5.1 Overview
Chapter 5 describes the internal structure and operation mechanism of ACCUVIX V10.
This chapter must be read for the product maintenance and upgrade.
ACCUVIX V10 is Software DSC-applied ultrasound video analyzer.
It not only adopted 17 inch LCD monitor and provides high resolution ultrasound
video, but also provides the premium grade system functions. To improve the
processing speed, MEDISON Co., Ltd. developed new interface to connect a latest
PC and the ultrasound system with its proprietary technology. The enhancement of
processor speed makes the system operations faster and reduces diagnosis time.
ACCUVIX V10 can use up to 192 Element probes and adopted Digital Beamforming
of TX 128 Channels. Ultrasound image is displayed on the LCD through the Front
End Part and Back End Part (including PC Part).
The resolution of the LCD monitor is 1280 X 1024 pixels and various image formats
are provided. The wide view angle of the LCD panel provides convenient work
environment for diagnosis. In addition, the arm-type monitor controller enables users
to control the monitor easily.
The DVD RW drive and USB port are placed on the front panel of the system for
easier image backup and software service. Since this system supports various
external storage devices such as USB MO, USB Flash Memory and external-type
USB HDD, upgrade becomes more easier.
ACCUVIX V10 consists of the following major components.
5.3.1 Overview
ACCUVIX V10 consists of Ultrasound System Part, PC Part, User Interface Part and
Power Part. However, it consists of Front End Part, Back End Part, User Interface
Part, and Power Part from the electronical view point.
The following is the description of electronical structure of ACCUVIX V10.
Front End Part refers to the CW (Continuous Wave) Board, PSA (Probe Select
Assembly), and BF (Beamformer) Board of the Ultrasound System Part. The Front
End Part delivers High Voltage Pulser to the probe so that ultrasound is generated,
amplifies the returned echo signal and processes Digital Beamforming. The RF
signal generated here is delivered to the Back End Part.
Back End Part refers to the BE(Back End) Board and PC of the Ultrasound System
Part. The RF signal generated in the BF(Beamformer) Board is processed to
diagnosis image such as BW, Color Doppler, PW Doppler, CW Doppler, and Power
Doppler and displayed on the monitor so that users can see it. In addition, new
technologies such as SCI, DMR and AIO are applied to provide various diagnosis.
User Interface Part refers to the LCD monitor and control panel.
Power Part consists of AC-DC Power Module and DC-DC Power Module. AC-DC
Power Module converts AC to DC voltage and supplies power to the DC-DC Power
Module and the PC Power of the PC Part. The DC-DC Power Module supplies
voltage to the boards of the Ultrasound System Part and the PC Power supplies
voltage to the PC Part.
5.3.3 PC Part
PC Part consists of Scan Converter and video output circuit so that the ultrasound
information generated by the Ultrasound System Part is displayed on the monitor. In
addition, it provides interface with the control panel.
The ultrasound information output from the Ultrasound System Part is connected to
the PC Part using the DMA mode of the PCI interface. The ultrasound image is
implemented by software DSC and VGA.
While conventional ultrasound system used Hardware DSC, ACCUVIX V10 used
software DSC and displayed ultrasound image on the LCD monitor.
PC Power Supply is Micro ATX and supplies power to the PC Part.
The PC Part consists of the following components.
y PCI(Peripheral Component Interconnect)
y DVI(Digital Video Interface)
y VGA(Video Graphics Array)
y PC Mother Board
y Rear Panel
y PC Power Supply
It removes unnecessary signal from the Echo returned through the PSA (Probe
Select Assembly)’s High Voltage Switch. Up to 180 Vpp Tx Pulses and a few mV
5.6.4.7 CW Sensing
It controls the TX voltage and current of CW(Continuous Wave).
1) CW TX Voltage Sense
If the voltage of the TX Pulser used by CW(Continuous Wave) fails to satisfy the
specification, the system stops with an error message.
2) CW TX Current Sense
If the current of the TX Pulser used by CW(Continuous Wave) fails to satisfy the
specification, the system stops with an error message.
5.8.3 Specification
y Interface between Ultrasound System Part and PC Part
y PCI Clock Generation
y 100 Pins Fast PCI Cable Connector
y DVI Cable Connector
5.10.2 Picture
Item Contents
Product Name nVIDIA GeForce 7600GS
Interface PCI-Express X 16
Memory Type DDR2
Memory Capacity 256MB
Memory Bus 126bit
Core Clock 400MHz
Memory Clock 800(400)MHz
RAMDAC 400MHz
Cooling System Cooling Fan & Heat Sink
Video Out TV-OUT, D-SUB, DVI
5.11.2 Picture
Item Contents
FSB 1066/800/533Mhz
IEEE1394 Supported
PCI Slot 3 PCI, 3 PCI Express x1, 1 PCI Express x16
Serial ATA 4 Ch
USB USB 2.0
Main Chip set Intel i965P
Sound SigmaTel STAC9227* 6Ch
Socket LGA775
Ethernet Intel 82566DC Gigabit Ethernet
Memory Dual Channel DDR2 800/667/533
Max Memory 8GB
Board Type ATX
AC Power
Connector
AC Voltage Out
Fuse
2) Output Voltage
AC/DC Output Voltage Use Range
110 ~ 120 V
External Device
AC 200 ~ 240 V
200 ~ 240 V PC Power
DC 24 V Ultrasound System Part
If current flows in the Over-Current Range exceeding the standard input and max
output current, the OLP is activated and all powers of DDM are cut off.