (drywall, dust, cement, etc. ) from settling on the pin seat.
The pin must slide freely in its seat. If required, thoroughly clean and lubricate with grease
“Molikote D".
PLEASE NOTE
Carefully check the bracket is parallel to the floor by using a water level.
xgenus ac/dc 19
5.4 ASSEMBLING THE POSITIONER ARM
1. Take out the positioner arm from the packaging (refer to Fig. 5)
2. Remove the arm and cap by unscrewing the fixing screw
3. Slide the arm guard slat
4. Insert the arm cable through the washer and then the positioner pin
Fig. 9
5. If required, clean the pin and the washer and lubricate with grease “Molikote D".
6. Insert the electric cable into the bracket housing
7. Assemble the guard slat
Insert the cable into the arm and push it until it reaches the pin outlet near the supply
8.
terminal board
20 xgenus ac/dc
5.5 CONNECTION TO THE TERMINAL BOARD
CAUTION
For electrical safety, it is essential that the ground conductors be correctly connected.
WARNING
The electric cable length is suitable for 82.5cm long arms.
DOT NOT tamper with the terminals arrangement when 41cm long arms are used.
The excess cable must be housed in the housing provided.
The 110cm long arms are equipped with a special extension cable.
Fig. 10
xgenus ac/dc 21
2. Proceed to the electric connection as shown in Fig. 11
PLEASE NOTE
The “xgenus® dc” unit, differently from “xgenus® ac”, is equipped with 2 communication cables.
Fig. 11
BROWN L (line)
YELLOW GREEN T (ground)
BLUE N (neutral)
BLACK
∗
COMMUNICATION
∗
RED
∗
COMMUNICATION
∗
∗
xgenus® dc only
22 xgenus ac/dc
5.6 ASSEMBLING THE RADIOGRAPHIC UNIT
ASSEMBLY INSTRUCTIONS
1. Take out the radiographic unit from the packaging (refer to Fig. 5)
2. Check that all the rating data match the power supply voltage
3. Remove both guards from the positioner arm by loosening the relevant screws (Fig. 12)
Fig. 12
4. Using a screw driver tip, separate the front coupling device (Fig. 13A)
5. Remove both guards (Fig. 13B)
xgenus ac/dc 23
6. Insert the arm unit pin into the x-ray head (Fig. 14A) and insert the supporting retainer “C” clip
(Fig. 14B)
7. Check that during insertion, the pin of the rotation prevention device correctly fits the seat located
on the x-ray head (Fig. 15)
Fig. 15
24 xgenus ac/dc
8. Couple the arm and x-ray head connectors and make sure that they are seated (Fig. 16)
Fig. 16
xgenus ac/dc 25
5.7 BALANCING THE POSITIONING ARM
CAUTION
The arm must be adjusted only when the radiographic unit is fully assembled and the x-
ray head is mounted.
WARNING
To prevent damage to the internal mechanism while performing adjustments and the balancing
tests, the adjustment wrench/key must not be left in place.
WARNING
The wrench/key provided must remain with the unit for future adjustments.
PLEASE NOTE
To reach the adjustment screw X the arm “A” must be put in vertical position.
To reach the adjustment screw Y the arm “B” must be put in horizontal position.
The adjustment key provided can be inserted only under the above conditions (Fig. 17)
arm “B”
Fig. 17
arm “A”
Fig. 17
26 xgenus ac/dc
Procedure to be followed to correctly balance the positioner arm (refer to Fig. 17):
PLEASE NOTE
The positioner type arm is supplied with arm “A” already factory adjusted.
The Arm “B” is supplied unloaded (no tension) for safety reasons.
if the head and arm do not hold the position you have selected:
PLEASE NOTE
Repeat the test and adjustment until the arm ”A” is steady and stable in all positions,
even when the arm “B” is completely extended.
