Академический Документы
Профессиональный Документы
Культура Документы
24 September 2009
Manufacturers putting medical devices on the subparagraph above leading to systematic
market in the European Union (EU) must have recall of devices of the same type by the
a vigilance system in place for collecting and manufacturer6
evaluating reported incidents and taking correc- Information from package inserts for approved morphological evaluations.9 If the
tive action if necessary to prevent the recurrence drugs1are
Table another
shows thesource of information.
sections of the directives that ularly a novel ocular drug—has no
of such incidents. The main purpose of a medical refer to A better approach,
vigilance however, is to develop
reporting. ously studied, ocular route studies
device vigilance system is to improve protection Incident
a robust drugreporting
development applies
plan to Class
based on I,pub-
IIa, IIb (typically nonrodents) are preferred
of patients’ and other users’ health and safety orlicly
III available
medical information
devices, activeand implantable
consultation with devices authorities. One species may be su
by reducing the likelihood that a device incident and in vitro diagnostics.
experts in this area of specialty. The next step is to proper justification, especially in c
will recur.1 At a higher level, the vigilance system meetInwith
addition, any incidents
the regulatory reported
authorities early tobygar-
medi- biologic homology in other species
is an integral part of a postmarket surveillance cal practitioners or medical institutions to the
ner buy-in for the initial drug development plan. is best to gain FDA buy-in at the p
(PMS) system. Competent Authorities should be reported to
The medical device directives establish Formanufacturer
the ophthalmic products, particular importance
or the authorized representa- Ocular pharmacokinetic studi
two principal mechanisms for providing is placed
tive on Based
as well. efficientupon
preclinical development
an assessment by the one species, are needed to examine
feedback and information about marketed and the pre-IND
Competent and subsequent
Authority together with meetings with
the manufac- distribution, metabolism and excre
medical devices.2 The first is PMS, which is the FDA.orDrugs
turer for ophthalmic
the authorized use in the US
representative, are
Member in the various ocular compartment
systematic review and analysis of postproduction States are obliged
regulated by FDA’stoDivision
inform of both the European
Anti-Infective and of study generally needed is ocular
information including clinical follow-ups, quality Commission and otherThe
Ophthalmic Products. Member
advisoryStates about the
committee that of the drug, including irritancy. Th
complaints, customer feedback, vigilance and incidents and any
provides expert corrective
opinions actions
for New Drugtaken.
Application are customarily conducted in rabb
new scientific information and literature. This approval of these products is the Dermatologic and toxicokinetic studies are needed on
information should be used to facilitate process Postmarket Surveillance
Ophthalmic Drugs Advisory Committee. tissues are evaluated. Ocular toxico
and product improvements to meet customers’
expectations. The second mechanism is related Nonclinical
The medical devicedevelopment
directives for novel ocular
collectively require are generally not needed. Systemic
to the vigilance system and includes the drugs is device
medical not wellmanufacturers
defined. Nonclinical safety assess-
to implement ies via another route of administra
requirement to notify the appropriate Competent systematic
ment of theprocedures to review
drug is crucial device
to protect experi-
subjects intravenous or oral are required in
Authorities about any incidents occurring in ence in the postproduction
participating in clinical trials, hence Itanisoverall
phase. 7
important
risk vitro genotoxicity tests are also req
their territory/country. This function must be toassessment
distinguish forbetween
humans is vigilance
required.reporting
9 and
It is crucial chromosomal mutations and clasto
performed by a person with the education and PMS. Annex II.3.1 of both the Medical Device
that a thorough assessment be conducted to ensure While a pre-IND meeting an
training needed to distinguish reportable from Directive (MDD 93/42/EEC8) and the Active
that the proper species are selected for testing and towards the end of Phase 2 develo
nonreportable incidents. Implantable Medical Device Directive (AIMDD
that the study
90/385/EEC 9 design is robust, including adequate
) and Annex III.5 of the In Vitro standard in drug development, m
communication with the agency i
Legal Basis to ensure that the development pr
The provisions for setting up vigilance and PMS in line with FDA guidelines. It is
systems are outlined in the following directives: idea to have the Division of Anti-
• Directive 93/42/EEC of 14 June 1993 Ophthalmic Products review the p
concerning medical devices3 (Article 10), pivotal studies prior to obtaining
as amended
and starting clinical trials. This w
• Directive 90/385/EEC of 20 June 1990
all FDA requirements are addresse
relating to active implantable medical
devices4 (Article 8), as amended ple, in glaucoma treatment, certai
• Directive 98/79/EC of 27 October 1998 on in must show significant treatment e
vitro diagnostic medical devices5 (Article 11) cific time points. If these are not
• Directive 2007/47 of 5 September 2007 into the research design, the drug
amending Directives 93/42/EEC and timelines may be significantly imp
90/385/EEC and Directive 98/8/EC company may have to repeat the
concerning the placing of biocidal products lyze existing data.
on the market
Regulatory Focus 25
50 June 2009
Table 1. Provisions for Vigilance Reporting for Medical Devices2
Diagnostic Device Directive (IVDD 98/79/EC10) revisited and updated to account for new infor-
clearly state that when a manufacturer submits mation and risk perspectives after product release.
an application for assessment of its quality sys-
tem with a Notified Body, the application must Vigilance Reporting
include a description of the PMS system. The
purpose of the PMS system is to implement and Vigilance/incident reporting, part of the larger
maintain a systematic procedure to review expe- PMS system for medical device manufacturers,
rience gained from devices in the postproduction is the responsibility of manufacturers, national
phase and apply any necessary corrective Competent Authorities and the users.
action.11 Manufacturers or their authorized repre-
Typical sources of information on device sentatives must notify the relevant Competent
experience include: Authority about incidents and file safety cor-
• customer surveys rective actions when the reporting criteria are
• user surveys met.15 In addition, manufacturers are responsible
• adverse incidents (vigilance system) for investigating incidents and taking corrective
• user reports action if necessary.
