07 calibration temperature and after 30 min carefully dilute to volume,
AOAC Official Method 942.06 with aid of capillary tube, by adding H2O previously boiled and Alcohol by Volume cooled to same temperature. Determine specific gravity of distillate in Distilled Liquors as in 945.06C (see 26.1.06). Obtain corresponding % alcohol by Pycnometer Method volume from 913.02 (see Appendix C). (This result is % alcohol by Final Action volume at 15.56 [60F].) (b) Samples containing more than 60% alcohol by volume.— Codex-adopted-AOAC Method* Proceed as in (a) with following changes: Calibrate 100 mL and 50 mL pycnometers, Figure 942.06, at 15.56, fill 50 mL pycnome- A. Apparatus ter with sample, and adjust to volume at 15.56. Add 50 mL cold Distillation apparatus.—500 mL flask, connected through bulb H2O to distilling flask before transfer of sample and collect to vertically assembled Liebig condenser with jacket 400 mm long, distillate in 100 mL pycnometer. Adjust to volume at 15.56. inner tube 9 1 mm id, with adapter. Joints may be live rubber or Obtain specific gravity of distillate, and from table, 913.02 (see standard taper. Heat with electric- or gas-operated unit. Appendix C), obtain % alcohol by volume in distillate. B. Determination C. Calculation (Some samples do not require distillation prior to determination, e.g., straight bourbon whisky; alcohol-H2O mixtures containing traces of volatile ingredients.) % alcohol by volume in sample at 15.56 = D W/W