If the head and arm do not hold the position you have selected:
xgenus ac/dc 27
PLEASE NOTE
Repeat the test and adjustment until the arm “A” is steady and stable in all positions,
even when the arm “B” is completely extended.
a. Insert the movable finish cover (A) between the arm guard and metal frame (Fig. 18)
Fig. 18
b. Insert the pins of one guard into the relevant seats; then position it and check that the
finish piece is coupled to the guards (Fig. 19)
Fig.19
28 xgenus ac/dc
5.8 ASSEMBLING “xgenus® ac/dc” X-RAY CONTROL
CAUTION
Check that the cable runs are arranged in the x-ray control installation wall; check the
compliance of the power supply with the installation specifications
(refer to § 4 “INSTALLATION SPECIFICATIONS”).
CAUTION
Check that the rating data match the power supply voltage.
Fig. 20
xgenus ac/dc 29
7. Remove the 26-pole connector from its seat to release both control guards (Fig. 21)
CONNECTOR 26 pole
Fig. 21
8. Insert the electric feed cables into the hole (Fig. 22)
INLET HOLE
FOR THE RADIOGRAPHIC UNIT(S)
AND WARNING LAMP(S) CABLE
Fig. 22
9. Insert the connection cables coming from the radiographic units into the rectangular slot
10. Insert the cables of the external signal lamps (IF INSTALLED) (Fig. 22)
11. Insert the cables of the remote control button (IF INSTALLED) (Fig. 22)
12. Attach the x-ray control base to the wall, matching the three anchor screws with the holes
13. Tighten the screws (refer to Fig. 20 ) with the relevant washers (refer to Fig. 20 ) into
the screw anchors
PLEASE NOTE
If the wall is uneven, use appropriate shims to prevent the x-ray control from warping.
PLEASE NOTE
Keep the x-ray control clean. Protect the board from dust, dirt or wire particles.
30 xgenus ac/dc
5.9 X-RAY CONTROL ELECTRICAL CONNECTION
CAUTION
Before proceeding to connections, the power supply must be turned off.
CAUTION
If the equipment is mounted on metal wall or studs, these must be connected to the
grounding circuit of the system.
WARNING
While carrying out electrical connection procedures, always respect the polarity PHASE – NEUTRAL.
WARNING
While stripping the cable, pay attention to the small copper wires that may fall on the printed circuit
board and cause short circuits or malfunctions.
6. Connect the cables coming from the operatory unit# 2 to the terminals XRAY2
∗
6a. Connect the communication cables from the tube head 2 to the terminals C21 and C22
∗
xgenus® dc only
xgenus ac/dc 31
1 2 3A-3B 4A-4B 5 6 7
Fig. 23
WARNING
Check the value of the fuses F1-F2 (see schemes on § 19) which must be in compliance with the type
of tube head connected to the XRAY1 and XRAY2 terminal boards:
∗
xgenus dc only
32 xgenus ac/dc
6. X-RAY CONTROL PANEL
The control panel is the user's interface, where the operator can:
xgenus ac/dc 33
control panel
LED INDICATOR
DISPLAY
X-RAY EXPOSURE
BUTTON
RADIOGRAPHIC
DISTANCE INDICATOR
(LONG CONE
RECOMMENDED-
MANDATORY OF DC
OPERATION)
KEY TO DECREASE
EXPOSURE TIME KEY TO INCREASE
EXPOSURE TIME
TUBEHEAD TYPE
TUBEHEAD
INDICATOR
SELECTION
SELECTION OF PATIENT’S
RADIOGRAPHIC SIZE
KV/VOLTAGE SELECTOR
MEMORY
MANDIBULARY
LOWER TEETH OCCLUSAL EXAM
X-RAY EMISSION
SIGNAL
BITE-WING EXAM
DIGITAL RADIOGRAPHIC