• user feedback National Competent Authorities must
• customer complaints appoint vigilance contact points and monitor the
• customer requirements, contract informa- incident investigations carried out by manufac-
tion and market needs turers. Competent Authorities must disseminate
• patient follow-up after clinical trials or any information necessary to prevent further inci-
investigations dents and make all parties, including end users,
• service and evaluation reports aware that their cooperation is vital in providing
• scientific papers in peer-reviewed journals the first link in the vigilance chain.16
• reports on similar products by competitors Users should report incidents to the
• compliance-related communications from manufacturer or national Competent Authority,
regulatory agencies depending upon national practice.
• changes to relevant standards and The person responsible for vigilance report-
regulations12 ing (manufacturer or authorized representative)
must submit an initial incident report to the
The PMS system is also part of the standard Competent Authority followed by a final report.
EN ISO 13485,13 which indicates manufacturers The initial report of an incident should contain as
should establish documented procedures for a much relevant detail as is immediately available,
feedback system to provide early warning of but should not be delayed for the sake of gath-
quality problems and input to the corrective and ering additional information. The timelines for
preventive action processes. These processes are, reporting incidents are given below; final reports
in turn, aimed at detecting and minimizing risk should be submitted as soon as data related to the
(ISO 14971:200714). incident become available but, at the latest, within
ISO 14971:2007 specifies a process for a 30 days of the initial report submission. Not all
manufacturer to identify the hazards associated incidents lead to corrective action; however, in
with medical devices, including in vitro diag- case of doubt about the reportability of an inci-
nostic (IVD) medical devices, to estimate and dent, there should be a predisposition to report.17
evaluate the hazards’ associated risks, to control
these risks and to monitor the effectiveness of Requirements for an Incident to be Reportable
the controls. The requirements of ISO 14971:2007 An event has occurred.
are applicable to all stages of a medical device’s The list below provides some examples of
lifecycle. typical incidents.
Typically, manufacturers perform reason- • malfunction or deterioration in device char-
ably well-documented initial risk assessments. acteristics or performance
Compliance issues mostly occur due to inad-
equate follow-up measures; old files are rarely
Regulatory Focus 27
• false positive or false negative test
result falling outside the declared test
performance
• unanticipated adverse reaction or side effect
• device degradation/destruction
• inaccuracy in labeling, instructions for use
and/or promotional materials
28 September 2009
BIO1003_RAPSAD_081309.indd 1 8/13/2009 1:35:09 PM
• serious public health threat: immediately
(without any delay that could not be justi-
fied) but not later than two calendar days
following the date of awareness of the event
• death or unanticipated serious deterioration
in state of health: immediately (without any
delay that could not be justified) but not
later than 10 elapsed calendar days follow-
ing the date of awareness of the event
• others: immediately (without any delay that
could not be justified) but not later than 30
elapsed calendar days following the date of
awareness of the event
Regulatory Focus 29
be reported to the national Competent Authority 6. Ibid 3.
in the country where the incident occurred. This 7. Ibid 2.
may pose difficulties where reporting in the local 8. Ibid 4.
9. Ibid 3.
language is foreseen or where in-person dialogue
10. Ibid 5.
with the local authority is necessary. 11. Ibid 4.
Notified Bodies play a key role in assessing 12. Ibid 2.
the quality and effectiveness of a manufacturer’s 13. International Organization for Standardization.
vigilance system. Many vigilance systems are EN ISO 13485:2003 Medical devices–Quality
not foolproof and leave room for improvement. management systems – Requirements for
Notified Bodies together with the Competent regulatory purposes.
Authorities should improve their assessment and 14. International Organization for Standardization. ISO
SO 14971:2007 Medical devices—Application of
monitoring systems to capture critical system
risk management to medical devices
shortcomings and defects. Only with stricter 15. Ibid 1.
monitoring and penalties will manufacturers be 16. Ibid 1.
motivated to invest in and improve their vigi- 17. Ibid 1.
lance and PMS systems. 18. Ibid 1.
19. Ibid 1.
References
1. MEDDEV 2.12/1: Guidelines on a Medical Devices Author
Vigilance System (rev 5) (April 2007) Salma Michor, PhD, RAC, is CEO and principal consultant
2. Fundamentals of EU Regulatory Affairs, 4th edition, at Michor Consulting e.U., her recently established regula-
2008, Regulatory Affairs Professionals Society, pp tory consulting business specializing in regulatory affairs and
compliance, Quality Assurance, risk management and interim
1-15.
management. Previously, she worked in the medical device
3. Council Directive 93/42/EEC of 14 June 1993 con- and pharmaceutical industries for companies including Croma
cerning medical devices Pharma, Wyeth and Chiesi. She is an independent expert for
4. Council Directive 90/385/EEC of 20 June 1990 on the European Commission and teaches regulatory affairs and
the approximation of the laws of the Member States clinical strategies at the postgraduate level. Michor is a mem-
relating to active implantable medical devices ber of the Board of Editors for Regulatory Focus and can be
5. Directive 98/79/EC of the European Parliament reached at michor_salma@hotmail.com.
and of the Council of 27 October 1998 on in vitro
diagnostic medical devices
30 September 2009