07 calibration temperature and after 30 min carefully dilute to volume,


AOAC Official Method 942.06 with aid of capillary tube, by adding H2O previously boiled and
Alcohol by Volume cooled to same temperature. Determine specific gravity of distillate
in Distilled Liquors as in 945.06C (see 26.1.06). Obtain corresponding % alcohol by
Pycnometer Method volume from 913.02 (see Appendix C). (This result is % alcohol by
Final Action volume at 15.56 [60F].)
(b) Samples containing more than 60% alcohol by volume.—
Codex-adopted-AOAC Method*
Proceed as in (a) with following changes: Calibrate 100 mL and
50 mL pycnometers, Figure 942.06, at 15.56, fill 50 mL pycnome-
A. Apparatus ter with sample, and adjust to volume at 15.56. Add 50 mL cold
Distillation apparatus.—500 mL flask, connected through bulb H2O to distilling flask before transfer of sample and collect
to vertically assembled Liebig condenser with jacket 400 mm long, distillate in 100 mL pycnometer. Adjust to volume at 15.56.
inner tube 9 1 mm id, with adapter. Joints may be live rubber or Obtain specific gravity of distillate, and from table, 913.02 (see
standard taper. Heat with electric- or gas-operated unit. Appendix C), obtain % alcohol by volume in distillate.
B. Determination
C. Calculation
(Some samples do not require distillation prior to determination,
e.g., straight bourbon whisky; alcohol-H2O mixtures containing
traces of volatile ingredients.) % alcohol by volume in sample at 15.56 = D W/W

(a) Samples containing 60% or less alcohol by volume.—Cali-


brate 100 mL pycnometer, Figure 942.06, as in 945.06B (see where D = % alcohol by volume in distillate at 15.56; W =
26.1.06), at one of temperatures specified in 913.02 (see Appendix weight H 2O at 15.56 in 100 mL pycnometer; and W = weight
C). Fill clean, dry pycnometer with sample and adjust to volume at H 2O at 15.56 in 50 mL pycnometer.
calibration temperature as in 945.06B (see 26.1.06).
Transfer contents of pycnometer to distilling flask, just previously References: Ind. Eng. Chem. Anal. Ed. 14, 237(1942). JAOAC
rinsed with cold H2O and containing few glass beads, or equivalent. 28, 88(1945); 41, 118(1958); 42, 329(1959); 53,
Rinse pycnometer 3 times, using total of 25 mL cold H2O (40 mL 338(1970) .
for cordials or wines), and add rinse H2O to flask. Place wet
pycnometer so that adapter extends just into bulb. Surround CAS-64-17-5 (alcohol)
pycnometer with ice or ice-H2O. Complete connections and pass
through condenser rapid stream of H2O kept at 25 at outlet. Distil ca © AOAC INTERNATIONAL
96 mL at uniform rate in 30 but 60 min, using longer times for higher
percentages of alcohol. Remove and stopper pycnometer, mix dis-
tillate by swirling, and wash down with H2O any drops that may be *Adopted as a Codex Defining Method for pycnometry of residual al-
above graduation mark. Immerse in constant temperature bath at cohol in vinegar

Figure 942.09—100 mL and 50 mL pycnometers