DUTY CICLE PAUSE TECHNIQUE (HOLD FOR 3
INDICATOR SEC.-ALWAYS USE LONG
CONE TECHNIQUE)
MALFUNCTIONING
INDICATOR
34 xgenus ac/dc
7. SYSTEM CONFIGURATION
The “xgenus ac/dc” radiographic systems is factory configured for the following operative “standard
mode” configuration
DIP
PARAMETER ON OFF
SWITCH
1 RADIOGRAPHIC UNIT N° 1 INSTALLED NOT INSTALLED
2 TYPE OF RADIOGRAPHIC UNIT N° 1 DC AC
3 RADIOGRAPHIC UNIT N° 2 INSTALLED NOT INSTALLED
4 TYPE OF RADIOGRAPHIC UNIT N° 2 DC AC
5 CONTROL BUTTON N° 2 INSTALLED NOT INSTALLED
LONG (12”) SHORT (8”)
6 CONE
INSTALLED INSTALLED
7 NOT AVAILABLE NOT AVAILABLE
8 NOT AVAILABLE NOT AVAILABLE
∗
xgenus® dc only
xgenus ac/dc 35
IF THE LONG 12” (31cm) CONE IS USED
ON OFF
1 ■ radiographic unit n° 1
∗
2 ■ ■ AC (=off) / DC (=on)
3 ■ radiographic unit n° 2
∗
4 ■ ■ AC (=off) / DC (=on)
5 ■ control button n° 2
6 ■ cone 12” (31cm)
7 ■ not available
8 ■ not available
ON OFF
1 ■ radiographic unit n° 1
∗
2 ■ ■ AC (=off) / DC (=on)
3 ■ radiographic unit n° 2
4 ■∗ ■ AC (=off) / DC (=on)
5 ■ control button n° 2
6 ■ cone 8” (20cm)
7 ■ not available
8 ■ not available
∗
xgenus® dc only
36 xgenus ac/dc
The configuration may be changed if:
PLEASE NOTE
“xgenus® dc” tube head allows the following combinations for the radiographic voltage and current:
∗
xgenus® dc only
xgenus ac/dc 37
8. CHANGING THE CONFIGURATION
CAUTION
This operation must be carried out by a qualified installer/technician.
WARNING
To install the changes you have made, turn off the x-ray control and then turn it on again.
To change the configuration, the dip-switch positions in the x-ray control must be changed:
DIP
PARAMETER ON OFF
SWITCH
1 RADIOGRAPHIC UNIT N° 1 INSTALLED NOT INSTALLED
2 TYPE OF RADIOGRAPHIC UNIT N° 1 DC AC
3 RADIOGRAPHIC UNIT N° 2 INSTALLED NOT INSTALLED
4 TYPE OF RADIOGRAPHIC UNIT N° 2 DC AC
5 CONTROL BUTTON N° 2 INSTALLED NOT INSTALLED
LONG (12”) SHORT (8”)
6 CONE
INSTALLED INSTALLED
7 NOT AVAILABLE NOT AVAILABLE
8 NOT AVAILABLE NOT AVAILABLE
a. If the radiographic unit is connected to the RX1 terminal board, put the dip-switch n° 1
to the ON position, otherwise to the OFF position.
b. If the radiographic unit is connected to the RX2 terminal board, put the dip-switch n° 3
to the ON position, otherwise to the OFF position.
a. if the tube head connected to the XRAY1 is “ac” type, move the dip-switch n° 2 to the
∗
OFF position; otherwise to the ON position.
b. if the tube head connected to the XRAY2 is “ac” type, move the dip-switch n° 4 to the
∗
OFF position; otherwise to the ON position.
To replace the 12” cone with the 8” cone, move the dip-switch n° 6 to the ON position.
Check that the led “8” lights up on the control panel.
PLEASE NOTE
After modification, the set exposure times are automatically changed.
∗
xgenus® dc only
38 xgenus ac/dc
To replace the 8” cone with the 12” cone, move the dip-switch n° 6 to the ON position.
Check that the led “12” lights up on the control panel.
PLEASE NOTE
After modification, the set exposure times are automatically changed.
To change the number of control buttons, move the dip-switch n° 5 to the ON position.
PLEASE NOTE
After modification, each radiographic unit is controlled by the relevant button.
PLEASE NOTE
It’s possible to connect to the timer:
n° 2 “xgenus® ac”
n° 2 “xgenus® dc”
n° 1 “xgenus® ac” + n°1 “xgenus® dc”
xgenus ac/dc 39
9. START UP
CAUTION
For further information, see the USER'S MANUAL.
b. Set the “MAIN SWITCH” located on the upper part of the control to the
position “I” (ON).
ATTENTION
If an error is detected when the system is turned on, the anomaly is indicated as
follows:
In this case, turn off the timer and then turn it back on again.
If the error should repeat itself, call for “Technical Assistance”.
PLEASE NOTE
The exposure time and radiographic parameters which appear on the display are the last that
were set before the x-ray control was turned off.
PLEASE NOTE
If installed outside the operatory, the RX signaling lamp corresponding to the selected tube head
turns on.
PLEASE NOTE
If the x-ray control remains unused and inactive for a few minutes, it switches to the stand-by
mode. Press any key of the CONTROL PANEL to restore it to the operative mode.
The previous settings will be retained.
40 xgenus ac/dc
10. CHECKING THE INSTALLATION
CAUTION
When assembly and all connections are completed, the installer must check the electrical
safety and function of the system for compliance.
OPERATING INSTRUCTIONS:
Check on the control panel that all LEDs corresponding to the required configuration are lit;
otherwise, change them.
(Refer to section. 7 “SYSTEM CONFIGURATION” and section.8 “CHANGING THE
CONFIGURATION”.)
a. Verify the correct operation of the control panel by selecting different exposure times
b. Verify the time on the display
c. When changing the selected radiographic unit, check that the corresponding external
RX signal lamps (if installed) turn on.
d. Press the “X-RAY KEY” and keep it pressed until the acoustic signal (beep)
stops and the yellow “X-RAY OUTPUT SIGNAL” Led turns off.
PLEASE NOTE
If the “EXPOSURE BUTTON” is released early, the exposure is immediately
interrupted and the E12 error message appears on the display.
e.
At the end of the exposure, verify the operation of the Led “PAUSE INDICATOR”
xgenus ac/dc 41
4. Checking the operation of the radiographic unit: CAUTION! Radiation will be
emitted
Carry out several exposures on the installed radiographic units installed and check for the
following:
5. Checking the power drawn by the radiographic system: CAUTION! Radiation will
be emitted.
To check the power drawn by the radiographic system, a meter must be used, with the
Ammeter in “AC”
PLEASE NOTE
The radiographic system complies with the requirements when:
To check the electrical system, a meter must be used, with the Voltmeter in “AC” mode.
PLEASE NOTE
The electric system complies with the requirements when the result of the formula is
as follows:
∗ ∗∗
(V0 – V1 ) / V0 ≤ 0.03 or 0.012
∗
xgenus® dc only
42 xgenus ac/dc
11. CHECKING THE EXPOSURE FACTORS
The radiographic high voltage is measured using a calibrated "non invasive" instrument with an initial
∗∗
delay set of 50ms . The kVp value is factory checked with a PMX-I D.
xgenus® ac xgenus® dc
Line Voltage Vn -5% / +10% Vn ±15%
Max Voltage drop 1.2% 3%
SET Nominal High Voltage 70 kV 60-70 kV
TECHNICAL Nominal current 8mA 4mA – 8 mA
FACTORS
Set exposure time 3.2 s 3.2 s
The radiographic VOLTAGE is: 70kVp ±15% for “xgenus® ac”
60kVp – 70kVp ±10% for “xgenus® dc”
3. Dose (mGy)
The dose in air is measured with a "non invasive" instrument, by positioning the detector at a spot-
skin distance = 310mm (12”). The mGy are factory checked with the instrument SOLIDOSE.
xgenus® ac xgenus® dc
Line Voltage Vn -5% / +10% Vn ±15%
Max Voltage drop 1.2% 3%
Nominal High Voltage 70 kV 60-70 kV
Nominal current 8mA 4mA – 8 mA
SET Set exposure time 1s 1s
TECHNICAL ∗∗
Dose in air 3.3 mGy/s ±30% (with 12” cone)
FACTORS
6.5 mGy/s ±30% (with 8” cone)
∗
Dose in air (60kVp – 4mA)=2.2mGy/s±30% (with 12” cone)
(70kVp – 4mA)=3mGy/s±30% (with 12” cone)
(60kVp – 8mA)=4.5mGy/s±30% (with 12” cone)
(70kVp – 8mA)=5.5mGy/s±30% (with 12” cone)
∗∗
xgenus® ac only
∗∗
xgenus® ac only
∗
xgenus® dc only equipped with a 12” cone. See §13 of the “USER’S MANUAL” for the dose in air with a 8” cone.
xgenus ac/dc 43
4. Exposure time (s)
∗∗
PLEASE NOTE
The “xgenus” x-ray control features the "self-compensating" technology.
The automatic correction or compensation of the exposure time offsets the effects of the
line voltage fluctuations on x-ray film quality.
Radiographic over/under exposure is eliminated.
The automatic correction system works as follows:
• A voltmeter located inside the timer constantly reads the mains voltage
while the operator selects the required exposure time.
• The last voltage measured (before the exposure) is sent to the
microprocessor. The selected exposure time is corrected to voltage by
applying a special algorithm.
• At the end of the exposure, during the pause time, the time-corrected value
is shown on the display of the x-ray control.
THE PAUSE TIME IS APPROX. 32 TIMES THE SET TIME.
The correct time is the actual duration of the exposure
The accuracy of the “actual time” is factory checked with instrument PMX-I.
xgenus® ac xgenus® dc
Line Voltage Vn -5% / +10% Vn ±15%
Max Voltage drop 1.2% 3%
SET Nominal High Voltage 70 kV 60-70 kV
TECHNICAL Nominal current 8mA 4mA – 8 mA
FACTORS
Set exposure time 0.2 s 1.6 s
®
The EXPOSURE TIME measured is: 0.2s ±10% for “xgenus ac”
1.6s ±10% for “xgenus® dc”
∗∗
5. Checking the exposure time (continued on p. 45: “Operating instructions”)
WARNING
While performing periodic quality checks, some differences between the set exposure time and
the measured exposure time may be detected.
The differences are due to the automatic time compensation that intervenes when the power
voltage differs from the nominal one.
The x-ray control is provided with two types of accuracy measurements for the exposure time:
A. Set exposure (is the exposure time set by the operator on the x-ray control).
The max error is +100% / - 100% when the voltage fluctuation is -5% / +10%.
B. After exposure: it’s the exposure duration (automatically adjusted). It’s calculated by the
microprocessor according to the line voltage, and it’s shown on the display.
∗∗
xgenus® ac only
44 xgenus ac/dc
OPERATING INSTRUCTIONS:
CAUTION
Radiations will be emitted.
To check the exposure times (sec) it is advisable to use calibrated “non invasive” instruments;
i.e.:
NERO by Victoreen
PMX by RTI
TRIAD QA by Keithley
a. Put the “X-ray detector” at the cone outlet; center the primary beam.
b. Turn on the x-ray control
c. Set one exposure time on the control.
d. Release the exposure.
e. At the end of the exposure, when the functional pause period starts, read the actual time of
exposure on the display.
f. To check the accuracy of the exposure time, use the chart below.
3° subtract 1° from 2°
PLEASE NOTE
The exposure time complies with the requirements when:
xgenus ac/dc 45
12. DIAGNOSTIC
The “xgenus ac/dc” radiographic systems allows the installer to set or display some functional
parameters of the x-ray control.
WARNING
Both in digital and in conventional mode, the minimum exposure time MUST NOT BE LESS than
∗∗
0,08s
BITE-WING
LINE VOLTAGE 105 Volt
POSTERIOR
∗∗
xgenus® ac only
46 xgenus ac/dc
∗
13. TUBE HEAD CALIBRATION∗
CAUTION
During this operation radiations will be emitted.
It is mandatory to carry out all the safety measures relevant to radioprotection.
INSTRUCTIONS
pressed until the acoustic signal (beeping) stops and the yellow led turns off.
6. Once the exposure has been taken, if the display does not show errors, the calibration has been
successfully done.
∗
xgenus® dc only
xgenus ac/dc 47
14. ERROR MESSAGES
The following chart gives a list of error messages that may appear while the “xgenus ac/dc”
radiographic systems is working.
The chart also includes the reason for the error messages and what to do to solve them.
ERROR
CAUSE SOLUTION
MESSAGES
RX1 TUBEHEAD
CALL TECHNICAL SUPPORT
IS NOT CONNECTED OR IS OUT OF ORDER
RX2 TUBEHEAD CALL TECHNICAL SUPPORT
IS NOT CONNECTED OR IS OUT OF ORDER
CORRUPTED EEPROM DATA CALL TECHNICAL SUPPORT
∗
xgenus® dc only
48 xgenus ac/dc
15. REPLACEMENT OF FUSES
The x-ray control electronic equipment is protected by 4 fuses located on the electronic P.C.B.
To replace them, proceed as follows (Fig. 25):
1. Turn off the power supply
2. Remove the front cover from the x-ray control by unscrewing the fixing screws
Fig. 26 Fig. 27
8. Replace the x-ray control cover.
9. Restore the power supply.
‡
n° 2 spare fuses are supplied in a plastic bag inside the XGT control.
xgenus ac/dc 49
16. MAINTENANCE
To guarantee the safety of the radiographic system, the following checks must be carried out:
CAUTION
The owner is responsible for arranging the maintenance schedule.
The radiographic system must undergo maintenance to check correct operation every
12 months.
50 xgenus ac/dc
17. CLEANING THE OUTER SURFACES
Use a soft cloth dampened with water and mild soap to clean the outer surfaces of the x-ray.
DO NOT permit water or any liquid to penetrate the machine or severe electrical component damage
may occur. Damage of this nature is not covered under warranty.
For protection against cross-infection, it is recommended that barrier protection be used and replaced
for each patient.
Germicides must be selected carefully, realizing that many are not suitable for use with such plastics
that are used in manufacturing. Damage of this nature is not covered under warranty.
TG Group, Inc.
3505 Laird Road, Unit # 7
Mississauga, Ontario L5L 5Y7
Canada
Tel. 1-877-557-4888
Fax 1-905-832-0732
Manufacturer:
de Götzen S.r.l.
Via Roma, 45
21057 OLGIATE OLONA VA - ITALY
E-mail: degotzen@degotzen.com
xgenus ac/dc 51
19. ATTACHMENTS
The manufacturer undertakes to supply, upon request, drawings, circuit diagrams, component parts
lists, instructions or other information needed by qualified technical personnel to perform repairs on
those parts of the “xgenus ac/dc” radiographic systems which may be repaired.
52 xgenus ac/dc
xgenus ac/dc 53
A B C
54 xgenus ac/dc
REF SPARE PARTS TUBEHEAD ‘xgenus® dc’ CODE Q.TY PCS
1 DC XG (1) FORK PLASTIC COVER 29700979 8
2 AC/DC XG(2) RING NUT SET FOR THE 29700980 5
FORK
3 AC/DC XG (3) FORK 29700777 3
4 AC/DC XG (4) ROTATION LIMITING 29700750 1
DEVICE
5 DC XG (5) WASHER 27500959 1
6 DC XG (6) AIR CARD 29700981 9
7 DC XG (7) AIR CARD COVER 29700982 11
8 DC XG (8) HEAD CABLE 39700385 1
xgenus ac/dc 55
REF SPARE PARTS ‘xgenus® ECB’ CODE Q.TY PCS
(OPTIONAL)
1 XG ECB (1) SUP & INF SHELL 29700978 4
2 MAIN BOARD WITH SWITCH AND LED 39700303 1
3 XG ECB (3) KEYLOCK SWITCH 35100298 1
56 xgenus ac/dc
REF SPARE PARTS xgenus® LIGHT CODE Q.TY PCS
(OPTIONAL)
1 XG LIGHT (1) SUP & INF SHELL 29700977 4
2 XG LIGHT (2) LAMP HOLDER 33400372 1
3 XG LIGHT (3) LAMP 230V 15W 33300352 1
3 XG LIGHT (3) LAMP 115V 15W 33300390 1
4 XG LIGHT (4) 230V BOARD 39200354 3
4 XG LIGHT (4) 115V BOARD 39200353 3
xgenus ac/dc 57
Ref SPARE PARTS TIMER ‘xgenus®’ ac/dc CODE Q.TY PCS
1 AC/DC XG TIMER (1) TEMPLATE 24200883 1
2 AC/DC XG TIMER (2) CONTROL BOARD 39200306 7
2 AC/DC XG TIMER (2) CONTROL BOARD US 39200307 7
3 AC/DC XG TIMER (3) XM COILED CABLE 32000325 1
4 AC/DC XG TIMER (4) XM DOUBLE COILED 32000359 1
CABLE
5 AC/DC XG TIMER (5) EXPOSURE BUTTON 29700974 4
HOUSING
6 AC/DC XG TIMER (6) EXPOSURE BUTTON 39700303 1
BOARD
7 AC/DC XG TIMER (7) LED BOARD 39200313 7
8 AC/DC XG TIMER (8) DEG KEYBOARD 39200308 1
8 AC/DC XG TIMER (8) SAT KEYBOARD 39200309 1
9 AC/DC XG TIMER (9) REAR COVER 29700975 1
10 AC/DC XG TIMER (10) DEG FRONT COVER 29700304 4
10 AC/DC XG TIMER (10) SAT FRONT COVER 29700976 4
11 AC/DC XG TIMER (11) KEYSWITCH LOCK 39700305 2
12 AC/DC XG TIMER (12) FILTER 36200382 3
13 GREEN BIPOLAR SWITCH 35000044 1
14 BIPOLAR SWITCH PROTECTION 33500228 1
58 xgenus ac/dc
REF SPARE PARTS POSITINER ARM CODE QTY PCS
‘xgenus®’ ac/dc
1 AC XG B (1) CABLE + 3 POLES CONNECTOR 39700384 1
2 DC XG B (2) CABLE + 5 POLES CONNECTOR 39700381 1
3 AC/DC XG B (3) HEAD SIDE SHELLS 29700990 9
4 POSITIONER ARM CONNECTING ROD ANT 29700460 4
5 POSITIONER ARM TIE ROD HOLDER 29700458 6
6 AC/DC XG B (6) CAPS (set) 29700991 7
7 AC/DC XG B (7) EXT COVER 27500757 1
8 AC/DC XG B (8) INT COVER 27500758 1
9 AC/DC XG B (9) CENTRAL SHELLS 29700993 8
10 POSITIONER ARM LOADING WRENCH 22000543 1
11 AC/DC XG B (11) INT COVER 27500758 1
12 AC/DC B (12) EXT COVER 27500757 1
13 POSITIONER ARM CONNECTING ROD POST 29700992 7
14 AC/DC XG B SAFETY KEY 22000176 1
15 AC/DC B (15) BRACKET SIDE SHELLS 29700994 8
xgenus ac/dc 59
REF SPARE PARTS WALL PLATE 350 CODE Q.TY PCS
1 WALL BOX COVER (1) DEG 29700985 5
1 AC/ XG WALL BOX (1) COVER SAT 29700986 5
2 AC/DC XG B & WALL PLATE SAFETY KEY 22000176 1
3 WALL PLATE (3) TEMPLATE 24200984 1
4 WALL PLATE (4) INSERT + CAPS 29701036 3
60 xgenus ac/dc
REF SPARE PARTS WALL PLATE 350 TOP Q.TY PCS
CODE
1 WALL PLATE (1) COVER DEG 29700985 5
1 AC/ XG WALL PLATE (1) COVER SAT 29700986 5
2 AC/DC XG B & WALL PLATE SAFETY KEY 22000176 1
3 WALL PLATE (3) TEMPLATE TOP MOUNT 24201002 1
4 WALL PLATE (4) INSERT + CAPS 29701036 3
5 WALL PLATE (5) WASHER TOP MOUNT 22000597 1
xgenus ac/dc 61
Ref ARM SPARE PARTS CODE Q.TY PCS
BRACKET 400 29700380 COMPLETO
BRACKET 800 29700378 COMPLETO
BRACKET 1100 29700379 COMPLETO
BRACKET 1100 DC XG 29701019 CPMPLETO
62 xgenus ac/dc
Dichiarazione di Conformità
Declaration of Conformity Statement - Déclaration de Conformité
Konformitäts-Erklärung - Déclaration de Conformidad de la
Impresa
Company
Società
Firma
Empresa
modello
type of machine
modèle de machine xgenus®
bauart der maschine
modelo de la máquina
dispositivi medici
medical devices
dispositives medicaux 93/42/CEE
medizinische geräte
dispositivos medicos
organismo notificato
notified body
organisme notizie 0434
verantwortliche gestaltung
organismo responsible
certificato n°
certificate No.
certificat n° 2002 – OSL – MDD – 0044
zertifikat Nr.
certificato n°
bassa tensione
low voltage
basse tension 73/23/CEE
niederspannung
baja tension
Il dispositivo definito di classe IIb è conforme ai requisiti dell’allegato II della direttiva 93/42/CEE
This product, definited as a class IIb device, is conformed to the requirements of annex II of the council directive 93/42/ECC
Le produit ci-mentionné, defini comme dispositif de classe IIb, est conforme aux requises de l’annex II de la directive 93/42/EEC
Dieses Gerät, bestimmt als Klasse IIb, ist den richtilinien 93/42/EWG, annex II gemäß
El producto aqui mencionado esta definido como dispositivo de clase IIb y esta conformado a los requisitos de l’annex II de la directiva 93/42/CEE
data di emissione
issuing date
date de èmission 18.03.2002
ausstelldatum
fecha de emisio
xgenus ac/dc 63
Dichiarazione di Conformità
Declaration of Conformity Statement - Déclaration de Conformité
Konformitäts-Erklärung - Déclaration de Conformidad de la
Impresa
Company
Società
Firma
Empresa
modello
type of machine
modèle de machine xgenus® dc
bauart der maschine
modelo de la máquina
dispositivi medici
medical devices
dispositives medicaux 93/42/CEE
medizinische geräte
dispositivos medicos
organismo notificato
notified body
organisme notizie 0434
verantwortliche gestaltung
organismo responsible
certificato n°
certificate No.
certificat n° 2002 – OSL – MDD – 0044
zertifikat Nr.
certificato n°
bassa tensione
low voltage
basse tension 73/23/CEE
Niederspannung
baja tension
Il dispositivo definito di classe IIb è conforme ai requisiti dell’allegato II della direttiva 93/42/CEE
This product, definited as a class IIb device, is conformed to the requirements of annex II of the council directive 93/42/ECC
Le produit ci-mentionné, defini comme dispositif de classe IIb, est conforme aux requises de l’annex II de la directive 93/42/EEC
Dieses Gerät, bestimmt als Klasse IIb, ist den richtilinien 93/42/EWG, annex II gemäß
El producto aqui mencionado esta definido como dispositivo de clase IIb y esta conformado a los requisitos de l’annex II de la directiva 93/42/CEE
data di emissione
issuing date
date de èmission 25.06.2002
ausstelldatum
fecha de emisio
64 xgenus ac/dc
NOTES
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………
……………………………………………………………………………………………………………………………………………………